Omron Colin VP-1000 plus, Colin BP-203RPE III User Manual

Thank you for purchasing the OMRON BP-203RPE3 unit.
Read all of the instructions in the manual before you operate the unit and keep the manual near the unit at all times for future reference.
Non-invasive Vascular Screening Device
3
1
Contents
1. Before Use
1-1. Exemptions .................................................................................................. 3
1-2. Intended Use................................................................................................ 4
1-3. Meaning of Symbols ................................................................................... 5
1-4. Safety Information....................................................................................... 6
Explanation of Symbols........................................................................... 6
1-5. Product and Accessories .........................................................................13
Main Unit ............................................................................................... 13
Standard Accessories............................................................................ 13
Accessories (Sold Separately) .............................................................. 15
Options .................................................................................................. 16
1-6. Name and Function of Each Part .............................................................17
Main Unit ............................................................................................... 17
Stand ..................................................................................................... 18
1-7. Installation/Moving .................................................................................... 19
Inspection Before Starting Work ............................................................ 19
Procedure for Moving ............................................................................ 20
2. Measurement Procedure
2-1. Preparing for Measurement ...................................................................... 22
Measurement Procedure .......................................................................22
Patient Information ................................................................................ 23
Initial Screen (ID Entry Screen) ............................................................. 24
Entering and Editing Patient Information............................................... 25
Displaying the Measurement History..................................................... 40
Attaching the Cuff and Sensor............................................................... 41
2-2. Basic Measurement................................................................................... 47
Viewing the Measurement Screen......................................................... 47
Contents of the Measurement Screen................................................... 48
Starting and Ending Measurement ........................................................ 51
2-3. Measurement Results ............................................................................... 53
Contents of the Measurement Results Screen...................................... 53
Measurement Results Reports .............................................................. 54
2-4. R-R Interval Examination ..........................................................................65
Starting and Ending R-R Interval Examination ...................................... 65
R-R Interval Examination Results.......................................................... 66
2-5. Stress Mode ............................................................................................... 67
Starting and Ending Stress Test............................................................. 67
Checking Numerical Values During Stress Test .................................... 71
Stress Test Results................................................................................ 73
3. Settings and Data Processing
3-1. Main Menu Screen..................................................................................... 74
3-2. User Default Settings ................................................................................76
Items That Can Be Set ..........................................................................76
User Default Settings Procedure ........................................................... 79
3-3. Print Default Settings................................................................................ 80
Items That Can Be Set .......................................................................... 80
Print Default Settings Procedure ........................................................... 85
2
3-4. Facility name / Doctor / Technician / Category Settings......................... 86
Selecting the List Entry Method............................................................. 86
List Entry Procedure .............................................................................. 87
Pre-selection Setting Procedure............................................................ 91
3-5. Date & Time Settings................................................................................. 93
3-6. Printing Reports and Editing Patient Information.................................. 94
Reprinting Measurement Data............................................................... 95
Editing Patient Information .................................................................... 97
Deleting Measurement Data.................................................................. 99
3-7. Printing a Trend Report .......................................................................... 101
3-8. Advanced Registration of Patient Information ..................................... 105
Selecting the Registration Method....................................................... 105
Registration Procedure........................................................................ 106
Editing Patient Information .................................................................. 109
Deleting Patient Information ................................................................. 111
3-9. Printing Usage Frequency / Facility Patient Reports ........................... 113
Types of Reports ................................................................................. 113
Procedure for Printing Usage Frequency /
Facility Patient Reports......................................................................... 115
3-10. Data Export / Import (USB Flash Drive)................................................. 116
Data Processing Items ......................................................................... 116
Data Processing Procedure................................................................. 117
3-11. Transferring Report Data (PC)................................................................ 120
4. Options
4-1. Options .....................................................................................................121
TBI package ........................................................................................ 121
HMC package...................................................................................... 122
Bar Code Reader Set ..........................................................................123
5. Maintenance
5-1. Routine Maintenance .............................................................................. 124
Maintenance Procedures......................................................................124
Supplies................................................................................................125
5-2. Replacing Cuffs .......................................................................................126
Replacing an Arm Cuff ........................................................................ 126
Replacing an Ankle Cuff ...................................................................... 126
5-3. Connections............................................................................................. 127
Connectors on the Device ................................................................... 127
Basic Connections................................................................................128
5-4. Changing the Arm Position ..................................................................... 129
5-5. Handling Errors ........................................................................................132
Types of Audible Alarms....................................................................... 132
5-6. Displaying System Information............................................................... 133
5-7. Maintenance Menu ................................................................................... 134
5-8. Specifications ...........................................................................................135
5-9. Guidance and Manufacturer's Declaration ............................................139
5-10. Explanation of Technical Terms.............................................................. 143
5-11. Disposal .................................................................................................... 147
3
1. Before Use
1-1. Exemptions
Disclaimer
Principles
Trademark
Our company assumes no responsibility for the following:
1.Failures, damage, or injuries due to maintenance or repair work performed by other than our company or a company that we specify.
2.Failures or damage to one of our products caused by a product of another manufacturer not delivered by us.
3.Failures, damage, or injuries due to maintenance or repair work using a repair part other than a part that we specify.
4.Failures, damage, or injuries caused by failure to observe the safety instructions and operational procedures given in this manual.
5.Use of the product in conditions that do not conform to the product usage conditions indicated in this manual, including power, installation, and storage conditions.
6.Failures, damage, or injuries due to modification or inappropriate repair of the product.
7.Failures, damage, or injuries due to natural calamities such as fire, earthquake, flooding, or lightning.
1.The contents of this manual are subject to change without notice.
2.Considerable care has been taken in the preparation of this manual. In the unlikely event that an error or other problem is discovered in the manual, please contact us.
3.Unauthorized reproduction of all or part of this manual is prohibited. Use other than individual (corporate) use without the permission of our company is prohibited by copyright.
Product brand names shown in this manual are likely to be the trademark or registered trademark of the company concerned.
4
1-2. Intended Use
Medical Purpose
This is a non-invasive diagnostic system designed to assist in the detection of peripheral vascular diseases.
Using Population
Legally certified medical experts, such as doctor, nurse and ME.
Patient Population
It is used on adult patients only.
Environment
The instrument is used in a vascular laboratory, clinic, hospital, doctor's office, and other medical facilities where the non-invasive peripheral vascular test is conducted.
Durable Period
5 years, provided that the appropriate maintenance has been done from production date.
(Self-certification through OMRON HEALTHCARE's own data)
Measurement Parameter
Non-invasive Blood Pressure
Heart Rate
Pulse Wave
Heart Sound
Calculating Parameter
ABI (Ankle Brachial Index)
Pulse Wave Velocity
Augmentation Index
Systolic Time Interval
Upstroke Time
Precautions for use
Warnings and cautions described in the instruction manual should be observed.
5
1-3. Meaning of Symbols
Start (of action) This way up
Stop (of action) Keep away from rain
Caution (Refer to safety information)
Stacking limit by number
Consult instructions for use Temperature limitation
Authorised representative in the European community
Catalogue number
Manufacturer Disposal instructions
Date of manufacture Use by
Do not reuse Batch code
Defibrillation-proof type BF applied part
Fragile; Handle with care
Defibrillation-proof type CF applied part
Gruene Punkt (Green point in German)
CE mark Serial number
6
1-4. Safety Information
The warning signs and symbol examples indicated below are intended to ensure safe use of the product and prevent damage and injury to you and others. The signs and symbols are explained below.
Explanation of Symbols
* Physical damage means serious damage to your house and household goods, and serious injury
to pets or other domestic animals.
Warning
Indicates a situation where incorrect handling may cause human death or serious injury.
Caution
Indicates a situation where incorrect handling may cause human injury or physical damage.*
indicates "mandatory" (an action that must be observed).
The actual action that is mandatory is indicated inside or next to . The icon at left indicates "disconnect the power plug".
indicates "forbidden" (an action that must not be taken).
The actual action that is forbidden is indicated inside or next to
.
The icon at left indicates that "disassembly is forbidden".
Note:
This indicates information that should be known when operating the device.
7
1-4. Safety Information
Warning:
Safety rules when using the product
Do not use a frayed or damaged power cord or plug. Otherwise electric shock, short circuiting, or fire may result.
Do not touch the power plug with wet hands. Otherwise electric shock or injury may result.
Be sure to plug the three-prong power plug into a grounded (three-prong) outlet for medical use (when a printer is included). Otherwise electric shock or current leakage may result.
Use a dedicated outlet. Otherwise electric shock or current leakage may result.
Do not poke or scratch the buttons or display with a sharp or pointed object. Incorrect diagnosis and treatment or an accident may result.
Do not connect the power plug to the outlet if the electrical ratings are outside the specified range below: 100 - 240 VAC. This may cause fire.
Do not touch the unit when discharging a defibrillator.
Safety rules when performing measurement
This device is only to be used by qualified medical personnel, or under the guidance of such personnel. Otherwise incorrect diagnosis and treatment or device failure may result.
The results of measurement should only be interpreted by a doctor. If you are concerned about a measurement result, consult your doctor. Otherwise incorrect diagnosis and treatment may result.
This device is intended to perform measurement for examination. Do not use the device for patient monitoring. Otherwise an accident may result.
If pressurization does not stop during measurement or another abnormal condition occurs, remove the cuff or air tube and disconnect the power. Otherwise peripheral nerve may be damaged.
Use only the specified supplies for the cord, cuff, USB devices, and other parts. Do not install other than specified options. Otherwise an accident may result.
Do not connect the air tube or cuff to any other device tubes attached to the body. Otherwise air may enter the blood vessels and an accident may result.
8
1-4. Safety Information
Do not attach the arm cuff for patient during medical treatment with intravenous drip or blood transfusion. This may result incorrect diagnosis and treatment.
Do not use this device on a patient that cannot express pain. If cuff pressurization does not stop in a specified time, internal hemorrhaging may occur in the upper arm or ankle.
This device cannot be used during MRI imaging. This may cause an accident.
Pay constant attention to patient's condition during measurement. Otherwise an accident may result.
Do not perform measurement without cleaning and sterilizing the device, cuffs and sensors after measuring for patient with infection.
Pay extra attention to a patient with pacemaker during measurement. Rate meters may continue to count the pacemaker rate during occurrence of cardiac arrest or some arrhythmias. Do not rely entirely upon rate meter alarm. Refer to the pacemaker pulse rejection capability in this manual.
Installation
As this device conforms to the IEC60601-1-2 standard, it can be used in combination with another medical device. While or after an electrosurgical knife or a device which generates noise being used near this device however, make sure that it is properly functioning. Otherwise incorrect diagnosis and treatment or device failure may result.
Do not use this device in the presence of a flammable gas such as a highly inflammable anesthetic, or in a high-pressure oxygen chamber or tent. Fire and explosion may result.
Do not install or store the device in a location where water or chemicals may splash on the device. Electric shock may result.
Do not connect any electric device, without the approval of IEC60601-1, or not fulfilling IEC60601-1-1. For the use of a PC or a printer with this device, the connecting device should be approved according to standards mentioned above. (This unit meets the restricted level of leakage current required for medical devices and it does not include all the connected devices. Connecting to other device is not allowed unless the total leakage current of such a combination is within the restricted level.) Neglecting this caution, it could cause electric shock to the device.
Maintenance
After cleaning the device, dry it completely before turning on the power again. Otherwise electric shock or current leakage may result.
9
1-4. Safety Information
Caution:
Safety rules when using the product
Observe the rules below when handling the power cord. Failure to observe these rules may result in electric shock or device failure.
•Do not damage the cord
•Do not break the cord
•Do not modify the cord
•Do not bend or pull on the cord with undue force
•Do not twist the cord
•Do not tightly coil the cord when in use
•Do not place heavy objects on the cord
•Do not let the cord become pinched
Insert the power plug all the way into the outlet. Otherwise electric shock, short circuiting, or fire may result.
Wipe dust off the power plug. Otherwise electric shock, short circuiting, or fire may result.
After using the device, turn the power switch to the "off" position and disconnect the power plug. Failure to do so may cause deterioration of the insulation and result in electric shock, current leakage, or fire.
When removing the power plug from the outlet, grasp and pull on the plug, not the cord. Pulling on the cord may break wires and cause a short circuit, resulting in fire or electric shock.
If a power failure occurs while using the device, turn the power switch to the "off" position and disconnect the power plug. Failure to do so may result in a product failure or other problem.
If the device becomes wet, wipe it completely dry with a soft cloth before use. Otherwise electric shock, short circuiting, or fire may result.
Safety rules when performing measurement
If the patient has any of conditions below, do not perform measurement. Otherwise incorrect diagnosis and treatment may result.
•The patient has an aneurysm
•The patient has insufficient peripheral circulation, noticeably low blood pressure, or low body temperature.
•The patient frequently has an irregular pulse.
In the following situations, check by auscultation or palpation. Otherwise incorrect diagnosis and treatment may result.
•When irregular pulse waves are indicated. External vibration or patient movement during measurement may cause an incorrect indication.
•When an error occurs or when a measured value is questionable.
10
1-4. Safety Information
If a power failure occurs during measurement, immediately remove the cuff. If the patient's ankle or upper arm is pressurized for a long time, internal hemorrhaging may result.
Do not attach the cuff on the measurement site below:
•Arm with intravenous drip
•Upper limb in which a shunt is placed in for hemodialysis Incorrect diagnosis and treatment or an accident may result.
If there is acute inflammation, a pyogenic ailment, or an external wound where the cuff is to be attached, follow the instructions of a doctor. Symptoms may become worse.
When there is impossibility of test or doubts about the measurement values, please confirm the patient's condition first. The patient's condition may have deteriorated to the point where measurement limits are exceeded. Always verify that the cuff and cuff hose are appropriately used and are not bent or blocked. If the display continues to show 0, the monitor's pressure may be 0. But if the cuff hose is blocked or bent there may be air remaining in the cuff. At this time disconnect the hose from the cuff to ensure that blood flow is not restricted and no disorders occur to the peripheral nerves.
When using the device in combination with another medical device, read the manual of the other device well and understand all warnings and cautions. Incorrect diagnosis and treatment may result.
Do not use the device in a location with loud noise. Incorrect diagnosis and treatment may result.
Attach the PCG sensor correctly. Incorrect diagnosis and treatment may result.
Take care that the PCG sensor does not fall on the patient. This may cause injury.
Wrap the cuff on bare skin or on a thin layer of clothing. Otherwise internal hemorrhaging may result.
Make sure that the ECG clips are attached in the correct positions. Otherwise incorrect diagnosis and treatment may result.
Do not use the device near a cellular phone. Incorrect diagnosis and treatment may result.
Do not use in an MRI, CT, X-ray room, an operating room, or other rooms where radio noise is generated. Incorrect diagnosis and treatment or an accident may result.
Do not use a worn or expired ECG clip electrode or PCG sensor pad. Otherwise incorrect diagnosis and treatment may result. (ECG clip electrode and PCG sensor pad have an expiry date. After the expiry date, the pad becomes dry and incapable of accurate measurement. Use only a pad or electrode whose indicated expiry date has not passed. Refer to page 12 for how to confirm the expiry date.)
11
1-4. Safety Information
On a bedridden patient, check for lower-limb deep venous thrombus before taking a measurement. Otherwise an accident may result.
The ECG clip electrode and PCG sensor pad are disposable supplies. Do not reuse them once they are removed. In case that they are used on moist, injured or infected skin, dispose them right away. Otherwise an infection may result.
The sensor box removal lever uses a strong spring mechanism. Take care not to injure yourself on the metal edges. Hold the sensor box firmly and press down hard on the lever from the exterior to remove the connector. This may cause injury with the metal edges.
Do not use an air hose of an arm cuff or an ankle cuff that is bent or collapsed. Incorrect diagnosis and treatment may result.
Installation
Do not install in a location where the temperature or humidity is outside the allowed range. This may cause malfunctioning or device failure.
Do not use in a moving vehicle such as an ambulance. Incorrect measurement may result. Incorrect diagnosis and treatment or an accident may result.
If you are not using the device stand, exercise sufficient caution when moving the device. The device is heavy and may slip out of your hands, causing injury.
Do not place objects on the device. This may cause injury.
Maintenance
Do not disassemble, repair or modify the device. This may cause electric shock.
Before cleaning or maintaining the device, disconnect the power plug. Otherwise electric shock or injury may result.
12
1-4. Safety Information
Important:
Before use
• Make sure that all cords and tubes are firmly connected.
• Inspect dials and buttons and verify that the device operates correctly.
• Check printer toner, paper, and other supplies.
• Confirm the expiry date of the ECG clip electrode or the PCG sensor pad. The expiry date can be calculated from the lot number as follows:
During use
• Do not pressurize when the cuff is not wrapped.
• Do not use a torn cuff.
• The device may be used on the patients below. If used on patients other than those described below, incorrect measurement may result or measurement may not be possible.
- Height: 120 to 210 cm
- Girth of arm: 20 to 32 cm (using the standard cuff) / 16 to 38 cm (using the optional cuff)
- Girth of ankle: 16 to 33 cm
• Do not disconnect the USB or LAN cable while data is being transmitted. This may corrupt the data.
• Do not turn off the power during printing, data transmission, or writing.
• Do not pull out the paper before printing is finished.
• Make sure the date and time setting is correct. If the date and time setting is not correct, the date and time of measurement will not be recorded correctly.
• Do not connect an ECG clip to another conductive part including the ground.
• The electrical energy from a defibrillator may damage the device. During defibrillation, remove the sensors from the patient and do not touch the device.
• When moving or using the device, do not drop or subject the device to shock. This may cause the electric components and precision mechanisms to fail.
After use
• Clean the device and accessories and arrange properly for storage.
• Do not wash or moisten the cuff
• Do not use solvents such as thinner, benzene, or concentrated alcohol to clean the device
• Do not use an autoclave, ultraviolet radiation, or gas disinfection (EOG, formaldehyde gas, concentrated ozone, etc.) to disinfect the device
• If a fuse blows, there may be a problem in the device. Contact a dealer or an Omron Healthcare technical support representative
• Do not install the unit in the following locations
- A location with prolonged exposure to direct sunlight
- A location with dusty or salt air
- On an inclined surface or a location subject to vibration, shock, or noise
- A location where chemicals are stored or gas is emitted
• If the patient has one of the following conditions, a correct measurement may not be obtained
- The patient has body movements due to convulsions caused by rheumatism or otherwise
- The patient has diabetic arteriosclerosis (blood pressure at leg joints tends to be high)
- The patient has false high blood pressure
- The patient has convulsions or tremors
• When the unit is incapable of ECG measurement, the message "Electrode removed" is displayed.
Lot Number:
Production Date: 2006 July 10th
Expiry Date: 2008 July
6071011
(2 years after the production date)
(Example)
13
1-5. Product and Accessories
Before using this product, make sure that no accessories are missing and that neither the unit nor the accessories are damaged. Contact a dealer or an Omron Healthcare technical support representative If any accessory is missing or damaged.
Main Unit
Non-invasive Vascular Screening Device BP-203RPE3
Standard Accessories
BP-203RPE3 stand
BP-203RPE3 stand cover
BP-203RPE3 hook-and-loop fastener for the sensor box
Power cord for main unit (0.8 m)
Power cord for printer (1.1 m)
BP-203RPE3 power tap
USB cable (for printer)
Items on this page will be assembled on delivery as follows:
14
1-5. Product and Accessories
BP-203RPE3 arm cuffs, left and right pair (M size: For arm girths from 20 to 32 cm)
BP-203RPE3 arm cuff hoses, left and right pair
BP-203RPE3 ankle cuffs, left and right pair
Sensor box
ECG clips
Phonocardiogram sensor (PCG sensor)
PCG weight
Sensor gel packet (consumable packet) 5 sets
• PCG sensor pad, 1 piece x 5
• ECG clip electrodes, 3 pieces x 5
Blood vessel model
BP-203RPE3 touch pen
BP-203RPE3 manual
Quick manuals
15
1-5. Product and Accessories
Accessories (Sold Separately)
Product description REF Model
Right arm cuff, S size 9999492-9 HEM-CS30-RIGHT
Right arm cuff, M size 9999490-2 HEM-CR30-RIGHT
Right arm cuff, L size 9999496-1 HEM-CL30-RIGHT
Left arm cuff, S size 9999494-5 HEM-CS30-LEFT
Left arm cuff, M size 9999491-0 HEM-CR30-LEFT
Left arm cuff, L size 9999498-8 HEM-CL30-LEFT
Arm cuff air tube (right) 9999505-4 HEM-CR30R-TUBE
Arm cuff air tube (left) 9999504-6 HEM-CR30L-TUBE
Ankle cuff (right) 9999500-3 HEM-CR31-RIGHT
Ankle cuff (left) 9999501-1 HEM-CR31-LEFT
ECG clips 9999503-8 HFA-RPE3-ECG
PCG sensor 9999507-0 HFA-RPE3-PCG
PCG weight 9999506-2 HFA-RPE3-W700
Sensor gel packet (20 sets) 9967933-0 HBP-FORM-101S
Blood vessel model 9999510-0 HFA-FORM-ARTVS
Touch pen 9996749-2 HBP-PEN
LAN cable (for hub use) 9997621-1 HFA-RPE3-LANS
LAN cable 9997622-0 HFA-RPE3-LANX
Toe standard disposable cuff 9957110-6 HBP-DCUFF-TBI31
Toe small disposable cuff 9957111-4 HBP-DCUFF-TBI32
Toe cuff tube (right) 9957112-2 HBP-FORM-TBICR
Toe cuff tube (left) 9957113-0 HBP-FORM-TBICL
Table-tap 9515310-5 HBP-RPE3-TAKO
Stand cover 9511950-0 HFA-RPE3-CVR
Hook-and-loop fastener for the sensor box 9511951-9 HFA-RPE3-SBF
16
1-5. Product and Accessories
Options
Software and unit options can be purchased to expand the functionality of the product. For details on the uses and functions of the options, see Chapter 4.Refer to "4-1. Options" (page 121)
TBI package
HMC package
BP-203RPE3 bar code reader set
• Bar code reader
• Attachment fittings
Product description REF Model
TBI package 9512233-1 HFA-TBI-ENG
HMC package 9512232-3 HFA-RPE3-HMCPE
USB cable for HMC package 9903144-6 HBP-RPE3-USB
Serial cable for HMC package 9903143-8 HBP-RPE3-SER
Bar Code Reader Set 9996743-3 HBP-RPE3-BAR
TBI package manual 9512236-6 HFA-RPE3-TBIME
HMC package manual 9512235-8 HFA-RPE3-HMCME
Affix S/N label
BP-203RPE3 Unit Serial Number
Product Package Classification
HMC Package
OMRON HEALTHCARE CO., LTD.
Non-invasive Vascular Screening Device BP-203RPE 3
Product Package License Certificate
This certificate is a registration certificate for the product package that you purchased, and certifies that you, the customer, have lawfully received a license from OMRON HEALTHCARE Co., Ltd. OMRON HEALTHCARE Co., Ltd. permits the customer to install and use this product package on the BP-203RPE 3 unit indicated in this license certificate.
17
1-6. Name and Function of Each Part
Main Unit
Enter patient information, configure settings, and perform measurement.
Top Back
1. LCD display (touch panel)
When configuring settings: Setting buttons appear. Touch the setting buttons with the touch pen to enter settings. During measurement: Measured values, measured waveforms, and operation buttons appear.
2. Jog dial
Settings can be entered using the jog dial. Turn the jog dial right or left to select an item and press to enter.
3. Display lamps
Off: Power is off Green (on): Power is on (normal mode) Orange (on): Power is on (sleep mode)
4. [START] button
Press to begin measurement.
5. [STOP] button
Press to interrupt and stop measurement. On the screen with the [BACK] button, the [STOP] button can be also used to return to the previous screen.
6. Brightness button
: This enables the "-" and "+" brightness buttons. : This darkens the screen. : This makes the screen brighter.
7. Power on/off switch
Turns the power on and off. The power is normally turned on and off with the power switch on the stand, so keep the main unit power switched on.
8. Connectors
For details on the connectors, see Chapter 5. Refer to "5-3. Connections" (page 127)
18
1-6. Name and Function of Each Part
Stand
1. Arm
The sensors are placed on this arm.
2. PCG sensor pocket
Stores the PCG sensor.
3. Main unit holder
The main unit is placed on this holder.
4. Casters
During examination, lock the casters to keep the stand from moving. Unlock before moving the stand.
5. Cable cover
This prevents dust from collecting on the connectors on the back of the main unit.
6. Arm stand / touch pen stand
Attach the arm either on the left or the right. Insert the touch pen stand on the side that is empty.
Refer to "5-4. Changing the Arm Position" (page 129).
7. Drawer
Store printing paper and other supplies in the drawer.
8. Cable hook
Hang the sensor box cable on this hook.
9. Tray
Stores supplies on the tray.
10.Laser printer holder
11.Cuff storage pocket
Store arm and ankle cuffs.
12.Handle
Hold this when moving the main unit.
13.Power on/off switch
Turns the power of the main unit, printer, and options on/off at the same time.
( ) position is to turn on the power. ( ) position is to turn off the power.
14.Back pocket
Stores manuals and the power cord (when moving the stand).
19
1-7. Installation/Moving
Inspection Before Starting Work
For safe and proper use, inspect the device at the start of each day. When installing the device, be sure to follow the instructions on pages 7 to 12.
Before turning the power on
Main Unit
• Is the sensor that directly touches the patient clean?
• Are the cords arranged properly?
• Is the device kept dry?
• Is there no damage on the device?
• Are the casters not broken or caught by any foreign object?
Power Cord
• Is the power cord placed safe from people stepping on?
• Are the power cords not damaged? Is power-line not exposed? Is the power cords not frayed?
Supplies
• Are ECG clip electrodes and PCG sensor pads ready?
• Does the printer have enough toner and paper?
• Is paper loaded in the printer?
After turning the power on
Main Unit
• No smoke or abnormal smell?
• No abnormal noises?
Check the date and time setting
• Are the date and time correct? If not, see page 93 to set the correct date and time.
Warning
Do not install or store the device in a location where water or chemicals may splash on the device. Electric shock may result.
Caution
Do not install in a location where the temperature or humidity is outside the allowed range. This may cause malfunctioning or device failure.
20
1-7. Installation/Moving
Procedure for Moving
When you need to move the device, follow the procedure below.
1. Turn the power switch to the off ( ) position.
2. Remove the power plug from the outlet and coil the
cord.
Please do not pull the power cord.
3. Store the sensors on the arm.
4. Store the cuffs in the cuff storage pocket.
5. Hang the sensor box cable on the cable hook.
21
1-7. Installation/Moving
6. Unlock the casters.
7. Grasp the handle and move the device.
Do not push or pull any other parts rather than the handle.
8. After moving the device, lock the casters to secure
the stand.
22
2. Measurement Procedure
2-1. Preparing for Measurement
Measurement Procedure
Turn the power on
Enter or check patient information
Attach the cuffs and sensors to the patient
Make sure that "ECG: OK" and
"PCG: OK" appear on the display
Press the [START] button
Measurement starts
To stop measurement, press the [STOP] button
Measurement ends.
Measurement results will be displayed and/or
measurement report will be printed.
Have the patient in a supine position on
the bed and remain quite
23
2-1. Preparing for Measurement
Patient Information
Patient information is required to properly maintain a measurement history. Input items marked "Required" must be entered.
*Entry is only required if it will be used with the initial screen set to order
number entry. Patient information must be stored in advance.
There are two methods for entering patient information. A Enter the patient's information on the touch panel at the time of measurement. B Enter the patient's information in advance from "Advanced Registration of Patient Information".
This is convenient for group examinations. Refer to "3-8. Advanced Registration of Patient Information" (page 105).
Item name Input Page
ID
Required
26
Sex
29
Birth Date
30
Height
31
Name
Optional
28
Weight
31
Waist
32
Disease
32
Order Number*
33
Measurement Sensors
34
Measurement Sites
35
Upper limit of pressurizing
36
Synchro measurement
38
Doctor
38
Technician
39
Category
39
24
2-1. Preparing for Measurement
Initial Screen (ID Entry Screen)
Following a brief interval after the power is turned on, the ID entry screen appears. This screen is called the "initial screen" in this manual. The ID is entered in this screen. You can select whether the ID input type is all numbers, or both numbers and alphabetical characters. It is recommended that you decide which type of ID will be used in advance. The first time the power is turned on, the left-hand screen below appears. This screen is for number entry. If you wish to use combined number and character IDs, change the "ID Input type" in advance. Refer to "3-2. User Default Settings" "ID Input Type" (page 76).
If you manage data by "Order Number" associated with the electronic chart, you can change the initial screen from the ID entry screen to the order number entry screen. To use order numbers, patient information must be stored in advance.
Refer to "3-2. User Default Settings" "Search Key" (page 76).
[MAIN MENU] button: Touch this to change to the main menu screen. The main menu screen is
used to configure basic settings (refer to page 74).
[SWITCH TO ID] button:If you are measuring a patient that does not have an order number, you
can change back to ID number entry.
Initial screen (ID input type: numbers)
Initial screen (ID input type: numbers and characters)
Initial screen: order number entry screen (Input type: numbers)
Initial screen: order number entry screen (Input type: numbers and characters)
25
2-1. Preparing for Measurement
Entering and Editing Patient Information
Enter the patient information and edit previously stored patient information.
Notes:
• Do not assign multiple IDs to a single patient. Even if all other information such as the name is the same, the system will treat different IDs as different patients. In this case, you will not be able to make full use of the device's functions for long-term storage of measurement histories and diagnosis support.
• The number of digits must be the same or the ID will treated as two separate IDs. Example: "300" and "0300". The above will be treated as two separate IDs.
• The [ID], [SEX], [BIRTH DATE], and [HEIGHT] must be entered. If not entered, measurement cannot be performed.
• Once ID number or order number is entered in the initial screen, the "ID" or "Order Number" cannot be changed.
• Take care to avoid mistakes when entering information and numerical values, as the examination results will be printed based on this information.
• If you return to the initial screen without saving patient information, the information will be cleared.
26
2-1. Preparing for Measurement
Entering the ID Number
Follow the steps below to enter an ID number for the first time or check existing ID numbers. Up to 13 characters (including hyphens "-") can be entered. ID number must be entered.
For entering an Order number, up to 16 digits of number (including hyphens"-") can be entered.
*The initial screen for "ID input type: numbers" is shown as an example. The key arrangement is different for "ID input
type: numbers and characters".
First-time patient
Patient information not registered
1 Enter a new ID number
1 Enter the ID number
2 Press the [SEARCH / NEXT] button
1 Enter the order number
2 Press the [SEARCH / NEXT] button
2 Press the [SEARCH / NEXT] button
3 Press the [SEARCH / NEXT] button
Patient for whom patient information is already registered
Patient information already registered
Patient for whom re-measurement is necessary due to inadequate measurement results
Previously measured (patient information already registered)
1 Press the [PREVIOUS ID] button
1 Press the [SWITCH TO ID]
button
2 Check the ID number
* Press the [SWITCH TO ID] button
to change to the ID entry screen
27
2-1. Preparing for Measurement
Register patient information screen
For the items to be entered, see pages 28 to 39.
For the items to be entered, see pages 28 to 39.
[BACK] button: Return to the initial screen. [HMC DATA PROC] button: Can be used when the optional HMC Package is connected. [NEXT] button: Proceed to the measurement screen.
[MEAS. HISTORY] button: Show the patient's measurement history.
[BACK] button: Return to the initial screen. [HMC DATA PROC] button:
Can be used when the optional HMC Package is connected.
[NEXT] button: Proceed to the measurement screen.
1 Enter patient information
2 Press the [NEXT] button
2 Press the [NEXT] button
1 Check the patient information. Enter any items that have not been entered as needed.
The measurement screen will appear. Attach the cuffs and sensors to the patient and start measuring. Refer to page 41.
Confirm patient information screen
28
2-1. Preparing for Measurement
Entering the Name
1. Press the [NAME] button.
2. Enter the name.
• Up to 40 characters can be entered.
• A space counts as one character.
• To enter a space, press the [SPACE] button.
• To switch between upper case and lower case letters, use the [UPPER] button / [lower] button. When upper case entry is selected, the [lower] button appears. When lower case entry is selected, the [UPPER] button appears. Each time the button is pressed, it changes between [UPPER] and [lower].
• To change a character, press the [ ] or [ ] button to move the cursor to that character, press the [DELETE] button, and enter the new character.
• To delete all characters that have been entered, press the [CLR ALL] button.
3. Press the [OK] button.
To cancel entry, press the [CANCEL] button.