
GAS MASS FLOWMETERS  
AND CONTROLLERS
For Clean Gases 
With Optional Integral Display
FMA-A2000 Series
Standard
U ±1% Accuracy 
U Linear Analog Output 
U Low Cost 
U Thermal Technology 
U For Flow Rates Up to 100 SLM 
U Power Supply Included
FMA-A2117 without display,  
shown smaller than actual size.
FMA-A2417 with display, 
shown smaller than actual size.
The FMA-A2000 Series electronic mass flowmeters/ 
controllers provide high performance, versatility, and 
state-of-the-art design in one compact package. The 
FMA-A2000 uses capillary-type thermal technology to 
directly measure mass flow of gases. No temperature, 
pressure, or square root corrections are required. 
The FMA-A2300/2400 Series comes with an LCD  
display, and all models have linear 0 to 5 Vdc and  
4 to 20 mA output. 
The FMA-A2000 measures the mass flow rate of 
gases in 24 ranges from 0 to 100 SLM. For the  
complete listing, refer to the range table on page 
D-20. The FMA-A2000 Series is compatible with most 
non-corrosive gases. The user is advised to check the 
wetted materials against gas compatibility.
OMEGA’s mass flow controllers use an internal  
electromagnetic proportional valve to control the  
mass flow rate. A command signal can be supplied 
either by the internal setpoint pot or by a 0 to 5 Vdc 
external source.
D-19

SPECIFICATIONS
154
(6.05)
(0.687)
(1.50)
1/4 FNPT
19 (0.75)
17
38
FLOW
INLET
OUTLET
154
(6.05)
(0.687)
(1.50)
1/4 FNPT
19 (0.75)
(3.50) 
(3.80)
(2.450)
(0.25)
(0.25)
0.10 Dia. x 0.25 DP
2 PLCS
(1.00)
17
38
25
62
6
6
89
97
Accuracy: ±1% of full scale  
including linearity 
Repeatability: ±0.15% full scale  
or better
Turndown Ratio: 100:1 
Response Time: 5 seconds 
Gas Ambient Temperature: 
0 to 50°C (32 to 122°F) 
Temperature Coefficient: 
0.05% full scale per 1°C or better 
Pressure Coefficient: 
0.01% full scale per psi or better
Minimum Pressure Requirements: 
 Meters: 1 inch of H 
 Controllers Up to 1 SLM: 1 psi 
 Controllers Greater than 
 20 SLM: 20 psi 
Differential Pressure for  
Mass Flow Controllers: 
<5 LPM, 5 to 50 psi standard 
>5 LPM, 20 to 40 psi standard 
Maximum Gas Pressure:  
 Polyacetal: 250 psig 
 316 SS: 500 psig 
Leak Integrity: 
-6
 cc/min of He or better
1 x10
Wetted Parts:  
 Standard: Polyacetal, FKM O-rings  
 Optional: 316 SS, FKM O-rings 
Connections: 
1
⁄4" compression fitting for all  
stainless steel models  
Input/Output Connection:  
9-pin sub-D connector
Control Signal: 0 to 5 Vdc 
Control Valve: Electromagnetic 
normally closed 
Output Signal: Linear 0 to 5 Vdc,  
1000 Ω min load, 4 to 20 mA,  
500 Ω loop resistance
Mounting Threads:
 316 SS Body: 
 Polyacetal: Self tap #4/6 screw
Input Power: 
  Meters: 24 Vdc @ 150 mA, 
115 Vac, 220 Vac with AC adaptor 
 Controller: 24 Vdc @ 300 mA,  
 115 Vac, 220 Vac with AC adaptor
Dimensions: See drawing 
Weight: 1.25 kg (2.75 lb) 
with power supply
2
O
1
⁄4 FNPT standard; 
6
" thread
⁄
32
Range Table
 Code  SCCM  Code  SLM  Code  SLM 
 00  0 to 10  06  0 to 1  14  0 to 35
 01  0 to 20  07  0 to 2  15  0 to 40 
 02  0 to 50  08  0 to 5  16  0 to 45 
 03  0 to 100  09  0 to 10  17  0 to 50 
 04  0 to 200  10  0 to 15  18 
 05  0 to 500  11  0 to 20  19 
 —  —  12  0 to 25  20 
 —  —  13  0 to 30  21 
 —  —  —  —  22 
 —  —  —  —  23
 FMA-A2100/2300 Series only
*
Dimensions: mm (inch)
  0 to 55 
  0 to 60 
  0 to 65 
  0 to 70 
  0 to 75 
  0 to100
* 
* 
* 
* 
*
*
 To Order Visit omega.com/fmaa2100_2200_2300_2400 for Pricing and Details
 Model No.  Description 
 FMA-A21(*)   Mass flowmeter without display, ranges 0 to 10 SCCM to 
 FMA-A22(*)-SS-(**)  Mass flow controller without display, with control valve, 
 FMA-A23(*)   Mass flowmeter with display, ranges 0 to 10 SCCM to 
 FMA-A24(*)-SS-(**)  Mass flow controller with display and control valve, ranges 
Compatible Meters; DPF50, DPF60, DPF300, DPF5500, DPF403. Visit omega.com for  
more details.
0 to 100 SLM
ranges 0 to 10 SCCM to 0 to 50 SLM, 316 SS body
0 to 100 SLM
0 to 10 SCCM 
to 0 to 50 SLM, 316 SS body
D
Tubing and fittings of various 
materials and styles are available. 
Visit omega.com
Accessories
 Model No.  Description 
 FMA-200PWA Replacement socket plug-in power supply for 90 to 260 Vac 
 SSLK-14-14  316 SS compression fitting ¹⁄₄ NPT for ¹⁄₄" OD tube
Comes complete with operator’s manual, 9-pin sub-D connector, power supply and  
NIST calibration (no data points).
 Insert range code from table at top of page to complete model number. 
*
For optional 316 stainless steel body, add suffix “-SS” to model number for additional cost 
1
(
" compression fittings standard). SS body standard on FMA-A22 and FMA-A24 controllers.
⁄4
Note: All controller flow ranges specified are for nitrogen or air at 20 psig; when used for other 
gases, a correlation factor is needed to determine the flow rate.
 Specify gas, inlet/outlet pressure, temperature.
**
For 5-point NIST calibration certificate, add suffix “-NIST5” to model number, no additional charge. 
For 10-point NIST calibration certificate, add suffix “-NIST10” to model number for additional cost. 
Ordering Examples: FMA-A2316, mass flowmeter with display, calibrated for nitrogen at 
20 psig inlet, room temperature for 0 to 45 SLM. ∆P: 5 to 20 psi. 
FMA-A2102-SS, 316 stainless steel body mass flowmeter without display, 0 to 50 SCCM.
D-20