Omega Products DP3002-JC1-C1-G User Manual

4¹⁄₂ Digit Thermocouple Meters
1
8 DIN
R
DP3002-J, -K, -T
U J, K or T Thermocouple Input U 0.05 deg/deg Cold-Junction Error U -138 to 1248°C Range U Five-Segment Linearization U 0.1°C or 0.1 °F Resolution,
Jumper-Selectable
U 1 mV/deg Linearized
Recorder Output
U 120 dB CMR, 70 dB NMR U Bright, 14 mm (0.56")
LED Display
U Display Hold and Test U 115/230 Vac Power U EMI/RFI Filter for AC Power U Screw-Terminal Barrier Strip U Short 104 mm (4.1") Deep ¹⁄₈ DIN Case
Options
U Isolated 9 to 32 Vdc Power U Isolated 26 to 56 Vdc Power U NEMA 4 (IP65) Splash-Proof
Lens Cover
DP3002-KC1, shown
larger than actual size.
The DP3002-J, -K and -T are low-cost 4½ digit (±19,999 count) thermocouple meters with 0.1°C or
0.1°F resolution for indication-only applications. High accuracy is provided by five-segment linearization and matched, graded components for each T/C type.
STANDARD FEATURES
OPTIONS
Power options include isolated 9 to 32 Vdc and 26 to 56 Vdc. A splash-proof lens cover which meets NEMA 4 (IP65) requirements allows the DP3002 to be used in harsh industrial environments.
R-1
Specifications
Input
Thermocouple: Type J (Iron-Constantan, Fe-CuNi)
Type K (Chromel-Alumel, NiCr-NiAl) Type T (Copper-Constantan,
Cu-CuNi)
Calibration: IEC 584-1 (IPTS-68) Configuration: Single-ended
(-TC lead tied to ANA GND)
Polarity: Bipolar Zero: Adjustable ±20°C (±36°F)
Cold-Junction Tempco:
±0.05 deg/deg
Sensor-Wire Resistance Effect Per Conductor:
Type J: 40 µdeg/deg/Ω, up to 2500 Ω Type K: 50 µdeg/deg/Ω, up to 2000 Ω Type T: 50 µdeg/deg/Ω, up to 2000 Ω
Sensor-Break Current: 300 nA Sensor-Break Indication:
Positive overrange (upscale)
Digital Inputs
Level: TTL or 5V CMOS compatible
Noise Rejection
NMR, SIG HI to SIG LO:
70 dB, 50/60 Hz
CMR, ANA GND to PWR GND: 120 dB, DC to 60 Hz
CMV, ANA GND to PWR GND: 1500 Vp per HV test, 354 Vp per IEC spacing
Accuracy at 25°C
Range for Rated Accuracy: Type J: -138.0 to 760.0°C
( -216.0 to 1400.0°F) Type K: -105.0 to 1248.0°C
( -157.0 to 1999.9°F) Type T: -112.0 to 400.0 °C
( -170.0 to 752.0°F)
Maximum Error: ±1.5°C (±2.7°F) ±5 cts Span Tempco:
±0.01% of reading/°C Resolution: 0.1°C or 0.1°F
(jumper-selectable)
Full-Scale Step Response: 1 s Warmup to Rated Accuracy: 20 min
Analog Output
Voltage: 1 mV/deg Calibration Error: ±1.5 mV on °C,
±2.7 mV on °F
Current: 1 mA max Source Resistance: 68 Ω
Analog-to-Digital Conversion
Technique: Dual-slope, average value
Signal-integration period: 100 ms
Read Rate: 2.5/s
Display
Type: 7-segment, red LED Height: 14.2 mm (0.56") Symbols: -1.8.8.8.8 Decimal Points: Four positions
programmable by jumpers behind the lens or at connector
Overrange Indication: Four least­significant digits flash
Power
AC Voltages: 115 or 230 Vac, ±15% AC Frequency: 49 to 440 Hz DC Voltages: 9 to 32 Vdc, isolated to
300 Vp; 26 to 56 Vdc, isolated to 300 Vp
Power Consumption: 3.7 W nominal Output Voltages: 4.7 Vdc and
-4.7 Vdc ±5%, 10 mA max
Environmental
Operating Temperature: 0 to 60°C Storage Temperature: -40 to 85°C Relative Humidity: 95% at 40°C
(non-condensing)
Mechanical
Bezel: 96 W x 48 H x 5.1 mm D
(3.78 x 1.89 x 0.20")
Depth Behind Bezel: 104 mm (4.09") Panel Cutout: 92 W x 45 mm H
(3.62 x 1.77")
Weight: 15 oz (425 g) Case Material: 94V-0 UL-rated
polycarbonate
To Order Visit omega.com/dp3002-jkt for Pricing and Details
Model No. Description
DP3002
-JC1
-JF1
-KC1
-KF1
-TC1
-TF1
*
-C1
-C3C
-C3E
NOTE: All combinations may not be valid, check online for valid part numbers. Ordering Examples: DP3002-KC1, 4½ digit Type K thermocouple meter, 115 Vac power, -105 to 1248°C input range.
4½ digit thermocouple meters
-138 to 760°C
-216 to 1400°F
-105 to 1248°C
-157 to 1999°F
-112 to 400°C
-170 to 752°F
(Nothing, leave field blank) 115 Vac, 50/60 Hz power
230 Vac, 50/60 Hz
Isolated 9 to 32 Vdc
Isolated 26 to 56 Vdc
*
(Nothing, leave field blank) red LED display
-G
Green LED display
R-2