The IVIS® Spectrum Imaging System is a high-sensitivity, low noise, in vivo
imaging technology platform that enables noninvasive visualization and
tracking of cellular and genetic activity within a living organism in real time
(Figure 1.1). The system provides both bioluminescence and fluorescence
imaging capability.
For fluorescence imaging, the instrument can operate in reflectance or
transillumination mode. Filtered light from a broad-band lamp provides the
excitation source in both modes. In the reflectance mode, light is delivered to
four reflectors that are located on the ceiling of the imaging chamber. In the
transillumination mode, the excitation light is delivered to an x-y translation
assembly under the stage and focused to a 2 mm diameter beam that can be
directed to a particular location on the underside of the animal subject. The
system includes ten excitation and 18 emission filters that enable spectral
scanning of reporters over the range from 480-850 nm.
3D TomographyThe system also includes a structured light projector that enables reconstruction
of the surface topography. The 3D location and concentration of fluorescent
sources can be computed from structured light and transillumination fluorescent
images. The 3D location and strength of luminescent sources is computed from
structured light and luminescent image data.
1
Page 8
1. Welcome
High resolution monitor
XWS-260 Workstation
CCD
camera
Imaging
chamber
Compartment that contains
the ThermoCube chiller unit
and camera power supply
Acquisition
computer
XGI-8 Gas
Anesthesia
System
Figure 1.1 IVIS® Spectrum Imaging System
The IVIS Spectrum is an integrated imaging system that includes a:
CCD camera mounted on a light-tight imaging chamber
•
Camera power supply
•
Thermoelectric cooling unit
•
Fluorescence light source
•
Windows®-based computer system for data acquisition and analysis.
•
The sound-limiting enclosure beneath the imaging chamber contains the
thermoelectric cooling unit, camera power supply, and other imaging chamber
functions. All components except the computer are integrated into a single,
moveable chassis.
This manual explains how to operate and maintain the equipment, and provides
guidelines for obtaining the best bioluminescent or fluorescent images. Before
using the IVIS Spectrum, please read this manual carefully to obtain safe,
optimum performance and a maximum service life from the unit. For
®
instructions on using the system software, please see the Living Image
Software User’s Manual.
If you have questions regarding this manual or the IVIS Imaging System, please
call IVIS Imaging Systems Technology Support.
When you see the caution, warning, or voltage symbol, pay particular attention
to the safety information presented. Observance of safety precautions will help
avoid actions that could damage or adversely affect the performance of the IVIS
Spectrum. If the equipment is used in a manner not specified in this manual, the
protection provided by the equipment may be impaired.
CAUTION! This symbol indicates a potentially hazardous situation
which, if not avoided, may result in minor or moderate injury and/or
mechanical damage. It is also used to alert against unsafe practices. It
reminds you that all safety instructions should be read and understood
before installation, operation, maintenance, or repair of this instrument.
WARNING! Used when an action or condition may potentially cause serious
personal injury or loss of life. Mechanical damage may also result.
VOLTAGE! Indicates high voltage or risk of electric shock.
5
Page 12
2. Important Safety Instructions
2.2 Instrument Labels
Pay careful attention to the labels on the instrument.
LabelDescription
The Caliper IVIS
The instrument itself is classified as a Class 1 device because the user is safely
protected from the laser beam. The potential hazard of exposure to service
personnel is clearly indicated with a warning label per IEC 60825 and CDRH
requirements.
Laser caution label
Note: The actual warning
label mounted on the laser
module at the service
access inside the unit uses
non-fluorescing colors and
materials.
Where located on the IVIS system, the read manual label indicates that important
information about the labeled area can be found in the IVIS Spectrum Hardware
Manual.
Read manual label
Where located on the IVIS system, the pinch hazard label indicates the location of
a potential pinch point.
®
Spectrum Imaging Instrument contains a Class 3R Laser module.
Pinch hazard label
2.3 Instructions
Read InstructionsRead and understand all the safety and operating instructions before you install,
operate, or perform maintenance on this product. Make sure that you fully
understand the following safety instructions, warnings, and disclaimers before
proceeding to the rest of the manual.
Retain InstructionsRetain the safety and operating instructions for future reference.
Follow InstructionsFollow all operating and handling instructions. Failure to follow operating or
handling instructions may void any warranty covering this product.
Heed WarningsAbide by all warnings on the product and in the operating instructions. Failure
to adhere to warnings or safety precautions may void any warranty covering this
product.
6
Page 13
®
WARNING
!
WARNING
!
Spectrum Hardware Manual
IVIS
2.4 Laser Safety
The IVIS Spectrum is equipped with Class 3R laser (532 nm (green) laser
source, 5 mW maximum continuous (CW), which is suitably embedded and
limited in power to less than 1mW where accessible in the imaging
compartment behind the front door.
WARNING! A laser beam can be harmful to the eyes. ABSOLUTELY DO
NOT look directly into the laser beam when the laser is turned on. Be aware
that the laser light can be easily reflected off of shiny surfaces and that the
reflected light can be dangerous. When the laser is on, remove all watches
and jewelry before placing hands in the imaging chamber.
2.5 Fluorescence Light Source Safety
WARNING! The IVIS Spectrum is equipped with a strong bottom illumination
fluorescence light source. Do not stare directly into this source.
2.6 Environmental Considerations for the System Components
Location for the IVIS®
Spectrum
HeatThe system should be situated away from heat sources such as open flames,
Before the IVIS® Spectrum is installed, consider the proper environment for the
components.
Install the equipment in an environment where:
The temperature does not change or fluctuate widely. Choose an
•
environment where the temperature is maintained between 15-25° C (5978° F).
The humidity does not exceed 80%.
•
No strong electric or magnetic fields exist.
•
No vibrations are present.
•
No corrosive gases are present.
•
High amounts of dust are not present.
•
No open flame is present.
•
There is sufficient space behind the IVIS Spectrum equipment. A
•
minimum space of six inches from the flat surface of the rear panel should
be provided behind the IVIS Spectrum to provide unobstructed air flow
and access to the main power on/off switch.
The floor is level and structurally sound.
•
radiators, heat registers, stoves, and other heat-generating electrical equipment.
7
Page 14
Water & Moisture
VOLTAGE
VOLTAGE
2. Important Safety Instructions
VOLTAGE! Do not use this product near water (for example, near a sink or
wet room) due to risk of electric shock, electrical damage, and/or failure.
Laboratory Space &
Ventilation Requirements
The IVIS® Spectrum requires a minimum of 100 square feet (9 square meters)
of laboratory space for adequate ventilation. At least of six inches of space is
required behind the rear flat panel to provide unobstructed air flow.
Allocate an area at least 36 inches wide and 51 inches deep for the imaging
system console to allow the door to be fully opened and provide the
recommended air space behind the instrument.
An additional area that is 48 inches wide should be provided (preferably on the
right hand side) for a table or cart to hold the computer and provide a work
surface.
2.7 Power Considerations
Power SourcesThe IVIS
installation locality that was specified at the time of order. If the system is
moved to another area, check to make sure that the same voltage requirements
exist.
An IVIS Spectrum that operates on 120VAC requires a dedicated 20 Amp
circuit that is not shared with other loads except the computer and the optional
XGI-8 Gas Anesthesia System. Make sure that all equipment is plugged into a
properly grounded AC supply.
The computer should share the same circuit to avoid ground loops. Never use
auxiliary power outlets on other equipment to supply the IVIS Spectrum. Since
the system contains internal surge protection, it must be powered directly from
the main electrical supply wall socket. Use the surge protector provided by
Caliper for the computer and other accessories.
An IVIS Spectrum configured for 230 VAC requires a dedicated minimum 10
Amp circuit subject to the same conditions noted above. For more details on the
operating requirements, see Electrical Power Requirements, page 21.
®
Spectrum is configured for the voltage requirements of the
VOLTAGE! The IVIS Spectrum can operate at multiple voltages (100, 120,
220, 240 VAC); however, you are not permitted to change the input voltage
to any of the system components. Several internal modifications are required
for voltage change. If the operating voltage must be changed, contact IVIS
Imaging Systems Technology Support.
Power Cord ProtectionPower supply cords should be routed so that they are unlikely to be walked on
or pinched by items placed upon or against them. Pay close attention to
receptacles and to points of connection between cords and equipment.
Lightning & Power Line
Surges
The IVIS® Spectrum is supplied with an internal surge protector that is not
customer serviceable. Contact IVIS Imaging Systems Technology Support
regarding equipment servicing.
8
Page 15
®
WARNING
!
VOLTAGE
Spectrum Hardware Manual
IVIS
A separate surge protector is provided for the computer and other accessory
equipment.
Facilities should be adequately wired according to local building codes.
Power OutagesIf the IVIS Imaging System experiences a loss of supply power, turn off the
power switch for all components and do not restart the system until reliable
power has been restored. For more details on how to restart the system, see
Startup Procedure, page 36.
Overloading
WARNING! Do not overload wall outlets, extension cords, or integral
convenience receptacles as this can result in a risk of fire or electric shock.
See Electrical Power Requirements, page 21 for requirements.
2.8 Cleaning or Moving the System Components
Cleaning/Liquid Entry
VOLTAGE! Do not use liquid or aerosol cleaners and never spill liquid of any
kind on any of the IVIS
into contact with the IVIS System may result in damage to the system or
electrocution. For more details on proper care of the system, see Cleaning the
IVIS Spectrum, page 51.
Moving the IVIS® SpectrumThe IVIS
®
Spectrum is a sensitive, scientific instrument and should be moved
®
Spectrum components. Sprays and liquids that come
with care. Pay particular attention when rolling the instrument on its casters to
avoid toppling the equipment.
The total height of the instrument is 83 inches (211cm). If necessary, the plastic
dome can be removed for easier passage through doorways. For more details on
removing the dome, see page 55.
If you have any questions about moving the instrument, contact IVIS Imaging
Systems Technology Support.
2.9 Other Equipment, Chemicals, or Supplies
Use of any equipment other than that recommended by this manual has not been
evaluated for safety and, therefore, is the sole responsibility of the user.
No chemicals are required for the operation of the IVIS Spectrum. Other usersupplied chemicals or materials may be required as part of your specific
biological testing procedures.
9
Page 16
2.10 Servicing
2. Important Safety Instructions
Refer all servicing to IVIS Imaging Systems Technology Support. If the product
is damaged and requires service, unplug the product from the outlet and contact
technical support. Servicing by anyone other than those authorized by Caliper
voids any warranty covering this product.
WARNING! DO NOT attempt to service the IVIS® Spectrum yourself. Contact
IVIS Imaging Systems Technology Support for electrical service needs.
Although there are no voltages in excess of 45V inside the imaging chamber,
local line voltages can be present inside the electronics cabinets.
CAUTION! If cleaning is necessary, wipe the exterior surfaces of the
IVIS Spectrum with a soft, damp cloth only. Do not allow fluids of any
kind to enter the system interior under any circumstances. For more
details on cleaning the system, see page 51.
11
Page 18
3. Warnings
CAUTION
!
Note: The actual warning label
mounted on the laser module at
the service access inside the unit
uses non-fluorescing colors and
materials
3.2 Burn Hazard & Eye Safety
Laser
The IVIS Spectrum Imaging Instrument contains an embedded Class 3R Laser
module: 532 nm (visible green) laser source, 5 mW maximum continuous
(CW). The instrument itself is classified as a Class 1 device because the user is
safely protected from the laser beam. The potential hazard of exposure to
service personnel is clearly indicated with a warning label per IEC 60825 and
CDRH requirements.
CAUTION! The use of controls or adjustments or performance of
procedures other than those specified herein may result in hazardous
radiation exposure.
NEVER remove back, side, front or internal panels of the instrument while the
laser is powered on. Note that laser light can be easily reflected off of shiny
surfaces. Remove all watches and jewelry before placing hands in the imaging
chamber.
Complies with FDA performance standards for laser products except for
deviations pursuant to Laser Notice No. 50, dated 6/24/2007.
Complies with IEC/EN 60825-1:2007.
Light Source ModuleThe IVIS Spectrum is equipped with a fluorescence light source module and
connecting fiber optic cables that are capable of producing intense light that
could be damaging to the eyes. Additionally, the system is equipped with a
Class 3R laser, which could be an eye hazard.
Figure 3.1 Laser beams for alignment grid
3.3 Mechanical Safety
The IVIS Spectrum has many internal motorized components that can move
only when the door is closed and locked. Because the imaging platform moves
frequently during routine use, keep loose objects away from the edges of the
platform where they could become jammed.
12
Page 19
®
CAUTION
!
CAUTION
!
Spectrum Hardware Manual
IVIS
Figure 3.2 IVIS Spectrum, imaging chamber door
CAUTION! DO NOT defeat any of the safety interlocks. Do not place
anything under the imaging platform.
The imaging platform cannot be moved unless the door is closed and locked. If
the platform moves when the door is unlocked, shut down the system and
contact IVIS Imaging Systems Technology Support.
The IVIS Spectrum contains several delicate optical and mechanical
components. Do not touch these components and avoid sprays or other
contaminates that may damage lenses or other optical-mechanical components.
CAUTION! The laser/galvanometer is a very delicate optical
component that is visible when the filter wheel side access panel is
removed
adjust it.
(Figure 3.3). DO NOT touch this device or attempt to clean or
13
Page 20
3. Warnings
WARNING
!
Filter wheel panel
Figure 3.3 Filter wheel removable panel
The laser beam is directed into the imaging chamber through an opening at the
top of the chamber. Never place any object, tool, or fingers into this opening,
otherwise damage to the galvanometer will result.
WARNING! DO NOT look directly into this opening when the laser is
operating.
14
Page 21
®
Spectrum Hardware Manual
IVIS
3.4 Chemical & Biological Safety
Normal operation may involve the use of test samples that are pathogenic, toxic,
or radioactive. It is your responsibility to ensure that all necessary safety
precautions are taken before such materials are used.
Dispose of all waste materials according to appropriate environmental health
and safety guidelines.
It is your responsibility to decontaminate the IVIS
service by IVIS Imaging Systems Technology Support. Ask your laboratory
safety officer to advise you about the level of containment required for your
application and about the proper decontamination or sterilization procedures to
follow.
Handle all infectious samples according to good laboratory procedures and
methods to prevent the spread of disease.
3.5 Panels, Covers & Modules
The imaging chamber door and the filter wheel access panel are the only user
removable panels. There are no user serviceable components in the electronics
cabinets of the IVIS
from the system console unless instructed by and under the supervision of IVIS
Imaging Systems Technology Support.
®
Spectrum before requesting
®
Spectrum. Do not remove the electronics cabinet covers
Caliper Life Sciences, Inc. ("Caliper") provides the following limited warranty
for each new IVIS
i. This Limited Warranty for the System extends to the original purchaser
("Customer") for a period of one (1) year following installation of the
System, and is not assignable or transferable to any successor.
ii. During the Limited Warranty, Caliper will repair or replace, at Caliper's
sole option, any defective parts if such repair or replacement is needed
because of System malfunction or failure to conform to published
specifications during normal usage in accordance with the instructions in
this manual. Repairs and replacements under the Limited Warranty will be
made at Caliper’s expense. Caliper’s limit of liability under the Limited
Warranty shall be the purchase price of the Imaging System. Caliper shall
not be liable for any other losses or damages. These remedies are the
Customer's exclusive remedies for breach of this Limited Warranty.
iii. No coverage or benefits shall be provided under this Limited Warranty if
any of the following conditions apply:
a. The System has been subjected to unauthorized modifications (e.g.
unauthorized installation of hardware or software), unauthorized
repair or servicing, misuse, neglect, abuse, accident, alteration, any use
inconsistent with or in contradiction to the instructions in this manual,
or other acts which are not the fault of Caliper.
b. Caliper was not advised in writing by the Customer of the alleged
defect or malfunction of the System within the earlier to occur of ten
(10) days after the expiration of the Limited Warranty period, or 15
days after becoming aware of the defect or malfunction.
c. If Customer moves the System from its installed location to another
location without Caliper's technical assistance, then any damage to the
System due to such movement shall not be covered under the initial
warranty or any extended warranty, and Licensee shall pay Caliper's
standard service rates for repair of such damage.
iv. If a problem develops during the Limited Warranty, the Customer shall
contact Caliper technical support for assistance.
v. THE FOREGOING LIMITED WARRANTY IS THE CUSTOMER'S
SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER
WARRANTIES, EXPRESS OR IMPLIED. CALIPER SHALL NOT BE
LIABLE FOR SPECIAL, INCIDENTAL OR CONSEQUENTIAL
DAMAGES, INCLUDING BUT NOT LIMITED TO LOSS OF
ANTICIPATED BENEFITS OR PROFITS, LOSS OF SAVINGS OR
REVENUE, PUNITIVE DAMAGES, LOSS OF USE OF THE SYSTEM
®
Spectrum CT ("System") as follows ("Limited Warranty"):
17
Page 24
4. Legal Notices
OR ANY ASSOCIATED EQUIPMENT, COST OF CAPITAL, COST OF
ANY SUBSTITUTE EQUIPMENT OR FACILITIES, DOWNTIME,
THE CLAIMS OF ANY THIRD PARTIES, INCLUDING
CUSTOMERS, AND INJURY TO PROPERTY, RESULTING FROM
THE PURCHASE OR USE OF THE SYSTEM OR ARISING FROM
BREACH OF THE WARRANTY, BREACH OF CONTRACT,
NEGLIGENCE, STRICT TORT, OR ANY OTHER LEGAL OR
EQUITABLE THEORY, EVEN IF CALIPER KNEW OF THE
LIKELIHOOD OF SUCH DAMAGES. CALIPER SHALL NOT BE
LIABLE FOR DELAY IN RENDERING SERVICE UNDER THE
LIMITED WARRANTY, OR LOSS OF USE DURING THE PERIOD
THAT THE SYSTEM IS BEING REPAIRED.
vi. Some countries, states or provinces do not allow the exclusion or limitation
of implied warranties or the limitation of incidental or consequential
damages for certain products or the limitation of liability for personal
injury, so the above limitations and exclusions may be limited in their
application to you. When any implied warranties are not allowed to be
excluded in their entirety, they will be limited to the duration of the
applicable written warranty. This Limited Warranty gives you specific
legal rights which may vary depending on local law.
vii. This Limited Warranty shall be governed by the laws of the State of
California, U.S.A., excluding its conflicts of laws principles and excluding
the United Nations Convention on Contracts for the International Sale of
Goods.
4.2 Patents
4.3 Trademarks
The detection and imaging of light originating within mammals is the subject of
several issued patents and pending patent applications in the United States and
around the world, including U.S. Patent Numbers 5,6200,135, 6,217,847,
6,649,143, 6,890,515, and European Patent Commission Number EP0861093,
for which Xenogen Corporation is the exclusive licensor. The use of an IVIS
®
Imaging System for such applications requires a sublicense from Xenogen
Corporation.
In addition, many of the hardware and software components of the Imaging
System are the subject of various issued patents and pending patent applications
owned by Xenogen, including: United States Patent Number 6,614,452
(Graphical User Interface for In Vivo Imaging) and 6,775,567 (Improved
Imaging Apparatus); and United States Patent Applications 09/905668 (Multiview Imaging Systems), 10/606976 (Method and Apparatus for 3-D
Reconstruction of Light Emitting Sources), 10/151463 (Method and Apparatus
for Determining Target Depth, Brightness, and Size Within a Body Region), 10/
189886 (Fluorescence illumination assembly for an imaging apparatus), and 10/
068573 (Light calibration device for use in low level light imaging systems).
IVIS and Living Image are registered trademarks or trademarks of Caliper Life
Sciences, Inc. The names of companies and products mentioned herein may be
the trademarks of their respective owners. Apple, Macintosh, and QuickTime
are registered trademarks of Apple Computer, Inc. Microsoft, PowerPoint and
Windows are either registered trademarks or trademarks of Microsoft
18
Page 25
®
Spectrum Hardware Manual
IVIS
Corporation in the United States and/or other countries. Adobe and Illustrator
are either registered trademarks or trademarks of Adobe Systems Incorporated
in the United States and/or other countries. Intel is a registered trademark of
Intel Corporation.
System of lenses and filters for optimum image acquisition.
•
Laser/galvanometer.
•
Photographic illumination system.
•
Integrated fluorescence system.
•
Imaging platform crash avoidance system.
•
These components are integrated into one freestanding instrument that is
equipped with casters
A pre-configured Windows
as well as acquire and analyze images.
(Figure 6.1).
®
-based computer enables you to control the system
ALERT! If you modify the IVIS Imaging System in any way, without prior
approval from Caliper, all warranties that cover this product are void. In
addition, the computer included with the IVIS Spectrum is specifically
configured to run all system-related applications. Any modification of existing
software or hardware voids all warranties.
If you have any questions, please contact IVIS Imaging Systems Technology
Support.
Compartment that
contains the ThermoCube
chiller unit and camera
power supply
Emission filter wheel
Excitation filter wheel
Top illumination
light source
Heated platform
Integrated fluorescent
imaging features:
Bottom illumination
light source
Figure 6.1 IVIS Spectrum components
The system is housed in a portable cart with 25 x 34” footprint.
6.1 CCD Camera
The camera is a scientific grade, thermoelectrically-cooled, back-thinned, backilluminated, large format CCD manufactured by Spectral Instruments (Figure
6.1). The camera uses a ThermoCube water chiller. Two flexible plastic water
lines connect the CCD camera to the ThermoCube chiller unit.
The CCD camera has a heated, vacuum sealing window to prevent condensation
that can obscure the image or damage other components.
The camera power supply is housed in the base of the instrument and maintains
the camera operating temperature when the computer is turned off. It has a 16bit digitized, low noise electronic readout for extremely low background
images.
6.2 ThermoCube Chiller Unit
The ThermoCube unit is a thermoelectric water chiller (Figure 6.2). Water from
the thermoelectrically cooled CCD camera is pumped to the ThermoCube unit
where the water is cooled and returned to the camera.
26
Page 33
®
WARNING
!
Spectrum Hardware Manual
IVIS
6.3 Imaging Chamber
Figure 6.2 ThermoCube chiller unit
The imaging chamber (Figure 6.3) is a highly specialized device consisting of the
imaging chamber housing, a heated, moveable platform, an auto focusing lens
system with f/stop control, and synchronized filter wheels that control the
spectral content of the luminescent or fluorescent images.
The LED-based illumination system and a laser-based galvanometer provide
the means to acquire photographic information and structured light before
bioluminescent or fluorescent imaging. The imaging chamber is equipped with
gas anesthesia inlet and outlet ports for use with the optional XGI-8 Gas
Anesthesia System.
The imaging chamber is light tight, so that no light penetrates from the outside
after the magnetically activated door is locked. The magnetic latch ensures the
door cannot be inadvertently opened during an imaging session.
The interior of the imaging chamber is constructed from materials that are nonphosphorescent and non-fluorescent to prevent internal light contamination that
could compromise sample measurements.
WARNING! Under no circumstances should you attempt to make any
mechanical modifications to the imaging chamber.
27
Page 34
6.4 Optical Components
NOTICE
Door locking hardware
1.5X
3.9 cm
A
2.5X
6.5 cm
B
5.0X
13 cm
C
8.8X
22.5 cm
D
Power
FOV
Position
6. IVIS® Spectrum Components
Figure 6.3 IVIS® Spectrum, imaging chamber
Imaging System/LensesThe IVIS
demagnification of 1.5X, 2.5X, 5X, and 8.8X power (Figure 6.4).The filter wheel
settings are selected in the Living Image® software. For more details, see the
Living Image Software Manual.
Figure 6.4 Example field of view obtained with different lens configurations
The automatic shutter and f/stop iris is integrated into the lens system. A
window at the top of the imaging chamber protects the lenses and other optical
components in the lens compartment from contamination. For more details on
the care of this window, see Cleaning the Lens Protection Window, page 52.
Do not touch the glass window or permit specimens to come in contact with
the window, otherwise image quality may be impaired. If the window is struck
by a hard object, it may crack or shatter.
®
Spectrum is equipped with a rotating lens carousel that provides
28
Page 35
®
Temperature controlled imaging platform
Spectrum Hardware Manual
IVIS
Emission & Excitation
Filter Wheels
There are two 11-position filter wheels. Filter settings are selected by software
control. For more details on how to access the filter wheels and change filters,
see page 53.
Figure 6.5 Filter wheel
®
For fluorescence studies, the IVIS
Spectrum Imaging System can illuminate a
specimen from the top of the imaging chamber or from the bottom of the stage.
A 12-position excitation filter wheel module is located in the back of the
instrument (not accessible to users). If the filters must be changed or the module
requires service, contact IVIS Imaging Systems Technology Support.
For more details on fluorescence imaging, see page 41.
6.5 Specimen Warming System
The temperature-controlled imaging platform keeps specimens warm during
imaging. The temperature is controlled by the software, and is only enabled
when the instrument is initialized from Living Image
temperature upon initializing the instrument is 37° C.
The imaging platform does not have active cooling. Allowing the platform to
cool from 37° C to ambient temperature may require up to 20 minutes.
Figure 6.6 Imaging chamber
®
software. The default
29
Page 36
6. IVIS® Spectrum Components
WARNING
!
IMPORTANT
!
6.6 Laser/Galvanometer System
The laser/galvanometer system comprises a Class 3R laser (532 nm (visible
green) laser source, 5 mW maximum continuous (CW). The laser beam is
directed into the imaging chamber by two computer-controlled mirrors.
WARNING! A laser beam can be harmful to the eyes. ABSOLUTELY DO
NOT look directly into the laser beam when the laser is turned on. Be aware
that laser light can be easily reflected off of shiny surfaces and that this
reflected light can be dangerous. When the laser is on, remove all watches
and jewelry when placing hands in the imaging chamber.
The laser/galvanometer system provides field of view alignment targets for
reproducible specimen positioning and structured light for 3D contouring and
auto-focusing.
6.7 Crash Avoidance System
The crash avoidance system uses a scanning laser to determine the subject
height and verify that it is within acceptable limits. This system prevents injury
to specimens and damage to delicate optical components such as the lens
window or the illumination diffusers for field of view A only.
You are prompted at the start of each imaging session to input a sample height
estimate. If the estimate is in error, the crash avoidance system protects the
specimen. Alternatively, the system can be configured to automatically
determine the sample height.
6.8 Base Enclosure
30
ALERT! Do not block the ventilation openings on the base or position the
system too close to a wall so that the exhaust flow is obstructed.
The base is equipped with four casters that enable you to move the instrument
(Figure 6.7). For more details on how to move the IVIS Spectrum, see page 55.
Page 37
®
Ventilation openings
Ventilation openings
Spectrum Hardware Manual
IVIS
Figure 6.7 IVIS Spectrum, back view
6.9 Acquisition Computer
6.10 Accessory Cable
The system includes a Windows®-based personal computer. The Living Image®
software as well as Microsoft Office software is installed on the computer. The
Living Image software controls the IVIS Spectrum, and displays and analyzes
the image data. A 20-inch monitor and a 10/100 Ethernet network adapter card
are also included. A printer may be attached to the computer.
The IVIS Spectrum is equipped with an electrical connector that enables the
transfer of signals to or from the back panel of the instrument. The back panel
connector
connector is located inside the imaging chamber
(Figure 6.8) is accessible at the rear of the instrument and the stage
(Figure 6.9). Table 6.1 shows
guidelines for pin usage.
Table 6.1 Connector pin usage
Pin NumberFunction
1 to 5These pins are optimized for transport of ECG (EKG) signals and use
6 to 9Use these pins to connect a pulse-oxymetry system (for example, the
10 and 11Use these pins to connect to an Infusion pump that you mount inside
shielded cabling that maximizes the signal-to-noise ratio. An off-theshelf ECG monitor may be connected to this connector (for example,
the THM100, Indus Instruments). Pin 1 (ECG ground) in this particular
case, may be left unconnected.
Nonin OEM III module (internal OEM Oxymetry), Nonin.com)
the IVIS Spectrum imaging chamber.
31
Page 38
6. IVIS® Spectrum Components
Note: A 15 pin D-sub male mating connector is required (supplied by user).
Note: Connector type in shelf--MS3119E16-26PS. Mating connector required (supplied by
customer). Manufacturer: Amphenol Aerospace, part no. MS3126F16-26P.
Table 6.1 Connector pin usage
Pin NumberFunction
12 and 13Use these pins to monitor the skin or rectal temperature of the animal.
The use of a copper-constantan thermocouple is recommended.
These pins may also be used to connect to the THM 100 ECG monitor
(Indus Instruments).
Figure 6.8 Back panel connector
Figure 6.9 Stage connector
32
Page 39
®
Spectrum Hardware Manual
IVIS
6.11 High Reflectance Hemisphere
You can special order a high reflectance hemisphere that is used to check for
light contamination in the imaging chamber
The hemisphere does not emit any photons when it is imaged in a light-tight and
contaminant-free imaging chamber. As a result, it is not visible in a luminescent
image. If a light leak or an internal phosphorescent contaminant is present in the
imaging chamber, the hemisphere reflects that light and is visible in a
luminescent image. If you suspect that there is light contamination in the
imaging chamber, see Light contamination-External, page 47.
The IVIS Spectrum is intended for use in biophotonic or fluorescent imaging
procedures. The system is designed to detect extremely low-level light
emissions that are orders of magnitude dimmer than can be detected by the
naked eye. The IVIS Spectrum allows you to monitor and record cellular and
genetic activity within a living organism, in real time. The imaging system
captures, quantifies, and images the light emitted by a sample.
7.1 Restarting the System
Electrical Control & Status
Lights
All components of the IVIS Spectrum should be left on unless the system will
not be used for more than 30 days. It is also important to leave the system on
to enable automatic overnight electronic background measurements.
Periodically rebooting the computer is permissible and does not affect the
camera operation.
The IVIS Spectrum console has two electrical switches and one status light
(Figure 7.1). The color of the status light indicates the state of the instrument
(Table 7.1). The main power on/off switch (circuit breaker) is located on the rear
panel and controls the AC power to the instrument.
The front panel contains a recessed toggle switch that distributes power within
the system console. This switch can be in the standby position (down) or on
position (up) and is equipped with a locking mechanism to prevent accidental
activation. The shaft of the switch must be pulled out to allow movement up or
down. The gas control valve is to the left of the toggle switch. For more details
on the gas control valve, see Gas Plumbing, page 37.
CAUTION! Placing the toggle switch in standby mode turns off the
ThermoCube cooling unit and camera power supply. All low voltage
equipment remains on.
35
Page 42
7. Operating the IVIS® Spectrum
Status light
Gas valve
Standby toggle switch
Bottom illumination
input optics
Figure 7.1 IVIS Spectrum
Table 7.1 IVIS Spectrum status light modes
Status Light Is...Indicates...
BlueThe system is in standby mode.
Blue and blinkingThe CCD camera cooling system is disabled due to a power
RedThe system is initializing or acquiring an image.
GreenThe system is ready.
interruption. Use the Living Image software to initialize the
system and enable the cooling system.
Startup ProcedureThe same procedure is used to restart the system whether the shut down was
intentional or due to a power outage.
1. Confirm that the rear panel switch is off, the front panel toggle switch is in
the down position (in standby), and the power cord is plugged into the
console and the wall socket.
2. Turn on the computer and monitor, and start the Living Image
®
software.
3. Turn on the main power switch on the rear panel.
— The status light is blue.
4. Turn on the front panel toggle switch by pulling the toggle switch out and
up.
— The status light is blue and blinking.
36
Page 43
®
WARNING
!
The temperature square is red at startup and turns green
when the operating temperature is reached. Click the
temperature square to display the demand and measured
temperature of the CCD and stage.
Temperature box color indicates:
System not initialized.
System is initialized, but CCD temperature is out of
range.
System is initialized and CCD is at or within
acceptable range of the demand temperature (-90° C)
and locked. The system is ready for imaging.
Spectrum Hardware Manual
IVIS
5. In the Living Image software, click Initialize IVIS system in the IVIS
System Control panel
(Figure 7.2).
— The status light is red.
6. Wait until the camera has cooled to operating temperature (requires about
ten minutes).
— The status light is green when the system is ready for operation.
You can monitor the system temperature in the Living Image software
(Figure 7.2).
7.2 Gas Plumbing
Figure 7.2 IVIS System control panel in the Living Image software
The instrument is ready for operation when the camera temperature is locked at -90
C, as indicated by the green temperature square in the IVIS System Control panel.
For operating instructions, see the Living Image Software Manual.
WARNING! Use only isoflurane with the IVIS® Spectrum. DO NOT USE
FLAMMABLE ANESTHESIA GAS.
37
Page 44
7. Operating the IVIS® Spectrum
CAUTION
!
NOTICE
CAUTION! Caliper recommends using the XGI-8 Gas Anesthesia System
(Figure 7.3) when imaging small animals. This system supplies a controlled
amount of isoflurane to the imaging chamber and continuously reduces the
build-up of isoflurane in the chamber. If you want to use a gas other than the
recommended isoflurane/oxygen gas mixture or pure air, contact IVIS
Imaging Systems Technology Support.
38
Figure 7.3 XGI-8 Gas Anesthesia System
Be careful to use only tubing and other plumbing fixtures that do not fluoresce
or phosphoresce (glow) in the imaging chamber. Contact IVIS Imaging
Systems Technology Support for a list of acceptable materials.
The IVIS Spectrum is equipped with plumbing connections for supplying
isoflurane anesthetic gas to the imaging chamber and exhausting gases out. The
flow of gas into the imaging chamber is manually controlled by the gas valve
on the front panel of the console. The imaging chamber does not include a gas
scavenging system. This is provided by suitable auxiliary equipment such as the
XGI-8 Gas Anesthesia System.
Gas ports are located on the rear of the system console and are labeled GAS IN
and GAS OUT. GAS IN means the direction of flow is into that port.
Similarly, the port labeled GAS OUT means that flow can be exhausted out
of this port. The flow of gas from the rear panel GAS IN port is directed to the
main gas control valve on the front of the system and proceeds to a distribution
block located on the imaging platform where it exits through the port labeled
GAS.
Page 45
®
NOTICE
WARNING
!
WARNING
!
Spectrum Hardware Manual
IVIS
7.3 Door Operation
The mouse anesthesia manifold (part no. 117706) with transparent nose cones
is provided with the IVIS
®
Spectrum. The manifold has capacity for up to five
mice and has a low profile (height) for operation with the system.
The IVIS Imaging System 100 Series and the XGI-8 Gas Anesthesia System
are equipped with a different five-station manifold (PN 118274) that is taller
than the low profile manifold supplied with the IVIS Spectrum. Only use the
low profile manifold with the IVIS Spectrum, which permits imaging the
smallest field of view at the highest platform height (FOV A).
The manifold port labeled GAS connects to the GAS port on the platform
distribution block. The anesthesia manifold also has a port labeled EXH that
should be connected to the EXH port on the platform distribution block. The
vacuum or exhaust circuit connects to the GAS OUT port on the rear of the
system console. No gases are exhausted from the imaging chamber unless gas
scavenging equipment such as the XGI-8 Gas Anesthesia System is used.
The door of the imaging chamber cannot be opened when the system status light
is red. The door is equipped with an electromagnetic lock to prevent accidental
opening during an imaging session.
WARNING! Never try to defeat the door interlock or force the door open
during image acquisition when the status light is red. Such action could
expose the user to a severe pinch hazard from the rapidly moving imaging
platform.
7.4 Centering a Subject in the Field of View
WARNING! The IVIS Spectrum is equipped with a laser similar to the type
used for laser pointers (532nm, output < 1mW). Users are protected from eye
damage due to the eye’s physiological response to bright light by blinking. A
laser beam can be harmful to the eyes. ABSOLUTELY DO NOT look directly
into the laser beam when the laser is turned on. Be aware that laser light can
be easily reflected off of shiny surfaces and that this reflected light can be
dangerous. When the laser is on, remove all watches and jewelry when
placing hands in the imaging chamber.
1. In the Living Image software, choose the Enable alignment grid option in
the IVIS System Control panel (Figure 7.4).
39
Page 46
7. Operating the IVIS® Spectrum
NOTICE
NOTICE
To display the alignment
grid, choose Alignment Grid On.
Figure 7.4 IVIS System Control panel
2. Open the imaging chamber door to see the alignment grid projected by the
laser/galvanometer on the imaging platform
(Figure 7.5).
— The field of view (FOV) setting defines the size of the squares in the
alignment grid.
3. To change the size of the squares in the alignment grid, select a different
FOV setting.
For example FOV settings, see Figure 6.4, page 28.
40
Figure 7.5 Laser alignment grid
The alignment laser turns off after 60 seconds. Repeat step 1 to turn on the
laser.
4. Using the alignment grid as a guide, place the specimen(s) at the locations
of interest, then proceed with image acquisition.
The Living Image® software controls the photographic illumination LED
intensity.
Page 47
®
Choose
luminescent or
fluorescent
imaging
Spectrum Hardware Manual
IVIS
7.5 Fluorescent Imaging
The IVIS® Spectrum has both bioluminescent and fluorescent imaging
capabilities. To change from bioluminescent to fluorescent operation, select
fluorescence mode in the IVIS System Control panel and select an appropriate
excitation and emission filter
Figure 7.6 IVIS System Control panel
(Figure 7.6).
A 150 Watt quartz tungsten halogen lamp with a dichroic reflector provides the
®
fluorescence excitation light. The Living Image
software controls the lamp
intensity level. The fluorescence light source is a module located behind the rear
panel and is not user serviceable.
The lamp output is delivered to the excitation filter wheel assembly located in
the rear compartment, where it is collimated and transmitted through a
fluorescence excitation filter. The 12 excitation filter wheel locations provide
10 fluorescence filters. Two filter slots contain a light block that is used during
bioluminescent imaging to prevent external light from entering the imaging
chamber. The 12-position excitation filter wheel is motorized and controlled
through the Living Image software.
Upon leaving the excitation filter, the light enters an optical switch that directs
the light to either the top of the imaging chamber or to the bottom of the stage.
When top illumination is selected, the light is equally split to four diffusers to
provide uniform stage illumination. When bottom or trans-illumination is
selected, the excitation light is directed to one of the 2 mm diameter holes in the
aperture plate located in the stage surface. This removable plate has a 15 x 23
array of 2mm diameter holes which provide discrete illumination locations that
are user selectable.
You can replace the array plate with a 3 mm thick glass plate or a well plate
holder to provide alternate illumination possibilities. A blank plate is also
provided for when the bottom illumination equipment is not in use. For more
details on how to use the top and bottom illumination systems, see the Living Image Software User’s Manual.
Selecting an Emission
Filter
The IVIS Spectrum has a factory-installed filter set. The filters are mounted on
a rotating filter wheel and can be removed or replaced with filters with other
spectral characteristics. For more details, see Emission & Excitation Filter
Wheels, page 29. For information on how to replace a filter, see page 53. See
®
the Living Image
Software Manual for more details on how to select an
emission filter.
41
Page 48
7. Operating the IVIS® Spectrum
Selecting an Excitation
Filter
In fluorescence mode, the excitation filter is automatically selected based on the
wavelength of the emission (imaging) filter selected from the IVIS System
Control panel. However, you can override the automatic selection in the IVIS
System Control panel.
The excitation filters are housed in a filter wheel module that is not accessible
to the user. Unlike the emission filters, the user cannot replace these filters.
Contact IVIS Imaging Systems Technology Support if an excitation filter must
be replaced.
7.6 Imaging Basics
Black PaperAlthough the platform is black anodized, it is recommended that you image
items on a high quality black paper, especially biological specimens. Caliper
Corporation has surveyed many types of paper and recommends Swarthmore,
Artagain, Black, part no. 445-109, size 8.5 inch x 11 inch. This paper prevents
illumination reflections and helps keep the stage clean.
Low Fluorescence MatWhen operating the system in the fluorescence mode, use the Low Fluorescence
Imaging Mat (part no. 117839) to reduce background fluorescence. The mat is
made from a low reflectance textured plastic that is easily cleaned.
Glowing MaterialsAlways keep in mind that nearly EVERYTHING glows (that is, has the
potential to phosphoresce and contaminate the image). Most plastics, almost all
tape, plants, paint, rodent food (mostly plants), mouse urine, and animal
bedding have been found to glow.
®
Use caution when introducing materials into the IVIS
to prescreen all items by imaging them alone, before imaging them with
samples under study. Caliper recommends using non-powdered gloves when
working with IVIS Spectrum equipment.
Spectrum. It is advisable
7.7 System Shut Down Procedure
Caliper Corporation does not recommend power cycling the IVIS Spectrum
(turning the system components on and off) on a daily basis. Leaving the system
on keeps the camera cold and the electronics systems stable. The software is
preconfigured to run automatic background accumulation and self-diagnostics
overnight. The platform temperature and the computer monitor may be turned
off.
If it is necessary to shut down the system for any reason, such as long term
storage (more than 30 days), it is important to follow the procedure below.
1. Close the Living Image
prompt.
2. Turn off the computer.
3. Turn off any gas to the imaging chamber and disconnect the gas supply if
necessary. If you are using the XGI-8 Gas Anesthesia system, follow the
shut down procedure in the XGI-8 Anesthesia System User Manual.
4. Turn off (standby) the toggle switch on the front of the IVIS Spectrum.
42
®
software and save any information of interest at the
Page 49
®
Spectrum Hardware Manual
IVIS
5. Turn off the main circuit breaker switch located on the rear of the IVIS
Spectrum and disconnect the power cord from the wall socket.
6. Cover the gas inlet and outlet ports with masking tape to prevent entry of
dust.
7. Remove any loose objects from the imaging chamber and close the door.
If you plan to move the system, see important information on page 55.
If you have any problems during the shut down procedure, please contact IVIS
Imaging Systems Technology Support for assistance.
43
Page 50
7. Operating the IVIS® Spectrum
[This page intentionally blank.]
44
Page 51
®
Click the temperature square to display
the demand and measured temperatures
of the camera and imaging chamber.
Spectrum Hardware Manual
IVIS
Troubleshooting
8
Measured Temperature Does Not Equal the Demand Temperature . 45
Accessing the Area Under the Imaging Platform . . . . . . . . . . . . 50
8.1 Measured Temperature Does Not Equal the Demand Temperature
At start up, the Living Image® software programs the CCD camera to maintain
the CCD at -90° C. If the camera power supply and ThermoCube chiller unit
remain on, the system maintains this temperature regardless of whether the
Living Image software is open or the computer is turned on.
■ To check the temperature of the CCD, click the Temperature square (red or
green) in the Camera Control Panel of the Living Image software (Figure
.
8.1)
Figure 8.1 IVIS System control panel in the Living Image software
ProblemPossible CauseCorrective Action
Measured temperature is
warmer than the demand
temperature.
A problem may exist with the
camera or the ThermoCube unit.
Contact IVIS Imaging Systems Technology
Support.
45
Page 52
8. Troubleshooting
8.2 Photographic Image Is Unacceptable
Default camera controls are programmed during the initialization of the Living
®
Image
image. Check that the exposure time, f/stop and binning levels are set to the
default conditions.
If the f/stop is open too wide, the image may be saturated and less focused.
Binning to a higher value or increasing the exposure time can produce an image
that is too bright and lacks resolution. If you have questions about the
photographic settings, please contact IVIS Imaging Systems Technology
Support.
Also, refer to the Living Image Software Manual for further details on acquiring
images.
ProblemPossible CauseCorrective Action
Image is streaked.Subject moved during the
exposure.
Image is blurry.Subject height is not correctly
entered in the Living Image
software.
software. Changes to these settings can greatly affect the photographic
Check to see if the subject may have moved. If the
subject is not on the sample stage, it is probably on the
floor of the imaging chamber. If the sample has moved,
locate and reanesthetize it. If gas anesthesia is being
used, confirm that the anesthesia is turned on and the
flow rate is appropriate.
If you are using the “Sample Height”setting to determine
the focus, confirm that the correct height of the sample
is entered. Alternatively, Auto-Focus may be used to
determine the sample height.
If you are using the “Manual”setting to determine the
focus, confirm that the focus has been manually
adjusted.
A white spot appears in the
center of the field of view.
Incorrect f/stop setting.The f/stop for photographs should be set to f/8. An f/stop
An excessively moist
environment in the imaging
chamber can result in
condensation on the CCD
window (Figure 8.2).
Figure 8.2 Condensation example
smaller than 8 reduces the depth of field in the
photograph.
Remove excess moisture in the imaging chamber and
allow it to thoroughly dry. Place a desiccant in the imaging
chamber to decrease the drying time.
Do not touch or let specimens touch the window at the
top of the imaging chamber. For more details on the care
of this window, see page page 52.
If the problem persists, contact IVIS Imaging Systems
Technology Support for assistance.
46
Page 53
®
Spectrum Hardware Manual
IVIS
8.3 Luminescent Image Is Unacceptable
Binning, f/stop, and exposure time affect the appearance of a luminescent
®
image. Please refer to the Living Image
Software Manual for instructions on
setting binning, exposure time, and f/stop values.
In order to function properly and reduce camera noise, the CCD camera must
be cooled to the demand temperature before acquiring an image. If the camera
is not cooled to the demand temperature, imaging may result in false positive
signals.
ProblemCorrective Action
Light contaminationInternal
Light contaminationExternal
Camera noiseVerify that the camera is cooled to the demand temperature.
Check to see that there are no extraneous light sources in the imaging chamber. Many
substances phosphoresce when exposed to light. Be especially cautious of plastics and
substances that contain pigment. Be sure to pre-screen any substance or material before
performing actual experiments.
A 2” diameter high reflectance hemisphere can be special ordered to help check for light leaks
(XRH-1, part number 118937, Figure 8.3).
To check for light leaks:
1. Place the high reflectance hemisphere in the imaging chamber on the stage using a subject
height of 3.5 cm at field of view D.
2. Take a luminescent image of the hemisphere using the luminescent settings: f/stop = 1,
Binning = Large (high sensitivity), and exposure time = 5 minutes.
If the hemisphere can be easily seen, there is a light leak. Contact IVIS Imaging Systems
Technology Support for assistance.
1. Check the measured temperature in the Camera Control panel to ensure that it is locked. If
the camera temperature is locked, the camera temperature box is green.
2. If the camera temperature box is red, click the red box to display the actual temperature. See
Measured temperature is warmer than the demand temperature., page 45.
Figure 8.3 High reflectance hemisphere
Wear powderless gloves when handling the hemisphere to prevent surface
contamination that may glow when using the IVIS® Spectrum.
47
Page 54
8. Troubleshooting
VOLTAGE
8.4 No Image Is Produced
If no image is produced, there may be an error in the Living Image software, a
problem with the physical connections to the camera, or a hardware failure.
1. Close the Living Image
2. Restart the Living Image software and try to acquire an image.
3. If after restarting the computer, you are still unable to produce an image,
8.5 Hardware Problems
Due to its integrated design, almost all of the components in the IVIS Spectrum
are not user-serviceable.
®
software and restart the computer.
contact IVIS Imaging Systems Technology Support for assistance.
VOLTAGE! Do not remove any panel from the console, except for the filter
wheel access panel. Removing a panel could expose the user to hazardous
voltages.
Circuit Breaker (Power
Entry Module)
If you suspect a problem with the instrument hardware, contact IVIS Imaging
Systems Technology Support.
Under normal operation, the circuit breaker that is incorporated into the rear
panel power entry module (main power switch) should not trip. Frequent
tripping of this circuit breaker may indicate an underlying electrical problem. If
this occurs, unplug the system and contact IVIS Imaging Systems Technology
Support.
If the circuit breaker is occasionally tripped:
1. Confirm that the rear panel switch is off, the front panel toggle switch is
down (standby), and the power cord is plugged into the console and the wall
socket.
®
2. Turn on the computer and, if necessary, start the Living Image
software.
3. Turn on the rear panel main power switch.
— The status light is blue.
4. Turn on the front panel toggle switch by pulling the toggle out and up.
— The status light is blue and blinking.
5. Initialize the system.
— The status light is red and turns green when the system is initialized.
Wait until the camera has cooled to the demand temperature before you
begin an imaging session.
Computer ConnectionsNever disconnect or wiggle the rear panel computer connections when the
®
Spectrum is powered on. If you suspect that any of the rear panel
IVIS
computer connections are loose, follow the procedure below.
1. Close the Living Image software and shut down the computer.
2. Turn off the front panel toggle switch (standby).
48
Page 55
®
VOLTAGE
Toggle switch
Front panel
Back panel
Spectrum Hardware Manual
IVIS
3. Turn off the rear panel circuit breaker switch.
Figure 8.4 IVIS Spectrum, front and back panels
4. Check the connections between the computer and the rear panel of the IVIS
Spectrum. Tighten any loose connection.
5. Turn on the computer and start the Living Image software.
6. Turn on the rear panel main power switch.
— The status light is blue.
7. Turn on the front panel toggle switch by pulling the toggle out and up.
— The status light is blue and blinking.
8. Initialize the system.
— The status light is red and turns green when initialization is complete.
Wait until the camera has cooled to the demand temperature before you
begin an imaging session.
Laser/GalvanometerThe laser/galvanometer is not user-serviceable. If you notice any changes to the
alignment grid (for example, change in the brightness or position of the grid),
contact IVIS Imaging Systems Technology Support.
Lamp ReplacementThe quartz halogen lamp that provides the fluorescent light source typically
operates for thousands of hours. This light source is housed in a compartment
that contains hazardous voltages and is not user accessible or serviceable.
VOLTAGE! DO NOT attempt to service the quartz halogen lamp.
If the lamp brightness is decreased or if lamp replacement is required, contact
IVIS Imaging Systems Technology Support.
49
Page 56
8. Troubleshooting
The stage lock down bar is located at
the bottom front edge of the imaging
platform.
Pull the bar down from the stowed
position and place it as shown above.
Return the bar to its stow position
before returning the imaging platform to
the operating position.
8.6 Accessing the Area Under the Imaging Platform
During normal operation, the imaging platform always returns to the loading
position before the door can be opened. This helps ensure user safety and
convenience when loading subjects. Occasionally you may need access to the
area beneath the platform for cleaning or to retrieve a subject. Since the
platform moves rapidly and produces strong forces, the area beneath the
platform is not accessible unless the normal operation mode is overridden.
To access the space below the imaging platform:
1. Remove all items from the imaging stage that are taller than 2 cm and close
the door.
2. In the IVIS System Control panel, select Service from the FOV drop-down
menu.
— The platform moves to the service position and the door can be opened.
3. Place the stage lock down bar in position as shown in Figure 8.5.
Be careful not to disturb the platform heater cable or the gas tubing located
in this area.
To return the platform to the operating position:
1. Return the lock down safety bar to the stowed position under the front edge
of the imaging platform and close the door.
2. In the IVIS System Control panel, select Load from the FOV drop-down
menu.
— The imaging platform returns to the normal operating position.
Figure 8.5 Imaging platform and the stage lock down bar
If you need to access the space under the imaging platform, use the lock down bar as a safety precaution.
If necessary, wipe the exterior surface of the IVIS Spectrum console with a soft
cloth. If the system requires more aggressive cleaning or sterilization, contact
IVIS Imaging Systems Technology Support.
DANGER! DO NOT use sprays or fluids to clean the exterior or interior of the
module. Do not allow fluids of any kind to penetrate the electronics cabinet
under any circumstances. Sprays and liquids that come into contact with the
IVIS Spectrum may result in damage to the system or electrocution.
Approved Cleaning
Solutions
WARNING! Do not use fluids or moistened towels to clean any part of the
system console where electrical or fiber optic cables make connections. Do
not use fluids of any kind near the filter wheel assembly. Turn off the electrical
power to the IVIS Spectrum by turning the front panel toggle switch to standby
and turning off the rear panel switch before engaging in cleaning operations
that use fluids. The imaging chamber power switch is located on the rear of
the system console.
The compounds shown in Table 9.1 do not damage the internal finish of the
®
Spectrum imaging chamber and are suitable as cleaners, if required. Do
IVIS
not use any solution not included in this list. In particular, avoid strong bases,
bleach, or acids that may potentially damage the unit and compromise its
operation.
ALERT! Do not spray cleaning solutions in the imaging chamber.
51
Page 58
9. Care & Maintenance
NOTICE
Table 9.1 Acceptable cleaning solutions for the IVIS® Spectrum imaging chamber
Cleaning SolutionManufacturer
®
Cidexplus
70% methyl alcohol/30% deionized
water solution
70% ethyl alcohol/30% deionized water
solution
Sporicidin
phenol)
Clidox-s
Solution (3.4% glutaraldehyde) Johnson & Johnson Medical
®
Sterilizing Solution (1.56%
®
DisinfectantPharmacal Research Laboratories, Inc.
Sporicidin International
Caliper makes no claims as to the sterility of the IVIS imaging chamber after
using the solutions in Table 9.1. Please refer to the manufacturer’s literature
for information as to the applicability of the compound for the organism of
interest. Consider dedicating an imaging system for immunodeficient animals.
This will reduce the risk of cross contamination.
Cleaning the Imaging
Platform
It is recommended that you use a lint-free wipe, such as Scott Pure® wipe or a
®
Kaydry EX-L
wipe to minimize the presence of particulate matter in the
imaging chamber.
After saturating a lint-free wipe, clean the internal surfaces using a gentle
circular motion. Do not pour or spray the solution directly onto internal
surfaces. Rinse surfaces using a wipe saturated with sterile deionized water. Do
not allow puddles of water to remain on the surfaces.
Avoid cleaning the ceiling of the image chamber except when necessary. When
cleaning the chamber ceiling, use extreme care to avoid damaging the lens
protection window and the light diffusers in this area.
To avoid any phosphorescence from the cleaner, be sure that the surfaces are
dry before using the imaging chamber.
9.2 Cleaning the Lens Protection Window
Contact IVIS Imaging Systems Technology Support for information about
cleaning or sterilizing any of the optical components or replacing the optical
filters. A clear window at the top of the imaging chamber protects the lenses and
other optical components from contamination. Do not touch or allow subjects
to touch this window, otherwise image quality may be impaired. Further, the
glass window could be cracked or broken if struck by a hard object. The lens
protection window should be cleaned periodically.
To clean the surface of the lens protection window:
1. Blow off any accumulated dust using a Dust-Off® compressed gas duster.
2. Dampen a lint-free cloth with isopropyl alcohol and gently swab the
window surface to remove any smudges or streaks.
3. Repeat step 2 using a clean area of the cloth or a new cloth until the window
is clean.
52
Page 59
®
NOTICE
NOTICE
Spectrum Hardware Manual
IVIS
4. Take an image to confirm that the image quality is restored.
9.3 Changing the Emission Filters Used in Fluorescence Imaging
The following steps explain how to change one or more of the optional emission
filters used for fluorescence imaging in the IVIS
®
Spectrum.
Before you begin, gather the following tools:
5/32″ hex key
•
#2 Phillips head screw driver
•
One pair of lint free, powderless gloves
•
1. In the Living Image
®
software, save any important data, then exit the
program.
2. Turn the front panel switch to off (standby) and turn off the main power at
the rear of the system console.
3. Using the 5/32″ hex key, remove the eight screws that hold the filter access
panel in place on the upper right side of the console.
— The panel is held in place by four pins until it is pulled forward. be
careful not to damage the light-sealing gasket when removing the panel.
The panel has a machined recess on the inner left-hand side (towards the
front of the console). Noting this orientation is helpful when you reinstall the
panel.
4. Notice that the 1/2″ shaft on the right side of the filter wheel compartment
has a carriage clamp lock with a handle. Turn the handle counter-clockwise
to loosen the clamp so that it can slide freely over the shaft.
Starting with step 5, wear lint-free, powderless gloves to prevent smudging
the filters.
5. Grip the filter wheel by the metal rim and pull the filter wheel carriage
forward until it stops. Be careful not to touch the glass filters or to dislodge
the two drive belts.
Since the carriage slides out about eight inches, only two or three of the
filters will be visible. The filter wheel holds the filters numbered 1-11.
6. To remove a filter from the filter wheel:
a. Slowly rotate the filter wheel until the filter of interest is exposed.
b. Use a #2 Phillips head screw driver to loosen the flat head screw ONE
TURN.
The screw has a captured O-ring that retains the glass emission filter.
c. Use two gloved fingers to remove the filter. Alternatively, gently push
the filter up with a gloved finger.
53
Page 60
9. Care & Maintenance
NOTICE
NOTICE
NOTICE
NOTICE
7. To clean a filter, use a Dust-Off® compressed gas duster to apply an air
stream at an angle to the filter surface, otherwise dust particles could be
driven into the filter coating.
Do not blow air onto a filter in the filter wheel as dust could be spread to other
optical components.
8. To install a filter, insert the filter into the filter wheel recess and tighten the
O-ring screw until it bottoms out (about one turn).
Make sure that the filter retaining screw and O-ring are fully seated to prevent
interference or rubbing against components above them. The top of the filter
metal ring should be nearly flush with the wheel. The filter encapsulating metal
ring has one side that is thinner than the other. The thin part of the ring should
face downward in the wheel recess so that the arrow that is printed on the side
of the filter points up.
9. Push the filter wheel carriage back into the console by placing one finger on
the rim of the upper filter wheel.
10. While you hold the carriage in the closed position, turn the carriage-locking
clamp in the clock-wise direction. Be careful not to dislodge any drive belts
off of their pulleys when locking the clamp.
11. Replace the access panel by positioning it onto the four pins. The inner
machined out recess should be on your left side.
12. Loosely install the eight screws, then lightly tighten the four central screws
and in a crisscross pattern. Lightly tighten the four outer screws in a
crisscross pattern.
Do not over tighten the screws. The panel is designed to seal out light without
large clamping force.
13. Restart the system following the procedure on page 35.
If you change a filter or a filter position, you must also change the
corresponding filter label in the Living Image® software. For more details, see
the Living Image Software Manual.
54
Page 61
®
CAUTION
!
Spectrum Hardware Manual
IVIS
9.4 Moving the System
Removing the DomeIt may be necessary to remove the plastic camera dome when you move the
IVIS® Spectrum though a doorway or other low overhang.
■ To remove the dome, wrap a flat blade screw driver with a thin cloth or
paper towel and use the screw driver to gently pry the dome off at the rear
of the console.
Moving the SystemIf the console is only repositioned a short distance within the range of the power
cord and cables, the power need not be turned off. However, be careful not to
accidentally turn off the power when gripping the system at the right rear.
®
When possible, two people should move the IVIS
and it is difficult to see potential obstacles when pushing the console from
behind. Make sure that the path is clear of items that could jolt the casters and
cause the console to tip. Move the console slowly so that you are always in
control of its motion.
To move the system:
1. Put the front panel toggle switch in the standby position.
2. Turn off the main power switch on the rear panel and disconnect the power
cord from the wall socket and the console.
3. Remove all cables from the rear panel of the console.
Spectrum. The console is tall
CAUTION! Any cables that dangle from the console could cause the system
to topple if caught by a caster.
4. Unlock the four casters.
5. Grip the console at a convenient location, but do not use the door handle as
a grip. Be careful when gripping the right rear of the console so that you do
not damage the connectors or fan grills.
6. After the system is positioned at the new location, lock the casters.
7. Reconnect the cables and power cord.
8. If necessary, reinstall the dome.
9. To restart the system:
a. Confirm that the rear panel switch is off, the front panel toggle switch is
down (standby), and all cables are reconnected.
b. Confirm that the power cord is plugged into the console and the wall
socket.
®
c. Turn on the computer and start the Living Image
software.
d. Turn on the rear panel main power switch.
— The status light is blue.
e. Turn on the front panel toggle switch by pulling the toggle out and up.
— The status light is blue and blinking.
f. Initialize the system.
55
Page 62
9. Care & Maintenance
— The status light is red.
The status light turns green when the system is initialized. Wait until the
camera has cooled to the demand temperature before you begin an
imaging session (about ten minutes).
56
Page 63
®
VOLTAGE
Spectrum Hardware Manual
IVIS
10
Servicing & Maintaining the ThermoCube Chiller Unit
Over time, the coolant level in the ThermoCube thermoelectric chiller unit may
diminish and require replenishing. It is recommended that you check the coolant
level in the ThermoCube unit every six months. This chapter explains how to
add coolant to the Thermocube unit.
The following tools and items are required, but not provided.
Allen Wrenches (hex head wrench): 3/32" and ¼" measured across the
•
flats (AF).
Measuring cup with a spout.
•
Spectral TE Camera Cooling Mixture (part no. 119219, includes a 3/32"
•
and ¼" allen wrench) or a solution of 25% pure ethanol, 75% distilled
water.
Paper towel or cloth for clean up.
•
10.2 Turning Off Electric Power to the Main Console
VOLTAGE! Do not remove any panels from the console until you turn off the
IVIS Imaging System disconnect the power cord. Otherwise, electric shock
could result.
1. Pull the front toggle switch out and down to the standby position.
2. Turn off the main power switch on the rear panel and disconnect the console
power cord from the wall socket.
3. Move any cables away from the lowest rear panel of the console so that the
panel can be removed
(Figure 10.1). (No other cables need to be removed.)
57
Page 64
10. Servicing & Maintaining the ThermoCube Chiller Unit
CAUTION
!
ThermoCube unit
Figure 10.1 Lower rear panel
Note that the power cord has been removed from the console.
10.3 Removing the Console Rear Panel
1. Use a 3/32" hex head wrench to remove the screws that secure the bottom
rear (fan) panel
(Figure 10.2).
2. Carefully remove the panel and lean it against the side of the console.
CAUTION! Be careful not to pull too hard on the rear fan panel, otherwise its plug
may become disconnected. If this happens, contact IVIS Imaging Systems
Technology Support).
.
Figure 10.2 Removing the lower rear panel using a hex head wrench (left) and ThermoCube unit exposed (right)
58
Page 65
®
NOTICE
Spectrum Hardware Manual
IVIS
10.4 Sliding the Tray Out
The ThermoCube unit sits on a tray that must be pulled out to access the unit.
1. Use the ¼" hex head (Allen) wrench to loosen and remove the bolt that
secures the tray under the ThermoCube unit to the console
(Figure 10.3).
Figure 10.3 Loosen the tray locking bolt
2. Move any cables or tubes out of the way so that the tray can be pulled out
3. Pull out the tray with the ThermoCube unit
The ThermoCube unit has its own power switch located at the rear of the unit.
The power switch is not visible until the unit is pulled out. Make sure that the
power switch is not accidentally turned off when you slide the tray out or back
into the console.
4. Remove the fill cap on the ThermoCube chiller and check coolant level. If
freely.
(Figure 10.4).
the coolant level is half way up the neck of the tank, no additional
coolant is required.
Figure 10.4 Removing the fill cap
59
Page 66
10. Servicing & Maintaining the ThermoCube Chiller Unit
CAUTION
!
NOTICE
WARNING
!
10.5 Filling the ThermoCube Unit
1. If additional coolant is required, wrap a piece of cloth or a paper towel
around the neck of the tank to absorb accidental spills
2. Using a spouted measuring cup, add small quantities of the 25% ethanol,
75% distilled water solution to the ThermoCube tank until the coolant level
is half way up the neck of the tank. Use care not to spill the liquid and wipe
any liquid from the top or walls of the chiller.
CAUTION! DO NOT OVERFILL the ThermoCube unit. Over filling the tank
can cause fluid to run down the chiller walls and contact the power plug.
(Figure 10.5).
60
Figure 10.5 Adding coolant to the ThermoCube unit
If the Caliper Spectral TE Camera Cooling Mixture (part no. 119219) is not
available, a suitable substitute can be made from ordinary 80 proof Vodka and
distilled water. To make 750 ml of the disinfecting coolant, mix 470 ml of 80
proof alcohol (40% ethanol) and 280 ml distilled water.
WARNING! If you prepare your own ethanol solution, adhere to the 25% (by
volume) pure ethanol concentration. A higher alcohol concentration may be
flammable and a lower concentration may not suppress organic growth in the
cooling system or camera.
Page 67
®
Spectrum Hardware Manual
IVIS
10.6 Reassembling the System
1. Replace the cap on the ThermoCube unit. Make sure that the cap is fully
tightened to prevent rapid coolant evaporation.
2. Slide the tray back into the console while you hold the cables and hoses out
of the way to prevent snagging.
3. Replace the tray bolt using the ¼" hex head (Allen) wrench
4. Confirm that the ThermoCube power switch is on and that no tools or other
hardware remain inside the IVIS.
5. Replace the lower rear panel using a 3/32" hex head wrench to tighten the
screws
10.7 Restarting the IVIS Spectrum
1. Confirm that the IVIS system rear panel power switch is off and that the
front panel toggle switch is in the down (standby) position.
2. Reconnect the power cord first to the console and then to the wall socket.
3. Confirm that all other cables are properly connected.
4. Turn on the computer.
5. Turn on the rear panel power switch.
— The front status light is blue.
6. Turn on the front panel toggle switch by pulling out and up.
— The status light is blue and blinking.
7. Start the Living Image software and initialize the system.
— The status light is red and turns green when the system is initialized.
Wait until the camera has cooled to the demand temperature before you
begin an imaging session (about 10 minutes).
(Figure 10.3).
(Figure 10.2).
61
Page 68
10. Servicing & Maintaining the ThermoCube Chiller Unit
[This page intentionally blank.]
62
Page 69
®
Spectrum Hardware Manual
IVIS
Appendix A Spare Parts List
DescriptionCaliper Part Number
Mouse Inhalation Nose Cone117150
Low Profile, 5 Station Manifold117706
XRH-1 High-Reflectance Hemisphere with container118937
Black Paper, Swarthmore, Artagain. 9x12”, 24 sheet pad117837
mat, low fluorescence 42
measured unequal to demand temperature 45
mechanical safety 12
moving the system 9, 55–56
N
no image produced 48
nose cone 63
O
optical components
lenses 28
optical specifications 22
other equipment, chemicals, or supplies 9
P
patents 18
photographic image unacceptable 46
pinch hazard label 6
power cord 8
power line surges 8
power outages 9
power overloading 9
power sources 8