This document and the information contained in it is proprietary and
confidential information of Philips Medical Systems (”Philips”) and
may not be reproduced, copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written
permission of the Philips Legal Department. Use of this document
and the information contained in it is strictly reserved for current
Philips personnel and Philips customers who have a current and
valid license from Philips for use by the customer’s designated in-house service employee on equipment located at the customer’s
designated site. Use of this document by unauthorized persons is
strictly prohibited. Report violation of these requirements to the Philips Legal Department. This document must be returned to Philips
when the user is no longer licensed and in any event upon Philips’
first written request.
Warranty Disclaimer Language:
Philips provides this DOCUMENT without warranty of any kind, implied or expressed, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
Limitations of Liability Language:
Philips has taken care to ensure the accuracy of this document.
However, Philips assumes no liability for errors or omissions and reserves the right to make changes without further notice to any products herein to improve reliability, function, or design. Philips may
make improvements or changes in the product(s) or program(s) described in this document at any time.
This documentation is valid for BuckyDiagnost TH2 / TF.
The table BuckyDiagnost TH2 / TF is part of the BuckyDiagnost family and can be installed together with the ceiling
suspension BuckyDiagnost CS, respectively the floor stand BuckyDiagnost FS and the X-ray generator OPTIMUS
RAD / OPTIMUS 30 single phase.
The table BuckyDiagnost TH2 is characterized by:
-- height-adjustable tablebase
-- floating tabletop
The table BuckyDiagnost TF characterized by:
-- fixed height tablebase
-- floating tabletop
2.Items supplied
BuckyDiagnost TH2
-- Table with floating tabletop, 750 mm x 2400 mm, Kevlar
-- Height sensor
-- Manual bucky unit, non-sensing
-- Grid for bucky unit
-- AMPLIMAT chamber
Options:
-- Manual bucky unit with format sensing
-- Automatic cassette loader (ACL4)
-- Tomo 2 device
-- Footswitch locking device
-- Second table control
-- Wider table top850 mm x 2400 mm
-- Shorter table top750 mm x 2000 mm
BuckyDiagnost TF
-- Table with floating tabletop, 750 mm x 2400 mm, Kevlar
-- Height sensor for BuckyDiagnost TH9848 500 2260x
-- Grid 36/8 FO 1009860 834 9010x
-- Grid 36/8 FO 1109896 010 2608x
-- Grid 36/8 FO 1409896 010 2609x
-- Grid 36/12 FO 1009860 836 9010x
-- Grid 36/12 FO 1109896 010 2600x
-- Grid 36/12 FO 1409896 010 2607x
-- Grid 36/12 FO 1809896 010 2604x
-- Grid 40/12 FO 1009860 846 9010x
-- Grid 40/12 FO 1109860 846 901 1x
-- Amplimat chamber9890 000 0161x
-- BuckyDiagnost CS9890 xxx xxxxx
-- BuckyDiagnost FS9890 xxx xxxxx
INTRODUCTION AND TECHNICAL DATA
Options:
-- Footswitch locking device9890 010 02712
-- Second table control9890 010 02731
Accessories:
-- Adjustable head clamp9804 662 33009
-- Pair of handgrips9804 662 43009
-- Ratchet compressor9804 666 22009
-- Symm. ratchet compressor9804 666 22109
-- Lateral cassette holder9804 666 32009
-- Cassette holder, hor. beam9804 668 10009
-- Pair of leg supports9804 668 50009
-- Infusion bottle holder9804 669 10009
-- Pair adj. positioners9804 669 20009
-- Parking frame for accessories9804 669 30009
Upgrades:
-- Upgrade f. bucky brake TH9890 010 02881
-- Upgrade sec. table contr. TH9890 010 02901
-- Upgrade footswitch look9890 010 02912
-- Upgrade kit tracking9890 010 80191
-- Upgrade kit high load9890 010 81251
-- Upgrade kit key switch for bucky TH4512 130 82901
CSIP Level 1 (05.0)1--2
E 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost TH2 / TF (MSB)
BD_th2_TF_MSB_1_BW
Page 8
BuckyDiagnost TH2 / TF (MSB)
INTRODUCTION AND TECHNICAL DATA
4.Technical data
4.1.Mechanical data
-- Mechanical dimensions BuckyDiagnost TH2 / TFsee 2Z--1.1 and 2Z--1.2
4.2.Environmental data
OperationStock / Transport
Temperature in °Celcius10 / + 40-- 25 / + 70
Temperature in °Celcius / HourN/AN/A
Humidity in % (non-- condensing)20 / 805/95
Gradient in % / hourN/AN/A
Vibrations / Shock range in Hz5 -- 5005 -- 500
Vibrations / Shock amplitude in mmN/AN/A
Vibrations / Shock acceleration in g0.25 peak1.0 peak
Shock acceleration in g5 peak30 peak
Shock pulse duration in msec1111
Air pressure in Hecto--pascal700 / 1100700 / 1100
Acoustic noise level:N / A
Air cooling:N / A
EMC:IEC 950
4.3.Electrical data
Equipment related:
Power required:max. 900 VA
Nominal voltages:230 VAC
Supply configurations:230 VAC
Voltage variation:N/A
Voltage impulse:N/A
Voltage surge:N/A
Voltage sag:N/A
Static frequency variation:50 or 60 Hz ± 1Hz
Dynamic frequeny variation:N/A
Harmonic voltage distortion (single, total) :N/A
Neutral to ground voltage variation:N/A
Neutral to ground volate impulse:N/A
Nominal current:6.0 A / 0.6 A (TH2 / TF)
Nominal frequency:50 Hz / 60 Hz
Delivered: N/A
To be ordered: N/A
To be arranged locally :-- TC 129, Tool kit, standard
-- Power tools (hammer drill with 12 mm carbide drill bits)
Installation
Delivered: Set of standard fixing material
Wooden table transport bars
To be ordered: N/A
To be arranged locally :-- TC 129, Tool kit, standard
-- Spirit level
-- TC 092, Notebook, service PC
-- TC 003, Service PC cable kit
-- TC 091, Multimeter
-- Service software X--Scope 1.4.2 or higher
-- Power tools (hammer drill with 12 mm carbid drill bits)
-- Measuring tape, 10 m
4.5.Network data
INTRODUCTION AND TECHNICAL DATA
N/A
4.6.Remote service data
N/A
4.7.Transport data
For information about transport data refer to the corresponding PRD.
4.8.Pre --installation requirements
-- Room layout (floor fixing)(Refer to PRD)
The preliminary planning work is restricted to the selection of a suitable room. The outfit of the room has to be
equipped with the conduits and pipes required and, in cases where the loading capacity of the bearing area
(380 kg on 2 x ≅ 500 cm
Where DIN specification are not applicable the minimum wall clearances should be determined in compliance with
the concerning national regulations.
2
) is insufficient, the floor preparation for the table.
4.9.Movements
-- Table height(see 2Z--1.1 and 2Z--1.2)
-- Tabletop(see 2Z--1.1 and 2Z--1.2)
4.10. Pre-installation data
For information about the pre--installation data refer to PRD ’Product planning’ BuckyDiagnost TH2 / TF.
5.Safety information
N/A
CSIP Level 1 (05.0)1--4
E 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost TH2 / TF (MSB)
BD_th2_TF_MSB_1_BW
Page 10
BuckyDiagnost TH2 / TF (MSB)
6.Compliance information
N/A
7.Equipment id entification an d labellin g
N/A
8.Abbreviations an d definitio n s
AbbreviationExplanation
CSCeiling Suspension
FSFloor Stand
MSBMini System Bus
PMSPhilips Medical Systems
POSTPower-On Self Test
PRDProduct Reference Data
SIDSource Image Distance
INTRODUCTION AND TECHNICAL DATA
TCTool Code
TFTable Fix
THTable Height adjustable
VEVertical
VRVertical Radiography
VTVertical Tiltable
9.Manual history
DateVe rsionNameReason of changes
01.02.2005 V1 REV AAM. BierstedtIntroduction of TH2 (MSB)
Refer to PRD room layout or local planning data and use
the drilling template from theSystem Manual
Installation.
S Mark the four fixing points at the floor.
S Drill ∅12 mm holes in the middle of the marked
area about 100 mm deep.
S Insert the four anchor dowels in the holes.
S Fix the anchor dowels.
S Lift the table and place it in its working position.
3.Height adjustment for installation (TH2)
INSTALLATION
S Connect cable MEX from table to the mains of
220 VAC to move the table upwards into a better
working position.
Also connect the protection lead (green/yellow).
S Press foot switchto move the table upwards.
S Press foot switchto move the table
downwards.
Warning!
If the table does not move downwards,
press on the tabletop with 20kp. Do not
adjust the collision switch! The table
automatically stops at the middle
position.
S Disconnect MEX from mains.
4.Alignment for BuckyDiagnost TH2 /TF
S Level the table in both directions by using a spirit
level or by using the laser alignment tool, use the
spacer, if necessary (1).
4x
S Insert the fixing block (2, 4x), the special washer
5.Change position of the AMPLIMAT chamber from left to right side
A manipulation of the connector at the AMPLIMAT
chamber is necessary, if the position of the
BuckyDiagnost TH2 / TF must be changed from left
(1) to right (2) side.
S Remove the tabletop by removing the four
screws (1).
S Mark the center line of the tabletop to the frame
(both sides).
S Remove the tabletop.
1
2
Left hand
version
Right hand
version
1
Left hand
version
1.5mm
S Dismantle the anti--scatter grid by unscrewing the
four metal holders (2).
S Remove the anti--scatter grid.
S Dismantle the AMPLIMAT chamber by loosening
the two grub screws (3).
S The wire assignment on the connector PH X1 must
be interchanged, the violett wire on X1:1 with the
black wire on X1:3.
S Place the measuring chamber at the new
position (4).
S Tighten the measuring chamber with the two grub
screws.
S Re--install the anti--scatter grid (5).
S Tighten the anti--scatter grid.
PH X1
2
4
3
5
6
Right hand
version
S Re--install the table cover at the new position (6).
2.Power supply of BuckyDiagnost TH2 / TF with ACL4 and tomo
Mains power of 220 VAC for
table via F1, L1 to X50 ...X56
table lift motor SNV
PCB power supply SN1 for
table control SZ2
bucky carriage, magnetic brake VAC
bucky controller SZ1
table tomo drive & control SAN1
ACL4
PCB power supply for SN2 for
automatic cassette loader ACL4
Power supply
230 VAC
F1/L1
X56
table lift motor
X50
X51
X52
X53
X54
X55
S
230 VAC
for
230 VAC
230 VAC
SZ2
X6
X14:2X14:1
GND
SN1
26 VDC
GND(--)
26VDC(+)
VAC
GND26 VDC
X6:2X6:1
SZ1
SAN1
SN2
X1
X2:2X2:1
GND
(--)
24 VDC
(+)
X3:2X3:1
ACL4
X5
X1MX2
X100
M
SNV
X43
X42
X44
X41
X40
2.1.Location of electrical components
ACL4
SAX10
VA
4
SN1
SX50 ... SX56
Rear side
SF1
SL1
SN2
SX40
...
SX45
SZ1
SAX100
SAN1
SX60
Front side
SZ2
CSIP Level 1 (05.0)3--2
E 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost TH2 / TF (MSB)
BD_th2_TF_MSB_3_BW
Page 24
FAULTFINDING
3.Power-ON sequence and stand-by
The BuckyDiagnost system runs through a selftest after
power--ON.
S Switch ON the system.
3.1.BuckyDiagnost TH2/TF selftest
S Raise the table to its uppermost position.
Press foot switchto move the table
upwards. The table stops automatically
(only TH2).
S Unscrew the service cover at the rear side, open it
for inspection.
BuckyDiagnost TH2 / TF (MSB)
S Press reset switch (1) to re-start the bucky
controller.
The red LED (2) (= bucky controller is not ready) is
illuminated until the selftest is ended successful.
The yellow LED (3) (= CAN bus is working) flashes after
successful selftest.
The green LED (4) (= bucky controller is ready) is
illuminated after successful selftest.
-- If the board SZ1 fails, replace it. Configure the
new board via program X-Scope.
During the initialization phase the bucky carriage moves
to its tomo center position. Afterwards the bucky
carriage moves to one side then to the other side and
stops in the middle position.
If the tomo unit selftest fails:
S Check brake release switch (1) and magnetic
brake (2) on the bucky carriage.
The brake release switch is currently closed and is
mechanically opened via the grip (3).
brake
release
SACS
BuckyDiagnost TH2 / TF (MSB)FAULTFINDING
1
2
3
X8
carriage
brake
SA
VAC
SA X100
X10
SAN1
CSIP Level 1 (05.0)3--4
E 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost TH2 / TF (MSB)
BD_th2_TF_MSB_3_BW
Page 26
FAULTFINDING
4.Table movement of BuckyDiagnost TH2
4.1.Table switches
S Check the function of the footswitch lock SAE.
Pressed switch = footswitches without function.
S Move the tabletop into the working position via
footswitches SC S6, S5 or via the optional second
table control SCH.
BuckyDiagnost TH2 / TF (MSB)
CH
AE
If the tabletop does not move
upwards
S Check closed 900 mm switch SNVS1 at
SZ2 X5:6/7 (24VDC) via 750 mm switch
SNVS3 at SZ2 X5:3/4.
S Check footswitch S5 for upwards movement at
SZ2 X4.
S Check optional handswitch SCH for upwards
movement at
SZ2 X9.
If the tabletop does not move
downwards
S Check closed safety switch SS10 at
SZ2 X7:2/1 (24 VDC) via closed 500 mm switch
SNVS2 at SZ2 X5:8/9 via 750 mm switch
SNVS3 at SZ2 X5:3/4.
S Check footswitch S6 for downwards movement at
SZ2 X4.
S Check optional handswitch SCH for downwards
movement at SZ2 X9.
S10
S
NVS1
NVS3
Z2
C
S
S6S5
C
NVS2
NV
S3
Front side
S2 S1
CH
1
1
X9
1
1
X7
Z2
X4
X5
S10
If the tabletop does not move upwards or
S Check the relay K1 or K2 on board SZ2. One of the
two relays must be activated by pressing either the
footswitch or the optional handswitch.
The relay contact supplies the voltage of 230 VAC to the
motor SNV via SZ2 X3:1/:2/:3.
During movement the tabletop stops in the middle
positionof thetableheight viathe 750 mm
switch SNV S3 on the table lift motor.
S Seealsodrawing4Z--2.
4.2.Collision switch
S Check the force of the table movement.
-- Pull up the tabletop with a spring balance.
The safety switch SS10 must open if a force
of 160 ... 240 N is reached. Otherwise:
-- Adjust the spring tension at the tension screw
with a 13 mm spanner
clockwise= force of collision
decreases
counterclockwise = force of collision
increases
S
C
Z2
S6
X4
0VDC
1 ... 2 mm
Clearence of retainer
W1
+24 VDC
K2
K1
K3
K7
160 ... 240 N
K4X5
S10
X7
X5
S3
750 mm
S2
500 mm
S1
900 mm
NV
CSIP Level 1 (05.0)3--6
E 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost TH2 / TF (MSB)
BD_th2_TF_MSB_3_BW
Page 28
FAULTFINDING
5.Tabletop movement
5.1.Floating table
S Press the footswitch SC S3. The tabletop must be
floatable. If not, check items:
-- footswitch lock SAE
(must not be active = pressed)
-- brakes for longitudinal and transverse direction
-- footswitch for tabletop floating
S Use the second table control to check functions as
described above, if possible.
5.2.Table carriage brakes
S Check the function of the brakes.
BuckyDiagnost TH2 / TF (MSB)
AE
S
Z2
C
S3S3
Front side
S Check the connection to SZ2 X10. Voltage of +24 V
is delivered via the footswitch S3 or via the optional
second table control switch.
If the footswitches are locked via the footswitch
lock, functions of tabletop movements via the
optional second table control are always possible.
The height sensor operates as a current source.
The range of the current value is 5 mA (table is in lowest
position) up to 18 mA (table is in uppermost position).
BuckyDiagnost TH2 / TF (MSB)FAULTFINDING
S
U
The current value can be shown :
S Connect a service PC to SZ1 X20.
S Call up program VT100 for monitoring.
S Type in:↵
-- superuser↵
-- set service output_errors_to_vt100
1↵
Every action is shown on the PC screen, also the current
depending on the topical table height.
-- show analogue↵
The analogue inputs is displayed on the PC screen.
-- show service sid↵
The SID parameters are displayed on the PC screen.
S Check the height sensor SU at SZ1 X4, measure
the current.
S
U
24 VDC
1
6
A
I = 5 ... 18 mA
U
X4
Rear
side
Z1
D1
X4
Z1
24 VDC
1
6
analog
S
-- I f I = 0 m A ,
check the voltage at X4:1 or find the reason
in the cable.
-- I f I = 2 0 m A ,
check the cable or replace the height sensor.