Omron BP762 User Manual

7 series Plus
Blood Pressure Monitor
with ComFit
TM
Cuff
Model BP762

INSTRUCTION MANUAL

ESPAÑOL
2
Before Using the Monitor
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Operating the Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Risk of Electrical Shock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Care and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Before Taking a Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Operating Instructions
Know Your Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Unit Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Display Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Irregular Heartbeat Symbol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Movement Symbol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Heartbeat Symbol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Average Value Symbol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Cuff Wrap Guide Symbol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Using the AC Adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Battery Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Setting the Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Applying the Arm Cuff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Taking a Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Cuff Wrap Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Special Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
Using the Memory Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
Averaging Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
To Display the Measurement Values . . . . . . . . . . . . . . . . . . . . . . . .31
To Delete All Values Stored in the Memory . . . . . . . . . . . . . . . . . .32
Care and Maintenance
Care and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
Error Indicators and Troubleshooting Tips . . . . . . . . . . . . . . . . . . . . . . . . .35
FCC Statement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39

TABLE OF CONTENTS

3
Thank you for purchasing the Omron®7 series Plus BP762 Automatic Blood Pressure Monitor with ComFit
TM
Cuff.

INTRODUCTION

Fill in for future reference.
DATE PURCHASED: __________________ SERIAL NUMBER: ____________________
Your new blood pressure monitor uses the oscillometric method of blood pressure measurement. This means the monitor detects your blood’s movement through your brachial artery and converts the movements into a digital reading. An oscillometric monitor does not need a stethoscope so the monitor is simple to use.
The BP762 comes with the following components:
• Monitor • ComFit
TM
Cuff • AC Adapter
• Storage Case
• Instruction Manual printed in English and Spanish
• Quick Start Guide The 7 series Plus BP762 Automatic Blood Pressure Monitor is intended
for home use.
Please read this instruction manual thoroughly before using the unit. Please keep for future reference. For specific information about your own blood pressure, CONSULT YOUR DOCTOR.
SAVE THESE INSTRUCTIONS
Before Using the Monitor
• Staple your purchase receipt here
• Register your product on-line at www.register-omron.com
4
To assure the correct use of the product basic safety measures should always be followed including the warnings and cautions listed in this instruction manual.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property.

OPERATING THE DEVICE

Contact your physician for specific information about your blood pressure. Self-diagnosis and treatment using measured results may be dangerous. Follow the instructions of your physician or licensed healthcare provider.
DO NOT adjust medication based on measurement results from this blood pressure monitor. Take medication as prescribed by your physician. Only a physician is qualified to diagnose and treat High Blood Pressure.
The monitor is not intended to be a diagnostic device. If battery fluid should get in your eyes, immediately rinse with
plenty of clean water. Contact a physician immediately. Read all of the information in the instruction manual and any other
literature in the box before operating the unit.

SAFETY INFORMATION

OPERATING THE DEVICE (continued)
This device is intended for use in measuring blood pressure and pulse rate in the adult population. Do not use this device on infants or persons who cannot express their intentions.
Read the Special Conditions section (page 29) of the instruction manual if your systolic pressure is known to be more than 220 mmHg. Inflating to a higher pressure than necessary may result in bruising where the cuff is applied.
Operate the device only as intended. Do not use the device for any other purpose.
Dispose of the device, components and optional accessories according to applicable local regulations. Unlawful disposal may cause environmental pollution.
Do not use a cellular phone near the device. It may result in an operational failure.
Use only Omron authorized parts and accessories. Parts and accessories not approved for use with the device may damage the unit.
Use only 1.5V alkaline batteries with this device. Do not use other types of batteries. This may damage the unit.
SAFETY INFORMATION
5

RISK OF ELECTRICAL SHOCK

Do not plug or unplug the power cord into the electrical outlet with wet hands.
Do not overload power outlets. Plug the device into the appropriate voltage outlet.

CARE AND MAINTENANCE

Do not subject the monitor to strong shocks, such as dropping the unit on the floor.
Do not submerge the device or any of the components in water. Store the device and the components in a clean, safe location. Changes or modification not approved by Omron Healthcare will
void the user warranty. Do not disassemble or attempt to repair the unit or components.
SAFETY INFORMATION
6
7

BEFORE TAKING A MEASUREMENT

To ensure a reliable reading follow these recommendations:
1. Avoid eating, drinking alcohol, smoking, exercising, and bathing for 30 minutes prior to taking a measurement. Rest for at least 15 minutes prior to taking the measurement.
2. Stress raises blood pressure. Avoid taking measurements during stressful times.
3. Measurements should be taken in a quiet place.
4. Remove tight-fitting clothing from your arm.
5. Sit on a chair with your feet flat on the floor. Rest your arm on a table so that the cuff is at the same level as your heart.
6. Remain still and do not talk during the measurement.
7. Wait 2-3 minutes between measurements. The wait time allows the arteries to return to the condition prior to taking the blood pressure measurement. You may need to increase the wait time depending on your individual physiological characteristics.
8. Keep a record of your blood pressure and pulse readings for your physician. A single measurement may not provide an accurate indication of your true blood pressure. You need to take and record several readings over a period of time. Try to measure your blood pressure at about the same time each day for consistency.
8

KNOW YOUR UNIT

Main Unit:
Air Jack
Battery Compartment
Start/Stop Button
USER ID Selection Switch
Memory ( ) Button
Display
AC Adapter Jack
Operating Instructions
Up/Down (
/
) Buttons
SET Button
KNOW YOUR UNIT
9
Components:
Arm Cuff
Air Tube
Air Plug (Connects to Air Jack)
AC Adapter
Storage Case
Housing
Cord
Quick Start Guide
Instruction Manual
INSTRUCTION MANUAL
7 series Plus
Blood Pressure Monitor
TM
with ComFit
Cuff
Model
BP762
ESPAÑOL
ENGLISH
10

UNIT DISPLAY

Diastolic Blood Pressure
Systolic Blood Pressure
Movement Symbol
Pulse Display / Memory Number
Date/Time Display
Low Battery Indicator
Deflation Symbol
Heartbeat Symbol
Average Value Symbol
Memory Symbol
USER ID Symbol (A/B)
Blood Pressure Level Indicator
Cuff Wrap Guide
Irregular Heartbeat Symbol
11
IRREGULAR HEARTBEAT SYMBOL ( )
When the monitor detects an irregular rhythm two or more times during the measurement, the Irregular Heartbeat Symbol ( ) will appear on the display with the measurement values.
An irregular heartbeat rhythm is defined as a rhythm that varies by less than 25% from the average rhythm or 25% more than the average rhythm detected while the monitor is measuring the systolic and diastolic blood pressure.
If the irregular heartbeat symbol ( ) displays with your measurement results, we recommend you consult your physician. Follow the directions of your doctor.
MOVEMENT SYMBOL ( )
The Movement Symbol is displayed if you move your body during the measurement. Please remove the arm cuff, and wait 2-3 minutes. Take another measurement, remain still during inflation.

DISPLAY SYMBOLS

Normal Heartbeat
Irregular Heartbeat
Pulse
Blood
pressure
Pulse
Blood
pressure
Short Long
Internationally recognized guidelines for upper limit of normal home blood pressure readings are as follows:
Home Guideline for Upper Limit of Normal BP
Systolic Blood Pressure 135 mmHg Diastolic Blood Pressure 85 mmHg This is a general guideline because blood pressure depends on
age and morbidity. A lower home blood pressure goal may be recommended for certain patients, including diabetic patients, pregnant women, and patients with renal failure.
CAUTION
The monitor is not intended to be a diagnostic device.
HEARTBEAT SYMBOL ( )
The Heartbeat Symbol flashes on the display at every heartbeat during the measurement.
AVERAGE VALUE SYMBOL ( )
The Average Value Symbol is displayed when you press the memory button. The most recent average reading appears on the display screen.
CUFF WRAPGUIDE SYMBOL ( / )
When the cuff is applied correctly, ( ) is displayed while taking a measurement or using the memory function. If the cuff is not applied correctly, ( ) is displayed.
DISPLAY SYMBOLS
12
IMPORTANT SAFETY INSTRUCTIONS
SAVE THESE INSTRUCTIONS
NOTES: • Use only the authorized Omron AC Adapter that came with
this monitor.
• We recommend you install the batteries even if the AC Adapter is used. If no batteries are installed, you may need to reset the date and time if the AC Adapter is disconnected. The measurement results will not be deleted.
CAUTION
Use only Omron authorized parts and accessories. Parts and accessories not approved for use with the device may damage the unit.

USING THE AC ADAPTER

13
- DANGER -
TO REDUCE THE RISK OF
FIRE OR ELECTRIC SHOCK,
CAREFULLY FOLLOW THESE INSTRUCTIONS.
For connection to a power supply not in the U.S.A., use an attachment plug adapter of the proper configuration for the power outlet.
The power unit (AC Adapter) is intended to be correctly oriented in a vertical or floor mount position.
TO CONNECT THE AC ADAPTER
Insert the AC Adapter Plug into
the AC Adapter Jack on the back side of the monitor as illustrated.
Plug the AC Adapter into a 120V
AC outlet (60 cycles).
TO DISCONNECT THE AC ADAPTER
Hold the Housing to remove the AC Adapter from the electrical
outlet.
Remove the AC Adapter plug from the monitor.
WARNING
Do not plug or unplug the power cord into the electrical outlet with wet hands.
CAUTION
Do not overload power outlets. Plug the device into the appropriate voltage outlet.
USING THE AC ADAPTER
14
Housing
Cord
1. Remove the battery cover by pulling the hook up while pushing it in the direction of the arrow.
2. Install 4 “AA” size batteries so the + (positive) and - (negative) polarities match the polarities of the battery compartment as indicated.
3. Replace the battery cover.

BATTERY INSTALLATION

15
Battery Cover
BATTERY REPLACEMENT
When the Low Battery Indicator appears on the display screen turn the monitor off and remove all the batteries. Replace with four new batteries at the same time. Longlife alkaline batteries are recommended.
CAUTION
If battery fluid should get in your eyes, immediately rinse with plenty of clean water. Contact a physician immediately.
NOTES: • Turn the unit off before replacing the batteries. If the
batteries are removed while the unit is still on, the date and time will be reset to that of the previous use. However, the measurement results are not deleted.
• When the batteries are replaced, you may need to reset the date and time. If “-:--” appears on the display, refer to “Setting the Date and Time”.
CAUTION
Dispose of the device, components and optional accessories according to applicable local regulations. Unlawful disposal may cause environmental pollution.
BATTERY INSTALLATION
16
Low Battery Indicator
17
Set the monitor to the current date and time before taking a measurement for the first time. If the date and time are not set correctly, the measurement values stored in the memory and the average values may not be correct.
If the date and time are not set, “-:--” appears during or after measurement.
The Blood Pressure Monitor automatically stores up to 60 measurement values per user with the dates and times.
The monitor can also calculate an average reading based on the three most recent measurements taken within 10 minutes of the last reading.
NOTE: • If only two readings are stored in the memory for the 10-
minute period, the average is based on the two readings.
• If one reading is stored in the memory for the 10-minute period, this reading is displayed as the average.

SETTING THE DATE AND TIME

1. TO START THE SETTING
Press the SET button to adjust the date and time. The year flashes on the display.
2. SETTING THE YEAR
The year can be set between 2010 and
2040. When the display reaches 2040, it will return to 2010.
Press the Up button ( ) to advance by one year.
Press the SET button to set the current year.
The month flashes on the display.
NOTES: • Press the Down button ( ) to decrease the date and time
values.
• Press and hold the Up or Down button to increase or decrease the date and time values faster.
SETTING THE DATE AND TIME
18
19
SETTING THE DATE AND TIME
3. SETTING THE MONTH
Press the Up button ( ) to advance by one month.
Press the SET button to set the current month.
The day flashes on the display.
4. SETTING THE DAY
Press the Up button ( ) to advance by one day.
Press the SET button to set the current day.
The hour flashes on the display.
20
5. SETTING THE HOUR
The time is set using AM or PM. Press the Up button ( ) to advance
by one hour. Press the SET button to set the current
hour.
The minute flashes on the display.
6. SETTING THE MINUTE
Press the Up button ( ) to advance by one minute. Press the SET button to set the minute setting.
7. TO TURN THE MONITOR OFF
Press the START/STOP button.
SETTING THE DATE AND TIME
CAUTION
Read all of the information in the instruction manual and any other literature in the box before operating the unit.
CAUTION
This device is intended for use in measuring blood pressure and pulse rate in the adult population. Do not use this device on infants or persons who cannot express their intentions.
APPLYING THE CUFF ON THE LEFT ARM
1. Make sure the air plug is securely inserted in the main unit.
2. Remove tight-fitting clothing from your upper left arm.

APPLYING THE ARM CUFF

21
22
3. Hold the thumb grip on the cuff securely with your right hand.
4. Turn the palm of your left hand upward.
5. Apply the cuff to your left upper arm so the blue stripe is on the inside of your arm and aligned with your middle finger. The air tube runs down the inside of your arm. The bottom of the cuff should be approximately 1/2" above your elbow.
APPLYING THE ARM CUFF
Thumb Grip
Cloth Fastener
1/2 inch (1 - 2 cm)
Air Tube
Blue Stripe
23
6. Wrap the cuff firmly in place around your arm using the cloth fastener.
APPLYING THE ARM CUFF
APPLYING THE CUFF ON THE RIGHT ARM
When taking a measurement using the right arm use this instruction for Step 5 on page 22.
Apply the cuff to your right upper arm so the thumb grip is centered on the inside of your inner arm. The bottom of the cuff should be approximately 1/2'' above your elbow.
NOTE: Be careful not to rest your arm on the air tube. This will restrict
the flow of air to the cuff.
APPLYING THE ARM CUFF
24
Thumb Grip
The monitor automatically determines your ideal inflation level. In rare circumstances when a higher inflation may be necessary, the monitor automatically re-inflates the cuff up to 30 mmHg higher than the initial inflation and restarts the measurement.
WARNING
Contact your physician for specific information about your blood pressure. Self-diagnosis and treatment using measured results may be dangerous. Follow the instructions of your physician or licensed healthcare provider.
CAUTION
Operate the device only as intended. Do not use the device for any other purpose.

CUFF WRAP GUIDE

This monitor checks whether the arm cuff is applied correctly during the inflation. When the cuff is applied correctly, is displayed while taking a measurement or using the memory function. If the cuff is not applied correctly, is displayed. If is displayed, refer to the section “Applying the Arm Cuff”, reapply the cuff and take another measurement.

TAKING A MEASUREMENT

25
1. Press the START/STOP button to turn the monitor on. All symbols appear on the display.
2. Select your USER ID (A or B).
3. Sit on a chair with your feet flat on the floor. Place your left arm on a table so the cuff is level with your heart.
TAKING A MEASUREMENT
26
4. Press the START/STOP button. The cuff starts to inflate automatically. As the cuff inflates, the monitor
automatically determines your ideal inflation level. This monitor detects the pulse during inflation. Do not move your arm and remain still until the entire measurement process is completed.
NOTE: To stop the inflation or measurement, press the START/STOP
button. The monitor will stop inflating, start deflating and will turn off.
5. Inflation stops automatically and the measurement is started. As the cuff deflates, decreasing numbers appear on the display.
The Heartbeat Symbol ( ) flashes at every heartbeat.
TAKING A MEASUREMENT
27
TAKING A MEASUREMENT
28
6. When the measurement is complete, the arm cuff completely deflates. Your blood pressure and pulse rate are displayed.
7. Press the START/STOP button to turn the monitor off.
NOTE: The monitor will automatically turn
off after two minutes.
WARNING
Contact your physician for specific information about your blood pressure. Self-diagnosis and treatment using measured results may be dangerous. Follow the instructions of your physician or licensed healthcare provider.
WARNING
DO NOT adjust medication based on measurement results from this blood pressure monitor. Take medication as prescribed by your physician. Only a physician is qualified to diagnose and treat High Blood Pressure.
WARNING
The monitor is not intended to be a diagnostic device.
SPECIAL CONDITIONS
CAUTION
Inflating to a higher pressure than necessary may result in bruising where the cuff is applied.
If your systolic pressure is known to be more than 220 mmHg, press and hold the START/STOP button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure.
NOTES: • Do not apply more pressure than necessary.
• The monitor will not inflate above 299 mmHg.
When the measurement is complete, the arm cuff completely deflates. Your blood pressure and pulse rate are displayed.
29
TAKING A MEASUREMENT
The monitor is designed to store the blood pressure and the pulse rate in the memory for two people (USER A and USER B) every time a measurement is completed.
The monitor automatically stores up to 60 sets of measurement values (blood pressure and pulse rate) for each user (A and B). When 60 sets of measurement values are stored, the oldest record is deleted to save the most recent values.

AVERAGING FUNCTION

The monitor calculates the average reading based on the three most recent sets of measurement values taken within 10 minutes of the most recent reading.
NOTES: • If two sets of measurement values are stored in the memory
for the 10 minutes period, the average is based on the two sets of measurement values. If one set of measurement values is stored, this is displayed as the average.
• When viewing the reading taken without setting the date and time, “-:--” is displayed instead of the date and time.

USING THE MEMORY FUNCTION

30

TO DISPLAY THE MEASUREMENT VALUES

1. Press the START/STOP button to turn the monitor on.
2. Select your USER ID (A or B).
3. Press the Memory button ( ). The Average Symbol ( ) appears on the display when the average
reading is shown.
4. Press the Down button ( ) to display the most recent set of measurement values on the screen. The values are displayed from the most recent to the oldest.
NOTES: • The date and time are alternately
displayed with the measurement values.
• The Memory number appears for a second before the pulse rate is displayed. The newest set is numbered “1”.
Press the Up or Down button to display the next set of values.
: To the older readings : To the more recent readings
NOTE: Press and hold the Up or Down button to display the values faster.
USING THE MEMORY FUNCTION
31
Alternating display
32
USING THE MEMORY FUNCTION

TO DELETE ALL VALUES STORED IN THE MEMORY

The values stored in the memory are deleted by USER ID. You cannot partially delete values stored in the memory. All values will be deleted.
1. Press the START/STOP button to turn the monitor on.
2. Select your USER ID (A or B).
3. While holding the Memory button ( ), press the START/STOP button down for more than 2 seconds until the display appears as it does in the following illustration.
33
To keep your blood pressure monitor in the best condition and protect the unit from damage, follow the directions listed below:
Keep the monitor in the storage case
when not in use. Make sure the AC adapter is placed under the main unit so that it does not damage the display. Avoid kinking or sharply bending the AC Adapter cord.
Do not forcefully bend the arm cuff or air tube. Do not fold tightly.
Clean the monitor with a soft dry cloth. Do not use any abrasive or volatile cleaners. Do not attempt to clean the cuff.
CAUTION
Do not submerge the device or any of the components in water. Do not subject the monitor to extreme hot or cold temperatures, humidity or direct sunlight.
CAUTION
Store the device and the components in a clean, safe location.
CAUTION
Do not subject the monitor to strong shocks, such as dropping the unit on the floor.
Remove the batteries if the unit will not be used for three months or longer. Always replace all the batteries with new ones at the same time.

CARE AND MAINTENANCE

Care and Maintenance

Use the unit consistent with the instruction provided in this manual.
CAUTION
Changes or modification not approved by Omron Healthcare will void the user warranty. Do not disassemble or attempt to repair the unit or components.
CAUTION
Use only Omron authorized parts and accessories. Parts and accessories not approved for use with the device may damage the unit.
CARE AND MAINTENANCE
34
35
ERROR INDICATORS

ERROR INDICATORS AND TROUBLESHOOTING TIPS

Refer to “Applying the Arm Cuff”.Cuff is not applied correctly.
Replace the four batteries. Refer to “Battery Installation”.
Batteries are worn
Air plug disconnected.
Insert the plug securely. Refer to “Applying the Arm Cuff”.
Arm cuff not applied correctly.
Apply the arm cuff correctly. Refer to “Applying the Arm Cuff”.
Air is leaking from the arm cuff.
Replace the cuff with the new one.
Cuff over or under inflated.
Repeat measurement. Remain still and do not talk during measurement. Refer to “Taking a Measurement”.
If “E2” appears repeatedly, inflate the cuff manually until it is 30 to 40 mmHg above your previous measurement result. Refer to “Taking a Measurement”.
The arm cuff was inflated above 299 mmHg when inflating the cuff manually.
Do not inflate the cuff above 299 mmHg. Refer to “Taking a Measurement”.
Movement during measurement.
Repeat measurement. Remain still and do not talk during measurement. Refer to “Taking a Measurement”.
Clothing is interfering with the arm cuff.
Remove any clothing interfering with the arm cuff. Refer to “Applying the Arm Cuff”.
Device error.
Contact Customer Service.
SYMBOL CAUSE CORRECTION
36
ERROR INDICATORS AND TROUBLESHOOTING TIPS
PROBLEM CAUSES AND SOLUTIONS
No power.
No display appears on the unit.
Replace all four batteries with new ones. Check the battery installation for proper placement of the battery polarities.
Measurement values appear too high or too low.
Blood pressure varies constantly. Many factors including stress, time of day, and how you wrap the cuff, may affect your blood pressure. Review the sections “Before Taking a Measurement” and “Taking a Measurement.”
TROUBLESHOOTING TIPS
37

FCC STATEMENT

NOTE:
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only) This product has been tested and found to comply with the limits for a Class B
digital device, pursuant to part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful
interference in a residential installation. The product generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If the product does cause harmful interference to radio or television reception, which can be determined by turning the product on and off, the user is encouraged to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the product and the receiver.
• Connect the product into an outlet on a circuit different from that to which
the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for Canada only) This digital apparatus does not exceed the Class B limits for radio noise
emissions from digital apparatus as set out in the interference-causing equipment standard entitled “Digital Apparatus”, ICES-003 of the Canadian Department of Communications.
Cet appareil numérique respecte les limites de bruits radioeléctriques applicables aux appareils numériques de Clase B prescrites dans la norme sur le materiel brouilleur: “Appareils Numériques”, ICES-003 édictée par le minister des communications.
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.

LIMITED WARRANTY

38
Your Omron®BP762 Automatic Blood Pressure Monitor, excluding the arm cuff, is warranted to be free from defects in materials and workmanship appearing within 5 years from the date of purchase, when used in accordance with the instructions provided with the monitor. The arm cuff is warranted to be free from defects in materials and workmanship appearing within one year from the date of purchase when the monitor is used in accordance with the instructions provided with the monitor. The above warranties extend only to the original retail purchaser. We will, at our option, repair or replace without charge any monitor or arm cuff covered by the above warranties. Repair or replacement is our only responsibility and your only remedy under the above warranties. To obtain warranty service contact Customer Service by calling 1-800-634-4350 for the address of the repair location and the return shipping and handling fee. Enclose the Proof of Purchase. Include a letter, with your name, address, phone number, and description of the specific problem. Pack the product carefully to prevent damage in transit. Because of possible loss in transit, we recommend insuring the product with return receipt requested.
THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY DISCLAIMS ANYOTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING IMPLIED WARRANTIES OF MERCHANTABILITYAND FITNESS FOR APARTICULAR PURPOSE. IMPLIED WARRANTIES AND OTHER TERMS THAT MAY BE IMPOSED BY LAW, IFANY, ARE LIMITED IN DURATION TO THE PERIOD OF THE ABOVE EXPRESS WARRANTY. OMRON SHALL NOT BE LIABLE FOR LOSS OFUSE OR ANY OTHER SPECIAL, INCIDENTAL, CONSEQUENTAIL OR INDIRECT COSTS, EXPENSES OR DAMAGES.
This warranty provides you with specific legal rights, and you may have other rights that vary by jurisdiction. Because of special local requirements, some of the above limitations and exclusions may not apply to you.
FOR CUSTOMER SERVICE
Call toll free: 1-800-634-4350
39
Model: BP762 (HEM-7220-ZCS) Display: LCD Digital Display Measurement Range: Pressure: 0 to 299 mmHg, Pulse: 40 to 180/min. Accuracy/Calibration: Pressure: ±3mmHg or 2% of reading
Pulse: ±5% of reading Inflation: Automatic by electric pump Deflation: Automatic pressure release valve Rapid Air Release: Automatic exhaust valve Measurement Method: Oscillometric method Power Source: 120V 60Hz AC adapter or 4 “AA” batteries Battery Life: Approx. 500 uses when used 6 times a day
with 4 new alkaline batteries Operating Temperature 50ºF to 104ºF (10ºC to 40ºC) /15 to 90% RH
/Humidity: Storage Temperature -4ºF to 140ºF (-20ºC to 60ºC) /10 to 95% RH
/Humidity/Air Pressure: /700 - 1060 hPa Main Unit Weight: Approximately 15 1/8 oz. (430 g)
not including batteries Main Unit Dimensions: Approx. 6 1/4" (l) × 4 7/8" (w) × 3 5/16" (h)
(158 mm × 124 mm × 86 mm) Cuff Size: Approximately 6" (w) (152 mm)
Cuff tube 23 5/8" (l) (600 mm) Cuff Circumference: Fits arm circumferences 9" to 17"
(220 mm to 420 mm) Memory: Up to 60 per user Contents: Main Unit, ComFit
TM
Cuff, AC Adapter,
Instruction Manual, Quick Start Guide, Storage Case
= Type B
NOTE: These specifications are subject to change without notice.

SPECIFICATIONS

Manufactured by OMRON HEALTHCARE Co., Ltd.
Made in China
Distributed by:
Omron Healthcare, Inc.
1200 Lakeside Drive Bannockburn, IL 60015 www.omronhealthcare.com
Copyright © 2010 Omron Healthcare, Inc.