PASCO TD-8578 User Manual

Instruction Manual and Experiment Guide for the PASCO scientific Model TD-8578
THERMAL EXPANSION
APPARATUS
012-07114D
© 1999 PASCO scientific
Table of Contents
Section Page
Copyright, Warranty and Equipment Return ................................................. ii
INTRODUCTION:
Equipment.................................................................................................. 1
Additional Equipment Required ................................................................ 1
Notes on Temperature Measurement ......................................................... 1
Accepted Values for Coefficient of Thermal Expansion .......................... 2
Changing Tubes ......................................................................................... 2
Replacement Parts ..................................................................................... 2
EXPERIMENT:
Measuring the Coefficient of Linear Expansion for Copper, Brass, and Aluminum:
Introduction .........................................................................................3
Procedure .............................................................................................3
Data and Calculations .......................................................................... 4
Questions ............................................................................................. 5
Thermistor Conversion Table: ............................................................. 5
Technical Support ............................................................................................ 7
i
Thermal Expansion Apparatus 012-07114D
Copyright, Warranty and Equipment Return
Please—Feel free to duplicate this manual subject to the copyright restrictions below.
Copyright Notice
The PASCO scientific 012-07114 Thermal Expansion manual is copyrighted and all rights reserved. How­ever, permission is granted to non-profit educational institutions for reproduction of any part of the manual providing the reproductions are used only for their laboratories and are not sold for profit. Reproduction under any other circumstances, without the written consent of PASCO scientific, is prohibited.
Limited Warranty
PASCO scientific warrants the product to be free from defects in materials and workmanship for a period of one year from the date of shipment to the customer. PASCO will repair or replace at its option any part of the product which is deemed to be defective in mate­rial or workmanship. The warranty does not cover damage to the product caused by abuse or improper use. Determination of whether a product failure is the result of a manufacturing defect or improper use by the customer shall be made solely by PASCO scientific. Responsibility for the return of equipment for war­ranty repair belongs to the customer. Equipment must be properly packed to prevent damage and shipped postage or freight prepaid. (Damage caused by im­proper packing of the equipment for return shipment will not be covered by the warranty.) Shipping costs for returning the equipment after repair will be paid by PASCO scientific.
Equipment Return
Should the product have to be returned to PASCO scientific for any reason, notify PASCO scientific by letter, phone, or fax BEFORE returning the product. Upon notification, the return authorization and ship­ping instructions will be promptly issued.
NOTE: NO EQUIPMENT WILL BE
ACCEPTED FOR RETURN WITHOUT AN AUTHORIZATION FROM PASCO.
When returning equipment for repair, the units must be packed properly. Carriers will not accept responsi­bility for damage caused by improper packing. To be certain the unit will not be damaged in shipment, observe the following rules:
The packing carton must be strong enough for the
item shipped.
Make certain there are at least two inches of
packing material between any point on the apparatus and the inside walls of the carton.
Make certain that the packing material cannot shift
in the box or become compressed, allowing the instrument come in contact with the packing carton.
Address: PASCO scientific
10101 Foothills Blvd.
Roseville, CA 95747-7100
Phone: (916) 786-3800
FAX: (916) 786-3292
email: techsupp@pasco.com
web: www.pasco.com
ii
Introduction
Introduction
The PASCO Model TD-8578 Thermal Expansion Apparatus provides easy and accurate measurements of the coefficient of linear expansion for brass, copper, and aluminum.
To make the measurement, the brass, copper, or aluminum tube is placed on the expansion base. The length of the tube is measured at room temperature, then steam is passed through it. The expansion of the metal is measured with 0.01 mm resolution using the built-in dial-gauge. Temperatures are measured to within 0.2 °C using a thermistor attached to the center of the tube. If you wish to investigate the expansion of the metals at additional temperatures, hot or cold water can be passed through the tubes.
Complete step by step instructions and a data sheet for results are provided on the following pages.
Equipment
Your TD-8578 Thermal Expansion Apparatus in­cludes:
• A 40 cm long expansion base with a built-in dial gauge and thermistor.
➤➤
➤ NOTE: The dial gauge can be removed or
➤➤
repositioned by loosening the set screw on the dial gauge mounting block.
• Three metal tubes — brass, copper (99.5% Cu,
0.5% Te), and aluminum (98.9% Al, 0.7% Mg,
0.4% Si): 6.4 mm outside diameter.
• A foam insulator to avoid heat loss at the ther­mistor connection point.
• Thermoplastic elastometer tubing with 6.4 mm I.D.
Additional Equipment Required
In addition to the TD-8578 Thermal Expansion Apparatus, the following items are needed to perform the experiment:
A source of steam or hot water, such as the PASCO
Model TD-8556A Steam Generator.
A digital ohmmeter such as PASCO Model SE-
9589 to measure the thermistor resistance. Leads should have banana plug connectors, such as PASCO Model SE-9750 or SE-9751 Patch Cords.
A container to catch the water as it drains out of the
tube.
If additional data points are desired you will also
need: a source of hot or cold water.
Notes on Temperature Measurement
A thermistor's resistance varies reliably with tempera­ture. The resistance can be measured with an ohmme­ter, and converted to a temperature measurement using the attached conversion table tag and also on the back page of this manual. Although the relationship between temperature and resistance is not linear, a linear approximation can be accurately used to interpo­late between table data points with an accuracy of approximately ±0.2 °C.
The thermistor used to measure the tube temperature is embedded in the thermistor lug. Once thermal equilib­rium has been reached, the heat is highly uniform along the length of the tube. The foam insulator is used to inhibit heat loss through the thermistor lug so the lug temperature closely follows the tube tempera­ture. The insulator does not have any appreciable effect on the local temperature of the tube itself.
1
Thermal Expansion Apparatus 012-07114D
Accepted Values for Coefficient of Thermal Expansion
Material
Copper
Brass
Aluminum
( x10-6/ C)
a
8
17
19
23
Changing Tubes
Caution: When changing tubes, be careful not to
pull the wires off of the thermistor.
Compress the capture spring by pressing the two
ends together.
Replacement Parts
The following parts can be ordered from PASCO scientific.
Item
Mod. Thermistor (100k )
V
PASCO #
150-031
AL Tube
Cu Tube
Brass Tube
Foam Insulator
Dial Gauge
648-07139
648-07141
648-07140
716-041
620-050
Insert the curved lug of the thermistor between the
capture spring and the thermal tube.
Release spring.
thermistor
capture spring
thermal rod
2
Experiment: Measuring the Coefficient of Linear
Expansion for Copper, Brass, and Aluminum
Introduction
Most materials expand when heated through a temperature range that does not produce a change in phase. The added heat increases the average amplitude of vibration of the atoms in the material, which increases the average separation between the atoms.
Suppose an object of length L undergoes a temperature change of magnitude T. If T is reasonably small, the change in length, L, is generally proportional to L and T. Stated mathematically:
L = αL T;
where α is called the coefficient of linear expansion for the material.
Materials that are not isotropic, such as an asymmetric crystal for example, α can have a different value depending on the axis along which the expansion is measured. α can also vary somewhat with temperature. Therefore, the degree of expansion depends not only on the magnitude of the temperature change, but on the absolute temperature as well.
In this experiment, you will measure α for copper, aluminum, and brass. These metals are isotropic, so it is necessary to measure α along only one dimension. Also, within the limits of this experiment, α does not vary with temperature.
Procedure
Measure L, the length of the copper tube at room temperature. Measure from the outer edge
of the stainless steel ring on one end, to the outer edge of the stainless steel ring at the other end (see Figure 1). Record your results in Table 1.
Mount the copper tube in the expan-
sion base as shown in Figure 2. The stainless steel ring on the tube fits into the slot on the slotted mounting block and the bracket on the tube presses against the spring arm of the dial gauge.
Use the capture spring on
the expansion tube to attach the thermistor lug somewhere in the middle of the copper tube. The lug will be aligned with the axis of the tube, as shown in Figure 2, so there is maximum contact between the lug and the tube.


Figure 1 Measuring Tube Length
stainless steel ring
capture spring
slotted mounting block
Figure 2 Equipment Setup (Top View)
L
stainless steel ring
thermistor
contact indicator
3
Thermal Expansion Apparatus 012-07114D
Place the foam insulator over the thermistor lug.
➄ Plug the leads of your ohmmeter into the banana plug connectors labeled THERMISTOR in
the center of the expansion base. The banana plug connectors are located at one end of this unit in a stop bracket.
Measure and record R
value in the table.
Use tubing to attach your steam generator to the end of the copper tube. Attach it to the end
closest to the dial gauge.
Place a container under the other end of the tube to catch the draining water.
Turn the outer casing of the dial gauge to align the zero point on the scale with the long
indicator needle. As the tube expands, the indicator needle will move in a counterclockwise direction.
➉ Turn on the steam generator. As steam begins to flow, watch the dial gauge and the ohmme-
ter. When the thermistor resistance stabilizes, record the resistance (R record the expansion of the tube length (L) as indicated by the displacement of the indicator on the dial gauge. (Each increment on the dial gauge is equivalent to 0.01 mm of tube expan­sion.)
➤ Repeat the experiment for the brass and aluminum tubes.
, the resistance of the thermistor at room temperature. Record this
rm
) in Table 1. Also
hot
Data and Calculations
TABLE 1 Data and Calculations
CALCULATIONS
8
rm
( C)
T
hot
L(mm)
R
rm
( )
V
DATA
D
L(mm)
R
hot
( )
V
T
copper
brass
aluminum
Use the Thermistor Conversion Table on page 5 of this manual or on the tag attached to the
apparatus. Convert your thermistor resistance measurements, Rrm and R measurements, Trm and T
➤ Note: To achieve 0.2°C accuracy for temperature, you must interpolate between the
temperature values listed in the Thermistor Conversion Table.
. Record your results in the table.
hot
, into temperature
hot
8
( C)
8
( C)
T
D
4
Calculate ∆T = T
– Trm. Record the result in the table.
hot
Using the equation L = αL T, calculate α for copper, steel, and aluminum.
αCu = __________________
α
= __________________
Brass
αAl = __________________
Questions
Look up the accepted values for the linear expansion coefficient for copper, brass, and
aluminum. Compare these values with your experimental values. What is the percentage difference in each case? Is your experimental error consistently high or low?
On the basis of your answers in question 1, speculate on the possible sources of error in
your experiment. How might you improve the accuracy of the experiment?
THERMISTOR CONVERSION TABLE:
Temperature versus Resistance
Res. Temp. (Ω) (°C)
351,020 0 332,640 1 315,320 2 298,990 3 283,600 4 269,080 5 255,380 6 242,460 7 230,260 8 218,730 9 207,850 10 197,560 11 187,840 12 178,650 13 169,950 14 161,730 15 153,950 16 146,580 17 139,610 18 133,000 19 126,740 20 120,810 21 115,190 22 109,850 23 104,800 24 100,000 25
Res. Temp. (Ω) (°C)
95,447 26 91,126 27 87,022 28 83,124 29 79,422 30 75,903 31 72,560 32 69,380 33 66,356 34 63,480 35 60,743 36 58,138 37 55,658 38 53,297 39 51,048 40 48,905 41 46,863 42 44,917 43 43,062 44 41,292 45 39,605 46 37,995 47 36,458 48 34,991 49 33,591 50 32,253 51
Res. Temp. (Ω) (°C)
30,976 52 29,756 53 28,590 54 27,475 55 26,409 56 25,390 57 24,415 58 23,483 59 22,590 60 21,736 61 20,919 62 20,136 63 19,386 64 18,668 65 17,980 66 17,321 67 16,689 68 16,083 69 15,502 70 14,945 71 14,410 72 13,897 73 13,405 74 12,932 75 12,479 76 12,043 77
Res. Temp. (Ω) (°C)
11,625 78 11,223 79 10,837 80 10,467 81 10,110 82
9,767.2 83 9,437.7 84 9,120.8 85 8,816.0 86 8,522.7 87 8,240.6 88 7,969.1 89 7,707.7 90 7,456.2 91 7,214.0 92 6,980.6 93 6,755.9 94 6,539.4 95 6,330.8 96 6,129.8 97 5,936.1 98 5,749.3 99 5,569.3 100
5
Thermal Expansion Apparatus 012-07114D
Notes
6
Technical Support
Feedback
If you have any comments about the product or manual, please let us know. If you have any sugges­tions on alternate experiments or find a problem in the manual, please tell us. PASCO appreciates any cus­tomer feedback. Your input helps us evaluate and improve our product.
To Reach PASCO
For technical support, call us at 1-800-772-8700 (toll-free within the U.S.) or (916) 786-3800.
fax: (916) 786-3292
e-mail: techsupp@pasco.com
web: www.pasco.com
Contacting Technical Support
Before you call the PASCO Technical Support staff, it would be helpful to prepare the following information:
If your problem is with the PASCO apparatus, note:
- Title and model number (usually listed on the
label);
- Approximate age of apparatus;
- A detailed description of the problem/sequence of
events (in case you can’t call PASCO right away, you won’t lose valuable data);
- If possible, have the apparatus within reach when
calling to facilitate description of individual parts.
If your problem relates to the instruction manual,
note:
- Part number and revision (listed by month and
year on the front cover);
- Have the manual at hand to discuss your
questions.
7
Loading...