ECONOMICAL GAS
MASS FLOWMETERS
For Clean Gases With
Optional Integral Display
FMA1700/1800 Series
Standard
U Reads Gas Mass Flow
Without Temperature
or Pressure Compensation
U Available in Economical
Aluminum or CorrosionResistant 316SS
U Tiltable LCD Display
for Easy Reading
U NIST Traceable Calibration
FMA1700/1800 Series are ideal
substitutes for many rotameter
The FMA1700/1800 Series
eletronic gas mass flowmeters provide
for monitoring the flow of a wide
variety of gases from 10 SCCM up
to 1000 SLM. Utilizing heat transfer
through a heated tube to
measure molecular gas flow rate,
the FMA1700/1800 provides
measurement of direct gas mass flow
rate, without the need to compensate
for variations in gas temperature or
pressure (within stated limits). They
are available in an economical
aluminum/brass construction for
typical gas flows and a 316 SS
construction for applications
requiring more corrosion resistance.
The FMA1700 series without integral
display is supplied with a field
selectable analog 0 to 5 Vdc or
4 to 20 mA output for remote
monitoring; the FMA1800 series has
both an integral 31⁄2 digit display
and analog output. The display is
tiltable over 90 degrees for viewing
convenience. The display is calibrated
standard to read out directly in
SCCM or SLM for nitrogen (other
gas calibrations available on
special order).
Due to their low cost, digital display,
analog output capability, and
insensitivity to variations in gas
temperature and pressure, the
applications. When used with the
FMA178BP portable battery pack,
the FMA1800 Series are ideal for
in-the-field calibration of flowmeters
or testing of air sampling equipment.
The FMA178BP battery pack
includes batteries, recharger, and
carrying case with shoulder strap
and belt loop. With the FMA178BP,
the flowmeter can operate in excess
of 40 hours; the batteries can be
recharged at least 200 times.
The FMA1700/1800 Series require
12 Vdc power @ 200 mA
maximum, which can be supplied by
the FMA178PW plug-in socket power
supply. The electronics are reverse
polarity protected and has externallyaccessible fusing. Model number
FMA178C (supplied separately)
provides a mating 9-pin “D”
connector with 3 feet of ribbon cable
for accessing the 0 to 5 Vdc output
signal and power input connections
(use FMA178C-MA for 4 to 20 mA
output signal connection).
The LCD supplied with the FMA1800
Series is connected to the lower
electronics via a modular plug. The
LCD can be remotely located by
purchasing an FMA18RC remote
cable assembly-you must then build
your own assembly for panel
mounting the LCD.
FMA1828,
shown smaller
than actual size.
SPECIFICATIONS
Accuracy: ±1.5% of full scale,
including linearity over 15 to 25°C and
5 to 60 psia (0.35 to 4.2 kg/cm
±3% of full scale, including linearity over
0 to 50°C and 1 to 500 psia
(0.07 to 10 kg/cm
Repeatability: ±0.5% of full scale
and for units ≥100 scm from 0 to 20%
of range
Temp Coefficient: 0.15% of full scale
per °C or better
Pressure Coefficient: 0.01% of full
scale per psi (0.07 bar)
Maximum Pressure Drop:
SLM in. W.C. SLM in W.C.
10 1 80 168
20 14 100 227
30 34 200 112
50 90 500 140
60 129 1000 252
Response Time: 800 msec time
constant; 2 seconds (typical) to within
±2% of set flow rate over 25 to 100% of
full scale
Max Gas Pressure: 500 psig
(35 kg/cm
Gas and Ambient Temp: 0 to 50°C
(0 to 122°F)
Leak Integrity: 1 x 10
helium max. to outside environment
Ma terials in Fluid Contact:
Aluminum Models: Anodized
aluminum, 316 SS, brass and
FKM O-rings
Stainless Steel Models: 316 SS
and FKM O-rings
2
2
)
gauge); 20 psig optimum
7
-
2
);
std cc/sec of
D-11
Dimensions: cm (in)
Output Signal
Linear 0 to 5 Vdc: 1000 Ω min load;
4 to 20 mA: 50 to 250 Ω loop resis
Transducer Power: 12 Vdc @
200 mA max.
Shipping Weight: 1.1 kg (2.5 lb)
Compliance: EN55011 class 1, class B;
EN50082-1
Connection Unit Max Lay Length Maximum Maximum Compression
Flow Rate with Fittings Height Width Fitting
10 SCCM to 10 SLM 12.8 (5.02) 14.2 (5.60) 2.5 (1.00) ¹⁄₄"
15 to 50 SLM 15.6 (6.15) 15.2 (5.98) 3.2 (1.25) ¹⁄₄"
60 to 100 SLM 15.9 (6.27) 15.2 (5.98) 3.2 (1.25) ³⁄₈"
200 SLM 22.4 (8.83) 16.8 (6.60) 4.4 (1.75) ³⁄₈"
500 SLM 24.6 (9.67) 19.3 (7.60) 7.6 (3.00) ¹⁄₂"
1000 SLM 18.5 (7.30) 21.8 (8.60) 10.2 (4.00) ³⁄₄ FNPT
To Order Visit omega.com/fma1700_1800 for Pricing and Details
Model No. Model No. Model No. Model No.
Aluminum/Brass Stainless Steel Aluminum/Brass Stainless Steel Range Maximum
Body with Display Body with Display Body without Display Body without Display Code** Flow Rate
FMA1802 FMA1802ST FMA1702 FMA1702ST 02 10 sccm
FMA1804 FMA1804ST FMA1704 FMA1704ST 04 20 sccm
FMA1806 FMA1806ST FMA1706 FMA1706ST 06 50 sccm
FMA1808 FMA1808ST FMA1708 FMA1708ST 08 100 sccm
FMA1810 FMA1810ST FMA1710 FMA1710ST 10 200 sccm
FMA1812 FMA1812ST FMA1712 FMA1712ST 12 500 sccm
FMA1814 FMA1814ST FMA1714 FMA1714ST 14 1 SLM
FMA1816 FMA1816ST FMA1716 FMA1716ST 16 2 SLM
FMA1818 FMA1818ST FMA1718 FMA1718ST 18 5 SLM
FMA1820 FMA1820ST FMA1720 FMA1720ST 20 10 SLM
FMA1823 FMA1823ST FMA1723 FMA1723ST 23 15 SLM
FMA1824 FMA1824ST FMA1724 FMA1724ST 24 20 SLM
FMA1826 FMA1826ST FMA1726 FMA1726ST 26 30 SLM
FMA1827 FMA1827ST FMA1727 FMA1727ST 27 40 SLM
FMA1828 FMA1828ST FMA1728 FMA1728ST 28 50 SLM
FMA1840 FMA1840ST FMA1740 FMA1740ST 40 60 SLM
FMA1841 FMA1841ST FMA1741 FMA1741ST 41 80 SLM
FMA1842 FMA1842ST FMA1742 FMA1742ST 42 100 SLM
FMA1843 FMA1843ST FMA1743 FMA1743ST 43 200 SLM
FMA1844 FMA1844ST FMA1744 FMA1744ST 44 500 SLM
FMA1845 FMA1845ST FMA1745 FMA1745ST 45 1000 SLM
Comes with dual
*
Insert range code, see table above.
**
Flow ranges specified are for nitrogen or air at 20 psig. When used for other gases, a multiplication factor is used to determine the flow rate,
and the digital display must be rescaled in the field. To request a custom calibration add the gas abbreviation and pressure as a suffix to the
model number.
3
4
⁄
FNPT connections instead of compression fittings.
D
*
Accessories
Model No. Description
FMA178PW Socket plug-in power supply for 115 Vac (recommended)
FMA178C Female 9-pin D-connector with 1 m (3') of ribbon cable, 0 to 5 Vdc output
FMA178C-MA Female 9-pin D-connector with 1 m (3') of ribbon cable, 4 to 20 mA output
FMA178PW-220VAC Socket plug-in power supply for 220 Vac, VDE plug type
FMA178BP Portable battery pack, with recharger for 115 Vac
FMA178BP-220VAC Portable battery pack, with recharger for 220 Vac
FMA18RC10 3 m (10') cable for remotely locating LCD of FMA1800
FMA18RC25 7.6 m (25') cable for remotely locating LCD of FMA1800
Comes complete with compression fittings, operator’s manual, NIST certificate. Power supplies sold separately.
For 24 Vdc powered units, add suffix “-24VDC” to model number, no additional price.
For oxygen cleaned units, add suffix “-02CLEAN” to model number for additional price.
Ordering Examples: FMA1712, AL/BR body flowmeter without an integral display, calibrated for nitrogen at 20 psig, ambient temperature
from 0 to 500 SCCM.
FMA1810, AL/BR body with display and FMA178PW, power supply.
D-12