Instruction Sheet
for the PASCO
Model SF-9288
Spectral Lamp Power Supply and Mount
➄➄
➄ Allow about five minutes warm up time.
Introduction:
Always unplug before
removing cover of lamp.
Loosen
thumbscrew
to adjust
lamp height.
The PASCO Model SF-9288 Spectral Lamp Power Supply and Mount is designed for use with Osram 9 pin DIN
base, 21 mm diameter spectral lamps. The Mercury lamp
(SF-9282) is available from PASCO scientific.
Mercury Spectral Lamp SF-9282
Sodium Spectral Lamp SF-9289
Cadmium Spectral Lamp SF-9290
Helium Spectral Lamp SF-9291
Neon Spectral Lamp SF-9292
Changing spectral lamps is easy, so a single power supply
can be used to power any one of the Osram lamps.
To use the Spectral Lamp Power Supply and Lamp:
➀➀
➀ Insert a spectral lamp as shown on the back of this
➀➀
page.
➁➁
➁ Plug the round, three-pronged plug from the lamp
➁➁
mount into the socket on the front of the power supply.
➂➂
➂ Plug the power supply power cord into a grounded
➂➂
107-128 VAC, 60 Hz or 206-242 VAC, 50 Hz outlet.
(See label on rear of unit).
➃➃
➃ Flip the ON/OFF Switch on the front of the power
➃➃
supply to ON.
➄➄
If, at any time, the lamp fails to come on, check the fuse
on the rear panel of the power supply. If the fuse is
blown, replace only with a similarly rated fuse:
Fuse: 2 Amp
➤➤
➤ NOTE: To insert or change the spectral lamp,
➤➤
follow the procedure shown on the back of this page.
CAUTION
RISK OF ELECTRIC SHOCK
DO NOT OPEN
CAUTION:
TO PREVENT THE RISK OF ELECTRIC SHOCK, DO
NOT REMOVE COVER ON UNIT. NO USER SERVICEABLE PARTS INSIDE. REFER SERVICING TO
QUALIFIED SERVICE PERSONNEL.
The lightning flash with arrowhead, within
an equilateral triangle, is intended to alert
the user of the presence of uninsulated
“dangerous voltage” within the product’s
enclosure that may be of sufficient magnitude to constitute a risk of electric shock to
persons.
The exclamation point within an equilateral triangle is intended to alert the user
of the presence of important operating
and maintenance (servicing) instructions in the literature accompanying the
appliance.
Hg Lamp Contains Mercury
Manage in accordance with
disposal laws.
See https://www.lamprecycle.org/ or 1-866-666-6850
012-03048D
10101 Foothills Blvd. Roseville, CA 95678-9011 USA
®
Phone (916) 786-3800 FAX (916) 786-8905 web: www.pasco.com
better
ways to
teach science
Spectral Lamp Power Supply and Mount 012-03048D
Replacing the Spectral Lamp
➤➤
➤ CAUTION HIGH VOLTAGE: Always
➤➤
unplug the lamp mount from the power supply
➀➀
➀ Unplug the Lamp Mount cord from
➀➀
the power supply, and allow several
minutes for the lamp to cool.
➁➁
➁ Unscrew and remove the top
➁➁
screws.
➂➂
➂ Lift off the top.
➂➂
➃➃
➃ Gently pull the spectral lamp up,
➃➃
out of the 9-pin socket.
➤➤
➤ CAUTION! WEAR GLOVES TO PROTECT
➤➤
YOUR HANDS FROM A HOT LAMP OR
POSSIBLE BREAKAGE.
➄➄
➄ Reverse the procedure to replace the lamp.
➄➄
➤➤
➤ CAUTION! WHEN USING THE MERCURY
➤➤
LAMP, SF-9282, DO NOT LOOK INTO LAMP.
UV RAYS MAY BE DAMAGING.
If the glass bulb is broken, immediately turn lamp
off and remove it to avoid possible injury. Replace
bulb prior to next use.
S1
117 VAC or
220 VAC
GROUND
COMMON
Schematic: Model SF-9288 Spectral Lamp Power
ON/OFF
F1
2 Amp
Supply and Mount
T2
T1
Limited Warranty
PASCO scientific warrants this product to be free from
defects in materials and workmanship for a period of one
year from the date of shipment to the customer. PASCO
will repair or replace, at its option, any part of the product
which is deemed to be defective in material or workmanship. This warranty does not cover damage to the product
caused by abuse or improper use. Determination of
whether a product failure is the result of a manufacturing
defect or improper use by the customer shall be made
solely by PASCO scientific. Responsibility for the return
of equipment for warranty repair belongs to the customer.
Equipment must be properly packed to prevent damage
and shipped postage or freight prepaid. (Damage caused
by improper packing of the equipment for return shipment will not be covered by the warranty.) Shipping
costs for returning the equipment, after repair, will be
paid by PASCO scientific.
Equipment Return
Should this product have to be returned to PASCO scientific, for whatever reason, notify PASCO scientific by
letter or phone BEFORE returning the product. Upon
notification, the return authorization and shipping instructions will be promptly issued.
➤➤
➤ NOTE:
➤➤
FOR RETURN WITHOUT AN AUTHORIZATION.
When returning equipment for repair, the units must be
packed properly. Carriers will not accept responsibility for
damage caused by improper packing. To be certain the unit
will not be damaged in shipment, observe the following rules:
➀➀
➀ The carton must be strong enough for the item
➀➀
shipped.
➁➁
➁ Make certain there is at least two inches of packing
➁➁
material between any point on the apparatus and the
inside walls of the carton.
➂➂
➂ Make certain that the packing material can not shift in
➂➂
the box, or become compressed, thus letting the instrument come in contact with the edge of the box.
NO EQUIPMENT WILL BE ACCEPTED
2
®