S IDE 1
S IDE 2
M anual Operation
l
Preparation
Insert Data Card (optional).
Turn Energy Select knob to .
Using Paddles
Apply conductive matter.
Apply paddles to patient's chest, using
the anterior-apex placement.
Using Pads
Apply pads as directed by packaging
instructions.
Connect the pads to the pads patient
cable.
l
Defibrillation
Select Energy.
1
Move the Energy Select knob to
the desired energy level.
Charge.
Press or yellow charge
button on apex paddle.
CHARGE
Shock.
Press . If using external
paddles or switched internal
paddles, simultaneously press
shock buttons on paddles.
To disarm, press .
l
Pacing
Select .
Apply electrodes and pads.
Press .
- adjust Rate.
- press Start.
- adjust Output if needed.
for AE D Operation
Manual On
Pacer
SHOCK
Manual On
DISARM
E CG M onitoring
1. Apply monitoring electrodes.
2. Select lead (optional).
Press to choose desired lead.
LEAD
SELECT
3. Set HR Alarm (optional).
Press to select alarm limits.
To disable alarm, press .
HR ALARM
HR ALARM
Electrode Placement: 3 - Lead
RA/R
ll
Electrode Placement: 5 - Lead
RA/R
RL/N
Limb Leads
LA/L
V1
LL/F
LA/L
l
lll
LL/F
1
V-LeadOptions
2
3
6
5
4
*M4735-97800*
*C*
M4735-97800 Rev C
If Patient is:
wUnresponsive
wNot Breathing
wPulseless
Connect Pads Patient Cable
Apply Pads
Insert Data Card (optional)
Turn Energy Select to
AED On
If Instructed, Press
If Shock Advised
Press
if shock series is set to greater than one
WARNING
Electric shock hazards exist internally. Do not remove assembly
screws. Refer servicing to qualified personnel.
Charge batteries between use or if the message LOW BATTERY
is displayed. Perform a Battery Capacity Test every 3 months.
The defibrillator will take longer to charge when used with only
AC power or a depleted battery. Use only with the M3516A
Battery Pack.
for Manual O pe ration, P acing & Monitoring
SHOCK
AE D Operation
ANALYZE
If No Shock Advised
Check Patient
If needed, press
and begin CPR
M4735A
PAUSE
if rhythm monitoring active