Paragon PC-620-2, PC-620-1 User Manual

Page 1
Operator’s Manual
PC-620-1 & PC-620-2
Tabletop
Ultrasonic Cleaners
Page 2
Warranty
Ultrasonic Cleaners, when used in accordance with manufacturer’s instructions and under normal use, are guaranteed for two years after date of shipment. Within the period guaranteed, we will repair or replace free of charge, at our sole discretion, all parts that are defective because of material or workmanship, not including costs for removing or installing parts.
Liability, whether based on warranty, negligence or other cause, arising out of and/or incidental to sale, use or operation of the transducer elements, or any part thereof, shall not in any case exceed the cost of repair or replacement of the defective equipment, and such repair or replacement shall be the exclusive remedy of the purchaser, and in no case will we be responsible for any and/or all consequential or incidental damages including without limitation, and/or all consequential damages arising out of commercial losses.
CAUTION
Do not place parts or containers directly on the bottom of the cleaning tank; use a tray, wire hook or other device to suspend items.
Do not allow the cleaning solution level to drop more than one inch below the top of the tank.
Do not ever use alcohol, gasoline or flammable solutions. Doing so could cause a fire or explosion. Use only water-based solutions.
Do not ever use mineral acids. These could damage the tank. Failure to comply with these cautions will void your warranty.
Page 3
Contents
Front Warranty Cover
1 Safety Precautions 3 Introduction
3 PC620 General Information 4 Accessories 4 Unpacking Your Cleaner 4 Installing Your Cleaner 5 Equipment Specifications 6 How Ultrasonic Cleaning Works
7 Operating Your Cleaner
7 Before You Start 7 Explanation of Controls 8 Getting Ready 8 Cleaning Items 9 Optimizing Your Cleaner 10 Application Hints 11 Cleaning Solutions
13 Troubleshooting
14 Performance 15 Service
Page 4
Page 5
Safety Precautions
Before using your Ultrasonic Cleaner, please read and thoroughly understand these safety precautions. Failure to follow them may result in serious personal injury or property damage.
To avoid electrical shock:
Do unplug from power source before filling or emptying the tank.
Do keep the switches and the area around the cleaner clean and dry -
- wipe up solution which spills over the tank brim. Water and high voltage can cause electrical shock.
Do not operate the cleaner without proper grounding.
Do not remove the grounding prong on the line cord plug.
Do not disassemble your cleaner -- high voltage inside the cleaner is
dangerous.
Do not immerse the cleaner in water.
To prevent personal and/or property damage:
Do use water-based solutions.
Do not ever use alcohol, gasoline or flammable solutions. Doing so
could cause a fire, explosion, or serious personal injury and will void your warranty. Use only water-based solutions.
Do not ever use mineral acids. These could damage the tank.
Do not touch the stainless steel tank or cleaning solution -- they may
be hot.
Do not allow fluid temperature to exceed 80ºC (175ºF).
Do not place your fingers or hands into the tank while the cleaner is
operating. Doing so may cause discomfort and possible skin irritation. Avoid contact with solutions and provide adequate ventilation.
Do not use solutions containing chlorine bleach.
1 Safety Precautions
Page 6
To prevent damage to the cleaner:
Do change your solution regularly.
Do not operate the cleaner dry.
Do not place parts or containers directly on the bottom of the cleaning
tank; use a tray, wire hook or other device to suspend items. Failure to comply may cause transducer damage and will void your warranty.
Do not allow the cleaning solution level to drop more than one inch
below the top of the tank with heat or ultrasonics on. Failure to comply may cause transducer and/or heater damage and will void your warranty.
2 Safety Precautions
Page 7
Introduction
PC620 General Information
This PC-620 ultrasonic cleaner is available in two models; the unheated PC-620-1 and the heated PC-620-2. Each model is constructed using durable industrial style 44 kHz transducers. These provide increased cleaning power and ensure uniform cleaning activity throughout the bath. Both models have drains and are supplied with tank drain kits. For customer convenience, both units are available in three voltages to meet requirements around the world. Be sure the unit you have purchased is correct for your area.
3 Introduction
Page 8
Accessories For Your Cleaner
A stainless steel cover and solid wall-perforated bottom parts basket are available for both models. Consult your distributor for price and delivery.
Unpacking Your Cleaner
Please check your cleaner and its carton carefully for any external or internal damage. If you find damage, contact your shipping carrier immediately, before contacting your distributor. Please retain your packaging for future use.
Installing Your Cleaner
Check the label on the back of the cleaner for correct input power requirements. Position your cleaner within easy reach of a standard grounded electrical outlet. Do not place the cleaner on a circuit which could become overloaded.
If your cleaner does not operate correctly, first refer to the troubleshooting section for possible causes, or contact an authorized service center listed at the back of this manual for additional information.
4 Introduction
Page 9
Equipment Specifications
Model Tank
PC620R-1
PC620R-2
PC-620E-1
PC620E-2
PC-620J-1
PC-620J-2
Volume
2.75 gal. (8.5 L)
2.75 gal. (8.5 L)
2.75 gal. (8.5 L)
Tank
Size
L: 19.5”
W: 5.8”
D: 6”
L: 19.5”
W: 5.8”
D: 6”
L: 19.5”
W: 5.8”
D: 6”
Overall
Size
L: 20.3”
W: 6.3”
H: 10.5”
L: 20.3”
W: 6.3”
H: 10.5”
L: 20.3”
W: 6.3”
H: 10.5”
Weight Heater
Power
23 lbs.
(10.5Kg)
24 lbs.
10.9 Kg)
24 lbs.
10.9 Kg)
None
180 Watts
None
180 Watts
None
135 Watts
Input
Power
115 V
1.4A
115 V
3.1A
230 V
.7A
230 V
1.6A
100 V
1.6A
100 V
3.1A
NOTE:
All models have a frequency of 44 kHz.
Cleaners have NRTL/CSA approval and comply with FCC regulations.
All 230V units meet CE standards.
Units may cause GFI outlets to trip.
All units have a ground leakage current less than .50ma.
5 Introduction
Page 10
How Ultrasonic Cleaning Works
Ultrasonic sound is sound transmitted at frequencies generally beyond the range of human hearing. In your ultrasonic cleaner, ultrasonic sound (sonics) is used for cleaning materials and parts. This is how it works:
As the sound waves from the transducer radiate through the solution in the tank, they cause alternating high and low pressure areas in the solution.
During the low pressure stage, millions of microscopic bubbles form and grow. This process is called CAVITATION.
During the high pressure stage, the bubbles collapse or “implode” releasing enormous amounts of energy. These implosions act like scrub brushes, eroding soils,
away. They work in all directions attacking every surface and invading all recesses and openings .
6 Introduction
Page 11
Operating Your Cleaner
If this is the first time you are using the cleaner, please read this whole section before proceeding.
Before You Start Cleaning
CAUTION
Do not place parts or containers directly on the bottom of the cleaning tanks; use a basket, wire hook or other device to suspend items.
Do not allow the cleaning solution level to drop more than one inch below the top of the tank with the cleaner on.
Do not ever use alcohol, gasoline or flammable solutions. Doing so could cause a fire or explosion. Use only water-based solutions.
Do not ever use mineral acids. These could damage the tank. Failure to comply with these cautions will void your warranty.
Operating Your PC-620-1 or PC-620-2 Cleaner
Explanation of Controls
Control Function
ULTRASONIC POWER SWITCH
HEATER SWITCH
Activates and de-activates ultrasonic cavitation in the tank
Activates the external heater on the cleaning tank. The heater power has been selected to provide the optimum temperature for most cleaning applications.
7 Operating Your Cleaner
Page 12
Getting Ready
Step Action
1 Select your cleaning chemistry (check with your chemical
supplier for solution effects on metals).
2 Allowing for the volume of the parts you will be cleaning and
cleaning chemistry, fill the tank with warm tap water to the
operating level (one inch from the top). 3 Add cleaning chemistry to the tank water. 4 Plug the cleaner into a grounded outlet. 5 For maximum efficiency, refer to page 9, “Optimizing Your
Cleaner” before proceeding.
NOTE:
If this is the first time you are running the cleaner, or if you have changed cleaning solution, you must degas the solution. This is done by setting the cleaner up for operation and allowing the ultrasonics to drive the warm solution for 5-10 minutes This will force out excess gas and assure optimum cleaning. You will likely notice a change in the sound of the unit as is degasses.
Cleaning Items
Step Action
1 Place the items to be cleaned into a basket or perforated tray. 2 If using a solid container, add cleaning solution to the container
to cover the items. 3 Slowly lower the tray into the tank. Do not allow items to
contact the tank bottom. Do not stir the solution. 4 When items are clean, slowly remove them from the cleaner. 5 Rinse the clean items with fresh water and dry them, if
necessary.
8 Operating Your Cleaner
Page 13
Optimizing Your Cleaner Tanks
Cleaning - check the tank for contamination whenever you change
solution. If necessary, remove contaminants with a nonabrasive cloth and water.
Emptying - always unplug the cleaner before emptying the tank. Dispose of the contents in an environmentally responsible way.
Filling - always unplug the line cord before filling the tank. Fill the cleaner to the operating level (one inch from the top with beaker/tray in place), using warm tap water.
Low solution level - will cause the cleaner to fail. When you remove heavy or bulky loads from the cleaner, the solution level may drop below the operating level. If so, be sure to replace lost solution and degas, if necessary, depending on the amount used.
Overload - do not rest any items on the tank bottom. Weight on the tank bottom dampens sound energy and will cause damage to the transducer. Instead, use a basket or other method to support all items. Allow at least one inch between the tank bottom and the parts or receptacle for adequate cavitation.
Covers - allow the cleaner to heat up faster, to a higher temperature, and avoid excessive liquid evaporation. However, leaving the cover on with heat and ultrasonics can cause the solution to boil, diminishing effectiveness.
Temperature Heater - the heater may cause some discoloration of the tank wall. This
is normal and will not affect the performance of the unit. Solution - the fastest method to get your cleaner to the best operating
temperature is to fill it with warm solution, turn on both the heat and ultrasonics, and use a cover.
9 Operating Your Cleaner
Page 14
Application Hints First time cleaning - first experiment with one piece, then proceed with
the remainder. Solution level - Be sure to maintain solution level within one inch of the
tank top. Surface activity will vary with liquid level. Load size - It is faster and more efficient to run several small loads
rather than a few big loads. Placing items - Never allow items to sit on the bottom of the tank.
Always place them in a basket or suspend in the solution. Rinsing items - After cleaning, use clean water to rinse away chemicals
adhering to items. Lubricating items - When necessary, re-lubricate items immediately
after cleaning. Drying items - Air drying at room temperature works for some items.
Place parts requiring faster drying time under hot air blowers or in ovens.
Please call your local distributor if you have application questions.
10 Operating Your Cleaner
Page 15
Cleaning Solutions
CAUTION
Do not use alcohol, gasoline, bleach, mineral acids, solutions with a flash point, semi-aqueous or combustible liquids in ultrasonic tanks, or you will void the warranty. Only use non-flammable and water-based solutions.
Solution Types
Water-based solutions are either slightly acidic or alkaline. They include detergents, soaps and industrial cleaners designed to remove specific soils.
Acidic water-based solutions: remove rust, tarnish or scale. They range from mild solutions that remove tarnish, to concentrated, inhibited acidic solutions that remove investment plaster, milk-stone, zinc oxide and rust from steel and cast iron as well as smut and heat-treat scale from hardened steel.
Alkaline water-based solutions: include carbonates, silicates and caustics. These cause emulsifying action, which keeps soil from redepositing on the cleaned surface, and improves cleaning action in hard water.
Alkaline strength Removes:
Mild Light oils and greases, cutting oils and coolant
compounds.
Mild to strong Heavy grease and oil, waxes, vegetable oils, inks,
wax or fat-base buffing and polishing compounds, milk residues and carbohydrates.
Heavy-duty Mill scale, heat-treat scale, corrosion or oxides.
11 Operating Your Cleaner
Page 16
Change the cleaning solution periodically. Cleaning solutions can become contaminated with soil particles which coat the tank bottom. This coating dampens the ultrasonic action and reduces cleaning efficiency. Certain solutions will cavitate better than others . Contact your local distributor for further information.
Heat and cavitation: increase the chemical activity of cleaning solutions. Some materials may be damaged by this stronger chemical action. When in doubt, test run samples of items to be cleaned.
Chemistry Concentrations
Chemistry concentrations may vary. The amount you use depends on the detergent and the type of soil to be removed. Follow instructions on the chemistry container and refer to the table below for the effects of chemistry on metals.
Chemicals Harmful to Your Tank
The following chemicals will harm your ultrasonic tank and the action of ultrasonics and higher operating temperatures will increase their chemical activity. Do not use these or similar chemicals directly or in dilution in your ultrasonic tank or you will void your warranty.
Acetophenone Chloracetic Acid Hydrocyanic Acid Aluminum Chloride Chloric Acid Hydrofluoric Acid Aluminum Fluoride Chlorine, Anhydrous Hydrofluosilicic Acid Aluminum Sulphate Chromic Acid Iodoform Ammonium Bifluoride Copper Chloride Mercuric Chloride Ammonium Chloride Copper Fluoborate Muriatic Acid Ammonium Hydroxide Ethyl Chloride Phosphoric (crude) Amyl Chloride Ferric Chloride Sodium Hypochlorite Antimony Trichloride Ferrous Chloride Potassium Chloride Aqua Regia Ferris Sulfate Stannic Chloride Bromine Fluoboric Acid Stannous Chloride Calcium Bisulfate Fluorine Sulfur chloride Calcium Bisulfite Hydrobromic Acid Sulfuric Acid Calcium Hypochloride Hydrochloric Acid Zinc Chloride
12 Operating Your Cleaner
Page 17
Troubleshooting
If your cleaner does not operate satisfactorily, please check the tables below for possible causes before calling your authorized service center.
WARNING
High voltage inside - dangerous shock hazard.
DO NOT attempt to disassemble or repair the cleaner.
Problem Cause What to do
Cleaner will not start.
Cleaner operates but does not heat solution Decreased ultrasonic activity.
Cleaner not plugged in properly.
Circuit board fuse blown Heater malfunctions. Call nearest authorized
Solution is not degassed.
Solution is spent.
Plug into functioning electrical outlet.
. Call nearest authorized
service center. service center.
Make sure that tank was filled with warm tap water plus cleaning chemistry and has run 5-10 minutes.
Change solution.
Solution level is incorrect for load.
Tank bottom is covered with soil particles.
Using deionized water in the tank.
13 Troubleshooting
Adjust solution +/- 3/8 inch from current level.
Empty, then clean tank with warm water. Wipe with a nonabrasive cloth. Deionized water does not cavitate as actively as soapy tap water.
Page 18
Performance
Check your cleaner periodically to test the level of activity of the ultrasonic cavitation. Frequency of testing will depend on your use of the cleaner, however, we suggest running this test monthly.
Glass Slide Test
You will need the following equipment:
Frosted microscope glass slide (1” x 3”), such as ESCO #2951F, or equivalent;
No. 2 lead pencil; and
General purpose dish detergent.
Test procedure:
1. Prepare a fresh solut ion with general purpose dish detergent (concentration 1%) and warm tap water (120º - 140ºF).
2. Fill the cleaner to within one(1) inch of the tank top.
3. Turn the ultrasonics on for at least five minutes to allow for degassing.
4. Prepare the glass slide by first wetting the frosted portion with tap water.
5. With the No. 2 pencil, on the frosted portion make an “X” from corner to corner.
6. Immerse the frosted end of the slide into the solution. Hold the slide vertically and center it in the solution.
7. Turn ultrasonic switch to “On”.
14 Troubleshooting
Page 19
The ultrasonics will begin immediately to remove the lead from the slide. All lead should be removed within 10 seconds. If your cleaner passes this test, its ultrasonic cavitation is acceptable.
NOTE:
To ensure consistency from test to test, be sure to repeat test conditions - use the same solution concentration, liquid level, temperature, type of pencil, length of degassing, etc.
Service
With normal use, your Ultrasonic Cleaner should not require servicing. However, if it fails to operate satisfactorily, first try to diagnose the problem by following the suggestions in the Troubleshooting Guide. If you find that your cleaner needs repair, carefully pack and return it to your local Service Center. If under warranty, remember to include proof of purchase. Your cleaner will be returned by ground service unless you specify otherwise.
Authorized Service Centers
Name Address Tel/Fax Number
Alpha Omega Electronics Corp.
Crystal Electronic Inc. 140 Centre Street
Paragon Electronics 6861 SW 196th Avenue
Master Sonics Repair Center
2821 National Drive Garland, TX 75041
Aurora, Ontario Canada L4G 1K1
Pembroke Pines, FL 33332 77 Whiting Street
Plainville, CT 06062
15 Troubleshooting