PerkinElmer aanalyst 200 User Manual

Page 1
AAnalyst 200
User’s Guide
Page 2
Release History
Part Number Release Publication Date
0993 6594 E April 2006
User Assistance PerkinElmer, Inc. 710 Bridgeport Avenue Shelton, Connecticut 06484-4794 U.S.A.
Or emailed to: info@perkinelmer.com
Notices
The information contained in this document is subject to change without notice.
Except as specifically set forth in its terms and conditions of sale, PerkinElmer makes no warranty of any kind with regard to this document, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
PerkinElmer shall not be liable for errors contained herein for incidental consequential damages in connection with furnishing, performance or use of this material.
Copyright Information
This document contains proprietary information that is protected by copyright. All rights are reserved. No part of this publication may be reproduced in any form whatsoever or translated into any language without the prior, written permission of PerkinElmer, Inc.
Copyright © 2006 PerkinElmer, Inc.
Trademarks
Registered names, trademarks, etc. used in this document, even whe n not specifically marked as such, are protected by law.
PerkinElmer is a registered trademark of PerkinElmer, Inc. AAnalysts and e-ssentials are trademarks of PerkinElmer, Inc.
Page 3
Table of Contents
Chapter 1 Introduction
About This Manual......................................................................................1-3
Conventions Used in this Manual................................................................1-5
Warning Labels on the Instrument..............................................................1-6
Chapter 2 Safety Practices
Chapter Overview........................................................................................2-3
Working Safely with the AAnalyst 200 ......................................................2-4
Precautions ..................................................................................................2-5
Hazards with Flame Atomization................................................................2-6
Safety Conventions in the Hardware Guide................................................2-8
Symbols Used on the Instrument.................................................................2-9
Safety Practices for Flame Atomization....................................................2-10
High Temperatures.................................................................................... 2-12
Electrical Safety ........................................................................................2-13
Compressed Gases..................................................................................... 2-14
Safe Handling of Gas Cylinders................................................................ 2-15
Operating Conditions ................................................................................2-18
Exhaust Venting........................................................................................2-19
Safe Handling of Chemicals......................................................................2-20
Decontamination ....................................................................................... 2-21
Waste Disposal..........................................................................................2-22
Electromagnetic Compatibility (EMC) .....................................................2-23
WEEE Instructions for PerkinElmer Products ..........................................2-24
References for Laboratory Safety Practice................................................2-25
Chapter 3 Preparing Your Laboratory
Introduction.................................................................................................3-3
Environmental Conditions...........................................................................3-4
General Laboratory Requirements ..............................................................3-5
Location and Space Requirements ..............................................................3-7
Exhaust Vent Requirements......................................................................3-12
Electrical Requirements ............................................................................3-20
Lamps........................................................................................................3-21
Chapter 4 System Description
Introduction.................................................................................................4-3
AAnalyst 200 Overview.............................................................................. 4-4
The AAnalyst 200 Spectrometer System ....................................................4-5
General Lamp Information.......................................................................... 4-6
The Deuterium Arc Background Corrector...............................................4-11
The Burner System....................................................................................4-12
Connecting the Electrical Supply..............................................................4-19
Mercury/Hydride Determination System (MHS-15).................................4-21
Page 4
Chapter 5 Installation
Introduction to the Installation Procedure...................................................5-3
When the Spectrometer Arrives ..................................................................5-4
Placing the Spectrometer on the Bench.......................................................5-6
AAnalyst 200 Shipping List........................................................................5-8
AAnalyst 200 Spares Kit............................................................................. 5-9
Installing the Touch Screen....................................................................... 5-10
Connecting the Gas Hoses......................................................................... 5-14
Setting up the Burner System.................................................................... 5-19
Installing the Burner Door......................................................................... 5-21
Installing the Sample Tray.........................................................................5-22
Connecting to the Electrical Supply.......................................................... 5-23
Switches and Connectors on the E-Box ....................................................5-24
Setting Up the AutoSampler......................................................................5-25
Connecting the Foot Switch ......................................................................5-26
Chapter 6 Operation of Instrument
Starting up the System................................................................................. 6-3
Setting up the Burner System...................................................................... 6-5
Installing Solvent-Resistant Components in the Burner............................6-11
User Interface............................................................................................6-19
Installing Lamps........................................................................................ 6-21
Switching on the Spectrometer..................................................................6-25
Safety Checks............................................................................................6-26
Using the Background Corrector...............................................................6-27
Optimization Procedures...........................................................................6-28
Aligning Lamp .......................................................................................... 6-29
Setting the Burner Height..........................................................................6-33
Igniting the Flame ..................................................................................... 6-35
Preadjusting the Nebulizer ........................................................................ 6-37
Optimizing the Burner System..................................................................6-39
Rinsing the Burner System........................................................................6-43
Chapter 7 Touch Screen
User Interface..............................................................................................7-3
Startup Dialog..............................................................................................7-5
Lamp Page...................................................................................................7-6
Flame page ................................................................................................ 7-11
Parameters Page ........................................................................................7-14
Analyze Page: Manual Data Display......................................................... 7-20
Analyze Samples Page: Manual Continuous Display ...............................7-23
Analyze Page: Autosampler......................................................................7-25
Tools Dialog.............................................................................................. 7-27
Recommended Conditions.........................................................................7-31
Network Dialog.........................................................................................7-32
Download ..................................................................................................7-41
Importing AAnalyst 200 into Microsoft Excel..........................................7-47
Page 5
Chapter 8 Analyzing Samples
Sequence of Operations...............................................................................8-3
Manual Flame Analyses..............................................................................8-4
Automated Flame Analyses.......................................................................8-15
Manual Analyses: Data Display MHS-Technique ....................................8-22
Running Samples in Emission Mode ........................................................ 8-23
Shutting Down The System....................................................................... 8-31
Emergency Shutdown................................................................................8-32
Rinsing the Burner System........................................................................8-33
Chapter 9 Maintenance
Introduction.................................................................................................9-3
General Maintenance Checklist...................................................................9-4
Instrument Maintenance Checklist..............................................................9-5
Checks.........................................................................................................9-7
Spectrometer Maintenance Checklists ........................................................9-8
Burner Maintenance Checklist....................................................................9-9
Changing the Air Filter..............................................................................9-11
Extinguishing the Flame............................................................................ 9-12
Cleaning the Burner Chamber................................................................... 9-13
Removing the Burner Door.......................................................................9-15
Replacing the Burner Door........................................................................9-16
Cleaning the Burner System......................................................................9-17
Cleaning the Burner Head.........................................................................9-19
Pre-Conditioning a New Burner or Drain Tube ........................................9-21
Removing the Burner Head....................................................................... 9-22
Removing the Burner Assembly from the Instrument...............................9-23
Disassembling the Burner Chamber..........................................................9-25
Reassembling the Burner System..............................................................9-28
Drain System Maintenance........................................................................9-32
Cleaning the Nebulizer.............................................................................. 9-36
Replacing Parts of Wear............................................................................9-45
E-Box ........................................................................................................9-46
Maintaining the Acetylene Supply............................................................9-50
Replacing the Lamps.................................................................................9-51
Replacing the Deuterium Arc Background Corrector Lamp.....................9-52
AS-90plus and AS-93plus Autosampler Maintenance..............................9-56
MHS-15 Mercury/Hydride System Maintenance......................................9-57
Replacement Parts and Accessories .......................................................... 9-60
Chapter 10 Troubleshooting
Introduction...............................................................................................10-3
Flame Analysis Troubleshooting...............................................................10-4
Flame/Burner Troubleshooting .................................................................10-8
Lamp Troubleshooting............................................................................10-11
Troubleshooting Deuterium Background Corrector Problems................10-13
Index
Page 6
Page 7
Introduction
1
Page 8
Page 9
AAnalyst 200 User’s Guide
About This Manual
For complete safety label information see AAnalyst 200 Quick Installation and Setup Guide (0993-6593). This User’s Guide is divided into the following chapters:
Chapter 1 Introduction
This chapter contains a brief introduction on the instrument, the conventions and warnings used in the manual.
Chapter 2 Safety Practices
Important safety information for the AAnalyst 200 is provided in this chapter.
Chapter 3 Preparing Your Laboratory
The preparation of your laboratory needed for the AAnalyst 200. Laboratory requirements for the instrument are reviewed in this chapter.
Chapter 4 System Description
This chapter contains information on the components of the AAnalyst 200, how it works and instrument specifications.
Chapter 5 Installation
Information on installing and re-installing your AAnalyst 200 should you ever need to move your system is provided.
Chapter 6 Setting Up Instrument
Detailed information on how to align, adjust, and optimize the AAnalyst 200 is provided in this chapter.
Chapter 7 Touch Screens
Detailed information on how to operate the AAnalyst 200 using the touch screen is provided in this chapter.
Chapter 8 Preparing for an Analyses
Detailed information on how to set up the AAnalyst 200 for analyses is provided in this chapter.
Chapter 9 Maintenance
Maintenance and cleaning procedures for the various components of your AAnalyst 200 are provided.
1-3
Page 10
Introduction
Chapter 10 Troubleshooting
Performance checks, troubleshooting information, software and system error messages are provided for the AAnalyst 200.
1-4
Page 11
AAnalyst 200 User’s Guide
Conventions Used in this Manual
Normal text is used to provide information and instructions.
Bold text refers to text that is displayed on the screen.
All eight digit numbers are PerkinElmer part numbers unless stated otherwise.
Notes, cautions and warnings
Three terms, in the following standard formats, are used to highlight special circumstances and warnings.
NOTE: A note indicates additional, significant information that is provided with some procedures.
CAUTION
WARNING
We use the term CAUTION to inform you about situations that could result in serious damage to the instrument or other equipment. Details about these circumstances are in a box like this one.
We use the term WARNING to inform you about situations that could result in personal injury to yourself or other persons. Details about these circumstances are in a box like this one.
1-5
Page 12
Introduction
Warning Labels on the Instrument
Caution, hot surface.
Caution, risk of electric shock.
Warning (refer to accompanying documents).
Figure A Label located on the front of the AAnalyst 200
1-6
Page 13
A
B
C
D
AAnalyst 200 User’s Guide
WARNING: Hot Lamp.
WARNING: To avoid electrical shock, disconnect power cord before servicing
WARNING: UV Light Source. Wear protective glasses
when working with this cover removed.
Nebulizer clamp must be lowered and locked over the nebulizer flange.
E
Never use oxygen. Read instructions before lighting burner. Drain trap must be filled.
F
WARNING
Hot surfaces
G
Emergency flame off switch
1-7
Page 14
Introduction
Figure B Label located on the right side of the AAnalyst 200
A
CAUTION: do not restrict air intake or exhaust
1-8
Page 15
Figure C Labels located on the back of the AAnalyst 200
AAnalyst 200 User’s Guide
N
Purge (translucent hose)
O
2
(blue hose)
Air (black
hose)
C2 H2 (red hose)
A
C
D
B
E
Burner Gases Connections
For more information see page 5-14.
1-9
Page 16
Introduction
Figure D Labels located on the E-Box front panel of the AAnalyst 200
A
B
Turn the power switch off and unplug the AC line cord from the AC outlet before loosening the screw.
WARNING
Grounding circuit continuity is vital for safe operation of equipment. Never operate equipment with grounding conductor disconnected.
1-10
Page 17
AAnalyst 200 User’s Guide
Figure E Labels located on the E-Box rear panel of the AAnalyst 200
A
B
C
WARNING
To prevent a shock hazard, turn the power switch off and unplug the AC line cord from the AC outlet.
WARNING
To reduce the chance of an electrical shock do not remove covers that require tool access. No user serviceable parts are inside. Refer servicing to PerkinElmer qualified service personnel.
1-11
Page 18
Introduction
1-12
Page 19
Safety
Practices
2
Page 20
Page 21
AAnalyst 200 User’s Guide
Chapter Overview
This chapter describes the general safety practices and precautions that must be observed when operating the AAnalyst 200.
This advice is intended to supplement, not supersede, the normal safety codes in the user's country. It is also a supplement to the PerkinElmer standard Safety and Health Policy. The information provided does not cover every safety procedure that should be practiced. Ultimately, maintenance of a safe laboratory environment is the responsibility of the analyst and the analyst's organization.
Please consult all manuals supplied with the AAnalyst 200 and accessories before you start working with the instrument. Carefully read the safety information in this chapter and in the other manuals supplied. When setting up the instrument or performing analyses or maintenance procedures, strictly follow the instructions provided.
2-3
Page 22
Safety Practices
Working Safely with the AAnalyst 200
Built-in safety
This instrument complies with international standards governing the safety of laboratory equipment. In routine use, the instrument poses virtually no risk to you. If you take some simple, common-sense precautions, you can maintain the continued safe operation of the instrument.
Safety interlocks
PerkinElmer provides a number of safety interlocks on its instruments that monitor such things as door locks, gas pressures, and the temperatures of various components. Do not attempt to defeat these interlocks; you may compromise your own, or someone else’s, safety.
Correct use of the instrument
Before you install or use your instrument, and in order to get the best results, you should be familiar with all of the instruments in the system and know how to operate them. You should also be aware of the safety procedures in force in your laboratory, especially those concerning atomic spectrometry instruments. Consult the guides supplied with the instruments before you start.
The persons who use this instrument and perform the routine maintenance must be suitably qualified and have received adequate training. If the instrument requires servicing, only a PerkinElmer service engineer or similarly trained and authorized person must do this.
If you use the instrument in a manner not specified in the user guides, or if you use it for a purpose other than that intended, you may damage the instrument, or compromise your own, or someone else’s, safety.
CAUTION
The protection provided by this equipment may be impaired if the equipment is used in a manner not specified by PerkinElmer. This equipment requires no specified inspection or preventive maintenance to ensure the continuous functioning of its safety features.
2-4
Page 23
AAnalyst 200 User’s Guide
Precautions
Be sure that all instrument operators read and understand the precautions listed below. It is advisable to post a copy of the precautions near or on the instrument.
WARNING
The following precautions must be observed when using the AAnalyst 200:
Be sure that the voltage of the AAnalyst 200 corresponds to the power line voltage used in
your laboratory.
Never remove the side panels of the AAnalyst 200 without shutting down the instrument
and disconnecting the instrument power cord from line power.
Only high quality purge gases should be used with the AAnalyst 200. Minimum purity of
99.9% is recommended. A high quality filter-dryer accessory is recommended for the removal of any moisture from the purge gases.
2-5
Page 24
Safety Practices
Hazards with Flame Atomization
This section contains a summary of the hazards associated with the flame atomization technique. For more detailed information see the remaining sections in this safety chapter.
High temperatures
The flame can generate temperatures of up to 2800 °C.
To avoid serious burns, never touch the burner head until it has cooled to room temperature.
WARNING
Toxic products
Toxic combustion products can be generated by the system, depending on the type of
analyses being performed. You must provide an efficient exhaust venting system to remove toxic products generated during instrument operation.
Combustion products may vary with the analysis of each sample. Acetylene (less than 1
liter) is released during the bleeding of the gas box prior to shutdown. An exhaust venting system is always required to remove those gases which are generated during the normal operation of the instrument.
The materials in sample and reagent containers, and material that you collect in waste
containers may include small amounts of toxic substances that were used in the analyses. You may have to dispose of these as hazardous waste.
UV radiation
You should be aware of the health hazard presented by UV radiation.
Always wear UV-absorbing eye protection when viewing:
The flame, especially the nitrous oxide-acetylene flame.
Hollow cathode or electrodeless discharge lamps.
The deuterium background correction lamp.
Keep the atomizer compartment door closed when the flame is burning and never directly
view the flame unless you are wearing UV-absorbing glasses.
2-6
Page 25
AAnalyst 200 User’s Guide
Acetylides
If you have aspirated high concentrations of copper, silver, or mercury solutions into an acetylene flame, unstable acetylides may have formed in the burner chamber. If permitted to dry, these compounds may explode
WARNING
1. Aspirate dilute acid (1% (v/v) HCl solution) for 5 minutes then aspirate with deionized water
for another five minutes before turning off the flame.
2. Extinguish the flame and allow the burner head to cool.
3. Remove the burner head.
4. Pour about 500 mL of deionized water slowly through the neck of the burner chamber to
thoroughly flush the chamber and the drain system.
5. Empty the drain vessel.
NOTE: Dispose of hazardous or corrosive solutions properly and refer to your local safety regulations
for proper disposal procedures
Cyanide solutions
If you intend to aspirate cyanide solutions, make sure that the pH of the liquid in the drain vessel is greater than pH 10; toxic hydrogen cyanide gas is formed when cyanides contact acidic solutions.
Never allow solutions containing cyanides to mix with acidic solutions.
2-7
Page 26
Safety Practices
Safety Conventions in the Hardware Guide
Any of the following safety conventions can be used throughout this guide:
This is an example of a warning, a situation that could lead to a personal injury.
WARNING
WARNING
WARNING
CAUTION
This symbol alerts you to the risk of
injury
to yourself or other persons.
electric shock that could result in personal
Details about these circumstances are in a box like this one.
This symbol alerts you to the risk of
injury
to yourself or other persons.
hot surfaces that could cause personal
Details about these circumstances are in a box like this one.
We use the term CAUTION to inform you about situations that could result in serious damage to the instrument or other equipment. Details about these circumstances are in a box like this one.
2-8
Page 27
Symbols Used on the Instrument
Any of the following symbols can be used on the instrument:
Caution, risk of electric shock.
Caution, hot surfaces.
AAnalyst 200 User’s Guide
Caution, risk of danger.
Documentation must be consulted to determine the nature o the potential hazard and any actions which have to be taken.
Indicates alternating current.
Indicates the off position of the main power switch.
Indicates the on position of the main power switch.
Indicates the protective conductor terminal.
2-9
Page 28
Safety Practices
Safety Practices for Flame Atomization
Safe operation of the flame
Before you ignite the flame make sure that:
The laboratory exhaust venting system is operating;
The burner head is installed correctly;
The burner end cap is secured firmly;
The burner unit contains the necessary solvent-resistant components if you intend to
analyze samples containing organic solvents;
The atomizer compartment door is closed;
The pH of the liquid in the drain vessel is greater than pH 10 if you intend to aspirate
cyanide solutions.
Do not leave the flame unattended. Always make sure that there is a fire extinguisher near
the instrument.
Never change the gas pressure or shut a gas valve while the flame is burning.
Do not allow the burner head slot to become blocked. This can cause a flashback of the
flame.
Do not place open containers of flammable liquids and solvents near to the flame. Be
especially careful with samples that contain highly volatile solvents.
When you have analyzed samples containing organic solvents, flush all traces of solvent
from the burner system.
The drain system:
Always place the drain vessel in a well ventilated place underneath the spectrometer, in
full view while you are working with the spectrometer. This prevents the build-up of potentially hazardous gases and allows you to see the liquid level.
Empty the drain vessel frequently; especially do not allow the drain vessel to fill with
organic solvent.
Thoroughly flush the drain system, as described in the guide, when you shut down the
system.
Check the condition of the drain tube regularly, especially if you use organic solvents.
Replace the drain tube when it first shows signs of cracking or discoloration.
2-10
Page 29
AAnalyst 200 User’s Guide
UV radiation
You should be aware of the health hazard presented by UV radiation.
When the instrument is on, do not remove any covers unless specifically instructed to do
so in the guide or you may be exposed to potentially hazardous UV radiation.
Always wear UV-absorbing eye protection when viewing:
The flame, especially the nitrous oxide-acetylene flame.
Hollow cathode or electrodeless discharge lamps.
The deuterium background correction lamp.
2-11
Page 30
Safety Practices
High Temperatures
Burner system
The flame can generate temperatures of up to 2800 °C.
To avoid serious burns, never touch the burner head until it has cooled to room
temperature.
Quartz tube atomizer
The quartz tube atomizer (QTA) used for the hydride-generation technique (option) can reach temperatures of up to 1000 °C.
To avoid serious burns, never touch any part of the QTA mount (flame-heated systems) or
the QTA until they have cooled to room temperature.
2-12
Page 31
AAnalyst 200 User’s Guide
Electrical Safety
The instrument has been designed to protect the operator from potential electrical hazards. This section describes some recommended electrical safety practices.
Lethal voltages are present at certain areas within the instrument. Internal maintenance of the instrument should only be performed by a PerkinElmer service engineer or similarly authorized and trained person. When the
WARNING
The instrument must be correctly connected to a suitable electrical supply. The supply must have a correctly installed protective conductor (earth ground) and must be installed and checked by a qualified electrician before connecting the instrument.
WARNING
instrument is connected to line power, opening the instrument covers is likely to expose live parts. Even when the power switch is off, high voltages can still be present. Capacitors inside the instrument may still be charged even if the instrument has been disconnected from all voltage sources.
Any interruption of the protective conductor (earth ground) inside or outside the instrument or disconnection of the protective conductor terminal is likely to make the instrument dangerous. Intentional interruption is prohibited.
When working with the instrument:
Connect the instrument to a correctly installed line power outlet that has a protective
conductor connection (earth ground).
Do not operate the instrument with any covers or internal parts removed.
Do not attempt to make internal adjustments or replacements except as directed in the
manuals.
Disconnect the instrument from all voltage sources before opening it for any
adjustment, replacement, maintenance, or repair. If afterwards, the opened instrument must be operated for further adjustment, maintenance, or repair, this must only be done by a PerkinElmer Service engineer or similarly authorized and trained person.
Whenever it is possible that the instrument is no longer electrically safe for use, make
the instrument inoperative and secure it against any unauthorized or unintentional operation. The electrical safety of the instrument is likely to be impaired if, for exam­ple, the instrument shows visible damage, has been subjected to prolonged storage under unfavorable conditions or has been subjected to severe stress during transportation.
2-13
Page 32
Safety Practices
Compressed Gases
NOTE: The permanent installation of gas supplies is the responsibility of the user and should conform to
local safety and building regulations.
Warning: Compressed Gases
High pressure gas cylinders can be dangerous if mishandled or misused. You
WARNING
NOTE: The responsible body must ensure that the type of connector used at the outlet side of the gas-
pressure regulator conforms to applicable national requirements.
Summary of gas hazards
Hazards associated with the different gases used in atomic absorption spectrometry are presented in the table below:
Gas Suffocation Explosion Spontaneous Decomposition
Air Argon x Nitrogen x Acetylene (ethyne, C2H2) x x x Nitrous Oxide (dinitrogen
monoxide, N
2
O)
must have a tank regulator on the gas cylinder.
or Combustion
x x
Contact the gas supplier for a material safety data sheet (MSDS) containing detailed information on the potential hazards associated with the gas.
WARNING
WARNING
Make sure that all gas connections are gas-tight. Check the connections and gas hoses regularly for leaks using a leak testing fluid.
Warning: Nitrous oxide is a strong oxidant and an asphyxiant
Nitrous oxide is an asphyxiant; it can be considered a potential hazard whenever it is stored or used in a confined space. Spontaneous combustion may occur if nitrous oxide comes into contact with grease, oil or other organic materials.
Store nitrous oxide cylinders in a well-ventilated area outside the laboratory.
Do not store nitrous oxide cylinders in close proximity to flammable gases (e.g. acetylene).
• Make sure tubing and fittings carry nitrous oxide are free of grease, oil and other organic material.
Warning: Acetylene-Explosion Hazard
Acetylene can decompose explosively at pressure higher than 103 kPa/1.03 bar/15 psig.
Always make sure that the acetylene outlet gauge pressure is below these values.
Acetylene can react with copper to form a readily explosiv e comp o und.
Never use copper tubing or fittings for acetylene lines.
2-14
Page 33
AAnalyst 200 User’s Guide
Safe Handling of Gas Cylinders
Gases commonly used with the AAnalyst 200 include acetylene, nitrous oxide, argon and nitrogen. The major hazard associated with these gases is suffocation. This can occur if the gas is allowed to escape in an enclosed area and displaces the oxygen in the air.
Contact the gas supplier for a material safety data sheet (MSDS) containing detailed information on the potential hazards associated with the gas.
Carefully use, store, and handle compressed gases in cylinders. Gas cylinders
WARNING
NOTE: The permanent installation of gas supplies is the responsibility of the user and should conform to
local safety and building codes.
If liquid nitrogen is used, the gas cylinder must be fitted with an over-pressure regulator, which will vent the cylinder as necessary to prevent it from becoming a safety hazard.
can be hazardous if they are mishandled.
Listed below are some general safety practices for the proper identification, storage, and handling of gas cylinders. Consult the following references for more detailed information and additional guidelines.
Compressed Gas Association (USA), "Safe Handling of Compressed Gases in
Containers," pamphlet no. P-1, 1984.
Compressed Gas Association (USA), "The Inert Gases – Argon, Nitrogen and Helium,"
pamphlet no. P-9, 1992.
Identification of Gas Cylinders
Legibly mark cylinders to identify their contents. Use the chemical name or commercially
accepted name for the gas.
Storing Cylinders
Review the following precautions to ensure the safe use and storage of gas cylinders.
Cylinders should be stored in accordance with the regulations and standards applicable to
the customer’s locality, state, and country.
When cylinders are stored indoors in storage rooms, the storage room should be well
ventilated and dry. Ensure that the ventilation is adequate to prevent the formation of dangerous accumulations of gas. This is particularly important in small or confined areas.
Do not store cylinders near elevators, gangways, or in locations where heavy moving
objects may strike or fall against them.
2-15
Page 34
Safety Practices
Use and store cylinders away from exits and exit routes.
Locate cylinders away from heat sources, including heat lamps. Compressed gas cylinders
should not be subjected to temperatures above 52 °C (126 °F).
Do not allow ignition sources in the storage area and keep cylinders away from readily
ignitable substances such as gasoline or waste, or combustibles in bulk, including oil.
Store cylinders standing upright, fastened securely to an immovable bulkhead or
permanent wall.
When storing cylinders outdoors, they should be stored above ground on a suitable floor
and protected against temperature extremes (including the direct rays of the sun).
Mark gas cylinders clearly to identify the contents and status (full, empty, etc.).
Do not attempt to refill gas cylinders.
Arrange gas hoses where they will not be damaged or stepped on and where things will
not be dropped on them.
Perform periodic gas leak tests by applying a soap solution to all joints and seals.
Handling Cylinders
Review the following precautions to ensure the safe use and handling of gas cylinders.
If it becomes necessary to move cylinders, do so with a suitable hand truck after insuring
that the container cap is secured and the cylinder is properly fastened to the hand truck.
Use only regulators, tubing and hose connectors approved by an appropriate regulatory
agency.
Arrange gas hoses where they will not be damaged or stepped on and where objects will
not be dropped on them.
Do not refill cylinders.
Check the condition of pipes, hoses and connectors regularly. Perform gas leak tests at all
joints and seals of the gas system regularly, using an approved gas leak detection system.
The responsible body must carry out appropriate leakage tests necessary for safety on
those gas and liquid connections which the operator is directed to assemble during installation, normal use, or maintenance.
Check that when the equipment is turned off, close all gas cylinder valves tightly at the
cylinder. Bleed the remainder of the line before turning the exhaust vent off.
2-16
Page 35
AAnalyst 200 User’s Guide
Safe use of burner gases
The seepage of fuel gas or fuel gas mixture from the burner system, the drain system, the gas control system, or the gas connections constitutes a serious fire hazard.
If you suspect that the gas box has been damamged (for example, using the acetylene tank below 85 psi) do not use the instrument until it has been inspected by a PerkinElmer Service Representative.
WARNING
Make sure that there are no breaks or leaks in any of these systems and that all the seals
are correctly installed and in good condition.
Regularly check for leaks at joints and seals using an approved leak test fluid.
When you perform any maintenance or installation procedures, follow the instructions
in the guide exactly.
Do not attempt to service the gas control system yourself. A PerkinElmer service
engineer, or similarly authorized and trained person, must perform the work.
When you shut down the instrument, for example at the end of the working day, shut all the gas lines at the cylinder or regulator valves. Bleed the lines between the regulator and instrument to atmosphere before switching off the exhaust venting system.
Warning: Flashback Hazard
PerkinElmer burner systems are designed for use with compressed air. The
WARNING
use of oxygen can cause an explosion in the burner system. And oxygen­enriched air can cause a flashback of flame.
Never use oxygen or oxygen-enriched air with PerkinElmer burner systems. Use only compressed air.
2-17
Page 36
Safety Practices
Operating Conditions
Warning: Explosive Atmosphere
This instrument is not designed for operation in an explosive atmosphere.
WARNING
Recommended operating conditions:
Indoors.
The location must be free of smoke, dust, corrosive fumes, direct sunlight and excessive
vibration.
Temperature: +15 °C to +35 °C (+59 °F to +95 °F). Maximum rate of change not exceeding
3 °C (5 °F) per hour.
Relative humidity: 20% to 80%, without condensation.
Altitude: in the range 0 m to 2,000 m (sea level to 6,562 feet).
Safe operating conditions:
Indoor.
Temperature: +5 °C to +40 °C (+41 °F to +104 °F).
Relative humidity: 20% to 80%, without condensation.
Altitude: in the range 0 m to 2,000 m (sea level to 6,562 feet).
Storage conditions:
Ambient temperature: –20 °C to +60 °C (–4 °F to +140 °F).
Relative humidity 20% to 80%, without condensation.
Altitude: in the range 0 m to 12,000 m (sea level to 39,370 feet).
2-18
Page 37
AAnalyst 200 User’s Guide
Exhaust Venting
Toxic combustion products, metal vapor, ozone, etc., can be generated by the system, depending on the type of analyses being performed.
You must provide an efficient exhaust venting system to remove toxic products generated
during instrument operation.
2-19
Page 38
Safety Practices
Safe Handling of Chemicals
Some chemicals used with the instrument may be hazardous or may become hazardous after completion of an analysis.
The responsible body must take the necessary precautions to ensure that the surrounding workplace is safe and that instrument operators are not exposed to hazardous levels of toxic substances (chemical or biological) as defined in the applicable national, state, and local health and safety regulations and laws. Venting for fumes and disposal of wastes must be in accordance with all national, state, and local health and safety regulations and laws.
2-20
Page 39
AAnalyst 200 User’s Guide
Decontamination
Before using any cleaning or decontamination methods except those specified by PerkinElmer, responsible bodies should check with PerkinElmer that the proposed method will not damage the equipment.
NOTE: The responsible body has the responsibility for carrying out appropriate decontamination if
hazardous material is spilt on or inside the equipment.
There are two documents on the PerkinElmer public website that responsible bodies can access and complete. One is “Decontamination of Instrumentation and Associated Sub-assemblies” and the other is the “Certificate of Decontamination.” They are used to certify the decontamination process was completed before equipment can be returned to PerkinElmer and may serve as a format for similar customer documents.
2-21
Page 40
Safety Practices
N
Waste Disposal
Contents of waste containers
The materials that you collect in waste containers may include small amounts of the substances that were analyzed and other chemicals used in the analyses. If these materials are toxic, corrosive, or contain organics, you may have to treat the collected effluent as hazardous waste. Refer to your local safety regulations for proper disposal procedures.
NOTE: The responsible body is responsible for the correct collection and disposal of waste materials.
This includes the necessity for a suitably sized waste container of appropriately resistant material for the collection of organic solvent waste and provision for the removal into an appropriate exhaust system of any gases or vapors which may be produced in hazardous concentrations.
Disposing of defective lamps
Hollow cathode lamps and electrodeless discharge lamps contain small quantities of the lamp element in a very pure form. If a lamp containing toxic elements becomes defective or otherwise unusable, you must treat it as hazardous waste and dispose of it accordingly. A licensed company in the chemical waste disposal business can provide lamp disposal in accordance with environmental regulatory requirements. Please note that PerkinElmer does not take back lamps for disposal. In addition, hollow cathode lamps and deuterium lamps are maintained under reduced pressure. Handle and dispose of them correctly to minimize the implosion risk. For information, the quantities of analyte material contained in the cathodes of hollow cathode lamps are listed in the table below:
Quantity of Material in the Cathode
less than 5 g As, Au, B, Ba, Be, Ca, Dy, Er,
5–10 g Ag, Al, Bi, Cd, Co, Cr, Mn,
10–15 g Cu, Fe, Hg, Nb, P, Pb, Ti, Tl all other multi-element HCLs
For EDLs, the quantity of analyte material used is much less than the quantities used in HCLs. Typically, only several milligrams of material are placed in the EDL bulb. No PerkinElmer EDLs (System 1 or 2) contain more than 0.05 g of analyte material.
Single-Element Hollow Cathode Lamps
Eu, Ga, Gd, Ge, Hf, Ho, In, Ir, K, La, Li, Mg, Na, Nd, Pd, Pr, Pt. Re, Rh, Ru, Sc, Se, Sm, Sn, Ta, Tb, Tm, Yb, Y
Mo, Ni, Sb, Si, Sr, Te, V, W, Zn, Zr
Multi-Element Hollow Cathode Lamps
a-K, Pt-Ru
Ca-Mg, Ca-Zn, Ag-Au, Sn-Te, Ca-Mg-Zn, Ca-Mg-Al
2-22
Page 41
AAnalyst 200 User’s Guide
Electromagnetic Compatibility (EMC)
Regulatory Information
United States (FCC)
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which user will be required to correct the interference at his own expense.
NOTE: Changes or modifications not expressly approved by PerkinElmer could void the user’s authority
to operate this product.
Europe
All information concerning EMC standards is in the Declaration of Conformity, and these standards may change as the European Union adds new requirements.
2-23
Page 42
Safety Practices
WEEE Instructions for PerkinElmer Products
or
A label with a crossed-out wheeled bin symbol and a rectangular bar indicates that the product is covered by the Waste Electrical and Electronic Equipment (WEEE) Directive and is not to be disposed of as unsorted municipal waste. Any products marked with this symbol must be collected separately, according to the regulatory guidelines in your area.
The objectives of this program are to preserve, protect and improve the quality of the environment, protect human health, and utilize natural resources prudently and rationally. Specific treatment of WEEE is indispensable in order to avoid the dispersion of pollutants into the recycled material or waste stream. Such treatment is the most effective means of protecting the customer’s environment.
Requirements for waste collection, reuse, recycling, and recovery programs vary by regulatory authority at your location. Contact your local responsible body (e.g., your laboratory manager) or authorized representative for information regarding applicable disposal regulations. Contact PerkinElmer at the web site listed below for information specific to PerkinElmer products.
Web address:
http://las.perkinelmer.com/OneSource/Environmental-directives.htm
Customer Care: 1-800-762-4000 (inside the USA) (+1) 203-925-4602 (outside the USA)
0800 40 858 (Brussels) 800 90 66 42 (Monza)
Products from other manufacturers may also form a part of your PerkinElmer system. These other producers are directly responsible for the collection and processing of their own waste products under the terms of the WEEE Directive. Please contact these producers directly before discarding any of their products.
Consult the PerkinElmer web site (above) for producer names and web addresses.
2-24
Page 43
AAnalyst 200 User’s Guide
References for Laboratory Safety Practice
1. Bretherik, L., Bretherik’s Handbook of Reactive Chemical Hazards, 4th ed., Butterworth &
Co. Ltd., London, UK, 1990.
2. Safe Practices in Chemical Laboratories,
Royal Society of Chemistry, London, UK, 1989.
3. Hazards in the Chemical Laboratory, 5th edition, Luxon, S.G., ed.,
Royal Society of Chemistry, London, UK, 1992.
4. CRC Handbook of Laboratory Safety, 4th edition, Furr, K., ed.,
The Chemical Rubber Co. Press, Florida, USA, 1995.
5. Prudent Practices for Handling Hazardous Chemicals in Laboratories, National Research
Council, National Academic Press, Washington D.C., USA, 1981.
6. Sax’s Dangerous Properties of Industrial Materials,
9th edition, Sax, N.I. and Lewis, R.J., eds.,Wiley, New York, USA, 1998.
7.
Pohanish, R.P. and Greene, S.A., Hazardous Materials Handbook,
Wiley, New York, USA, 1996.
Compressed Gas Association, Inc., Arlington, VA 22202, USA, various publications.
8.
9.
Data Sheets provided by chemical manufacturers, e.g.:
-USA, Material Safety Data Sheets (MSDS),
-FRG, DIN-Sicherheitsblätter,
-UK, Hazard Data Sheets
.
2-25
Page 44
Safety Practices
2-26
Page 45
Preparing
Your Lab
3
Page 46
Page 47
AAnalyst 200 User’s Guide
Introduction
The AAnalyst 200 spectrometer has some special requirements that you must consider when setting up the laboratory. The items shown in the following checklist need to be considered when preparing the laboratory for the instrument.
Environmental Conditions
General Laboratory Requirements
Laboratory Space Requirements
Exhaust Vent Requirements
Gas Supply Requirements
Electrical Requirements
Lamps
Make sure your laboratory meets all of the requirements before you begin to install the system.
3-3
Page 48
Preparing Your Lab
Environmental Conditions
The laboratory in which the AAnalyst 200 is located must meet the following conditions:
A corrosive-free environment.
The instrument will operate with a laboratory temperature between 15 and 35 °C (59-95 °F)
with a maximum rate of change of 3 °C (5 °F) per hour.
The environment should be relatively dust-free to avoid sample and instrument
contamination problems.
Free of excessive vibration.
Free of direct sunlight.
The relative humidity should be between 20% and 80%, noncondensing.
Two types of flames are commonly used for atomic absorption flame analysis: air-acetylene
or nitrous oxide-acetylene
. Air acetylene is the preferred flame for the determination of about 35 elements by atomic absorption. The temperature of the air-acetylene flame is approximately 2300 °C (4200 °F). For most air-acetylene flames, the acetylene flow used is about 4 liters per minute ( or 0.14 cubic feet/min). Using a heat combustion value of 1,450 BTU per cubic foot, the heat given off would be approximately 12,300 BTU per hour.
The nitrous oxide-acetylene flame has a maximum temperature of approximately 2800 °C (5100 °F) and is also used for the determination of elements which form refractory oxides. It is also used to overcome chemical interferences that may occur in flames of lower temperatures. For the nitrous oxide-acetylene flame, the acetylene flow is about 14 liters/min (or 0.5 cubic feet per min.). Using a heat of combustion value of 1,450 BTU per cubic foot, the heat given off would be approximately 43,000 BTU per hour.
3-4
Page 49
AAnalyst 200 User’s Guide
General Laboratory Requirements
Operating conditions
Warning: Explosive Atmosphere
This instrument is not designed for operation in an explosive atmosphere.
WARNING
Recommended operating conditions:
Indoors.
The location must be free of smoke, dust, corrosive fumes, direct sunlight and excessive
vibration.
Temperature: +15 °C to +35 °C (+59 °F to +95 °F). Maximum rate of change not exceeding
3 °C (5 °F) per hour.
Relative humidity: 20% to 80%, without condensation.
Altitude: in the range 0 m to 2,000 m (sea level to 6562 feet).
Safe operating conditions:
Indoor.
Temperature: +5 °C to +40 °C (+41 °F to +104 °F).
Relative humidity: 20% to 80%, without condensation.
Altitude: in the range 0 m to 2,000 m (sea level to 6562 feet).
Storage conditions:
Ambient temperature: –20 °C to +60 °C (–4 °F to +140 °F).
Relative humidity 20% to 80%, without condensation.
Altitude in the range 0 m to 12,000 m (sea level to 39,370 feet).
3-5
Page 50
Preparing Your Lab
NOTE: When you remove the instrument from storage and before you put it into operation, allow it to
stand for at least a day under the recommended operating conditions.
Laboratory hygiene
Keep the work area scrupulously clean to avoid contaminating your samples and to maintain
a safe working environment. Clean up spilled chemicals immediately and dispose of them properly.
Do not allow waste to accumulate in the work area. Dispose of waste correctly.
Do not allow smoking in the work area. Smoking is a source of significant contamination
and also a potential route for ingesting harmful chemicals.
Do not store, handle, or consume food in the work area.
Working with chemicals
Some chemicals used with the instrument may be hazardous or may become hazardous after completion of an analysis.
Use, store, and dispose of chemicals in accordance with the manufacturer’s
recommendations and the applicable national, state, and/or local regulations.
Do not put open containers of solvent near the instrument.
Store solvents in an approved cabinet (with the appropriate ventilation) away from the
instrument.
Wear appropriate eye protection at all times while handling chemicals. Depending on the
types of chemicals you are handling, wear safety glasses with side shields, or goggles, or a full-face shield.
Wear suitable protective clothing, including gloves if necessary, resistant to the chemicals
you are handling.
When preparing chemical solutions, always work in a fume hood that is suitable for the
chemicals you are using.
Perform sample preparation away from the instrument to minimize corrosion and
contamination.
Clean up spills immediately using the appropriate equipment and supplies, such as spill
cleanup kits.
3-6
Page 51
AAnalyst 200 User’s Guide
Location and Space Requirements
Bench and location requirements
The location you choose for the spectrometer must fulfill the criteria listed below:
Place the spectrometer on a sturdy bench or trolley (cart) capable of sustaining the weight.
The work surface must be level, flat, clean, dry, and free from vibration. It should be strong and stiff enough to bear the weight of the spectrometer system over a long period without warping or bending. A movable table is very convenient, but make sure that the wheels can be securely locked.
Position the bench or trolley out of direct sunlight and away from radiators and heaters.
Position the system near to the electricity and gas supply points.
Position the system so that the atomizer compartment is located beneath a suitable exhaust
venting system.
The instrument should not be against a wall, as the instrument must be accessible from the
back for servicing purposes. An accessible space of at least 25.4 cm (10 in.) should be available behind the instrument.
Make sure that there is space at the rear and sides of the system for air to circulate freely. Do
not block the fan operating on the right side of the instrument.
Make sure that there is sufficient room for you to work comfortably with the instrument, and
that you can reach the connections at the rear of the instruments and the electricity and gas supply points.
Laboratory Space Requirements
The illustrations on the following pages show the dimensions of the spectrometer and the space that you require to accommodate the spectrometer, and ancillary equipment.
Instrument Dimensions
AAnalyst 200 is 700 mm wide (27.6 in.), 760-mm deep (29.9 in.), 650-mm (25.5 in.) high. The instrument weighs 49 kg (108 pounds).
The main power cable length is 2 m (6 ft).
To provide access for servicing and space for ventilation, allow at least 25.4 cm (10 in.) of space between the instrument and the laboratory walls.
3-7
Page 52
Preparing Your Lab
170 mm
700 mm
650 mm
473 mm
760 mm
910 mm
Figure 3- 1 Outside dimensions of instrument
3-8
Page 53
AAnalyst 200 User’s Guide
Bench Requirement
The benchtop or area in which the AAnalyst 200 instrument is to be installed should be large enough to accommodate the instrument and any accessories you’ll be using. The instrument itself weighs 49 kg (108 lb.). If you are considering the use of other accessories, the bench you choose to use should be able to sustain as much as 136 kg (300 lbs.).
Location
Typically, the system is installed on a laboratory work bench or table. The table should be smooth, level, and free from vibration. The dimensions for the instrument are listed below and shown in the next figure. For other accessories, refer to the manuals supplied with the accessory.
The bench should be located near the required electricity and gas supplies. Allow additional power outlets for all your accessories. A distance of at least 6 inches (153 mm) should be surrounding both sides and rear of the instrument for proper ventilation. If not possible, the bench on which the instrument is placed should be on wheels so that it can be easily moved.
Accessories
Make sure that there is sufficient room near the spectrometer for any accessories. For example, we recommend that you place the (optional) flame autosampler and mercury hydride system on the left side of the instrument or on a roll cart in front of the instrument.
Printer
Refer to the manuals supplied with your printer for dimensions. The printer may be placed on the bench next to the instrument.
3-9
Page 54
Preparing Your Lab
Approx. 170 mm
525 mm
700 mm
Figure 3- 2 Space requirements for the spectrometer system – front view
3-10
Page 55
AAnalyst 200 User’s Guide
Figure 3- 3 Space requirements for the spectrometer system – top view
3-11
Page 56
Preparing Your Lab
Exhaust Vent Requirements
Exhaust venting is important for the following reasons.
It protects laboratory personnel from toxic vapors that may be produced by some samples.
It helps to protect the instrument from corrosive vapors that may originate from the
sample(s).
It removes dissipated heat produced by the instrument and power supply.
It will tend to reduce the effects of room drafts and the laboratory atmosphere on flame
stability.
The use of the AAnalyst 200 without adequate ventilation to outside air may constitute a health hazard. For example, the
WARNING
combustion of halogenated hydrocarbons produces toxic vapors.
Temperatures
The following temperatures have been measured with a nitrous oxide-acetylene flame: 310 °C at the vent intake, 160 °C at 2.4 meters (8 feet) from the vent intake, 105 °C at the blower intake, and 50 °C at the blower motor housing near the front bearing.
To withstand these temperatures stainless-steel tubing must be used.
NOTE: Local electrical codes do not allow PerkinElmer Service Engineers to install the blower and vent
assembly.
Exhaust ducting
Alternatively, smooth stainless-steel tubing should be used instead of flexible stainless steel tubing where flexibility is not required to reduce system friction loss or "drag." If smooth stainless steel is used, there must be a way to move the vent hood out of the way for servicing. A length of smooth stainless steel ducting has 20-30% less friction loss than a comparable length of flexible ducting. When smooth stainless steel tubing is used, elbows must be used to turn corners. These elbows should turn at a centerline radius of 150mm with a maximum bend angle of 45 degrees to reduce friction losses, and the number of elbows should be minimized.
You must provide an efficient exhaust venting system to remove heat, toxic combustion fumes and vapors that can be generated by atomic spectroscopy instruments. We strongly recommend that you do not place the spectrometer in a chemical hood.
3-12
Page 57
AAnalyst 200 User’s Guide
Additional recommendations on the venting system include the following items:
Make sure the duct casing is installed using fireproof construction. Route ducts away from
sprinkler heads.
The duct casing and venting system should be made of materials suitable for temperatures
greater than 70 °C (158 °F). They should be installed to meet local building code requirements.
The exhaust hood opening should be about 170 mm (6.7 inches) above the top of the burner
shield.
Locate the discharge outlet as close to the blower as possible. All joints on the discharge
side should be airtight, especially if toxic vapors are being vented.
Equip the blower with a pilot light located near the instrument to indicate to the operator
when the blower is on.
Equip the outlet end of the system with a back draft damper and take the necessary
precautions to keep the exhaust outlet away from open windows or inlet vents. In addition, extend it above the roof of the building for proper dispersal of the exhaust.
Equip the exhaust end of the system with an exhaust stack to improve the overall efficiency
of the system.
Make sure the length of the duct that enters into the blower is a straight length at least ten
times the duct diameter. An elbow entrance into the blower inlet causes a loss in efficiency.
Position the discharge outlet away from open windows and inlet vents.
Exhaust capacity
The system should provide a flow rate of 5600-8400 liters/minute (200–300 cubic
feet/minute). Measure the flow rate at the inlet to the hood with a suitable air-flow meter to ensure that the system is drawing properly.
Provide make-up air to the laboratory in the same quantity as is exhausted. An ‘airtight’
laboratory will cause a loss in efficiency in the exhaust volume.
Blower information
The required capacity of the blower depends on the duct length and the number of elbows or bends used to install the system. If an excessively long duct system or a system with many bends is used, a stronger blower may be necessary to provide sufficient exhaust volume.
The blower should be located at least 4 meters (13.1 ft) and not more than 6.5 meters (21.3 ft) from the inlet directly above the instrument, and should exhaust to the atmosphere or into a considerably wider exhaust duct.
3-13
Page 58
Preparing Your Lab
In many countries, PerkinElmer offers an accessory blower and vent kit that will fulfill the exhaust requirements for all atomic spectroscopy instruments, or your local PerkinElmer office can supply you with the addresses of suitable suppliers.
In many countries electrical codes do not permit PerkinElmer service engineers to install the blower and vent assembly.
Blower and vent kit
The blower and vent kit (P/N 0303-0406) available from PerkinElmer contains a rotary blower with associated hardware, a hood to be located above the instrument, an adapter to connect the blower and hood to suitable metal ducting, and installation instructions. The blower requires a line voltage of 120 V or 230 V. The adapter and hood are made of stainless steel. The metal ducting required to connect the hood to the blower and for the exhaust from the blower is not included in the kit.
11 cm I.D. (4.3 in.)
10 cm
(4 in.)
30 cm
(12 in.)
15 cm
(6 in.)
Figure 3- 4 Dimensions of a typical exhaust hoodGas Supply Requirements
The permanent installation of gas supplies is the responsibility of the user and should conform to local safety and building regulations. PerkinElmer recommends that cylinders be stored outside the laboratory and that gases be piped to the spectrometer system in approved gas lines.
The user must provide the gas supplies, regulators, connectors, and valves.
An air compressor is recommended for air. A double-stage or heated regulator is recommended for nitrous oxide. Either the gas supplier or PerkinElmer can provide the correct regulators and valves.
You must be able to reach the on/off valves easily and see the pressure indicators.
3-14
Page 59
AAnalyst 200 User’s Guide
Recommendations for compressed gas systems
Gas installations
Legibly mark gas lines and outlets to clearly identify the contents. Use the chemical name or
commercially accepted name for the gas.
Use only approved regulators, tubing, and hose connectors.
Arrange gas hoses where they will not be damaged or stepped on and where things will not
be dropped on them.
Check the condition of pipes, hoses, and connectors regularly, and replace any damaged
parts.
Perform periodic gas leak tests at all joints and seals of the gas system by applying an
approved gas leak detection fluid.
Using gas cylinders
Store cylinders in accordance with the applicable national, state, and/or local regulations and
standards.
Legibly mark cylinders to clearly identify the contents and status—full or empty. Use the
chemical name or commercially accepted name for the gas.
When gas cylinders are stored in confined areas, such as a storage room, make sure that
ventilation is adequate to prevent toxic or explosive accumulations of gas.
Do not store acetylene in close proximity to oxidizing gases, such as nitrous oxide.
Store all gas cylinders upright, with the valve cap in place, in a dry, well ventilated area,
protected against temperature extremes.
Do not locate gas cylinders near exits and exit routes, elevators, gangways, or in locations
where heavy moving objects may strike or fall against them.
Locate gas cylinders away from heat sources, including the direct rays of the sun, and away
from ignition sources and readily ignitable substances, such as gasoline, oil, or combustible waste. Compressed gas cylinders should not be subjected to temperatures above 52 °C (126 °F).
Locate all gas cylinders only in a vertical position, fastened securely to an immovable
bulkhead or a permanent wall.
Move cylinders with a suitable hand truck after making sure that the valve cap is securely in
place and that the cylinder is properly fastened to the hand truck.
Do not attempt to refill gas cylinders.
3-15
Page 60
Preparing Your Lab
N
Two types of flame are commonly used for atomic absorption analyses with flame atomization: the air-acetylene flame and the nitrous oxide-acetylene flame. The specifications and requirements for the burner gases are given below:
Gas Purity Outlet Gauge Pressure
Acetylene (ethyne, C2H2) 99.6%, atomic
kPa Bar psig
Pmin Pmax Pmin Pmax Pmin Pmax
80 96 0.8 1.0 12 14 absorption grade in acetone (propan-2-one)
Air
oxygen < 20%; particle
300 400 3.0 4.0 44 58
free; oil free; water free
itrous oxide (dinitrogen
monoxide, N
2
O)
99% 300 400 3.0 4.0 44 58
Never set the outlet pressure to a value exceeding the maximum pressure (Pmax).
Warning: Acetylene – Explosion Hazard
Acetylene can decompose explosively at pressures higher than 103 kPa/ 1.03
WARNING
bar/15 psig.
Α
lways make sure that the acetylene outlet gauge pressure is below these
values.
Air supply requirements
For the burner system, an air supply that is clean, dry, and oil-free is required. The air supply should provide a minimum of 28 liters/min (1 cubic foot/min) at a minimum pressure of 300 kPa (44 psig). The air supply should be suitably filtered to remove oil, water, and solid particles.
Air compressor
A small oil-less air compressor meeting the air supply requirements is available from PerkinElmer. The use of an air dryer filter is strongly recommended to remove entrained water, oil, water aerosols, and solid particles from compressed air lines. Water in the air supply can cause severe damage to the pneumatics.
3-16
Page 61
AAnalyst 200 User’s Guide
M
CAUTION
Risk of damage to the instrument
Water in the air supply can cause damage to the pneumatics.
Use only dry air. Bleed compressor tanks to remove water at least once a week
Air compressors are generally uncomfortably noisy to have in the immediate vicinity of the instrument. Whenever possible, locate them at some distance from the laboratory in an area providing suitable ventilation.
Air cylinders
Cylinders of compressed air can also be used but are recommended only as an emergency or short-term solution. A standard #1 size air cylinder will last only about five hours. It is common to find other than 20% oxygen in air cylinders. This can cause erratic burner operation and non­reproducible analytical results and, in extreme cases, may cause a potential safety hazard.
In general, if cylinder air is to be used, it is important to specify compressed air rather than breathing air and that the oxygen content is less than 20%.
Warning: Flashback Hazard
PerkinElmer burner systems are designed for use with compressed air. The use
WARNING
of oxygen can cause an explosion in the burner system, and oxygen-enriched air can cause a flashback of the flame.
Never use oxygen or oxygen-enriched air with PerkinElmer burner systems. Use only compressed air
Nitrous oxide supply requirements
Warning: Nitrous oxide is a strong oxidant and an asphyxiant
WARNING
Nitrous oxide is an asphyxiant; it can be considered a potential hazard whenever it is stored or used in a confined space. Spontaneous combustion may occur if nitrous oxide comes into contact with grease, oil, or other organic material.
Store nitrous oxide cylinders in a well-ventilated area outside the
laboratory.
Do not store nitrous oxide cylinders in close proximity to flammable gases
(e.g., acetylene).
.
ake sure tubing and fittings carrying nitrous oxide are free of grease, oil,
or other organic material.
3-17
Page 62
Preparing Your Lab
Cylinder information
The nitrous oxide supply should be at least 99% pure. Cylinders of nitrous oxide are available from local suppliers. Nitrous oxide is supplied in the liquid state, initially at a pressure of about 5000 kPa (725 psig). A size 1A cylinder of nitrous oxide contains about 14800 liters (523 cubic feet) and will typically last for 10–12 hours of burning time.
Heated regulator
Since the nitrous oxide is in liquid form, the pressure gauge does not give a true indication of how much nitrous oxide remains in the cylinder until the pressure starts to fall rapidly as the residual gas is drawn off. When nitrous oxide is rapidly removed from the cylinder, the expanding gas causes cooling of the cylinder pressure regulator and the regulator diaphragm can freeze. This can cause erratic flame conditions or, in the most extreme case, a flashback.
It is therefore advisable to use a double-stage or heated regulator for nitrous oxide to prevent freezing of the diaphragm and a loss of pressure regulation. A suitable nitrous oxide pressure regulator is available from PerkinElmer.
Acetylene supply requirements
Warning: Acetylene – Explosion Hazard
Acetylene can decompose explosively at pressures higher than 103 kPa/
1.03 bar/15 psig.
Always make sure that the acetylene outlet gauge pressure is below these
WARNING
Grade
Suitable acetylene (ethyne), such as ‘atomic absorption grade’ or welding grade, typically has a minimum purity specification of 99.6% with the actual assay being about 99.8%. Acetylene is supplied dissolved in acetone (propan-2-one) to prevent explosive decomposition. You should not use acetylene that is dissolved in any other solvent. Experience indicates that solvents can be carried through the regulator with the acetylene and can weaken, to the point of rupture, the walls of the tubing carrying the acetylene to the burner, causing a potential hazard.
Cylinder information
A size 1A acetylene cylinder contains about 8500 liters (300 cubic feet) of acetylene and usually lasts about 30 hours of burning time with an air-acetylene flame. With a nitrous oxide-acetylene flame, the same size cylinder will last about 10 hours.
values.
Acetylene can react with copper to form a readily explosive compound.
Never use copper tubing or fittings for acetylene gas lines.
3-18
Page 63
AAnalyst 200 User’s Guide
Flash arrestor
PerkinElmer recommends the use of the Flash Arrestor (P/N N930-0068). The flash arrestor provides a positive gas shutoff by checking the reverse flow under all conditions. It also stops the gas supply when a flashback occurs. The flash arrestor can be directly connected to the acetylene fuel line.
Piping/tubing/fittings
Use galvanized iron, steel, or wrought iron tubing. Never use copper tubing with acetylene. Joints may be welded or made up of threaded or flanged fittings, typically stainless steel or aluminum, or brass composed of less than 65% copper. Rolled, forged, or cast steel or malleable iron fittings may also be used. Cast iron fittings cannot be used safely for acetylene lines.
Safe use of acetylene
Periodically check for the presence of acetylene in the laboratory atmosphere, especially
near the ceiling.
Never allow the outlet gauge pressure to exceed 103 kPa (1.03 bar, 15 psig); acetylene can
explode spontaneously above this pressure.
Prevent acetylene from coming into contact with copper, silver, mercury or gaseous
chlorine.
Solvent carryover.
If the acetylene cylinder pressure falls to below 600 kPa / 6.0 bar / 87 psig (at 20 °C), some of the solvent used to stabilize the acetylene (such as acetone) may be carried over into the burner. This could influence the characteristics of the burner with a resulting influence on the analytical results. Change to a new acetylene cylinder when the cylinder pressure falls below this value.
3-19
Page 64
Preparing Your Lab
Electrical Requirements
The following section describes the power requirements for all the components that make up the AAnalyst 200 system.
The AAnalyst 200 spectrometer requires a stable single phase electrical supply of 100–230 VAC (± 10%), 50 Hz or 60 Hz (± 1%), capable of providing 300VA maximum, that has a correctly wired protective earthing system (ground connection).
The line power supply should be free of line transients in excess of 50 V peak. If the electrical supply voltage produces large AC line voltage fluctuations, a qualified electrician should install a voltage regulator between the electrical outlet and the instrument.
The line power supply point must be within 2 meters of the rear of the spectrometer. We recommend that you connect the spectrometer, computer, printer, and any accessories to the same phase of the line power supply and the same protective earth. The line power supply must conform to local safety regulations and be checked by a qualified electrician before you connect the instrument to line power.
Installation Category
The AAnalyst 200 is able to withstand transient overvoltage according to Installation Category II as defined in IEC 1010-1.
Pollution Degree
The AAnalyst 200 will operate safely in environments that contain nonconductive foreign matter up to Pollution Degree 2 in IEC 1010-1.
Instrument
Grounding circuit continuity is vital for safe operation of equipment. Never operate equipment with grounding conductor disconnected.
WARNING
The instrument power cord is to be connected to a circuit with a protective circuit breaker or fuse.
NOTE: NEMA: The National Electrical Manufacturer’s Association; CEE: International Commission on
Rules for the Approval of Electrical Equipment.
3-20
Page 65
AAnalyst 200 User’s Guide
Lamps
These are often called primary source lamps, atomic absorption source lamps, spectroscopic lamps, or line sources.
You need to provide the lamps for your particular applications. Please refer to the PerkinElmer Instruments e-ssentials catalog for a list of available lamps.
3-21
Page 66
Preparing Your Lab
3-22
Page 67
System
Description
4
Page 68
Page 69
AAnalyst 200 User’s Guide
Introduction
The AAnalyst 200 instrument line is self-contained with its built-in touch screen and burner system to perform Flame AA determinations.
There are three versions of the AAnalyst 200 instrument. Its basic features (AAnalyst 200 P/N B315-0050) include a touch screen, a single lamp and a burner system with automatic, computer­programmed flame gas control. It is designed for optimized, cost-effective, automatic flame and mercury/hydride analyses. The second version, AAnalyst 200 (P/N B315-0060), has all the basic features plus it contains a built in deuterium background corrector and supports one EDL lamp. The third version, AAnalyst 200 (P/N B315-0070), features an automated lamp turret and supports two EDL lamps.
This chapter gives a complete description overview of the AAnalyst 200. The optional autosamplers are described later in this manual.
4-3
Page 70
System Description
AAnalyst 200 Overview
Figure 4- 1 The AAnalyst 200 Instrument
The AAnalyst 200 Atomic Absorption Spectrometer is a double-beam atomic absorption spectrometer for flame and mercury/hydride analysis. It is a sophisticated analytical system capable of performing automated single element determinations.
4-4
Page 71
AAnalyst 200 User’s Guide
The AAnalyst 200 Spectrometer System
Concept
The AAnalyst 200 spectrometer from PerkinElmer is a compact benchtop flame atomic absorption spectrometer featuring:
Double beam optical system optimized for maximum radiation throughput.
Burner system for flame atomization.
Deuterium background correction system (optional).
System components
The major components of the system are:
The spectrometer containing the optical system, the electronics, and the atomizer
compartment.
Τhe touch screen is the interface between you and the spectrometer, enabling you to
control the spectrometer in order to perform your analyses easily and conveniently.
The burner system.
The printer. (This is optional but strongly recommended.)
System control
To control the spectrometer and perform your analyses, you use the touch screen. When you switch on the spectrometer system it performs an initialization procedure to check the instrument. While the spectrometer is operating, the control system continually monitors many instrument functions. It generates appropriate messages when a problem occurs and prevents potentially hazardous situations from arising.
4-5
Page 72
System Description
General Lamp Information
There are a number of different lamps that can be used with the AAnalyst 200 instrument: the Lumina Hollow Cathode Lamp (HCL) and the Electrodeless Discharge Lamp (EDL) (coded and non-coded). The Intensitron Hollow Cathode Lamp (coded and non-coded), when used with an adapter, will function on the AAnalyst 200 instrument. All three types of lamps can also be used in the AAnalyst instrument.
The type of primary source lamp required for your instrument depends on the elements to be determined. Either hollow cathode or electrodeless discharge lamps may be used. Hollow cathode lamps (HCLs) are excellent for most elements; however, there are a number of “difficult” elements for which an improved light source is desirable. Electrodeless discharge lamps (EDLs) are more intense than the HCLs. Most EDLs provide better lamp life, stability and in some cases better sensitivity.
The 3 types of lamps that can be used in the AAnalyst 200 are listed:
the Lumina hollow cathode lamp (HCL)
the coded Electrodeless discharge lamp (EDL System II)
the coded and non-coded Intensitron hollow cathode lamp (HCL) with adaptors
The basic AAnalyst 200 instrument has a mount for a single lamp. It can, however, be purchased with an automatic four-lamp turret.
An Explanation: Hollow Cathode Lamps (HCLs)
The specific determination of one element in the presence of others is possible with the use of special light sources and careful wavelength selection in a spectrometer. A single-element lamp emits specific wavelengths of light for the element from which the cathode is constructed. There are single-element lamps for all elements commonly measured by atomic absorption. Hollow cathode lamps (HCLs) are ideal for determining most elements by atomic absorption spectroscopy. "
Refer to figure below. The basic components of an HCL are a glass cylinder filled with an inert gas (argon or neon) at low pressure, an anode, a cathode, and a quartz window with a graded seal to be used at wavelengths below 230 nanometers (nm), or a UV-transmitting glass window for lamps to be used at longer wavelengths. The cathode is a hollow cylinder of the metal whose spectrum is to be produced. Both the anode and cathode are sealed in the glass cylinder. The quartz graded seal or glass window is fused to the glass cylinder.
Figure 4- 2 The workings of a hollow cathode lamp (HCL).
4-6
Page 73
AAnalyst 200 User’s Guide
Hollow cathode lamps emit light by the following process: the fill gas is ionized when an electrical potential is applied between the anode and the cathode. The positively charged ions collide with the negatively charged cathode and dislodge individual metal atoms in a process known as “sputtering.” These sputtered metal atoms are excited through impact with fill gas ions. Light of the specific wavelengths for that element is emitted when the atom “decays” from the excited state to a more stable state.
Lumina Hollow Cathode Lamps
The Lumina hollow cathode lamps were designed for use with the AAnalyst series of atomic absorption spectrometers. They are the latest innovation in optimizing light sources for AA spectrometers. These lamps provide fully automated wavelength, current and slit selection which results in reduced setup times for easy and accurate operation.
Lumina Hollow Cathode Lamp
Socket
Figure 4- 3 Drawing of a Lumina hollow cathode lamp.
The Lumina HCLs are manufactured without cables. When they are used with the AAnalyst instrument, they are installed simply by sliding the lamp into the single lamp mount or turret. Lamp alignment occurs automatically if you have the model with the automatic four-lamp turret feature.
Each Lumina HCL includes a milliampere-hour timer in its base. This allows you to monitor the lamp usage easily and accurately for Good Laboratory Practices compliance as well as lamp replacement scheduling. With a range of 0 to 10,000 milliampere-hours, the meter measures the number of milliampere-hours (i.e. lamp current X hours of operation) that your lamp has operated.
Plug
Intensitron Hollow Cathode Lamps
The AAnalyst series spectrometers can also be used with the older coded or non-coded Intensitron hollow cathode lamps (HCLs). With the use of the proper adapter cable, Intensitron HCLs can be used with AAnalyst instruments.
4-7
Page 74
System Description
If you have a coded Intensitron HCL, there is a universal adapter cable (N305-0196) to be used with the lamp. However, lamp coding is lost, therefore lamp and element parameters must be set manually when coded HCLs and universal adapter cables are used.
Intensitron
Hollow Cathode
Lamp
(coded)
Figure 4- 4 The Intensitron hollow cathode lamp with universal adapter cable (N305-0196).
Universal Adapter Cable
for the Coded
Intensitron HCL's
(N305-0196)
Non-coded HCLs can also be used with the appropriate adapter cable (N305-0197); however, lamp and element parameters must be set manually when non-coded lamps are used.
Intensitron
Hollow Cathode
Lamp
(non-coded)
Adapter Cable
for the Non-coded
Intensitron HCL's
(N305-0197)
Figure 4- 5 The noncoded Intensitron hollow cathode lamp with adapter cable (N305-0197).
4-8
Page 75
AAnalyst 200 User’s Guide
Electrodeless Discharge Lamps (EDL System II)
While hollow cathode lamps are an ideal source for determining most elements by atomic absorption, for volatile elements, an electrodeless discharge lamp (EDL) is recommended instead. EDLs provide greater light output for improved detection limits and precision, along with increased sensitivity and working curve linearity. EDLs also have a longer life than corresponding hollow cathode lamps. For certain elements (such as arsenic and selenium), EDLs will also provide improved sensitivity.
Lamp Sleeve
Lamp Driver
Locking Pin
Note: This cable connects to the "Lamp1" or "Lamp2" receptacle on the ba ck of the ED L Pow er Su pp ly.
Note: This co ding plug is not used with the AAnalyst Instruments.
Note: Use this coding plug with the
AAnalyst
Instruments.
Figure 4- 6 An Electrodeless Discharge Lamp (EDL).
An EDL consists of a quartz bulb under an inert gas atmosphere, containing the element or a salt of the element for which the lamp is to be used. When an RF field is applied to the bulb, the inert gas is ionized and the coupled energy excites the vaporized analyte atoms inside the bulb, causing them to emit their characteristic spectrum. The emission spectra produced is typically much more intense than that from hollow cathode lamps, frequently tenfold larger or more.
Lamp Installation
When you first receive
your instrument, there is no lamp source installed. For initial lamp installation details see, Chapter 6, Operation of the Instrument in this manual. If you need to replace an existing lamp see, Chapter 9, Maintenance later in this manual contains the various lamp replacement
procedures.
4-9
Page 76
System Description
Lamp Alignment
Refer to Chapter 6, Operation of the Instrument, in this manual for the procedures to align a lamp in an AAnalyst instrument.
Lamp Maintenance and Cleaning
There are really no maintenance procedures for any of the PerkinElmer hollow cathode lamps. If you notice fingerprints on the “window” of your lamp, wipe them off with a soft, dry, lint free cloth. Do not use chemicals or water. See the section below regarding lamp life and when to replace them.
Lamp Life
Both hollow cathode and electrodeless discharge lamps have a finite lifetime. Electrodeless discharge lamps typically last longer than hollow cathode lamps for the same element. If, while aligning a lamp, you are not able to obtain a satisfactory energy value, this may indicate that the lamp is starting to deteriorate. As a hollow cathode lamp reaches the end of its useful life, the current on the lamp may be increased up to the maximum current recommended on the label. Once the intensity of a lamp begins to deteriorate, it should be replaced.
Lamp Troubleshooting
Refer to Chapter 10, Troubleshooting, in this manual for lamp troubleshooting information.
To avoid damages to the instrument, do not insert EDL’s or non­linear lamps too far into the lamp holder.
CAUTION
4-10
Page 77
AAnalyst 200 User’s Guide
The Deuterium Arc Background Corrector
An Explanation: The Deuterium Arc Background Corrector
The Deuterium Arc Background Corrector is an internally mounted assembly that provides correction for molecular absorption and light-scattering. The spectrometer uses a deuterium arc lamp to correct for background absorbance caused by non-atomic species in the atom cloud.
When the deuterium arc background corrector is used, two alternating sets of measurements are made: one with the primary source and a second with the background corrector source. The primary source measures a combination of background and atomic absorption. The continuum source measures only the background. The difference between the two measurements yields the atomic absorption.
A deuterium arc source is used for maximum wavelength coverage and intensity. The AAnalyst can accurately handle combined atomic plus background absorption signals as large as 3 absorbance units.
The background corrector is internally installed, with its power supplied from the instrument. Power to the source is automatically and continuously adjusted to be in the optimum range for operation. The deuterium arc lamp is easily accessible, allowing convenient replacement of the lamp by the users. Instructions to replace the lamp are included in the Maintenance chapter.
Deuterium Alignment
Refer to Chapter 6, Operation of the Instrument, in this manual for the procedures to align the background corrector in an AAnalyst instrument.
4-11
Page 78
System Description
The Burner System
The burner system used in the spectrometer system is of the premix design. The measurement solution is sprayed by a pneumatic nebulizer into the burner chamber in the form of an aerosol. This aerosol is thoroughly premixed with the burner fuel and oxidant gases in the burner chamber. This gas mixture then passes through the burner slot into the flame.
The burner system comprises the following major components:
Nebulizer
Burner chamber (including the flow spoiler and end cap)
Burner head
Drain system
Gas control system (incorporated in the spectrometer)
Flow Spoiler
Impact Bead Structure
Spacer
O-Rings
Nebulizer Bo dy
Nebulizer Sid e Arm
Burner
Head
Burner
(Mixing)
Chamber
End Cap
Drain
Figure 4- 7 Major components of the burner system
4-12
Page 79
AAnalyst 200 User’s Guide
How the Burner System Works
The basic instrument includes a burner system for flame sampling operation. A system block diagram is shown in the next figure. As previously mentioned, the heat produced by the burner converts the sample into free atoms that can absorb the light from the source lamp for the element(s).
A sample solution is aspirated through the nebulizer and sprayed as a fine aerosol mist into the burner chamber. Here, the sample mist is thoroughly mixed with the fuel and either air or nitrous oxide gases. Fuel gas enters the chamber through the fuel inlet hose while the oxidant enters through the nebulizer side-arm and the AUX oxidant inlet.
To prevent incomplete sample atomization, a plastic flow spoiler is placed inside the chamber directly in front of the nebulizer. The nebulizer utilizes a ceramic, one-piece, impact-bead structure which improves nebulization efficiency. Larger sample droplets not carried around the flow spoiler by the gas flow are forced by the spoiler to the bottom of the chamber. This, in turn, allows only the finer sample droplets to be carried into the chamber. The larger droplets are removed through the drain.
Once the sample mist is thoroughly mixed in the chamber, it is carried to the burner head. Once there, the flame causes dislocation and atomization.
The gas control system monitors and controls fuel and oxidant flows. Burner operation, including flame ignition, flame shutoff and gas flow settings, is controlled through the instrument touch screen (AAnalyst 200).
The next figure shows a “block diagram” of how the AAnalyst burner system flows, from the gas supplies, to the gas box, to the burner hardware itself and out to the drain vessel.
4-13
Page 80
System Description
Burner
Assembly
Ne bulizer O x idant
Auxiliary
Oxidant
Fuel
Sample
Capillary
to
(int ernal
Gas Box
instrument)
2
NO In
Air In
Ignit or A r m
End Cap
Nebulizer
Tube
Dra in Tube
Drain
Vessel
Sample
Acet y lene In
Red
Acetylene
Adapt er (0990-3215)
(CGA No. 300 or 510)
Black
Tank
Blue
Air/A rgon Hos e
Assem bly
(0057-0567)
Nitro us Oxide
Tank
Air Dry er F ilter (0047-0652)
or Air/Acetylene Filter (N930-1398)
Nitrous Oxide Regul at or (includes
built -in heat er) (0303 -0204, 117V;
0303-0349, 230V)
Nitrous Oxide Hose Assembly (0047-0258)
Ace ty l ene Regulator
(0303-0106)
Acetylene Hose Assembly (0057-0559)
Flash Arre sto r
(N930-0068)
Fitting
Air
Supply
(0990-3946)
Male Connector
1/4 in. Pipe to 5/16 in.
5/16 in.
Com pre ssio n
Compression Fitting
Figure 4- 8 Block diagram of the complete burner system.
4-14
Page 81
AAnalyst 200 User’s Guide
The Burner Head
All the burner heads are made from titanium. The 10-cm, air-acetylene burner head is standard with the AAnalyst instruments. All the available burner heads are listed next in numerical order.
N040-0100: 5-cm, single-slot burner head (for nitrous oxide-acetylene operation).
N040-0101: 5-cm, single-slot burner head (for air-acetylene operation). It features a wider slot
than the nitrous oxide burner head.
N040-0102: 10-cm, single-slot burner head (for air-acetylene operation - standard with all
burners).
N040-0103: 10-cm, three-slot burner head (for analysis of high dissolved solids content
solutions with air-acetylene flames).
The 5-cm single-slot air-acetylene burner head is rotatable to 90 degrees.
N040-0100, 5-cm (2-in.) Nitrous Oxide Operation
N040-0102, 10-cm (4-in.) Air-Acetylene Operation
Positioning
Notches
Magnetic Identification Strip
N040-0101, 5-cm (2-in.) Air-Acetylene Operation
90°
60°
45°
N040-0103, 10-cm (4-in.) 3-Slot Air-Acetylene Operation
Figure 4- 9 The four available burner heads.
4-15
Page 82
System Description
The Burner Mount
The quick change burner mount provides horizontal, vertical, and rotational adjustment. The quick-disconnect release lever simplifies removal of the burner assembly.
The Ignitor Assembly
The ignitor assembly includes the glow plug, the ignitor solenoid and ignitor arm, the burner head interface circuit, and the flame detect circuit. When the ignitor is activated (automatically), the glow plug is energized and heats up. The ignitor solenoid, which is also energized, swings the ignitor arm into place. The ignitor arm is a gas-capture tube through which the gases flow from the burner to the glow plug. The gases in the tube are ignited by the heat of the glow plug, which, in turn, ignite the flame at the burner head slot.
The gas controls
The gas controls provide control of the flow rates of the fuel gas (acetylene) and the oxidant (air or nitrous oxide).You select the gas flow rates and optimize the flame via the touch screen. The gas control system is incorporated in the spectrometer. The system includes safety interlocks for the flame. If the fuel or oxidant pressure falls below a minimum value the flame is shut down in a safe sequence. The system also includes automatic sequencing of the gases when the nitrous oxide-acetylene flame is ignited or extinguished.
The burner will be shut down in a safe sequence in the event of a power failure.
NOTE: In case of emergency, use the red emergency button located on the upper left hand side of the
instrument.
The drain system
During flame operation, measurement solution that condenses out in the spray chamber runs to waste through the siphon interlock drain trap and is collected in the drain vessel.
The siphon interlock drain trap is integral with the burner end cap and includes an internal trap to make certain that back pressure is applied to the burner chamber. This ensures that a fuel/oxidant mixture cannot enter the drain vessel. There is a magnetic float in the drain trap that shuts down the flame if the drain system is not functioning properly. If the liquid level in the trap is too low, the switch is activated and the flame is shut down.
4-16
Page 83
AAnalyst 200 User’s Guide
A
n
A
l
y
a
t
s
2
0
0
A
i
t
m
o
c
A
s
b
o
o
r
t
p
i
e
p
n
S
o
t
r
c
m
r
e
e
t
Spiral Wrap
Interlock Cable
Drain Interlock Cable Connector at Burner
Drain Tubing
Retainer Cap
Drain interlock Float Switch Assembly
Drain Loop
Waste Vessel
Figure 4- 10 The burner drain system and its parts
Burner Installation
The entire installation of the burner is covered in Chapter 6, Operation of this User’s Guide Manual. The basic steps are bulleted below for your reference. Directions on how to perform each step are discussed in detail during installation.
installing the nebulizer on the burner
checking that the burner end cap is secure
connecting and securing the gas hoses to the burner
installing and securing the burner head
4-17
Page 84
System Description
installing the quick-change burner mount assembly into the instrument
installing and connecting the burner drain system
Burner System Maintenance
Refer Chapter 9, Maintenance, in this User’s Guide manual for extensive burner system maintenance information.
Burner System Troubleshooting
We recommend that you consult the troubleshooting information included in this manual before you call a PerkinElmer service engineer. Refer to Chapter 10, Troubleshooting, for your burner system troubleshooting information. Included are suggestions to help you determine systematically whether instrument problems are due to improper analytical techniques, improper selection of instrument parameters, or a malfunction of the instrument.
Burner Spare Parts and Accessories
Refer to the end of Chapter 9, Maintenance, for burner system spare parts information.
4-18
Page 85
AAnalyst 200 User’s Guide
Connecting the Electrical Supply
A Review of the Electrical Specifications
The electrical supply voltage must be between: 100-230 VAC ± 10%
The electrical supply voltage should be free of line transients in excess of 50V peak. If the supply voltage produces large AC line voltage fluctuations, install a voltage regulator between the AC outlet and the instrument.
The spectrometer requires either a 100/120 VAC standard 15 amp grounded outlet, or a 230 VAC outlet. In addition, the instrument consumes 300 volt amps (maximum) when configured for background correction.
The AAnalyst 200 requires a grounded AC outlet providing voltage corresponding to the voltage range marked on the instrument (100-230 VAC).
WARNING
Do not remove or alter the ground prong on a three-prong power plug.
4-19
Page 86
System Description
Installing the AC line Cord
Grounding circuit continuity is vital for safe operation of equipment. Never operate equipment with grounding conductor disconnected. Do not remove or alter the ground pin on a three-
WARNING
The line cord should have the electrical rating and plug appropriate for your location. See the figure below for the correct cord.
1. Select the proper line cord for your location.
2. Insure that the power switch is off and insert one side of the line cord plug into the instrument
and the other side into the AC supply.
prong line cord.
North America NEMA-5-15 0998-8986
Europe CEE 7 "Schuko" 0999-1415
Europe Switzerland 0999-1413
Europe Italy 0999-1422
Denmark 0999-1416
Old British Standard BS 546 India 0999-1423
British Standard BS 1363 United Kingdom 0999-1414
Austra lia ETSA S/86 0999-1417
Israel 0999-1424
Japan 0999-1425
Figure 4- 11 The AC line cord selections for the AAnalyst instrument.
4-20
Page 87
AAnalyst 200 User’s Guide
r
Mercury/Hydride Determination System (MHS-15)
Mercury/Hydride Systems are proven accessories for high sensitivity determinations of mercury and metallic hydride-forming elements, including As, Se, Sb, Te, Bi and Sn by atomic absorption spectroscopy.
The basic components of a mercury/hydride system are the reaction vessel and sampling cell. The sample solution reacts with a reducing agent in the reaction vessel. A spontaneous reaction occurs to form the gaseous hydride or to reduce mercury to its elemental form. The gaseous products are carried to the sample cell, a quartz cell placed in the light path of the spectrometer. The quartz cell is heated to dissociate the hydride gas into free atoms, or in the case of mercury, to prevent condensation of the water vapor.
Plunge
Reaction Flask Socket with Bayonet Lock
Figure 4- 12 The MHS-15 and its parts.
Lugs
Reaction Flask
4-21
Page 88
System Description
t
The MHS-15 is a manual system. The quartz cell is heated by the flame. Key operating parameters are controlled by the instrument touch screen. Data collection, calibration, and calculation functions are performed by the instrument.
The MHS-15 analyzer assembly is freestanding and is located near the sample compartment of the spectrometer. This assembly includes a reaction flask, a reservoir for the reducing agent, and all pneumatic components for control of the carrier gas and transport of metallic vapors to the quartz cell.
The cell assembly consists of a quartz cell and a mount. The mount installs on the AAnalyst burner head, positioned above the single slot in the burner head, permitting the quartz cell to be heated in the flame.
Quartz
Cell
Sample Inle
Transfer Hose
Burner
Head
Burne r Chamber
Figure 4- 13 The MHS-15 sample cell mounted on the burner.
For detailed information on the MHS-15, refer to the manual that came with the System.
4-22
Page 89
Installation
5
Page 90
Page 91
AAnalyst 200 User’s Guide
Introduction to the Installation Procedure
Before you start to install your AAnalyst 200 spectrometer, please take time to read through this self­installation guide and become familiar with the steps involved. Please make sure your laboratory meets all of the requirements listed in this chapter, before you begin to install the system.
NOTE: A CD-ROM is provided with the spectrometer that contains a video showing the complete installation
procedure. We strongly recommend that you view this video before you attempt to install your spectrometer.
Summary of the installation procedure
This summary lists the steps you must perform to install the spectrometer.
1. Inspect the shipment.
2. Unpack the spectrometer.
3. Place the spectrometer on the bench.
4. Install the touch screen.
5. Connect the gas hoses.
6. Set up the burner system.
7. Install the burner door.
8. Install the sample tray.
9. Connect the spectrometer to the electrical supply.
10. Perform the post-installation checks.
Tools required for the installation
You will need the following tools during installation of the spectrometer:
small flat-bladed screwdriver
two wrenches for gas fittings
5-3
Page 92
Installation
When the Spectrometer Arrives
Inspect the shipment
Upon receipt of the instrument, inspect the exterior of the shipping carton for obvious signs of damage. The carton should arrive unopened and undamaged. If the carton has been damaged, file a claim with the authorized carrier immediately and inform your nearest PerkinElmer office. A representative of the shipping carrier must be present for insurance purposes.
How to unpack the spectrometer
Carefully read these instructions before you unpack the AAnalyst 200 to avoid damaging the instrument. The AAnalyst 200 is shipped in a single carton. The carton contains the instrument in the lower portion of the box with an inner tray designed to hold items in the upper portion of the box. The touch screen, atomizer compartment door, drain bottle, sample shelf, user guides, and small parts are contained in the inner tray.
1. Place the instrument carton on the floor in front of the bench on which the instrument will be placed.
2. Cut the two bands that hold the carton together, lift off the top of the carton.
3. Pull out the two foam pieces that secure the instrument in the box.
4. Remove the three small cardboard boxes from the carton.
5. Lift up and remove the carton sleeve (the sides of the box).
6. Next remove the protective wrapping around the instrument (not shown).
7. Prepare to lift the instrument out of the carton. You will need two people to do this. The bench
should be nearby.
The instrument weighs 49 kg (108 lb). Use proper lifting posture when lifting the instrument out of the carton. Be aware that most of the weight is located at
the right front corner of the instrument. It will want to tip forward when it is
WARNING
lifted out of the carton, so be prepared to support that corner.
5-4
Page 93
AAnalyst 200 User’s Guide
Shipping Carton Top
Molded Foam Padding
Cardboard Boxed Items (as required)
Carton Sleeve
Shipping Carton Base
Figure 5- 1Exploded view of the shipping carton
5-5
Page 94
Installation
Placing the Spectrometer on the Bench
Risk of injury through lifting. The instrument weighs approximately 49 kg (108 lbs.).Improper lifting can cause painful and sometimes permanent back injury.
Use proper lifting postures and/or mechanical lifting aids to lift or move the instrument.
WARNING
1. Prepare to lift the instrument out of the carton. You will need two people to do this. The bench should be
nearby.
2. Position a person on each side of the instrument as shown. Squat down and place both hands underneath
the instrument – palms up. Spread your hands as far apart as possible (against the foam corners) while grasping the underside of the instrument.
3. With knees bent, simultaneously lift the instrument out of the carton as you rise to a standing position.
Carefully place the instrument on the bench.
Foam Forms
Shipping Carton Base
IMPORTANT!
Use proper lifting posture when taking the instrument out of the shippi ng carton base.
Figure 5- 2 Lifting the instrument out of the box
5-6
Page 95
AAnalyst 200 User’s Guide
Carton recycling instructions
The carton can be folded down for easy storage. You can store the carton for future use (to transport the instrument), or break it down completely and give it to your local recycling center.
5-7
Page 96
Installation
AAnalyst 200 Shipping List
The following parts are included in the shipment.
Table 5- 1 Shipping List
Description
AAnalyst 200 Instrument
Burner Door (in a box)
Touch Screen (in a box)
Drain Assembly (including the drain vessel)
Installation Kit (including the multimedia CD)
5-8
Page 97
AAnalyst 200 User’s Guide
AAnalyst 200 Spares Kit
The following parts are included in the AAnalyst 200 Spares Kit (P/N B315-0021). The kit is shipped with the instrument.
Table 5- 2 Spares Kit
Part No. Description QTY
N315-0188 Nebulizer Kit 1 B019-1060 Nebulizer tubing, 1000mm 1 B315-0458 Nebulizer tubing, 300mm (for
manual sampling) B050-1696 Air Filter 1 0941-0063 Serial Cable 1 0990-3004 Hose Clamp 1 0998-8986 Line Cord-North America 1 0999-1415 Line Cord Europe 1 0303-1573 Burner Cleaning Tool 1 N315-0188 AA 200/400 Nebulizer Kit 1 N101-0235 Quick Start CD 1 0990-9390 AA Newsletter Subscription
Form 0993-6643 AAnalyst 400 Safety Booklet 1 0943-2046 Software Utilities CD 1 0993-6550 AA Instrument Maintenance
Log 0993-8893 Software/ Spectrometer
Registration Card
1
1
1
1
5-9
Page 98
Installation
Installing the Touch Screen
The touch screen is packed separately for shipping. To facilitate installation of the touch screen you must temporarily remove the lamp door.
Figure 5- 3 Removing lamp door
Remove the lamp door and install the touch screen
Remove the lamp door
1. Loosen the lower securing screw with a screwdriver and place in a secure place. Gently push out the
upper part of the door from the upper pin and maneuver the lower part of the door out from the instrument. Keep the screw for when you remount the lamp door.
2. Lift out the lamp door and put it in a secure location.
5-10
Page 99
Install the touch screen
AAnalyst 200 User’s Guide
Figure 5- 4 Installing the touch screen
Locate the touch screen (packed separately).
1. Thread the connector cord through the slot above the lamp compartment.
Screen 5- 1 Slot above the lamp compartment
5-11
Page 100
Installation
Slide the touch screen over the slot in the instrument cover and gently move into place.
2.
Screen 5- 2 Mounting the touch screen
3. Locate the screw above the lamp compartment and tighten it.
Screen 5- 3 Screw above the lamp compartment
4. Thread the connector cord into the channel.
Screen 5- 4 Cable Channel
5-12
Loading...