PerkinElmer AAnalyst 700 User Manual

Page 1
A
TOMIC SPECTROMETRY
Return to Document Menu
AAnalyst 700
Atomic Absorption Spectrometer
User’s Guide
Page 2
Page 3
AAnalyst 700
Atomic Absorption Spectrometer
User’s Guide
Page 4
PerkinElmer Bodenseewerk is registered for the design and manufacture of laboratory analytical equipment under the quality requirements of BS EN ISO 9001.
Certificate No. FM 22178
Notice
The information contained in this document is subject to change without notice.
Release history
Part Number Release Publication Date
0993-5255 1
2 3 4 5
January 1998
September 1998
June1999
March 2000
September 2000
Trademarks
PerkinElmer is a trademark of PerkinElmer, Inc. AAnalyst, AA WinLab, Lumina, and HGA are trademarks of affiliates of PerkinElmer Instruments LLC.
Windows is a trademark and Microsoft is a registered trademark of Microsoft Corporation.
Registered names, trademarks, etc. used in this document, even when not specifically marked as such, are not to be considered unprotected by law.
Copyright information
This document contains proprietary information that is protected by copyright. All rights are reserved. No part of this document may be reproduced in any form whatsoever or translated into any language without the prior written permission of PerkinElmer Instruments LLC or one of its affiliates.
Copyright ©1998–2000 by affiliates of PerkinElmer Instruments LLC
Printed in the Federal Republic of Germany
Technical Documentation PerkinElmer Bodenseewerk D-88647 Ueberlingen, Federal Republic of Germany
Page 5
Contents
Chapter 1 Safety and Regulatory Information
Safety practices and conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Correct use of the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Electrical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Operating conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Laboratory hygiene . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Laboratory ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Safe handling of chemicals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Waste disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
UV radiation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
High temperatures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-12
Handling compressed gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
Safety practices for flame atomization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-16
Safety practices for electrothermal atomization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-22
Safety practices for FIAS and mercury hydride systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-24
References for laboratory safety practice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-26
Regulatory information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-27
IEC 1010 compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-27
Electrical protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-27
Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-28
Symbols used on the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-29
Warning Markings on the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-30
Warning markings at the front of the spectrometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-30
Warning markings on the rear panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-32
Warning markings on the deuterium lamp cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-35
page
0993-5255
C-1
Page 6
Contents
Chapter 2 Spectrometer Installation
Before you install the spectrometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Moving and reinstalling the spectrometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Preparing the spectrometer to be moved . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Moving the spectrometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Reinstalling the spectrometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Connecting the burner gases. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Connecting the furnace gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
To connect the drain system to the burner . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
To install the furnace autosampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
Connecting cooling sources . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
Electrical connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
Chapter 3 Preparing the System for Analyses
Switching on the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Installing lamps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Installing sample trays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Shutting down the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Emergency shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
page
Chapter 4 Spectrometer: System Maintenance
PerkinElmer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Maintenance checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Spectrometer maintenance procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Cleaning the windows of the optical system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Maintaining the electrical components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
Changing the air filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
Replacing the deuterium lamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
Maintaining the acetylene supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
Chapter 5 Parts and Supplies
Parts provided . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Replacement parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Obtaining supplies, replacement parts, and accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
C-2
0993-5255
Page 7
Contents
Chapter 6 Laboratory Requirements
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Operating conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Space requirements for the spectrometer system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Electrical requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
Gas supply requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
Burner gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
Furnace gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
Cooling the spectrometer system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
Lamps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-18
Computer and printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-18
Chapter 7 Spectrometer: System Description
The AAnalyst 700 Spectrometer System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
The optical system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-6
Switches and connectors on the spectrometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-14
page
Chapter 8 Translations of Warnings
Translations of Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
Symbols used on the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-45
Index
Customer Service
0993-5255
C-3
Page 8
Contents
C-4
0993-5255
Page 9

Safety and Regulatory Information

1Safety and Regulatory Information
1
Contents
Safety practices and conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Correct use of the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Electrical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Operating conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Laboratory hygiene . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Laboratory ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Safe handling of chemicals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Waste disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
UV radiation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
High temperatures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-12
Handling compressed gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
Safety practices for flame atomization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-16
Safety practices for electrothermal atomization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-22
Safety practices for FIAS and mercury hydride systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-24
References for laboratory safety practice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-26
Regulatory information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-27
IEC 1010 compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-27
Electrical protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-27
Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-28
Symbols used on the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-29
Warning Markings on the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-30
Warning markings at the front of the spectrometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-30
Warning markings on the rear panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-32
Warning markings on the deuterium lamp cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-35
page
0993-5255
1-1
Page 10
1-2
0993-5255
Page 11

Safety practices and conventions

This instrument has been designed to comply with a wide variety of international standards governing the safety of laboratory equipment. In routine use, the instrument poses virtually no risk to you. If you take some simple, common-sense precautions, you can maintain the continued safe operation of the instrument.
The guide provided with the instrument contains information and warnings that must be followed by the user to ensure safe operation and to maintain the instrument in a safe condition. This advice is intended to supplement, not supersede, the normal safety code of behavior prevailing in the country of operation.
The information provided does not cover every safety procedure that should be practiced. Ultimately, maintenance of a safe laboratory environment is the responsibility of the user and the user’s organization.
Possible hazards that could harm the user or result in damage to the instrument are clearly stated at appropriate places throughout this guide.
Any of the following safety conventions can be used throughout this guide:
Safety/Regulatory Information
0993-5255
This symbol alerts you to situations that could result in personal injury to yourself or other persons. Details about these circumstances are in a box like this one.
This symbol alerts you to the risk of
injury
to yourself or other persons.
Details about these circumstances are in a box like this one.
This symbol alerts you to the risk of
injury
to yourself or other persons.
Details about these circumstances are in a box like this one.
electric shock that could result in personal
hot surfaces that could cause personal
1-3
Page 12
Safety/Regulatory Information
This symbol alerts you to the risk of ultraviolet radiation that could cause
eye damage to yourself or other persons.
Details about these circumstances are in a box like this one.
Caution
The term CAUTION alerts you to situations that could result in serious damage
to the instrument
Details about these circumstances are described in a message similar to this one.
or other equipment.

Correct use of the instrument

Before you install or use your instrument, and in order to get the best results, you should be familiar with all of the instruments in the system and know how to operate them. You should also be aware of the safety procedures in force in your laboratory, especially those concerning atomic spectrometry instruments. Consult the guides supplied with the instruments before you start.
If you use the instrument in a manner not specified in the guide, or if you use it for a purpose other than that intended, you may damage the instrument, or compromise your own, or someone else’s, safety.
This instrument should only be operated by persons who are suitably qualified and have received adequate training.
1-4
0993-5255
Page 13

Electrical safety

Safety/Regulatory Information
Warning: Unauthorized Adjustments and Servicing
Only a PerkinElmer service engineer or similarly trained and authorized person should be permitted to service the instrument.
Do not attempt to make adjustments, replacements, repairs, or modifications
to this instrument except as described in the documentation supplied with the instrument.
Warning: Electrical Hazard
Any interruption of the protective conductor inside or outside the instrument or disconnection of the protective conductor (earth/ground) terminal is likely to make the instrument dangerous. Intentional interruption is prohibited.
Lethal voltages are present in the instrument
Even with the power switch off, line power voltages can still be present within
the instrument.
When the instrument is connected to line power, terminals may be live, and
opening covers or removing parts (except those to which access can be gained without use of a tool) is likely to expose live parts.
Capacitors inside the instrument may still be charged even if the instrument
has been disconnected from all voltage sources.
W1.2
W2.1
0993-5255
To ensure satisfactory and safe operation of the instrument, it is essential that
the line power cord is connected to a properly installed line power outlet that has a protective conductor (earth/ground).
Do not attempt to make internal adjustments or replacements except as
directed in the guide provided with the instrument.
Do not operate the instrument with any covers or parts removed.
Disconnect the instrument from all voltage sources before opening it for any
adjustment, replacement, maintenance, or repair. If, afterwards, the opened instrument must be operated for further adjustment, maintenance, or repair, this must only be done by a skilled person who is aware of the hazard involved.
1-5
Page 14
Safety/Regulatory Information
Use only fuses with the required current rating and of the specified type for
replacement.
Do not use makeshift fuses or short-circuit the fuse holders.
Whenever it is likely that the instrument is no longer electrically safe for use,
make the instrument inoperative and secure it against any unauthorized or unintentional operation.
The instrument is likely to be electrically unsafe when it:
Shows visible damage.
Fails to perform the intended measurement.
Has been subjected to prolonged storage under unfavorable conditions.
Has been subjected to severe transport stresses.
Note: The spectrometer is provided with an operational on/off switch and with a circuit breaker. The operational on/off switch turns the instrument on and off, but it does not disconnect the instrument from the line power supply. The circuit breaker functions as a ‘disconnecting device’ and disconnects the instrument entirely from the line power supply. Further information on these two switches is provided in ‘Switches and connectors on the spectrometer’ starting at page 7-6.
1-6
0993-5255
Page 15

Operating conditions

Safety/Regulatory Information
Warning: Explosive Atmosphere
This instrument is not designed for operation in an explosive atmosphere.
The instrument will operate correctly under the following conditions:
W1.3
Indoors.
Ambient temperature +15 C to +35 C (+59 F to +95 F), with a maximum
change not exceeding 2.8 C (5 F) per hour.
Ambient relative humidity 20% to 80%, without condensation.
Altitude in the range 0 m to 2000 m (sea level to 6500 feet).
The location must be free of dust, smoke, and corrosive fumes.
You can store the instrument safely under the following conditions:
Ambient temperature –20 C to +60 C (–4 F to +140 F).
Ambient relative humidity 20% to 80%, without condensation.
Altitude in the range 0 m to 12000 m (sea level to 40 000 feet).
Note: When you remove the instrument from storage and before you put it into operation, allow it to stand for at least a day under the approved operating conditions.
0993-5255
1-7
Page 16
Safety/Regulatory Information

Laboratory hygiene

Keep the work area scrupulously clean to avoid contaminating your samples
and to maintain a safe working environment. Clean up spilled chemicals immediately and dispose of them properly.
Do not allow smoking in the work area. Smoking is a source of significant
contamination and also a potential route for ingesting harmful chemicals.
Do not store, handle, or consume food in the work area.

Laboratory ventilation

Toxic combustion products, metal vapor, ozone, etc., can be generated by the system, depending on the type of analyses being performed.
You must provide an efficient laboratory ventilation system to remove toxic
products generated during instrument operation.
1-8
0993-5255
Page 17

Safe handling of chemicals

Some chemicals used with the instrument may be hazardous or may become hazardous after completion of an analysis.
1
The responsible body surrounding workplace is safe and that instrument operators are not exposed to hazardous levels of toxic substances (chemical or biological) as defined in the applicable national, state, and local health and safety regulations and laws. Venting for fumes and disposal of wastes must be in accordance with all national, state, and local health and safety regulations and laws.
Use, store, and dispose of chemicals in accordance with the manufacturer’s
recommendations and the applicable national, state, and/or local regulations.
Wear appropriate eye protection at all times while handling chemicals.
Depending on the types of chemicals you are handling, wear safety glasses with side shields, or goggles, or a full-face shield.
Wear suitable protective clothing, including gloves if necessary, resistant to
the chemicals you are handling.
must take the necessary precautions to ensure that the
Safety/Regulatory Information
When preparing chemical solutions, always work in a fume hood that is
suitable for the chemicals you are using.
Perform sample preparation away from the instrument to minimize corrosion
and contamination.
Clean up spills immediately using the appropriate equipment and supplies,
such as spill cleanup kits.
Do not put open containers of solvent near the instrument.
Store solvents in an approved cabinet (with the appropriate ventilation) away
from the instrument.
1. Definitions from IEC 1010-1:
Responsible body: ‘individual or group responsible for the use and maintenance of
equipment, and for ensuring that operators are adequately trained.’
Operator: ‘person operating equipment for its intended purpose.’
0993-5255
1-9
Page 18
Safety/Regulatory Information

Waste disposal

Contents of waste containers
The materials that you collect in waste containers may include small amounts of the substances that were analyzed and other chemicals used in the analyses. If these materials are toxic, corrosive, or contain organics you may have to treat the collected effluent as hazardous waste. Refer to your local safety regulations for proper disposal procedures.
Disposing of defective lamps
Hollow cathode lamps and electrodeless discharge lamps contain small quantities of the lamp element in a very pure form. If a lamp containing toxic elements becomes defective or otherwise unusable, you must treat it as hazardous waste and dispose of it accordingly. A licensed company in the chemical waste disposal business can provide lamp disposal in accordance with environmental regulatory requirements. Please note that PerkinElmer does not take back defective lamps for disposal. In addition, hollow cathode lamps and deuterium lamps are maintained under reduced pressure. Handle and dispose of them correctly to minimize the implosion risk.
1-10
For information, the quantities of analyte material contained in the cathodes of hollow cathode lamps are listed in the table below:
Quantity of material in the cathode
less than 5 g
5–10 g
10–15 g Cu, Fe, Hg, Nb, P, Pb, Ti, Tl all other multi-element HCLs
Single-element hollow cathode lamps
As, Au, B, Ba, Be, Ca, Dy, Er, Eu, Ga, Gd, Ge, Hf, Ho, In, Ir, K, La, Li, Mg, Na, Nd, Pd, Pr, Pt. Re, Rh, Ru, Sc, Se, Sm, Sn, Ta, Tb, Tm, Yb, Y
Ag, Al, Bi, Cd, Co, Cr, Mn, Mo, Ni, Sb, Si, Sr, Te, V, W, Zn, Zr
Multi-element hollow cathode lamps
Na-K, Pt-Ru
Ca-Mg, Ca-Zn, Ag-Au, Sn-Te, Ca-Mg-Zn, Ca-Mg-Al
For EDLs, the quantity of analyte material used is much less than the quantities used in HCLs. Typically, only several milligrams of material are placed in the EDL bulb. No PerkinElmer EDLs (System 1 or 2) contain more than 0.05 g of analyte material.
0993-5255
Page 19

UV radiation

You should be aware of the health hazard presented by UV radiation.
When the instrument is on, do not remove any covers unless specifically
Always wear UV-absorbing eye protection when viewing any of these
Safety/Regulatory Information
instructed to do so in the guide since otherwise you may be exposed to potentially hazardous UV radiation.
sources:
The flame, especially the nitrous oxide/acetylene flame.
The graphite furnace when it is heated to incandescence.
Hollow cathode or electrodeless discharge lamps.
The deuterium background correction lamp.
0993-5255
1-11
Page 20
Safety/Regulatory Information

High temperatures

Burner system
The flame can generate temperatures of up to 2800 °C.
Do not touch the burner head until it has cooled to room temperature.
Graphite furnace
The HGA graphite furnace can generate temperatures of up to 3000 C.
Do not touch any part of the graphite furnace until it has cooled to room
temperature.
Do not attempt to inject a sample into the graphite tube when the furnace is
hot. You may be subjected to fumes from the sudden vaporization of the sample, and may damage the pipet.
Quartz tube atomizer cell
The quartz tube atomizer cell (QTA-cell) used for the hydride-generation technique (option) can reach temperatures of up to 1000 C.
1-12
Do not touch any part of the heating mantle (electrically heated systems), or
the cell holder (flame-heated systems), or the QTA-cell until they have cooled to room temperature.
0993-5255
Page 21

Handling compressed gases

Note: The permanent installation of gas supplies is the responsibility of the user and should conform to local safety and building regulations.
Summary of gas hazards
Hazards associated with the different gases used in atomic absorption spectrometry are presented in the table below:
Gas Suffocation Explosion
Air – Argon x – Nitrogen x – Acetylene
(ethyne, C
Nitrous Oxide (dinitrogen monoxide, N
2H2
)
2
O)
Safety/Regulatory Information
Spontaneous Decomposition or Combustion
xx x
x– x
0993-5255
Contact the gas supplier for a safety data sheet containing detailed information on the potential hazards associated with the gas.
Warning: Compressed Gases
High pressure gas cylinders can be dangerous if mishandled or misused.
Always handle gas cylinders with caution and observe your local regulations
for the safe handling of gas cylinders.
Identifying cylinders
Legibly mark cylinders to clearly identify the contents and status (full, empty,
etc.). Use the chemical name or commercially accepted name for the gas.
1-13
Page 22
Safety/Regulatory Information
Storing cylinders
Store cylinders in accordance with the applicable national, state, and/or local
regulations and standards.
When gas cylinders are stored in confined areas, such as a storage room, make
sure that ventilation is adequate to prevent toxic or explosive accumulations of gas. The storage room should be well ventilated and dry. This is particularly important in confined areas.
Do not store cylinders near elevators, gangways, or in locations where heavy
moving objects may strike or fall against them.
Use and store cylinders away from exits and exit routes.
Locate gas cylinders away from heat sources, including heat lamps.
Compressed gas cylinders should not be subjected to temperatures above 52 °C (125 °F).
Do not allow ignition sources in the storage area and keep cylinders away
from readily ignitable substances such as gasoline or waste, or combustibles in bulk, including oil.
1-14
Store all gas cylinders only in a vertical position, with the valve cap in place,
and fastened securely to an immovable bulkhead or a permanent wall.
If you are storing cylinders outdoors, store them above ground on a suitable
floor where they are protected against temperature extremes (including the direct rays of the sun).
0993-5255
Page 23
Safety/Regulatory Information
Handling cylinders
Move cylinders with a suitable hand truck after making sure that the valve cap
is securely in place and that the cylinder is properly fastened to the hand truck.
Use only approved regulators, tubing, and hose connectors.
Arrange gas hoses where they will not be damaged or stepped on and where
things will not be dropped on them.
Do not attempt to refill gas cylinders.
Check the condition of pipes, hoses, and connectors regularly, and replace any
damaged parts.
Perform periodic gas leak tests at all joints and seals of the gas system by
applying an approved gas leak detection solution.
0993-5255
1-15
Page 24
Safety/Regulatory Information

Safety practices for flame atomization

Safety interlocks
PerkinElmer provides a number of safety interlocks on the burner system to monitor gas pressure and check for the proper setup of the burner head, nebulizer, and drain system. In addition a flame sensor checks that the flame is burning.
Do not attempt to defeat these interlocks; you may compromise your own, or
someone else’s, safety.
Safe use of burner gases
The seepage of fuel gas or fuel gas mixture from the burner system, the drain system, the gas control system, or the gas connections constitutes a serious fire hazard.
Make sure that there are no breaks or leaks in any of these systems and that
all the seals are correctly installed and in good condition.
Regularly check for leaks at joints and seals using an approved leak test
solution.
1-16
When you perform any maintenance or installation procedures, follow the
instructions in the guide exactly.
Do not attempt to service the gas control system yourself.
A PerkinElmer service engineer, or similarly authorized and trained person, must perform the work.
When you shut down the instrument, for example at the end of the working
day, shut all the gas lines at the cylinder or regulator valves. Bleed the lines between the regulator and instrument to atmosphere before switching off the ventilation system.
0993-5255
Page 25
Safety/Regulatory Information
Air supply
An air compressor is recommended for the air supply to the burner system. Cylinders of compressed air should only be used as an emergency or short-term solution.
If you use cylinder air, make sure that it is compressed air and not oxygen-
enriched air (e.g. breathing air); the use of oxygen-enriched air can cause a flashback of the flame.
Never use oxygen as the oxidant since this can cause an explosion.
Nitrous oxide
(dinitrogen monoxide, N2O)
Tubing and fittings carrying nitrous oxide must be free of all oil, grease, and
organic materials. Spontaneous combustion may occur if nitrous oxide comes into contact with these materials.
Use a double-stage or heated regulator for nitrous oxide to prevent freezing of
the diaphragm and a loss of pressure regulation.
0993-5255
Do not store nitrous oxide in close proximity to flammable gases, such as
acetylene or hydrogen.
1-17
Page 26
Safety/Regulatory Information
Acetylene
Always use ‘Atomic Absorption Grade’ acetylene (ethyne) dissolved in
acetone (propan-2-one). This is satisfactory for all PerkinElmer AA spectrometers.
Do not store acetylene in close proximity to oxidizing gases, such as nitrous
oxide.
Use approved tubing and fittings for acetylene. Never use copper fittings
since acetylene forms an explosive compound with copper.
Prevent acetylene from coming into contact with copper, silver, mercury or
gaseous chlorine.
Periodically check for the presence of acetylene in the laboratory atmosphere,
especially near the ceiling.
For acetylene, set an outlet gauge pressure of between 90 kPa and 100 kPa
(0.9–1.0 bar, 13–14.5 psig). Never allow the outlet gauge pressure to exceed 103 kPa (1.03 bar, 15 psig); acetylene can explode spontaneously above this pressure.
Solvent Carryover.
If the acetylene cylinder pressure falls to below 600 kPa / 6.0 bar / 87 psig (at 20 °C), some of the solvent used to stabilize the acetylene (such as acetone) may be carried over into the burner. This could influence the characteristics of the burner with a resulting influence on the analytical results.
Change to a new acetylene cylinder when the cylinder pressure falls to below this value.
1-18
0993-5255
Page 27
Safety/Regulatory Information
Safe operation of the flame
Before you ignite the flame make sure that:
The laboratory fume ventilation system is operating;
The burner head is installed correctly;
The burner end cap is secured firmly;
The fuel and oxidant tubing fittings are properly connected;
The burner unit contains the necessary solvent-resistant components if
you intend to analyze samples containing organic solvents;
The atomizer compartment door is closed;
The pH of the liquid in the drain vessel is greater than pH 10 if you intend
to aspirate cyanide solutions.
Do not leave the flame unattended. Always make sure that there is a fire
extinguisher near the instrument.
Never change the gas pressure or shut a gas valve while the flame is burning.
Do not allow the burner head slot to become blocked. This can cause a
flashback of the flame.
0993-5255
Do not place open containers of flammable liquids and solvents near to the
flame. Be especially careful with samples that contain highly volatile solvents.
When you have analyzed samples containing organic solvents, flush all traces
of solvent from the burner system.
1-19
Page 28
Safety/Regulatory Information
The drain system
Always place the drain vessel in a well ventilated place underneath the
spectrometer, prevents the build-up of potentially hazardous gases, and allows you to see the liquid level.
Never use a glass drain vessel.
Do not kink, fold, or loop the drain tube.
Never place the drain tube directly into a laboratory sink.
Empty the drain vessel frequently; especially do not allow the drain vessel to
fill with organic solvent.
Thoroughly flush the drain system, as described in the guide, when you shut
down the system.
Check the condition of the drain tube regularly, especially if you use organic
solvents. Replace the drain tube when it first shows signs of cracking or discoloration.
Flashbacks
in full view while you are working with the spectrometer. This
1-20
A flashback is an explosion of the premixed gases in the spray chamber. The safety interlocks normally prevent conditions that could cause a flashback, but if a flashback should ever occur:
Check that the burner head slot is clean.
Make sure that you are using compressed air and not oxygen-enriched air.
Check the flow spoiler or impact bead for damage, and replace any damaged
parts.
Carefully check and tighten all connections to, and seals on, the burner unit,
and check the gas flow settings before you attempt to ignite the flame. Seepage of fuel and oxidant from the spray chamber may cause a flashback.
Check that the burner head is seated correctly.
0993-5255
Page 29
Safety/Regulatory Information
Hazards with Flame Atomization
Toxic combustion products:
Toxic combustion products can be generated by the system, depending on the type of analyses being performed.
You must provide an efficient laboratory ventilation system to remove toxic
products generated during instrument operation.
High temperatures:
The flame can generate temperatures of up to 2800 °C.
To avoid serious burns, never touch the burner head until it has cooled to
room temperature.
UV radiation:
The flame, especially the nitrous oxide/acetylene flame, emits ultraviolet radiation.
Keep the atomizer compartment door closed when the flame is burning and
never directly view the flame unless you are wearing UV-absorbing glasses.
0993-5255
Acetylides:
If you have aspirated high concentrations of copper, silver, or mercury solutions into an acetylene flame, unstable acetylides may have formed in the spray chamber. If permitted to dry, these compounds may explode.
Aspirate solution continuously to prevent residues from drying.
Thoroughly flush the spray chamber and drain system with water
immediately after such an analysis.
Cyanide solutions:
If you intend to aspirate cyanide solutions, make sure that the pH of the liquid in the drain vessel is greater than pH 10; toxic hydrogen cyanide gas is formed when cyanides contact acidic solutions.
Never allow solutions containing cyanides to mix with acidic solutions.
1-21
Page 30
Safety/Regulatory Information

Safety practices for electrothermal atomization

Safety interlocks
PerkinElmer provides a number of safety interlocks on the graphite furnace (electrothermal atomization) system that monitor the gas pressure and the temperature of certain system components.
Do not attempt to defeat these interlocks; you may compromise your own, or
someone else’s, safety.
Inert gas
The graphite furnace requires a supply of inert gas. PerkinElmer recommends
argon.
Although nitrogen can be used, it is not recommended for the inert gas supply. At furnace temperatures above 2300 °C, nitrogen reacts with graphite to form toxic cyanogen gas (CN)
Safety checks
.
2
1-22
Before you start an analysis, make sure that:
The laboratory fume ventilation system is operating;
The inert gas supply is connected and set to the correct pressure;
The cooling water supply is turned on;
The autosampler waste bottle is empty.
0993-5255
Page 31
Safety/Regulatory Information
Hazards with electrothermal atomization
Toxic products:
Toxic metal vapors, ozone, etc., can be generated by the system, depending on the type of analyses being performed.
You must provide an efficient laboratory ventilation system to remove toxic
products generated during instrument operation.
If you use nitrogen as the inert gas, toxic cyanogen gas (CN)
will be produced
2
when the furnace temperature is above 2300 °C.
High temperatures:
The HGA graphite furnace can generate temperatures of up to 3000 °C
To avoid serious burns, never touch the graphite components until the furnace
has cooled to room temperature.
Do not attempt to inject a sample into the graphite tube when the furnace is
hot. You may be subjected to fumes from the sudden vaporization of the sample, and may damage the pipet.
UV radiation:
The graphite furnace emits ultraviolet radiation when it is heated to incandescence.
Never directly view the graphite furnace unless you are wearing UV-
absorbing glasses.
0993-5255
1-23
Page 32
Safety/Regulatory Information

Safety practices for FIAS and mercury hydride systems

These systems are offered as options for the spectrometer.
Safety checks
Before you start an analysis, make sure that:
The laboratory fume ventilation system is operating;
The tubing in the fluid system is not constricted since this could lead to a
pressure build-up in the system;
The drain tubing is inserted in the waste container and that this container
is empty.
If you are using a flame to heat the QTA-cell or if you intend to perform
flame-FIAS analyses, refer to the information under Safety Practices for Flame Atomization before you ignite the flame.
1-24
0993-5255
Page 33
Safety/Regulatory Information
Hazards with FIAS and mercury hydride systems
Toxic products:
Toxic metal vapors, etc., can be generated by the system, depending on the type of analyses being performed.
You must provide an efficient laboratory ventilation system to remove toxic
products generated during instrument operation.
High temperatures:
The QTA-cell used for the hydride-generation technique can reach temperatures of up to 1000 C.
Do not touch any part of the heating mantle (electrically heated systems), or
the cell holder (flame-heated systems), or the QTA-cell until they have cooled to room temperature.
Hazardous chemicals:
Some of the chemicals required for analyses are corrosive and/or toxic.
Refer to the information under Safe Handling of Chemicals and to the specific
warnings in the documentation provided with the system before you start an analysis.
0993-5255
1-25
Page 34
Safety/Regulatory Information

References for laboratory safety practice

Bretherik, L., Bretherik’s Handbook of Reactive Chemical Hazards, 4th ed.,
Butterworth & Co. Ltd., London, UK, 1990.
Safe Practices in Chemical Laboratories,
Royal Society of Chemistry, London, UK, 1989.
Hazards in the Chemical Laboratory, 5th edition, Luxon, S.G., ed.,
Royal Society of Chemistry, London, UK, 1992.
CRC Handbook of Laboratory Safety, 3rd edition, Furr, K., ed.,
The Chemical Rubber Co. Press, Florida, USA, 1990.
Prudent Practices for Handling Hazardous Chemicals in Laboratories,
National Research Council, National Academic Press, Washington D.C., USA, 1981.
Sax’s Dangerous Properties of Industrial Materials,
9th edition, Sax, N.I. and Lewis, R.J., eds.,Wiley, New York, USA, 1998.
Pohanish, R.P. and Greene, S.A., Hazardous Materials Handbook,
Wiley, New York, USA, 1996.
1-26
Compressed Gas Association, Inc., Arlington, VA 22202, USA, various
publications.
Data Sheets provided by chemical manufacturers, e.g.:
– USA, Material Safety Data Sheets (MSDS), – FRG, DIN-Sicherheitsblätter, – UK, Hazard Data Sheets.
0993-5255
Page 35

Regulatory information

IEC 1010 compliance

This instrument has been designed and tested in accordance with EN 61010­1:1993 (IEC 1010-1): Safety requirements for electrical equipment for measurement, control, and laboratory use, including Amendments 1 and 2 to this standard, and EN 61010-2-061:1996 (IEC 1010-2-061:1995): Particular
requirements for laboratory atomic spectrometers with thermal atomization and ionization.

Electrical protection

Insulation
Class I as defined in IEC 1010-1.
Installation category
This instrument is able to withstand transient overvoltage according to Installation Category II as defined in IEC 1010-1 and IEC 664.
Safety/Regulatory Information
0993-5255
Pollution degree
This equipment will operate safely in environments that contain nonconductive foreign matter and condensation up to Pollution Degree 2 as defined in IEC 1010-1 and IEC 664.
1-27
Page 36
Safety/Regulatory Information

Electromagnetic compatibility (EMC)

European Union (EMC Directives)
This instrument has been designed and tested to meet the requirements of the EU Directives 89/336/EEC and 92/31/EEC. It complies with the generic EMC standard EN 61326-1:1997 Electrical equipment for measurement, control, and laboratory use – EMC requirements – General requirements.
This instrument has passed the following EMC tests:
Emission of conducted
Electromagnetic compatibility
and radiated noise
EN 55011:96 EN 61000-3-2:95 EN 61000-3-3:95
EN61000-4-2:95 EN61000-4-3:95 EN61000-4-4:95 EN61000-4-5:95 EN61000-4-6:96
United States (FCC)
This instrument is classified as a digital device used exclusively as industrial, commercial, or medical equipment. It is exempt from the technical standards specified in Part 15 of FCC Rules and Regulations, based on Section 15.103 (c).
Japan (FCC)
This instrument has been tested and found to comply with the limits of a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense. Changes or modifications not expressly approved by the manufacturer could void the user’s authority to operate the equipment.
1-28
0993-5255
Page 37

Symbols used on the instrument

Any of the following symbols can be used on the instrument:
This symbol indicates that there is a potential hazard and that the user must look in the guide for an explanation of the hazard and how to avoid it.
This symbol indicates the risk of electric shock.
This symbol indicates the risk of hot surfaces.
This symbol indicates the off position of the main power switch or circuit breaker. With the switch in this position, the instrument is disconnected entirely from the line power supply.
Safety/Regulatory Information
0993-5255
This symbol indicates the on position of the main power switch or circuit breaker. With the switch in this position, the instrument is connected to the line power supply but is not necessarily switched on and operational.
This symbol indicates the operational on/off switch.
This symbol indicates alternating current.
1-29
Page 38
Safety/Regulatory Information

Warning Markings on the instrument

Warning markings at the front of the spectrometer

A
C A
A
Risk of hot surfaces Risque de surfaces chaudes Gefahr durch heisse
Oberflächen
B
Burner system
B
Warning: Moving Parts – Risk of Injury
Some moving parts of the instrument are accessible in normal operation. Keep hands, clothing and other objects away from the moving parts of the instrument.
Danger: Parties en mouvement – Risque d'accident corporel
Certaines parties en mouvement de l'instrument sont accessibles en fonctionnement normal. Tenir les mains, vêtements et autres objets éloignés des parties en mouvement de l'instrument.
Warnung: Bewegliche Teile – Verletzungsgefahr
Bei normalem Betrieb sind bewegliche Teile des Geräts zugänglich. Hände, Kleidung und andere Gegenstände von den beweglichen Teilen des Geräts fernhalten.
C
Nebulizer clamp must be lowered and locked over the nebulizer flange. La pince du nébuliseur doit être abaissée et verrouillée audessus de la bride du nébuliseur. Die Zerstäuber-Befestigungsplatte muss nach unten geschoben und über dem Flansch des
Zerstäubers befestigt sein.
1-30
0993-5255
Page 39
A
Risk of hot surfaces Risque de surfaces chaudes Gefahr durch heisse
Oberflächen
Safety/Regulatory Information
A
B
HGA furnace
B
Warning: Moving Parts – Risk of Injury
Some moving parts of the instrument are accessible in normal operation. Keep hands, clothing and other objects away from the moving parts of the instrument.
Danger: Parties en mouvement – Risque d'accident corporel
Certaines parties en mouvement de l'instrument sont accessibles en fonctionnement normal. Tenir les mains, vêtements et autres objets éloignés des parties en mouvement de l'instrument.
Warnung: Bewegliche Teile – Verletzungsgefahr
Bei normalem Betrieb sind bewegliche Teile des Geräts zugänglich. Hände, Kleidung und andere Gegenstände von den beweglichen Teilen des Geräts fernhalten.
0993-5255
1-31
Page 40
Safety/Regulatory Information
A

Warning markings on the rear panel

Furnace Gases
Gaz du Four
Ofengase
P
P
min
max
350
400
kPa
3.5
4.0
bar
Ar / N
50.0
58.0
psig
2
Special Gas
Normal Ga s
Gaz Spécial
Gaz Normal
Spezialgas
Normalg as
Burner Gases Gaz du Bruleur Brennergase
P
P
minPminPmin
maxPmaxPmax
kPa
100
90
450
500
kPa
450
500
kPa
4.5
5.0
bar
bar
1.0
0.9
4.5
5.0
bar
65.0
72.5
psig
psig
14.5
13.0
65.0
72.5
psig
CH22NO
Air / Luft
2
HO
Furnace Cooling
Refroidissement du Four
Ofenkühl un g
Instrument
Communic ati on
230 V 50/60 Hz
230 V 50/60 Hz
Cooling System Système de Refroidissement Umlaufkühlung
InIn EntréeEntrée EingangEingang
P
P
PP
min
2
max
250
350
kPa
2.5
3.5
bar
36
50
psig
0
OutOut
SortieSortie
AusgangAu sgang
230 V
230 V
50/60 Hz
50/60 Hz
F4 4 A (T)F34 A (T)F24 A (T)F14 A (T)
I
LR 45111SA***
Warning
ll field wiring must have insula tion
suitable for at least 300 V
Avertissement
Le cablage à pied d 'ou vre doi t êtr e certifiés pour un mini m um de 30 0 V
D
F3 4 A (T)
230 V 50/60 Hz
F4 4 A (T)
230 V
50/60 Hz
F1
230 V
50/60 Hz
4 A (T)
E
Coo ling Sy stem Système de Refroidissement Umlaufkühlung
F2 4 A (T)
230 V 50/60 Hz
D
Power outlets
Power outlets for a printer, computer, or monitor. Only connect a printer, a computer, or a monitor to the respective outlet when the
instrument requires 230 V, 50/60 Hz.
E
Cooling system outlet
Connect the cooling system to this power outlet only. Do not connect the cooling system to any other supply. Do not connect other equipment to this power outlet.
1-32
0993-5255
Page 41
Safety/Regulatory Information
A
A
Burner Gases
Gaz du Bruleur
Brennergase
Furnace Cool ing
Refroidissement du Four
Ofenkühlung
P
P
min
Instrument
Communication
230 V
230 V
230 V
50/60 Hz
50/60 Hz
50/60 Hz
F4
230 V 50/60 Hz
Cooling Sy st em Système de Re froi diss emen t Umlaufkühl ung
0
OutOut
InIn
SortieSortie
EntréeEntrée
AusgangAusgang
EingangEingang
HO
P
PP
P
min
max
2
350
kPa
250
3.5
bar
2.5 50
psig
36
4 A (T)F34 A (T)F24 A (T)F14 A (T)
I
LR 45111SA***
Warning
ll field wiring must have insulation suitable for at least 300 V
Avertissement
Le cablage à pied d'ouvre doit être certifiés pour un minimum de 300 V
Furnace Gases Gaz du Four
Ofengase
P
P
min
max
350
400
kPa
3.5
4.0
bar
Ar / N
2
50.0
58.0
psig
Special Gas
Normal Gas Gaz Normal
Gaz Spécial Spezialgas
Normalgas
Burner Gases Gaz du Bruleur Brennergas e
P
P
minPminPmin
maxPmaxPmax
90
100
kPa
450
500
kPa
450
500
kPa
4.5
5.0
bar
0.9
1.0
bar
4.5
5.0
bar
65.0
72.5
psig
13.0
14.5
psig
65.0
72.5
psig
CH22NO
Air / Luft
2
F
450
4.5
65.0
max
500
5.0
72.5
ir / Luft
kPa bar psig
P
minPmax
450
4.5
65.0
NO
500
5.0
72.5
2
kPa bar psig
P
minPmax
90
0.9
13.0
CH
100
1.0
14.5
22
kPa bar psig
F
Burner Gases Never set the outlet pressure to a value exceeding the maximum pressure (P
Warning: Acetylene – Explosion Hazard
Acetylene can decompose explosively at pressures higher than 103 kPa / 1.03 bar / 15 psig. Always make sure that the acetylene outlet gauge pressure is below this value.
Danger: Acétylène – Risque d'explosion
L'acétylène peut se décomposer de manière explosive à des pressions supérieures à 103 kPa / 1.03 bar / 15 psig. Toujours vérifier que la pression de l'acétylène sur le manomètre de sortie est inférieure à cette valeur.
Warnung: Acetylen – Explosionsgefahr
Stellen Sie für Acetylen keinen höheren Ausgangsdruck als 103 kPa (1,03 bar) ein. Bei höheren Drücken kann Acetylen spontan zerfallen und explodieren
max
).
Warning: Flashback Hazard
PerkinElmer burner systems are designed for use with compressed air. The use of oxygen can cause an explosion in the burner system, and oxygen-enriched air can cause a flashback of the flame. Never use oxygen or oxygen-enriched air with PerkinElmer burner systems. Use only compressed air.
Danger: Risque de retour de flamme
Les systèmes de brûleurs PerkinElmer sont conçus pour être utilisés avec de l'air comprimé. L'utilisation d'oxygène peut provoquer une explosion dans le système de brûleur, et l'air suroxygéné peut provoquer un retour de flamme. Ne jamais utiliser d'oxygène ou d'air suroxygéné avec les systèmes de brûleurs PerkinElmer. N'utiliser que l'air comprimé.
Warnung: Sauerstoff und mit Sauerstoff angereicherte Luft
Verwenden Sie für PerkinElmer Brenner nur Druckluft. Sauerstoff oder sogenannte ‘Atemluft’ sind nicht zulässig. Bei Verwendung von Sauerstoff besteht Explosionsgefahr. Mit Sauerstoff angereicherte Luft kann einen Flammenrückschlag verursachen.
0993-5255
1-33
Page 42
Safety/Regulatory Information
A
HO
Furnace Cooling
Refroidissement du Four
Ofenkühl un g
Instrument
Communic ati on
230 V 50/60 Hz
230 V 50/60 Hz
Cooling System Système de R ef roid i ssem en t Umlaufkühlung
0
InIn
OutOut
EntréeEntrée
SortieSortie
EingangEingang
AusgangAu sgang
P
P
PP
min
max
2
250
350
kPa
2.5
3.5
bar
36
50
psig
230 V
230 V
50/60 Hz
50/60 Hz
F4 4 A (T)F34 A (T)F24 A (T)F14 A (T)
I
Warning LR 45111SA***
ll field wiring must hav e ins ula tion suitable for at least 300 V
Avertissement
Le cablage à pied d'ouvre doi t être
certifiés pour un mi nim um de 30 0 V
Furnace Gases
Gaz du Four
Ofengase
P
min
P
max
350
400
kPa
3.5
4.0
bar
Ar / N
2
50.0
58.0
psig
Special Gas
Normal Ga s
Gaz Spécial
Gaz Normal
Spezialgas
Normalg as
Burner Gases Gaz du Bruleur Brennergase
P
P
minPminPmin
maxPmaxPmax
kPa
kPa
100
500
kPa
90
450
450
500
bar
bar
1.0
5.0
bar
0.9
4.5
4.5
5.0 psig
psig
14.5
72.5
psig
13.0
65.0
65.0
72.5
CH22NO
Air / Luf t
2
Warning Disconnect supply before servicing Avertissement Couper l’alimentation avant l’entretien
Warning All field wiring must have insulation suitable for at least 300 V Avertissement Le cablage à pied d’ouvre doit être certifiés pour un minimum de 300 V
1-34
0993-5255
Page 43
Safety/Regulatory Information

Warning markings on the deuterium lamp cover

The warning markings below are visible when you open the top cover of the spectrometer.
D
2
High Temperatures – Risk of Burns
The lamp is hot.
Températures élevées
La lampe est chaude.
Verbrennungsgefahr
Die Lampe ist heiß.
UV
0993-5255
UV Radiation – Risk of Eye Damage
The lamp emits intense UV radiation which can damage your eyes. Do not gaze into a lighted lamp.
Rayonnement UV
La lampe émet un rayonnement UV intense susceptible de provoquer des lésions oculaires. Ne pas regarder fixement une lampe allumée.
UV-Strahlung – Gefährdung der Augen
Die Lampe emittiert starke UV-Strahlung und kann dadurch Augenschäden verursachen. Nicht mit ungeschützten Augen direkt in die leuchtende Lampe schauen.
Electrical Hazard
The deuterium lamp operates at high voltage. To avoid electrical shock, disconnect the line power cord before opening the cover.
Risque d’électrocution
Pout éviter les chocs électriques, débrancher le cordon d’alimentation avant d’ouvrir le couvercle.
Gefährdung durch elektrischen Strom
Die Deuteriumlampe benötigt zum Betrieb gefährlich hohe Spannung. Um elektrischen Stromschlag zu vermeiden, ziehen Sie die Netzanschlußleitung bevor Sie die Abdeckung öffnen.
1-35
Page 44
Safety/Regulatory Information
1-36
0993-5255
Page 45

Spectrometer Installation

2Spectrometer Installation
Marker for header
2
Contents
Before you install the spectrometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Moving and reinstalling the spectrometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Preparing the spectrometer to be moved . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Moving the spectrometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Reinstalling the spectrometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Connecting the burner gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Connecting the furnace gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
To connect the drain system to the burner . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
To install the furnace autosampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
Connecting cooling sources. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
Electrical connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
page
0993-5255
2-1
Page 46
Spectrometer Installation
2-2
0993-5255
Page 47

Before you install the spectrometer

When the spectrometer arrives
Check the exterior of the shipping crate for obvious signs of damage. If the
crate has been damaged, file a claim with the authorized carrier immediately and inform your nearest PerkinElmer office.
Contact your local PerkinElmer office to arrange for the installation.
Before the service engineer arrives
You must prepare a suitable site for the spectrometer and any accessories.
See Chapter 6, Laboratory Requirements, and the guides for any accessories.
You must make sure that the fume ventilation system and the necessary gas
supply lines and regulators are installed.
Although you should not perform the initial installation yourself, we
recommend that you read the installation instructions to learn about the spectrometer and its requirements before you start using it. These instructions are also provided in case you need to alter any of the connections or move the instrument.
Spectrometer Installation
0993-5255
When the service engineer arrives
The service engineer will remove the spectrometer from the crate and check
that all parts have been delivered. If anything is missing or damaged, the service engineer will make sure that the respective parts are supplied.
The service engineer will remove the shipping locks, perform pre-installation
checks, install the spectrometer, and make sure that everything functions correctly.
2-3
Page 48
Spectrometer Installation

Moving and reinstalling the spectrometer

We recommend that only a PerkinElmer service engineer or a PerkinElmer trained and authorized person moves and reinstalls the spectrometer. Should you attempt to do this yourself, PerkinElmer cannot be held responsible for any damage or malfunctioning of the spectrometer. The instructions here are primarily intended for occasions where you may want to move the spectrometer Read the instructions carefully and be sure you understand them thoroughly before you attempt to move the spectrometer. Read the detailed information given in the relevant sections of this chapter before you reinstall the spectrometer. If you intend to transport the spectrometer over a longer distance, contact your local PerkinElmer office since internal shipping locks must be set via the service software, inside the spectrometer.

Preparing the spectrometer to be moved

1. Close the AA WinLab application, then shut down the Windows operating system.
short distances to a new location within the laboratory.
2-4
2. Switch off the spectrometer, computer and all the accessories.
3. Disconnect all the electrical cables.
4. Remove the furnace autosampler from the spectrometer (see ‘Removing the autosampler from the spectrometer’ in the Graphite Furnace User’s Guide).
5. Disconnect the burner gas supplies.
6. Disconnect the drain system from the burner.
7. Disconnect the furnace gas supplies.
8. Disconnect the cooling water supply.
9. Prepare the computer and other system components for transport as described in their own guides.
10. Attach the original handles securely. Only use the original handles and screws.
Note: Make sure that the handles are properly secure before you move the spectrometer.
0993-5255
Page 49

Moving the spectrometer

Spectrometer Installation
Warning: Risk of injury through lifting
W1.4
The instrument weighs approximately 150 kg. Improper lifting can cause painful and sometimes permanent back injury.
Use proper lifting techniques and always use the handles to lift or move the
instrument.
Caution: Risk of damage to the instrument
The spectrometer is a precision optical instrument. Move the spectrometer very carefully and protect it from vibrations and jolts. Do not tilt the spectrometer. If the spectrometer must be tilted or mechanically lifted, contact your PerkinElmer service representative before moving the spectrometer. The shipping locks must be properly secured to prevent any damage occurring if the spectrometer is tilted or jolted.
Carefully lift the spectrometer onto a movable table to transport it to the new location.
0993-5255
2-5
Page 50
Spectrometer Installation

Reinstalling the spectrometer

Reinstall the spectrometer by following exactly the procedures described in this chapter, as summarized in the steps below:
1. Prepare the new site for the spectrometer; see Chapter 6, Laboratory Requirements. Make sure that there will be enough space at the rear and sides of the spectrometer system for air to circulate freely.
2. Remove the handles. Save the handles and screws. Insert the plastic plugs provided (B015-3803) into the screw holes.
3. Connect the burner and furnace gas supplies.
4. Connect the drain system to the burner.
5. Install the furnace autosampler.
6. Connect the cooling water supply.
7. Make sure that the main power switches of all the instruments are in the off position, then connect the electrical cables.
2-6
0993-5255
Page 51

Connecting the burner gases

The connectors for the burner gases are located on the rear panel of the spectrometer.
What you need to provide:
Suitable supplies of acetylene, air, and nitrous oxide. The gases must conform
with the requirements listed in Chapter 6, Laboratory Requirements.
Suitable pressure regulators and on/off valves near to the spectrometer for the
gas supplies.
The gas supply systems must comply with your local safety regulations. Refer also to the information given in Chapter 1, Safety and Regulatory Information.
Spectrometer Installation
Warning: Flashback Hazard
PerkinElmer burner systems are designed for use with compressed air. The use of oxygen can cause an explosion in the burner system, and oxygen-enriched air can cause a flashback of the flame.
Never use oxygen or oxygen-enriched air with PerkinElmer burner systems.
Use only compressed air.
W5.9
What is provided with the spectrometer:
Gas hose assemblies for acetylene, air, and nitrous oxide.
The fittings on the gas hose assemblies are suitable for connection to most commercially available pressure regulators. These are R 3/8'' standard threaded fittings. The fuel (acetylene) fitting has a left-hand thread, while the oxidant (air, nitrous oxide) fittings have right-hand threads. If you require an adapter for your regulator, contact the gas supplier or the regulator manufacturer, or your local PerkinElmer office.
0993-5255
2-7
Page 52
Spectrometer Installation
To connect the gases
Refer to Figure 2-1 when you connect the burner gases.
Air
1. Make sure that the air supply is turned off.
2. Use the black hose assembly to connect a source of clean, dry, and filtered air to the fitting labelled
Nitrous oxide
1. Make sure that the nitrous oxide supply is turned off.
Air/Luft on the rear panel of the spectrometer.
2. Use the blue N oxide to the fitting labelled
Acetylene
O hose assembly to connect a source of clean, dry nitrous
2
N2O on the rear panel of the spectrometer.
1. Make sure that the acetylene supply is turned off.
2. Use the red C the fitting labelled
hose assembly to connect a source of clean, dry acetylene to
2H2
C2H2 on the rear panel of the spectrometer.
2-8
0993-5255
Page 53
Furnace Cooling
A
Refroidissement du Four
Ofenkühlung
Spectrometer Installation
Instrument
Communication
230 V
230 V
230 V
50/60 Hz
50/60 Hz
50/60 Hz
F4
230 V 50/60 Hz
Cooling System Système de Refroidissement Umlaufkühlung
InIn EntréeEntrée EingangEingang
HO
P
PP
P
min
max
2
350
250
3.5
2.5 50
36
0
OutOut SortieSortie
AusgangAusgang
kPa bar psig
4 A (T)F34 A (T)F24 A (T)F14 A (T)
I
Warning
LR 45111SA***
ll field wiring must have i nsulation suitable for at least 300 V
Avertissement
Le cablage à pied d'ouv re doit être certifiés pour un minimum de 300 V
Furnace Gases
Gaz du Four Ofengase
P
P
min
max
350
400
kPa
3.5
4.0
bar
Ar / N
2
50.0
58.0
psig
Normal Gas
Speci al G as Gaz Spécial
Gaz Normal
Spezialgas
Normalgas
Burner Gases Gaz du Bruleur Brennergase
P
P
minPminPmin
maxPmaxPmax
90
100
kPa
450
500
kPa
450
500
kPa
4.5
5.0
bar
0.9
1.0
bar
4.5
5.0
bar
65.0
72.5
psig
13.0
14.5
psig
65.0
72.5
psig
CH22NO
Air / Luft
2
Air
Nitrous oxide: N2O
Acetylene: C2H
Figure 2-1. Connecting the Burner Gases
Air/Luft
NO
2
CH
22
Black (B015-4987)
Blue (B015-4990)
Red (B015-4984)
2
0993-5255
2-9
Page 54
Spectrometer Installation

Connecting the furnace gases

The connectors for the furnace gases are located on the rear panel of the spectrometer.
Note: The furnace will operate correctly only if there is gas pressure at both the Normal Gas and Special Gas inlets. If you are using an inert gas only you must connect the gas to both inlets using the Y-connector.
What you need to provide:
A suitable supply of argon (or nitrogen) and if necessary a suitable supply of a special gas. The gases must conform with the requirements listed in Chapter 6, Laboratory Requirements.
Suitable pressure regulators and on/off valves near to the spectrometer for the
inert and special gas supplies.
A tube cutter for the special gas tube. This is an optional accessory; Part No.
0992-3183.
The gas supply systems must comply with your local safety regulations. Refer also to the information given under Chapter 1, Safety and Regulatory Information.
2-10
What is provided with the spectrometer:
About 10 m of special gas tube, B050-8264.
A Y-connector with automatic push-fit connectors, 0992-0179.
Two screw connectors, R ¼'' for the gas regulator. These are not supplied with
all instrument versions.
0993-5255
Page 55
Spectrometer Installation
To connect argon (or nitrogen) only
Refer to Figure 2-2 on the following page when you connect the furnace gas.
1. Make sure that the argon (or nitrogen) supply is turned off.
2. Use the special tube cutter to cut two, 10 cm lengths of tube from the special gas tube. The ends of the tubes must be cut squarely.
3. Connect the long gas tube to the argon (or nitrogen) pressure regulator. Use the R ¼'' screw connector provided or use a connector or adapter suitable for the gas regulator. Make sure that all the connections are secure.
4. Pass the gas tube to the rear of the spectrometer. Make sure that it cannot be damaged or strained.
5. Push the two short tubes and the long tube into the connectors on the Y­connector as shown on the following page.
6. Push the free ends of the two short gas tubes into the
connectors on the manifold as shown.
Gas
Normal Gas and Special
Push the tubes into the connectors until the tubes butt up to the end of the connector.
Note: To disconnect the gas tube from a connector: Carefully push the black locking ring of the connector into the connector and pull the tube out of the connector.
0993-5255
2-11
Page 56
Spectrometer Installation
A
Instrument
Communication
230 V 50/60 Hz
230 V 50/60 Hz
Cooling System Système de Refroidissement Umlaufkühlung
I
0
OutOut SortieSortie
AusgangAusgang
kPa bar psig
Furnace Cooling
Refroidissement du Four
Ofenkühlung
InIn EntréeEntrée EingangEingang
HO
P
PP
P
min
max
2
250
350
2.5
3.5
36
50
Black locking ring
Black locking ring
Furnace Gases
Gaz du Four
Ofengase
P
P
min
230 V
230 V
50/60 Hz
50/60 Hz
F4 4 A (T)F34 A (T)F24 A (T)F14 A (T)
Warning
LR 45111SA***
ll field wiring must have i nsulation suitable for at least 300 V
Avertissement
Le cablage à pied d'ouv re doit être certifiés pour un minimum de 300 V
max
350
400
kPa
3.5
4.0
bar
Ar / N
2
50.0
58.0
psig
Speci al G as
Normal Gas Gaz Normal
Gaz Spécial
Normalgas
Spezialgas
Burner Gases Gaz du Bruleur Brennergase
P
P
minPminPmin
maxPmaxPmax
90
100
kPa
450
500
kPa
450
500
kPa
4.5
5.0
bar
0.9
1.0
bar
4.5
5.0
bar
65.0
72.5
psig
13.0
14.5
psig
65.0
72.5
psig
CH22NO
Air / Luft
2
Y-connector
Argon
(or nitrogen)
Figure 2-2. Connecting Argon (or Nitrogen) only
2-12
0993-5255
Page 57
Spectrometer Installation
To connect both argon and a special gas
Refer to Figure 2-3 on the following page when you connect the furnace gases.
1. Make sure that the gas supplies are turned off.
2. Use the special tube cutter to cut the special gas tube into two pieces for the two gas supplies. The ends of the tubes must be cut squarely.
3. Connect the gas tubes to the argon (or nitrogen) and special gas pressure regulators. Use the R ¼'' screw connectors provided or use a connector or adapter suitable for the gas regulator. Make sure that all the connections are secure.
4. Pass the gas tubes to the rear of the spectrometer. Make sure that they cannot be damaged or strained.
5. Push the free ends of the gas supply tubes into the
Gas
connectors on the manifold as shown. Push the tubes into the connectors until the tubes butt up to the end of the connector.
Note: To disconnect the gas tube from a connector: Carefully push the black locking ring of the connector into the connector and pull the tube out of the connector.
Normal Gas and Special
0993-5255
2-13
Page 58
Spectrometer Installation
A
Instrument
Communication
230 V 50/60 Hz
230 V 50/60 Hz
Cooling System Système de Refroidissement Umlaufkühlung
I
0
OutOut SortieSortie AusgangAusgang
kPa bar psig
Furnace Cooling
Refroidissement du Four
Ofenkühlung
InIn EntréeEntrée EingangEingang
HO
P
PP
P
min
max
2
350
250
3.5
2.5 50
36
Black locking ring
Furnace Gases
Gaz du Four
Ofengase
P
P
min
230 V
230 V
50/60 Hz
50/60 Hz
F4 4 A (T)F34 A (T)F24 A (T)F14 A (T)
Warning
LR 45111SA***
ll field wiring must have i nsulation suitable for at least 300 V
Avertissement
Le cablage à pied d'ouv re doit être certifiés pour un minimum de 300 V
max
350
400
kPa
3.5
4.0
bar
Ar / N
2
50.0
58.0
psig
Speci al G as
Normal Gas Gaz Normal
Gaz Spécial
Normalgas
Spezialgas
Burner Gases Gaz du Bruleur Brennergase
P
P
minPminPmin
maxPmaxPmax
90
100
kPa
450
500
kPa
450
500
kPa
4.5
5.0
0.9
bar
1.0
bar
4.5
5.0
bar
65.0
72.5
13.0
psig
14.5
psig
65.0
72.5
psig
CH22NO
Air / Luft
2
Special gas
Argon
(or nitrogen)
Figure 2-3. Connecting Argon (or Nitrogen) and a Special Gas
2-14
0993-5255
Page 59

To connect the drain system to the burner

Procedures for connecting the drain system to the burner are described in Maintaining the drain system in the Burner System User’s Guide.

To install the furnace autosampler

Procedures for demounting and reinstalling the furnace autosampler are described in Installing the furnace autosampler in the Graphite Furnace User’s Guide.
Spectrometer Installation
0993-5255
2-15
Page 60
Spectrometer Installation

Connecting cooling sources

Installing the cooling system
PerkinElmer offers a recirculating cooling system as an option for the spectrometer system. This cooling system includes a pressure-sensitive safety device that switches off the pump if the internal pressure in the circulatory system increases to above a safe value (see ‘Cooling the furnace’ in the Graphite Furnace User’s Guide).
Connecting and disconnecting the water hoses
The inlet and outlet hoses are permanently attached to the cooling system. They have automatic connectors which are marked with the direction of the water flow.
1. Place the cooling system in a convenient position
on the floor underneath
the spectrometer. Do not place the cooling system on the same bench as the spectrometer since vibrations can adversely influence the spectrometer.
Make sure that:
You can reach the on/off switch at the rear.
There is space at the rear and sides of the cooling system for air to
circulate freely.
There is space to remove the stopper and fill the system.
You can see the water level indicator.
2. Pass the two hoses from the cooling system to the rear of the spectrometer. Make sure they are not strained.
3. Push the hose connectors onto the correct fittings on the spectrometer as shown on the following page. The connectors are spring loaded. Push them straight onto the fittings until they click into place.
Note: To disconnect the water hoses: Carefully pull the knurled locking ring down slightly, to release the lock, and pull the connector off the fitting on the spectrometer.
2-16
0993-5255
Page 61
Spectrometer Installation
A
A
Locating the cooling system
Connecting the hoses
WATER IN
Furnace Cooling
Refroidissement du Four
Ofenkühlung
InIn EntréeEntrée
EingangEingang
HO
P
P
min
2
250
2.5 36
Cooling System
Instrument
Communication
230 V 50/60 Hz
230 V 50/60 Hz
Cooling System Système de Refroidissement Umlaufkühlung
I
0
OutOut SortieSortie AusgangAusgang
PP
max
350
kPa
3.5
bar
50
psig
WATER OUT
Furnace Gases Gaz du Four
Ofengase
P
P
min
max
350
400
230 V
230 V
50/60 Hz
50/60 Hz
F4 4 A (T)F34 A (T)F24 A (T)F14 A (T)
LR 45111SA***
Warning
ll field wiring must have insulation suitable for at least 300 V
LR 45111SA***
Warning
Disconnect supply before servicing
vertissement
Le cablage à pied d'ouvre doit être
Avertissement
certifiés pour un minimum de 300 V
Couper l'alimentation avant l'entretien
kPa
3.5
4.0
bar
Ar / N
2
50.0
58.0
psig
Specia l G as
Normal Gas
Gaz Spécial
Gaz Normal
Spezialgas
Normalgas
Burner Gases Gaz du Bru leu r Brennergas e
P
P
minPminPmin
maxPmaxPmax
450
500
kPa
90
100
kPa
450
500
kPa
4.5
5.0
bar
0.9
1.0
bar
4.5
5.0
bar
65.0
72.5
psig
13.0
14.5
psig
65.0
72.5
psig
CH22NO
Air / Luft
2
Figure 2-4. Connecting the Cooling System
0993-5255
2-17
Page 62
Spectrometer Installation
Filling the cooling system
The cooling system is self-priming; air in the system is removed automatically as the cooling system operates. The coolant is a 1:10 solution of glycerol in water (1 part glycerol plus 9 parts water).
Note: The cooling system contains 0.5 L glycerol when it leaves the factory. (See ‘Cooling system maintenance procedures’ in the Graphite Furnace User’s Guide for the procedure to follow when you empty and refill the cooling system. Do not use antifreeze.)
.
1. Remove the stopper from the top of the cooling system – twist and pull.
2. Fill the system to the sediment-free drinking water – about
4.5 L. Do not use deionized water since this can cause corrosion.
3. Replace the stopper securely.
Max mark with
Max
Min
2-18
Water level indicator
0993-5255
Page 63
Spectrometer Installation
Connecting external cooling sources
If you intend to run the spectrometer system without the PerkinElmer cooling system, you must connect another suitable source of cooling, such as the main water supply or a laboratory thermostat. If you connect the spectrometer system to the main water supply, make sure that the water pressure is as given in Table 2-1.
Pressure relief valve
Drain hose
Flow indicator
0993-5255
Water In
Water Out
Figure 2-5. Connecting the Main Water Supply
A water hose kit with pressure relief valve is provided with the spectrometer system which you must install in the water supply line to the spectrometer. This valve opens if the water pressure to the furnace system exceeds 4 bar. This feature eliminates the risk of damage to the furnace system through excess water pressure.
The cooling water must conform with the requirements listed in ‘External cooling’ on page 6-16.
2-19
Page 64
Spectrometer Installation
Caution Water Pressure – Leaking hose fittings or burst water hoses
If the water inlet pressure is too high, hose fittings may leak or hoses may even burst.
For reliable and safe operation, you must install the pressure relief valve kit
B313-0124 in the water supply line.
The hoses provided in the kit are fitted with automatic connectors that incorporate stop valves; these valves prevent water from running out when the hoses are dis­connected. The hoses are marked with the direction of the water flow.
Table 2-1. Pressure for the Cooling Water Supply
Outlet Pressure
P
min
P
max
250 350 kPa
2.5 3.5 bar
36 50 psig
Check the water pressure to your laboratory and make sure that the water inlet
pressure is set to a value between the minimum value (P maximum value (P
max
).
) and the
min
Make sure that the outlet pressure is never set to a value exceeding the
maximum pressure (P
1. The PerkinElmer service engineer will connect the pressure relief valve to the water tap (faucet).
2. Make sure that the drain hose from the pressure relief valve hangs down into a suitable laboratory sink or drain.
3. Connect the inlet hose (marked pressure relief valve.
4. Connect the other end of the inlet hose to the
In on the spectrometer.
The connectors are spring loaded. Push them straight onto the fittings until they click into place.
max
).
WATER IN) to the fitting at the bottom of the
Furnace Cooling fitting marked
2-20
0993-5255
Page 65
Spectrometer Installation
A
A
5. Connect the outlet hose (marked WATER OUT) to the Furnace Cooling fitting marked
Out on the spectrometer.
Note: To disconnect the water hoses: Carefully pull the knurled locking ring down slightly, to release the lock, and pull the connector off the fitting on the spectrometer.
6. We recommend that you fit a proprietary flow indicator in the outlet hose so that you can easily monitor that cooling water is flowing.
7. Place the free end of the outlet hose in a suitable laboratory sink or drain. Make sure that the hose cannot slip out.
Pressure relief valve
HO
2
Furnace Cooling
Refroidissement du Four
Ofenkühlung
Connect the ‘Water In’ hose to this fitting
Water In Water Out
Figure 2-6. Connecting Cooling Water Hoses
Instrument
Communication
230 V
230 V
230 V
50/60 Hz
50/60 Hz
50/60 Hz
F4
230 V 50/60 Hz
Cooling System Système de Refroidissement Umlaufkühlung
InIn EntréeEntrée EingangEingang
P
P
PP
min
max
350
250
3.5
2.5 50
36
0
OutOut SortieSortie AusgangAusgang
kPa bar psig
4 A (T)F34 A (T)F24 A (T)F14 A (T)
I
LR 45111SA***
Warning
ll field wiring must have insulation suitable for at least 300 V
LR 45111SA***
Warning
Disconnect supply before servicing
vertissement
Le cablage à pied d'ouvre doit être
Avertissement
certifiés pour un minimum de 300 V
Couper l'alimentation avant l'entretien
Furnace Gases
Gaz du Four
Ofengase
P
P
min
max
350
400
kPa
3.5
4.0
bar
Ar / N
2
50.0
58.0
psig
Special Ga s
Normal Gas
Gaz Spécial
Gaz Normal
Spezialgas
Normalgas
Burner Gases
Gaz du Bruleur
Brennergase
P
P
minPminPmin
maxPmaxPmax
90
450
500
450
100
kPa
500
kPa
kPa
0.9
4.5
5.0
4.5
1.0
bar
5.0
bar
bar
13.0
65.0
72.5
65.0
14.5
psig
72.5
psig
psig
CH22NO
Air / Luft
2
0993-5255
2-21
Page 66
Spectrometer Installation

Electrical connections

Warning: Electrical Hazard
To prevent potential injury to yourself and damage to the instrument, switch
OFF all instruments in the system and disconnect them from the line power supply before you perform the procedures described below.
Read the following instructions very carefully before connecting any cables.
See Chapter 6, Laboratory Requirements for the specification of the line power supply.
Summary
1. Make sure that the spectrometer and the cooling system (where applicable) are switched off.
2. Switch the circuit breaker on the rear panel of the spectrometer to the off position (marked ).
3. Switch off the computer and all other instruments that you will connect to the spectrometer or computer. See the guides for the individual instruments for the procedure to use.
4. Disconnect the line power cords of the spectrometer and all other instruments from the line power supply.
5. Connect the instrument control cables and power cords as shown on the following page.
6. When you have completed all connections, switch on the system (see ‘Switching on the system’, page 3-3).
W2.3
2-22
0993-5255
Page 67
Computer
Connections panel on the rear of the spectrometer
Instrument
Communication
230 V
230 V
50/60 Hz
50/60 Hz
I
0
Monitor
F2 4 A (T)
230 V 50/60 Hz
Cooling System Système de R efroidiss ement Umlaufkühlung
F1 4 A (T)
Spectrometer Installation
230 V 50/60 Hz
F3 4 A (T)
F4 4 A (T)
Use these power outlets only if the computer and printer require 230 V
Spectrometer line power cord
Communication cable
Printer
0993-5255
Cooling system
Connect to the outlets at the rear of the spectrometer or to a suitable line power supply that uses the same protective ground (earth) as the line power supply to the spectrometer
Figure 2-7. Electrical Connections
Communication interface
2-23
Page 68
Spectrometer Installation
2-24
0993-5255
Page 69

Preparing the System for Analyses

3Preparing the System for Analyses
Marker for header
3
Contents
Switching on the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Installing lamps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Installing sample trays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Shutting down the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Emergency shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
page
0993-5255
3-1
Page 70
Preparing the System
3-2
0993-5255
Page 71

Switching on the system

1. Read the safety information at the front of this guide before you operate the system.
2. Make sure that the spectrometer system and other ancillary instruments are correctly installed and connected to their power supplies. Make sure that the coolant hoses are connected correctly to the spectrometer. Where applicable, make sure that the cooling system is filled to the mark (see ‘Filling the cooling system’, page 2-18).
3. Switch on the fume ventilation system for the work area.
4. Turn on the burner gases and adjust the outlet gauge pressures to the recommended values (see Table 3-1 on page 3-5). Always set the same outlet gauge pressure for air and nitrous oxide.
5. Turn on the inert (and special) gas for the furnace and adjust the outlet gauge pressures to the recommended values (see Table 3-2 on page 3-6).
Note: You must always turn on the air supply since air pressure is required for the internal pneumatic system. If you intend to work with only one atomizer, you only need to additionally turn on the gas supplies for that atomizer.
6. Turn on the cooling water to the system:
1. If you are using the circulatory cooling system, switch it on by pressing
the power switch located on the rear panel.
You can leave this switch permanently in the on position since power is supplied by the spectrometer.
2. If you are using the main water supply, set a flowrate of approximately
2.5 L/min.
Note: If the water is too cold or flowing too fast, moisture may condense on the graphite contacts or on the furnace windows. Moisture on the graphite contacts leads to shorter lifetimes of the graphite tube and contacts, while moisture on the windows leads to a reduction in the radiant energy.
7. Switch the circuit breaker on the rear panel of the spectrometer to the on position (marked ).
You can leave this switch permanently in the on position during routine operation; you switch the spectrometer system on and off via the operational on/off switch located on the front panel.
Preparing the System
Max
0993-5255
3-3
Page 72
Preparing the System
8. Switch on the computer.
9. Install a Lumina hollow cathode lamp in any holder in the lamp compartment
10. Switch on the spectrometer system with the operational on/off switch located
11. If you need the diluter, autosampler, or flow injection system, switch on the
12. If you have changed the configuration of the system, such as installed a
13. Start AA WinLab: In the AA WinLab group, double click on the AA WinLab
14. Set up the various parts of the system:
If your computer has a power management system, switch this off.
(procedure: page 3-7).
A hollow cathode lamp must be installed in at least one holder before you switch on the spectrometer.
on the front panel.
respective systems.
different autosampler or added or removed any accessories, use the
Reconfigure utility to reconfigure the software (see the Online Help in AA
WinLab).
icon.
1. Install any further lamps that you require for the analyses
(procedure: page 3-7).
2. Set up the burner system
(see ‘Setting up the system for analyses using the flame technique’ in the Burner System User’s Guide).
3. Set up the graphite furnace system
(see ‘Setting up the system for analyses using the furnace technique’ in the Graphite Furnace User’s Guide).
3-4
Procedures for shutting down the system are described on page 3-12. In an emergency, you can shut down the system as described on page 3-14.
0993-5255
Page 73
Table 3-1. Required Pressures for the Burner Gases
Gas Outlet Gauge Pressure
kPa bar psig
Preparing the System
Acetylene (C
)
2H2
P
min
90 100 0.9 1.0 13.0 14.5
P
max
P
min
P
max
P
min
P
max
Air 450 500 4.5 5.0 65.0 72.5
Nitrous oxide (N
O)
2
450 500 4.5 5.0 65.0 72.5
Set the outlet gauge pressure for each gas to a value just below the maximum
value (P
max
).
Always set the same outlet gauge pressure for air and nitrous oxide.
Never set the outlet gauge pressure to a value exceeding the maximum
pressure (P
max
).
Always turn on the air supply since air pressure is required for the
spectrometer’s pneumatic system.
Warning: Acetylene – Explosion Hazard
Acetylene can decompose explosively at pressures higher than 103 kPa/
1.03 bar/15 psig.
Always make sure that the acetylene outlet gauge pressure is below this
value.
W5.7
0993-5255
3-5
Page 74
Preparing the System
Table 3-2. Required Pressures for the Furnace Gases
Gas Outlet Gauge Pressure
kPa bar psig
P
min
P
max
P
min
P
max
P
min
P
max
Argon 350 400 3.5 4.0 50.0 58.0
Nitrogen 350 400 3.5 4.0 50.0 58.0
Special 350 400 3.5 4.0 50.0 58.0
Set the outlet gauge pressure for each gas to a value just below the maximum
value (P
max
).
Never set the outlet gauge pressure to a value exceeding the maximum
pressure (P
Warning: Toxic Gases
If you use nitrogen as the inert gas, toxic cyanogen gas (CN)2 will be produced when the furnace temperature is above 2300 °C.
Make sure that the main laboratory fume ventilation system removes all
gases and fumes emitted by the furnace.
max
).
W6.1
3-6
0993-5255
Page 75

Installing lamps

Lamp compartment
The lamp compartment is located on the left-hand side of the spectrometer, beneath a hinged cover. The compartment is easily accessible from the front when the cover is open.
Lamp compartment cover
Preparing the System
0993-5255
The spectrometer is designed to operate with PerkinElmer Lumina hollow cathode lamps (HCLs) and coded electrodeless discharge lamps (EDLs).
The lamp compartment has holders for up to eight lamps. You can install Lumina HCLs in all eight holders. The holders incorporate the receptacles for connection of the Lumina lamps. You can install EDLs in holders 1–4 only. The sockets for the power supplies to the EDLs are located on the left-hand side of the lamp compartment.
3-7
Page 76
Preparing the System
3
2
4
5
6
7
8
Lamp holders numbered 1 through 8
Sockets for EDLs
Figure 3-1. Lamp Holders in the Lamp Compartment
Installing and connecting lamps
HCLs
Note: A hollow cathode lamp must be installed in at least one holder before you switch on the spectrometer.
1. Open the lamp compartment cover by lifting it at the front and swinging it back.
3-8
2. Slide the Lumina lamp for the required element into the selected lamp holder and make sure that the integrated socket slides fully home over the receptacle on the lamp holder.
3. Close the lamp compartment cover.
Lumina hollow cathode lamp
3
2
4
5
6
7
8
0993-5255
Page 77
Preparing the System
EDLs
You must use the lamp holders numbered 1 through 4 for EDLs.
1. Open the lamp compartment cover by lifting it at the front and swinging it back.
2. If necessary, assemble the lamp by sliding the lamp driver into the lamp sleeve for the required element. Depress the locking pin as the sleeve slides into place and make sure that the locking pin extends through the locating hole in the sleeve.
Lamp sleeve
Lamp driver
Locking pin
Coding plug
3. Slide the lamp carefully into the selected lamp holder until it butts up against the end-stop inside the holder.
4. Insert the coding plug over the receptacle on the lamp holder.
0993-5255
5. Insert the lamp connector into the correct EDL socket at the left-hand side of the lamp compartment. Use the receptacle that has the same number as the lamp holder.
6. Close the lamp compartment cover.
Coding plug
EDL socket
3
2
4
5
6
7
8
3-9
Page 78
Preparing the System
Removing lamps
Do not unplug and remove lamps while they are switched on.
1. On the
Toolbar, click on Lamps. The Align Lamps window appears.
2. Make sure that the lamps you want to remove are switched off: The
On buttons must not be green. If necessary, click on the respective On
buttons.
3. Close the Align Lamps window.
4. Unplug and remove the lamps. For EDLs, remove the coding plugs as well.
Tips for installing lamps
Use only PerkinElmer lamps. You can use both hollow cathode and
electrodeless discharge lamps.
Do not touch the front window of the lamp; perspiration or other
contamination can reduce the intensity of the radiation.
If you install more than one lamp containing a particular element, for
example, a single-element copper lamp and a multi-element lamp containing copper, the system will use the lamp in the holder with the lower number when you perform determinations of that element. Make sure that you place the preferred lamp in the holder with the lower number.
With EDLs, connect each lamp plug to the socket that has the same number
as the holder of the lamp.
With EDLs, make sure that you use the correct coding plug and connect it to
the receptacle on the lamp holder containing the EDL.
3-10
If you use the wrong coding plug or connect the coding plug to the wrong
receptacle, you will cause the spectrometer to malfunction and may cause irreparable damage to the lamp.
0993-5255
Page 79

Installing sample trays

A sample tray can be inserted into one of three locations in front of the spectrometer – on the left-hand side, in front of the atomizer compartment, and on the right-hand side. You can leave a tray permanently installed on the left-hand side if you wish. You cannot install a tray in front of the atomizer compartment or on the right-hand side as long as the furnace autosampler is installed. When the autosampler is swivelled to the standby position at the right, you can install a tray in front of the atomizer compartment.
The sample trays are robust and can support an accessory, such as a diluter, or a flame autosampler, or a flow injection system.
To install the tray
1. Hold the tray with both hands and position it in front of the location into which you wish to insert it.
2. Insert the tray support posts into the guides at the base of the instrument and slide in the tray until it hits the stop. Then gently lift the front of the tray and slide it finally into place. You can feel it lock into place.
Preparing the System
0993-5255
To remove the tray
1. Gently lift the tray at the front and then pull it back to slide it out of the guides.
3-11
Page 80
Preparing the System

Shutting down the system

Note: In an emergency, switch off the spectrometer using the operational on/off switch on the front panel. The flame is automatically extinguished and the gases shut down in a safe sequence. See page 3-14 for the procedure to follow after an emergency shutdown.
You must shut down each atomizer correctly before you shut down the spectrometer.
To shut down the flame system
1. With the flame still burning, aspirate the correct rinsing solutions to rinse the
burner system (see ‘Extinguishing the flame’ in the Burner System User’s Guide for detailed information). Either aspirate the solutions manually or use the flame autosampler.
2. Extinguish the flame and bleed the gas lines:
1. In the
2. Shut down the gases to the spectrometer at source.
3. In the
Flame Control window, click on the Flame icon.
The system uses a predetermined shutdown sequence to extinguish the flame safely.
Flame Control window, click on Bleed Gases to depressurize the gas
lines.
3-12
To shut down the furnace system
1. Flush the furnace autosampler rinsing system (see ‘Autosampler maintenance procedures’ in the Graphite Furnace User’s Guide for detailed information).
2. Shut down the furnace gases to the spectrometer at source.
0993-5255
Page 81
Preparing the System
To shut down the spectrometer
1. Exit AA WinLab: In the
File menu, click on Exit.
2. Switch off the spectrometer at the operational on/off switch on the front panel.
3. Switch off any accessories.
4. If you do not want to use the computer for other tasks, shut down the computer and printer as described in their user’s guides.
5. Empty the burner drain vessel (see ‘Emptying the drain vessel’ in the Burner System User’s Guide).
6. Empty the furnace autosampler waste bottle.
Note: Dispose of hazardous or corrosive solutions properly and refer to your local safety regulations for proper disposal procedures.
0993-5255
3-13
Page 82
Preparing the System

Emergency shutdown

If a situation arises where you have to shut down the system quickly and you do not have time to carry out the correct procedure, you can use the procedure described below.
Note: This procedure is for use in an emergency only. Do not use it as a routine shutdown procedure.
1. Switch off the spectrometer at the operational on/off switch on the front panel.
2. Shut down the gases to the spectrometer at source.
When you return to the spectrometer, proceed as follows to shut down the system properly:
3. If you were previously running concentrated salt solutions, organic solvents, or solutions containing cyanide, copper, silver, or mercury:
1. Remove the burner head (see the Burner System User’s Guide).
2. Pour a large quantity of deionized water slowly through the neck of the
spray chamber to thoroughly flush the chamber and the drain system.
4. Check the burner head. If there are deposits along the slot, clean the burner head (see the Burner
System User’s Guide).
3-14
5. Reinstall the burner head.
6. Switch on and set up the spectrometer system.
7. Ignite the flame.
8. Extinguish the flame properly.
0993-5255
Page 83

Spectrometer: System Maintenance

4Spectrometer: System Maintenance
Marker for header
Contents
PerkinElmer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Maintenance checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Spectrometer maintenance procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Cleaning the windows of the optical system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Maintaining the electrical components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Changing the air filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
Replacing the deuterium lamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
Maintaining the acetylene supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
4
page
0993-5255
4-1
Page 84
System Maintenance
4-2
0993-5255
Page 85

PerkinElmer Service

System Maintenance
Warning: Unauthorized Adjustments and Servicing
Only a PerkinElmer service engineer or similarly trained and authorized person should be permitted to service the instrument.
Do not attempt to make adjustments, replacements, repairs, or modifications
to this instrument except as described in the documentation supplied with the instrument.
W1.2
If the spectrometer system does not function correctly…
Switch off the spectrometer system and make sure that it cannot be operated.
Contact your local PerkinElmer office.
You should only perform the maintenance procedures described in this guide. For any other maintenance or service contact your local PerkinElmer office to arrange for a service engineer to visit. In particular you should only allow a PerkinElmer service engineer or similarly trained and authorized person to perform any work on:
electrical components inside the instrument;
the gas connections and gas system pneumatics;
the monochromator and all other optical system components.
0993-5255
Before the service engineer arrives:
1. Make sure that the instrument and work area are clean.
2. Empty corrosive solutions from the burner drain vessel. Then add water to the drain trap to activate the function.
3. Empty the furnace autosampler waste bottle.
Troubleshooting
See the Troubleshooting Guide supplied with the AA WinLab operating system. This guide includes information on problems that can arise with both the software and the spectrometer system.
4-3
Page 86
System Maintenance

Maintenance checklist

Maintenance checklists for the spectrometer system are presented below. Maintenance checklists for the burner system are presented in the Burner System User’s Guide and for the graphite furnace, cooling system, and the furnace autosampler in the Graphite Furnace User’s Guide.
General preventative maintenance procedures
Instrument and work area
For safety reasons, and to avoid contaminating new samples, you should make sure that the instrument and the work area are always absolutely clean. This is especially important when you are working with toxic substances and when you are measuring trace amounts of any elements. Wipe up spills immediately before they can cause further contamination or damage. Before you leave the instrument standing idle for any period of time, make sure that all the components are clean.
Fume ventilation
Environmen­tal conditions
Acetylene cylinder
Make sure that the main fume ventilation system functions correctly. Always switch it on when you are using the instrument.
Make sure that the environmental conditions are as described in Chapter 6, Laboratory Requirements.
Always change to a new acetylene cylinder when the cylinder pressure falls to below 600 kPa (6 bar; 87 psig) (more information: page 4-14).
Spectrometer, maintenance checklist
External surfaces
Wipe over the surfaces with a lint-free cloth moistened with a dilute solution of laboratory detergent. The surfaces are resistant to dilute acids and alkalis, and to a lesser extent to strong acids and alkalis and organic solvents.
Air filters Fit new filters yearly or more often in a dusty environment
(procedure: page 4-10).
Atomizer compartment windows, lamp end
These windows do not normally require cleaning but if they become marked, e.g., with fingerprints, you can clean them (procedure: page 4-5).
windows
4-4
0993-5255
Page 87

Spectrometer maintenance procedures

Procedures described
Cleaning the windows.
Changing the air filters.
Resetting the circuit breaker and changing the fuses.
Replacing the deuterium lamp.

Cleaning the windows of the optical system

Caution Risk of damage to optical components
Never touch or clean the surfaces of the grating or mirrors.
If mirror surfaces are scratched they must be resurfaced. Do not allow cleaning fluids to come into contact with any of the reflecting surfaces.
System Maintenance
0993-5255
Most of the optical components are under the main cover of the spectrometer and are protected by the black optics covers. Usually these components require no maintenance, provided the laboratory fume ventilation system is working correctly and the environmental conditions are as described in Chapter 6, Laboratory Requirements.
The only optical parts that are exposed are:
The atomizer compartment windows.
The windows of the lamps.
The furnace windows. Cleaning the furnace windows is described under
System Maintenance in the Graphite Furnace User’s Guide.
These windows are important parts of the whole optical system. Handle them carefully, as you would any sensitive optical component. Do not directly touch the windows. If mirror surfaces need cleaning, this must be done by a PerkinElmer service engineer or a similarly authorized and trained person who knows the special techniques required.
4-5
Page 88
System Maintenance
To clean the windows:
This must be done only by a skilled person who knows the special techniques required.
Note: The more you rub the surface of the windows, the more chance there is of scratching them. You must change scratched windows.
Warning: UV Radiation – Risk of Eye Damage
The lamps may emit UV radiation which can damage your eyes.
Do not gaze into a lighted lamp.Always wear UV-absorbing safety glasses when looking at the radiation from
the lamps.
W3.1
1. Blow dust off optical surfaces with clean dry air before you attempt to clean them. Do not rub the surfaces.
2. Clean the window surfaces with a soft lint-free cloth moistened with spectroscopic grade alcohol. The window must dry quickly to avoid leaving drying marks.
Keep the cleaning solutions and water away from other parts of the instrument.
4-6
0993-5255
Page 89

Maintaining the electrical components

System Maintenance
Warning: Electrical Hazard
To prevent potential injury to yourself and damage to the instrument, switch
OFF all instruments in the system and disconnect them from the line power supply before you perform the procedures described below.
W2.3
Parts that may be serviced by the user
The only parts that you should attempt to maintain are those described below. The electronic components in the instrument do not contain any parts which you can maintain yourself. If there is a problem, contact your local PerkinElmer office. If any fuses blow frequently or if the automatic circuit breaker switch is activated frequently, contact your local PerkinElmer office for advice.
0993-5255
4-7
Page 90
System Maintenance
Resetting the automatic circuit breaker
The spectrometer has an automatic circuit breaker which prevents the system from drawing too much current. When the circuit breaker is activated, the switch flips to the off position (marked ). This can happen while an analysis is running.
Changing the fuses
To reset the switch, wait a few seconds then put the switch to the on position (marked ). If this switch is activated frequently, contact your local PerkinElmer office for advice.
Warning: Electrical Hazard
Use only fuses with the correct current rating and of the specified type for
replacement.
Do not use makeshift fuses and do not short-circuit the fuse holders.
1. Switch off the spectrometer and all other instruments in the system as described in the relevant guides.
2. Switch the circuit breaker on the rear panel of the spectrometer to the off position (marked ).
3. Disconnect the spectrometer’s line power cord from the line power supply.
4. Disconnect all the line power cords from the sockets at the rear of the spectrometer.
5. To change the fuses numbered F1, F2, F3, F4:
1. Insert the tip of a suitable screwdriver in the slot of the respective fuse
holder and unscrew the fuse holder.
2. Remove the blown fuse, then insert a new one of the correct type and
rating. The correct rating is shown on the label of the fuse on the rear panel of the spectrometer.
3. Screw the fuse holder into the correct socket.
6. Reconnect the line power cords correctly.
W2.9
4-8
0993-5255
Page 91
System Maintenance
A
Fuses, F1, F2, F3, F4
InIn EntréeEntrée EingangEingang
P
HO
P
PP
min
max
2
350
250
kPa
Furnace Cooling
3.5
2.5
bar
Refroidissement du Four
50
36
psig
Ofenkühlung
Instrument
230 V 50/60 Hz
Cooling System Système de Refr oidissement Umlaufkühlung
Instrument
Communication
230 V
230 V
230 V
50/60 Hz
50/60 Hz
50/60 Hz
F4
230 V 50/60 Hz
Cooling System Système de Refroidissement Umlaufkühlung
0
OutOut SortieSortie AusgangAusgang
4 A (T)F34 A (T)F24 A (T)F14 A (T)
I
Warning
LR 45111SA***
ll field wiring must have i nsulation suitable for at least 300 V
Avertissement
Le cablage à pied d'ouv re doit être certifiés pour un minimum de 300 V
Communication
F1 4 A (T)
230 V 50/60 Hz
F2 4 A (T)
230 V 50/60 Hz
F3 4 A (T)
Furnace Gases
Gaz du Four Ofengase
P
P
min
max
350
400
kPa
3.5
4.0
bar
Ar / N
2
50.0
58.0
psig
Speci al G as
Normal Gas Gaz Normal
Gaz Spécial Spezialgas
Normalgas
Burner Gases Gaz du Bruleur Brennergase
P
P
minPminPmin
maxPmaxPmax
90
100
kPa
450
500
kPa
450
500
kPa
4.5
5.0
bar
4.5
5.0
bar
0.9
1.0
bar
65.0
72.5
psig
65.0
72.5
psig
13.0
14.5
psig
CH22NO
Air / Luft
2
230 V 50/60 Hz
F4 4 A (T)
I
0
Automatic circuit breaker
Figure 4-1. Automatic Circuit Breaker and Fuses on the rear panel
0993-5255
4-9
Page 92
System Maintenance
A

Changing the air filters

To make sure that the electrical components in the spectrometer are properly cooled, the air filters must allow sufficient air into the instrument. Check the filters regularly and change them if they are dirty. You should not need to change the filters more than once every year if the environmental conditions are as described under Chapter 6, Laboratory Requirements. Replacement filters are available; see Chapter 5, Parts and Supplies for the part number.
1. Carefully pull off the snap-on plastic grid that holds the air filter in place.
2. Replace the filter and push the grid back into place.
HO
2
Furnace Cooling
Refroidissement du Four
Ofenkühlung
Instrument
Communication
230 V 50/60 Hz
230 V 50/60 Hz
Cooling System Système de Refroidissement Umlaufkühlung
0
InIn
OutOut
EntréeEntrée
SortieSortie
EingangEingang
AusgangAusgang
P
PP
P
min
max
350
kPa
250
3.5
bar
2.5 50
psig
36
230 V
230 V
50/60 Hz
50/60 Hz
F4 4 A (T)F34 A (T)F24 A (T)F14 A (T)
I
LR 45111SA***
Warning
ll field wiring must have i ns u la tio n suitable for at leas t 300 V
Avertissement
Le cablage à pied d'ouvre doit être certifiés pour un minimum de 300 V
Furnace Gases
Gaz du Four
Ofengase
P
P
min
max
350
400
kPa
3.5
4.0
bar
Ar / N
2
50.0
58.0
psig
Special Gas
Normal Gas Gaz Normal
Gaz Spécial Spezialgas
Normalgas
Burner Gases
Gaz du Bruleur
Brennergase
P
P
minPminPmin
maxPmaxPmax
90
100
kPa
450
500
kPa
450
500
kPa
0.9
1.0
bar
4.5
5.0
bar
4.5
5.0
bar
13.0
14.5
psig
65.0
72.5
psig
65.0
72.5
psig
CH22NO
Air / Luft
2
Air filters
4-10
0993-5255
Page 93

Replacing the deuterium lamp

The deuterium lamp for the background corrector has a limited lifetime. After a prolonged period of operation the lamp will exhibit decreasing radiant intensity and will ultimately fail. The lifetime of the deuterium lamp will be prolonged if you do not switch it on and off frequently.
You can replace the deuterium lamp assembly yourself. The replacement lamp is fitted in a prealigned mount.
Deuterium lamp cover
System Maintenance
0993-5255
Figure 4-2. Location of Deuterium Lamp inside the Spectrometer
W2.3
Warning: Electrical Hazard
To prevent potential injury to yourself and damage to the instrument, switch
OFF all instruments in the system and disconnect them from the line power supply before you perform the procedures described below.
1. Switch off the spectrometer at the operational on/off switch on the front panel.
2. Switch the circuit breaker on the rear panel of the spectrometer to the off position (marked ).
3. Disconnect the spectrometer from the line power supply.
4-11
Page 94
System Maintenance
Captive screws
Thumbscrew
Deuterium lamp assembly
4-12
Deuterium lamp plug
Figure 4-3. Deuterium Lamp Cover and Bracket Assembly
4. Open the top cover of the spectrometer:
1. Remove the two securing screws; one on the left and one on the right.
2. Swing back the cover to the open position.
5. Remove the deuterium lamp assembly:
1. Undo the two captive screws on the deuterium lamp cover.
2. Carefully lift up the deuterium lamp cover. The deuterium lamp assembly is attached to a bracket underneath the cover.
3. Carefully place the complete assembly on top of the black optics cover.
4. Pull out the deuterium lamp plug from the socket in the lamp compartment.
Do not pull on the deuterium lamp leads.
6. Slacken the thumbscrew on the bracket and remove the old lamp assembly. Save the thumbscrew for use with the new lamp assembly.
0993-5255
Page 95
System Maintenance
7. Unpack the new lamp assembly, taking care to hold it only by the metal mount to prevent fingermarks on the lamp envelope.
8. Slip the slot at the base of the lamp mount over the stud on the lamp bracket and secure the lamp assembly with the thumbscrew saved in step 6.
9. Wipe the lamp envelope with a soft cloth moistened with alcohol to remove dirt, since this would otherwise be burned in when the lamp is hot.
10. Insert the deuterium lamp plug into the socket in the lamp compartment. The plug is asymmetric and can only be inserted in one direction.
11. Reinstall the lamp assembly:
1. Carefully lower the deuterium lamp assembly into the lamp compartment
until the lamp cover fits in place.
2. Screw in the two captive screws on the lamp cover.
12. Close the top cover:
1. Swing down the cover to the closed position.
2. Screw in the two securing screws; one on the left and one on the right.
Since the deuterium lamp is fitted with as prealigned mount, you do not normally need to align the lamp.
Note: Do not attempt to make adjustments to the lamp bracket assembly. If there is indication that the alignment is not correct (e.g., low energy for the BG bar graph in the Align Lamps or Continuous Graphics windows), please contact your local PerkinElmer office.
0993-5255
4-13
Page 96
System Maintenance

Maintaining the acetylene supply

Warning: Acetylene – Explosion Hazard
Acetylene can decompose explosively at pressures higher than 103 kPa / 1.03 bar / 15 psig.
Always make sure that the acetylene outlet gauge pressure is below this
value.
W5.7
You should depressurize both the fuel and oxidant gas hoses at the end of the
working day or if you are not going to use the instrument for an extended period. Close the cylinder valves to avoid the possibility of gas hoses being subjected to the full cylinder pressure in the event that a pressure regulator should fail.
Acetylene is supplied dissolved in a solvent such as acetone, and a small
amount of solvent carryover with the acetylene is normal. However, as tank pressure falls, the relative amount of solvent entering the gas stream increases and can give erratic results, particularly for elements such as calcium, tin, chromium, molybdenum and others whose sensitivity is highly dependent on the fuel:oxidant ratio.The pressure drop is not linear with gas removal since the acetylene is dissolved in a solvent, and a pressure of 600 kPa (6 bar; 85 psig) at ambient temperature indicates that the cylinder is nearly empty. For this reason, replace acetylene cylinders when the cylinder pressure drops to about 600 kPa (6 bar; 85 psig).
Always store and use acetylene cylinders vertically, rather than horizontally,
to prevent liquid acetone from reaching the cylinder valve.
4-14
The practice of ‘cracking the valve’ of an acetylene tank (that is, opening the
valve slightly for a very short period previous to attaching the regulator) is not recommended. While such an action will clear the valve opening of dust or dirt particles and may remove acetone from the cylinder valve, it is a potentially hazardous practice and one which should never be attempted in the presence of an open flame, sparks or other possible sources of ignition.
0993-5255
Page 97

Parts and Supplies

5Parts and Supplies
Marker for header
5
Contents
Parts provided . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Replacement parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Obtaining supplies, replacement parts, and accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
page
0993-5255
5-1
Page 98
Parts and Supplies
5-2
0993-5255
Page 99

Parts provided

PerkinElmer reserves the right to alter the schedule of parts provided without prior notice.
Qty. Item Part No.
1 AAnalyst 700 atomic absorption spectrometer. . . . . . . . B312-0082
1 Read cable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B312-1302
3 Special line cords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B310-0003
1 pack Replacement air filters . . . . . . . . . . . . . . . . . . . . . . . . . . B050-2706
4 Plugs, for handle screw holes . . . . . . . . . . . . . . . . . . . . . B015-3803
1 pack Replacement fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B312-1303
1 AS-800 furnace autosampler . . . . . . . . . . . . . . . . . . . . . B300-1500
1 Swing-arm mount for autosampler . . . . . . . . . . . . . . . . . B312-0922
1 Fuel hose assembly (red) . . . . . . . . . . . . . . . . . . . . . . . . B015-4984
1 Air hose assembly (black) . . . . . . . . . . . . . . . . . . . . . . . B015-4987
1 Nitrous oxide hose assembly (blue) . . . . . . . . . . . . . . . . B015-4990
10 m Special gas tube (FLEXO) . . . . . . . . . . . . . . . . . . . . . . . B050-8264
1 Y-connector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0992-0179
1 Hose kit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B311-0036
1 Drain vessel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B312-1306
1 Sample tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B015-6272
5 Corkpreen gaskets for burner end cap . . . . . . . . . . . . . . 0047-2014
1 O-ring for burner end cap . . . . . . . . . . . . . . . . . . . . . . . . 0990-2147
1 O-ring for the spray chamber . . . . . . . . . . . . . . . . . . . . . 0990-2219
1 Burner head cleaning tool. . . . . . . . . . . . . . . . . . . . . . . . 0303-1573
1 pack Graphite components . . . . . . . . . . . . . . . . . . . . . . . . . . . B300-0408
5 Graphite tubes with integrated platforms . . . . . . . . . . . . B300-0407
1 Tube alignment tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . B008-0112
1 set Tools for changing graphite contacts . . . . . . . . . . . . . . . B312-1301
1 Documentation package B312-0299
Parts and Supplies
Further parts provided with the furnace autosampler are listed under ‘Parts and Supplies’ in the Graphite Furnace User’s Guide.
0993-5255
5-3
Page 100
Parts and Supplies

Replacement parts

Obtaining supplies, replacement parts, and accessories

Supplies, replacement parts, and accessories can be ordered directly from PerkinElmer, using the part numbers quoted in the guides provided with the instrument.
e-ssentials
PerkinElmer Instruments on-line catalog, your single source for quality supplies and accessories. http://essentials.perkinelmer.com/
PerkinElmer’s catalog service offers a full selection of high-quality supplies. To place an order for supplies and many replacement parts, request a free catalog, or ask for information:
– If you are located within the U.S., call toll free 1-800-762-4002, 8 a.m. to
8 p.m. EST. Your order will be shipped promptly, usually within 24 hours.
– If you are located outside of the U.S., call your local PerkinElmer sales or
service office.
5-4
What is listed in this section
This section lists the most important replacement parts for your spectrometer system. Parts listed under Parts provided are not repeated in this section. Information about other parts and accessories for the system is given in the current PerkinElmer brochures and price lists.
Background corrector
Description Part No.
Deuterium lamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B014-8615
0993-5255
Loading...