DRC-4710
Bridge Input Signal Conditioner
INSTRUCTION
SHEET
M4738/0109
Shop online at: omega.com e-mail: info@omega.com
For latest product manuals: omegamanual.info
Description:
The DRC-4710 is a DIN Rail mount, self-contained DC
powered module designed for load cell, strain gage, or
single ended use. The DRC-4710 contains a precision
instrumentation amplifier, filtered output and a builtin 10V excitation supply capable of driving a 350 ohm
bridge. The 0 to 30mV input range makes it compatible
with most strain gage based load cell or pressure
transducer outputs. The DRC-4710 provides an output of
0 to 10Vdc. Connections are made via easily accessible
screw clamp terminal blocks. Zero and Span adjustment
potentiometers are located externally as well for easy
access.
Amplifier
Gain
Adjustment Range
Input for 10V Output
333 to 1000
10mV to 30mV
Linearity: 0 to 10V Out ±0.01%
Temperature Coefficient 0.05% / ºC
Input Offset Voltage
Temperature Coefficient
±70µV
0.7µV / ºC
Common Mode Voltage 0 to +5 Volts
Common Mode Rejection - DC 100 dB
Input Noise
0.1Hz to 10Hz
0.3µV pp Typ
Output
Output Range 0V to +10V
Load Current 5mA Max
Frequency Response
2 Pole Filter
DC to 10Hz
Total RMS Gain
Temperature Coefficient
0.007% / ºC
Bridge Supply
Fixed Output +10V ±1.5%
Temperature Coefficient 0.05% / ºC
Load Current 30mA Max
Power Requirements
Voltage 18 to 26 VDC
Input Current (1 - 350 Ohm Bridge) 40 mA
Environment
Operating Temperature 0ºC to +55ºC
Storage Temperature -40ºC to +80ºC
Size
Size
(Not including Universal Foot for
standard DIN EN Rails)
1.65”H x 1.06”W x 3.78”L
(42 x 27 x 96mm)
Weight 3 oz. (85g)
Agency Approval
UL, CE UL508, UL1604, CE
Pending
Features:
Low Cost
Din Rail Mount
Load Cell or Single Ended Application
0-10V
•
•
•
•
2
FULL BRIDGE CONNECTION
SINGLE ENDED
1. Hook Up Procedure
A. Connect the +out of the load cell to the +INPUT, pin 3.
B. Connect the -out of the load cell to the -INPUT, pin 4.
C. Connect the +excitation of the load cell to +EXCITATION,
pin 1.
D. Connect the -excitation of the load cell to -EXCITATION,
pin 2.
E. Connect the +24 VDC power supply to +24V, pin 8 and
COM, pin 7.
2. Turn on Procedure
A. Verify that the hook up procedure is complete.
B. Turn on the +24 VDC power source connected to the
DRC-4710.
3. Calibration Procedure
A. Jumper the +INPUT and the -INPUT terminals, pins 3
and 4, together.
B. Check the Gain Switch Table, and set SW1-1 and SW1-2
to the expected full scale output of the load cell.
Getting Started with the DRC-4710
SW1-1 SW1-2
FULL SCALE
LOAD CELL INPUT
OFF OFF 30 mV
ON OFF 20 mV
ON ON 10 mV
C. Connect a voltmeter across the output, pins 5 and 6.
D. Adjust the Zero Adjustment potentiometer for the
desired output.
E. Remove the jumper from the +INPUT and -INPUT
terminals.
F. With no load on the load cell, readjust the zero output.
G. Apply a known load to the load cell; in most cases it
would be 100% of full scale.
H. Adjust the SPAN ADJUSTMENT potentiometer for the
desired full scale output.
I. Repeat steps F thru H until the desired settings are
obtained.
DRC-4710 BRIDGE INPUT SIGNAL CONDITIONER