Philips HF 365 Service Manual

Philips Domestic Appliances and Personal Care
Electronic clinical thermometer
HF 365
PRODUCT INFORMATION
Features :
Memory Last measured temperature appears
on the display after pressing the on/off button for at least 2 seconds.
Waterproof The thermometer may be immersed
Battery low indication On display
Automatic switch - off After 30 minutes
The thermometer meets the Medical Device Directive 93/42/EEC for medical appliances
The buzzer sounds 3 times and the ºC symbol stops blinking.
in disinfectant containing water for maximally one hour. The water temperature must be lower than 60 ºC.
Technical Data
Power consumption Less than 60 micro-Watts, which
means more than 3000 measurements with one battery.
Service
Display LCD display with 4.8 mm character
Measuring range 32.00 ºC - 42.00 ºC
Measuring time Approximately 60 seconds
Measuring system Peak hold system
Measurement ready Measurement is ready when the
Accuracy 0.1 ºC between 35.5 ºC and 42.0 ºC
Battery Button-cell 1.5 V type LR41
Materials Sensor-tip: Stainless steel
For this thermometer no spare parts will be available in the after sales period. When the product is defective, it has to be replaced.
height
temperature does not change during 20 seconds.
0.2 ºC between 32.0 ºC and 35.5 ºC
Housing parts and protective box:ABS
Display informations p.t.o.
Replacing the battery p.t.o.
Cleaning instructions p.t.o.
PublishedbyPhilipsDomesticAppliancesandPersonalCare PrintedintheNetherlands
4822 729 22062 98/11
PCS 99 803
©
Copyrightreserved Subjecttomodification
DISPLAY INFORMATIONS
1
Error message after “ measurement complete signal .“
Blinking dashes and ºC symbol, indicating the temperature measured is lower than 32 ºC
2
3
4
Blinking ºC symbol the temperature measured is higher than 42 ºC
The battery must be replaced
Replacing battery
When the symbol shown in fig. 3 appears on the display, the battery (type LR41) must be replaced (see fig. 4). Hand in the empty battery at a chemical waste collection point.
5
Lysetol V
¥
Aldasan 2000
¥
Lysofobminn 2000
¥
Gigasept
¥
Lysethol FF
¥
Gigasept FF
¥
Sekusept Forte
¥
Kohrdorin
¥
Alcohol
¥
Isopropyl Alcohol
¥
Cleaning
Clean the thermometer with water that is not hotter than 60 ºC or with medicinal alcohol. For hospitals : Immerse in water containing the prescribed disinfectant (fig. 5) for 1 hour at the most.
PCS 99 803