ULTRA HIGH-PRESSURE TRANSDUCER
WITH HIGH OPERATING TEMPERATURE RANGE
mV/V Output 
0-5000 to 0-200,000 psi 
0-350 to 0-14,000 bar
1 bar = 14.5 psi 
1 kg/cm2 = 14.22 psi 
1 atmosphere = 14.7 psi = 29.93 
inHg = 760.2 mmHg = 1.014 bar
PX91 Series
Standard
U All Stainless Steel 
Construction
U High Operating 
Temperature, Up to  
204°C (400°F)
U Hermetically Sealed for 
Harsh Environments
OMEGA’s PX91 is a rugged,  
ultra-high pressure transducer 
series. Hermetically sealed, with 
all stainless steel construction, it is 
designed for use in test chambers. 
The PX91 is available in sealed 
gage pressures up to 200,000 psi.  
It has a high operating temperature  
up to 204°C (400°F). The pressure 
connection is via a high-pressure 
ferrule connection. 
Cable style.
PX91N0-10KSV, 
shown smaller 
than actual size.
To Order
 RANGE    MODEL NO. 
 CABLE STYLE, SEALED GAGE PRESSURE
 0 to 5000 psi  0 to 345 bar  PX91N0-5KSV  DP41-S, DP25B-S 
 0 to 10,000 psi  0 to 689 bar  PX91N0-10KSV  DP41-S, DP25B-S 
 0 to 15,000 psi  0 to 1034 bar  PX91N0-15KSV  DP41-S, DP25B-S* 
 0 to 20,000 psi  0 to 1379 bar  PX91N0-20KSV  DP41-S, DP25B-S* 
 0 to 30,000 psi  0 to 2068 bar PX91N0-30KSV DP41-S, DP25B-S* 
 0 to 35,000 psi  0 to 2413 bar PX91N0-35KSV DP41-S, DP25B-S* 
 0 to 50,000 psi  0 to 3447 bar PX91N0-50KSV DP41-S, DP25B-S* 
 0 to 60,000 psi  0 to 4137 bar PX91N0-60KSV DP41-S, DP25B-S* 
 0 to 80,000 psi  0 to 5516 bar PX91P0-80KSV DP41-S, DP25B-S* 
 0 to 100,000 psi  0 to 6895 bar PX91P0-100KSV DP41-S, DP25B-S* 
 0 to 120,000 psi  0 to 8274 bar PX91P0-120KSV DP41-S, DP25B-S* 
 0 to 150,000 psi 
 0 to 200,000 psi 
Comes complete with 5-point NIST traceable calibration.   
Metric ranges available – consult Engineering.
* 4-digit meter.  ** See omega.com for compatible meters.  
Ordering Examples: PX91P0-100KSV, 100,000 psi sealed gage model with 1-14 NS fitting 
and 3 m (10') cable. 
PX91N0-15KSV, 15,000 psi sealed gage model with 
 Visit omega.com/px91-mv for Pricing and Details
COMPATIBLE METERS**
0 to 10,342 bar 
0 to 13,790 bar
 PX91P0-150KSV DP41-S, DP25B-S* 
 PX91P0-200KSV DP41-S, DP25B-S*
1
⁄2-14 NPSM fitting and 3 m (10') cable.
B-287

AR TWORK/PRODUCT  ART/ OMEGADYNE/OD-PX91SERIES
HIGH-TEMPERATURE/HIGH-PRESSURE TRANSDUCER
13 
(0.50) 
19
(0.75) 
5,000-70,000 psi
(Standard)
32 
(1.25) 
25
(1.0) 
8 (0.315) DIA.
1/2-14 NPS
+ 0
150°
– 2
59°
+ 0 
– 0.5
23
(0.910) 
DIA.
Wiring Code 
Green
+ OUTPUT
White
– OUTPUT
Black
– INPUT
Red
+ INPUT
38 
(1.50) 
DIA.
3 m (10') 4-CONDUCTOR 
SHIELDED PTFE CABLE 18 
3/8-18 NPT
(0.69) 
N Style Pressure Port
45 HEX
(1.75) 
105
(4.13) 
P Style Pressure Port
44 (1.75)
HEX
PRESSURE TRANSDUCERS
SPECIALTY
Dimensions: mm (inch)
SPECIFICATIONS
mV/V Output
Excitation: 10 Vac/Vdc (15V max) 
Output: 1 mV/V ±0.5% 
Bridge Resistance: 350 Ω nominal 
Zero Balance: ±2.5% 
Linearity: 
 ≤70K psi: ±0.35% FSO  
 >70K psi: ±0.50% FSO 
Hysteresis: ±0.10% FSO 
Repeatability: ±0.10% FSO 
Operating Temp Range: 
-56 to 204°C (-70 to 400°F)
Compensated Temp Range: 
21 to 218°C (70 to 425°F)
Thermal Effects:  
 Span: 0.01% rdg/°F  
 Zero: 0.01% FSO/°F
Proof Pressure:  
 ≤70K psi: 120% of rated   
 pressure 
 >70K psi: 110% of rated    
 pressure
Burst Pressure:  
 ≤70K psi: 150% of rated   
 pressure 
 >70K psi: 120% of rated     
 pressure
80,000-200,000 psi
(Standard)
1-14 NS-2
Wetted Parts:  
 ≤70K psi: 17-4 PH stainless steel  
 >70K psi: 455 stainless steel
Pressure Port:  
 ≤70K psi: 
1
⁄2-14 NPSM  
 (use with 0.315" dia. 59° ferrule) 
 >70K psi: 1-14 NS-2 
 (use with 0.910" dia 150° ferrule)
Electrical Connection: 3 m (10') 
4-conductor shielded cable
Mating 
Connector 
PT06F10-6S
CUSTOM CONFIGURATIONS
 PX91  PRESSURE ELECTRICAL  RANGE 
 SERIES PORT CONNECTION  (psi)  UNITS OUTPUT  OPTIONS
  [ 1 ]  [ 2 ]      [ 3 ]  [ 4 ]  [ 5 ]  [ 6 ]
Ranges ≤70,000 psi
1
N = 
⁄2-14 NPSM (use with 
  0.315" dia. 59° ferrule)
Ranges >70,000 psi 
P= 1-14 NS-2 (use with 0.910" 
  dia. 150° ferrule)
5K 
10K
S = Sealed  
 gage
V = mV/V
15K 
20K 
30K 
35K
See omega.com for 
additional options.
50K
0 = 3 m (10') PTFE cable
60K 
80K 
100K 
120K 
150K 
200K
 To order a custom configuration:
 1. Select a pressure port
 2. Select electrical connection
 3. Select a pressure range
 4. Select pressure units
Metric ranges available – consult Engineering.
 5. Select output
To order extra cable on cable style models, add suffix “-(length in feet) FT” for additional cost per foot beyond the standard 3 m (10') length. 
Ordering Examples: PX91N0-50KSV-25FT, 50,000 psi sealed gage pressure range model with mV/V output, 1⁄2 -14 NPSM fitting and 
7.6 m (25') cable.  
PX91P0-100KSV, 100,000 psi sealed gage pressure range model with mV output, 1-14 NS-2 fitting and 3 m (10') cable. 