z Copyright law allows no part of this instruction manual to be reproduced without
permission.
z The content of this manual are subject to change without notice.
z The contents of this manual should be correct. If, for some reason, there are any
questionable points, please do not hesitate to contact our service center.
z The manual will be replaced if any pages are missing or collation is incorrect.
Warranty
z Please contact your local distributor about the warranty period.
z Device failure or damage related to the following situations during the guarantee period
is not covered by this warranty:
z Installation, transfer installation, maintenance and repairs by any person other than
an authorized Omron Healthcare Co., Ltd. employee or technician specified by
Omron Healthcare Co., Ltd..
z Damage sustained to the Omron Healthcare product(s) caused by product(s) from
another company excluding products delivered by Omron Healthcare Co., Ltd.
z Damage – caused by mishandling and/or misuse – is the responsibility of the user.
z Maintenance and repairs utilizing maintenance components that are not specified
by Omron Healthcare Co., Ltd.
z Device modifications or use of accessories not recommended by Omron
Healthcare Co., Ltd.
z Damage caused by accidents or natural disasters (earthquakes, flooding, etc.).
z Damage resulting from usage where caution statements and operating instructions
shown in this manual have not been followed.
z Damage due to neglect of specified maintenance checks.
z This warranty only covers the hardware of the BP-S510. The warranty does not cover the
following selections:
z Whatever damage or loss results from the attachment of accessories or their
operation.
z In the event of a defect in the product, contact our sales outlet or EU representative
as noted on the back cover.
z The BP-S510 conforms to the EMC standard IEC60601-1-2.
Note that mobile phones should not be used in the vicinity of the BP-S510.
Note, however, any device not complying to the EMC standard that is used with the BP-S510
renders the BP-S510 as non-compliable to the EMC standard.
Revision History
Product brand names shown in this manual are likely to be the trademark or registered
trademark of the company concerned.
Front Case Disassembly (B) ...............................................................................................................73
Rear Case Disassembly (C)................................................................................................................77
SPARE PARTS ...........................................................................................................................................87
Figure 11. Air Leakage Test ................................................................................................................................................................. 29
Figure 12. The access of Service Menu via Set-up menu ................................................................................................................... 43
Figure 13. Entering the pass code....................................................................................................................................................... 44
Figure 14. Service Menu...................................................................................................................................................................... 44
Figure 20. Front Case Disassembly – LCD Bracket, Side Key Case .................................................................................................. 73
Figure 21. Front Case Disassembly – LCD, Insulation, Inverter Board, Front Key Board,.................................................................. 74
Figure 22. Front Case Disassembly – Bezel, LCD Window, Front Key Array, Side Key Array............................................................ 76
Figure 23. Rear Case Disassembly – Patient Sensor Case, Water Trap, CO
Figure 24. Rear Case Disassembly – TRX-02 Module Option Cover, Recorder, Main Board............................................................. 79
Figure 25. Rear Case Disassembly – ECG Board, SpO
Module, CPU Module................................................................................. 80
2
Figure 26. Rear Case Disassembly – NIBP Module, Hub Board, ECG Wire, CO
Figure 27. Rear Case Disassembly – In-Case..................................................................................................................................... 82
Figure 28. Rear Case Disassembly – SMPS....................................................................................................................................... 83
Figure 29. Rear Case Disassembly – VESA Bracket, Handle............................................................................................................. 84
Figure 32. System Block Diagram ..................................................................................................................................................... 101
Figure 33. Power Unit Block Diagram................................................................................................................................................ 102
Figure 34. Process Unit Block Diagram............................................................................................................................................. 102
Figure 35. User-Control Unit Block Diagram...................................................................................................................................... 103
Figure 36. Sound Unit Block Diagram................................................................................................................................................ 103
Figure 37. Communication Unit Block Diagram................................................................................................................................. 103
Figure 38. GUI Unit Block Diagram.................................................................................................................................................... 104
Figure 39. Thermal Recorder Unit Block Diagram............................................................................................................................. 104
Figure 40. NIBP Unit Block Diagram.................................................................................................................................................. 104
Figure 41. ECG Unit Block Diagram .................................................................................................................................................. 105
Figure 42. Respiration Unit Block Diagram........................................................................................................................................ 105
Figure 43. SpO
Unit Block Diagram.................................................................................................................................................. 105
2
Figure 44. Temperature Unit Block Diagram...................................................................................................................................... 106
Figure 45. IBP Unit Block Diagram .................................................................................................................................................... 106
Figure 46. CO
Unit Block Diagram ................................................................................................................................................... 107
2
Figure 47. TRX-02 Unit Block Diagram.............................................................................................................................................. 107
Table 5. Patient Leakage Current Values............................................................................................................................................. 36
Table 6. Patient Leakage Current Values - Mains Voltage on Applied Part ......................................................................................... 36
Table 7. Test Lead Combinations......................................................................................................................................................... 37
Table 9. Service menu ......................................................................................................................................................................... 45
T able 10. Audible Alarm Characteristics .............................................................................................................................................. 46
Table 11. System Test.......................................................................................................................................................................... 48
Table 16. Problem Categories.............................................................................................................................................................. 56
Table 18. Alarm Messages and Check Items....................................................................................................................................... 62
Table 19. Part Descriptions – Front Case and Rear Case Assembly .................................................................................................. 72
Table 20. Part Descriptions – LCD Bracket, Side Key Case................................................................................................................ 73
Table 21. Part Descriptions – LCD, Insulation, Inverter Board, Front Key Board, ............................................................................... 74
Table 22. Part Descriptions – Bezel, LCD Window, Front Key Array, Side Key Array ......................................................................... 76
Table 23. Part Descriptions – Patient Sensor Case, Water Trap, CO2 Absorber & Holder, AC Cord Clasp & Holder......................... 77
Table 24. Part Descriptions – TRX-02 Module Option Cover, Recorder, Main Board.......................................................................... 79
Table 25. Part Descriptions – ECG Board, SpO2 Module, CPU Module............................................................................................. 80
Table 26. Part Descriptions – NIBP Module, Hub Board, ECG Wire, CO2 Module............................................................................. 81
Table 27. Part Descriptions – In Case.................................................................................................................................................. 82
Table 28. Part Descriptions – SMPS.................................................................................................................................................... 83
Table 29. Part Descriptions – VESA Bracket, Handle.......................................................................................................................... 84
Table 30. Spare Part List...................................................................................................................................................................... 88
iv
SAFETY INFORMATION
General Safety Information
This section contains important safety information related to general use of the BP-S510
multi-parameter patient monitor. Other important safety information appears throughout
the manual. The BP-S510 will be referred to as the monitor throughout this manual.
Important! Before use, carefully read this manual, accessory directions for use, all
precautionary information and specifications.
Warnings
Warnings are identified by the WARNING symbol shown above.
Warnings alert you to potential serious outcomes (death, injury, or adverse events)
to the patient or user.
WARNING: Explosion hazard. Do not use the monitor in the presence of flammable
anesthetics or gases.
WARNING: Do not spray, pour, or spill any liquid on the monitor, its accessories,
connectors, switches, or openings in the chassis.
WARNING: Do not immerse the monitor or its accessories in liquid or clean with
caustic or abrasive cleaners.
WARNING: Ensure that conductive portions of the electrodes, leads, and cable do
not come into contact with any other conductive parts,
WARNING: Before attempting to open or disassemble the monitor, disconnect the
power cord from the monitor.
WARNING: Chemicals from a broken LCD display panel are toxic when ingested.
Use caution when handling a monitor with a broken display panel.
WARNING: The use of accessories, transducers, and cables other than those
specified may result in increased emission and/or decreased immunity of the
monitor.
WARNING: Do not silence the monitor audible alarm or decrease its volume if
patient safety could be compromised.
WARNING: During the safety test, AC main voltage will be present on the applied
part terminals, Exercise caution to avoid electrical shock hazard.
WARNING: Do not place the monitor into operation after repair or maintenance has
been performed, until all Performance Tests and Safety Tests listed in
Performance Verification section of this service manual have been performed.
Failure to perform all tests could result in erroneous monitor readings.
WARNING: High voltage is generated by the LCD backlight driver. Exercise
caution when operating monitor with covers open.
1
Cautions
Cautions are identified by the CAUTION symbol shown above.
Caution statements identify conditions or practices that could result in damage to
the equipment or other property.
CAUTION: Observe ESD (electrostatic discharge) precautions when working within
the unit and/or when disassembling and reassembling the monitor and when
handling any of the components of the monitor.
CAUTION: When reassembling the monitor, over-tightening could strip out the
screw holes in the cases, rendering it unusable.
CAUTION: If any problem with the monitor built in an optional recorder, check a
recorder’s door is closed well. Operating error may be caused if the cover is not
closed correctly.
CAUTION: If internal battery cable has been disconnected, pay particular attention
to polarity of the cable before reattaching. If battery cable polarity is reversed, it is
likely that circuit damage will occur.
CAUTION: Ferrite Cores are used for electromagnetic compatibility. Please do not
remove Ferrite Cores while disassembling or reassembling, otherwise the monitor
can be affected by electromagnetic interference and measure inaccurate data to be
displayed or stored.
CAUTION: For continued protection against risk of fire, replace only with same type
and rating of fuse.
2
Manual Overview
This manual contains information for servicing the BP-S510 monitor.
The monitor subsequently referred to as the monitor throughout this manual. Only
qualified service personnel should service this product. Before servicing the monitor, read
the operator’s manual carefully for a thorough understanding of safe operation.
Read and understand all safety warnings and service notes printed in this service manual
and the operator’s manual part number A7072
Related Documents
To perform test and troubleshooting procedures and to understand the principles of
operation and circuit analysis sections of this manual, you must know how to operate the
monitor. Refer to the monitor operator’s manual part number A7072
To understand the various SpO
probes, IIBP transducer and CO
individual directions for use that accompany these accessories.
sensors, ECG leads, blood pressure cuffs, temperature
2
accessories that work with the monitor, refer to the
2
Intended Use for the BP-S510 Monitor
The BP-S510 is intended to be used to monitor electrocardiography, heart rate,
noninvasive blood pressure (systolic, diastolic and mean arterial pressures), functional
arterial oxygen saturation, pulse rate, respiration, temperature, invasive blood pressure
and capnography for adult and neonatal patients in all areas of a hospital and hospitaltype facilities. Monitor users should be skilled at the level of a technician, doctor, nurse or
medical specialist.
Note: Hospital use typically includes such areas as general care floors, operating rooms,
special procedure areas, intensive and critical care area, within the hospital.
Hospital-type facilities include physician office-based facilities, sleep labs, skilled
nursing facilities, surgical centers, and sub acute care centers.
Identifying the BP-S510 monitor Configurations
The following table identities BP-S510 monitor configurations and how they are indicated.
The model-option number and serial number are located on the back of the monitor.
All information in this manual, including the illustrations, is based on a monitor configured
with Capnography (EtCO
exist, please verify your unit configuration.
and InCO2), IBP and recorder. If the relevant functions do not
2
Configuration REF No. Description
BP-S510 112580 Standard
(ECG, NIBP, SpO
BP-S510P 112581 Standard + Recorder
BP-S510C 112582 Standard + Capnography
BP-S510PC 112583 Standard + Capnography + Recorder
S510-IBP 131324 2-channel IBP module
, 2-channel Temperature, Respiration)
2
3
Front Panel Components
1 Visual alarm indicator
2 Alarm silence button
3 NIBP interval button
4 NIBP start/stop button
5 LCD
6 Jog dial
7 Power on indicator
8 Power on/off button
9 Battery charging indicator 1
10 Battery charging indicator 2
11 Record button
12 Trend button
13 Home button
1 ECG connector 7 NIBP connector
2 Temperature channel 1 8 AC power connector
3 Temperature channel 2 9 Water trap (option)
4 SpO
5 IBP channel 1 (option) 11 CO
6 IBP channel 2 (option) 12 CO
connector 10 CO2 connector (option)
2
filter (option)
2
exhaust port (option)
2
Figure 3. Left Panel Components
6
Right Panel Components
1 USB port 4Battery cover
2 LAN port 5Recorder (option)
3 RS-232 port
Figure 4. Right Panel Components
7
This page is intentionally left blank.
8
ROUTINE MAINTENANCE
WARNING: Do not spray or pour any liquid on the monitor or its accessories. Do
not immerse the monitor or its accessories in liquid or clean with caustic or
abrasive cleaners.
Cleaning
The monitor may be surface-cleaned by using a soft cloth dampened with either a
commercial, nonabrasive cleaner or one of the solution listed in the below. Lightly wipe
the top, bottom and front surfaces of the monitor lightly.
z 70% Isopropyl alcohol
z 10% Chlorine bleach solution
For cables, sensors, cuffs, transducer and probes, follow cleaning instructions in the
directions for use shipped with those components.
Avoid spilling liquid on the monitor, especially in connector areas. If liquid is accidentally
spilled on the monitor, clean and dry thoroughly before reuse. If in doubt about monitor
safety, refer the unit to qualified service personnel for checking.
Periodic Safety and Functional Checks
The following performance verification tests may be used following repair or during routine
maintenance (if required by your local institution). The following checks should be
performed at least every year by qualified service personnel.
1. Inspect labels for legibility. If the labels are not legible, contact your local supplier.
2. If the monitor has been visibly damaged or subjected to mechanical shock (for example,
if dropped), perform the performance tests as described in Performance Verification
section. If the unit fails these performance tests, refer to Troubleshooting section.
3. Perform the electrical safety tests described in Performance Verification section. If
the unit fails these electrical safety tests, do not attempt to repair. Contact your local
supplier.
4. Inspect the fuses for proper value and rating.
qty 2, 6.3 A, 250 volts for AC mains
9
Batteries
If the monitor has not been used for a long period of time, more than 2 months, the battery
will need charging. To charge the battery, connect the monitor to an AC outlet as
described in Battery Charge paragraph in this service manual or the Battery Operation
section of the operator’s manual.
Note: Storing the monitor for a long period without charging the battery may degrade the
battery capacity. The battery may require a full charge/discharge cycle to restore
normal capacity. Omron Healthcare recommends that the monitor’s sealed, Ni-MH
batteries be replaced at 2 year intervals. Refer to Disassembly Guide section.
CAUTION: If the monitor is to be stored for a period of 2 months or longer, it is
recommended to notify service personnel to remove the battery from the monitor
prior to storage. To replace or remove the battery, refer to Disassembly Guide
section. Recharging the battery is strongly recommended when the battery has not
been recharged for 2 or more months.
CAUTION: If the battery shows any signs of damage, leakage, or cracking, it must
be replaced immediately.
CAUTION: Discarded battery may explode during incineration. Recycle used
batteries properly. Do not dispose of batteries in refuse containers.
Environmental Protection
Follow local governing ordinances and recycling plans regarding disposal or recycling
batteries and other device components.
Note: The monitor should be disposed of separately from the municipal waste stream via
designated collection facilities appointed by the government or the local authorities.
Note: The correct disposal of your old appliance will help prevent potential negative
consequences for the environment and human health.
Note: For more detailed information about disposal of your old appliance, please contact
your city office, waste disposal service or the shop where you purchased the
monitor.
10
PERFORMANCE VERIFICATION
General
This section discusses the tests used to verify performance following repairs or during
routine maintenance. All tests can be performed without removing the monitor covers. All
tests except the battery charge and battery discharge tests must be performed as the last
operation before the monitor is returned to the user.
If the monitor fails to perform as specified in any test, repairs must be made to correct the
problem before the monitor is returned to the user.
Required Equipment
Table 1 lists the equipment required for performance verifications.
Table 1. Required Equipment
Equipment Description
Digital multimeter (DMM) Fluke Model 87 or equivalent
ECG cable ECG cable No.8
ECG 3 lead wires ECG 3 lead wires No.5
ECG 5 lead wires ECG 5 lead wires No.6 (optional)
NIBP cuff for neonatal NIBP cuff No.11 (3cm)
NIBP cuff for adult NIBP cuff No.3 (12cm)
NIBP cuff hose for neonatal NIBP cuff hose No.3 (3.5m)
NIBP cuff hose for adult NIBP cuff hose No.1 (3.5m)
NIBP rigid PVC vessel 9 cm diameter
NIBP rigid PVC vessel 5 cm diameter
SpO2 extension cable DOC-10
SpO2 sensor (durable) DS-100A
Temperature probes YSI-400 series
IBP extension cable IBP extension cable
ECG simulator Metron PS-420 or equivalent
SpO2 simulator Nellcor SRC-MAX simulator
NIBP simulator Bio-Tek BP Pump 2 or equivalent
Sphygmomanometer small Common used sphygmomanometer
Y tube Omron Y tube
Cuff joint Omron cuff joint
Inflation bulb Omron inflation bulb
Temperature simulator Metron PS-420 or equivalent
IBP simulator Metron PS-420 or equivalent
IBP test cable Omron IBP test cable
CO2 gas flow meter STEC SEF-21A or equivalent
Safety analyzer Metron QA-90 or equivalent
Data interface cable RS-232 cable
LAN cable Common used LAN cable
USB memory Common used USB memory
Stopwatch Manual or electronic
Note: The sphygmomanometer small shall be calibrated periodically. The correct value
can not be found if the sphygmomanometer has not been calibrated.
Note: Contact your local supplier for pricing and ordering information.
11
System Tests
The monitor must be placed in the service menu. For a detailed explanation to access the
service menu, refer to Service Menu and Factory Default Settings section.
1. Rotate the jog dial to select System Test in the service menu, and then press the jog
Switch/LED Test
This tests the buttons, jog dial and the visual alarm indicator.
dial.
Figure 5. Switch/LED Test
1. Select Switch/LED Test in the system test. The image screen will appear as Figure
5.
2. Verify that the alarm visual indicator flashes with red and yellow.
3. Press all the buttons one by one, except for the power on/off button.
When a button is pressed, the same button on the image screen will turn blue.
For example, when the jog dial is pressed, the jog dial on the screen turns blue.
When the jog dial is rotated, the sign indicating the rotate direction will appear.
4. After finishing the test, press the jog dial twice to exit.
Pass/Fail Results
If a pressed button turns blue on the screen, the button is in a normal state.
12
LCD Test
This tests the LCD display.
1. Select LCD Test in the system test.
2. The screen color will change over the following sequence:
3. After testing the test, press jog dial twice to exit.
Pass/Fail Results
When the color of the test screen changes in the order from Red, Green, Blue, White to
Black, the LCD display is in a normal state.
Alarm Audible Test
This tests the alarm tones by displaying the level of the alarm tone on the screen.
Red Æ Green Æ Blue Æ White Æ Black … every two seconds
Figure 6. Alarm Audible Test
1. Select Alarm Audible Test in the system test.
2. The level of the alarm tone will appear on the screen as the alarm tone sounds. Then,
the level goes up gradually. When the tone reaches the maximum level 8, it returns to
the minimum level 1.
3. After finishing the test, press the jog dial to select Cancel. The menu box will
disappear.
Pass/Fail Results
When the alarm tone changes to eight steps, the alarm tone is in a normal state.
13
Tone Audible Test
This tests HR/PR tones, key beeps and completion sounds.
1. Select Tone Audible Test in the system test.
2. Rotate the jog dial to select HR/PR Tone.
3. Press Cancel to finish the test.
4. Rotate the jog dial to select Key Beep or Completion Sound.
5. Verify that Key Beep or Completion sound sounds properly.
6. After finishing the test, press the jog dial to select Cancel. The menu box will
Pass/Fail Results
When the HR/PR tone changes to seven steps and the pitch changes to three steps, the
HR/PR tone is normally set.
When the key beep changes to seven steps, the key beep is normally set.
When the completion sound is heard, it is normally set.
Figure 7. Tone Audible Test
The HR/PR tone sounds intermittently as the level goes up gradually. When the tone
reaches the maximum level 7, it returns to the minimum 1.
When the tone returns to the minimum, the pitch will change automatically.
There are three pitches – High, Med and Low.
disappear.
14
Recorder Test
This tests the printing condition.
Figure 8. Recorder Test
1. Select Recorder Test in the system test.
The recorder condition is displayed as follows:
Paper: The “ Mounted” sign appears when the paper is installed properly. If the paper
runs out or the printer door is open, “empty” sign will appear in red.
Head Temperature: The “OK” sign appears when the head temperature is normal. If
the head temperature is abnormal, the ”error” sign appears.
Vp Voltage: The “OK” sign appears when the voltage is normal. If the voltage is
abnormal, the “error” sign appears.
2. Select Grid Print. Verify that printing can be done without meandering.
3. After checking, select Cancel to stop printing.
4. Select Contact Print. Verify that printing can be done at an equal density.
5. After checking, select Cancel to stop printing.
Pass/Fail Results
When the printed line is free from meandering on unevenness in density, the printing
function is normal.
15
Backup RAM Clear
When Backup RAM Clear is set to Yes, all settings of the monitor including the service
settings return to the factory defaults from the next cycle.
Note: Set the monitor to appropriate AC line frequency (50Hz or 60Hz) again after Back
Figure 9. Backup RAM Clear
Up RAM clearance.
16
Performance Tests
The battery charge and battery discharge test should be performed before monitor repairs
whenever the battery is suspected as being a source of the problems. All other tests may
be used following repairs or during routine maintenance (if required by your local
institution). Before performing the battery discharge test, ensure that the battery is fully
charged. This section is written using factory defaults as power-up. If your institution has
preconfigured custom defaults, those values will be displayed.
Power
1. Connect the monitor to AC power source using proper power cord.
2. Verify that Battery charging indicator is lit.
3. Press Power on/off button.
4. Verify that the monitor is turned on and that Power on indicator is lit.
5. After the monitor operates in normal mode, disconnect the power cord.
6. Verify that Battery status icon appears on the screen instead of lighting Battery
charging indicator.
7. Press Power on/off button over 1 second, and then verify that the monitor is turned off.
Battery Charge
1. Connect the monitor to AC power source using proper power cord.
2. Verify that Battery charging indicator is lit with orange.
3. Charge the battery fully until Battery charging indicator 1 and 2 are changed to green.
It will be take about 12 hours.
4. To check for a full charge, perform the procedure in paragraph “Battery Discharge”.
Note: The battery may require a complete charge/discharge cycle to restore its normal
Battery Discharge
1. Disconnect the power cord from the monitor with a fully charge battery.
2. Turn on the monitor by pressing Power on/off button.
3. Verify that Battery status icon appears at bottom of the screen after power-on self-test
is completed. The bar in battery status icon should be filled, indicating battery is
charged.
4. Connect the SpO
5. Connect the NIBP simulator to the monitor via NIBP cuff hose.
capacity, depending on its previous usage.
simulator to monitor via the SpO2 extension cable.
2
6. Set the SpO
simulator as follows: SpO2 of 95% and pulse rate of 60bpm.
2
7. Set the NIBP simulator to simulate pressure setting of 120/80 mmHg and heart rate of
80 bpm.
8. Set NIBP auto mode interval to 5 minutes.
9. The monitor must operate for 1 hour with a fully charged battery. The monitor must
operate for at least 15 minutes before the monitor powers down due to the low battery
condition.
10. Verify that the low priority alarm occurs and the alarm message “System:Low battery” is displayed about 15 minutes before battery fully discharges.
17
11. Allow monitor to operate until it automatically powers down due to the low battery
condition. Verify that the high priority alarm occurs and the alarm message “System: Critically low-battery condition” is displayed about 5 minutes before the monitor
automatically shuts down.
12. If the monitor passes this test, immediately recharge the battery. (paragraph “Battery
Charge”)
Power-On Self-Test (POST)
1. Connect the monitor to AC power source and verify that Battery charging indicator is
lit.
2. Observe the monitor’s LCD screen. With the monitor off, press Power on/off button.
The monitor must perform the following sequence.
a. The monitor progresses the checksum for the flash memory and displays the
bar while the checksum is proceeding.
b. After the checksum for flash memory is completed, the copyright screen
appears and all indicators are lit for a few seconds. The copyright screen
displays the company logo, the version of system and the current time.
c. Then the monitor performs the power-on self-test (POST).
d. Upon successful completion of power-on self-test (POST), the post pass tone
sounds and the monitor will be in normal monitoring screen.
Note: Power-on self-test (POST) including the checksum for the flash memory takes
approximately 13 seconds to complete.
Note: If an error condition occurs during the POST, the monitor will be display an error
message.
Note: During the POST, the integrity of all programming is checked first. If software testing
is successful, hardware tests are initiated. If all testing is successful, the monitor is
ready for use. If an error message is displayed during the POST, please refer to
Troubleshooting section.
18
General Operation Tests
Alarms and Alarm Silence
1. Connect the monitor to an AC power source.
2. Press Power on/off button to turn on the monitor.
3. Connect the SpO
simulator to the SpO2 extension cable and connect the cable to the
2
monitor.
4. Set the SpO
simulator as follows: SpO2 of 75% and pulse rate of 60bpm.
2
5. Verify following the monitor reaction:
a. The pulse amplitude indicator begins to track artificial pulse signal from the
simulator.
SpO
2
b. After about 10 to 20 seconds, the monitor displays oxygen saturation and pulse
rate as specified by simulator. Verify values are within following tolerances:
z Tolerance of Oxygen Saturation : ±2 %
z Tolerance of Pulse Rate : ±3 bpm
c. Audible alarm sounds and “SpO
displayed and %SpO
numerical area will flash, indicating the parameter has
6. Press Alarm silence button on the monitor’s front panel. Audible alarm will be
temporarily silenced.
7. Verify the following:
a. An audible alarm remains silenced.
b. Alarm silence icon appears in SpO
numerical area on the screen.
2
c. %SpO
d. Audible alarm returns in approximately 120 seconds.
HR/PR Tone Volume Control
1. Press Power on/off button to turn on the monitor.
2. Connect the SpO
monitor.
3. Set the SpO
4. Verify %SpO
5. Press Alarm silence button on front panel of the monitor to temporarily silence audible
alarm.
6. Select Setup icon to display Setup menu.
7. Select HR/PR tone Volume in Setup menu.
8. Set HR/PR tone volume level 1 to level 7 and return to the monitoring screen. Verify
beeping pulse rate tone increases.
9. Set HR/PR tone volume level 7 to level 1 and return to the monitoring screen. Verify
beeping pulse rate tone decreases.
display continues flashing.
2
simulator to SpO2 extension cable and connect the cable to the
2
simulator as follows: SpO2 of 75% and pulse rate of 60bpm.
2
and pulse rate values are correctly displayed.
2
19
10. Set HR/PR tone volume to Off and return to the monitoring screen. Verify beeping
pulse rate tone is no longer audible.
11. Return HR/PR tone volume to a comfortable level.
Sensor LED Excitation Test
This procedure uses normal system components to test circuit operation. An SpO
DS-100A is used to examine LED intensity control. The red LED is used to verify intensity
modulation caused by the LED intensity control circuit.
1. Connect the monitor to an AC power source.
2. Press the Power on/off button to turn on the monitor.
3. Connect the SpO
4. Connect the SpO
extension cable to the monitor.
2
sensor to the SpO2 extension cable.
2
5. Leave the sensor open with the LEDs and photo detector visible.
6. After the monitor completes its normal power-up sequence, verify that the sensor LED
is brightly lit.
7. Slowly move sensor LED in proximity of photodetector element of the sensor (close the
sensor slowly). Verify; as LED approaches the optical sensor, that the LED intensity
decreases.
8. Open the sensor and notice that the LED intensity increases.
9. Repeat step 7 and intensity will again decrease. This variation is an indication that the
microprocessor is in proper control of LED intensity.
10. Press Power on/off button to turn off the monitor.
Restoring Power-On Default Settings
sensor,
2
The following test procedures will verify that alarms are activated at the level of factory
default alarm limits and that any changed settings are saved and in effect when the user
changes alarm limit settings and saves the current settings as a power default.
1. Turn on the monitor at the factory default settings.
2. Select Alarm limits icon to display Alarm limits menu.
3. Verify that alarm limits are set to as shown in Table 2.
4. Change Patient mode in the setup menu Adult to Neonatal, then verify that alarm limits
are set to as shown in Table 2.
5. Change alarm limit value via Alarm limits menu.
6. Save the changed alarm limit values as a power on default setting via Service menu
(see Service Menu and Factory Defaults section). Turn off the monitor.
7. Press Power on/off button to turn on the monitor.
8. Verify alarm limits are set to the changed alarm limit values.
Recorder Test (Option)
If Recorder option is installed in the monitor, the following test procedures will verify the
recorder performance.
1. Connect the monitor to an AC power source.
2. Press Power on/off button to turn on the monitor.
3. Connect all necessary simulators to the monitor.
4. Select Setup icon on the screen to display Setup menu.
5. Test #1: 20 sec printing
a. Set Wave record time to 20 sec via setup menu.
b. Press Record button when all the parameter signals display normally.
c. Verify that the parameter values and waveforms are printed out for 20
seconds.
6. Test #2: Continuous printing
a. Set Wave record time to Continuous via setup menu
b. Press Record button when all the parameter signals display normally.
c. Verify the parameter values and waveforms are printed out continuously.
d. Verify printing stops with pressing Record button again.
21
7. Test #3: Record speed
a. Set Record speed to 25 mm/s.
b. Press Record button when all the parameter signals display normally.
c. Verify the parameter values and waveforms are printed out with 25 mm/s.
d. Set Record speed to 50 mm/s.
e. Press Record button when all the parameter signals display normally.
f. Verify that the parameter values and waveforms are printed out with 50 mm/s.
8. Test #4: Wave Record Select
a. Set Wave Record Select to ECG1+ECG2 via Setup menu.
b. Press Record button when all the parameter signals display normally.
c. Verify the two ECG waveforms are printed out.
d. Repeat this test for other selections.
9. Test #5: Record on Alarm
a. Set Record on Alarm to ON via Setup menu.
10. Test#6: Auto List Record
Note: If no recorder is installed in the monitor, the record setup menus will be grayed out
Note: If there is no recorder paper left or recorder paper places improperly, the monitor
Serial Interface Test
Perform the following procedure to test the serial port. The serial connector DSub-15,
located on the monitor’s right panel, identified with the data interface symbol (RS-232).
1. Turn on the monitor, and select the service menu via the setup menu.
2. Enter the service code, 6, 1, 6, in order.
b. Set Heart rate of ECG simulator to 30 bpm.
c. Verify “HR: Lower limit violated” alarm is activated and the parameter
values and waveforms are printed out.
a. Set Auto list record to ON via Setup menu.
b. Set Save time interval to 0.5 minutes via Tabular trend menu.
c. Verify that the monitor automatically prints out the data after stored 8 data in
the trend memory.
in Setup menu.
will display an informative message.
22
3. Connect the serial interface cable between the monitor and PC COM1 port.
4. Run “BP-S510_Comm_Test.exe” on PC.
5. Click Serial on the left side of dialog box for the serial interface test.
6. Click Connect, then verify that log message “RS232C [Com1] Connected” appears on
the PC.
7. Click Send, then verify the followings:
- The monitor displays three dashes (-) with the beep sound.
- The log message “RS232 Test OK!!!” appears on the PC.
8. Click Disconnect to stop the serial interface test.
23
Network Test
Perform the following procedure to test the Network. The Network connector is located on
the monitor’s right panel, identified with the Network symbol.
Prior to the network test, configure the network setting on PC with the followings.
- IP address 192.168.124.1.
- Gateway 192.168. 124. 254.
- Subnet Mask 255.255.255.0.
When completed the network setting on the PC properly, follow the below test procedure.
1. Turn on the monitor, and select the service menu via the setup menu.
2. Enter the service code, 6, 1, 6, in order.
3. Connect a network line to the monitor. (Do not connect PC to the monitor directly)
4. Run “BP-S510_Comm_Test.exe” on PC.
5. Click TCP/IP on the left side of the dialog box for the network test.
6. Click Connect, then verify that log message “TCP/IP [192.168.124.2] Connected”
appears on the PC.
24
Loading...
+ 89 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.