PerkinElmer Caliper IVIS Spectrum Hardware Manual

Page 1
IVIS Spectrum
®
In Vivo and In Vitro 3D Optical Imaging
Bioluminescence and Fluorescence
Spectral Unmixing
Transmission and Reflectance Imaging
March 2012
Page 2
© 2007-2012 Caliper Corporation. All rights reserved.
121450_Rev01
Caliper LIfe Sciences
68 Elm Street Hopkinton, MA 01748 Main Phone: 1.508.435.9500 Fax: 1.508.435.3439 E-mail: tech.support@caliperls.com www.caliperls.com
Discovery in the Living Organism, IVIS Imaging System and Living Image are either registered trademarks or trademarks of Xenogen Corporation. The names of companies and products mentioned herein may be the trademarks of their respective owners. Apple, Macintosh and QuickTime are registered trademarks of Apple Computer, Inc. Microsoft, PowerPoint and Windows are either registered trademarks or trademarks of Microsoft Corporation in the United States and/or other countries. Adobe and Illustrator are either registered trademarks or trademarks of Adobe Systems Incorporated in the United States and/or other countries.
Page 3
®
Spectrum Hardware Manual
IVIS
Contents
1 Welcome . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 About the IVIS Spectrum Imaging System . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 IVIS Imaging Systems Technology Support . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2 Important Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.1 Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2 Instrument Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.3 Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.4 Laser Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.5 Fluorescence Light Source Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.6 Environmental Considerations for the System Components . . . . . . . . . . . . . . . . . . . 7
2.7 Power Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.8 Cleaning or Moving the System Components . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.9 Other Equipment, Chemicals, or Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.10 Servicing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.1 Electrical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2 Burn Hazard & Eye Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.3 Mechanical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.4 Chemical & Biological Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.5 Panels, Covers & Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4 Legal Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.1 Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.2 Patents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.3 Trademarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
5 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.1 Electrical Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.2 CCD Camera . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.3 Optics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.4 Fluorescence Imaging Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.5 Environmental Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
i
Page 4
Contents
5.6 Weights & Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
6IVIS
®
Spectrum Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6.1 CCD Camera . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.2 ThermoCube Chiller Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.3 Imaging Chamber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
6.4 Optical Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
6.5 Specimen Warming System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
6.6 Laser/Galvanometer System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.7 Crash Avoidance System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.8 Base Enclosure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.9 Acquisition Computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
6.10 Accessory Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
6.11 High Reflectance Hemisphere . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
7 Operating the IVIS
®
Spectrum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
7.1 Restarting the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
7.2 Gas Plumbing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
7.3 Door Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
7.4 Centering a Subject in the Field of View . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
7.5 Fluorescent Imaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
7.6 Imaging Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
7.7 System Shut Down Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
8 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
8.1 Measured Temperature Does Not Equal the Demand Temperature . . . . . . . . . . . . . . . 45
8.2 Photographic Image Is Unacceptable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
8.3 Luminescent Image Is Unacceptable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
8.4 No Image Is Produced . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
8.5 Hardware Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
8.6 Accessing the Area Under the Imaging Platform . . . . . . . . . . . . . . . . . . . . . . . . 50
9 Care & Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
9.1 Cleaning the IVIS Spectrum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
9.2 Cleaning the Lens Protection Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
9.3 Changing the Emission Filters Used in Fluorescence Imaging . . . . . . . . . . . . . . . . . 53
ii
Page 5
®
Spectrum Hardware Manual
IVIS
9.4 Moving the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
10 Servicing & Maintaining the ThermoCube Chiller Unit . . . . . . . . . . . . . . . . 57
10.1 Required Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
10.2 Turning Off Electric Power to the Main Console . . . . . . . . . . . . . . . . . . . . . . . 57
10.3 Removing the Console Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
10.4 Sliding the Tray Out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
10.5 Filling the ThermoCube Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
10.6 Reassembling the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
10.7 Restarting the IVIS Spectrum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Appendix A Spare Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
iii
Page 6
Contents
iv
Page 7
®
Spectrum Hardware Manual
IVIS

Welcome

1
About the IVIS Spectrum Imaging System . . . . . . . . . . . . . . . . 1
IVIS Imaging Systems Technology Support . . . . . . . . . . . . . . . 3

1.1 About the IVIS Spectrum Imaging System

The IVIS® Spectrum Imaging System is a high-sensitivity, low noise, in vivo imaging technology platform that enables noninvasive visualization and tracking of cellular and genetic activity within a living organism in real time
(Figure 1.1). The system provides both bioluminescence and fluorescence
imaging capability.
For fluorescence imaging, the instrument can operate in reflectance or transillumination mode. Filtered light from a broad-band lamp provides the excitation source in both modes. In the reflectance mode, light is delivered to four reflectors that are located on the ceiling of the imaging chamber. In the transillumination mode, the excitation light is delivered to an x-y translation assembly under the stage and focused to a 2 mm diameter beam that can be directed to a particular location on the underside of the animal subject. The system includes ten excitation and 18 emission filters that enable spectral scanning of reporters over the range from 480-850 nm.
3D Tomography The system also includes a structured light projector that enables reconstruction
of the surface topography. The 3D location and concentration of fluorescent sources can be computed from structured light and transillumination fluorescent images. The 3D location and strength of luminescent sources is computed from structured light and luminescent image data.
1
Page 8
1. Welcome
High resolution monitor
XWS-260 Workstation
CCD camera
Imaging chamber
Compartment that contains the ThermoCube chiller unit and camera power supply
Acquisition computer
XGI-8 Gas Anesthesia System
Figure 1.1 IVIS® Spectrum Imaging System
The IVIS Spectrum is an integrated imaging system that includes a:
CCD camera mounted on a light-tight imaging chamber
Camera power supply
Thermoelectric cooling unit
Fluorescence light source
Windows®-based computer system for data acquisition and analysis.
The sound-limiting enclosure beneath the imaging chamber contains the thermoelectric cooling unit, camera power supply, and other imaging chamber functions. All components except the computer are integrated into a single, moveable chassis.
This manual explains how to operate and maintain the equipment, and provides guidelines for obtaining the best bioluminescent or fluorescent images. Before using the IVIS Spectrum, please read this manual carefully to obtain safe, optimum performance and a maximum service life from the unit. For
®
instructions on using the system software, please see the Living Image Software User’s Manual.
If you have questions regarding this manual or the IVIS Imaging System, please call IVIS Imaging Systems Technology Support.
2
Page 9
®
Spectrum Hardware Manual
IVIS

1.2 IVIS Imaging Systems Technology Support

For technical support, please contact Caliper at:
Telephone
E-mail
Fax
Mail
1.877.522.2447 (US)
1.508.435.9500
tech.support@caliperls.com
1.508.435.3439
Caliper Life Sciences
US Corporate Headquarters
68 Elm Street
Hopkinton, MA 01748
3
Page 10
1. Welcome
[This page intentionally blank.]
4
Page 11
®
CAUTION
!
WARNING
!
VOLTAGE
Spectrum Hardware Manual
IVIS

Important Safety Instructions

2
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Instrument Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Laser Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Environmental Considerations for the System Components . . . . . . 7
Power Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Cleaning or Moving the System Components . . . . . . . . . . . . . . 9
Other Equipment, Chemicals, or Supplies . . . . . . . . . . . . . . . . 9
Servicing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

2.1 Definitions

When you see the caution, warning, or voltage symbol, pay particular attention to the safety information presented. Observance of safety precautions will help avoid actions that could damage or adversely affect the performance of the IVIS Spectrum. If the equipment is used in a manner not specified in this manual, the protection provided by the equipment may be impaired.
CAUTION! This symbol indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury and/or mechanical damage. It is also used to alert against unsafe practices. It reminds you that all safety instructions should be read and understood before installation, operation, maintenance, or repair of this instrument.
WARNING! Used when an action or condition may potentially cause serious personal injury or loss of life. Mechanical damage may also result.
VOLTAGE! Indicates high voltage or risk of electric shock.
5
Page 12
2. Important Safety Instructions

2.2 Instrument Labels

Pay careful attention to the labels on the instrument.
Label Description
The Caliper IVIS The instrument itself is classified as a Class 1 device because the user is safely protected from the laser beam. The potential hazard of exposure to service personnel is clearly indicated with a warning label per IEC 60825 and CDRH requirements.
Laser caution label Note: The actual warning
label mounted on the laser module at the service access inside the unit uses non-fluorescing colors and materials.
Where located on the IVIS system, the read manual label indicates that important information about the labeled area can be found in the IVIS Spectrum Hardware Manual.
Read manual label
Where located on the IVIS system, the pinch hazard label indicates the location of a potential pinch point.
®
Spectrum Imaging Instrument contains a Class 3R Laser module.
Pinch hazard label

2.3 Instructions

Read Instructions Read and understand all the safety and operating instructions before you install,
operate, or perform maintenance on this product. Make sure that you fully understand the following safety instructions, warnings, and disclaimers before proceeding to the rest of the manual.
Retain Instructions Retain the safety and operating instructions for future reference.
Follow Instructions Follow all operating and handling instructions. Failure to follow operating or
handling instructions may void any warranty covering this product.
Heed Warnings Abide by all warnings on the product and in the operating instructions. Failure
to adhere to warnings or safety precautions may void any warranty covering this product.
6
Page 13
®
WARNING
!
WARNING
!
Spectrum Hardware Manual
IVIS

2.4 Laser Safety

The IVIS Spectrum is equipped with Class 3R laser (532 nm (green) laser source, 5 mW maximum continuous (CW), which is suitably embedded and limited in power to less than 1mW where accessible in the imaging compartment behind the front door.
WARNING! A laser beam can be harmful to the eyes. ABSOLUTELY DO NOT look directly into the laser beam when the laser is turned on. Be aware that the laser light can be easily reflected off of shiny surfaces and that the reflected light can be dangerous. When the laser is on, remove all watches and jewelry before placing hands in the imaging chamber.

2.5 Fluorescence Light Source Safety

WARNING! The IVIS Spectrum is equipped with a strong bottom illumination
fluorescence light source. Do not stare directly into this source.

2.6 Environmental Considerations for the System Components

Location for the IVIS® Spectrum
Heat The system should be situated away from heat sources such as open flames,
Before the IVIS® Spectrum is installed, consider the proper environment for the components.
Install the equipment in an environment where:
The temperature does not change or fluctuate widely. Choose an
environment where the temperature is maintained between 15-25° C (59­78° F).
The humidity does not exceed 80%.
No strong electric or magnetic fields exist.
No vibrations are present.
No corrosive gases are present.
High amounts of dust are not present.
No open flame is present.
There is sufficient space behind the IVIS Spectrum equipment. A
minimum space of six inches from the flat surface of the rear panel should be provided behind the IVIS Spectrum to provide unobstructed air flow and access to the main power on/off switch.
The floor is level and structurally sound.
radiators, heat registers, stoves, and other heat-generating electrical equipment.
7
Page 14
Water & Moisture
VOLTAGE
VOLTAGE
2. Important Safety Instructions
VOLTAGE! Do not use this product near water (for example, near a sink or wet room) due to risk of electric shock, electrical damage, and/or failure.
Laboratory Space & Ventilation Requirements
The IVIS® Spectrum requires a minimum of 100 square feet (9 square meters) of laboratory space for adequate ventilation. At least of six inches of space is required behind the rear flat panel to provide unobstructed air flow.
Allocate an area at least 36 inches wide and 51 inches deep for the imaging system console to allow the door to be fully opened and provide the recommended air space behind the instrument.
An additional area that is 48 inches wide should be provided (preferably on the right hand side) for a table or cart to hold the computer and provide a work surface.

2.7 Power Considerations

Power Sources The IVIS
installation locality that was specified at the time of order. If the system is moved to another area, check to make sure that the same voltage requirements exist.
An IVIS Spectrum that operates on 120VAC requires a dedicated 20 Amp circuit that is not shared with other loads except the computer and the optional XGI-8 Gas Anesthesia System. Make sure that all equipment is plugged into a properly grounded AC supply.
The computer should share the same circuit to avoid ground loops. Never use auxiliary power outlets on other equipment to supply the IVIS Spectrum. Since the system contains internal surge protection, it must be powered directly from the main electrical supply wall socket. Use the surge protector provided by Caliper for the computer and other accessories.
An IVIS Spectrum configured for 230 VAC requires a dedicated minimum 10 Amp circuit subject to the same conditions noted above. For more details on the operating requirements, see Electrical Power Requirements, page 21.
®
Spectrum is configured for the voltage requirements of the
VOLTAGE! The IVIS Spectrum can operate at multiple voltages (100, 120, 220, 240 VAC); however, you are not permitted to change the input voltage to any of the system components. Several internal modifications are required for voltage change. If the operating voltage must be changed, contact IVIS Imaging Systems Technology Support.
Power Cord Protection Power supply cords should be routed so that they are unlikely to be walked on
or pinched by items placed upon or against them. Pay close attention to receptacles and to points of connection between cords and equipment.
Lightning & Power Line Surges
The IVIS® Spectrum is supplied with an internal surge protector that is not customer serviceable. Contact IVIS Imaging Systems Technology Support regarding equipment servicing.
8
Page 15
®
WARNING
!
VOLTAGE
Spectrum Hardware Manual
IVIS
A separate surge protector is provided for the computer and other accessory equipment.
Facilities should be adequately wired according to local building codes.
Power Outages If the IVIS Imaging System experiences a loss of supply power, turn off the
power switch for all components and do not restart the system until reliable power has been restored. For more details on how to restart the system, see
Startup Procedure, page 36.
Overloading
WARNING! Do not overload wall outlets, extension cords, or integral
convenience receptacles as this can result in a risk of fire or electric shock. See Electrical Power Requirements, page 21 for requirements.

2.8 Cleaning or Moving the System Components

Cleaning/Liquid Entry
VOLTAGE! Do not use liquid or aerosol cleaners and never spill liquid of any
kind on any of the IVIS into contact with the IVIS System may result in damage to the system or electrocution. For more details on proper care of the system, see Cleaning the
IVIS Spectrum, page 51.
Moving the IVIS® Spectrum The IVIS
®
Spectrum is a sensitive, scientific instrument and should be moved
®
Spectrum components. Sprays and liquids that come
with care. Pay particular attention when rolling the instrument on its casters to avoid toppling the equipment.
The total height of the instrument is 83 inches (211cm). If necessary, the plastic dome can be removed for easier passage through doorways. For more details on removing the dome, see page 55.
If you have any questions about moving the instrument, contact IVIS Imaging Systems Technology Support.

2.9 Other Equipment, Chemicals, or Supplies

Use of any equipment other than that recommended by this manual has not been evaluated for safety and, therefore, is the sole responsibility of the user.
No chemicals are required for the operation of the IVIS Spectrum. Other user­supplied chemicals or materials may be required as part of your specific biological testing procedures.
9
Page 16

2.10 Servicing

2. Important Safety Instructions
Refer all servicing to IVIS Imaging Systems Technology Support. If the product is damaged and requires service, unplug the product from the outlet and contact technical support. Servicing by anyone other than those authorized by Caliper voids any warranty covering this product.
10
Page 17
®
WARNING
!
CAUTION
!
Spectrum Hardware Manual
IVIS

Warnings

3

3.1 Electrical Safety

Electrical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Burn Hazard & Eye Safety . . . . . . . . . . . . . . . . . . . . . . . . 12
Mechanical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Chemical & Biological Safety . . . . . . . . . . . . . . . . . . . . . . 15
Panels, Covers & Modules . . . . . . . . . . . . . . . . . . . . . . . . 15
WARNING! DO NOT attempt to service the IVIS® Spectrum yourself. Contact IVIS Imaging Systems Technology Support for electrical service needs. Although there are no voltages in excess of 45V inside the imaging chamber, local line voltages can be present inside the electronics cabinets.
CAUTION! If cleaning is necessary, wipe the exterior surfaces of the IVIS Spectrum with a soft, damp cloth only. Do not allow fluids of any kind to enter the system interior under any circumstances. For more details on cleaning the system, see page 51.
11
Page 18
3. Warnings
CAUTION
!
Note: The actual warning label mounted on the laser module at the service access inside the unit uses non-fluorescing colors and materials

3.2 Burn Hazard & Eye Safety

Laser
The IVIS Spectrum Imaging Instrument contains an embedded Class 3R Laser module: 532 nm (visible green) laser source, 5 mW maximum continuous (CW). The instrument itself is classified as a Class 1 device because the user is safely protected from the laser beam. The potential hazard of exposure to service personnel is clearly indicated with a warning label per IEC 60825 and CDRH requirements.
CAUTION! The use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure.
NEVER remove back, side, front or internal panels of the instrument while the laser is powered on. Note that laser light can be easily reflected off of shiny surfaces. Remove all watches and jewelry before placing hands in the imaging chamber.
Complies with FDA performance standards for laser products except for deviations pursuant to Laser Notice No. 50, dated 6/24/2007.
Complies with IEC/EN 60825-1:2007.
Light Source Module The IVIS Spectrum is equipped with a fluorescence light source module and
connecting fiber optic cables that are capable of producing intense light that could be damaging to the eyes. Additionally, the system is equipped with a Class 3R laser, which could be an eye hazard.
Figure 3.1 Laser beams for alignment grid

3.3 Mechanical Safety

The IVIS Spectrum has many internal motorized components that can move only when the door is closed and locked. Because the imaging platform moves frequently during routine use, keep loose objects away from the edges of the platform where they could become jammed.
12
Page 19
®
CAUTION
!
CAUTION
!
Spectrum Hardware Manual
IVIS
Figure 3.2 IVIS Spectrum, imaging chamber door
CAUTION! DO NOT defeat any of the safety interlocks. Do not place
anything under the imaging platform.
The imaging platform cannot be moved unless the door is closed and locked. If the platform moves when the door is unlocked, shut down the system and contact IVIS Imaging Systems Technology Support.
The IVIS Spectrum contains several delicate optical and mechanical components. Do not touch these components and avoid sprays or other contaminates that may damage lenses or other optical-mechanical components.
CAUTION! The laser/galvanometer is a very delicate optical component that is visible when the filter wheel side access panel is removed adjust it.
(Figure 3.3). DO NOT touch this device or attempt to clean or
13
Page 20
3. Warnings
WARNING
!
Filter wheel panel
Figure 3.3 Filter wheel removable panel
The laser beam is directed into the imaging chamber through an opening at the top of the chamber. Never place any object, tool, or fingers into this opening, otherwise damage to the galvanometer will result.
WARNING! DO NOT look directly into this opening when the laser is operating.
14
Page 21
®
Spectrum Hardware Manual
IVIS

3.4 Chemical & Biological Safety

Normal operation may involve the use of test samples that are pathogenic, toxic, or radioactive. It is your responsibility to ensure that all necessary safety precautions are taken before such materials are used.
Dispose of all waste materials according to appropriate environmental health and safety guidelines.
It is your responsibility to decontaminate the IVIS service by IVIS Imaging Systems Technology Support. Ask your laboratory safety officer to advise you about the level of containment required for your application and about the proper decontamination or sterilization procedures to follow.
Handle all infectious samples according to good laboratory procedures and methods to prevent the spread of disease.

3.5 Panels, Covers & Modules

The imaging chamber door and the filter wheel access panel are the only user removable panels. There are no user serviceable components in the electronics cabinets of the IVIS from the system console unless instructed by and under the supervision of IVIS Imaging Systems Technology Support.
®
Spectrum before requesting
®
Spectrum. Do not remove the electronics cabinet covers
15
Page 22
3. Warnings
[This page intentionally blank.]
16
Page 23
®
Spectrum Hardware Manual
IVIS

Legal Notices

4

4.1 Limited Warranty

Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Patents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Trademarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Caliper Life Sciences, Inc. ("Caliper") provides the following limited warranty for each new IVIS
i. This Limited Warranty for the System extends to the original purchaser
("Customer") for a period of one (1) year following installation of the System, and is not assignable or transferable to any successor.
ii. During the Limited Warranty, Caliper will repair or replace, at Caliper's
sole option, any defective parts if such repair or replacement is needed because of System malfunction or failure to conform to published specifications during normal usage in accordance with the instructions in this manual. Repairs and replacements under the Limited Warranty will be made at Caliper’s expense. Caliper’s limit of liability under the Limited Warranty shall be the purchase price of the Imaging System. Caliper shall not be liable for any other losses or damages. These remedies are the Customer's exclusive remedies for breach of this Limited Warranty.
iii. No coverage or benefits shall be provided under this Limited Warranty if
any of the following conditions apply:
a. The System has been subjected to unauthorized modifications (e.g.
unauthorized installation of hardware or software), unauthorized repair or servicing, misuse, neglect, abuse, accident, alteration, any use inconsistent with or in contradiction to the instructions in this manual, or other acts which are not the fault of Caliper.
b. Caliper was not advised in writing by the Customer of the alleged
defect or malfunction of the System within the earlier to occur of ten (10) days after the expiration of the Limited Warranty period, or 15 days after becoming aware of the defect or malfunction.
c. If Customer moves the System from its installed location to another
location without Caliper's technical assistance, then any damage to the System due to such movement shall not be covered under the initial warranty or any extended warranty, and Licensee shall pay Caliper's standard service rates for repair of such damage.
iv. If a problem develops during the Limited Warranty, the Customer shall
contact Caliper technical support for assistance.
v. THE FOREGOING LIMITED WARRANTY IS THE CUSTOMER'S
SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED. CALIPER SHALL NOT BE LIABLE FOR SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES, INCLUDING BUT NOT LIMITED TO LOSS OF ANTICIPATED BENEFITS OR PROFITS, LOSS OF SAVINGS OR REVENUE, PUNITIVE DAMAGES, LOSS OF USE OF THE SYSTEM
®
Spectrum CT ("System") as follows ("Limited Warranty"):
17
Page 24
4. Legal Notices
OR ANY ASSOCIATED EQUIPMENT, COST OF CAPITAL, COST OF ANY SUBSTITUTE EQUIPMENT OR FACILITIES, DOWNTIME, THE CLAIMS OF ANY THIRD PARTIES, INCLUDING CUSTOMERS, AND INJURY TO PROPERTY, RESULTING FROM THE PURCHASE OR USE OF THE SYSTEM OR ARISING FROM BREACH OF THE WARRANTY, BREACH OF CONTRACT, NEGLIGENCE, STRICT TORT, OR ANY OTHER LEGAL OR EQUITABLE THEORY, EVEN IF CALIPER KNEW OF THE LIKELIHOOD OF SUCH DAMAGES. CALIPER SHALL NOT BE LIABLE FOR DELAY IN RENDERING SERVICE UNDER THE LIMITED WARRANTY, OR LOSS OF USE DURING THE PERIOD THAT THE SYSTEM IS BEING REPAIRED.
vi. Some countries, states or provinces do not allow the exclusion or limitation
of implied warranties or the limitation of incidental or consequential damages for certain products or the limitation of liability for personal injury, so the above limitations and exclusions may be limited in their application to you. When any implied warranties are not allowed to be excluded in their entirety, they will be limited to the duration of the applicable written warranty. This Limited Warranty gives you specific legal rights which may vary depending on local law.
vii. This Limited Warranty shall be governed by the laws of the State of
California, U.S.A., excluding its conflicts of laws principles and excluding the United Nations Convention on Contracts for the International Sale of Goods.

4.2 Patents

4.3 Trademarks

The detection and imaging of light originating within mammals is the subject of several issued patents and pending patent applications in the United States and around the world, including U.S. Patent Numbers 5,6200,135, 6,217,847, 6,649,143, 6,890,515, and European Patent Commission Number EP0861093, for which Xenogen Corporation is the exclusive licensor. The use of an IVIS
®
Imaging System for such applications requires a sublicense from Xenogen Corporation.
In addition, many of the hardware and software components of the Imaging System are the subject of various issued patents and pending patent applications owned by Xenogen, including: United States Patent Number 6,614,452 (Graphical User Interface for In Vivo Imaging) and 6,775,567 (Improved Imaging Apparatus); and United States Patent Applications 09/905668 (Multi­view Imaging Systems), 10/606976 (Method and Apparatus for 3-D Reconstruction of Light Emitting Sources), 10/151463 (Method and Apparatus for Determining Target Depth, Brightness, and Size Within a Body Region), 10/ 189886 (Fluorescence illumination assembly for an imaging apparatus), and 10/ 068573 (Light calibration device for use in low level light imaging systems).
IVIS and Living Image are registered trademarks or trademarks of Caliper Life Sciences, Inc. The names of companies and products mentioned herein may be the trademarks of their respective owners. Apple, Macintosh, and QuickTime are registered trademarks of Apple Computer, Inc. Microsoft, PowerPoint and Windows are either registered trademarks or trademarks of Microsoft
18
Page 25
®
Spectrum Hardware Manual
IVIS
Corporation in the United States and/or other countries. Adobe and Illustrator are either registered trademarks or trademarks of Adobe Systems Incorporated in the United States and/or other countries. Intel is a registered trademark of Intel Corporation.
19
Page 26
4. Legal Notices
[This page intentionally left blank.]
20
Page 27
®
Spectrum Hardware Manual
IVIS

Specifications

5
Electrical Power Requirements . . . . . . . . . . . . . . . . . . . . . 21
CCD Camera . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Optics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Fluorescence Imaging Components . . . . . . . . . . . . . . . . . . . 22
Environmental Requirements . . . . . . . . . . . . . . . . . . . . . . 22
Weights & Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . 23

5.1 Electrical Power Requirements

Electrical Power Requirements Specification
Voltage Required (main console) or Recommended Dedicated Circuit
100-120VAC 50/60Hz 15A 220-240VAC 50/60Hz 10A 20 Amps for 120VAC or 10 Amps minimum for 230VAC For more details, see Power Considerations,
page 8.

5.2 CCD Camera

CCD Camera Specification
Sensor Type Back illuminated
CCD Format 2048 x 2048 pixels
CCD Size 26 x 26 mm
Effective CCD Format 1920 x 1920 pixels
Pixel Dimensions 13.5 x 13.5 μm
Quantum Efficiency ~85% 400-700nm
>50% 350-900nm
Readout Noise <5 electrons RMS
Dark Current <100 electrons/s/cm
Minimum Detectable Luminance
CCD Temperature Nominal -90° C
<70 photons/s/sr/cm
2
(-90° C)
2
21
Page 28
5. Specifications

5.3 Optics

Optical Component Specification
Lens f/stop f/1- f/8
Field of View (FOV) 3.9, 6.5, 13, 19.5, and 26 cm
Resolution >60 μm (FOV = 3.9, f/1)
Scanning Laser <1 mW at 532 nm

5.4 Fluorescence Imaging Components

Fluorescence Imaging Component
Excitation Filters 12 positions, 25 mm diameter, 10 filters supplied
Emission Filters 22 positions, 60 mm diameter, 18 filters supplied
Background (Autofluorescence and
Leakage)
Lamp 150 W quartz halogen

5.5 Environmental Requirements

Environmental Component Specification
Temperature 15 -25° C (59-78° F)
Humidity 0-80% non-condensing
Type of Use Indoor
Sound Level 61 dB < 500 Hz
Stage Temperature Ambient to 40° C
Altitude Rating <2000 meters (6560 ft.)
Specification
1 part in 10
21 V 3250° Kelvin
7
typical
22
Page 29
®
Spectrum Hardware Manual
IVIS

5.6 Weights & Dimensions

Item Specification
Weight 275 kg (600 lbs)
Depth 77 cm (30 inches)
Width 65 cm (25.5 inches)
Height 211 cm (83 inches)
Imaging Chamber Internal Dimensions
51 x 51 x 66 cm (D x W x H)
23
Page 30
5. Specifications
[This page intentionally blank.]
24
Page 31
®
IMPORTANT
!
Spectrum Hardware Manual
IVIS

IVIS® Spectrum Components

6
CCD Camera . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
ThermoCube Chiller Unit . . . . . . . . . . . . . . . . . . . . . . . . . 26
Imaging Chamber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Optical Components . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Specimen Warming System . . . . . . . . . . . . . . . . . . . . . . . 29
Laser/Galvanometer System . . . . . . . . . . . . . . . . . . . . . . . 30
Crash Avoidance System . . . . . . . . . . . . . . . . . . . . . . . . . 30
Base Enclosure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Acquisition Computer . . . . . . . . . . . . . . . . . . . . . . . . . . 31
High Reflectance Hemisphere . . . . . . . . . . . . . . . . . . . . . . 33
Accessory Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
The IVIS® Spectrum includes the following components:
Scientific imaging charged coupled device (CCD) camera.
Imaging chamber with specimen warming system.
Thermoelectric camera cooling system.
System of lenses and filters for optimum image acquisition.
Laser/galvanometer.
Photographic illumination system.
Integrated fluorescence system.
Imaging platform crash avoidance system.
These components are integrated into one freestanding instrument that is equipped with casters
A pre-configured Windows as well as acquire and analyze images.
(Figure 6.1).
®
-based computer enables you to control the system
ALERT! If you modify the IVIS Imaging System in any way, without prior approval from Caliper, all warranties that cover this product are void. In addition, the computer included with the IVIS Spectrum is specifically configured to run all system-related applications. Any modification of existing software or hardware voids all warranties.
If you have any questions, please contact IVIS Imaging Systems Technology Support.
25
Page 32
6. IVIS® Spectrum Components
CCD camera
Scanning laser assembly
Imaging chamber
Custom lens, 11-position filter wheel assembly, laser light alignment
Gas anesthesia manifold
Compartment that contains the ThermoCube chiller unit and camera power supply
Emission filter wheel
Excitation filter wheel
Top illumination light source
Heated platform
Integrated fluorescent imaging features:
Bottom illumination light source
Figure 6.1 IVIS Spectrum components
The system is housed in a portable cart with 25 x 34” footprint.

6.1 CCD Camera

The camera is a scientific grade, thermoelectrically-cooled, back-thinned, back­illuminated, large format CCD manufactured by Spectral Instruments (Figure
6.1). The camera uses a ThermoCube water chiller. Two flexible plastic water
lines connect the CCD camera to the ThermoCube chiller unit.
The CCD camera has a heated, vacuum sealing window to prevent condensation that can obscure the image or damage other components.
The camera power supply is housed in the base of the instrument and maintains the camera operating temperature when the computer is turned off. It has a 16­bit digitized, low noise electronic readout for extremely low background images.

6.2 ThermoCube Chiller Unit

The ThermoCube unit is a thermoelectric water chiller (Figure 6.2). Water from the thermoelectrically cooled CCD camera is pumped to the ThermoCube unit where the water is cooled and returned to the camera.
26
Page 33
®
WARNING
!
Spectrum Hardware Manual
IVIS

6.3 Imaging Chamber

Figure 6.2 ThermoCube chiller unit
The imaging chamber (Figure 6.3) is a highly specialized device consisting of the imaging chamber housing, a heated, moveable platform, an auto focusing lens system with f/stop control, and synchronized filter wheels that control the spectral content of the luminescent or fluorescent images.
The LED-based illumination system and a laser-based galvanometer provide the means to acquire photographic information and structured light before bioluminescent or fluorescent imaging. The imaging chamber is equipped with gas anesthesia inlet and outlet ports for use with the optional XGI-8 Gas Anesthesia System.
The imaging chamber is light tight, so that no light penetrates from the outside after the magnetically activated door is locked. The magnetic latch ensures the door cannot be inadvertently opened during an imaging session.
The interior of the imaging chamber is constructed from materials that are non­phosphorescent and non-fluorescent to prevent internal light contamination that could compromise sample measurements.
WARNING! Under no circumstances should you attempt to make any mechanical modifications to the imaging chamber.
27
Page 34

6.4 Optical Components

NOTICE
Door locking hardware
1.5X
3.9 cm A
2.5X
6.5 cm B
5.0X 13 cm C
8.8X
22.5 cm D
Power FOV Position
6. IVIS® Spectrum Components
Figure 6.3 IVIS® Spectrum, imaging chamber
Imaging System/Lenses The IVIS
demagnification of 1.5X, 2.5X, 5X, and 8.8X power (Figure 6.4).The filter wheel settings are selected in the Living Image® software. For more details, see the Living Image Software Manual.
Figure 6.4 Example field of view obtained with different lens configurations
The automatic shutter and f/stop iris is integrated into the lens system. A window at the top of the imaging chamber protects the lenses and other optical components in the lens compartment from contamination. For more details on the care of this window, see Cleaning the Lens Protection Window, page 52.
Do not touch the glass window or permit specimens to come in contact with the window, otherwise image quality may be impaired. If the window is struck by a hard object, it may crack or shatter.
®
Spectrum is equipped with a rotating lens carousel that provides
28
Page 35
®
Temperature controlled imaging platform
Spectrum Hardware Manual
IVIS
Emission & Excitation Filter Wheels
There are two 11-position filter wheels. Filter settings are selected by software control. For more details on how to access the filter wheels and change filters, see page 53.
Figure 6.5 Filter wheel
®
For fluorescence studies, the IVIS
Spectrum Imaging System can illuminate a specimen from the top of the imaging chamber or from the bottom of the stage. A 12-position excitation filter wheel module is located in the back of the instrument (not accessible to users). If the filters must be changed or the module requires service, contact IVIS Imaging Systems Technology Support.
For more details on fluorescence imaging, see page 41.

6.5 Specimen Warming System

The temperature-controlled imaging platform keeps specimens warm during imaging. The temperature is controlled by the software, and is only enabled when the instrument is initialized from Living Image temperature upon initializing the instrument is 37° C.
The imaging platform does not have active cooling. Allowing the platform to cool from 37° C to ambient temperature may require up to 20 minutes.
Figure 6.6 Imaging chamber
®
software. The default
29
Page 36
6. IVIS® Spectrum Components
WARNING
!
IMPORTANT
!

6.6 Laser/Galvanometer System

The laser/galvanometer system comprises a Class 3R laser (532 nm (visible green) laser source, 5 mW maximum continuous (CW). The laser beam is directed into the imaging chamber by two computer-controlled mirrors.
WARNING! A laser beam can be harmful to the eyes. ABSOLUTELY DO NOT look directly into the laser beam when the laser is turned on. Be aware that laser light can be easily reflected off of shiny surfaces and that this reflected light can be dangerous. When the laser is on, remove all watches and jewelry when placing hands in the imaging chamber.
The laser/galvanometer system provides field of view alignment targets for reproducible specimen positioning and structured light for 3D contouring and auto-focusing.

6.7 Crash Avoidance System

The crash avoidance system uses a scanning laser to determine the subject height and verify that it is within acceptable limits. This system prevents injury to specimens and damage to delicate optical components such as the lens window or the illumination diffusers for field of view A only.
You are prompted at the start of each imaging session to input a sample height estimate. If the estimate is in error, the crash avoidance system protects the specimen. Alternatively, the system can be configured to automatically determine the sample height.

6.8 Base Enclosure

30
ALERT! Do not block the ventilation openings on the base or position the system too close to a wall so that the exhaust flow is obstructed.
The base is equipped with four casters that enable you to move the instrument
(Figure 6.7). For more details on how to move the IVIS Spectrum, see page 55.
Page 37
®
Ventilation openings
Ventilation openings
Spectrum Hardware Manual
IVIS
Figure 6.7 IVIS Spectrum, back view

6.9 Acquisition Computer

6.10 Accessory Cable

The system includes a Windows®-based personal computer. The Living Image® software as well as Microsoft Office software is installed on the computer. The Living Image software controls the IVIS Spectrum, and displays and analyzes the image data. A 20-inch monitor and a 10/100 Ethernet network adapter card are also included. A printer may be attached to the computer.
The IVIS Spectrum is equipped with an electrical connector that enables the transfer of signals to or from the back panel of the instrument. The back panel connector connector is located inside the imaging chamber
(Figure 6.8) is accessible at the rear of the instrument and the stage
(Figure 6.9). Table 6.1 shows
guidelines for pin usage.
Table 6.1 Connector pin usage
Pin Number Function
1 to 5 These pins are optimized for transport of ECG (EKG) signals and use
6 to 9 Use these pins to connect a pulse-oxymetry system (for example, the
10 and 11 Use these pins to connect to an Infusion pump that you mount inside
shielded cabling that maximizes the signal-to-noise ratio. An off-the­shelf ECG monitor may be connected to this connector (for example, the THM100, Indus Instruments). Pin 1 (ECG ground) in this particular case, may be left unconnected.
Nonin OEM III module (internal OEM Oxymetry), Nonin.com)
the IVIS Spectrum imaging chamber.
31
Page 38
6. IVIS® Spectrum Components
Note: A 15 pin D-sub male mating connector is required (supplied by user).
Note: Connector type in shelf--MS3119E16-26PS. Mating connector required (supplied by
customer). Manufacturer: Amphenol Aerospace, part no. MS3126F16-26P.
Table 6.1 Connector pin usage
Pin Number Function
12 and 13 Use these pins to monitor the skin or rectal temperature of the animal.
The use of a copper-constantan thermocouple is recommended. These pins may also be used to connect to the THM 100 ECG monitor (Indus Instruments).
Figure 6.8 Back panel connector
Figure 6.9 Stage connector
32
Page 39
®
Spectrum Hardware Manual
IVIS

6.11 High Reflectance Hemisphere

You can special order a high reflectance hemisphere that is used to check for light contamination in the imaging chamber
The hemisphere does not emit any photons when it is imaged in a light-tight and contaminant-free imaging chamber. As a result, it is not visible in a luminescent image. If a light leak or an internal phosphorescent contaminant is present in the imaging chamber, the hemisphere reflects that light and is visible in a luminescent image. If you suspect that there is light contamination in the imaging chamber, see Light contamination-External, page 47.
Figure 6.10 High reflectance hemisphere
(Figure 6.10).
33
Page 40
6. IVIS® Spectrum Components
[This page intentionally blank.]
34
Page 41
®
NOTICE
CAUTION
!
Spectrum Hardware Manual
IVIS

Operating the IVIS® Spectrum

7
Restarting the System . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Gas Plumbing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Door Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Centering a Subject in the Field of View . . . . . . . . . . . . . . . . 39
Fluorescent Imaging . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Imaging Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
System Shut Down Procedure . . . . . . . . . . . . . . . . . . . . . . 42
The IVIS Spectrum is intended for use in biophotonic or fluorescent imaging procedures. The system is designed to detect extremely low-level light emissions that are orders of magnitude dimmer than can be detected by the naked eye. The IVIS Spectrum allows you to monitor and record cellular and genetic activity within a living organism, in real time. The imaging system captures, quantifies, and images the light emitted by a sample.

7.1 Restarting the System

Electrical Control & Status Lights
All components of the IVIS Spectrum should be left on unless the system will not be used for more than 30 days. It is also important to leave the system on to enable automatic overnight electronic background measurements. Periodically rebooting the computer is permissible and does not affect the camera operation.
The IVIS Spectrum console has two electrical switches and one status light
(Figure 7.1). The color of the status light indicates the state of the instrument (Table 7.1). The main power on/off switch (circuit breaker) is located on the rear
panel and controls the AC power to the instrument.
The front panel contains a recessed toggle switch that distributes power within the system console. This switch can be in the standby position (down) or on position (up) and is equipped with a locking mechanism to prevent accidental activation. The shaft of the switch must be pulled out to allow movement up or down. The gas control valve is to the left of the toggle switch. For more details on the gas control valve, see Gas Plumbing, page 37.
CAUTION! Placing the toggle switch in standby mode turns off the ThermoCube cooling unit and camera power supply. All low voltage equipment remains on.
35
Page 42
7. Operating the IVIS® Spectrum
Status light
Gas valve
Standby toggle switch
Bottom illumination
input optics
Figure 7.1 IVIS Spectrum
Table 7.1 IVIS Spectrum status light modes
Status Light Is... Indicates...
Blue The system is in standby mode.
Blue and blinking The CCD camera cooling system is disabled due to a power
Red The system is initializing or acquiring an image.
Green The system is ready.
interruption. Use the Living Image software to initialize the system and enable the cooling system.
Startup Procedure The same procedure is used to restart the system whether the shut down was
intentional or due to a power outage.
1. Confirm that the rear panel switch is off, the front panel toggle switch is in the down position (in standby), and the power cord is plugged into the console and the wall socket.
2. Turn on the computer and monitor, and start the Living Image
®
software.
3. Turn on the main power switch on the rear panel.
— The status light is blue.
4. Turn on the front panel toggle switch by pulling the toggle switch out and up.
— The status light is blue and blinking.
36
Page 43
®
WARNING
!
The temperature square is red at startup and turns green when the operating temperature is reached. Click the temperature square to display the demand and measured temperature of the CCD and stage.
Temperature box color indicates:
System not initialized.
System is initialized, but CCD temperature is out of range.
System is initialized and CCD is at or within acceptable range of the demand temperature (-90° C) and locked. The system is ready for imaging.
Spectrum Hardware Manual
IVIS
5. In the Living Image software, click Initialize IVIS system in the IVIS
System Control panel
(Figure 7.2).
— The status light is red.
6. Wait until the camera has cooled to operating temperature (requires about ten minutes).
— The status light is green when the system is ready for operation.
You can monitor the system temperature in the Living Image software
(Figure 7.2).

7.2 Gas Plumbing

Figure 7.2 IVIS System control panel in the Living Image software
The instrument is ready for operation when the camera temperature is locked at -90 C, as indicated by the green temperature square in the IVIS System Control panel. For operating instructions, see the Living Image Software Manual.
WARNING! Use only isoflurane with the IVIS® Spectrum. DO NOT USE FLAMMABLE ANESTHESIA GAS.
37
Page 44
7. Operating the IVIS® Spectrum
CAUTION
!
NOTICE
CAUTION! Caliper recommends using the XGI-8 Gas Anesthesia System
(Figure 7.3) when imaging small animals. This system supplies a controlled
amount of isoflurane to the imaging chamber and continuously reduces the build-up of isoflurane in the chamber. If you want to use a gas other than the recommended isoflurane/oxygen gas mixture or pure air, contact IVIS Imaging Systems Technology Support.
38
Figure 7.3 XGI-8 Gas Anesthesia System
Be careful to use only tubing and other plumbing fixtures that do not fluoresce or phosphoresce (glow) in the imaging chamber. Contact IVIS Imaging Systems Technology Support for a list of acceptable materials.
The IVIS Spectrum is equipped with plumbing connections for supplying isoflurane anesthetic gas to the imaging chamber and exhausting gases out. The flow of gas into the imaging chamber is manually controlled by the gas valve on the front panel of the console. The imaging chamber does not include a gas scavenging system. This is provided by suitable auxiliary equipment such as the XGI-8 Gas Anesthesia System.
Gas ports are located on the rear of the system console and are labeled GAS IN and GAS OUT. GAS IN means the direction of flow is into that port. Similarly, the port labeled GAS OUT means that flow can be exhausted out of this port. The flow of gas from the rear panel GAS IN port is directed to the main gas control valve on the front of the system and proceeds to a distribution block located on the imaging platform where it exits through the port labeled GAS.
Page 45
®
NOTICE
WARNING
!
WARNING
!
Spectrum Hardware Manual
IVIS

7.3 Door Operation

The mouse anesthesia manifold (part no. 117706) with transparent nose cones is provided with the IVIS
®
Spectrum. The manifold has capacity for up to five
mice and has a low profile (height) for operation with the system.
The IVIS Imaging System 100 Series and the XGI-8 Gas Anesthesia System are equipped with a different five-station manifold (PN 118274) that is taller than the low profile manifold supplied with the IVIS Spectrum. Only use the low profile manifold with the IVIS Spectrum, which permits imaging the smallest field of view at the highest platform height (FOV A).
The manifold port labeled GAS connects to the GAS port on the platform distribution block. The anesthesia manifold also has a port labeled EXH that should be connected to the EXH port on the platform distribution block. The vacuum or exhaust circuit connects to the GAS OUT port on the rear of the system console. No gases are exhausted from the imaging chamber unless gas scavenging equipment such as the XGI-8 Gas Anesthesia System is used.
The door of the imaging chamber cannot be opened when the system status light is red. The door is equipped with an electromagnetic lock to prevent accidental opening during an imaging session.
WARNING! Never try to defeat the door interlock or force the door open during image acquisition when the status light is red. Such action could expose the user to a severe pinch hazard from the rapidly moving imaging platform.

7.4 Centering a Subject in the Field of View

WARNING! The IVIS Spectrum is equipped with a laser similar to the type
used for laser pointers (532nm, output < 1mW). Users are protected from eye damage due to the eye’s physiological response to bright light by blinking. A laser beam can be harmful to the eyes. ABSOLUTELY DO NOT look directly into the laser beam when the laser is turned on. Be aware that laser light can be easily reflected off of shiny surfaces and that this reflected light can be dangerous. When the laser is on, remove all watches and jewelry when placing hands in the imaging chamber.
1. In the Living Image software, choose the Enable alignment grid option in
the IVIS System Control panel (Figure 7.4).
39
Page 46
7. Operating the IVIS® Spectrum
NOTICE
NOTICE
To display the alignment grid, choose Alignment Grid On.
Figure 7.4 IVIS System Control panel
2. Open the imaging chamber door to see the alignment grid projected by the laser/galvanometer on the imaging platform
(Figure 7.5).
— The field of view (FOV) setting defines the size of the squares in the
alignment grid.
3. To change the size of the squares in the alignment grid, select a different FOV setting.
For example FOV settings, see Figure 6.4, page 28.
40
Figure 7.5 Laser alignment grid
The alignment laser turns off after 60 seconds. Repeat step 1 to turn on the laser.
4. Using the alignment grid as a guide, place the specimen(s) at the locations of interest, then proceed with image acquisition.
The Living Image® software controls the photographic illumination LED intensity.
Page 47
®
Choose luminescent or fluorescent imaging
Spectrum Hardware Manual
IVIS

7.5 Fluorescent Imaging

The IVIS® Spectrum has both bioluminescent and fluorescent imaging capabilities. To change from bioluminescent to fluorescent operation, select fluorescence mode in the IVIS System Control panel and select an appropriate excitation and emission filter
Figure 7.6 IVIS System Control panel
(Figure 7.6).
A 150 Watt quartz tungsten halogen lamp with a dichroic reflector provides the
®
fluorescence excitation light. The Living Image
software controls the lamp intensity level. The fluorescence light source is a module located behind the rear panel and is not user serviceable.
The lamp output is delivered to the excitation filter wheel assembly located in the rear compartment, where it is collimated and transmitted through a fluorescence excitation filter. The 12 excitation filter wheel locations provide 10 fluorescence filters. Two filter slots contain a light block that is used during bioluminescent imaging to prevent external light from entering the imaging chamber. The 12-position excitation filter wheel is motorized and controlled through the Living Image software.
Upon leaving the excitation filter, the light enters an optical switch that directs the light to either the top of the imaging chamber or to the bottom of the stage. When top illumination is selected, the light is equally split to four diffusers to provide uniform stage illumination. When bottom or trans-illumination is selected, the excitation light is directed to one of the 2 mm diameter holes in the aperture plate located in the stage surface. This removable plate has a 15 x 23 array of 2mm diameter holes which provide discrete illumination locations that are user selectable.
You can replace the array plate with a 3 mm thick glass plate or a well plate holder to provide alternate illumination possibilities. A blank plate is also provided for when the bottom illumination equipment is not in use. For more details on how to use the top and bottom illumination systems, see the Living Image Software User’s Manual.
Selecting an Emission Filter
The IVIS Spectrum has a factory-installed filter set. The filters are mounted on a rotating filter wheel and can be removed or replaced with filters with other spectral characteristics. For more details, see Emission & Excitation Filter
Wheels, page 29. For information on how to replace a filter, see page 53. See
®
the Living Image
Software Manual for more details on how to select an
emission filter.
41
Page 48
7. Operating the IVIS® Spectrum
Selecting an Excitation Filter
In fluorescence mode, the excitation filter is automatically selected based on the wavelength of the emission (imaging) filter selected from the IVIS System Control panel. However, you can override the automatic selection in the IVIS System Control panel.
The excitation filters are housed in a filter wheel module that is not accessible to the user. Unlike the emission filters, the user cannot replace these filters. Contact IVIS Imaging Systems Technology Support if an excitation filter must be replaced.

7.6 Imaging Basics

Black Paper Although the platform is black anodized, it is recommended that you image
items on a high quality black paper, especially biological specimens. Caliper Corporation has surveyed many types of paper and recommends Swarthmore, Artagain, Black, part no. 445-109, size 8.5 inch x 11 inch. This paper prevents illumination reflections and helps keep the stage clean.
Low Fluorescence Mat When operating the system in the fluorescence mode, use the Low Fluorescence
Imaging Mat (part no. 117839) to reduce background fluorescence. The mat is made from a low reflectance textured plastic that is easily cleaned.
Glowing Materials Always keep in mind that nearly EVERYTHING glows (that is, has the
potential to phosphoresce and contaminate the image). Most plastics, almost all tape, plants, paint, rodent food (mostly plants), mouse urine, and animal bedding have been found to glow.
®
Use caution when introducing materials into the IVIS to prescreen all items by imaging them alone, before imaging them with samples under study. Caliper recommends using non-powdered gloves when working with IVIS Spectrum equipment.
Spectrum. It is advisable

7.7 System Shut Down Procedure

Caliper Corporation does not recommend power cycling the IVIS Spectrum (turning the system components on and off) on a daily basis. Leaving the system on keeps the camera cold and the electronics systems stable. The software is preconfigured to run automatic background accumulation and self-diagnostics overnight. The platform temperature and the computer monitor may be turned off.
If it is necessary to shut down the system for any reason, such as long term storage (more than 30 days), it is important to follow the procedure below.
1. Close the Living Image prompt.
2. Turn off the computer.
3. Turn off any gas to the imaging chamber and disconnect the gas supply if
necessary. If you are using the XGI-8 Gas Anesthesia system, follow the shut down procedure in the XGI-8 Anesthesia System User Manual.
4. Turn off (standby) the toggle switch on the front of the IVIS Spectrum.
42
®
software and save any information of interest at the
Page 49
®
Spectrum Hardware Manual
IVIS
5. Turn off the main circuit breaker switch located on the rear of the IVIS Spectrum and disconnect the power cord from the wall socket.
6. Cover the gas inlet and outlet ports with masking tape to prevent entry of dust.
7. Remove any loose objects from the imaging chamber and close the door.
If you plan to move the system, see important information on page 55.
If you have any problems during the shut down procedure, please contact IVIS Imaging Systems Technology Support for assistance.
43
Page 50
7. Operating the IVIS® Spectrum
[This page intentionally blank.]
44
Page 51
®
Click the temperature square to display the demand and measured temperatures of the camera and imaging chamber.
Spectrum Hardware Manual
IVIS

Troubleshooting

8
Measured Temperature Does Not Equal the Demand Temperature . 45
Photographic Image Is Unacceptable . . . . . . . . . . . . . . . . . . 46
Luminescent Image Is Unacceptable . . . . . . . . . . . . . . . . . . 47
No Image Is Produced . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Hardware Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Accessing the Area Under the Imaging Platform . . . . . . . . . . . . 50

8.1 Measured Temperature Does Not Equal the Demand Temperature

At start up, the Living Image® software programs the CCD camera to maintain the CCD at -90° C. If the camera power supply and ThermoCube chiller unit remain on, the system maintains this temperature regardless of whether the Living Image software is open or the computer is turned on.
To check the temperature of the CCD, click the Temperature square (red or
green) in the Camera Control Panel of the Living Image software (Figure
.
8.1)
Figure 8.1 IVIS System control panel in the Living Image software
Problem Possible Cause Corrective Action
Measured temperature is warmer than the demand temperature.
A problem may exist with the camera or the ThermoCube unit.
Contact IVIS Imaging Systems Technology Support.
45
Page 52
8. Troubleshooting

8.2 Photographic Image Is Unacceptable

Default camera controls are programmed during the initialization of the Living
®
Image image. Check that the exposure time, f/stop and binning levels are set to the default conditions.
If the f/stop is open too wide, the image may be saturated and less focused. Binning to a higher value or increasing the exposure time can produce an image that is too bright and lacks resolution. If you have questions about the photographic settings, please contact IVIS Imaging Systems Technology Support.
Also, refer to the Living Image Software Manual for further details on acquiring images.
Problem Possible Cause Corrective Action
Image is streaked. Subject moved during the
exposure.
Image is blurry. Subject height is not correctly
entered in the Living Image software.
software. Changes to these settings can greatly affect the photographic
Check to see if the subject may have moved. If the subject is not on the sample stage, it is probably on the floor of the imaging chamber. If the sample has moved, locate and reanesthetize it. If gas anesthesia is being used, confirm that the anesthesia is turned on and the flow rate is appropriate.
If you are using the “Sample Height”setting to determine the focus, confirm that the correct height of the sample is entered. Alternatively, Auto-Focus may be used to determine the sample height.
If you are using the “Manual”setting to determine the focus, confirm that the focus has been manually adjusted.
A white spot appears in the center of the field of view.
Incorrect f/stop setting. The f/stop for photographs should be set to f/8. An f/stop
An excessively moist environment in the imaging chamber can result in condensation on the CCD window (Figure 8.2).
Figure 8.2 Condensation example
smaller than 8 reduces the depth of field in the photograph.
Remove excess moisture in the imaging chamber and allow it to thoroughly dry. Place a desiccant in the imaging chamber to decrease the drying time.
Do not touch or let specimens touch the window at the top of the imaging chamber. For more details on the care of this window, see page page 52.
If the problem persists, contact IVIS Imaging Systems Technology Support for assistance.
46
Page 53
®
Spectrum Hardware Manual
IVIS

8.3 Luminescent Image Is Unacceptable

Binning, f/stop, and exposure time affect the appearance of a luminescent
®
image. Please refer to the Living Image
Software Manual for instructions on
setting binning, exposure time, and f/stop values.
In order to function properly and reduce camera noise, the CCD camera must be cooled to the demand temperature before acquiring an image. If the camera is not cooled to the demand temperature, imaging may result in false positive signals.
Problem Corrective Action
Light contamination­Internal
Light contamination­External
Camera noise Verify that the camera is cooled to the demand temperature.
Check to see that there are no extraneous light sources in the imaging chamber. Many substances phosphoresce when exposed to light. Be especially cautious of plastics and substances that contain pigment. Be sure to pre-screen any substance or material before performing actual experiments.
A 2” diameter high reflectance hemisphere can be special ordered to help check for light leaks (XRH-1, part number 118937, Figure 8.3).
To check for light leaks:
1. Place the high reflectance hemisphere in the imaging chamber on the stage using a subject height of 3.5 cm at field of view D.
2. Take a luminescent image of the hemisphere using the luminescent settings: f/stop = 1, Binning = Large (high sensitivity), and exposure time = 5 minutes.
If the hemisphere can be easily seen, there is a light leak. Contact IVIS Imaging Systems Technology Support for assistance.
1. Check the measured temperature in the Camera Control panel to ensure that it is locked. If the camera temperature is locked, the camera temperature box is green.
2. If the camera temperature box is red, click the red box to display the actual temperature. See
Measured temperature is warmer than the demand temperature., page 45.
Figure 8.3 High reflectance hemisphere
Wear powderless gloves when handling the hemisphere to prevent surface contamination that may glow when using the IVIS® Spectrum.
47
Page 54
8. Troubleshooting
VOLTAGE

8.4 No Image Is Produced

If no image is produced, there may be an error in the Living Image software, a problem with the physical connections to the camera, or a hardware failure.
1. Close the Living Image
2. Restart the Living Image software and try to acquire an image.
3. If after restarting the computer, you are still unable to produce an image,

8.5 Hardware Problems

Due to its integrated design, almost all of the components in the IVIS Spectrum are not user-serviceable.
®
software and restart the computer.
contact IVIS Imaging Systems Technology Support for assistance.
VOLTAGE! Do not remove any panel from the console, except for the filter wheel access panel. Removing a panel could expose the user to hazardous voltages.
Circuit Breaker (Power Entry Module)
If you suspect a problem with the instrument hardware, contact IVIS Imaging Systems Technology Support.
Under normal operation, the circuit breaker that is incorporated into the rear panel power entry module (main power switch) should not trip. Frequent tripping of this circuit breaker may indicate an underlying electrical problem. If this occurs, unplug the system and contact IVIS Imaging Systems Technology Support.
If the circuit breaker is occasionally tripped:
1. Confirm that the rear panel switch is off, the front panel toggle switch is down (standby), and the power cord is plugged into the console and the wall socket.
®
2. Turn on the computer and, if necessary, start the Living Image
software.
3. Turn on the rear panel main power switch.
— The status light is blue.
4. Turn on the front panel toggle switch by pulling the toggle out and up.
— The status light is blue and blinking.
5. Initialize the system.
— The status light is red and turns green when the system is initialized.
Wait until the camera has cooled to the demand temperature before you begin an imaging session.
Computer Connections Never disconnect or wiggle the rear panel computer connections when the
®
Spectrum is powered on. If you suspect that any of the rear panel
IVIS computer connections are loose, follow the procedure below.
1. Close the Living Image software and shut down the computer.
2. Turn off the front panel toggle switch (standby).
48
Page 55
®
VOLTAGE
Toggle switch
Front panel
Back panel
Spectrum Hardware Manual
IVIS
3. Turn off the rear panel circuit breaker switch.
Figure 8.4 IVIS Spectrum, front and back panels
4. Check the connections between the computer and the rear panel of the IVIS Spectrum. Tighten any loose connection.
5. Turn on the computer and start the Living Image software.
6. Turn on the rear panel main power switch.
— The status light is blue.
7. Turn on the front panel toggle switch by pulling the toggle out and up.
— The status light is blue and blinking.
8. Initialize the system.
— The status light is red and turns green when initialization is complete.
Wait until the camera has cooled to the demand temperature before you begin an imaging session.
Laser/Galvanometer The laser/galvanometer is not user-serviceable. If you notice any changes to the
alignment grid (for example, change in the brightness or position of the grid), contact IVIS Imaging Systems Technology Support.
Lamp Replacement The quartz halogen lamp that provides the fluorescent light source typically
operates for thousands of hours. This light source is housed in a compartment that contains hazardous voltages and is not user accessible or serviceable.
VOLTAGE! DO NOT attempt to service the quartz halogen lamp.
If the lamp brightness is decreased or if lamp replacement is required, contact IVIS Imaging Systems Technology Support.
49
Page 56
8. Troubleshooting
The stage lock down bar is located at the bottom front edge of the imaging platform.
Pull the bar down from the stowed position and place it as shown above. Return the bar to its stow position before returning the imaging platform to the operating position.

8.6 Accessing the Area Under the Imaging Platform

During normal operation, the imaging platform always returns to the loading position before the door can be opened. This helps ensure user safety and convenience when loading subjects. Occasionally you may need access to the area beneath the platform for cleaning or to retrieve a subject. Since the platform moves rapidly and produces strong forces, the area beneath the platform is not accessible unless the normal operation mode is overridden.
To access the space below the imaging platform:
1. Remove all items from the imaging stage that are taller than 2 cm and close the door.
2. In the IVIS System Control panel, select Service from the FOV drop-down
menu.
— The platform moves to the service position and the door can be opened.
3. Place the stage lock down bar in position as shown in Figure 8.5.
Be careful not to disturb the platform heater cable or the gas tubing located in this area.
To return the platform to the operating position:
1. Return the lock down safety bar to the stowed position under the front edge of the imaging platform and close the door.
2. In the IVIS System Control panel, select Load from the FOV drop-down
menu.
— The imaging platform returns to the normal operating position.
Figure 8.5 Imaging platform and the stage lock down bar
If you need to access the space under the imaging platform, use the lock down bar as a safety precaution.
50
Page 57
®
DANGER
!
WARNING
!
IMPORTANT
!
Spectrum Hardware Manual
IVIS

Care & Maintenance

9
Cleaning the IVIS Spectrum . . . . . . . . . . . . . . . . . . . . . . . 51
Cleaning the Lens Protection Window . . . . . . . . . . . . . . . . . 52
Changing the Emission Filters Used in Fluorescence Imaging . . . . 53
Moving the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55

9.1 Cleaning the IVIS Spectrum

If necessary, wipe the exterior surface of the IVIS Spectrum console with a soft cloth. If the system requires more aggressive cleaning or sterilization, contact IVIS Imaging Systems Technology Support.
DANGER! DO NOT use sprays or fluids to clean the exterior or interior of the module. Do not allow fluids of any kind to penetrate the electronics cabinet under any circumstances. Sprays and liquids that come into contact with the IVIS Spectrum may result in damage to the system or electrocution.
Approved Cleaning Solutions
WARNING! Do not use fluids or moistened towels to clean any part of the
system console where electrical or fiber optic cables make connections. Do not use fluids of any kind near the filter wheel assembly. Turn off the electrical power to the IVIS Spectrum by turning the front panel toggle switch to standby and turning off the rear panel switch before engaging in cleaning operations that use fluids. The imaging chamber power switch is located on the rear of the system console.
The compounds shown in Table 9.1 do not damage the internal finish of the
®
Spectrum imaging chamber and are suitable as cleaners, if required. Do
IVIS not use any solution not included in this list. In particular, avoid strong bases, bleach, or acids that may potentially damage the unit and compromise its operation.
ALERT! Do not spray cleaning solutions in the imaging chamber.
51
Page 58
9. Care & Maintenance
NOTICE
Table 9.1 Acceptable cleaning solutions for the IVIS® Spectrum imaging chamber
Cleaning Solution Manufacturer
®
Cidexplus
70% methyl alcohol/30% deionized water solution
70% ethyl alcohol/30% deionized water solution
Sporicidin phenol)
Clidox-s
Solution (3.4% glutaraldehyde) Johnson & Johnson Medical
®
Sterilizing Solution (1.56%
®
Disinfectant Pharmacal Research Laboratories, Inc.
Sporicidin International
Caliper makes no claims as to the sterility of the IVIS imaging chamber after using the solutions in Table 9.1. Please refer to the manufacturer’s literature for information as to the applicability of the compound for the organism of interest. Consider dedicating an imaging system for immunodeficient animals. This will reduce the risk of cross contamination.
Cleaning the Imaging Platform
It is recommended that you use a lint-free wipe, such as Scott Pure® wipe or a
®
Kaydry EX-L
wipe to minimize the presence of particulate matter in the
imaging chamber.
After saturating a lint-free wipe, clean the internal surfaces using a gentle circular motion. Do not pour or spray the solution directly onto internal surfaces. Rinse surfaces using a wipe saturated with sterile deionized water. Do not allow puddles of water to remain on the surfaces.
Avoid cleaning the ceiling of the image chamber except when necessary. When cleaning the chamber ceiling, use extreme care to avoid damaging the lens protection window and the light diffusers in this area.
To avoid any phosphorescence from the cleaner, be sure that the surfaces are dry before using the imaging chamber.

9.2 Cleaning the Lens Protection Window

Contact IVIS Imaging Systems Technology Support for information about cleaning or sterilizing any of the optical components or replacing the optical filters. A clear window at the top of the imaging chamber protects the lenses and other optical components from contamination. Do not touch or allow subjects to touch this window, otherwise image quality may be impaired. Further, the glass window could be cracked or broken if struck by a hard object. The lens protection window should be cleaned periodically.
To clean the surface of the lens protection window:
1. Blow off any accumulated dust using a Dust-Off® compressed gas duster.
2. Dampen a lint-free cloth with isopropyl alcohol and gently swab the
window surface to remove any smudges or streaks.
3. Repeat step 2 using a clean area of the cloth or a new cloth until the window
is clean.
52
Page 59
®
NOTICE
NOTICE
Spectrum Hardware Manual
IVIS
4. Take an image to confirm that the image quality is restored.

9.3 Changing the Emission Filters Used in Fluorescence Imaging

The following steps explain how to change one or more of the optional emission filters used for fluorescence imaging in the IVIS
®
Spectrum.
Before you begin, gather the following tools:
5/32 hex key
#2 Phillips head screw driver
One pair of lint free, powderless gloves
1. In the Living Image
®
software, save any important data, then exit the
program.
2. Turn the front panel switch to off (standby) and turn off the main power at the rear of the system console.
3. Using the 5/32 hex key, remove the eight screws that hold the filter access panel in place on the upper right side of the console.
— The panel is held in place by four pins until it is pulled forward. be
careful not to damage the light-sealing gasket when removing the panel.
The panel has a machined recess on the inner left-hand side (towards the front of the console). Noting this orientation is helpful when you reinstall the panel.
4. Notice that the 1/2shaft on the right side of the filter wheel compartment has a carriage clamp lock with a handle. Turn the handle counter-clockwise to loosen the clamp so that it can slide freely over the shaft.
Starting with step 5, wear lint-free, powderless gloves to prevent smudging the filters.
5. Grip the filter wheel by the metal rim and pull the filter wheel carriage forward until it stops. Be careful not to touch the glass filters or to dislodge the two drive belts.
Since the carriage slides out about eight inches, only two or three of the filters will be visible. The filter wheel holds the filters numbered 1-11.
6. To remove a filter from the filter wheel:
a. Slowly rotate the filter wheel until the filter of interest is exposed.
b. Use a #2 Phillips head screw driver to loosen the flat head screw ONE
TURN.
The screw has a captured O-ring that retains the glass emission filter.
c. Use two gloved fingers to remove the filter. Alternatively, gently push
the filter up with a gloved finger.
53
Page 60
9. Care & Maintenance
NOTICE
NOTICE
NOTICE
NOTICE
7. To clean a filter, use a Dust-Off® compressed gas duster to apply an air stream at an angle to the filter surface, otherwise dust particles could be driven into the filter coating.
Do not blow air onto a filter in the filter wheel as dust could be spread to other optical components.
8. To install a filter, insert the filter into the filter wheel recess and tighten the O-ring screw until it bottoms out (about one turn).
Make sure that the filter retaining screw and O-ring are fully seated to prevent interference or rubbing against components above them. The top of the filter metal ring should be nearly flush with the wheel. The filter encapsulating metal ring has one side that is thinner than the other. The thin part of the ring should face downward in the wheel recess so that the arrow that is printed on the side of the filter points up.
9. Push the filter wheel carriage back into the console by placing one finger on the rim of the upper filter wheel.
10. While you hold the carriage in the closed position, turn the carriage-locking clamp in the clock-wise direction. Be careful not to dislodge any drive belts off of their pulleys when locking the clamp.
11. Replace the access panel by positioning it onto the four pins. The inner machined out recess should be on your left side.
12. Loosely install the eight screws, then lightly tighten the four central screws and in a crisscross pattern. Lightly tighten the four outer screws in a crisscross pattern.
Do not over tighten the screws. The panel is designed to seal out light without large clamping force.
13. Restart the system following the procedure on page 35.
If you change a filter or a filter position, you must also change the corresponding filter label in the Living Image® software. For more details, see the Living Image Software Manual.
54
Page 61
®
CAUTION
!
Spectrum Hardware Manual
IVIS

9.4 Moving the System

Removing the Dome It may be necessary to remove the plastic camera dome when you move the
IVIS® Spectrum though a doorway or other low overhang.
To remove the dome, wrap a flat blade screw driver with a thin cloth or
paper towel and use the screw driver to gently pry the dome off at the rear of the console.
Moving the System If the console is only repositioned a short distance within the range of the power
cord and cables, the power need not be turned off. However, be careful not to accidentally turn off the power when gripping the system at the right rear.
®
When possible, two people should move the IVIS and it is difficult to see potential obstacles when pushing the console from behind. Make sure that the path is clear of items that could jolt the casters and cause the console to tip. Move the console slowly so that you are always in control of its motion.
To move the system:
1. Put the front panel toggle switch in the standby position.
2. Turn off the main power switch on the rear panel and disconnect the power cord from the wall socket and the console.
3. Remove all cables from the rear panel of the console.
Spectrum. The console is tall
CAUTION! Any cables that dangle from the console could cause the system to topple if caught by a caster.
4. Unlock the four casters.
5. Grip the console at a convenient location, but do not use the door handle as a grip. Be careful when gripping the right rear of the console so that you do not damage the connectors or fan grills.
6. After the system is positioned at the new location, lock the casters.
7. Reconnect the cables and power cord.
8. If necessary, reinstall the dome.
9. To restart the system:
a. Confirm that the rear panel switch is off, the front panel toggle switch is
down (standby), and all cables are reconnected.
b. Confirm that the power cord is plugged into the console and the wall
socket.
®
c. Turn on the computer and start the Living Image
software.
d. Turn on the rear panel main power switch.
— The status light is blue.
e. Turn on the front panel toggle switch by pulling the toggle out and up.
— The status light is blue and blinking.
f. Initialize the system.
55
Page 62
9. Care & Maintenance
— The status light is red.
The status light turns green when the system is initialized. Wait until the camera has cooled to the demand temperature before you begin an imaging session (about ten minutes).
56
Page 63
®
VOLTAGE
Spectrum Hardware Manual
IVIS
10

Servicing & Maintaining the ThermoCube Chiller Unit

10.1 Required Tools

Required Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Turning Off Electric Power to the Main Console . . . . . . . . . . . . 57
Removing the Console Rear Panel . . . . . . . . . . . . . . . . . . . 58
Sliding the Tray Out . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Filling the ThermoCube Unit . . . . . . . . . . . . . . . . . . . . . . . 60
Reassembling the System . . . . . . . . . . . . . . . . . . . . . . . . 61
Restarting the IVIS Spectrum . . . . . . . . . . . . . . . . . . . . . . 61
Over time, the coolant level in the ThermoCube thermoelectric chiller unit may diminish and require replenishing. It is recommended that you check the coolant level in the ThermoCube unit every six months. This chapter explains how to add coolant to the Thermocube unit.
The following tools and items are required, but not provided.
Allen Wrenches (hex head wrench): 3/32" and ¼" measured across the
flats (AF).
Measuring cup with a spout.
Spectral TE Camera Cooling Mixture (part no. 119219, includes a 3/32"
and ¼" allen wrench) or a solution of 25% pure ethanol, 75% distilled water.
Paper towel or cloth for clean up.

10.2 Turning Off Electric Power to the Main Console

VOLTAGE! Do not remove any panels from the console until you turn off the
IVIS Imaging System disconnect the power cord. Otherwise, electric shock could result.
1. Pull the front toggle switch out and down to the standby position.
2. Turn off the main power switch on the rear panel and disconnect the console power cord from the wall socket.
3. Move any cables away from the lowest rear panel of the console so that the panel can be removed
(Figure 10.1). (No other cables need to be removed.)
57
Page 64
10. Servicing & Maintaining the ThermoCube Chiller Unit
CAUTION
!
ThermoCube unit
Figure 10.1 Lower rear panel
Note that the power cord has been removed from the console.

10.3 Removing the Console Rear Panel

1. Use a 3/32" hex head wrench to remove the screws that secure the bottom rear (fan) panel
(Figure 10.2).
2. Carefully remove the panel and lean it against the side of the console.
CAUTION! Be careful not to pull too hard on the rear fan panel, otherwise its plug may become disconnected. If this happens, contact IVIS Imaging Systems Technology Support).
.
Figure 10.2 Removing the lower rear panel using a hex head wrench (left) and ThermoCube unit exposed (right)
58
Page 65
®
NOTICE
Spectrum Hardware Manual
IVIS

10.4 Sliding the Tray Out

The ThermoCube unit sits on a tray that must be pulled out to access the unit.
1. Use the ¼" hex head (Allen) wrench to loosen and remove the bolt that secures the tray under the ThermoCube unit to the console
(Figure 10.3).
Figure 10.3 Loosen the tray locking bolt
2. Move any cables or tubes out of the way so that the tray can be pulled out
3. Pull out the tray with the ThermoCube unit
The ThermoCube unit has its own power switch located at the rear of the unit. The power switch is not visible until the unit is pulled out. Make sure that the power switch is not accidentally turned off when you slide the tray out or back into the console.
4. Remove the fill cap on the ThermoCube chiller and check coolant level. If
freely.
(Figure 10.4).
the coolant level is half way up the neck of the tank, no additional coolant is required.
Figure 10.4 Removing the fill cap
59
Page 66
10. Servicing & Maintaining the ThermoCube Chiller Unit
CAUTION
!
NOTICE
WARNING
!

10.5 Filling the ThermoCube Unit

1. If additional coolant is required, wrap a piece of cloth or a paper towel around the neck of the tank to absorb accidental spills
2. Using a spouted measuring cup, add small quantities of the 25% ethanol,
75% distilled water solution to the ThermoCube tank until the coolant level is half way up the neck of the tank. Use care not to spill the liquid and wipe any liquid from the top or walls of the chiller.
CAUTION! DO NOT OVERFILL the ThermoCube unit. Over filling the tank can cause fluid to run down the chiller walls and contact the power plug.
(Figure 10.5).
60
Figure 10.5 Adding coolant to the ThermoCube unit
If the Caliper Spectral TE Camera Cooling Mixture (part no. 119219) is not available, a suitable substitute can be made from ordinary 80 proof Vodka and distilled water. To make 750 ml of the disinfecting coolant, mix 470 ml of 80 proof alcohol (40% ethanol) and 280 ml distilled water.
WARNING! If you prepare your own ethanol solution, adhere to the 25% (by volume) pure ethanol concentration. A higher alcohol concentration may be flammable and a lower concentration may not suppress organic growth in the cooling system or camera.
Page 67
®
Spectrum Hardware Manual
IVIS

10.6 Reassembling the System

1. Replace the cap on the ThermoCube unit. Make sure that the cap is fully tightened to prevent rapid coolant evaporation.
2. Slide the tray back into the console while you hold the cables and hoses out
of the way to prevent snagging.
3. Replace the tray bolt using the ¼" hex head (Allen) wrench
4. Confirm that the ThermoCube power switch is on and that no tools or other
hardware remain inside the IVIS.
5. Replace the lower rear panel using a 3/32" hex head wrench to tighten the
screws

10.7 Restarting the IVIS Spectrum

1. Confirm that the IVIS system rear panel power switch is off and that the front panel toggle switch is in the down (standby) position.
2. Reconnect the power cord first to the console and then to the wall socket.
3. Confirm that all other cables are properly connected.
4. Turn on the computer.
5. Turn on the rear panel power switch.
— The front status light is blue.
6. Turn on the front panel toggle switch by pulling out and up.
— The status light is blue and blinking.
7. Start the Living Image software and initialize the system.
— The status light is red and turns green when the system is initialized.
Wait until the camera has cooled to the demand temperature before you begin an imaging session (about 10 minutes).
(Figure 10.3).
(Figure 10.2).
61
Page 68
10. Servicing & Maintaining the ThermoCube Chiller Unit
[This page intentionally blank.]
62
Page 69
®
Spectrum Hardware Manual
IVIS

Appendix A Spare Parts List

Description Caliper Part Number
Mouse Inhalation Nose Cone 117150
Low Profile, 5 Station Manifold 117706
XRH-1 High-Reflectance Hemisphere with container 118937
Black Paper, Swarthmore, Artagain. 9x12”, 24 sheet pad 117837
Low Fluorescence Imaging Mat, 11.6” square 117839
63
Page 70
A. Spare Parts List
[This page intentionally blank.]
64
Page 71
®
Spectrum Hardware Manual
IVIS
Index
A
accessory cable 31
back connector 32
stage connector 32 anesthesia delivery system 38 anesthesia manifold 39 array plate 41
B
back connector 32 base enclosure 30 black paper 42, 63 burn hazard 12
C
camera noise 47 caution definition 5 CCD camera 26
specifications 21 centering a subject 39–40 changing emission filters 53–54 chemical & biological safety 15 chiller unit 26 circuit breaker 48 cleaning
lens protection window 52–53
system 9 cleaning solutions 51, 52 computer 31
connections 48 crash avoidance system 30
emission filter
selecting 41 emission filter wheel 29 emission filters
changing 53–54 environmental considerations 7
heat 7
location of system 7
space and ventilation 8
water and moisture 8 environmental requirements 22 excitation filter
selecting 42 excitation filter wheel 29 eye safety 12
F
field of view
centering a subject 39–40 filter wheel 29 filter wheel access panel 15 fluorescence imaging
changing emission filters 53–54 fluorescence imaging components specifications 22 fluorescent imaging 41
emission filter 41
excitation filter 42
G
gas in/out 38 gas plumbing 37 glowing materials 42
D
definitions
caution 5 voltage 5
warning 5 dome 55 door operation 39
E
electrical power requirements 21 electrical safety 11
H
heat safety 7 high reflectance hemisphere 33, 47, 63
I
image
blurry 46 streaked 46
imaging chamber 27
door 39
65
Page 72
Index
gas plumbing 37
imaging platform
accessing space below 50 lock down bar 50 warming system 29
imaging system
proper location 7
information on manual instructions 6 instrument labels
caution 6 pinch hazard 6 read manual 6
isoflurane gas anesthesia 37
L
lamp 49 laser caution label 6 laser safety 7 laser/galvanometer 30, 49 lenses 28 light contamination 47 lightning 8 limited warranty 17 location for imaging system 7 location safety 7 low fluorescence imaging mat 63 low fluorescence mat 42 low profile manifold 63 luminescent image unacceptable 47
M
mat, low fluorescence 42 measured unequal to demand temperature 45 mechanical safety 12 moving the system 9, 55–56
N
no image produced 48 nose cone 63
O
optical components
lenses 28
optical specifications 22 other equipment, chemicals, or supplies 9
P
patents 18 photographic image unacceptable 46 pinch hazard label 6 power cord 8 power line surges 8 power outages 9 power overloading 9 power sources 8
Q
quartz halogen lamp 49
R
read manual label 6 restarting the system 35–37
S
safety
burn hazard 12 chemical & biological 15 electrical 11 eye 12 laser 7 mechanical 12
water and moisture 8 servicing system 10 servicing the ThermoCube unit 57–61 shut down procedure 42–43 space and ventilation safety 8 spare parts 63
black paper 63
low fluorescence imaging mat 63
low profile manifold 63
nose cone 63
XRH-2 high reflectance hemisphere 63 specifications
CCD camera 21
electrical power requirements 21
environmental requirements 22
fluorescence components 22
optics 22
weights and dimensions 23 stage connector 32 starting the system 35–37 status lights 35–36
66
Page 73
®
Spectrum Hardware Manual
IVIS
system
cleaning 9 moving 9, 55–56 servicing 10 shut down 42–43
T
temperature
measured unequal to demand 45
ThermoCube
servicing 57–61 ThermoCube chiller unit 26 trademarks 18 transillumination 41 troubleshooting
camera noise 47
light contamination 47
measured temperature does not equal demand
temperature 45 no image produced 48 streaked image 46 unacceptable luminescent image 47 unacceptable photographic image 46
V
voltage definition 5
W
warming system 29 warning definition 5 water and moisture safety 8 weights and dimensions 23
X
XGI-8 Gas Anesthesia System 38
67
Page 74
Index