Technical Support ....................................................................Inside Back Cover
i
Sonometer012-03489E
Copyright, Warranty and Equipment Return
Please—Feel free to duplicate this manual
subject to the copyright restrictions below.
Copyright Notice
The PASCO scientific Model WA-9611 Sonometer
manual is copyrighted and all rights reserved. However,
permission is granted to non-profit educational institutions for reproduction of any part of this manual providing the reproductions are used only for their laboratories
and are not sold for profit. Reproduction under any other
circumstances, without the written consent of PASCO
scientific, is prohibited.
Limited Warranty
PASCO scientific warrants this product to be free from
defects in materials and workmanship for a period of one
year from the date of shipment to the customer. PASCO
will repair or replace, at its option, any part of the product
which is deemed to be defective in material or workmanship. This warranty does not cover damage to the product
caused by abuse or improper use. Determination of
whether a product failure is the result of a manufacturing
defect or improper use by the customer shall be made
solely by PASCO scientific. Responsibility for the return
of equipment for warranty repair belongs to the customer.
Equipment must be properly packed to prevent damage
and shipped postage or freight prepaid. (Damage caused
by improper packing of the equipment for return shipment will not be covered by the warranty.) Shipping
costs for returning the equipment, after repair, will be
paid by PASCO scientific.
Equipment Return
Should the product have to be returned to PASCO
scientific for any reason, notify PASCO scientific by
letter, phone, or fax BEFORE returning the product.
Upon notification, the return authorization and
shipping instructions will be promptly issued.
ä
NOTE: NO EQUIPMENT WILL BE
ACCEPTED FOR RETURN WITHOUT AN
AUTHORIZATION FROM PASCO.
When returning equipment for repair, the units
must be packed properly. Carriers will not accept
responsibility for damage caused by improper
packing. To be certain the unit will not be
damaged in shipment, observe the following rules:
➀ The packing carton must be strong enough for the
item shipped.
➁ Make certain there are at least two inches of
packing material between any point on the
apparatus and the inside walls of the carton.
➂ Make certain that the packing material cannot shift
in the box or become compressed, allowing the
instrument come in contact with the packing
carton.
Credits
This manual authored by: Clarence Bakken
This manual edited by: Dave Griffith
Teacher’s guide written by: Eric Ayars
The PASCO scientific Model WA-9611 Sonometer is an
enhanced version of the classic sonometer. You can perform the standard, qualitative sonometer experiments,
varying the tension, length, and linear density of the string
and observing the effects on the pitch of the plucked string.
Also, you can perform quantitative experiments, verifying
the equations for wave motion on a string by adding the
WA-9613 Driver/Detector Coils, a function generator capable of delivering 0.5 A of current, and an oscilloscope (or
a computer interface and power amplifier) where,
l = wavelength
L = length of string
Equipment
The WA-9611 Sonometer comes with the following
equipment (see Figure 1):
•Sonometer base with tensioning lever
•Two bridges
•10 wires (guitar strings), 2 each of the following diameters (linear densities):
To perform qualitative experiments, you will also need a
mass hanger and no more than 1.75 kg of mass to hang
from the tensioning lever.
Bridge
WA-9613
DRIVER
String
WA-9611
SONOMETER
WA-9613
DETECTOR
KEEP WEIGHTS AS NEAR TO FLOOR
AS POSSIBLE IN THE EVENT THE
SONOMETER WIRE SHOULD BREAK
n = number of antinodes
V = velocity of wave propagation
T = string tension
m = linear density of string
n = frequency of wave
The driver and detector coil can be placed anywhere along
the string. The driver coil drives string vibrations at any
frequency your function generator (or computer-compatible power amplifier) will produce. The detector coil allows
you to view the vibration of the string on your oscilloscope
or computer interface. With a dual trace oscilloscope or a
computer interface, you can examine phase differences between the driving frequency and the string vibrations.
Recommended Equipment
If you wish to accurately measure the frequency and
wavelength of the string vibrations, you will also need:
•WA-9613 Driver/Detector coils
•CI-6550 or CI-6565 Computer Interface and a
Power Amplifier (CI-6552)
OR
•Series 6500 Computer Interface, CI-6508 Input
Adapter Box, and a function generator capable of
producing 0.5 A
OR
•dual trace oscilloscope and a function generator
capable of producing 0.5 A
Optional Equipment (for use with function
generator)
•banana plug patch cords and BNC-to-banana
adapter (for connecting the function generator to
the BNC connector on an oscilloscope)
Tensioning
lever
CAUTION!
1.75 kg MAXIMUM
LOAD ON LEVER
Sonometer base
Figure 1 The Sonometer and Suggested Accessories
1
Sonometer012-03489E
Setup and Operation
To setup the sonometer (see Figure 2):
➀ Choose one of the ten strings and place the brass
string retainer into the slot on the tensioning lever.
➁ Loosen the string adjustment screw and place the
crimped lug that is attached to the other end of the
string over the screw head, as shown.
Crimped lug
Brass string retainer
➂ Tighten the string adjustment screw until the
tensioning lever hangs level.
➃ Place the bridges in any locations you wish, to deter-
mine the length of the string.
➄ Hang a mass (approximately 1 kg) from the tensioning
lever to produce the desired tension, then adjust the
string adjustment screw as needed so that the
tensioning lever is level. See Figure 3. (The lever must
be level to accurately determine the string tension
from the hanging mass.)
String tension is determined as shown in Figure 3. If
you hang a mass “M” from slot one of the lever, the
tension of the string is equal to Mg, where g is the
gravitational constant (9.8 m/s
2
). If you hang the mass
from slot two, the tension equals 2Mg; if you hang it
from slot three, the tension is 3 Mg, etc.
➅ You can now:
-Vary the tension of the string by hanging the mass
from different slots in the tensioning lever. (Always
adjust the string adjustment screw so the lever remains
level.)
String adjustment screw
Figure 2 Sonometer Setup
String tension (T)
Slot123 45
Figure 3 Setting the Tension
Tensioning lever
Bridges
5Mg
4Mg
3Mg
2Mg
1Mg
Hanging
Mass
(mass = M)
(weight = Mg)
-Vary the length of the string by adjusting the distance
between the bridges.
-Vary the linear density of the string by changing
strings.
-Pluck the string to observe how each of these variables effects the resonant frequency.
2
End plate
String
Adjustment
screw
Aluminum Cylinder
Figure 4 Reversing the End Plates
End plate
012-03489ESonometer
➤NOTE: At some lab stations, you may want the
tensioning lever to hang over the left end of the
table instead of the right (see Figure 4). In this case,
you can switch the end plates so that, when performing the experiment, the metric scale will still
be right side up. To switch the endplates:
➀ Loosen the string adjustment screw and remove
the string.
➁ Unscrew the two screws that hold each end plate
onto the sonometer and remove the end plates.
➂ Slide the aluminum cylinder out of the slot.
➃ Slide the cylinder into the slot on the other end
of the sonometer, then switch the end plates.
Using the Sonometer and the WA-9613
Driver/Detector Coils:
Sonometer and Driver/Detector Coils with a function
generator and oscilloscope:
➀ Connect the Driver and Detector Coils to the function
generator and oscilloscope as shown in the diagram.
Connect the driver coil directly to the output of the
PASCO PI-9587B Digital Function Generator. Connect the detector coil directly to channel two of an oscilloscope that has a BNC connector. You can use banana plug patch cords and a BNC-to-banana plug
adapter to connect the output of the function generator
to channel one of an oscilloscope that has a BNC connector. (If you are using a single trace oscilloscope,
connect only the detector coil to the oscilloscope.)
➁ Position the driver coil approximately 5 cm from one
of the bridges.
Depending on the wave pattern you are trying to pro-
duce, you might want to place the driver at some other
position. It will drive the string best if it is placed at an
antinode of the wave pattern. However, if you place it
near one of the bridges, it will work reasonably well
for most frequencies.
➂ Position the detector midway between the bridges ini-
tially, though for some patterns you may want to reposition it to best pick up the signal. As with the driver
coil, it works best when positioned near an antinode of
the wave pattern.
➃ Set the gain on channel-one of the oscilloscope to 5
mV/cm. Adjust the oscilloscope so it triggers on the
signal from the function generator.
➄ Set the function generator to produce a sine wave. Set
the frequency to a value between 100 and 200 Hz. Adjust the amplitude to about 5 V (approximately half of
maximum). Slowly vary the frequency of the function
generator output. When you reach a resonant frequency, you should see the motion of the string and
the sound produced by the vibrating string should be a
maximum. The wave pattern shown on the oscilloscope should become a clean sine wave. If you can’t
see or hear the string, raise the amplitude of the function generator output slightly and try again.
Driver coilDetector coil
WA-9613
DRIVER
FREQUEN
CY
HE
RTZ
PI-9587B
DIGITAL FUNCTION
GENERATOR AMPLIFIER
WAVEFO
RA
NGE
RM
EXTERN
AL
INPUTG
ND
AMPLITU
ADJU
ST
DE
M
MI
AX
N
Function generator
WA-9611
SONOMETER
OUTP
UT
TT
L
HI
Ω
G
ND
LO
Ω
WA-9613
DETECTOR
KEEP WEIGHTS AS NEAR TO FLOOR
AS POSSIBLE IN THE EVENT THE
SONOMETER WIRE SHOULD BREAK
200 Mhz OSCILLISCOPE
BK PRECISION
Channel 1 (trigger)Channel 2
Figure 5 Using the Driver and Detector Coils
CAUTION!
1.75 kg MAXIMUM
LOAD ON LEVER
INTENSITY
TRACE NOTATION
POS
AC
DC
CH 1
∞
400V
MODEL
MAX
2120
Oscilloscope
3
FOCUS
TRIG LEVEL
-
MANUAL AUTO
NORM
CH1
EXT
EXT
CH2
VERTICAL MODE
CH 1
VOLTZ/DIV
CAL
V
mV
VARVAR
PULL XSPULL XS
+
T XY
T XY
LINE
CH1
CH2
CH 2
VOLTZ/DIV
V
COUPLE SOURCE
AC
mV
CH1
X-POS
λ - Y
SLOPE
+
CH2
-
ALT
EXT
POSNORM
TIME/DI
V
VAR SWEEP
AC
CAL
DC
CAL
CH4
CAL EXT
CH 2
∞
200V
POWER
400V
MAX
MAX
Sonometer012-03489E
Table 1
ComputerInterfaceDevice to drive coilSoftwareFFT?
Apple IIAI-6501Power AmplifierPower Amplifier (Apple II)no
Apple IIAI-6501function generatorData Monitor (Apple II)no
DOS - PCCI-6500Power AmplifierPower Amplifier (MS-DOS)no
DOS - PCCI-6500function generatorData Monitor (MS-DOS)yes
MacintoshCI-6550Power AmplifierScience Workshop (Mac)yes
MacintoshCI-6550function generatorScience Workshop (Mac)yes
Windows - PC CI-6565Power AmplifierScience Workshop (Windows)yes
Windows - PC CI-6565function generatorScience Workshop (Windows)yes
Windows - PC CI-6500function generatorData Monitor (Windows)yes
Sonometer and Driver/Detector Coils with a PASCO
Computer Interface
There are several ways to use a PASCO Computer Interface with the sonometer. The method you use depends on
the kind of computer, the interface (e.g., CI-6500, CI6550, etc.), the device to control the coil, and whether you
wish to do frequency analysis (Fast Fourier Transform or
FFT) of the standing waves. See Table 1.
Using the Power Amplifier with a Series 6500
Computer Interface:
➀ Connect the Power Amplifier DIN plug to channel C
of the interface. Connect the Sonometer Driver Coil to
the output of the Power Amplifier.
➤ CAUTION: Do not turn on the power amplifier
until you have set the output amplitude from within
the program.
➁ Connect the BNC plug on the Sonometer Detector
Coil to the BNC jack on the CI-6508 Input Adapter
Box, and the DIN plug on the Adapter Box to channel
A of the interface. Turn the amplification select switch
on the CI-6508 to 100X. (See Figure 5.1.)
➂ Start the Power Amplifier program and set the output
to a 3-5 V sine wave; then turn on the power amplifier. Show channel A and channel C on the screen, so
you can see both the driving force and the resultant
motion of the wire.
➤ NOTE: The Power Amplifier program does not
have a frequency analysis feature (Fast Fourier
Transform or FFT).
Using the Power Amplifier with a CI-6550 or CI-6565
Computer Interface:
The Science Workshop program that comes with the CI-
6550 or CI-6565 interface allows you to do frequency
analysis (Fast Fourier Transform, or FFT) of the standing
waves. This can be used for an in-depth analysis of the
harmonics present in a standing wave, analysis of noise,
or observation of multiple simultaneous resonances.
➀ Connect the Power Amplifier DIN plug to channel C
of the interface. Connect the Sonometer Driver Coil to
the output of the Power Amplifier.
➤ CAUTION: Do not turn on the power amplifier
until you have set the output amplitude from within
the program.
CI-6502
POWER AMPLIFIER
FOR USE WITH PASCO SERIES 6500 INTERFACES
ON
CAUTION!
PASCO
SERIES
WHEN LIGHT IS ON
WAVEFORM IS DISTORTED.
6500
INTERFACE
DECREASE AMPLITUDE!
SYSTEM
Power Amplifier
KEEP WEIGHTS AS NEAR TO FLOOR
AS POSSIBLE IN THE EVENT THE
SONOMETER WIRE SHOULD BREAK
Interface
A ▲
ON
GAIN = 1,10,100
ISOLATED
WA-9613
DETECTOR
ANALOG CHANNELS
B ■ C ●
=
GAIN
1
ISOLATED
GAIN = 1
REF TO GND
CAUTION!
1.75 kg MAXIMUM
LOAD ON LEVER
ANALOG INPUT
PASCO
SERIES
6500
INTERFACE
SYSTEM
INPUT ADAPTOR
FOR USE WITH PASCO SERIES 6500 INTERFACES
(±10V MAX)
Model CI-6508
GAIN SELECT
X 100
X 10
X 1
NOTE: SWITCH
FUNTIONS ONLY WHEN
ADAPTOR IS
CONNECTED TO INPUT
MARKED ▲ ON THE
SIGNAL INTERFACE
SIGNAL OUTPUT
0 to ±10 V
1 A MAX
WA-9613
DRIVER
WA-9611
SONOMETER
CI-6510
SIGNAL INTERFACE
+
FOR USE WITH PASCO SERIES 6500 SENSORS
DIGITAL CHANNELS
1234
PASCO
SERIES
6500
INTERFACE
SYSTEM
Adapter Box
Figure 5.1 Using the Power Amplifier and Series 6500 Interface
4
Loading...
+ 16 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.