Omron RS8 User Manual

Thank you for purchasing the OMRON RS8 Wrist Blood Pressure
Wrist Blood
Pressure Monitor
Model RS8
Instruction Manual
IM-HEM-6310F-E-EN-01-11/2012
1883294-6A
A
E
D
H
J I
C
F
G
B
A
B
C
D
E
F
G I
J
K
P
Q
R
M
N
O
L
H
S
U
T
Year
Month
Day
Hour
Minute
1) Push the button to change.
• Hold down to advance the digit rapidly.
1
2
Cuff Wrapping Guide
START INFLATE DEFLATE COMPLETE
Alternating date/time display
Monitor.
The OMRON RS8 is a compact and easy to use blood pressure monitor, operating on the oscillometric principle. It measures your blood pressure and pulse rate simply and quickly. For comfortable controlled inflation without the need of pressure pre-setting or re-inflation the device uses its advanced “IntelliSense” technology.
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia or arteriosclerosis. Please read this section carefully before using the unit.
• Indicates a potentially hazardous situation which, if not avoided,
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results
• People with severe blood flow problems, or blood disorders, should
(Battery Usage)
• If battery fluid should get in your eyes, immediately rinse with
• Indicates a potentially hazardous situation which, if not avoided,
(General Usage)
• Do not leave the unit unattended with infants or persons who
• Do not use the unit for any purpose other than measuring blood
• Do not use a mobile phone or other devices that emit
• Do not disassemble the unit or wrist cuff.
• Do not operate the unit in a moving vehicle (car, airplane).
(Battery Usage)
• If battery fluid should get on your skin or clothing, immediately rinse
• Use only two “AAA” alkaline (LR03) batteries with this unit. Do not
• Do not insert the batteries with their polarities incorrectly aligned.
• Replace old batteries with new ones immediately. Replace both
• Remove the batteries if the unit will not be used for three months or
• Do not use new and used batteries together.
General Precautions
• Do not apply strong shocks and vibrations to or drop the unit.
• Do not take measurements after bathing, drinking alcohol or
• Do not inflate the wrist cuff when it is not wrapped around your
• Read and follow the “Important information regarding Electro
• Read and follow the “Correct Disposal of This Product” in the
Introduction
Please read this instruction manual thoroughly before using the unit. Please keep for future reference. For specific information about your own blood pressure, CONSULT YOUR DOCTOR.
Important Safety Information
Warning:
could result in death or serious injury.
and self-treatment are dangerous.
consult a doctor before using the unit, as cuff inflation can cause internal bleeding.
plenty of clean water. Consult a doctor immediately.
Caution:
may result in minor or moderate injury to the user or patient or damage to the equipment or other property.
cannot express their consent.
pressure.
electromagnetic fields, near the unit. This may result in incorrect operation of the unit.
with plenty of clean water.
use other types of batteries.
batteries at the same time.
more.
caffeine, smoking, exercising or eating.
wrist.
Magnetic Compatibility (EMC)” in the Technical Data Section.
Technical Data Section when disposing of the device and any used accessories or optional parts.
1. Overview
Main unit:
A. Wrist cuff B. Battery compartment C. Display D. User ID selection switch E. Weekly average ( ) button F. Memory ( ) button G. Transfer area H. START/STOP button I. Date/Time setting ( ) button
J. Da te/Time setting ( ) button
Display:
A. Memory symbol B. Systolic blood pressure C. Diastolic blood pressure D. Battery low symbol E. Heartbeat symbol (Flashes during measurement) F. Irregular heartbeat symbol G. Average value symbol H. Date/Time display I. Morning average symbol J. E vening average symbol K. Blood pressure trend graph L. Blood pressure level indicator M. Positioning indicator N. Positioning symbol O. Cuff wrapping guide P. User ID symbol (A or B) Q. OK symbol R. Transfer indicator S. Deflation symbol T. Pulse display and Memory number U. Movem ent error symbol
2. Preparation
2.1 Installing/Replacing the Batteries
1. Remove the battery cover.
2. Insert two 1.5 V “AAA”
alkaline (LR03) batteries as indicated in the battery compartment and then replace the battery cover.
Notes:
• If the battery low symbol ( ) appears on the display, turn off the unit then replace both batteries at the same time.
• The measurement values continue to be stored in memory even after the batteries are replaced.
Disposal of used batteries should be carried out in accordance with the national regulations for the disposal of batteries.
2.2 Setting the Date and Time
1. Press the button.
2. Set the monitor to the correct date and time
before taking a measurement for the first time.
2) Push the button to confirm.
• Next setting appears.
3. Press the START/STOP button to store the setting.
Notes:
• If the batteries have been removed, the date and time setting will need to be reset.
• If the date and time are not set, “-:--” appears during or after measurement.
3. Using the Unit
3.1 Applying the Wrist Cuff
Do not apply over clothing.
1. Put your arm through the
cuff loop.
Your palm should face upward.
2. Wrap the wrist cuff around
your wrist.
Wrap the wrist cuff securely around the wrist for taking accurate measurements.
Make sure that the wrist cuff does not cover the protruding part of the wrist bone (ulna) on the outside of the wrist.
Notes:
• You can take a measurement on either your left or right wrist.
• If the wrist cuff comes off, assemble as shown in the figure.
3.2 How to Sit Correctly
To take a measurement, you need to be relaxed and comfortably seated, under comf ortable room temperature. No bathing, drinking alcohol or caffeine, smokin g, exercising or eating 30 minutes before taking a measurement.
• Sit on a chair with your feet flat on the floor.
• Sit upright with your back straight.
• The cuff should be at the same level as your heart.
• Relax your wrist and hand. Do not bend your wrist back, clench your fist, or bend your wrist forward.
3.3 Taking a Reading
Notes:
• To cancel a measurement, press the START/STOP button at any time during measurement.
• Remain still while taking a measurement.
1. Select your User ID.
2. Press the START/STOP button.
All the symbols appear on the display. While your position is improper, the Positioning symbol and the Positioning indicator flash, when you are in proper position the Positioning symbol lights on the display.
The monitor is in proper position.
After the Positioning symbol lights for more than 2 seconds, the wrist cuff will start to inflate automatically.
Cuff Wrapping Guide
The Cuff Wrapping Guide is a unique feature that indicates if the cuff is not wrapped tightly enough around the wrist. Even when the blood pressure reading will be taken.
Note: This reading is NOT reliable due to the incorrect
3. Undo the wrist cuff and remove the unit.
4.
Press the START/STOP button to turn off the monitor.
The monitor automatically stores the measurement in its memory. It will automatically turn off after two minute s.
Important:
• Recent research suggests that the following values can be used as a guide to high blood pressure for measurements taken at home.
Systolic Blood Pressure Above 135 mmHg
Diastolic Blood Pressure Ab ove 85 mmHg
This criteria is for home blood pressure measurement.
The monitor is higher than the heart.
wrapping of the cuff. Please wrap the cuff again, taking care to wrap it correctly and take the measurement
again. When the is displayed, the cuff is correctly wrapped tightly enough on the wrist and the reading is accurate and reliable.
The monitor is lower than the heart.
is displayed, a
• Your blood pressure monitor includes an irregular heartbeat feature. Irregular heartbeats can influence the results of the measurement. The irregular heartbeat algorithm automatically determines if the measurement is usable or needs to be repeated. If the measurement results are affected by irregular heartbeats but the result is valid, the result is shown together with the irregular heartbeat symbol ( ). If the irregular heartbeats cause the measurement to be invalid, no result is shown. If the irregular heartbeat symbol ( ) is shown after you have taken a measurement, repeat
the measurement. If the irregular heartbeat symbol ( ) is shown frequently, please make your doctor aware of it.
• If you move during measuremen t, the movement error symbol ( ) will appear on the display. Keep still and repeat the measurement.
Notes:
• Wait 2 - 3 minutes before taking another blood pressure measurement. Waiting between readings allows the arteries to return to the condition prior to taking the blood pressure measurement.
• The blood pressure can differ between the right arm and the left arm, and therefore also the measured blood pressure values can be different. OMRON recommends to always use the same arm for measurement. If the values between the two arms differ substantially, please check with your physician which arm to use for your measurement.
Warning:
Self-diagnosis of measured results and treatment are dangerous. Please follow the instructions of your doctor.
3.4 Setting the Positioning Sensor
1. Press and hold the
button to reset the Positioning sensor. The Positioning sensor is set “on” as default.
2. Press the START/STOP button to store the setti ng.
3.5 Using the Memory Function
The monitor automatically stores the result up to 100 sets for each user (A and B). It can also calculate an average reading based on the measurements from the last three readings taken within 10 minutes. If there are only two readings in memory for that period, the average will be based on two readings. If there is one reading in memory for that period, the average will be based on one reading.
Notes:
• If the memory is full, the monitor will delete the oldest readings.
• When viewing the reading taken without setting the date and time, “-:--” is displayed instead of the date and time.
To View the Average Value
1. Select your User ID.
2. Press the button.
Note: If there are no measurements
results stored in the memory, the screen to the right is displayed.
Alternating date/time display
This week
Morning average
Note: If t here are not enough
measurements taken in a week to calculate the average the display will appear as illustrated.
This week
Evening average
Last week
Morning average
Last week
Evening average
7 weeks ago
Evening average
7 weeks ago
Morning average
First measurement of the morning
Within 10 minutes
11:59
4:00
Measurements of the morning
Up to 3 measurements
Morning Blood Pressure
Measurement
Measurement
Measurement
Measurement
Measurement
Measurement
Sunday Monday Tuesday Wednesday Thursday Friday Saturday
Measurements of the evening
Last measurement of the evening
Within 10 minutes
1:59
19:00
Up to 3 measurements
Evening Blood Pressure
Measurement
Measurement
Measurement
Measurement
Measurement
Measurement
Sunday Monday Tuesday
Wednesday
Thursday Friday Saturday
Note:It flashes to show
which week average value is displayed.
Systolic blood
pressure
above169 mmHg
152 - 168 mmHg
136 - 151 mmHg
119 - 135 mmHg
118 mmHg or less
3 weeks ago 7 weeks ago
2 weeks ago 6 weeks ago
Last week
5 weeks ago
This week
4 weeks ago
This week - 3 weeks ago
This week
morning SYS average
118 mmHg
This week
evening SYS average
140 mmHg
4 weeks ago - 7 weeks ago
7 weeks ago
morning SYS average
112 mmHg
7 weeks ago
evening SYS average
138 mmHg
HEM-6310F-E.book Page 2 Monday, December 10, 2012 4:14 PM
To View the Readings Stored in Memory
1. Pre ss the button, while the average value is
displayed.
The Memory number appears for a second before the pulse rate is displayed. The newest set is numbered “1”.
2. Press the button repeatedly to view the
readings stored in memory.
To View the Weekly Averages
This unit calculates and displays weekly averages for the measurements taken in the morning and evening within 8 weeks for each user.
Note: T he week begins Sunday at 2:00.
1. Sele ct your User ID.
2. Press the button once.
The morning average for the current week “THIS WEEK” appears on the display.
3. Press the button again.
The evening average for the current week “THIS WEEK” appears on the display.
4. Contin ue to press the button to disp lay the
previous weeks.
The unit displays “-1 WEEK” for the previous week to “-7 WEEK” for the oldest set of averages.
5. Press th e START/STOP button to turn off the
monitor.
Morning Weekly Average
This is the average for the measurements taken during the morning (4:00 - 11:59) between Sunday and Saturday. An average for each day is calculated for up to three measurements taken within 10 minutes of the first measurement of the morning.
Evening Weekly Average
This is the average for the measurements taken during the evening (19:00 - 1:59) between Sunday and Saturday. An average for each day is calculated for up to three measurements taken within 10 minutes of the last measurement of the evening.
The Blood Pressure Trend Graph
This unit also displays the Blood Pressure Trend Graph which shows the weekly averages for the Systolic Blood Pressure (SYS) taken in the morning and evening within 8 weeks for each user. The graph is displayed 4 weeks averages so that you can see the continuous blood pressure change.
To Delete All the Values Stored in Memory
The values stored in the memory are deleted by User ID.
When the memory symbol ( ) appears, first press the button. Then while holding it down, press the START/STOP button simultaneously for about 2 - 3 seconds.
First Second
Note: You cannot partially delete the stored readings.
All values for the user you select will be deleted.
4. Troubleshooting and Maintenance
4.1 The Icons and Error Messages
Error Display Cause Remedy
Irregular heartbeats are detected.
Movement during measurement.
The batteries are
Blink
low.
The batteries are
Lit
exhausted.
Wrist cuff not applied correctly.
Movement during measurement.
Wrist cuff not applied correctly, or movement during measurement.
Arm position changed during measurement.
Device error.
Note: T he irregular heartbeat symbol ( ) and the m ovement error
symbol ( ) may also be displayed with error messages.
Remove the wrist cuff. Wait 2 - 3 minutes and then take another measurement. Repeat the steps in section
3.3. If this error continues to appear, contact your doctor.
Carefully read and repeat the steps in section 3.3.
You should replace them with new ones ahead of time. Refer to section 2.1.
You should replace them with new ones at once. Refer to section 2.1.
Apply the wrist cuff correctly. Refer to section 3.1.
Repeat Measurement while remaining still and refraining from talking during the measurement. Refer to section 3.3.
Apply the wrist cuff correctly and repeat measurement while remaining still and refraining from taking during the measurement. Refer to section 3.1 and 3.3.
Remain still until the measurement is complete. Refer to section 3.3.
Contact your OMRON retail outlet or distributor.
4.2 Troubleshooting
Problem Cause Remedy
The wrist cuff is not at heart level.
The cuff is not wrapped snugly
The reading is extremely low (or high).
Wrist cuff pressure does not rise.
Wrist cuff deflates too soon.
The blood pressure is different each time. The reading is extremely low (or high).
The unit loses power during measurement.
Nothing happens when you press the buttons.
Other problems.
4.3 Transferring the Data to the Online
To begin using the software for the first time, please visit www.omron-healthcare.com for the initial set-up instructions.
4.4 Maintenance
To protect your unit from damage, please observe the following:
around the wrist.
The arms and shoulders are tense.
Movement or talking during measurement.
Air is leaking from the wrist cuff.
The wrist cuff is loose.
The batteries are exhausted.
The batteries have been inserted incorrectly.
• Press the START/STOP button and repeat measurement.
• If the problem continues, try replacing the batteries with new ones.
If this still does not solve the problem, contact your OMRON retail outlet or distributor.
Solution
• Do not subject the main unit and the cuff to extreme temperatures, humidity, moisture or direct sunlight.
• Do not disassemble the unit.
• Do not subject the unit to strong shocks or vibrations (for example, dropping the un it on the floor).
• Do not use volatile liquids to clean the main unit.
• Do not wash the cuff or immerse it in water.
• Do not use petrol, thinners or similar solvents to clean the cuff.
• Do not carry out repairs of any kind yourself. If a defect occurs, consult your OMRON retail out let or distributor as mentioned on the packaging.
• The unit should be cleaned with a soft, dry cloth.
• Use a soft, moistened cloth and neutral soap to clean the cuff.
• Keep the unit in its storage case when not in use.
• Fold the cuff into the storage case.
Measure while in the correct posture. Refer to section 3.2.
Wrap the cuff correctly. Refer to section 3.1.
Relax and try taking the measurement again. Refer to section 3.3.
Remain still and do not talk during Measurement. Refer to section 3.3.
Consult your OMRON retail outlet or distributor.
Apply the cuff correctly so that it is firmly wrapped around the wrist. Refer to section 3.1.
Blood pressure readings constantly vary with time of day and how relaxed you are. Take several deep breaths and try to remain relaxed before taking a measurement.
Replace the batteries with new ones. Refer to section 2.1.
Insert the batteries with the correct (+/-) polarity. Refer to section 2.1.
Do not store the unit in the following si tuations:
• If the unit is wet.
• Locations exposed to extreme temperatures, humidity, direct sunlight, dust or corrosive vapours.
• Locations exposed to vibrations, shocks or where it will be at an angle.
Calibration and Service
• The accuracy of this blood pressure monitor has been carefully tested and is designed for a long service life.
• It is generally recommended to have the unit inspected every two years to ensure correct functioning and accuracy. Please consult your authorised OMRON dealer or the OMRON Customer Service at the address given on the packaging or attached literature.
5. Technical Data
Product Description Wrist Blood Pressure Monitor
Model OMRON RS8 (HEM-6310F-E)
Display LCD Digital Display
Measurement Method Oscillometric method
Transmission Method NFC-F (conforming to ISO/IEC)
Measurement Range Pressure: 0 to 299 mmHg
Accuracy Pressure: ±3 mmHg
Inflation Automatic inflation by pump
Deflation Automatic rapid deflation
Memory 100 Measurements with date and time for
Power Source 2 x 1.5V (LR03, AAA alkaline batteries)
Battery Life Approx. 300 measurements with new
Applied Part = Type B
Protection Against Electric Shock
Operating temperature/ Humidity
Storage temperature/ Humidity/Air pres sure
Console Weight Approximately 80 g without batteries
Outer Dimensions Approximately 89 (w) mm x 61 (h) mm x
Measurable circumference
Cuff Material Nylon and polyester
Package Content Main unit, storage case, battery set,
Note: S ubject to technical modification without prior notice.
• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device Directive).
• This blood pressure monitor is designed according to the European Standard EN1060, Non-invasive sphygmomanometers Part 1: General Requirements and Part 3: Supplementary requirements for electromechanical blood pressure measuring systems.
• This OMRON product is produced under the strict quality system of OMRON HEALTHCARE Co. Ltd., Japan. The Core component for OMRON blood pressure monitors, which is the Pressure Sensor, is produced in Japan.
Pulse: 40 to 180 beats/min.
Pulse: ±5% of display reading
each user (A and B)
alkaline batteries at a room temperature of 23°C
Internally powered ME equipment
+10 to +40°C / Maximum: 30 to 85% RH
-20 to +60°C / Maximum: 10 to 95% RH / 700 to 1060 hPa
13 (d) mm (without the wrist cuff)
Approximately 13.5 to 21.5 cm
instruction manual, guarantee card, blood pressure pass
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC’s and mobile (cellular) telephones, medical devices in use may be susceptible to electromagnetic interference from other devices. Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation. Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the aim to prevent unsafe product situations, the EN60601-1-2:2007 standard has been implemented. This standard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this EN60601-1-2:2007 standard for both immunity and emissions. Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate strong electrical or electromagnetic fields, near the medical device. This may result in incorrect operation of the unit and create a potentially unsafe situation. Recommendation is to keep a minimum distance of 7 m. Verify correct operation of the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is available at OMRON HEALTHCARE EUROPE at the address mentioned in this instruction manual. Documentation is also available at www.omron-healthcare.com.
Correct Disposal of This Product (Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates that it should not be disposed of, with other household wastes at the end of its working life. To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate this product from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this product, or their local government office, for details of where and how they can return this item for environmentally safe recycling. Business users should contact their supplier and check the terms and conditions of the purchase contract. This product should not be mixed with other commercial wastes for disposal.
Manufacturer
EU-representative
Production Facility
Subsidiary
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp THE NETHERLANDS www.omron-healthcare.com
OMRON (DALIAN) CO., LTD.
Dalian, CHINA
OMRON HEALTHCARE UK LTD.
Opal Drive, Fox Milne, Milton Keynes, MK15 0DG U.K.
OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH
John-Deere-Str. 81a 68163 Mannheim, GERMANY www.omron-medizintechnik.de
OMRON SANTÉ FRANCE SAS
14, rue de Lisbonne, 93561 Rosny-sous-B ois Cedex, FRANCE
Made in China