Omron NL-200, NL-220, NL-220F User Manual

Page 1
NL-200, NL-220, NL-220F
LED Lighting Controller
User Manual
84-100022 Rev A
Page 2
NL-200 Series Controller User Manual
1-1
Except as prohibited by law:
• All hardware, software and documentation is provided on an “as is” basis.
• It is essential that the user ensures that the operation of the product is suitable for their application.
• The user must ensure that incorrect functioning of this equipment cannot cause any dangerous situation or signicant nancial loss to occur.
• Gardasoft Vision Ltd and Gardasoft Products Ltd will not accept any liability for consequential loss of any kind.
All trademarks acknowledged. Hardware, software and documentation are Copyright 2002 - 2009 Gardasoft Products Ltd. Hardware manufactured by Gardasoft Vision Ltd under licence.
Latest Manual Version
For the latest version of this manual, see the Download Center on our web site at:
www.microscan.com.
Technical Support
For technical support, email: helpdesk@microscan.com.
Omron Microscan Systems, Inc.
Renton Headquarters
Tel: 425.226.5700 / 800.251.7711 Fax: 425.226.8250
Nashua Ofce
Tel: 603.598.8400 Fax: 603.577.5818
Omron Microscan Europe
Tel: 011 31 172 423360 Fax: 011 31 172 423366
Omron Microscan Asia Pacific
Tel: 65 6846 1214 Fax: 65 6846 4641
Page 3
NL-200 Series Controller User Manual
2-1
2 Getting Started
The NL-200 current controller provides repeatable intensity control of LED lighting for machine vision applications. It includes the intensity control, timing and triggering functions required for machine vi­sion systems.
LED lighting needs a constant current supply as small variations in voltage can cause large variations in light output. The NL-200 can set currents in steps of 0.1% (with a lower limit of 2.5mA steps to give very ne control of intensity
.
Before you start consult your NERLITE specification or product labeling for it’s Current Rating. All specifications can be found on the Omron Microscan website.
NOTE: This current rating is critical to the safe operation of the NERLITE, care must be taken
to determine the correct rating for the light to be connected to the NL-2XX controller.
NOTE: Lighting must not be connected to anything else and must not have a common con­ nection to other lights
NOTE: The NL series has SafePowerTM low heat technology so that in most cases
heatsinking is not required. SafePower also automatically “steps-up” the output volt age, so that from a 24V supply a higher voltage (eg 36V) can be output.
1.) Read the sections on Safety (Section 3) and Specications (Appendix A) and check the
NL-2XX fulls your requirements.
2.) Mount the NL-2XX as described in Section 4.
3.) Connect the NL-2XX up to a supply as described in Connections (Section 5). When the front
panel is present, the NL-2XX should show two alternating lines on the display to indicate that it is operating properly.
4.) Read Lighting Setup (Section 7) and then enter the current rating for the lights you need to
connect. Then connect the lights. NOTE: The current release of the NL-2XX does not sup
port Voltage Rating, use only Current Rating to operate your light(s).
5.) Set up the NL-2XX for the desired operation as described in the Conguration sections
(Section 8, 10 or 11).
Page 4
2-2
NL-200 Series Controller User Manual
Getting Started
2.1 Summary of Features
Throughout this manual, references to the NL-2XX refer to all variants in the NL range unless other­wise stated.
The convention for the part number is:
NL-cd0-vv
NL-cd0F-vv
where:
NL product range name
c Number of channels: 2, 8.
d Conguration option: 0 = front panel, 2 = Ethernet
F Option for fast pulsing
vv Maximum current rating in amps: 2, 20
The following table lists the features on each model.
Feature NL200 NL220 NL220F
Number of channels 2 2 2
Front panel conguration Yes Yes Yes
Ethernet conguration No Yes Yes
Fast pulsing No No Yes
Page 5
NL-200 Series Controller User Manual
3-1
3 Safety
Read this before using the NL-2XX. Always observe the following safety precautions. If in doubt, contact your distributor or Omron Microscan Systems. The following symbols mean:
Warning: read instructions to understand possible hazard
Warning: Possible hazardous voltage
Warning. Surface may get hot
Where these symbols appear in the manual, refer to the text for precautions to be taken.
3.1 Heat
The NL-2XX can dissipate up to 10W and so can get hot. It should be positioned where personnel
cannot accidentally touch it and away from ammable materials.
Read the Mounting (Section 4). Do not exceed the power ratings given in the manual. Note that at the maximum ratings the case temperature can reach 65°C.
Ensure that any ventilation holes are not blocked. Allow free ow of air around the unit.
3.2 Electrical
The lighting connections can exceed 46.7V but should not exceed 70V. Pulse peak voltages above 46.7V are considered hazardous. The lighting connections must be shielded from being touched along the whole length of the cable and in the light.
The user must ensure that the potential difference between any combination of input signals does not exceed 46.7V. WARNING: Higher voltages may cause a danger to personal health.
The NL-2XX does not have complete tracking isolation of inputs and outputs.
Transients caused by inductive loads must be suppressed external to the NL2XX.
The NL-2XX outputs high energy pulses. Care must be taken to connect the outputs correctly and protect the output wiring and load from inadvertent short-circuits. When switched off, there is still energy stored in the NL-2XX for about 15 seconds.
3.3 General
The NL-2XX must not be used in an application where its failure could cause a danger to personal health or damage to other equipment.
If the equipment is used in a manner not specied by the manufacturer, the protection provided by the equipment may be impaired.
Page 6
NL-200 Series Controller User Manual
4-1
4 Mounting
In order to provide xing points to mount the unit onto a at surface or bracket M3 nuts must be inserted into one or more of the slots in the base, see illustration below. The quantity and position of these nuts is dependant upon the users requirements. To t a nut remove one of the end covers. The nuts can then be slid into the slots and the cover replaced. Ensure that the xing screws used do not extend past the lower base surface by more than 5.5mm.
See Addendum for installation of optional DIN Rail Mounting Kit.
4.1 Environmental considerations
The NL-2XX enclosure is a re enclosure as long as it is mounted so that none of the connectors are facing downwards.
If a re enclosure is used, the enclosure should be metal or plastic (with a ammability rating of UL94 V1 or better); with no holes below or to the sides of the NL-2XX when mounted. Cable entries below the NL-2XX should be via glands that have a ammability rating as before. The NL-2XX should be at least 10mm from any other part or side of the enclosure.
The NL-2XX does not have an IP rating and should be mounted so that moisture and dirt cannot enter the unit.
Page 7
NL-200 Series Controller User Manual
5-1
5 Connections
See the Specication (Appendix A) for information on connection ratings.
5.1 Power Supply
To avoid a re hazard from the NL-2XX or the power supply consider the implica­tions of overheating in the unlikely event of a fault in the NL200. The power dissipa­tion in the NL-2XX in a fault condition can be up to:
<Power supply voltage> * <max current delivered by power supply>
Either limit the power supply output current so that not more than 30W can be dissipated in the NL200, or mount the unit in a re enclosure.
Choose a PSU that limits its output current by design, by setting the current limit on the supply (if this feature exists) or use fuses. Remember to derate the fuse, if mounted in an enclosure, as the temperature will be higher than ambient. The external power supply will need to be able to supply at least the average output power for all active channels.
The use of a regulated power supply with 100% short circuit protection is recommended. If however a non-regulated power supply is used, then the maximum ripple voltage of this power supply must not exceed 10% of the actual DC value.
Route low voltage and mains wiring separately. If they must be loomed together ensure that low volt­age insulation rating is sufcient or that supplementary insulation is used.
Pin Power Input Connector
1 +24V to +48V
2 GND
5.2 Lighting Output
The lighting connections can exceed 46.7V but should not exceed 70V DC. Pulse peak voltages above 46.7V are considered hazardous. The lighting connections must be shielded from being touched along the whole length of the cable and in the light.
Make sure you set the current rating for a light before connecting it. See the Lighting Setup (Section 7) for details on this.
Light output is on 4-way pluggable screw terminal sockets. It is possible to use two 2-way connectors in a 4-way socket.
The lighting output connections must not be commoned or grounded in any way.
Page 8
5-2
NL-200 Series Controller User Manual
Connections
Pin LED12
1 CH1 lighting
out +ve
2 CH1 lighting
out -ve
3 CH2 lighting
out +ve
4 CH2 lighting
out -ve
5.3 Trigger Inputs
There is one trigger input per channel.
The trigger inputs are opto-isolated 3V to 24V input, drawing a minimum of 3mA.
Pin ET12 connector
1 TRIG1 -ve
2 TRIG1 +ve
3 TRIG2 -ve
4 TRIG2 +ve
5.4 12V Power Output
One or two 12V power supply outputs may be present. These can supply up to 1A at 12V for power­ing external cameras and other devices. Do not connect inductive loads or devices that take large peak currents. Do not exceed the current rating as these outputs are not fused.
Pin PW1 connector PW2 connector
1 GND GND
2 +12V +12V
5.4.1 Ethernet Connection (NL220, NL-220F)
The RJ45 Ethernet connector requires a straight through cable to connect into a network switch, hub or router. It runs at 10Mbits per second.
Page 9
NL-200 Series Controller User Manual
5-3
Connections
5.4.2 Connectors
The NL-2XX is supplied with mating connectors for the power supply input, trigger inputs, 12V power output and lighting output.
Should spare parts be required these can be obtained as follows:
Connector Description Wuerth part Farnell part
number number www.we-online.com www.farnell.com
Power Wuerth 351 series 691-351­Input 2W screw terminal free skt 500-002 164-1952
Trigger Wuerth 361 series 691-361- 1642-001 Input 4W screw terminal free socket 100-004
12V Wuerth 348 series 691-348- N/A Output 2W screw terminal free plug 500-002
Lighting Wuerth 348 series 691-348- N/A Output 4W screw terminal free plug 500-004
Page 10
NL-200 Series Controller User Manual
6-1
6 General Description
The NL-2XX current controller provides repeatable intensity control of LED lighting for machine vision applications. It includes the intensity control, timing and triggering functions required for machine vi­sion systems.
LED lighting needs a constant current supply as small variations in voltage can cause large variations in light output. The NL-2XX can set currents in steps of 0.1% (with a lower limit of 2.5mA steps) to give very ne control of intensity.
Four modes of operation are provided separately for each channel:
Continuous (“SCo”):
In continuous mode the output is a continuous brightness.
Pulse (Strobe) (SPu”):
In pulse mode output is pulsed once per trigger. One trigger input is used as a trigger. The delay from trigger to pulse, the pulse duration and the brightness can be set.
Switched (“SOn”):
In switched mode a trigger input can be used to switch the output current on and off. The output is only enabled when the trigger input has a voltage on it.
Selected (“SSe”):
In selected mode a trigger input can be used to select between two different intensities..
The NL-2XX is set up using the push buttons and display on the front of the unit. The set up is non­volatile, so the NL-2XX will resume the same operation after a power cycle.
6.1 Output Modes
The trigger inputs are used as follows:
Mode Trigger Input Output
Continuous Unused Output is on.
Switched Trigger = off Output is off
Trigger = on Output is on
Selected Trigger = off Output is continuous brightness 2
Trigger = on Output is continuous brightness 1
Pulsed Trigger rising edge Pulse is triggered if P ag = 1
Trigger falling edge Pulse is triggered if P ag = 0
Note that the P ag inverts the sense of the trigger input.
Page 11
6.1.1 Continuous Output, Switched Output, Selected Output
In continuous mode the output current Is xed and continuous. Switched mode uses a trigger input to switch the output on or off. Selected mode uses a trigger input to select between two different brightnesses.
The output current can be varied from 0% to 100% of full brightness.
6.1.2 Pulsed Output
The output is off by default. When the NL-2XX is triggered it will wait for a delay and then pulse the output. The delay, pulse width, retrigger delay and pulse intensity are all congurable.
Retrigger delay is the minimum allowed time from one trigger to the next. Any triggers that happen too soon after the previous trigger are ignored. The retrigger delay is set in multiples of 100us.
In pulsed mode, the brightness can be set up to 1000% of its rating, but only for short periods and at low duty cycles, so that the lighting does not overheat and get damaged. The duty cycle is limited by ignoring triggers which are too soon after the previous trigger.
Output Brightness Allowed Pulse Width Allowed Duty Cycle
0 to 100% 999ms 100%
101% to 200% 30ms 30%
201% to 300% 10ms 20%
301% to 500% 2ms 10%
501% to 1000% 1ms 5%
So for example, if the brightness is set to 250%, then the NL-2XX will not allow pulses greater than 10ms long. With 10ms pulses, if a trigger occurs within 50ms of a previous trigger (so that the duty cycle would be greater than 20%) the trigger is ignored.
If necessary the NL-2XX will limit the duty cycle by increasing the retrigger delay.
WARNING: To prevent possible overcurrent condition when operating NERLITE Strobe models, initiate the
brightness setting to 10%, then increase current appropriately.
6.1.3 Fault Detection
The NL-2XX detects the following errors. When the output current is less than 100mA, some fault detection is turned off.
Error Reason
PO Internal power dissipation is too high. Output turned off.
OP Output current to lighting is too low. The light is open circuit or there is not
enough supply voltage for the requested output current.
SH If the output voltage is too low, the controller detects that the output is short circuited.
HI The voltage required for the lighting has increased too much. Check for ageing of the
lighting or a failed LED.
LO The voltage required for the lighting has decreased too much. Check for ageing of the
lighting or a failed LED.
The user must press SEL to cancel the error and the NL-2XX will then re-sense the light.
6-2
NL-200 Series Controller User Manual
General Description
Page 12
NL-200 Series Controller User Manual
6-3
General Description
6.2 Cold Start
The NL-2XX conguration can be cleared to the default settings, which clears the lighting ratings and sets all channels to 50% brightness continuous operation. This can be done from the front panel or by sending the CL command using Ethernet.
To clear the conguration from the front panel, turn on the NL-2XX while holding the SEL and DOWN buttons, for about 5 seconds, until ‘COL’ is displayed.. The controller can be cold booted even when the keypad is locked. The lock is removed by the cold boot.
Page 13
NL-200 Series Controller User Manual
7-1
7 Lighting Setup
Lighting can be labelled with either a voltage or current rating or both. All NERLITES will have a current rating, consult specications or product label. This will be used to setup the rating for the light. This rating is the supply to the lighting that should be used to get 100% continuous brightness from the light.
WARNING!
When connecting a strobe light to an NL-2XX Series Lighting Controller, you must set the current rat­ing to 10% of the current specied for the light in its conguration guide.
The NL-2XX Series Controller allows the operator to set the brightness (current) to 1000% in strobe mode. By setting the initial current rating to 10% of the light’s specied current, a brightness setting of 1000% results in the light receiving 100% of its rated current. This will provide maximum light output without damaging the light.
Note: Certain lights require both channels of the NL-2XX Series Lighting Controller. Channel 1 and Channel 2 may have different current specications on some models. Be sure each channel is set correctly as specied in the light’s conguration guide.
The NL-2XX is compatible with both current and voltage rated lighting.
Before connecting a light the rating of the light must be entered. If a light is replaced with a different type of light, then the rating must be set rst. If a light is replaced with the same type of light then the previous rating will still apply.
Consult the specication or labelling for the light.
The current rating can be set from 0.01A to 4A in steps of 0.01A.
Voltage and current rated lights are both driven with a constant current. This gives better brightness stability and allows the NL-2XX to prevent the light being driven with too much power.
7.1 Setting the Rating for a Light
To set the rating of a light on the Ethernet versions of the NL2XX, use the VL command or the internal webpages.
To set the rating of a light from the front panel, press and hold the SEL button for 1 second. The display will show “CH1”. Press DOWN to show “rAt” then press SEL. Use UP and DOWN to select the channel to be set up. Then press SEL and select “Cur” to enter the rated current for each channel. Then press SEL and set the current. Press SEL to save the rating.
Page 14
7-2
NL-200 Series Controller User Manual
Lighting Setup
7.2 Light Auto-Sensing
When a channel does not have a light connected, the NL-2XX continually tries to put out a very small current.
When a light is connected, it will ash for a short time (the light will not be damaged by this) until the NL-2XX detects that it is connected. For voltage rated lights the light is briey driven at an increasing current until 100% output is achieved in order to sense the current rating of the light.
Note: Voltage Rating is not a suported feature.
Only use Current “Cur” to enter the rating.
Page 15
NL-200 Series Controller User Manual
8-1
8 Front Panel
Conguration
(NL-200, NL-220, NL-220F)
8.1 Startup
For controllers with a front panel, On power up, the NL­2XX will display “b r” for 5 seconds, then ‘8.8.8.’ to test the display is working, then ‘NL2’, then ‘00’, followed by the rmware version number, eg ‘001’, and then will be ready for operation. To show that the unit is operating normally, an alternating pattern is drawn on the display. Once a light is connected, the controller will sense it, as described below in Lighting Setup.
Note, if light(s) are connected and have no rating specied for them, then an error is indicated. To set the rating refer to section 7.1.
The overall structure of conguration is given to the right.
To congure the controller from the keypad, press and hold SEL for 1 second. ‘CH1’ will be displayed. Use UP and DOWN to select which feature to set up.
Pressing and holding MODE at any time will cancel the operation.
Using the keypad it is possible to set the conguration for each channel, set a keylock code so that unauthorised users cannot change any settings, set the internal trigger timer, view trigger status and set the voltage or current rat­ing of each of the channels.
8.2 Setting the Lighting Rating
This is described in Lighting Setup (Section 7).
Page 16
8-2
NL-200 Series Controller User Manual
Front Panel Conguration (NL-200, NL-220, NL-220F)
8.3 Default Display
When the NL-2XX is not being congured the display shows the status of all the channels. The seg­ments and their meanings are given below.
Errors are usually caused by a light being connected but the rating for that light has not been set. The rating of a light can be set as described in Section 7.1.
8.4 Setting Flags
The polarity of the trigger signal (P ag) can be set for some output modes and error checking (E ag) can be disabled. During conguration the user will be allowed to select between the following:
F E Error checking enabled
F - E Error checking disabled
F P Trigger is active high
F - P Trigger is active low
Page 17
NL-200 Series Controller User Manual
8-3
Front Panel Conguration (NL-200, NL-220, NL-220F)
8.5 Setting Up Continuous Output
Continuous output is set up as follows. Select “SCo” for continuous operation.
Page 18
8-4
NL-200 Series Controller User Manual
Front Panel Conguration (NL-200, NL-220, NL-220F)
8.6 Setting Up Switched Output
Switched output is set up as follows. Select “SOn” for switched operation.
Page 19
NL-200 Series Controller User Manual
8-5
Front Panel Conguration (NL-200, NL-220, NL-220F)
8.7 Setting Up Selected Output
Select “SSe” mode then enter the two brightness settings.
Page 20
8-6
NL-200 Series Controller User Manual
Front Panel Conguration (NL-200, NL-220, NL-220F)
8.8 Setting Up Pulsed Operation
Pulsed operation can be set up as follows. If a brightness greater than 100% is selected, the pulse width will be limited to a safe value as in the table in Section 6.1.2.
WARNING: To prevent possible overcurrent condition when operating NERLITE Strobe models, initiate the
brightness setting to 10% , then increase current appropriately.
Page 21
NL-200 Series Controller User Manual
8-7
Front Panel Conguration (NL-200, NL-220, NL-220F)
8.9 Setting Pulse Delay and Width Times
When the NL-2XX displays numeric values for the user to change, the right hand digit will be ashing to indicate that the Up and Down buttons can be used to change the value.
To be able to set pulse delay and pulse width values a scheme is used where the exponent (power of ten) of the value is set. The exponent values are as follows:
Exp Multiplier Number Range of values value format
E-2 0.01 9.99 Values are displayed in seconds from 10ms to 5
seconds in steps of 10ms.
E-3 0.001 999. Values are displayed in milliseconds from 1ms to
999ms in steps of 1ms.
E-4 0.0001 99.9 Values are displayed in milliseconds from 0.1ms
to 99.9ms in steps of 0.1ms.
E-5 0.00001 9.99 Values are displayed in milliseconds from
0.01ms to 9.99ms in steps of 0.01ms
E-6 0.000001 999. Values are displayed in microseconds from
1us to 999us in steps of 1us
The ow diagram for entering timings on the keypad is given below.
When a light is pulsed, the display shows that a trigger has occurred by showing ‘PUL’ on the display.
Page 22
8-8
NL-200 Series Controller User Manual
Front Panel Conguration (NL-200, NL-220, NL-220F)
8.10 Setting the Internal Trigger Timer
An internal timer is available for triggering channels in pulse mode. When this timer is turned on, all channels in pulse mode will trigger on this timer. The delay and pulse width will be the same as for external triggers. External triggers will also work.
The period of the timer (the time between triggers) is set as follows.
8.11 Key Lock
The keypad can be locked so that unauthorised users cannot change the conguration. This can be by setting a lock code. Users who know the lock code can unlock the keypad. The lock code is from 0 to 255, providing moderate protection. The lock code is entered by using the UP/DOWN buttons to change this number.
Page 23
NL-200 Series Controller User Manual
8-9
Front Panel Conguration (NL-200, NL-220, NL-220F)
8.12 Viewing Trigger Status
When a channel is in pulse mode, it is possible to view the trigger input status and whether the light is pulsing. To enter this mode, press and hold SEL for 1 second. Then use UP/DOWN to select “trg” then press SEL. The display below will be shown. Press SEL to cancel this display.
8.13 View Output Current
An approximate measure of the output current can be viewed from the keypad. The current is up­dated approximately every second, but can be slower for pulse mode.
Page 24
NL-200 Series Controller User Manual
9-1
9 Ethernet
Communication
(NL-220, NL-220F)
You may need to ask your network administrator for advice about setting up the Ethernet connection.
Ethernet set up is not affected by cold booting the NL220.
9.1 Connection
The Ethernet link uses a 10 base-T connection on an RJ45 connector. The NL-220 Series controller can be connected to a network switch (or hub or router). It is also possible to connect it direct into the network port on a PC by using a crossover cable.
9.2 IP Address
The NL-220 Series controller needs an IP address to communicate over Ethernet. There are two ways to get an IP address; either programmed into the unit or using DHCP.
Most networks use a DHCP server. If there is a PC on the network, You may be able to nd out whether a PC on the same network uses DCHP as follows:
• Go to Control Panel
• Select Network Connections
• Right click on Local Area Connection. Select Properties
• From the list, select Internet Protocol (TCP/IP), press Properties
If “Obtain an IP address automatically” is set, then DHCP is probably used. However, there may be an alternative xed IP address on the “Alternative Conguration” tab.
You can nd out what IP address is being used by a PC at any time by:
• Go to Control Panel
• Select Network Connections
• Right click on Local Area Connection. Select Status
• Select the Support tab. The IP address is displayed
When using a xed IP address, you must ensure that you use an IP address that is not being used by any other device on the network. It is usual to keep the rst three numbers of the IP address the same as other devices and to change only the last number. For example, if you have a network consisting of a PC (IP address 192.168.1.35) and two NL200s, you might give them addresses 192.168.1.201 and 192.168.1.202.
Page 25
9-2
NL-200 Series Controller User Manual
Ethernet Communication (NL-220, NL-220F)
9.2.1 Programmed IP Address and DHCP
For DHCP mode, the NL-220 Series acquires its IP address, subnet mask and gateway address from a DHCP server. Otherwise the NL-220 Series has a xed IP address, subnet mask and gateway
address.
DHCP mode or the IP address can be set and read via the NL-2XX Series Controller Configuration Program available for download at www.microscan.com.
9.2.2 Automatic Sensing
All the features below are implemented in the NL Series Lighting Controller Configuration Program from www.microscan.com.
The NL-220 Series will send out a message on three events:
• On power up
• When an IP address is received or renewed by DHCP
• When an enquiry message is received
On the rst two events, the message is broadcast. On the third it is a reply to a single IP address.
An enquiry message is a UDP packet from source port 30310, destination port 30311 with the mes-sage body “
Omron Microscan Search” (8-bit ASCII, 13 characters).
The message output by the NL-220 Series is a UDP packet from source port 30311, destination port
30310. It is formatted as:
Omron Microscan, NL220,000000,111111111111,22222222
(8-bit ASCII, 44 characters, where
000000 the serial number of the unit
111111111111 the MAC address in 6 HEX bytes
22222222 the IP address in 4 HEX bytes
For example for NL-220 Series serial number 12345, IP address 192.168.1.103, MAC address
00.0B.75.01.80.99 the packet will contain
Omron Microscan, NL220,012345,000B75018099,C0A80167
Page 26
NL-200 Series Controller User Manual
10-1
10 Web Page
Configuration (NL-220, NL-220F)
The NL-220 Series has a webserver inside, so that it can be congured from a standard web browser, such as Internet Explorer.
The IP address of the NL-220 Series must be known (see section 9 on Ethernet Communication. Open a web browser window and type the IP address (for example 192.168.1.71) of the NL-220 Se­ries into the URL box at the top. The main page of the NL-220 Series webserver should be shown.
10.1 Main Page
The main page shows general information about the NL220. Links are provided to the conguration pages.
10.2 General Setup Page
The General Conguration page allows the webpage protection password to be set or cleared and the internal trigger to be set up. Also any Ethernet command from Section 11 can be entered.
“Test Mode” referred to on this page is the internal trigger timer.
Page 27
10.3 Channel Conguration Pages
There is one Channel Conguration Page for each output channel. All the parameters for each output channel can be set up. Press the Submit button to update the NL-220 Series and save the changes to non-volatile memory.
The current rating for the light can be changed. Use this with care.
Some measured voltages and the actual output current are displayed on this page.
10-2
NL-200 Series Controller User Manual
Webpage Conguration (NL-220, NL-220F)
Page 28
NL-200 Series Controller User Manual
11-1
11 Command
Conguration
(NL-220, NL-220F)
The NL-2XX can be congured via the Ethernet connection using UDP or TCP/IP. A Conguration Program with source code can be downloaded from www.gardasoft.com.
11.1 Ethernet Communication (NL-220, NL-220F)
For TCP, commands from a host should be sent to destination port 30313. Replies will be to destina­tion port 30312. For UDP, commands from a host should be sent from source port 30312 to destina­tion port 30313. Replies will be sent from source port 30313 to destination port 30312.
A TCP/IP connection will timeout and close if it is idle for more than 10 seconds. The host must send regular “heartbeat” commands (eg “VR”) to keep the link open.
11.2 Command Structure
Communication consists of commands sent by the host (controlling PC). All output generated by the command is returned in reply UDP or TCP/IP packets. The last character sent is “>” (“greater than” symbol). Once this is received, the host knows that the command has been completed.
It is recommended that the host waits for the “>” symbol before sending the next command. UDP communications are not guaranteed to arrive, so the host software must be able to cope with lost messages.
Using the GT command, a host can request that a message is sent to it whenever an error occurs.
Several commands can be put into one command line by separating them by a semi-colon (“;”). A car­riage return character should be sent to terminate the command line. The NL-220 will send any replies to the commands and then send a ‘>’ character to indicate that the command line has been completed.
Commands comprise a code of two letters followed by the parameters (if any) needed for the com­mand. Spaces in the commands are ignored.
Numeric parameters are separated by a comma (“,”). For a parameter which is a time period the default units are milliseconds. “s”, “ms” or “us” can be added to the end of the number to indicate seconds, milliseconds or microseconds. For currents, “a” or “ma” can be added to indicate “amps” or milliamps”. The default is amps.
Note that parameters are in “USA/UK” format so that a half is written “0.5” not “0,5”
Page 29
11-2
NL-200 Series Controller User Manual
Command Conguration (NL-220, NL-220F)
For example:
Parameter Meaning
0.1 0.1 milliseconds
200us 200 microseconds
0.1s 0.1 seconds
100ma 100mA
2.45A 2.45A
2.3 2300mA or 2.3A
The command codes and their meaning are described below. The upper case commands are shown, followed by lower case letters denoting the numeric argument.
Error number Reason
Err 1 A parameter value is invalid
Err 2 Command not recognised
Err 3 Numeric value is wrong format
Err 4 Wrong number of parameters
Err 5 (This is only a warning) A timing parameter was
out of range and has been adjusted to a valid value.
Any changes made using Ethernet commands are not saved permanently until the AW command has been issued.
11.2.1 General Commands
Report the version of rmware running in the NL220F VR
This command returns the rmware version. For example: NL-220F Series (HW001) V002
Set the rating of a light
This command sets the current or voltage rating for a light. If a current rating is being set, then the voltage rating value should be 0.
VLo,v,c
Where:
o = output channel (1 or 2) v = voltage rating (NOT SUPPORTED Set to 0) c = current rating (0 or 10mA to 4A)
Page 30
NL-200 Series Controller User Manual
11-3
Command Conguration (NL-220, NL-220F)
Set continuous mode
The output is set to continuous mode at a percentage of full brightness. RSc,s Where:
c = output channel (1 or 2) s = setting in percent (s = 0 to 100)
Set switched mode
The output is set to switched mode at a percentage of full brightness. RWc,s Where:
c = output channel (1 or 2) s = setting in percent (s = 0 to 100)
Set selected mode
The output is set to selected mode with two brightness settings. RUc,s,t Where:
c = output channel (1 or 2) s = brightness 1 setting in percent (s = 0 to 100) t = brightness 2 setting in percent (t = 0 to s)
Set pulse mode
The output can be set up to pulse on a trigger input. The delay from trigger to the start of the pulse, the length of the pulse and the brightness are congurable.
An error is generated if the brightness setting requires a current greater than 20A or if the combination of pulse width and setting is not allowed.
RTc,p,d,s RTc,p,d,s,r
Where:
c = output channel (1 or 2)
p = pulse width in milliseconds (0.02 to 999) d = delay from trigger to pulse in milliseconds (0.02 to 999) s = setting in percent (s = 0 to 999) r = retrigger delay. This parameter is optional
Set the Option Flags
REc,m
Where:
c = output channel (1 or 2) m = ags: bit 1 = 0 E ag set (error detection enabled)
1 E ag cleared (error detection disabled)
bit 2 = 0 P ag set (positive triggers)
1 P ag cleared (negative triggers)
Page 31
11-4
NL-200 Series Controller User Manual
Command Conguration (NL-220, NL-220F)
Set the Trigger Input
This command sets which input is used for pulse and switch output modes.
RPc,p
Where:
c = output channel (1 or 2) p = trigger input (1 or 2)
Set Internal Trigger
Enable or disable the internal trigger. When enabled, all outputs are triggered simultaneously using an internal trigger signal. This setting can be saved to non-volatile memory using the AW command.
TT0 Disable internal trigger TT1 Enable internal trigger (uses previously set period) TT1,p Enable internal trigger and set the period
Where:
p= period of the triggers in microseconds
For example:
TT1,200 Set the internal trigger to 200ms (5Hz) TT1,500US Set the internal trigger to 500us (2KHz)
Save the settings to memory.
AW
Once the settings are saved to memory they are then retained when the unit is switched off. If this is not done, changes to the settings are volatile, and if the unit is switched off they revert to those in force when the last AW command was issued.
Simulate an Input Trigger
TRc
c which input channel (1 or 2) Simulates a trigger pulse. If the channel is in pulse mode it will pulse and show “PUL” on the display.
Enable Ethernet Messages
GTm
m = 0 to disable Ethernet messages
= 1 to enable Ethernet messages When Ethernet messages are enabled, any error reports are sent to the most recent UDP or TCP address from which a command has been received. Messages are of the form:
Evtc,e Where c zero for no channel or channel number (1 or 2)
v event value:
32 to 47 Lighting er or code 128 Light detected and waiting for current rating 129 Light detected and not waiting for current rating
Page 32
NL-200 Series Controller User Manual
11-5
Command Conguration (NL-220, NL-220F)
Clear any Errors
GR
If Ethernet messages are not enabled, the last event or error number can be read by this command. Any error displayed on the unit is cleared, so if there was a lighting error, the NL­2XX will resume auto-sensing on that channel.
The reply will be in he same form as the GT command above. If there are no outstanding events or errors, then only the prompt “>” is returned.
Set/Clear the Webpage Password
EY
EY asc1, asc2, asc3, asc4, asc5, asc6
This command sets the password required to access the webpages. If EY is entered on its own then the password is cleared. There are six optional parameters, which are decimal ASCII values for a password from one to six letters. A value of 65 is ‘A’, 66 is ‘B’, etc to 90 is ‘Z’.
Disable Keypad
KBd,c
d = 0 Enable keypad
d = 1 Disable keypad, but allow unlock code
d = 2 Disable keypad, disable unlock code
c Unlock code
In some applications it may be necessary to lock the keypad so that un-authorised operators cannot change the settings.
A unlock code can be set. This can be used as a low-security way of allowing trusted users to unlock the keypad. Ethernet commands and the web pages still work. The setting of this command is restored after a power cycle.
Page 33
11-6
NL-200 Series Controller User Manual
Command Conguration (NL-220, NL-220F)
11.2.2 Command Summary
Command Example Effect
AW AW Save changes
GT GT1 Enable Ethernet messages
GR GR Clear any error condition.
EY EY65,66 Set webpage password to “AB”
VR VR Read the rmware version
VL VL1,0,0.5 Set the rating of channel 1 to 0.5A
RS RS2,65 Set channel 2 to 65% brightness continuous
RW RW1,50 Set channel 1 to 50%, switch mode
RU RU1,75,25 Set channel 1 to selected mode at 75% and 25%
RT RT2,3,4,50 Set channel 2 to 3ms pulses, delayed by 4ms, at 50% brightness
RP RP1,2 Output channel 1 is triggered using input 2
RE RE1,3 Set channel 1 to ignore lighting errors and not prompt for the
current rating of a light when it is connected
TT TT1,1ms Set internal triggers every 1ms
KB KB1,23 Disable the buttons on the front panel. Set the unlock code to 23.
TR TR2 Trigger channel 2
NOTE: If the above commands are executed using the Web Page feature, they will not be executed in the command entry area of the Web Page.
Page 34
NL-200 Series Controller User Manual
A-1
A: Specifications
Output channels Up to 4A continuous or 20A pulsed. Maximum 48V output,
60W average total output power
Lighting Rating 12V to 36V in steps of 1V
100mA to 4A in steps of 10mA
Operating modes Continuous, strobe, switch and selected
Trigger inputs opto-isolated, standard or inverted operation. 3V to 24V
Pulse width 20 microsecond to 1 second. Repeatable to (NL-200, NL-220) 0.1 microsecond
Pulse width 2 microsecond to 1 second. Repeatable to (NL-220F) 0.1 microsecond (See Restrictions, Appendix B)
Trigger delay 20 microseconds to 1 second. Repeatable to (NL-200, NL-220) 2 microseconds
Trigger delay 2 microseconds to 1 second. Repeatable to (NL-220F) 2 microseconds
Internal trigger timer 0.2Hz to 5KHz
Supply voltage 24VDC to 48VDC regulated
Heat Dissipation 2W per channel typical at maximum output
Page 35
NL-200 Series Controller User Manual
B-1
B:
Restrictions
The following timings and restrictions apply for rmware revision V002. The restrictions are applied when settings are saved. Later rmware versions may have lesser limits.
B.1 Continuous Mode
The maximum output current is 4A.
B.2 Switched Mode
The maximum delay from a trigger input changing to the output current being turned on or off is 10us.
The maximum output current is 0.5A.
B.3 Selected Mode
The maximum delay from a trigger input changing to the output current being turned on or off is 5ms.
The maximum output current is 0.5A.
B.4 Pulse Mode
The maximum output current is 20A. For high currents pulses the following limits are recommended. These limits will be improved in future.
Pulses of 2A or more for pulse widths longer than 2ms might cause an error or have a lower current towards the end of the pulse.
The minimum pulse delay is about 2us. When overdriving or using the retrigger delay, the minimum delay is around 5us.
For pulse widths less than approximately 70us the output voltage and current cannot be measured. Because of this, fault detection does not happen and the following restrictions apply:
For pulse currents greater than 0.5A, the duty cycle will be restricted to no more than 1%
For pulse currents less than or equal to 0.5A, the duty cycle will be restricted to no more than 10%
Pulse Current Pulse Length Limit
20A 100us 12A 400us 10A 500us 5A 1ms
Page 36
NL-200 Series Controller User Manual
C-1
C:
Fatal Error Codes
Error number Reason
FAt, Err 44 The NL-200 is too hot. The NL-200 has a thermal cutout which
operates around 65°c to 70°c, depending on conditions.
FAC, Err 40, One channel is outputting more current than expected. Err 41, Err 45
Page 37
NL-200 Series Controller User Manual
D-1
D:
Error Codes
Error number Reason
Err 1 A parameter value is invalid
Err 2 Command not recognised
Err 3 Numeric value is wrong format
Err 4 Wrong number of parameters
Err 5 This is a warning, not an error. One of the parameters is out of range.
The value of the parameter has been adjusted. For example, sending an NL command with a delay of 0 will get a reply of “Err 5”. The command will be accepted and the delay set to the minimum allowed.
Err 19 A light has been connected to a channel but the rating has not been set.
Err 8, 12 EEPROM corrupt. The conguration has been cleared.
Err 9, 20 Couldn’t save settings to EEPROM.
Err 27 Can’t read Ethernet settings from EEPROM, so these may be incorrect.
PO, Err 34, 39 Internal power dissipation is too high. Output turned off.
OP, Err 35 Output current to lighting is too low.
OP, Err 43 There is not enough supply voltage for the requested output current.
SH, Err 36 The output is short circuit.
SH, Err 42 The output current is too high.
HI, Err 37 The voltage required for the lighting has increased too much.
Check for ageing of the lighting or a failed LED.
SH, Err 38 The voltage required for the lighting has decreased too much.
Check for ageing of the lighting or a failed LED.
Any other errors are internal errors.
Page 38
NL-200 Series Controller User Manual
E-1
E
: Event Codes
Event messages are sent when a light is connected or an error occurs. The format of these is
Evt<channel>,<event code>;
These event messages are only sent after the GT1 command has been sent.
Event number Reason
1 to 127 An error has occurred. The error code is given by the event number.
128 A light has been connected and is working.
129 A light has been connected but doesn’t have a current or voltage rating.
130 Over temperature error occurred ( FAT)
131 Over current error occurred (FAC)
132 An error has occurred while autosensing the rating of a light.
Page 39
NL-200 Series Controller User Manual
A
Addendum
Installation – see gure
Remove either end cover by removing the four screws.
The clips shown indicate the two orientations available for mounting.
Select the orientation required, a pair of clips is used for each (AX or AY).
Slide nuts (B) into the appropriate slots.
Replace the end cover.
Screws (C) are passed through mounting clips (A) and into slot nuts (B).
Position the loose assemblies as required and tighten the screws.
DIN Rail Mounting Kit for NL Range
Pack Contents
Description Drawing Ref Quantity
Mounting Clip A 2
M3 Nut B 2
M3x8mm Countersunk screw C 2