PerkinElmer 5184 Instruction Manual

Page 1
Model 5184
Ultra Low-Noise
Voltage Preamplifier
Instruction Manual
222516-A-MNL-D
Copyright © 2000 PERKINELMER CORPORATION
Page 2
FCC Notice
This equipment generates, uses, and can radiate radio-frequency energy and, if not installed and used in accordance with this manual, may cause interference to radio communications. As temporarily permitted by regulation, operation of this equipment in a residential area is likely to cause interference, in which case the user at his own facility will be required to take whatever measures may be required to correct the interference.
Company Names
PerkinElmer Instruments is part of Perkin Elmer, Inc., and includes the businesses formerly trading as EG&G Princeton Applied Research, EG&G Instruments and EG&G Signal Recovery.
Declaration of Conformity
This product conforms to EC Directives 89/336/EEC Electromagnetic Compatibility Directive, amended by 92/31/EEC and 93/68/EEC, and Low Voltage Directive 73/23/EEC amended by 93/68/EEC.
This product has been designed in conformance with the following IEC/EN standards:
EMC: BS EN55011 (1991) Group 1, Class A (CSPIR 11:1990)
BS EN50082-1 (1992):
IEC 801-2:1991 IEC 801-3:1994 IEC 801-4:1988
Safety: BS EN61010-1: 1993 (IEC 1010-1:1990+A1:1992)
Trademarks
is a trademark of PerkinElmer, Inc.
Page 3

Table of Contents

Table of Contents
Chapter One, Description
1.1 Description ................................................................................................................................................. 1-1
1.2 Options....................................................................................................................................................... 1-2
Chapter Two, Initial Checks
2.1 Introduction ................................................................................................................................................ 2-1
2.2 Equipment Needed ...................................................................................................................................... 2-1
2.3 Procedure.................................................................................................................................................... 2-1
Chapter Three, Operating Instructions
3.1 Introduction ................................................................................................................................................ 3-1
3.1.01 Power Switch.....................................................................................................................................3-1
3.1.02 Input..................................................................................................................................................3-1
3.1.03 Output ............................................................................................................................................... 3-1
3.2 Battery Operation........................................................................................................................................ 3-2
3.3 External Supply Operation..........................................................................................................................3-2
Appendix A, Specifications
Appendix B, External Power Supply
Warranty
................................................................................................................................. End of Manual
i
Page 4
TABLE OF CONTENTS
ii
Page 5
Description
1.1 Description
Chapter
The model 5184 is a medium-impedance (5 MΩ), AC-coupled, voltage preamplifier which features an ultra low-noise FET input stage (800 pV/√Hz at 1 kHz). It has a frequency range of 0.5 Hz to 1 MHz and a fixed gain of 60 dB (×1000). The 5184 is ideal for use with medium impedance cryogenic sources and IR detectors such as HgCdTe, InSb and InAs.
The model 5184 incorporates a special input stage which can be floated to give the ground loop immunity normally associated with differential inputs, without the usual accompanying noise penalty of the input devices required in differential circuits.
The model 5184 can be powered from its own internally housed (alkaline) batteries, an external low voltage supply (±15 V or ±18 V DC) or from a line power supply via the model PS0108 (optional extra). It can also be powered from most PerkinElmer Instruments lock-in amplifiers. Connection of external power to the instrument is by a 5-way DIN connector on the rear panel.
1
Figure 1-1, Model 5184 Front Panel
Figure 1-2, Model 5184 Rear Panel
1-1
Page 6
Chapter 1, DESCRIPTION
1.2 Options
The model 5184 is supplied complete with four 9 V alkaline batteries. The following optional extras are available.
Model PS0108
Model PS0109
Model C0218
Model K0304
Remote Line Power Supply for the model 5184. See Appendix B for details of this unit.
Pack of 20 alkaline batteries for the model 5184
Power cable. This cable allows the user to power the model 5184 from any one of the following PerkinElmer Instruments Lock-in Amplifiers: Model 5102, 5104, 5109, 5110, 5205, 5206, 5207, 5208, 5209, 5210, 5302, 7220, 7225, 7260, 7265 or 7280.
Rack mounting kit. This allows 1 or 2 model 5184 preamplifiers to be mounted in a standard 19" rack.
1-2
Page 7
Initial Checks
2.1 Introduction
The following procedure is provided to facilitate initial performance checking of the Model 5183. In general, the procedure should be performed after inspecting the instrument for shipping damage, but before using it experimentally. If any damage is noted, PerkinElmer Instruments should be notified immediately and a claim filed with the carrier. The shipping container should be saved for inspection by the carrier.
Should any difficulty be encountered in carrying out these checks, contact the factory or one of its representatives.
2.2 Equipment Needed
1) General purpose laboratory oscilloscope.
2) Signal generator capable of providing a 10 mV pk-pk sine wave at 1 kHz.
2.3 Procedure
Chapter
2
1) Set the front-panel press the adjacent indicating that the internal batteries are installed and functioning properly. If the
POWER TEST
the 5184s battery requirements.
Those users who have purchased the PerkinElmer Instruments external
Note:
power supply option can use the external supply to perform the initial checks, in which case step 1 is replaced by the following operations.
a) Make sure that the voltage selector switch on the external power supply is in
the position indicating the line voltage to be used (110 V AC or 240 V AC).
b) Plug the line cord into the external power supply and the external power
supply into the instrument power socket.
c) Set the
adjacent indicating that the external supply is functioning properly.
2) Set the 5184 input position.
POWER
POWER
POWER TEST
indicator does not light, refer to Section 3.2 for a discussion of
switch to the up (
POWER TEST
GROUND / ISOLATE
switch to the down (
button. The associated indicator should light,
EXTERNAL
button. The associated indicator should light,
BATTERY
) position. Then press the
switch to the down (
) position. Then
GROUND
)
3) Connect the oscilloscope to the
4) Allow approximately 1 minute for the model 5184 circuitry to stabilize
OUTPUT
BNC connector.
2-1
Page 8
Chapter 2, INITIAL CHECKS
5) Set the signal generator to 1 kHz, 10 mV pk-pk and connect it’s output to the
6) Monitor the 5184 output; the output level should be 10 V pk-pk.
INPUT
amplitude settings so as to obtain consistency between input settings and output readings.
BNC connector. Use the oscilloscope to monitor the signal generator
7) Set the 5184 input position.
8) The output level should still be 10 V pk-pk.
9) Return the
This completes the initial checks. If the instrument performed as indicated, one can be reasonably sure that it has arrived in good working order and is functioning properly.
POWER
GROUND / ISOLATE
switch to the center (
switch to the up (
) position.
OFF
ISOLATE
)
2-2
Page 9
Operating Instructions
3.1 Introduction
The Model 5184 Preamplifier can be powered from an external power source or from internal alkaline cells as selected from a front-panel switch. In operation, the voltage to be measured is connected to the AC-coupled preamplifier amplifies the signal voltage by a fixed 60 dB (1000 times) which is then output from the
OUTPUT
3.1.01 Power Switch
BNC connector.
INPUT
Chapter
BNC connector. The
3
The three position selected. To operate the 5184 from an external power source, set the switch to the up position. To operate the 5184 from internal batteries, set the switch to the down position. In the center position, the 5184 is unpowered. The status of the selected power source can be ascertained by pressing the LED will light if the power source voltage, whether of the external power source or the 5184 internal batteries, is above the minimum required value for reliable operation.
POWER
switch allows
EXTERNAL, OFF
POWER TEST
, or
BATTERY
button; the adjacent
3.1.02 Input
The 5184 input circuit is of an asymmetrical differential configuration. When the
GROUND / ISOLATE
the BNC input socket is grounded and the input may be used in the conventional single-ended mode. When the (
ISOLATE
pseudo-differential mode. This gives the ground loop immunity normally associated with differential inputs without the accompanying noise penalty associated with differential input stages. In pseudo-differential mode the signal voltages should be connected to the BNC on the front panel. The maximum common mode input voltage is 300 mV peak to peak and the maximum voltage difference between the shell of the BNC signal input connector and the front panel ground terminal is ±600 mV.
) position, the input stage is floated and the input may be used in the
switch is set to the down (
GROUND / ISOLATE
INPUT
socket and the signal ground to the
GROUN
D) position, the screen of
switch is set to the up
GROUND
to be
terminal
3.1.03 Output
The 5184 output can generate greater than 10 V peak to peak signals into loads greater than 100 kΩ. The 450 ohm output impedance provides a convenient 10:1 output attenuator if the amplifier is loaded with 50 ohms, which can be useful if the 5184 is required to drive coax cable greater than one meter in length at signal frequencies approaching 1 MHz.
3-1
Page 10
Chapter 3, OPERATING INSTRUCTIONS
3.2 Battery Operation
In battery powered operation, the 5184 requires four 9 V alkaline cells (type NEDA 1604, PP3 or equivalent). Each battery fits into its own removable tray accessible from the rear panel. When they are exhausted, all four batteries should be replaced together. Care should be taken when exchanging the batteries to observe the correct polarity which is marked on the inside of the battery trays and the top of the batteries. The batteries will provide up to eight hours operation at 1 V rms. output level.
Nickel-cadmium rechargeable batteries can be used if preferred, but there will be a penalty of a reduction in the operating time obtained. In addition it will be necessary to recharge these batteries in an external charger.
3.3 External Supply Operation
External power can be provided by a stand-alone external power supply (see Appendix B) providing ±18 V DC, or the Preamplifier Power Output of most PerkinElmer Instruments lock-in amplifiers which provide ±15 V DC. Connection is by a DIN 5-way socket on the rear panel of the 5184.
Figure 3-2, External Power Input Connector Pinout
Figure 3-1, Battery Type
3-2
Page 11
Specifications
General
Input
Appendix
AC coupled voltage amplifier with fixed x1000 (60dB) voltage gain and a maximum frequency response extending from 0.5 Hz to 1 MHz. Pseudo-differential input and single-ended output via BNC connectors.
Battery powered from internal alkaline batteries or external DC power supplies.
Modes Asymmetrical differential. Front panel ground
terminal provided. Coupling AC Impedance 5 MΩ //50 pF Frequency Response 0.5 Hz - 1 MHz C.M.R.R. > 80 dB (100 Hz to 1 kHz) Max differential input voltage 10 mV pk-pk Max common-mode input voltage 300 mV pk-pk Max signal low potential w.r.t. ground terminal
±600 mV Max input without damage ±15 V DC or 10 V rms. AC @ 50 Hz Noise See Figure A-1 ; typ 800 pV/√Hz @ 1 kHz
A
Output
Figure A-1, Model 5184 Noise Figure Contours (Typical)
Gain x1000 (60 dB) fixed Gain Accuracy ±1% Gain Stability ±800 ppm/° C
A-1
Page 12
Appendix A, SPECIFICATIONS
Power
Dimensions
Impedance 450 Max voltage swing >10 V pk-pk Slew rate > 22 V/µs Polarity Non-inverting Distortion < 0.1% T.H.D.
Internal Four 9 V alkaline batteries provide
External
a) ±15 V or ±18 V DC @
b) 110 V AC or 240 V AC via optional
(excluding connectors) 8.25" wide × 11" deep × 3.5" high
Weight 5.3 lbs. (2.4 kg) excluding optional power supply
approximately 8 hours of use
40 mA
external model PS0108 power supply
(210 mm wide × 279 mm deep × 89 mm high)
A-2
Page 13
Power Supply Unit (Model PS0108)
Appendix
B
Figure B-1, Model PS0108 External Power Supply
Fuse Rating
115 V operation 200 mA 230 V operation 100 mA
Fuse Type 5 mm × 20 mm, 250 V, Slow Blow
B-1
Page 14
Appendix B, POWER SUPPLY UNIT (MODEL PS0108)
B-2
Page 15

WARRANTY

PerkinElmer Instruments Corporation warrants each instrument of its own manufacture to be free of defects in material and workmanship for a period of ONE year from the date of delivery to the original purchaser. Obligations under this Warranty shall be limited to replacing, repairing or giving credit for the purchase, at our option, of any instruments returned, shipment pr epaid, to our Service Department for that purpose, provided prior authoriz ation for such r eturn has been given by an authorized representative of PerkinElmer Instruments Corporation.
This Warranty shall not apply to any instrument, which our inspection shall disclose to our satisfaction, to have become defective or unusable due to abuse, mishandling, misuse, accident, alteration, negligence, improper installation, or other causes beyond our control. This Warranty shall not apply to any instrument or component not manufactured by PerkinElmer Instruments Corporation. When products manufactured by others are included in PerkinElmer Instruments Corporation equipment, the original manufacturers Warranty is extended to PerkinElmer Instruments customers.
PerkinElmer Instruments Corporation reserves the right to make changes in design at any time without incurring any obligation to install same on units previously purchased.
THERE ARE NO WARRANTIES WHICH EXTEND BEYOND THE DESCRIPTION ON THE FACE HEREOF. THIS WARRANTY IS IN LIEU OF, AND EXCLUDES ANY AND ALL OTHER WARRANTIES OR REPRESENTATIONS, EXPRESSED, IMPLIED OR STATUTORY, INCLUDING MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, AS WELL AS ANY AND ALL OTHER OBLIGATIONS OR LIABILITIES OF PERKINELMER INSTRUMENTS CORPORATION, INCLUDING, BUT NOT LIMITED TO, SPECIAL OR CONSEQUENTIAL DAMAGES. NO PERSON, FIRM OR CORPORATION IS AUTHORIZED TO ASSUME FOR PERKINELMER INSTRUMENTS CORPORATION ANY ADDITIONAL OBLIGATION OR LIABILITY NOT EXPRESSLY PROVIDED FOR HEREIN EXCEPT IN WRITING DULY EXECUTED BY AN OFFICER OF PERKINELMER INSTRUMENTS CORPORATION.
SHOULD YOUR EQUIPMENT REQUIRE SERVICE
A. Contact your local PerkinElmer Instruments (Signal Recovery) office, agent, representative or distributor to discuss the
problem. In many cases it may be possible to expedite servicing by localizing the problem to a particular unit or cable.
B. We will need the following information, a copy of which should also be attached to any equipment which is returned
for service.
1. Model number and serial number of instrument
2. Your name (instrument user)
3. Your address
4. Address to which the instrument should be returned
5. Your telephone number and extension
C. If you experience any difficulties in obtaining service please contact:
PerkinElmer Instruments (Signal Recovery) 801 South Illinois Avenue Phone: +1 423 483 2121 Oak Ridge Fax: +1 423 483 0396 TN 37831-2011 E-mail: service.sigrecovery@perkinelmer.com USA
or
PerkinElmer Instruments (Signal Recovery) Sorbus House Phone: +44 (0)118 977 3003 Mulberry Business Park Fax: +44 (0)118 977 3493 WOKINGHAM RG41 2GY E-mail: service.sigrecovery@perkinelmer.com United Kingdom
6. Symptoms (in detail, including control settings)
7. Your purchase order number for repair charges (does not apply to repairs in warranty)
8. Shipping instructions (if you wish to authorize shipment by any method other than normal surface transportation)