InTemp® CX400 Series Temperature Logger Manual
InTemp CX400 series loggers measure temperature in indoor monitoring applications. Designed to meet World Health Organization (WHO) and Centers for Disease Control and Prevention (CDC) guidelines, this logger is ideal for clinical applications, such as vaccine storage and pharmaceutical manufacturing. This Bluetooth® Low Energy-enabled logger is designed for wireless communication with a mobile device. Using the InTemp app, you can easily configure the logger with one of four preset profiles designed for ambient storage, clinical refrigerator, freezer, or refrigerator monitoring or you can set up a custom profile for other applications. You can also quickly perform daily logger checks, download reports, and monitor tripped alarms. Or, you can use InTempConnect® to configure and download CX series loggers via the CX5000 Gateway. The InTempVerify™ app is also available to easily download loggers and automatically upload reports to InTempConnect. Use the built-in LCD screen on the logger to check the current temperature, daily maximum or minimum temperatures, logging status, battery use, and more. Once data is uploaded to InTempConnect, you can track logger configurations and automatically upload logger data to build custom reports for further analysis. The CX402 model uses a 2- or 4-meter probe and is available with a 5, 15, or 30 mL glycol bottle (bottle holder included). It also contains an internal sensor for ambient
InTemp CX400 Temperature Loggers
temperature monitoring. The CX403 model is available with an internal sensor only.
Specifications
Models:
CX402-T205 & CX402-VFC205, 2 meter probe and 5 mL glycol bottle
CX402-T215 & CX402-VFC215, 2 meter probe and 15 mL glycol bottle
CX402-T230 & CX402-VFC230, 2 meter probe and 30 mL glycol bottle
CX402-T405 & CX402-VFC405,
4 meter probe and 5 mL glycol bottle
CX402-T415 & CX402-VFC415, 4 meter probe and 15 mL glycol bottle
CX402-T430 & CX402-VFC430 4 meter probe and 30 mL glycol bottle
CX402-T2M, CX402-B2M, &
CX402-VFC2M, 2 meter probe
CX402-T4M, CX402-B4M, &
CX402-VFC4M, 4 meter probe
CX403
Included Items:
Double-sided tape (for models with glycol bottles)
Two AAA 1.5 V alkaline batteries
Battery door and screw
NIST Certificate of Calibration
Required Items:
InTemp app
Device with iOS or Android™
External Temperature Probe
Range -40° to 100°C (-40° to 212°F)
Accuracy CX40x-Txxx and CX402-BxM:
±1.0°C from -40° to -15°C (±1.8°F from -40° to 5°F) ±0.5°C from -15° to 80°C (±0.9°F from 5° to 176°F) ±0.7°C from 80° to 100°C (±1.3°F from 176° to 212°F)
CX40x-VFCxxx: ±1.0°C from -40° to -22°C (±1.8°F from -40° to -8°F) ±0.5°C from -22° to 50°C (±0.9°F from -8° to 122°F) ±1.0°C from 50° to 100°C (±1.8°F from 122° to 212°F)
Resolution 0.024°C at 25°C (0.04°F at 77°F)
Drift <0.1°C (0.18°F) per year
NIST Calibration CX40x-Txx and CX402-BxM: Single point NIST calibration, probe and
logger body CX40x-VFCxxx: Single point NIST calibration, probe only
CX403: Single point NIST calibration
Cable Length 2 or 4 meter (6.56 or 13.12 feet) flat ribbon cable
Probe Dimensions Food-grade stainless steel probe with pointed tip, 53.34 mm (2.1 inches)
long, 3.18 mm (0.125 inches) diameter
Ambient Temperature Sensor
Range -30° to 70°C (-22° to 158°F)
Accuracy CX40x-Txxx and CX402-BxM:
±0.5°C from -15° to 70°C (±0.9°F from 5° to 158°F) ±1.0°C from -30° to -15°C (±1.8°F from -22° to 5°F)
CX40x-VFCxxx: ±1.0°C from -30° to -22°C (±1.8°F from -22° to -8°F) ±0.5°C from -22° to 50°C (±0.9°F from -8° to 122°F) ±1.0°C from 50° to 70°C (±1.8°F from 122° to 158°F)
Resolution 0.024°C at 25°C (0.04°F at 77°F)
Drift <0.1°C (0.18°F) per year
Logger
Radio Power 1 mW (0 dBm)
Transmission Range Approximately 30.5 m (100 ft) line-of-sight
Wireless Data Standard Bluetooth Low Energy (Bluetooth Smart)
Logger Operating Range -30° to 70°C (-22° to 158°F), 0 to 95% RH (non-condensing)
Logging Rate 1 second to 18 hours
19996-Q
InTemp CX400 Series Temperature Logger Manual
Specifications (continued)
Time Accuracy ±1 minute per month at 25°C (77°F)
Battery Type Two AAA 1.5 V alkaline or lithium batteries, user replaceable
Battery Life 1 year, typical with logging interval of 1 minute. Faster logging intervals,
Memory 128 KB (84,650 measurements, maximum)
Full Memory Download Time
LCD LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or
Dimensions 9.4 x 4.5 x 2.59 cm (3.7 x 1.77 x 1.02 inches)
Weight 90.2 g (3.18 oz)
Environmental Rating IP54
Passed RTCA DO160G, part 21H
Logger Components and Operation
Audible alarm speaker
Alarm LED
LCD
remaining connected with the InTemp app, excessive report generation, numerous audible alarms, and paging all impact battery life.
Approximately 60 seconds; may take longer the farther the device is from the logger
go blank in temperatures outside this range
The CE Marking identifies this pr directives in the European Union (EU).
See last page
oduct as complying with all relevant
LCD: This screen shows the latest temperature reading and other status information. The LCD screen refreshes at the same rate as the logging interval. The example shows all symbols illuminated on the LCD screen followed by a table with descriptions of each symbol.
Start button
External temperature probe (CX402 models only)
Temperature sensor (behind rectangular plug)
One of four magnets (on back of logger)
Mute or Next button
Start Button: Press this button for 1 second to start the logger when it is configured to start “on button push.”
Mute or Next Button: Press this button for 1 second to mute a beeping alarm (see Logger Alarms). For CX402 loggers, press this button for 1 second to switch between the external probe and ambient internal sensor temperature readings.
Magnets: There are four magnets on the back of the logger (not shown in diagram) for mounting.
Temperature Sensor: This internal sensor measures ambient temperature.
External Temperature Probe: This is the probe that is plugged into the logger for measuring temperature (CX402 models only).
Audible Alarm Speaker: This is the speaker for the audible alarm that beeps when an alarm is tripped or the external probe is removed (if applicable).
Alarm LED: This LED blinks every 5 seconds when an alarm is tripped or the external probe is removed (if applicable). See Logger Alarms.
LCD Symbol Description
An alarm has tripped because the temperature reading is outside the specified range
The logger has bee twice daily checks (twice daily is shown in this example), but no checks have been performed yet.
The once daily or t this example) has been performed.
This indicates how much me for the current configuration. In this example, approximately 40 percent of the memory has been used.
This shows the appr remaining.
The logger is cu
The logger is cu tablet via BlueTooth. The more bars there are, the stronger the signal.
The logger is waiting to be star the Start button for 3 seconds to start the logger.
n configured to perform daily or
wice daily logger check (twice in
mory has been used
oximate battery power
rrently logging.
rrently connected to a phone or
ted. Press and hold
2
InTemp CX400 Series Temperature Logger Manual
LCD Symbol Description
This is an example of a temperature reading from an external probe.
This is an example of a temperature reading from the internal sensor.
This is an example of a minimum temperature, which is the lowest probe temperature reading from the day (CX402 models) or the lowest ambient temperature reading (CX403 model) within the current 24-hour period (midnight from one day to midnight the next day) if the logger was set to record logger checks (see Performing Logger Checks). Otherwise the minimum and maximum readings represent the entire logging period and only reset when the logger is downloaded and restarted or stopped and reconfigured.
This is an example of a maximum temperature, which is the highest probe temperature reading from the day (CX402 models) or the lowest ambient temperature reading (CX403 model) within the current 24-hour period (midnight from one day to midnight the next day) if the logger was set to record logger checks (see Performing Logger Checks). Otherwise the minimum and maximum readings represent the entire logging period and only reset when the logger is downloaded and restarted or stopped and reconfigured.
The external prob
and
(if applicable).
MUTE indicates an alar
or
beeping alarm by pressing the Mute button. The LCD then changes to MUTED.
The audible alarm has been
with your phone or tablet and then download reports, which are saved in the app and automatically uploaded to InTempConnect. The CX5000 Gateway is also available to automatically configure and download loggers and upload data to InTempConnect. Or, anyone can download a logger using the InTempVerify app if the loggers are enabled to be used with InTempVerify. If you do not need to access logged data via the cloud-based InTempConnect software, then you also have the option to use the logger with the InTemp app only
Follow these steps to begin using the loggers with InTempConnect and the InTemp app.
e is not connected to the logger
m is beeping. Turn off the
muted.
The logger has been configur a delay. The display will count down in days, hours, minutes, and seconds until logging begins. In this example, 5 minutes and 38 seconds remain until logging will begin.
The profile settings ar logger.
An erro
r occurred while loading the profile settings onto the logger. Try reconfiguring the logger.
The logger was paged from the
The logger has bee with the InTemp app or because the memory is full.
The logger is being
n downloaded and stopped
updated with new
ed to start logging on
e being loaded onto the
InTemp app.
firmware.
Note: If the logger has stopped logging because the memory is full, the LCD screen will remain on with “STOP” displayed until the logger is downloaded to your mobile device. Once the logger has been downloaded, the LCD will turn off automatically after 2 hours. The LCD will turn back on the next time the logger connects with your device.
Getting Started
InTempConnect is web-based software where you can monitor CX400 series logger configurations and view downloaded data online. Using the InTemp app, you can configure the logger