Omega Products FMA-A2105 User Manual

GAS MASS FLOWMETERS AND CONTROLLERS
For Clean Gases With Optional Integral Display
FMA-A2000 Series
Standard
U ±1% Accuracy U Linear Analog Output U Low Cost U Thermal Technology U For Flow Rates Up to 100 SLM U Power Supply Included
FMA-A2117 without display, shown smaller than actual size.
FMA-A2417 with display,
shown smaller than actual size.
The FMA-A2000 Series electronic mass flowmeters/ controllers provide high performance, versatility, and state-of-the-art design in one compact package. The FMA-A2000 uses capillary-type thermal technology to directly measure mass flow of gases. No temperature, pressure, or square root corrections are required. The FMA-A2300/2400 Series comes with an LCD display, and all models have linear 0 to 5 Vdc and 4 to 20 mA output.
The FMA-A2000 measures the mass flow rate of gases in 24 ranges from 0 to 100 SLM. For the complete listing, refer to the range table on page D-20. The FMA-A2000 Series is compatible with most non-corrosive gases. The user is advised to check the wetted materials against gas compatibility.
OMEGA’s mass flow controllers use an internal electromagnetic proportional valve to control the mass flow rate. A command signal can be supplied either by the internal setpoint pot or by a 0 to 5 Vdc external source.
D-19
SPECIFICATIONS
154
(6.05)
(0.687)
(1.50)
1/4 FNPT
19 (0.75)
17
38
FLOW
INLET
OUTLET
154
(6.05)
(0.687)
(1.50)
1/4 FNPT
19 (0.75)
(3.50) (3.80)
(2.450)
(0.25)
(0.25)
0.10 Dia. x 0.25 DP
2 PLCS
(1.00)
17
38
25
62
6
6
89
97
Accuracy: ±1% of full scale including linearity Repeatability: ±0.15% full scale or better
Turndown Ratio: 100:1 Response Time: 5 seconds Gas Ambient Temperature:
0 to 50°C (32 to 122°F) Temperature Coefficient:
0.05% full scale per 1°C or better Pressure Coefficient:
0.01% full scale per psi or better
Minimum Pressure Requirements: Meters: 1 inch of H Controllers Up to 1 SLM: 1 psi Controllers Greater than 20 SLM: 20 psi Differential Pressure for Mass Flow Controllers:
<5 LPM, 5 to 50 psi standard >5 LPM, 20 to 40 psi standard Maximum Gas Pressure: Polyacetal: 250 psig 316 SS: 500 psig Leak Integrity:
-6
cc/min of He or better
1 x10
Wetted Parts: Standard: Polyacetal, FKM O-rings Optional: 316 SS, FKM O-rings Connections:
1
⁄4" compression fitting for all stainless steel models Input/Output Connection: 9-pin sub-D connector
Control Signal: 0 to 5 Vdc Control Valve: Electromagnetic
normally closed Output Signal: Linear 0 to 5 Vdc, 1000 Ω min load, 4 to 20 mA, 500 Ω loop resistance
Mounting Threads:
316 SS Body: Polyacetal: Self tap #4/6 screw
Input Power: Meters: 24 Vdc @ 150 mA,
115 Vac, 220 Vac with AC adaptor Controller: 24 Vdc @ 300 mA, 115 Vac, 220 Vac with AC adaptor
Dimensions: See drawing Weight: 1.25 kg (2.75 lb)
with power supply
2
O
1
⁄4 FNPT standard;
6
" thread
32
Range Table
Code SCCM Code SLM Code SLM 00 0 to 10 06 0 to 1 14 0 to 35
01 0 to 20 07 0 to 2 15 0 to 40 02 0 to 50 08 0 to 5 16 0 to 45 03 0 to 100 09 0 to 10 17 0 to 50 04 0 to 200 10 0 to 15 18 05 0 to 500 11 0 to 20 19 — 12 0 to 25 20 — 13 0 to 30 21 — 22 — 23
FMA-A2100/2300 Series only
*
Dimensions: mm (inch)
0 to 55 0 to 60 0 to 65 0 to 70 0 to 75 0 to100
* * * * *
*
To Order Visit omega.com/fmaa2100_2200_2300_2400 for Pricing and Details
Model No. Description FMA-A21(*) Mass flowmeter without display, ranges 0 to 10 SCCM to
FMA-A22(*)-SS-(**) Mass flow controller without display, with control valve,
FMA-A23(*) Mass flowmeter with display, ranges 0 to 10 SCCM to
FMA-A24(*)-SS-(**) Mass flow controller with display and control valve, ranges
Compatible Meters; DPF50, DPF60, DPF300, DPF5500, DPF403. Visit omega.com for more details.
0 to 100 SLM
ranges 0 to 10 SCCM to 0 to 50 SLM, 316 SS body
0 to 100 SLM
0 to 10 SCCM
to 0 to 50 SLM, 316 SS body
D
Tubing and fittings of various materials and styles are available. Visit omega.com
Accessories
Model No. Description FMA-200PWA Replacement socket plug-in power supply for 90 to 260 Vac SSLK-14-14 316 SS compression fitting ¹⁄₄ NPT for ¹⁄₄" OD tube
Comes complete with operator’s manual, 9-pin sub-D connector, power supply and NIST calibration (no data points).
Insert range code from table at top of page to complete model number.
*
For optional 316 stainless steel body, add suffix “-SS” to model number for additional cost
1
(
" compression fittings standard). SS body standard on FMA-A22 and FMA-A24 controllers.
4
Note: All controller flow ranges specified are for nitrogen or air at 20 psig; when used for other gases, a correlation factor is needed to determine the flow rate.
Specify gas, inlet/outlet pressure, temperature.
**
For 5-point NIST calibration certificate, add suffix “-NIST5” to model number, no additional charge. For 10-point NIST calibration certificate, add suffix “-NIST10” to model number for additional cost. Ordering Examples: FMA-A2316, mass flowmeter with display, calibrated for nitrogen at
20 psig inlet, room temperature for 0 to 45 SLM. P: 5 to 20 psi. FMA-A2102-SS, 316 stainless steel body mass flowmeter without display, 0 to 50 SCCM.
D-20
Loading...