DRC-4720
Bridge Input Signal Conditioner
INSTRUCTION
SHEET
M4739/0109
Shop online at: omega.com e-mail: info@omega.com
For latest product manuals: omegamanual.info
Description:
The DRC-4720 is a DIN Rail mount, self-contained DC
powered module designed for load cell, strain gage, or
single ended use. The DRC-4720 contains a precision
instrumentation amplifier, filtered output and a builtin 10V excitation supply capable of driving a 350 ohm
bridge. The 0 to 30mV input range makes it compatible
with most strain gage based load cell or pressure
transducer outputs. The DRC-4720 provides an output of
4 to 20mA. Connections are made via easily accessible
screw clamp terminal blocks. Zero and Span adjustment
potentiometers are located externally as well for easy
access.
Amplifier
Gain
Input for a 20mA Output 10mV to 30mV
Linearity: 4 to 20mA Out ±0.01
Temperature Coefficient 0.05% / ºC
Input Offset Voltage
Temperature Coefficient
±70μV
0.7μV / ºC
Common Mode Voltage 0 to +5 Volts
Common Mode Rejection - DC 100 dB
Input Noise
0.1Hz to 10Hz
0.3μV pp Typ
Output
Output Range 4 to 20mA
Compliance Voltage 5.5V Max
Loop Resistance 250Ω Max
Frequency Response
2 Pole Filter
DC to 10Hz
Total RMS Gain
Temperature Coefficient
0.007% / ºC
Bridge Supply
Fixed Output +10V ±1.5%
Temperature Coefficient 0.05% / ºC
Load Current 30mA Max
Power Requirements
Voltage 18 to 26 Volts
Input Current (1 - 350 Ohm Bridge) 55 mA
Environment
Operating Temperature 0ºC to +55ºC
Storage Temperature -40ºC to +80ºC
Size
Size
(Not including Universal Foot for
standard DIN EN Rails)
1.65”H x 1.06”W x 3.78”L
(42 x 27 x 96mm)
Weight 3 oz. (85g)
Agency Approval
UL, CE UL508, UL1604, CE
Pending
Features:
Low Cost
Din Rail Mount
Load Cell or Single Ended Application
4-20mA
•
•
•
•
2
DRC-4720 BRIDGE INPUT SIGNAL CONDITIONER
FULL BRIDGE CONNECTION
SINGLE ENDED
1. Hook Up Procedure
A. Connect the +out of the load cell to the +INPUT, pin 3.
B. Connect the -out of the load cell to the -INPUT, pin 4.
C. Connect the +excitation of the load cell to +EXCITATION,
pin 1.
D. Connect the -excitation of the load cell to -EXCITATION,
pin 2.
E. Connect the +24 VDC power supply to +24V, pin 8 and
COM, pin 7.
2. Turn on Procedure
A. Verify that the hook up procedure is complete.
B. Turn on the +24 VDC power source connected to the
DRC-4720.
3. Calibration Procedure
A. Jumper the +INPUT and the -INPUT terminals, pins 3
and 4, together.
B. Check the Gain Switch Table, and set SW1-1 and SW1-2
to the expected full scale output of the load cell.
Getting Started with the Model DRC-4720
SW1-1 SW1-2
FULL SCALE
LOAD CELL INPUT
OFF OFF 30 mV
ON OFF 20 mV
ON ON 10 mV
C. Connect a current meter across the output, pins 5 and 6.
D. Adjust the Zero Adjustment potentiometer for the
desired zero current.
E. Remove the jumper from the +INPUT and -INPUT
terminals.
F. With no load on the load cell, readjust the zero output.
G. Apply a known load to the load cell; in most cases it
would be 100% of full scale.
H. Adjust the SPAN ADJUSTMENT potentiometer for the
desired full scale output current.
I. Repeat steps F thru H until the desired settings are
obtained.