PASCO Spark Timer ME-9243A Instruction Sheet

Instruction Sheet for the PASCO Model ME-9243A
012-05131B
7/93
$1.00
Spark Timer
TIME
INTERVAL FREQ
(Hz)
(sec)
100
0.01 50
0.02
0.04
Frequency
selector knob
0.1
0.2
0.4
RATE
SELECTOR
Power ON LED
(flashes at selected
frequency rate)
USE THIS RATE
FOR SETUP!
POWER ON – RATE INDICATOR
25 10
PUSH
TO SPARK
5
2.5
WARNING
VOLTAGE!
Introduction
The PASCO Model ME-9243A Spark Timer provides accurate, phase-locked sparking at 7 different frequencies. The spark track is clear and simple, without any double sparking. The high voltage (HV) cable and the black ground cable are each 1 meter long. Also included with the ME-9243A is a 6' long power cord and this instruction sheet.
Safety Summary
Power source:
This apparatus is intended to operate from a power source of 90-130 VAC (rms) OR 180-260 VAC (rms), 43-62 Hz. A protective ground connection by way of the grounding connector in the power cord is essential for safe operation.
Grounding the Spark Timer:
To avoid electrical shock, plug the power cord into a properly wired receptacle.
Danger arising from loss of ground:
Upon loss of the protective ground connection, all accessible conductive parts (including knobs and controls that may appear to be insulated) can render an electric shock. Use a power cord that is in good condition.
SPARK TIMER
Ground cable
(1 m)
HIGH
High Voltage (HV)
cable (1 m)
Use the proper fuse:
To avoid fire hazard, use only a 0.5 A fuse, type 3AG, rated at 250 VAC.
Do Not Operate in Explosive Atmospheres:
To avoid explosion, do not operate this product in an explosive atmosphere.
Connect Case Ground to Apparatus Frame Before Energizing:
Attach the alligator clip end of the black ground cable to the apparatus being used.
Operating Instructions
Safety:
Refer to the Safety Summary for power source, grounding, and other important safety matters pertaining to the use of this equipment. Before connecting the instrument to a power source, read the following instructions about line voltages, power cords, and fuses; then verify that the correct power-input fuse is installed.
© 1993 PASCO scientific
012-05131B
CAUTION
RISK OF ELECTRIC SHOCK
DO NOT OPEN
TO AVOID ELECTRIC SHOCK THE POWER CORD PROTECTIVE GROUNDING CONDUCTOR MUST BE CONNECTED TO GROUND.
DO NOT REMOVE COVER. REFER SERVICING TO QUALIFIED PERSONNEL.
DISCONNECT INPUT POWER BEFORE REPLACING FUSE. FOR CONTINUED FIRE PROTECTION REPLACE ONLY
WITH 0.5 A, 250 V, TYPE T FUSE, (PER IEC 127-2/III).
LINE VOLTAGE RANGE: 90 - 130 VAC OR 180 - 260 VAC
POWER: 54 W MAX FREQ.: 47 - 63 HZ
FUSE
ON
OFF
Power Cord:
The Model ME-9243A is delivered with a three-wire power cord. At the end of the cord is a three-contact plug for connection to the power source and to protective ground. For electrical shock protection, insert this plug only into a power source outlet that has a securely grounded protective-ground contact.
Replacing the Line Fuse:
This instrument fuse holder is located in the power cord receptacle. Verify that the proper fuse is installed and/or replace a defective fuse by performing the following steps.
Unplug the power cord from the power source and the
power input module.
Slide the window over to the left and eject the fuse for
inspection.
Note the fuse value and verify proper type and rating.
0.5 A, type 3AG, 250 VAC, fast blow.
Replace the fuse, slide the window back and reconnect
the power cord.
Spark Gap Distance
When used with the PASCO ME-9226 Air Track and Spark Timer System or the ME-9228 Spark Timer Kit, the recommended spark gap distance between the electrodes and the wires or paper mounting strip is 2 mm. Making the gap larger causes the spark to “wander” and leave multiple marks. A very large gap can cause a dangerous shock and may damage the Spark Timer.
Setup
Ground the Spark Timer being used with the black
ground cable. Make this connection secure. Then at­tach the HV cable to the apparatus.
Supply power to the Spark Timer as discussed previ-
ously. Press the power switch to the ON position. Set the rate selector to 2.5 Hz to test your setup.
Select the desired output frequency for the high voltage
pulses.
Depress the red spring-loaded pushbutton to energize
the high voltage pulses.
NOTE: Care must be taken to provide adequate ground return for the external High Voltage circuit to the instrument case ground. A shock hazard will exist if this is not done.
Timing of the output HV spark is derived from an internal oscillator, which is set to 100.00 Hz ± .02 Hz at time of manufacture. This will hold for the following conditions:
Voltage: 90-130 VAC OR 180-260 VAC; 50/60 Hz Temperature: 23°C ± 15°C
Other
The PASCO ME-9243A Spark Timer is included in the ME-9226 Air Track and Spark Timer System. This consists of the SF-9214 Air Track with Accessories, the SF-9216 Variable Output Air Supply, the ME-9228 Air Track Spark Timer Kit, and an Instruction Manual and Experiments Guide (part number 012-01783).
Limited Warranty
PASCO scientific warrants this product to be free from defects in materials and workmanship for a period of one year from the date of shipment to the customer. PASCO will repair or replace, at its option, any part of the product which is deemed to be defective in material or workman­ship. This warranty does not cover damage to the product caused by abuse or improper use. Determination of whether a product failure is the result of a manufacturing defect or improper use by the customer shall be made solely by PASCO scientific. Responsibility for the return of equipment for warranty repair belongs to the customer. Equipment must be properly packed to prevent damage and shipped postage or freight prepaid. (Damage caused by improper packing of the equipment for return shipment will not be covered by the warranty.) Shipping costs for returning the equipment, after repair, will be paid by PASCO scientific.
To Reach PASCO
For Technical Support call us at 1-800-772-8700 (toll­free within the U.S.) or (916) 786-3800.
NOTE: The Power ON LED in the front panel will light up and flash at the rate of the selected frequency.