PerkinElmer 5182 Instruction Manual

Page 1
Model 5182
Low-Noise Current
Preamplifier
Instruction Manual
222514-A-MNL-C
Copyright © 2000 PERKINELMER CORPORATION
Page 2
FCC Notice
This equipment generates, uses, and can radiate radio-frequency energy and, if not installed and used in accordance with this manual, may cause interference to radio communications. As temporarily permitted by regulation, operation of this equipment in a residential area is likely to cause interference, in which case the user at his own facility will be required to take whatever measures may be required to correct the interference.
Company Names
PerkinElmer Instruments is part of Perkin Elmer, Inc., and includes the businesses formerly trading as EG&G Princeton Applied Research, EG&G Instruments and EG&G Signal Recovery.
Declaration of Conformity
This product conforms to EC Directives 89/336/EEC Electromagnetic Compatibility Directive, amended by 92/31/EEC and 93/68/EEC, and Low Voltage Directive 73/23/EEC amended by 93/68/EEC.
This product has been designed in conformance with the following IEC/EN standards:
EMC: BS EN55011 (1991) Group 1, Class A (CSPIR 11:1990)
BS EN50082-1 (1992):
IEC 801-2:1991 IEC 801-3:1994 IEC 801-4:1988
Safety: BS EN61010-1: 1993 (IEC 1010-1:1990+A1:1992)
Trademarks
is a trademark of PerkinElmer, Inc.
Page 3

Table of Contents

Table of Contents
Chapter One, Description
1.1 Description ................................................................................................................................................. 1-1
1.2 Options....................................................................................................................................................... 1-2
Chapter Two, Initial Checks
2.1 Introduction ................................................................................................................................................ 2-1
2.2 Equipment Needed ...................................................................................................................................... 2-1
2.3 Procedure.................................................................................................................................................... 2-1
Chapter Three, Operating Instructions
3.1 Introduction ................................................................................................................................................ 3-1
3.1.01 Power Switch.....................................................................................................................................3-1
3.1.02 Inputs................................................................................................................................................. 3-1
3.1.03 Sensitivity Switch............................................................................................................................... 3-1
3.1.04 Outputs.............................................................................................................................................. 3-1
3.2 Operating the Model 5182........................................................................................................................... 3-2
3.2.01 Introduction........................................................................................................................................ 3-2
3.2.02 Signal Voltage and Gain..................................................................................................................... 3-2
3.3 Battery Operation........................................................................................................................................ 3-2
3.4 External Supply Operation..........................................................................................................................3-3
Appendix A, Specifications
Appendix B, External Power Supply
Warranty
................................................................................................................................. End of Manual
i
Page 4
TABLE OF CONTENTS
ii
Page 5
Description
1.1 Description
Chapter
The 5182 is a current input preamplifier designed for use whenever the signal source is a current source - for example, with electron multipliers, ion collectors, photo multipliers and photo diodes. The sensitivity is switch selectable with five settings enabling the amplifier, on its most sensitive range, to detect fractions of a picoamp without noise degradation, while on its least sensitive range it can sink currents up to 9 mA without overload.
-8
The instrument has a bandwidth of 25 kHz on its most sensitive range (10 1 MHz on its least sensitive range (10 noise ranging from 45 fA /√Hz on the 10 range. An additional low noise range is provided - denoted 10 features a reduced bandwidth of 10 kHz but with even lower noise.
-5
A/V), while still maintaining low input current
-8
A/V range to 10 pA /√Hz on the 10-5A/V
-8
A/V Low Noise, which
A/V) and
1
Front Panel
Rear Panel
Figure 1-1, Model 5182 Low-Noise Current Preamplifier
1-1
Page 6
Chapter 1, DESCRIPTION
1.2 Options
The model 5182 is supplied complete with four 9 V alkaline batteries. The following optional extras are available.
Model PS0108
Model PS0109
Model C0218
Model K0304
Remote Line Power Supply for the model 5182. See Appendix B for details of this unit.
Pack of 20 alkaline batteries for the model 5182
Power cable. This cable allows the user to power the model 5182 from any one of the following PerkinElmer Instruments (formerly EG&G) Lock-in Amplifiers: Model 5102, 5104, 5109, 5110, 5205, 5206, 5207, 5208, 5209, 5210, 5302, 7220, 7225, 7260, 7265 or 7280.
Rack mounting kit. This allows 1 or 2 model 5182 preamplifiers to be mounted in a standard 19" rack.
1-2
Page 7
Initial Checks
2.1 Introduction
The following procedure is provided to facilitate initial performance checking of the Model 5182. In general, the procedure should be performed after inspecting the instrument for shipping damage (any noted to be reported to the carrier and to PerkinElmer Signal Recovery), but before using it experimentally. Should any difficulty be encountered in carrying out these checks, contact the factory or one of its representatives.
2.2 Equipment Needed
1) General purpose laboratory oscilloscope.
2) Signal generator capable of providing a 10 V, 1 V, 100 mV and 10 mV pk-pk sine wave at 1 kHz, with a variable DC output offset facility.
3) A 1 MΩ resistor mounted inside a screened enclosure and connected between the inner connecting pins of male and female BNC connectors. The outer shield, or ground, terminals of these connectors should be connected together.
Chapter
2
2.3 Procedure
1) Set the front-panel press the adjacent indicating that the internal batteries are installed and functioning properly. If the
POWER TEST
5182’s battery requirements.
Those users who have purchased the PerkinElmer Instruments external
Note:
power supply option can use the external supply to perform the initial checks, in which case step 1 is replaced by the following operations.
a) Make sure that the voltage selector switch on the external power supply is in
the position indicating the line voltage to be used (110 V AC or 240 V AC).
b) Plug the line cord into the external power supply and the external power
supply into the instrument power socket.
c) Set the
adjacent indicating that the external supply is functioning properly.
POWER
POWER
POWER TEST
indicator doesn’t light, refer to section 3.3 for a discussion of the
switch to the up (
POWER TEST
switch to the down (
button. The associated indicator should light,
EXTERNAL
button. The associated indicator should light,
BATTERY
) position. Then press the
) position. Then
2) Select the AC sensitivity to
3) Connect the oscilloscope to the AC
10
-5
A/V
.
OUTPUT
BNC connector.
2-1
Page 8
Chapter 2, INITIAL CHECKS
4) Set the signal generator to 1 kHz, 10 V pk-pk, and connect it via the 1 MΩ BNC
5) Monitor the output; the output level should be 1 V pk-pk.
6) Set the signal generator to 1 kHz, 1 V pk-pk. Then set the 5182 AC sensitivity to
module to the
IINPUT
BNC connector. Use the oscilloscope to monitor the signal generator amplitude settings so as to obtain consistency between input settings and output readings.
-6
and monitor the output; the output level should be 1 V pk-pk.
A/V
10
7) Set the signal generator to 1 kHz, 100 mV pk-pk. Then set the 5182 AC sensitivity
-7
to
10
and monitor the output; the output level should be 1 V pk-pk.
A/V
8) Apply a 5 V DC offset to signal generator output signal and monitor the
DC OUTPUT
BNC connector with the oscilloscope; the output level should be
5 V DC.
9) Remove the DC offset from the signal generator output and reconnect the oscilloscope to the
AC OUTPUT
BNC connector.
10) Set the signal generator to 1 kHz, 10 mV pk-pk. Then set the 5182 AC sensitivity
-8
to
10
and monitor the output; the output level should be 1 V pk-pk.
A/V
11) Set the signal generator to 1 kHz, 10 mV pk-pk. Then set the 5182 AC sensitivity
-8
to
A/V LOW NOISE
10
and monitor the output; the output level should be 1 V
pk-pk.
12) Return the power switch to the center (
OFF
) position.
This completes the initial checks. If the instrument performed as indicated, one can be reasonably sure that it has arrived in good working order and is functioning properly.
2-2
Page 9
Operating Instructions
3.1 Introduction
The Model 5182 has been designed to be extremely easy to setup and use. It can be powered from an external source or from internal alkaline cells as selected from a front-panel switch. In operation, the signal to be measured is to the input BNC connector. The output is cabled directly to the input of the lock-in amplifier or other instrument with which the 5182 is to be used.
3.1.01 Power Switch
Chapter
3
The three-position power switch allows selected. To operate the Model 5182 from an external power source, set the switch to the up position. To operate the 5182 from internal batteries, set the switch to the down position. In the center position, the 5182 is unpowered. The status of the selected power source can be ascertained at a glance by pressing the the adjacent LED will light if the power source voltage, whether the external supply or internal batteries, is above the minimum required value for reliable operation.
EXTERNAL, OFF
, or
BATTERY
POWER TEST
3.1.02 Inputs
The 5182 sinking considerably more DC current than the AC current which it is measuring. The current source should be connected to the input by means of a low noise coaxial cable and BNC connector. Since cable capacitance can cause frequency response peaking, the cable length should be kept as short as possible, particularly on the more sensitive ranges.
INPUT
connector is at virtual ground. It is DC coupled and is capable of
3.1.03 Sensitivity Switch
The five position sensitivity switch provides a trade-off facility between sensitivity, noise, bandwidth and DC overload capability. AC sensitivity is switchable in decade steps from 10 corresponding DC sensitivity ranges from 10
-5
A/V to 10-8A/V with an additional low noise 10-8A/V setting. The
-3
A/V to 10-8A/V.
to be
button;
3.1.04 Outputs
Two BNC output connectors are provided; one representing the AC signal current and one representing the DC signal current. The AC Output can deliver >10 V pk-pk into loads >100 kΩ. The 450 Ω output impedance provides a convenient 10:1 output attenuator if the amplifier is loaded with 50 Ω, which can be useful if the 5182 is required to drive coaxial cable lengths of greater than one meter at signal frequencies approaching 1 MHz. The DC output provides a useful bias current monitor in several applications. It can output > ±9 V from an output resistance of 10 kΩ into loads >1 M
3-1
Page 10
Chapter 3, OPERATING INSTRUCTIONS
3.2 Operating the Model 5182
3.2.01 Introduction
The instrument is powered as required either from the self-contained batteries, from the preamplifier power-socket of a compatible lock-in amplifier (Models 7265, 7220, 5110, 5210, etc), or from an external power supply, and the signal to be amplified is applied to the input connector; the amplified signal is available at the AC BNC connector through a resistance of 450 Ω.
Note: Before operating from the external power supply (Appendix A), make sure the voltage selector switch on the external power supply is in the position indicating the line voltage to be used, and be sure the proper size line fuse is installed (200 mA for 110 V operation or 100 mA for 240 V operation). Operating from too high a line voltage will blow the line fuse and possibly damage the power transformer and circuit components.
3.2.02 Signal Voltage and Gain
The gain selector accurately sets the gain to the indicated level. The maximum output that the amplifier can provide is 10 V pk-pk into a high impedance load. For maximum input currents, refer to the specifications.
OUTPUT
3.3 Battery Operation
In battery powered operation, the 5182 requires four 9 V alkaline cells (type NEDA 1604, PP3 or equivalent). Each battery fits into its own removable tray accessible from the rear panel. When they are exhausted, all four batteries should be replaced together. Care should be taken when exchanging the batteries to observe the correct polarity which is marked on the inside of the battery trays and the top of the batteries. The batteries will provide up to fifteen hours operation at 1 V rms. output level.
Nickel-cadmium rechargeable batteries can be used if preferred, but there will be a penalty of a reduction in the operating time obtained. In addition it will be necessary to recharge these batteries in an external charger.
3-2
Figure 3-1, Required Battery Type
Page 11
3.4 External Supply Operation
External power can be provided by a stand-alone external power supply (see Appendix B) providing ±18 V DC, or the Preamplifier Power Output of most PerkinElmer Instruments lock-in amplifiers which provide ±15 V DC. Connection is by a DIN 5-way socket on the rear panel of the 5182.
Figure 3-2, External Power Input Connector Pinout
Chapter 3, TECHNICAL DESCRIPTION
3-3
Page 12
Chapter 3, OPERATING INSTRUCTIONS
3-4
Page 13
Specifications
General
Inputs
Appendix
DC coupled current to voltage amplifier with adjustable sensitivity and a maximum frequency response extending from DC to 1 MHz. Single-ended virtual ground input and single-ended DC and AC coupled outputs via BNC connectors.
Battery powered from internal alkaline batteries or external DC power supplies.
Mode Single-ended virtual ground Coupling DC Sensitivity Switch selectable (5 settings)
AC Output 10 DC Output 10 Accuracy ±2% Stability ±300 ppm/° C
Impedance see Figure A-1
-5
, 10-6, 10-7, 10-8, 10-8 low noise A/V
-3
, 10-4, 10-6, 10-7, 10-8 A/V
A
Figure A-1, Input Impedance vs. Frequency
Max input w/o damage ±15 V DC or 10 V rms. AC @ 50 Hz Noise see Table A-1
Sensitivity (A/V) Max DC Input Current at
-5
10
-6
10
-7
10
-8
10 10-8, low noise 90 nA
Table A-1, Max DC Input and Noise vs. Gain
1 kHz 9 mA 900 µA 9 µA 900 nA
Noise Current at 1 kHz
10 pA/√Hz 5 pA/√Hz 135 fA/√Hz 45 fA/√Hz 15 fA/√Hz
A-1
Page 14
Appendix A, SPECIFICATIONS
Frequency Response lower limit 0.5 Hz, (AC Output) upper limit
Max DC current at input see Table A-1
depends on sensitivity setting, see Figure A-2
Figure A-2, Frequency Response
Outputs
Power
AC Output
Impedance 450 Max voltage swing > 10 V pk-pk Slew rate > 22 V/µs
DC Output
Impedance 10 k Max voltage swing > ±9 V Polarity Current flowing into the input produces
Internal Four 9 V alkaline batteries provide
External
a) ±15 V or ±18 V DC @ 25 mA b) 110 V AC or 240 V AC via optional
Dimensions (excluding connectors) 8.25" wide × 11" deep × 3.5" high
Weight 5.3 lbs. (2.4 kg) excluding optional power supply
positive output voltage
approximately 15 hours of use
external model PS0108 power supply
(210 mm wide × 279 mm deep × 89 mm high)
A-2
Page 15
Power Supply Unit (Model PS0108)
Appendix
B
Figure B-1, Model PS0108 External Power Supply
Fuse Rating
115 V operation 200 mA 230 V operation 100 mA
Fuse Type 5 mm × 20 mm, 250 V, Slow Blow
B-1
Page 16
Appendix B, POWER SUPPLY UNIT (MODEL PS0108)
B-2
Page 17

WARRANTY

PerkinElmer Instruments Corporation warrants each instrument of its own manufacture to be free of defects in material and workmanship for a period of ONE year from the date of delivery to the original purchaser. Obligations under this Warranty shall be limited to replacing, repairing or giving credit for the purchase, at our option, of any instruments returned, shipment pr epaid, to our Service Department for that purpose, provided prior authoriz ation for such r eturn has been given by an authorized representative of PerkinElmer Instruments Corporation.
This Warranty shall not apply to any instrument, which our inspection shall disclose to our satisfaction, to have become defective or unusable due to abuse, mishandling, misuse, accident, alteration, negligence, improper installation, or other causes beyond our control. This Warranty shall not apply to any instrument or component not manufactured by PerkinElmer Instruments Corporation. When products manufactured by others are included in PerkinElmer Instruments Corporation equipment, the original manufacturers Warranty is extended to PerkinElmer Instruments customers.
PerkinElmer Instruments Corporation reserves the right to make changes in design at any time without incurring any obligation to install same on units previously purchased.
THERE ARE NO WARRANTIES WHICH EXTEND BEYOND THE DESCRIPTION ON THE FACE HEREOF. THIS WARRANTY IS IN LIEU OF, AND EXCLUDES ANY AND ALL OTHER WARRANTIES OR REPRESENTATIONS, EXPRESSED, IMPLIED OR STATUTORY, INCLUDING MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, AS WELL AS ANY AND ALL OTHER OBLIGATIONS OR LIABILITIES OF PERKINELMER INSTRUMENTS CORPORATION, INCLUDING, BUT NOT LIMITED TO, SPECIAL OR CONSEQUENTIAL DAMAGES. NO PERSON, FIRM OR CORPORATION IS AUTHORIZED TO ASSUME FOR PERKINELMER INSTRUMENTS CORPORATION ANY ADDITIONAL OBLIGATION OR LIABILITY NOT EXPRESSLY PROVIDED FOR HEREIN EXCEPT IN WRITING DULY EXECUTED BY AN OFFICER OF PERKINELMER INSTRUMENTS CORPORATION.
SHOULD YOUR EQUIPMENT REQUIRE SERVICE
A. Contact your local PerkinElmer Instruments (Signal Recovery) office, agent, representative or distributor to discuss the
problem. In many cases it may be possible to expedite servicing by localizing the problem to a particular unit or cable.
B. We will need the following information, a copy of which should also be attached to any equipment which is returned
for service.
1. Model number and serial number of instrument
2. Your name (instrument user)
3. Your address
4. Address to which the instrument should be returned
5. Your telephone number and extension
C. If you experience any difficulties in obtaining service please contact:
PerkinElmer Instruments (Signal Recovery) 801 South Illinois Avenue Phone: +1 423 483 2121 Oak Ridge Fax: +1 423 483 0396 TN 37831-2011 E-mail: service.sigrecovery@perkinelmer.com USA
or
PerkinElmer Instruments (Signal Recovery) Sorbus House Phone: +44 (0)118 977 3003 Mulberry Business Park Fax: +44 (0)118 977 3493 WOKINGHAM RG41 2GY E-mail: service.sigrecovery@perkinelmer.com United Kingdom
6. Symptoms (in detail, including control settings)
7. Your purchase order number for repair charges (does not apply to repairs in warranty)
8. Shipping instructions (if you wish to authorize shipment by any method other than normal surface transportation)