Omron RX-3 User Manual

RX3 Plus English
Introduction
The OMRON RX3 Plus measures your blood pressure and pulse simply and quickly from the wrist, without the use of an inflation bulb or stethoscope. The measurement values are stored in the memory of the OMRON RX3 Plus. The OMRON RX3 Plus fits wrist circumferences from 13.5 to 19.5 cm.
1
RX3 Plus
Digital Automatic
Blood Pressure Monitor
Instruction manual page 1
RX3 Plus English
Table of contents
1 How to obtain meaningful blood pressure readings 3 2 Overview 4 3 Preparation 5
3.1 Insert batteries 5
3.2 Fit wrist cuff 6
4 Operation 7
4.1 Control functions 7
4.2 Set date and time 8
4.3 Measurement 9
4.4 Read display 10
5 Memory 11
5.1 Recall average value 11
5.2 Recall previous values 11
5.3 Erase memory 12
6 Storage 13 7 Faults: causes and rectifications 13 8 Maintenance and spare parts 14 9 Technical data 15
10 General information about blood pressure 16
2
RX3 Plus English
1 How to obtain meaningful blood pressure readings
General
• The OMRON RX3 Plus is not suitable for measuring the frequency of cardiac pacemakers.
• Consult your doctor during pregnancy, arrhythmia and arteriosclerosis.
• Avoid eating, drinking (alcohol), smoking, sport and taking a bath before measurement.
• Be calm and relaxed before and during measurement.
• You should never change the dose of medicines prescribed by your doctor.
Tips for use
• Check your blood pressure at least twice a day (before breakfast, after work).
• Do not measure your blood pressure while you are in a vehicle.
• Always measure on the same wrist, prefer your left wrist for a comfortable use.
• Wrap the wrist cuff around your wrist before start of measurement.
Caution
• Operating temperature between 10°C and 40°C.
• Operating environment must be free from excessive vibrations, shocks, magnetic fields, electrical noise, etc.
• Stay out of sunlight during measuring.
• Keep portable phones 5 meters away during measurement.
• Do not wrap the cuff other than around your wrist.
• Do not drop the OMRON RX3 Plus.
3
2 Overview
A
F
E
D
G
C
B
RX3 Plus English
A Display B Wrist Cuff C Set button D Adjust button
E Memory button F On/off button G Battery Compartment
4
3 Preparation
3.1 Insert batteries
RX3 Plus English
1 Open the battery compartment. 2 Insert two batteries in the battery
3 Close the battery compartment.
5
compartment. Caution! Use two identical 1.5V alkaline batteries type AAA!
3.2 Fit wrist cuff
S Y
S mmH
g
D
I Am
m
H
g
P UL
S
E /
m i n
M
A D
JU S
T
S E
T
DATE/
TI
M E
SET
SYSmmHg
DIAmmHg
PULSE/
mi
n
M
A DJUS
T
S
E T
DATE/
TI
ME
SET
SYSm
m
Hg
DI
AmmH
g
P
ULS
E
/m
in
M
A
D
JU
S
T
S
E
T
D
A
TE
/TIME
S
E
T
Caution
! Do not press on/off button when the wrist cuff is not wrapped
around the wrist!
RX3 Plus English
1 Bare your wrist.
Caution! Push up your sleeve:
avoid constricting the blood flow!
3 Hold the end of the wrist cuff and
wrap it around your wrist. Caution! Be sure the wrist cuff fits closely; do not pinch your arm!
2 Put the monitor on the inner side
of your wrist.
Note: attach the monitor upside down for reliable m
easurement at
your right wrist.
6
SYSmmHg
DIAmmHg
P
ULSE/min
M
ADJUSTSET
DATE/TIME SET
C
D
E
F
H
K
J
P
L Q
M
N
T
R
S
4 Operation
4.1 Control functions
RX3 Plus English
C Set button D Adjust button E Memory button F On/off button
7
H Date. J Systolic blood pressure in
mmHg.
K Diastolic blood pressure in
mmHg.
L Memory display: 21 values in
memory.
M Average value of the latest 3
measurements displayed.
N Monitor blood pressure: flash -
measurement takes place, light up - finish measurement.
P Pulse: heartbeats per minute. Q Time. R Inflation: wrist cuff inflates and
measurement starts.
S Deflation: wrist cuff deflates and
measurement finishes.
T Battery exhausted: batteries are
weak/exhausted.
RX3 Plus English
S
YSmmHg
DIAmmHg
PULSE/min
M
ADJUSTSET
DATE/TIME SET
SYSmmHg
DIAmmHg
P
ULSE/min
M
ADJUSTSET
DATE/TIME SET
SYSmmHg
DIAmmHg
PULSE/min
M
ADJUSTSET
DATE/TIME SET
SYSmmHg
DIAmmHg
PULSE/min
M
ADJUSTSET
D
ATE/TIME SET
SYSmmHg
DIAmmHg
PULSE/min
M
ADJUSTSET
D
ATE/TIME SET
SYSmmHg
DIAmmHg
PULSE/min
M
ADJUSTSET
DATE/TIME SET
SYSmmHg
DIAmmHg
PULSE/min
M
ADJUSTSET
DATE/TIME SET
4.2 Set date and time
Note
: each time you press adjust button year (2001-2030), month (1-12), day (1-31), hour (0-23) and minutes (0-59) advance by one unit. Keep button pressed to advance fast.
Note: use the grey coloured button to carry out the instruction.
1 Press set or adjust button. 2 Press set button for 2 seconds.
Press adjust button to set the current year.
3 Press set button to proceed to
month setting. Press adjust button to set the current month.
4 Press set button to proceed to
day setting. Press adjust button to set the current day.
8
RX3 Plus English
SYSmmHg
DIAmmHg
P
ULSE/min
M
ADJUSTSET
DATE/TIME SET
SYSmmHg
DIAmmHg
P
ULSE/min
M
ADJUSTSET
DATE/TIME SET
SYSmmHg
D
IAmmHg
PULSE/min
M
ADJUSTSET
DATE/TIME SET
SYSmmHg
D
IAmmHg
PULSE/min
M
ADJUSTSET
DATE/TIME SET
SYSmmHg
D
IAmmHg
P
ULSE/min
M
ADJUSTSET
DATE/TIME SET
SYSmmHg
DIAmmHg
PULSE/min
M
ADJUSTSET
DATE/TIME SET
S
YSmmHg
D
IAmmHg
PULSE/min
M
ADJUSTSET
DATE/TIME SET
5 Press set button to proceed to
hour setting. Press adjust button to set the current hour.
6 Press set button to proceed to
minute setting and press adjust button to set the current minutes.
7 Press set button to complete setting. 8 Press on/off button.
4.3 Measurement
! Allow an interval of at least three minutes between two
Caution
successive measurements!
1 Sit comfortably upright with your feet flat on the floor.
Caution! The OMRON RX3 Plus should always be at heart height during
measurement!
9
RX3 Plus English
SYSmmHg
DIAmmHg
P
ULSE/min
M
ADJUSTSET
DATE/TIME SET
SYSmmHg
DIAmmHg
PULSE/min
M
ADJUSTSET
DATE/TIME SET
SYSmmHg
DIAmmHg
PULSE/min
M
ADJUSTSET
DATE/TIME SET
SYSmmHg
DIAmmHg
PULSE/min
M
ADJUSTSET
DATE/TIME SET
2 Press on/off button to start measurement.
The OMRON RX3 Plus measures during inflation! Caution! Sit still, do not move, do not speak and do not touch the OMRON RX3 Plus!
4.4 Read display
1 Read the values of blood pressure and pulse from the display.
Note: display shows alternately date, blood pressures and time, heartbeats.
2 Note date, time and values in your
personal blood pressure pass.
3 Press on/off button to turn off
power.
10
RX3 Plus English
L M
SYSmmHg
DIAmmHg
P
ULSE/min
M
ADJUSTSET
DATE/TIME SET
S
YSmmHg
DIAmmHg
PULSE/min
M
ADJUSTSET
DATE/TIME SET
H Q
SYSmmHg
DIAmmHg
PULSE/min
M
ADJUSTSET
DATE/TIME SET
SYSmmHg
DIAmmHg
PULSE/min
M
ADJUSTSET
DATE/TIME SET
SYSmmHg
DIAmmHg
PULSE/min
M
ADJUSTSET
DATE/TIME SET
5 Memory
5.1 Recall average value
L Memory display: 21 values in memory. M Average value of the latest 3 measurements displayed.
1 Press memory button.
Note: display shows alternately blood pressures and heartbeats.
5.2 Recall previous values
H/Q Date/time of measurement. 1 Press memory button for latest and previous values.
Press on/off button to turn off power. Note: display shows alternately date, blood pressures and time, heartbeats.
11
SYSmmHg
D
IAmmHg
P
ULSE/min
M
ADJUSTSET
DATE/TIME SET
SYSmmHg
DIAmmHg
P
ULSE/min
M
ADJUSTSET
DATE/TIME SET
5.3 Erase memory
RX3 Plus English
1 Press memory button to turn on
power.
2 Press memory button and
simultaneously on/off button to erase memory. Caution! All stored values will be erased! Press on/off button to turn off power.
12
RX3 Plus English
6 Storage
• Store the OMRON RX3 Plus in the hard case in a dry, closed place with a
temperature between -20°C and 60°C.
• Remove the batteries if you not intend to use the OMRON RX3 Plus for 3
months or more.
7 Faults: causes and rectifications
Fault Cause Rectification
Correct reading not possible
Low voltage Batteries
Display empty
Inexplicable values Wrong working
13
Movement, sound vibration, weak pulse
weak/exhausted
method
Wait for at least three minutes to repeat measurement
Fit two new identical 1.5V alkaline batteries type AAA
Follow instructions, repeat measurement
8 Maintenance and spare parts
RX3 Plus English
Cleaning instruction
• Use a soft, slightly moistened
cloth to clean the OMRON RX3 Plus and the wrist cuff. Caution! Do not use petrol, thinners or similar solvents! Caution! Do not wash the wrist cuff!
• Do not carry out repairs of any
kind yourself. If a defect occurs consult your OMRON retail outlet or distributor as mentioned on the packaging.
Calibration
• It is recommended to have the
OMRON RX3 Plus inspected every two years to ensure correct function and accuracy. Consult your OMRON retail outlet or OMRON distributor as mentioned on the packaging.
Replace batteries
• Remove the battery out of the
battery compartment and insert two new identical 1.5V alkaline batteries type AAA.
Spare parts
• Due to high company quality standards, OMRON considers the main unit as a non-serviceable part because the need for proper calibration after replacement of high tech components.
• If the cuff needs to be replaced have this done by an authorised expert. Consult your OMRON retailer or OMRON distributor as mentioned on the packaging.
Attention! Disposal of batteries and the OMRON RX3 Plus should be carried out in accord­ance with the national regulations for the disposal of electronic products.
14
RX3 Plus English
9 Technical data
Product name OMRON Digital Blood Pressure Monitor Model RX3 Plus Display Digital LCD Measurement range Pressure: 0 - 299 mmHg / Pulse: 40 - 180 /minute Accuracy Pressure: ± 3 mmHg / Pulse: ± 5% of display reading Inflation Automatic by electric pump Deflation Automatic rapid deflation system Pressure detection Electrostatic capacity pressure sensor (capacitive) Measurement method Oscillometric method Memory 21 Measurements Power supply 2 x 1.5V Alkaline batteries AAA Battery life New batteries will last for approx. 400 measurements Operating temperature /humidity +10°C to +40°C / 30% to max. 85% relative humidity Operating environment Free from excessive vibrations, shocks, magnetic
Storage temperature / humidity -20°C to +60°C / 10% to max. 95% relative humidity Weight Approx. 160 g including batteries Outer dimensions Approx. 79 mm (W) x 47 mm (H) x 71 mm (D) Wrist circumference Approx. 13.5 cm to 19.5 cm Accessories Instruction manual, guarantee card, blood pressure
Note Subject to technical modification without prior notice
= Type B 0197
Manufacturer OMRON HEALTHCARE Co., Ltd.
EU-representative OMRON HEALTHCARE EUROPE B.V.
This device fulfils the provisions of the EC directive 93/42/EEC (Medical Device Directive). This blood pressure monitor is designed according to the European standard EN1060, Non-invasive sphygmomanometers Part 1: General Requirements and Part 3: Supplementary requirements for electromechanical blood pressure measuring systems.
fields, electrical noise, etc.
pass, 2 alkaline batteries, storage box
24, Yamanouchi Yamanoshita-cho, Ukyo-ku, Kyoto, 615-0084 Japan
Kruisweg 577, NL-2132 NA Hoofddorp
15
RX3 Plus English
Optimal blood
pressure
(target value)
Normal blood
pressure
Normal systolic value
Mild hypertension
Moderate hypertension
Severe hypertension
Systolic blood pressure (mmHg)
Diastolic blood pressure (mmHg)
80
180
160
140
130
120
85 90 100 110
10 General information about blood pressure
Blood circulation
The blood circulation is responsible for supplying the body with oxygen. Blood pressure is the pressure exerted on the arteries. The systolic blood pressure value (higher pressure or top value)represents the blood pressure produced by contraction of the heart muscle. The diastolic blood pressure value (lower pressure or lower value) represents the blood pressure produced by relaxation of the heart muscle.
Classification of blood pressure
The World Health Organisation (WHO) and the International Society of Hypertension (ISH) developed the blood pressure classification. This classification is based on the blood pressure values measured while sitting on a chair in the outpatient department of a hospital.
According to the blood pressure classification by the WHO/ISH.
Health and blood pressure
The incidence of hypertension increases with age. In addition, a lack of exercise, excess body fat and high levels of cholesterol (LDL), which sticks to the inside in blood vessels, reduces elasticity of these vessels.
16
RX3 Plus English
Hypertension accelerates arterial sclerosis which can lead to very serious conditions such as stroke and myocardial infarction. For these reasons it is very important to know whether the blood pressure is within a healthy range. Blood pressure fluctuates from minute to minute, throughout the day. Therefore it is essential to take regular measurements to help you identify an average blood pressure.
Symptoms of high blood pressure
High blood pressure can go unnoticed for a long time, since it does not cause noticeable symptoms. The following are all possible causes of abnormally high blood pressure:
• Overweight
• High cholesterol level
• Smoking
• Excessive alcohol consumption
• Stress and emotional upset
• Excessive consumption of salt
• Lack of physical exercise
• Genetic/hereditary predisposition
• Underlying illnesses, such as kidney disorders or endocrine disturbance
Blood pressure measurement
Measuring your blood pressure daily enables you to recognize high blood pressure at an early stage and obtain medical treatment.
The oscillometric method of measurement determines your blood pressure by measuring the pressure fluctuations caused by the pulse waves.
Since your blood pressure fluctuates within a day (see graphic next page), you should always measure it at the same time each day to ensure that the measurements are genuinely comparable. At the doctor’s surgery, nervousness and tension may cause your blood pressure to be higher than in familiar circumstances at home. This is known as “White Coat Hypertension”.
17
RX3 Plus English
mmHg
200
1
50
100
50
6 12 18 24
Upper curve: systolic blood pressure
Lower curve: diastolic blood pressure
Time of day
Treatment of high blood pressure
Example: fluctuation within a day (male, 35 years old)
If your blood pressure reaches upper values of 140-160 mmHg and lower values of 90 - 95 mmHg in repeated measurements over several days, you should consult your doctor for detailed medical examination. You can assist the treatment prescribed by your doctor in the following ways:
• Loose weight and lower your cholesterol level by reducing the calories and fat in your diet. Cut down on animal fats and eat more fruit and vegetables.
• Reduce the consumption of alcohol.
• Reduce the intake of salt: the German League for Combating High Blood Pressure recommends six grams of salt per day (i.e. one teaspoon) on average.
• Stop smoking.
• Take regular exercise.
• Monitor your blood pressure.
18