ULTRA HIGH-PRESSURE TRANSDUCER
WITH HIGH OPERATING TEMPERATURE RANGE
mV/V Output
0-5000 to 0-200,000 psi
0-350 to 0-14,000 bar
1 bar = 14.5 psi
1 kg/cm2 = 14.22 psi
1 atmosphere = 14.7 psi = 29.93
inHg = 760.2 mmHg = 1.014 bar
PX91 Series
Standard
U All Stainless Steel
Construction
U High Operating
Temperature, Up to
204°C (400°F)
U Hermetically Sealed for
Harsh Environments
OMEGA’s PX91 is a rugged,
ultra-high pressure transducer
series. Hermetically sealed, with
all stainless steel construction, it is
designed for use in test chambers.
The PX91 is available in sealed
gage pressures up to 200,000 psi.
It has a high operating temperature
up to 204°C (400°F). The pressure
connection is via a high-pressure
ferrule connection.
Cable style.
PX91N0-10KSV,
shown smaller
than actual size.
To Order
RANGE MODEL NO.
CABLE STYLE, SEALED GAGE PRESSURE
0 to 5000 psi 0 to 345 bar PX91N0-5KSV DP41-S, DP25B-S
0 to 10,000 psi 0 to 689 bar PX91N0-10KSV DP41-S, DP25B-S
0 to 15,000 psi 0 to 1034 bar PX91N0-15KSV DP41-S, DP25B-S*
0 to 20,000 psi 0 to 1379 bar PX91N0-20KSV DP41-S, DP25B-S*
0 to 30,000 psi 0 to 2068 bar PX91N0-30KSV DP41-S, DP25B-S*
0 to 35,000 psi 0 to 2413 bar PX91N0-35KSV DP41-S, DP25B-S*
0 to 50,000 psi 0 to 3447 bar PX91N0-50KSV DP41-S, DP25B-S*
0 to 60,000 psi 0 to 4137 bar PX91N0-60KSV DP41-S, DP25B-S*
0 to 80,000 psi 0 to 5516 bar PX91P0-80KSV DP41-S, DP25B-S*
0 to 100,000 psi 0 to 6895 bar PX91P0-100KSV DP41-S, DP25B-S*
0 to 120,000 psi 0 to 8274 bar PX91P0-120KSV DP41-S, DP25B-S*
0 to 150,000 psi
0 to 200,000 psi
Comes complete with 5-point NIST traceable calibration.
Metric ranges available – consult Engineering.
* 4-digit meter. ** See omega.com for compatible meters.
Ordering Examples: PX91P0-100KSV, 100,000 psi sealed gage model with 1-14 NS fitting
and 3 m (10') cable.
PX91N0-15KSV, 15,000 psi sealed gage model with
Visit omega.com/px91-mv for Pricing and Details
COMPATIBLE METERS**
0 to 10,342 bar
0 to 13,790 bar
PX91P0-150KSV DP41-S, DP25B-S*
PX91P0-200KSV DP41-S, DP25B-S*
1
⁄2-14 NPSM fitting and 3 m (10') cable.
B-287
AR TWORK/PRODUCT ART/ OMEGADYNE/OD-PX91SERIES
HIGH-TEMPERATURE/HIGH-PRESSURE TRANSDUCER
13
(0.50)
19
(0.75)
5,000-70,000 psi
(Standard)
32
(1.25)
25
(1.0)
8 (0.315) DIA.
1/2-14 NPS
+ 0
150°
– 2
59°
+ 0
– 0.5
23
(0.910)
DIA.
Wiring Code
Green
+ OUTPUT
White
– OUTPUT
Black
– INPUT
Red
+ INPUT
38
(1.50)
DIA.
3 m (10') 4-CONDUCTOR
SHIELDED PTFE CABLE 18
3/8-18 NPT
(0.69)
N Style Pressure Port
45 HEX
(1.75)
105
(4.13)
P Style Pressure Port
44 (1.75)
HEX
PRESSURE TRANSDUCERS
SPECIALTY
Dimensions: mm (inch)
SPECIFICATIONS
mV/V Output
Excitation: 10 Vac/Vdc (15V max)
Output: 1 mV/V ±0.5%
Bridge Resistance: 350 Ω nominal
Zero Balance: ±2.5%
Linearity:
≤70K psi: ±0.35% FSO
>70K psi: ±0.50% FSO
Hysteresis: ±0.10% FSO
Repeatability: ±0.10% FSO
Operating Temp Range:
-56 to 204°C (-70 to 400°F)
Compensated Temp Range:
21 to 218°C (70 to 425°F)
Thermal Effects:
Span: 0.01% rdg/°F
Zero: 0.01% FSO/°F
Proof Pressure:
≤70K psi: 120% of rated
pressure
>70K psi: 110% of rated
pressure
Burst Pressure:
≤70K psi: 150% of rated
pressure
>70K psi: 120% of rated
pressure
80,000-200,000 psi
(Standard)
1-14 NS-2
Wetted Parts:
≤70K psi: 17-4 PH stainless steel
>70K psi: 455 stainless steel
Pressure Port:
≤70K psi:
1
⁄2-14 NPSM
(use with 0.315" dia. 59° ferrule)
>70K psi: 1-14 NS-2
(use with 0.910" dia 150° ferrule)
Electrical Connection: 3 m (10')
4-conductor shielded cable
Mating
Connector
PT06F10-6S
CUSTOM CONFIGURATIONS
PX91 PRESSURE ELECTRICAL RANGE
SERIES PORT CONNECTION (psi) UNITS OUTPUT OPTIONS
[ 1 ] [ 2 ] [ 3 ] [ 4 ] [ 5 ] [ 6 ]
Ranges ≤70,000 psi
1
N =
⁄2-14 NPSM (use with
0.315" dia. 59° ferrule)
Ranges >70,000 psi
P= 1-14 NS-2 (use with 0.910"
dia. 150° ferrule)
5K
10K
S = Sealed
gage
V = mV/V
15K
20K
30K
35K
See omega.com for
additional options.
50K
0 = 3 m (10') PTFE cable
60K
80K
100K
120K
150K
200K
To order a custom configuration:
1. Select a pressure port
2. Select electrical connection
3. Select a pressure range
4. Select pressure units
Metric ranges available – consult Engineering.
5. Select output
To order extra cable on cable style models, add suffix “-(length in feet) FT” for additional cost per foot beyond the standard 3 m (10') length.
Ordering Examples: PX91N0-50KSV-25FT, 50,000 psi sealed gage pressure range model with mV/V output, 1⁄2 -14 NPSM fitting and
7.6 m (25') cable.
PX91P0-100KSV, 100,000 psi sealed gage pressure range model with mV output, 1-14 NS-2 fitting and 3 m (10') cable.
B
B-288