Owon PDS, PDS5022S, PDS5022T, PDS6042S, PDS6062S User Manual

PDS Series
User Manual
PDS5022S
PDS5022T
PDS6042S
PDS6062S
www.owon.com.cn
www.owon.com.hk
Mar. 2012 edition Ver1.2
Copy Right in this Manual © Lilliput Company. All rights have been reserved.
The Lilliput’s products are under the protection of the patent rights in America and other countries,
including ones which have already obtained the patent rights and those which are applying for. The
information in this manual will replace all that in the materials published originally.
The information in this manual was correct at the time of printing. However, OWON will continue to
improve products and reserves the rights to changes specification at any time without notice.
OWON is the registered trademark of the Lilliput Company.
Xiamen Lilliput Technology Co.,Ltd.: the 5th floor, B Area, Chuangxin Mansion,
Software Park,
ZhenZhuWan,
Huandao RD, Xiamen, Fujian, China
Tel:+86-592-2575666 Fax:+86-592-2575669
Web : www.owon.com.cn
Mail: sales@owon.com.cn
Fujian Lilliput Optoelectronics Technology Co.,Ltd.: The mansion of optoelectronics hengsan road,
Lantian industrial zone ,Zhangzhou,Fujian,China
Tel:+86-596-2130430 Fax:+86-596-2109272
Web:www.owon.com.hk Mail:
Business Consulting: sales@owon.com.hk
Sale service: service@owon.com.hk
User Manual of PDS Series Oscilloscope
1
General Warranty
The Lilliput warrants that the product will be free from defects in materials and workmanship for a period of three years from the date of purchase of the product by the original purchaser from the Lilliput company. And the warranty period of accessories such as probe, battery, adapter is one year. This warranty only applies to the original purchaser and is not transferable to the third party. If the product proves defective during the warranty period, Lilliput either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Parts, modules and replacement products used by Lilliput for warranty work may be new or reconditioned to like new performance. All replaced parts, modules and products become the property of Lilliput.
In order to obtain service under this warranty, Customer must notify Lilliput of the defect before the expiration of the warranty period. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Lilliput, and with a copy of customer proof of purchase.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Lilliput shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Lilliput representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Lilliput supplies; or d) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product.
Please contact the nearest Lilliput’s Sales and Service Offices for services or a complete copy of the warranty statement.
Excepting the after-sales services provided in this summary or the applicable warranty statements, Lilliput will not offer any guarantee for maintenance definitely declared or hinted, including but not limited to the implied guarantee for marketability and special-purpose acceptability. Lilliput should not take any responsibilities for any indirect, special or consequent damages.
User Manual of PDS Series Oscilloscope
2
Table of Contents
General Safety Information................................................................................................ 4
Safety Terms and Symbols................................................................................................... 5
General features .................................................................................................................. 7
Junior User's Guidebook..................................................................................................... 8
Introduction to the Front Panel and the User's Interface of the PDS series Oscilloscope.... 9
Check the Package Contents .................................................................................................... 13
To perform the Function Inspection ........................................................................................ 14
To perform the Probe Compensation....................................................................................... 16
To set the Probe Attenuation Coefficient................................................................................. 17
To use the Probe Safely ............................................................................................................. 18
To perform Self-calibration ...................................................................................................... 18
Introduction to the Vertical System ......................................................................................... 19
Introduction to the Horizontal System .................................................................................... 20
Introduction to the Trigger System.......................................................................................... 21
Advanced User's Guidebook ............................................................................................. 22
To set the Vertical System ......................................................................................................... 23
Perform Mathematical Manipulation Function ..................................................................... 26
Using FFT function ................................................................................................................................ 26
Application of VERTICAL POSITION and VOLTS/DIV Knobs......................................... 30
To set the Horizontal system ..................................................................................................... 31
To set the Trigger System.......................................................................................................... 33
Trigger Control.......................................................................................................................... 34
To operate the Function Menu ................................................................................................. 39
To perform Sampling Setup...................................................................................................... 39
User Manual of PDS Series Oscilloscope
3
To set the Display System.......................................................................................................... 40
To Save and Recall a Waveform............................................................................................... 42
To perform the Auxiliary System Function Setting................................................................ 44
To perform the Automatic Measurement ................................................................................ 46
To perform the Cursor Measurement...................................................................................... 49
Use Autoscale ............................................................................................................................. 51
Use Executive controls............................................................................................................... 53
Quick Start Guide.............................................................................................................. 54
Example 1: Measuring a Simple Signal................................................................................... 54
Example 2: Working out the Gain of the Amplifier in the Metering Circuit....................... 55
Example 3: Capture the Single Signal ..................................................................................... 56
Example 4: Analyze the Details of a Signal............................................................................. 57
Example 5: Examine the Phase shift between two related signals ........................................ 59
Example 6: Video Signal Trigger.............................................................................................. 60
Troubleshooting.................................................................................................................61
Appendix A: T echnical Specifications.............................................................................. 62
Appendix B: Package Contents......................................................................................... 67
Appendix C: Maintenance, Cleaning and Repairing....................................................... 67
Appendix D: Battery Using Guide.................................................................................... 68
User Manual of PDS Series Oscilloscope
4
General Safety Information
Before any operations, please read the following safety precautions to avoid any possible bodily injury and prevent this product or any other products connected from damage. In order to avoid any contingent danger, this product is only used within the range specified.
Only the qualified technicians can implement the maintenance.
To avoid Fire or Personal Injury:
Connect the probe correctly. The grounding end of the probe corresponds to the
grounding phase. Please don't connect the grounding end to the positive phase.
Use Proper Power Cord. Use only the power cord supplied with the product and
certified to use in your country.
Connect or Disconnect Correctly. When the probe or test lead is connected to a
voltage source, please do not connect and disconnect the probe or test lead at random.
Product Grounded. This instrument is grounded through the power cord grounding
conductor. To avoid electric shock, the grounding conductor must be grounded. The product must be grounded properly before any connection with its input or output terminal.
When powered by AC power, it is not allowed to measure AC power source directly, because the testing ground and power cord ground conductor are connected together, otherwise, it will cause short circuit.
When powered by battery, the product must ground connection. To avoid electric shock, there must be a ground wire connect between ground and the ground port (on the back of product panel).
Check all Terminal Ratings. To avoid fire or shock hazard, check all ratings and
markers of this product. Refer to the user's manual for more information about ratings before connecting to the instrument.
Do not operate without covers. Do not operate the instrument with covers or panels
removed.
Use Proper Fuse. Use only the specified type and rating fuse for this instrument. Avoid exposed circuit. Do not touch exposed junctions and components when the
instrument is powered.
Do not operate if in any doubt. If you suspect damage occurs to the instrument, have
it inspected by qualified service personnel before further operations.
Use your Oscilloscope in a well-ventilated area. Make sure the instrument installed
with proper ventilation, refer to the user manual for more details.
Do not operate in wet conditions. Do not operate in an explosive atmosphere. Keep product surfaces clean and dry.
User Manual of PDS Series Oscilloscope
5
Safety Terms and Symbols
Safety Terms
Terms in this manual. The following terms may appear in this manual:
Warning: Warning indicates the conditions or practices that could result in
injury or loss of life.
Caution: Caution indicates the conditions or practices that could result in
damage to this product or other property.
Terms on the product. The following terms may appear on this product:
Danger: It indicates an injury or hazard may immediately happen.
Warning: It indicates an injury or hazard may be accessible potentially.
Caution: It indicates a potential damage to the instrument or other property might occur.
Safety Symbols
Symbols on the product. The following symbol may appear on the product:
Hazardous Voltage Refer to Manual
Protective Earth Terminal Chassis Ground
Test Ground
User Manual of PDS Series Oscilloscope
6
To avoid body damage and prevent product and connected equipment damage, carefully read the following safety information before using the test tool. This product can only be used in the specified applications.
Warning:
The two channels of the oscilloscope are non-isolated electrically. The channels should adopt common basis during measuring. To prevent short circuits, the 2 probe ground must not be connected to 2 different non-isolated DC level.
Warning:
The channels should adopt common basis during measuring. To prevent short circuits, the 2 probe ground must not be connected to 2 different non-isolated DC level. The diagram of the oscilloscope ground wire connection:
It is not allowed to measure AC power when charging the oscilloscope with adapter.
Warning:
To avoid fire or electrical shock, when the oscilloscope input signal connected is more than 42V peak (30Vrms) or on circuits of more than 4800VA, please take note of below items:
z Only use accessory insulated voltage probes and test lead. z Check the accessories such as probe before use and replace it if
there are any damages.
z Remove probes, test leads and other accessories immediately after
use.
z Remove USB cable which connects oscilloscope and computer. z Do not apply input voltages above the rating of the instrument
because the probe tip voltage will directly transmit to the oscilloscope. Use with caution when the probe is set as 1:1.
z Do not use exposed metal BNC or banana plug connectors. z Do not insert metal objects into connectors.
User Manual of PDS Series Oscilloscope
7
General features
Model Bandwidth Sample Rate
PDS5022S/T 25MHz 100MS/s
PDS6042S 40MHz 250MS/s
PDS6062S 60MHz 250MS/s
Record length of 6,000 points for each channel; Reading-out with the cursor; Twenty automatic measurement functions; Autoscale function; Color liquid crystal display of high resolution and high contrast; Storage and recall waveforms; Automatic setting function provided capable of fast setting; Multiple-waveform calculation function; Built-in FFT function; Perform detecting the average and peak values of the waveform; Real-time Digital Oscilloscope; Multiple triggering ,including Edge, video and Alternating triggering; RS232 or USB communication ports for PC; Different continuous displaying time; Multiple language User interface.
User Manual of PDS Series Oscilloscope
8
Junior User's Guidebook
This chapter deals with the following topics mainly:
z Introduction to the front panel and the user’s interface of the PDS series oscilloscope z To perform the general inspection z To perform the function inspection z To perform the probe compensation z To set the probe attenuation coefficient z To use the probe safely z To perform an self-calibration z Introduction to the vertical system z Introduction to the horizontal system z Introduction to the trigger system
User Manual of PDS Series Oscilloscope
9
Introduction to the Front Panel and the User's Interface of the PDS
series Oscilloscope
When you get a new-type oscilloscope, you should get acquainted with its front panel at first and the PDS series digital storage oscilloscope is no exception. This chapter makes a simple description of the operation and function of the front panel of the PDS series oscilloscope, enabling you to be familiar with the use of the PDS series oscilloscope in the shortest time. The PDS series oscilloscope offers a simple front panel with distinct functions to users for their completing some basic operations, in which the knobs and function pushbuttons are included. The knobs have the functions similar to other oscilloscopes. The 5 buttons in the column on the right side of the display screen are menu selection buttons (defined as F1 to F5 from top to bottom respectively), through which, you can set the different options for the current menu. The other pushbuttons are function buttons, through which, you can enter different function menus or obtain a specific function application directly.
User Manual of PDS Series Oscilloscope
10
Fig.1 The Front Panel of a PDS series Oscilloscope
Function Buttons
Fig.2 Explanatory Drawing for Operations of the PDS series Oscilloscope
Adjustment
Menu Selection
Vertical Control Connector
Horizontal
Control
Trigger Control
User Manual of PDS Series Oscilloscope
11
Fig.3 Illustrative Drawing of Display Interfaces
1. The Trigger State indicates the following information:
Auto: The oscilloscope is under the Automatic mode and is collecting the waveform under the non-trigger state.
Trig’d: The oscilloscope has already detected a trigger signal and is collecting the after-triggering information.
Ready: All pre-triggered data have been captured and the oscilloscope has been already ready for accepting a trigger.
Scan: The oscilloscope captures and displays the waveform data continuously in the scan mode.
Stop: The oscilloscope has already stopped the waveform data acquisition.
2. Waveform Viewing Area.
3. The purple pointer indicates the horizontal trigger position, which can be adjusted
by the horizontal position control knob.
4. The reading shows the time deviation between the horizontal trigger position and
the screen centre line, which equals 0 in the screen center.
5. This reading shows the time deviation between the horizontal trigger position and
the window centre line, which is regarded as 0 in the window center.
6. It indicates the current function menu.
7. It indicates the operation options for the current function menu, which changes
User Manual of PDS Series Oscilloscope
12
with the function menus.
8. The purple pointer shows the trigger level position.
9. The reading shows the trigger level value.
10. The reading shows the trigger source.
11. It shows the selected trigger type:
Rising edge triggering
Falling edge triggering
Video line synchronous triggering
Video field synchronous triggering
12. The reading shows the window time base set value.
13. The reading shows the main time base set value.
14. The two yellow dotted lines indicate the size of the viewing expanded window.
15. The icon shows the coupling mode of the CH2 channel.
“—“ indicates the direct current coupling
” indicates the AC coupling
” indicates GND coupling.
16. The reading shows the vertical scale factor (the Voltage Division) of the CH2
channel.
17. The icon indicates the coupling mode of the CH1 channel:
The icon "–" indicates the direct current coupling
The icon "" indicates the AC coupling
The icon "
” indicates GND coupling.
18. The reading indicates the vertical scale factor (the Voltage Division) of the CH1
channel.
19. The information shows the zero point positions of CH1 or CH2 channel.
20. The yellow pointer shows the grounding datum point (zero point position) of the
waveform of the CH2 channel. If the pointer is not displayed, it shows that this channel is not opened.
21. The red pointer indicates the grounding datum point (zero point position) of the
waveform of the CH1 channel. If the pointer is not displayed, it shows that the channel is not opened.
22. The positions of two purple dotted line cursors measurements.
User Manual of PDS Series Oscilloscope
13
Check the Package Contents
After you get a new PDS series oscilloscope, it is recommended that you should make a check on the instrument according to the following steps:
1. Check whether there is any damage caused by transportation.
If it is found that the packaging carton or the foamed plastic protection cushion has suffered serious damage, do not throw it away first till the complete device and its accessories succeed in the electrical and mechanical property tests.
2. Check the Accessories
The supplied accessories have been already described in the Appendix B “Accessories” of this Manual. You can check whether there is any loss of accessories with reference to this description. If it is found that there is any accessory lost or damaged, please get in touch with the distributor of LILLIPUT responsible for this service or the LILLIPUT’s local offices.
3. Check the Complete Instrument
If it is found that there is damage to the appearance of the instrument, or the instrument can not work normally, or fails in the performance test, please get in touch with the LILLIPUT’s distributor responsible for this business or the LILLIPUT’s local offices. If there is damage to the instrument caused by the transportation, please keep the package. With the transportation department or the LILLIPUT’s distributor responsible for this business informed about it, a repairing or replacement of the instrument will be arranged by the LILLIPUT.
User Manual of PDS Series Oscilloscope
14
To perform the Function Inspection
Make a fast function check to verify the normal operation of the instrument, according to the following steps:
1. Connect the Instrument to the Power and Push down the Power Switch Button.
The instrument carries out all self-check items and shows the prompt “Press any Key Enter the Operating Mode”. Press the “UTILITY” button to get access to the “FUNCTION” menu and push down F2 the menu selection button to call out the function “Recall Factory”. The default attenuation coefficient set value of the probe in the menu is 10X, shown as Fig.4.
Fig.4 Power on the Instrument
2. Set the Switch in the Oscilloscope Probe as 10X and Connect the Oscilloscope with CH1 Channel.
Align the slot in the probe with the plug in the CH1 connector BNC, and then tighten the probe with rotating it to the right side.
Connect the probe tip and the ground clamp to the connector of the probe compensator, shown as Fig.5.
User Manual of PDS Series Oscilloscope
15
Probe comp
CH1
Fig.5 Connection of the Probe
3. Press the AUTO SET Button.
The square wave of 1 KHz frequency and 5V peak-peak value will be displayed in several seconds (see Fig.6).
Fig.6 Auto set
Check CH2 by repeating Step 2 and Step 3.
User Manual of PDS Series Oscilloscope
16
To perform the Probe Compensation
When connect the probe with any input channel for the first time, make this adjustment to match the probe with the input channel. The probe which is not compensated or presents a compensation deviation will result in the measuring error or mistake. For adjusting the probe compensation, please carry out the following steps:
1. Set the attenuation coefficient of the probe in the menu as 10X and that of the
switch in the probe as 10X, and connect the oscilloscope probe with the CH1 channel. If a probe hook tip is used, ensure that it keeps in close touch with the probe. Connect the probe tip with the signal connector of the probe compensator and connect the reference wire clamp with the ground wire connector of the probe connector, and then press the button AUTOSET (see Fig.5).
2. Check the displayed waveforms and regulate the probe till a correct
compensation is achieved (see Fig.7 and Fig.8).
Fig.7 Displayed Waveforms of the Probe Compensation
3. Repeat the steps mentioned if necessary.
Fig.8 Adjust Probe
User Manual of PDS Series Oscilloscope
17
To set the Probe Attenuation Coefficient
The probe has several attenuation coefficients, which will influence the vertical scale factor of the oscilloscope.
If it is required to change (check) the set value of the probe attenuation coefficient, press the function menu button of the channels used, then push down the selection button corresponding to the probe till the correct set value is shown.
This setting will be valid all the time before it is changed again.
Note: The attenuation coefficient of the probe in the menu is preset to 10X when
the oscilloscope is delivered from the factory.
Make sure that the set value of the attenuation switch in the probe is the same as the menu selection of the probe in the oscilloscope.
The set values of the probe switch are 1X and 10X (see Fig.9).
Fig.9 Attenuation Switch
Note: When the attenuation switch is set to 1X, the probe will limit the bandwidth
of the oscilloscope in 5MHz. If it is needed to use the whole bandwidth of the oscilloscope, the switch must be set to 10X.
User Manual of PDS Series Oscilloscope
18
To use the Probe Safely
The safety guard ring around the probe body protects your finger against the electric shock, shown as Fig.10.
Fig.10 Finger Guard
Warning: In order to avoid suffering from the electric shock, please keep your
finger behind the safety guard ring of the probe body during the operation.
In order to protect you from suffering from the electric shock during your using the
probe, do not touch the metal part of the probe tip when the probe is connected to the power supply.
Before making any measurements, please connect the probe to the instrument and
connect the ground terminal to the earth.
To perform Self-calibration
The self-calibration application can make the oscilloscope reach the optimum condition rapidly to obtain the most accurate measurement value. You can carry out this application program at any time, but when the range of variation of the ambient temperature is up to or over 5 , this program must be executed.
For the performing of the self-calibration, all probes or wires should be disconnected with the input connector first. Then, press the “UTILITY” button to call out the FUNCTION menu; push down the F3 menu selection button to choose the option “ Perform Self-Calibration”; finally, run the program after confirming that everything is ready now.
User Manual of PDS Series Oscilloscope
19
Introduction to the Vertical System
Shown as Fig.11, there are a series of buttons and knobs in VERTICAL CONTROLS. The following practices will gradually direct you to be familiar with
the using of the vertical setting.
Fig.11 Vertical Control Zone
1. Use the button “VERTICAL POSITION” knob to show the signal in the center of
the waveform window. The “VERTICAL POSITION” knob functions the regulating of the vertical display position of the signal. Thus, when the “VERTICAL POSITION” knob is rotated, the pointer of the earth datum point of the channel is directed to move up and down following the wave form.
Measuring Skill
If the channel is under the DC coupling mode, you can rapidly measure the DC component of the signal through the observation of the difference between the wave form and the signal ground.
If the channel is under the AC mode, the DC component will be removed by filtration. This mode helps you display the AC component of the signal with a higher sensitivity.
2. Change the Vertical Setting and Observe the Consequent State Information
Change.
With the information displayed in the status bar at the bottom of the waveform window, you can determine any changes in the channel vertical scale factor.
z Rotate the vertical “VOLTS/DIV” knob and change the “Vertical Scale Factor
(Voltage Division)”, it can be found that the scale factor of the channel corresponding to the status bar has been changed accordingly.
z Press buttons of “CH1 MENU”, “CH2 MENU” and “MATH MENU”, the
operation menu, symbols, wave forms and scale factor status information of the corresponding channel will be displayed in the screen.
Loading...
+ 49 hidden pages