Omega Products PX91N0-35KSV User Manual

ULTRA HIGH-PRESSURE TRANSDUCER
WITH HIGH OPERATING TEMPERATURE RANGE
mV/V Output 0-5000 to 0-200,000 psi 0-350 to 0-14,000 bar
1 bar = 14.5 psi 1 kg/cm2 = 14.22 psi 1 atmosphere = 14.7 psi = 29.93 inHg = 760.2 mmHg = 1.014 bar
PX91 Series
Standard
U All Stainless Steel
Construction
U High Operating
Temperature, Up to 204°C (400°F)
U Hermetically Sealed for
Harsh Environments
Cable style.
PX91N0-10KSV, shown smaller than actual size.
To Order
RANGE MODEL NO. CABLE STYLE, SEALED GAGE PRESSURE
0 to 5000 psi 0 to 345 bar PX91N0-5KSV DP41-S, DP25B-S 0 to 10,000 psi 0 to 689 bar PX91N0-10KSV DP41-S, DP25B-S 0 to 15,000 psi 0 to 1034 bar PX91N0-15KSV DP41-S, DP25B-S* 0 to 20,000 psi 0 to 1379 bar PX91N0-20KSV DP41-S, DP25B-S* 0 to 30,000 psi 0 to 2068 bar PX91N0-30KSV DP41-S, DP25B-S* 0 to 35,000 psi 0 to 2413 bar PX91N0-35KSV DP41-S, DP25B-S* 0 to 50,000 psi 0 to 3447 bar PX91N0-50KSV DP41-S, DP25B-S* 0 to 60,000 psi 0 to 4137 bar PX91N0-60KSV DP41-S, DP25B-S* 0 to 80,000 psi 0 to 5516 bar PX91P0-80KSV DP41-S, DP25B-S* 0 to 100,000 psi 0 to 6895 bar PX91P0-100KSV DP41-S, DP25B-S* 0 to 120,000 psi 0 to 8274 bar PX91P0-120KSV DP41-S, DP25B-S* 0 to 150,000 psi 0 to 200,000 psi
Comes complete with 5-point NIST traceable calibration. Metric ranges available – consult Engineering.
* 4-digit meter. ** See omega.com for compatible meters. Ordering Examples: PX91P0-100KSV, 100,000 psi sealed gage model with 1-14 NS fitting
and 3 m (10') cable. PX91N0-15KSV, 15,000 psi sealed gage model with
Visit omega.com/px91-mv for Pricing and Details
COMPATIBLE METERS**
0 to 10,342 bar 0 to 13,790 bar
PX91P0-150KSV DP41-S, DP25B-S* PX91P0-200KSV DP41-S, DP25B-S*
1
2-14 NPSM fitting and 3 m (10') cable.
B-287
AR TWORK/PRODUCT ART/ OMEGADYNE/OD-PX91SERIES
PX91 Series
HIGH-TEMPERATURE/HIGH-PRESSURE TRANSDUCER
13
(0.50)
19
(0.75)
5,000-70,000 psi
(Standard)
32
(1.25)
25
(1.0)
8 (0.315) DIA.
1/2-14 NPS
+ 0
150°
– 2
59°
+ 0 – 0.5
23
(0.910)
DIA.
Wiring Code Green
+ OUTPUT
White
– OUTPUT
Black
– INPUT
Red
+ INPUT
38
(1.50)
DIA.
3 m (10') 4-CONDUCTOR SHIELDED PTFE CABLE 18
3/8-18 NPT
(0.69)
N Style Pressure Port
45 HEX
(1.75)
105
(4.13)
P Style Pressure Port
44 (1.75)
HEX
PRESSURE TRANSDUCERS
SPECIALTY
Dimensions: mm (inch)
SPECIFICATIONS
mV/V Output
Excitation: 10 Vac/Vdc (15V max) Output: 1 mV/V ±0.5% Bridge Resistance: 350 Ω nominal Zero Balance: ±2.5% Linearity:
70K psi: ±0.35% FSO >70K psi: ±0.50% FSO Hysteresis: ±0.10% FSO Repeatability: ±0.10% FSO Operating Temp Range:
-56 to 204°C (-70 to 400°F)
Compensated Temp Range:
21 to 218°C (70 to 425°F)
Thermal Effects: Span: 0.01% rdg/°F Zero: 0.01% FSO/°F
Proof Pressure: 70K psi: 120% of rated
pressure >70K psi: 110% of rated pressure
Burst Pressure: 70K psi: 150% of rated
pressure >70K psi: 120% of rated pressure
80,000-200,000 psi
(Standard)
1-14 NS-2
Wetted Parts: 70K psi: 17-4 PH stainless steel >70K psi: 455 stainless steel
Pressure Port: 70K psi:
1
2-14 NPSM (use with 0.315" dia. 59° ferrule) >70K psi: 1-14 NS-2 (use with 0.910" dia 150° ferrule)
Electrical Connection: 3 m (10') 4-conductor shielded cable
Mating Connector PT06F10-6S
CUSTOM CONFIGURATIONS
PX91 PRESSURE ELECTRICAL RANGE SERIES PORT CONNECTION (psi) UNITS OUTPUT OPTIONS
[ 1 ] [ 2 ] [ 3 ] [ 4 ] [ 5 ] [ 6 ]
Ranges 70,000 psi
1
N =
2-14 NPSM (use with
0.315" dia. 59° ferrule)
Ranges >70,000 psi P= 1-14 NS-2 (use with 0.910"
dia. 150° ferrule)
5K 10K
S = Sealed gage
V = mV/V
15K 20K 30K 35K
See omega.com for additional options.
50K
0 = 3 m (10') PTFE cable
60K 80K 100K 120K 150K 200K
To order a custom configuration:
1. Select a pressure port
2. Select electrical connection
3. Select a pressure range
4. Select pressure units
Metric ranges available – consult Engineering.
5. Select output
To order extra cable on cable style models, add suffix “-(length in feet) FT” for additional cost per foot beyond the standard 3 m (10') length. Ordering Examples: PX91N0-50KSV-25FT, 50,000 psi sealed gage pressure range model with mV/V output, 1⁄2 -14 NPSM fitting and
7.6 m (25') cable. PX91P0-100KSV, 100,000 psi sealed gage pressure range model with mV output, 1-14 NS-2 fitting and 3 m (10') cable.