Omega Products PX409-16HBV User Manual

HIGH ACCURACY TRANSDUCERS
Product Label
Hex 22 (0.87) AF
18 NPT
Pressure Adaptor
22
(0.88)
79 (3.1)
79 (3.1)
18 NPT
Pressure Adaptor
Hex 2.5 (1.0) AF
Conduit back end PXM409C
Integral cable back end PXM409
28
(1.1)
5
(0.2)
Twist lock back end PXM459
Vent with
porous plug
(gage units only)
mini DIN back end PXM419
5
(0.2)
M1 2x1
Ø 20.2
0.8
5
(0.2)
22
(0.88)
Vent with porous plug (gage units only)
Glass to medal sealed pins PXM449
5
MICRO-MACHINED SILICON FOR BAROMETRIC PRESSURE MEASUREMENT
Barometric Pressure Absolute Pressure Outputs: mV/V, 0 to 5V, 0 to 10V or 4 to 20 mA 0 to 32, 16 to 32 or 26 to 32 inHg 0 to 1100, 550 to 1100, 880 to 1100 hPa
PX409 Series
Standard
U High Accuracy
±0.08% BSL Includes Linearity, Hysteresis, and Repeatability
U Broad Temperature
Compensated Range
-18 to 85°C (0 to 185°F)
U Premium Temperature
Performance Span: ±0.5% Over Compensated Range
U 5-Point NIST Traceable
Calibration Included
U All Stainless Steel
Wetted Parts
U Fast Response Time U Solid State Reliability
and Stability
U 400% Proof Pressure
OMEGA’s PX409 series electronic barometers are high-accuracy barometers available in three ranges: 16 to 32 inHg, 26 to 32 inHg, or 0 to 32 inHg (absolute pressure) for laboratory or industrial test applications. Models with integral cables (PX409) are IP67 while
IP65 rated PX429-26BI
connector styles (PX419 and PX429) are IP65 rated. They feature all stainless steel construction, which makes them ideal for harsh industrial environments. Their high operating temperature, broad compensated range, and excellent temperature compensation allow
Metric threads available,
see omega.com/config
Dimensions: mm (inch)
B-133
Cable style
IP67 rated PX409-32BV
Twist-lock style
M12 Connector
also Available See
To Order Chart
IP65 rated PX419-16B5V
All shown actual size.
stable readings in applications with fluctuating temperatures commonly found where barometric pressure must be monitored. Models are available with millivolt, voltage or current outputs.
mini DIN style
HIGH ACCURACY, MICRO-MACHINED SILICON TRANSDUCERS
SPECIFICATIONS
Output:
Millivolt: 10 mV/V (100 mV @ 10 Vdc)
Amplified Voltage: 0 to 5 Vdc or 0 to 10 Vdc Current Loop: 4 to 20 mA
Power Requirements:
Millivolt: 5 to 10 Vdc (5 mA @ 10 Vdc)
Amplified Voltage:
0 to 5 Vdc Supply Voltage:
10 to 30 Vdc @ 10 mA
0 to 10 Vdc Supply Voltage:
15 to 30 Vdc @ 10 mA Current Loop: 9 to 30 Vdc [max loop res = (Vs-9) x 50]; [9 to 20 Vdc above 105°C (229°F)]
Accuracy (Combined Linearity, Hysteresis and Repeatability):
±0.08% BSL
Zero Balance: ±0.5% FS typical, 1% max Span Setting: ±0.5% FS typical,
1% max; calibrated in vertical direction with fitting down
Operating Temperature Range:
-45 to 121°C (-49 to 250°F) [-45 to 115°C (-49 to 240°F) for voltage or current outputs]
Compensated Temperature: Ranges > 5 psi: -29 to 85°C (-20 to 185°F) Ranges 5 psi: -18 to 85°C (0 to 185°F)
Thermal Effects Span (Over Compensated Range):
Ranges > 5 psi: ±0.5% span Ranges 5 psi: ±1% span
Long Term Stability (1-Year): ±0.1% FS typical
Shock: 50 g, 11 mS half sine, vertical and horizontal axis
Vibration: 5-2000-5 Hz, 30 minute cycle, Curve L, Mil-Spec 810 figure 514-2-2, vertical and horizontal axis
Response Time: <1 ms Bandwidth: DC to 1 kHz typical Proof Pressure: 400% of span Burst Pressure: 6000 psia Electrical Termination: PX409: Integral 2 m (6') cable
PX419: mini DIN PX429: Twist-lock
Mating Connectors: PX419: CX5302 (included)
PX429: PT06F10-6S
(sold separately)
Environmental Protection: PX409: IP67
PX419: IP65 PX429: IP65 PX459: IP65
Wetted Parts: 316 SS Pressure Port: Weight: 115 to 200 g (4 to 7 oz)
depending upon configuration
PX419-26B5V
1
4-18 NPT male
To Order Visit omega.com/px409_bar for Pricing and Details
CALIBRATED RANGE
BAROMETRIC RANGES (Absolute Pressure)
0 to 32 inHg PX4[*]9-32BV PX4[*]9-32B5V PX4[*]9-32BI
16 to 32 inHg PX4[*]9-16BV PX4[*]9-16B5V PX4[*]9-16BI
26 to 32 inHg PX4[*]9-26BV PX4[*]9-26B5V PX4[*]9-26BI
0 to 1100 hPa PX4[*]9-32HBV PX4[*]9-32HB5V PX4[*]9-32HBI
550 to 1100 hPa PX4[*]9-16HBV PX4[*]9-16HB5V PX4[*]9-16HBI
880 to 1100 hPa PX4[*]9-26HBV PX4[*]9-26HB5V PX4[*]9-26HBI
ACCESSORIES
MODEL NO. DESCRIPTION
PT06F10-6S Mating connector for PX429 series, sold separately CX5302 Spare mating connector for PX419 series; one included with each unit
M12C-PVC-4-S-F-5 PVC cable, straight 4-pin M12 female connector one end, flying leads one end, 5 m (16.4') long, fits PX459
M12C-PVC-4-R-F-5 PVC cable, right angled 4-pin M12 female connector one end, flying leads one end, 5 m (16.4') long, fits PX459
-MB Mounting bracket, factory installed
** For 0 to 10 Vdc output replace “5V” with “10V” in model number. [*] Select Electrical Termination (no extra charge).
Insert “0” for 2 m (6') cable with flying leads. For ½ NPT conduit connector order PX409CS. Insert “1” for Mini DIN connector (mating connector included). Insert “2” for Twist-lock connector (mating connector PT06F10-6S sold separately). Insert “5” for M12 connector, M12 cables sold separately. Custom configurations also available, see omega.com/pxconfig
Ordering Examples: PX409-32BI, cable termination, 0 to 32 inHg barometric range, 4 to 20 mA output. PX459-32HB5V, M12 connector, 0 to 1100 hPa barometric range, 0 to 5 Vdc output. PX419-16HBV mini DIN termination, 550 to 1100 hPa barometric range, mV/V output.
mV/V
OUTPUT
B-134
5 Vdc**
OUTPUT
4 to 20 mA
OUTPUTinHg hPa
mini DIN style
PRESSURE TRANSDUCERS
VOLTAGE OUTPUT
B
Loading...