Omega Products DP25-S Installation Manual

Page 1
User’s Guide
Shop on line at
DP25-S and DP25B-S
Strain Gage Panel Meter
omega.com
e-mail: info@omega.com
For latest product manuals
omegamanual.info
MADE IN
Page 2
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, patient-connected applications.
This device is marked with the international caution symbol. It is important to read the Setup Guide before installing or commissioning this device as the guide contains important information relating to safety and EMC.
This device is marked with the international caution symbol. It is important to read the Setup Guide before installing or commissioning this device as the guide contains important information relating to safety and EMC.
!
®
®
OMEGAnet® On-Line Service
www.omega.com
Internet e-mail
info@omega.com
Servicing North America:
USA: One Omega Drive, P.O. Box 4047
ISO 9001 Certified Stamford CT 06907-0047
 TEL: (203) 359-1660 FAX: (203) 359-7700  e-mail: info@omega.com
Canada: 976 Bergar  Laval (Quebec) H7L 5A1  TEL: (514) 856-6928 FAX: (514) 856-6886  e-mail: info@omega.ca
For immediate technical or application assistance:
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
 Customer Service: 1-800-622-2378 / 1-800-622-BEST  Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
Mexico and TEL: (001)800-TC-OMEGA® FAX: (001) 203-359-7807 Latin American: En Espa
ñol: (001) 203-359-7803
 e-mail: espanol@omega.com
®
®
®
Servicing Europe:
Czech Republic: Frystatska 184, 733 01 Karviná
 TEL: +420 59 6311899 FAX: +420 59 6311114  e-mail: info@omegashop.cz
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany  TEL: +49 7056 9398-0 FAX: +49 7056 9398-29  Toll Free in Germany: 0800 639 7678  e-mail: info@omega.de
United Kingdom: One Omega Drive
ISO 9002 Certified River Bend Technology Centre
 Northbank, Irlam Manchester M44 5BD United Kingdom  TEL: +44 161 777 6611 FAX: +44 161 777 6622  Toll Free in England: 0800 488 488  e-mail: sales@omega.co.uk
Page 3
i
PREFACE
Manual Objectives
This manual shows you how to set up and use the Programmable Digital Meter.
Standard Procedures:
* Checking voltage jumpers, or changing voltage power * Mounting the panel * Selecting the input type * Selecting a decimal point position * Scaling with known loads (on-line calibration) * Scaling without known loads * Selecting ratiometric/non-ratiometric operation * Displaying the filtered/unfiltered input signal * Selecting a display color * Setting the setpoint's active band * Selecting a latched or unlatched operation * Setting setpoint deadbands * Enabling/disabling setpoint changes * Enabling/disabling the RESET button in the Run Mode
Optional Procedures:
* Setting input resolution * Enabling/disabling analog output * Selecting analog output as current or voltage * Assigning the output to net/gross reading * Scaling analog output
Features with are for the “B” version which has three-color programmable “Big” LED display - All segment characters shown are for the “B” version.
For first-time users: Refer to the QuickStart Manual for basic operation and set­up instructions.
Page 4
ii
Table A-1. Sections of the Manual
If you want to read about:
Unpacking; safety considerations
Meter description and features
Main board power jumpers; panel mounting, sensor input, main power and analog and relay output
Input type; decimal point position; reading scale & offset; reading configuration; display color; setpoint configuration; setpoint deadbands; output configuration (analog output); analog output scaling; lock out configuration; display brightness
Display messages
Meter menu/sub-menu messages
Setpoint configuration messages
Specifications
Factory Preset Values
1
2
3
4
5
6
7
8
9
Refer to section
Introduction
About the Meter
Getting Started
Configuring the Meter
Display Messages
Menu Configuration
Setpoint Configuration Displays
Specifications
Factory Default Setup as Shipped
Page 5
iii
Table of Contents
Table of Contents
Section Page
SEC 1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.2 Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
SEC 2 ABOUT THE METER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
2.1 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
2.2 Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
2.3 Available Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
2.4 Front of the Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
2.5 Back of the Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
2.6 Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
SEC 3 GETTING STARTED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
3.1 Rating/Product Label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
3.2 Main Board Power Jumpers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
3.3 Panel Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
3.4 Connecting Sensor Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
3.5 Connecting Main Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
3.6 Connecting External Tare Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
3.7 Connecting Analog and Relay Outputs . . . . . . . . . . . . . . . . . . . . . . . . . .19
SEC 4 CONFIGURING THE METER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
4.1 Selecting the Input Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
4.2 Selecting a Decimal Point Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
4.3 Selecting Reading Scale and Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
4.3.1 Scaling with Known Loads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
4.3.2 Scaling without Known Loads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
4.4 Using Reading Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
4.4.1 Selecting Ratiometric/Non-Ratiometric Operation . . . . . . . . . . . . . . . . . .28
4.4.2 Setting Input Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
4.4.3 Displaying the Filtered/Unfiltered Input Signal . . . . . . . . . . . . . . . . . . . . .29
4.4.4 Selecting Gross/Net or Peak Display . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
Page 6
Table of Contents
iv
Table of Contents
Section Page
4.5 Selecting a Display Color . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
4.6 Using Setpoint 1 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
4.6.1 Setting Setpoint 1's Active Band . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
4.6.2 Selecting if Setpoint 1 is Latched or Unlatched . . . . . . . . . . . . . . . . . . . .31
4.6.3 Assigning Setpoint 1 Values to Net or Gross Readings . . . . . . . . . . . . . .32
4.7 Using Setpoint 2 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
4.7.1 Setting the Setpoint 2’s Active Band . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
4.7.2 Selecting if Setpoint 2 is Latched or Unlatched . . . . . . . . . . . . . . . . . . . .33
4.7.3 Assigning Setpoint 2 Values to Net or Gross Readings . . . . . . . . . . . . . .33
4.8 Setting the Setpoint 1 Deadband . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
4.9 Setting the Setpoint 2 Deadband . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
4.10 Using Output Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
4.10.1 Enabling or Disabling the Analog Output . . . . . . . . . . . . . . . . . . . . . . . . .37
4.10.2 Selecting Analog Output as Current or Voltage . . . . . . . . . . . . . . . . . . . .38
4.10.3 Selecting Analog Output or Proportional Control . . . . . . . . . . . . . . . . . . .38
4.11 Using Output Scale and Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
4.11.1 Examples of Output Scale and Offset . . . . . . . . . . . . . . . . . . . . . . . . . . .41
4.12 Using Lock Out Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42
4.12.1 Enabling or Disabling the RESET Button in the Run Mode . . . . . . . . . . .42
4.12.2 Enabling or Disabling Setpoint Changes . . . . . . . . . . . . . . . . . . . . . . . . .42
4.12.3 Setpoint Display Function: Firmware Version or Setpoint Value . . . . . . .43
4.13 Using Display Brightness Configuration . . . . . . . . . . . . . . . . . . . . . . . . .43
4.13.1 Changing Brightness Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
SEC 5 DISPLAY MESSAGES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44
SEC 6 MENU CONFIGURATION DISPLAYS . . . . . . . . . . . . . . . . . . . . . . . . . .45
SEC 7 SETPOINT CONFIGURATION DISPLAYS . . . . . . . . . . . . . . . . . . . . . . .49
SEC 8 SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
SEC 9 FACTORY PRESET VALUES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
SEC 10 CE APPROVAL INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
Page 7
v
Table of Contents
List of Figures
Figure Page
2-1 Front-Panel with “Big” LED Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
2-2 Front Panel with Standard LED Display . . . . . . . . . . . . . . . . . . . . . . . . . .5
2-3 Connectors (AC-Powered and DC-Powered) . . . . . . . . . . . . . . . . . . . . . .8
3-1 Main Board Power Jumpers (W1, W2, W3) . . . . . . . . . . . . . . . . . . . . . .11
3-2 Main Board Jumper Positions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
3-3 Upper Isolated Analog Output Option Board Installation . . . . . . . . . . . . .12
3-4 Meter - Exploded View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
3-5 Panel Cut-Out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
3-6 Meter-powered Bridge Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
3-7 Externally-powered Bridge Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
3-8 4-Wire DC Input Connections with External Excitation . . . . . . . . . . . . . .16
3-9 Current Input Connections with Internal Excitation . . . . . . . . . . . . . . . . .16
3-10 Current Input Connections with External Excitation . . . . . . . . . . . . . . . . .17
3-11 Main Power Connections - AC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
3-12 Main Power Connections - DC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
3-13 External Tare Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
3-14 Analog Output Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
3-15 Relay Output Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
3-16 Isolated Analog Output Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4-1 Alarm Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
8-1 Meter Dimensions/Panel Cutout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
Page 8
Table of Contents
vi
List of Tables
Table Page
A-1 Sections of the Manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ii
2-1 Accessories and Add-Ons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
2-2 Connector Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
2-3 DIP Switch Positions/Input Range & Excitation . . . . . . . . . . . . . . . . . . . .10
3-1 S3 Jumper Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
3-2 AC-Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
4-1 Range Selection Dip Switch Positions for Regular Voltage Input . . . . . .23
4-2 Range Selection Dip Switch Positions for Millivolt/Milliamp Input . . . . . .23
4-3 Natural Gain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
4-4 Input Resolution Multiplier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
5-1 Display Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44
6-1 Menu Configuration Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
6-2 Run Mode Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
7-1 Setpoint Configuration Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49
8-1 Color Chart for DC Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
9-1 Factory Preset Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
Page 9
vii
Notes, Warnings and Cautions
NOTES, WARNINGS and CAUTIONS
Information that is especially important to note is identified by three labels:
• NOTE
• WARNING
• CAUTION
• IMPORTANT
NOTE: provides you with information that is important to successfully setup
and use the Programmable Digital Meter.
CAUTION or WARNING: tells you about the risk of electric shock.
CAUTION, WARNING or IMPORTANT: tells you of circumstances or
practices that can effect the meter's functionality and must refer to accompanying documents.
TIP: Provides you helpful hints.
Note
Page 10
Notes
viii
Page 11
1
Introduction
1
SECTION 1. INTRODUCTION
1.1 UNPACKING
Remove the Packing List and verify that all equipment has been received. If there are any questions about the shipment, use the phone numbers listed on the back cover to contact the Customer Service Department nearest you.
Upon receipt of shipment, inspect the container and equipment for any signs of damage. Take particular note of any evidence of rough handling in transit. Immediately report any damage to the shipping agent.
The carrier will not honor any claims unless all shipping material is saved for their examination. After examining and removing contents, save packing material and carton in the event reshipment is necessary.
Verify that you receive the following items in the shipping box:
QTY DESCRIPTION
1 Programmable Digital Meter indicator/controller with all applicable
connectors attached.
1 Owner's Manual
1 Set Mounting brackets
If you ordered any of the available options (except the "BL" Blank Lens option), they will be shipped in a separate container to avoid any damage to your indicator/controller.
Note
Note
Page 12
1
Introduction
2
1.2 SAFETY CONSIDERATIONS
This device is marked with the international caution symbol. It is important to read this manual before installing or commissioning this device as it contains
important information relating to Safety and EMC (Electromagnetic Compatibility).
This instrument is a panel mount device protected in accordance with EN 61010­1:2001, electrical safety requirements for electrical equipment for measurement, control and laboratory. Installation of this instrument should be done by qualified personnel. In order to ensure safe operation, the following instructions should be followed.
This instrument has no power-on switch. An external switch or circuit-breaker shall be included in the building installation as a disconnecting device. It shall be marked to indicate this function, and it shall be in close proximity to the equipment within easy reach of the operator. The switch or circuit-breaker shall not interrupt the Protective Conductor (Earth wire), and it shall meet the relevant requirements of IEC 947–1 and IEC 947-3 (International Electrotechnical Commission). The switch shall not be incorporated in the main supply cord.
Furthermore, to provide protection against excessive energy being drawn from the main supply in case of a fault in the equipment, an overcurrent protection device shall be installed.
Do not exceed voltage rating on the label located on the top of the instrument housing.
Always disconnect power before changing signal and power connections.
Do not use this instrument on a work bench without its case for safety reasons.
Do not operate this instrument in flammable or explosive atmospheres.
Do not expose this instrument to rain or moisture.
Unit mounting should allow for adequate ventilation to ensure instrument does not exceed operating temperature rating.
Use electrical wires with adequate size to handle mechanical strain and power requirements. Install without exposing bare wire outside the connector to minimize electrical shock hazards.
EMC Considerations
Whenever EMC is an issue, always use shielded cables.
Never run signal and power wires in the same conduit.
Use signal wire connections with twisted-pair cables.
Install Ferrite Bead(s) on signal wires close to the instrument if EMC problems persist.
Failure to follow all instructions and warnings may result in injury!!
Note
Page 13
3
About The Meter
2
SECTION 2. ABOUT THE METER
2.1 DESCRIPTION
The Digital Programmable Strain meter is a value packed indicator/controller. Four full digits and broad scaling capability allow for display in virtually all engineering units. A wide variety of DC current and voltage input ranges cover typical strain applications. Standard features include sensor excitation and front panel or remote tare. Your meter may be a basic indicator or it may include analog output or dual relay output. Analog or dual relay output must be ordered at time of purchase. Analog output is fully scalable and may be configured as a proportional controller, or to follow your display. Dual 5 amp, form C relays control critical processes. A mechanical lockout has been included to guard against unauthorized changes.
2.2 FEATURES
The following is a list of standard features:
* 4-digit, three color Programmable, Big LED display or
4-digit, Standard LED display * NEMA 4/Type 4 Front Bezel * ±0.03 % accuracy * 8 DC input ranges: 0-100 mV, ±50 mV, 0-5 V, 1-5 V,
0-10 V, ±5 V, 0-20 mA, and 4-20 mA * 5, 10, 12, or 24 Vdc sensor excitation * Peak detection * Front panel and remote tare function * Nonvolatile memory-no battery backup * 115 or 230 Vac 50/60 Hz power supply or
10-32 Vdc or 26-56 Vdc
The following is a list of optional features:
* Dual 5 amp, form C relay outputs * Scalable analog output * Proportional control * Easy setup for proportional control
Page 14
2
About The Meter
4
2.3 AVAILABLE ACCESSORIES
Table 2-1. Accessories and Add-Ons
Add-On Options
FS Special Calibration/Configuration SPC4 NEMA-4 Splash Proof Cover SPC18 NEMA-4 Splash Proof Cover, NEW
Accessories
TP1A Trimplate panel adaptor.
Adapts DIN1A/DIN2A cases to larger panel cutouts RP18 19-In. Rack Panel for one (1) 1/8 DIN instrument RP28 19-In. Rack Panel for two (2) 1/8 DIN instruments RP38 19-In. Rack Panel for three (3) 1/8 DIN instruments
Page 15
5
About The Meter
2
2.4 FRONT OF THE METER
Figure 2-1 shows each part of the front of the three-color programmable “Big” LED
display meter (Version B).
Figure 2-1. Front-Panel with Big Display
Figure 2-2 shows each part of the front of the standard LED display meter.
Figure 2-2. Front-Panel wtih Standard Display
These meter display windows (both versions) light when appropriate:
1 Setpoint 1 status
2 Setpoint 2 status
5 Pushbuttons for programming the meter.
//

••••
12
RESETMENU
TARENT/GRSSETPTS
Digital LED Display:
-
1.9.9.9 or 9.9.9.9 4-digit three color programmable, 21 mm (0.83") high LED display with programmable decimal point.
Digital LED Display:
-
1.9.9.9. or 9.9.9.9. 14 segment, 13.8 mm (0.54") high LED display with programmable decimal point.
Page 16
2
About The Meter
6
METER BUTTONS
SETPTS Button
In the Run Mode, this button will sequentially recall the previous setpoint settings. As necessary, use the /NT/GRS and /TARE buttons to alter these settings, then press the SETPTS button to store new values.
Unless you press the SETPTS,
/TARE, or /NT/GRS button within 20 seconds, the
meter will scroll to setpoint 2 and then to the Run Mode.
If the dual relay option is not installed or if L.3=1 is displayed on the LK.CF menu, pressing the SETPTS button will display the meter's firmware version.
/NT/GRS Button
In the Run Mode, this button will toggle between net/gross readings or peak readings depending upon setup.
In the Configuration Mode, press this button to change the value of the flashing digit shown on the display and/or toggle between menu choices, such as R.1=R or R.1=N on RD.CF menu. When configuring your setpoint values, press the /NT/GRS button to advance the flashing digit's value from 0 to 9 by 1.
/TARE Button
In the Run Mode press the /TARE button to tare your reading (zeroing).
In the Configuration Mode, press the this button to scroll to the next digit.
Note
Page 17
7
About The Meter
2
MENU Button
In the Run Mode, press the MENU button to terminate the current measuring strain and enter you into the Configuration Mode.
Only if you have not installed the lock out jumpers on the main board.
In the Configuration Mode, press the MENU button to store changes in the nonvolatile memory and then advance you to the next menu item.
RESET Button
If you hard reset (press the MENU button followed by the RESET button) or power off/on the meter, it shows RST, followed by STRN.
In the Run Mode, press the RESET button to reset the latched setpoints. The meter shows SP.RS and returns to the Run Mode.
In the Configuration Mode, press the RESET button once to review the previous menu. Press the RESET button twice to perform a hard reset and return to the Run Mode.
In the Peak Mode, press the RESET button to reset peak values. The meter shows PK.RS and returns to the Run Mode.
In the Setpoint Mode, press the RESET button to reset the latched setpoint. The meter shows SP.RS and enters the Run Mode.
In the Tare Mode, press the RESET button to reset. The meter shows T.RST.
When in setpoint or Configuration Mode, if the meter shows 9999 or -1999 with all flashing digits, the value has overflowed. Press the
/NT/GRS
button to start a new value.
Note
Note
Page 18
2
About The Meter
8
2.5 BACK OF THE METER
Figure 2-3 shows the label describing the connectors on the back of the meter. Table 2-2
on the following page gives a brief description of each connector at the back of the meter.
Figure 2-3. Connectors (AC-Powered and DC-Powered Detail)
POWER & INPUT CONNECTIONS(STRAIN)
Page 19
9
About The Meter
2
2.5 BACK OF THE METER (Continued)
Table 2-2. Connector Description
Connector Description
TB1-1 Setpoint 1: Normally open (N.O.1) connection TB1-2 Setpoint 1: Normally closed (N.C.1) connection TB1-3 Setpoint 1: Common (COM1) connection TB1-4 Setpoint 2: Normally open (N.O.2) connection TB1-5 Setpoint 2: Normally closed (N.C.2) connection TB1-6 Setpoint 2: Common (COM2) connection TB1-7 AC line connection (no connections on DC-powered units) TB1-8 AC neutral connection (+ Input on DC-powered units) TB1-9 AC earth ground (DC-power return on DC-powered units) TB1-10 Analog voltage output TB1-11 Analog current output TB1-12 Analog return TB2-1 -E: Negative excitation connection from meter (5, 10, 12 V) TB2-2 +E: Positive excitation connection from meter (5, 10, 12 V) TB2-3 +20 mA connection for analog input TB2-4 +R (Not used) TB2-5 +24 V output connection TB2-6 +S: Positive signal input TB2-7 -S: Negative signal input and return for +20 mA or +24 V TB2-8 -R (Not used) TB5-1 Isolated Analog Voltage Output TB5-2 Isolated Analog Current Output TB5-3 Isolated Analog Output Return J1 (1-2) Remote tare connection with a momentary switch
Page 20
2
About The Meter
10
The DIP switches are located at the S1 position (refer to Figure 3-2). Use a small instrument, such as a paper clip, to change the switches from open to closed. Table 2-3 lists DIP switch settings at the S1 position required to complete the setup of your meter.
Table 2-3. DIP Switch Positions/Input Range & Excitation
Function S1 DIP Switch Positions
C= Closed 1 2 3 4 5 6 7 8 O= Open
Settings for Excitation Voltage
Internal 5/10/12 excitation C - - - - - - -
External 5/10/12 excitation O - - - - O O ­Internal 12 Vdc excitation C - - - - O O ­Internal 10 Vdc excitation C - - - - C O -
Internal 5 Vdc excitation C - - - - C C -
Settings for Input Ranges
0-100 mV DC - O C O O - - O ±50 mV DC - O C O C - - O ±5 Vdc - C O O C - - C 0-10 Vdc - C O O O - - C 0-20 mA DC - O C C O - - O
The display must also be configured to the selected input type after setting the DIP switches (see Section 4.1, Selecting the Input Type)
2.6 DISASSEMBLY
You may need to open up the meter for one of the following reasons:
To check or change the 115 or 230 Vac power jumpers.
To install or remove jumpers on the main board.
Disconnect the power supply before proceeding.
To remove and access the main board, follow these steps:
Disconnect the main power from the meter.
Remove the back case cover.
Lift the back of the main board upwards and let it slide out of the case.
Note
Note
Page 21
11
Getting Started
3
SECTION 3. GETTING STARTED
Caution: The meter has no power-on switch, so it will be in operation as
soon you apply power.
If you power off/on the meter, or perform a hard reset (press the RESET button twice), the meter shows RST , followed by STRN.
3.1 RATING/PRODUCT LABEL
This label is located on top of the meter housing (refer to Figure 3-4).
3.2 MAIN BOARD POWER JUMPERS (refer to Figure 3-1)
Important: If you want to change the Factory preset jumpers, do the following steps; otherwise go to Section 3.3.
Warning: Disconnect the power from the unit before proceeding. This
device must only be reconfigured by a specially trained electrician with corresponding qualifications. Failure to follow all instructions and warnings may result in injury!
1. Remove the main board from the case. Refer to Section 2.6.
2. Locate the solder jumpers W1, W2, and W3 (located near the edge of the main board alongside the transformer).
3. If your power requirement is 115 Vac, solder jumpers W1 and W3 should be
wired, but jumper W2 should not. If your power requirement is 230 Vac, solder jumper W2 should be wired, but jumpers W1 and W3 should not.
Note: W4 jumper is not used.
Figure 3-1 shows the location of solder jumpers W1 through W3.
Figure 3.1 Main Board Power Jumpers
Note
W3 W2 W1
T1A
DISPLAY BOARD
Page 22
3
Getting Started
12
3.2 MAIN BOARD POWER JUMPERS (Continued)
Figure 3-2 shows the location jumper positions on the main board.
Figure 3-2. Main Board Jumper Positions
Figure 3-3. Upper Isolated Analog Output Option Board Installation
D
IS
P
L
A
Y
B
O
A
R
D
T1
S3
MAIN BOARD
S4
TB1
{
TB2
{
S1
S2
J1
TP1 TP2 TP3 TP4 TP5 TP6 TP7 TP8 TP9 TP10
EDCBA
Refer to table 3-1
S2
DCBA
S3
TP11
Attach cable to P1.
TB5
P1
Page 23
13
Getting Started
3
3.2 MAIN BOARD POWER JUMPERS (Continued)
S2 jumpers are used for testing purposes. Do not use as reading errors may result.
S3 jumpers are used for the following (refer to Figure 3-2):
* To enable or disable the front panel push-buttons * To allow for an extremely low resistance load for analog output * To disable the MENU button * To perform factory calibration procedure
Test pins TP1 - TP11 are for testing purposes. Do not use as reading errors may result.
S4-A Factory default jumper is removed.
Table 3-1. S3 Jumper Functions
Jumper Description
S3-A Install to enable front panel push-buttons.
Remove to disable all front panel push-buttons. S3-B Removed. For factory calibration only. S3-C Removed. Not used. S3-D Installed for external ratiometric. S3-E If installed without S3-B, the MENU button locks out.
If you press the MENU button, the meter shows LOCK .
Page 24
3
Getting Started
14
3.3 PANEL MOUNTING
Figure 3-4. Meter - Exploded View
1. Cut a hole in your panel, as
shown in Figure 3-4. For specific dimensions refer to Figure 3-5.
2. Insert the meter into the hole.
Be sure the front bezel gasket is flush to the panel.
3. Slide on mounting bracket to
secure.
4. Proceed to Section 3.4 to
connect your sensor input and main power.
Figure Figure 3-5. Panel Cut-Out
CONNECTOR
LABEL
REAR COVER
(REMOVED)
PANEL
CUT-OUT
"NEW" STYLE
MOUNTING
BRACKET
PRODUCT
LABEL
GASKET
FRONT BEZEL
CASE
"OLDER" STYLE
MOUNTING
BRACKET 2 PCS
PANEL THICKNESS
6,4 (.25) MAX 0,8 (.03) MIN
1,5
R(.06)
4 PLCS
45,00 + 0,61/-0,00
(1.772 + .024/–.000)
92,00 + 0,81/–0,00
(3.622 + .032/–.000)
NOTE: Dimensions in Millimeters (Inches)
Page 25
15
Getting Started
3
3.4 CONNECTING SENSOR INPUTS
Figures 3-6 shows excitation supplied from the meter’s internal supply
(50 mA maximum) Select 5, 10, or 12 volt excitation at DIP switch.
Figure 3-6. Meter-powered Bridge Input
Figures 3-7 shows the connections required for an externally-powered bridge input: the
external supply is brought to the meter’s buffer circuits to permit ratiometric readings. Set S1 DIP switch for external excitation for Figure 3-7 and 3-8.
Figure 3-7. Externally-powered Bridge Input
-E +E +20mA +R
+24 +S -S -R
1234
5678
+
-
TB2
Connections with “typical wire colors”
+E = Positive Excitation (red)
-E = Negative Excitation (black)
+S = Positive Signal Input (green)
-S = Negative Signal Input (white)
Connections with “typical wire colors”
+E = Positive Excitation (red)
-E = Negative Excitation (black)
+S = Positive Signal Input (green)
-S = Negative Signal Input (white)
-E +E
+20mA +R
1234
+24 +S -S -R
5678
+EX
TB2
+
-
-EX
Page 26
3
Getting Started
16
3.4 CONNECTING SENSOR INPUTS (Continued)
Figure 3-8. 4-Wire DC Input Connections with External Excitation
Figure 3-9. Current Input Connections with Internal Excitation
Page 27
17
Getting Started
3
3.4 CONNECTING SENSOR INPUTS (Continued)
Figure 3-10. Current Input Connections with External Excitation
3.5 CONNECTING MAIN POWER
Connect the AC main power connections as shown in Figure 3-11.
WARNING: Do not connect AC power to your device until you have
completed all input and output connections. This device must only be installed by a specially trained electrician with corresponding qualifications. Failure to follow all instructions and warnings may result in injury!
Figure 3-11. Main Power Connections - AC
Page 28
3
Getting Started
18
3.5 CONNECTING MAIN POWER (Continued)
Table 3-2 shows the wire color and respective terminal connections for both USA and
Europe.
Table 3-2. AC-Power Connections
Connect the DC main power connections as shown in Figure 3-12.
When using DC power, refer to the Table 8-1 Color Chart in the Specifications Section for Display Color, Intensity, Excitation Voltage and Current, and Analog Output Isolated Option. Failure to use proper ratings may result in damaging the unit.
Figure 3-12. Main Power Connections - DC
WIRE COLORS
TB1 AC POWER EUROPE USA
7 ~ AC Line Brown Black 8 ~ AC Neutral Blue White 9 ~ AC Earth Green/Yellow Green
TB1 TB2
1234 12 3456
78910 567811 12
+ DC
DC POWER
- DC
Page 29
19
Getting Started
3
3.6 CONNECTING EXTERNAL TARE SWITCH
Connect external tare connections as shown in Figure 3-13.
Figure 3-13. External Tare Connections
3.7 CONNECTING ANALOG AND RELAY OUTPUTS
If you have purchased a meter with analog or dual relay or isolated analog output, refer to the following drawings for output connections.
Figure 3-14. Analog Output Connections
TB1 TB2
1234 12 3456
78910 56 7811 1 2
S1
OPEN
CLOSED
EXTERNAL TARE SWITCH
J1
1 2
3 4
TB1 TB2
1234 123456
78910 567811 12
+
ANALOG
VOLTAGE
0-10V dc
+
-
ANALOG
CURRENT
0-20mA
or
4-20mA
ANALOG RTN
-
Page 30
3
Getting Started
20
3.7 CONNECTING ANALOG AND RELAY OUTPUTS (Continued)
Figure 3-15. Relay Output Connections.
Figure 3-16. Isolated Analog Output Connections.
Relay 1 Relay 2
N.O.1
}
N.C.1
COM1
N.O.2
}
N.C.2
COM2
NO2
COM2
EXTERNAL LOAD
FUSE
N
L
TB1
123456
789101112
TB1 TB2
1234 12 34
TB2
1234
5678
0—10VDC
56
0—20mA
OR
4—20mA
12 3
TB5
789101112
5678
Page 31
21
Configuring The Meter
4
SECTION 4. CONFIGURING THE METER
Refer to Table 6-1 for a summary list of menu configuration.
For first-time users: Refer to the QuickStart Manual for basic operation and set-up instructions.
4.1 SELECTING THE INPUT TYPE INPT
To select your appropriate input type signal, follow these steps:
Before proceeding, set the input DIP switch settings at the back of your meter. (Refer to Table 2-3).
1. Press the MENU button. The meter shows INPT.
2. Press the TARE button. The meter flashes one of the following:
100M (for 0-100 mV dc) (Default)
±50M (for ±50 mV dc)
10V (for 0-10 Vdc)
±5V (for ±5 Vdc)
0-20 (for 0-20 mA dc)
3. Press the ßNT/GRS button to scroll through available choices.
4. Press the MENU button to store your choice. The meter momentarily shows
STRD , followed by DEC.P (Decimal Point).
Note
Note
Page 32
4
Configuring The Meter
22
4.2 SELECTING A DECIMAL POINT POSITION DEC.P
Refer to Table 6-1 for a summary list of menu configuration.
To select a decimal point display position, follow these steps:
1. Press the MENU button until the meter shows DEC.P.
2. Press the TARE button. The meter shows one of the following:
FFF.F
FF.FF
F.FFF
FFFF
(
Default
)
3. Press the ßNT/GRS button to scroll between available choices.
4.
Press the MENU button to store your choice. The meter momentarily shows STRD , followed by the next menu RD.S.O (Reading Scale and Offset). Or you can press the RESET button to abort and go back to the DEC.P menu.
4.3 SELECTING READING SCALE AND OFFSET RD.S.0
Refer to Table 6-1 for a summary list of menu configuration.
To scale the meter to show readings in engineering units. There are two methods. One method is to scale with known inputs. Another method is to scale without known inputs: you calculate input values based on the transducer specifications and manually enter them through the keyboard.
Note
Note
Page 33
23
Configuring The Meter
4
4.3.1 Scaling with Known Loads (On-Line Calibration)
For maximum resolution, find the maximum signal that will be applied to the meter input.
• For regular voltage input, refer to the main body of Table 4-1.
• For millivolt or milliamp input, refer to the main body of Table 4-2.
Set the DIP switch positions as indicated at the top of either Table 4-1 or 4-2. The numbers 1 through 8 in the top row of either table represent dip switches 1 through 8, and the O, C or X directly below the number indicates the correct position of each switch.
‘O' Switch should be open or up.
‘C' Switch should be closed or down.
‘X' Switch is used to control excitation (refer to Table 2-3 to determine
correct position of these switches).
Once Dip switches have been positioned correctly, apply power. Proceed to the RD.CF (Reading Configuration) and set R2 equal to the value in the right hand column of the chart.
Table 4-1. Range Selection Dip Switch Positions For Regular Voltage Input
Table 4-2. Range Selection Dip Switch Positions For Millivolt/ Milliamp Input
Table 4-1. Range Selection Dip Switch Positions For Regular Voltage Input
Table 4-2. Range Selection Dip Switch Positions
For Millivolt and Milliamp Input
p
* Reading Configuration
12345678 12345678 12345678 RD.CF*
XOCOOXX0 XOCOCXX0 XOCCOXX0 R.2=
0 - 100 mV ±50 mV 0 - 20 mA 4
0 - 50 mV ±50 mV 0 - 10 mA 3 0 - 30 mV ±30 mV 0 - 6 mA 2 0 - 20 mV ±20 mV 0 - 4 mA 1 0 - 10 mV ±10 mV 0 - 2 mA 0
12345678 12345678 RD.CF*
XCOOOXXC XCOOCXXC R.2=
0 - 10 V ±5 V 4
0 - 5 V ±5 V 3 0 - 3 V ±3 V 2 0 - 2 V ±2 V 1 0 - 1 V ±1 V 0
Note
Page 34
4
Configuring The Meter
24
4.3.1 Scaling with Known Loads (On-Line Calibration) (Continued)
To scale with known inputs: apply known loads to a transducer connected to a meter, or simulate the transducer output with a voltage or current simulator. To scale with known inputs, follow these steps:
1. Apply a known load equal to approximately 0% of the transducer range.
2. Press the MENU button until the meter shows RD.S.O .
3. Press the TARE button. The meter shows IN1 (Input 1).
IN1 (Input 1) is the unscaled display reading at minimum input.
4. Press the TARE button again. The meter shows last stored value for Input 1.
5. Press the TARE button once more. The meter shows the actual signal being
received.
6. Press the MENU button to store this value as IN1 (Input 1). The meter shows
RD1 (Read 1).
RD1 (Read 1) is the desired display reading at Input 1.
7. Press the TARE button. The meter shows the last stored value for Read 1.
8. Press the ßNT/GRS button to change the value of your digits.
9. Press the TARE button to scroll horizontally to the next digit.
10. Press the MENU button to store value as RD1 . The meter shows IN!2 (Input 2).
IN!2 (Input 2) is the unscaled display reading at maximum input.
Note
Note
Note
Page 35
25
Configuring The Meter
4
4.3.1 Scaling with Known Loads (On-Line Calibration) (Continued)
11. Apply a known load equal to approximately 100% of the transducer range.
12. Press the TARE button again. The meter shows the last stored value for Input 2.
13. Press the TARE button once more. The meter shows the actual signal being
received.
14. Press the MENU button to store Input 2 value. The meter shows RD!2 (Read 2).
RD!2 (Read 2) is the desired display reading at input 2.
15. Press the TARE button. The meter shows the last stored value for Read 2.
16. Press the ßNT/GRS button to change the value of your digits.
17. Press the TARE button to scroll horizontally to the next digit.
18. Press the MENU button to store value as RD!2 (Read 2). The meter momentarily shows STRD , followed by RD.CF . Meter scaling is now complete.
Note
Page 36
4
Configuring The Meter
26
4.3.2 Scaling Without Known Loads
To scale without known inputs, calculate input values based on the transducer specifications and manually enter them on the front-panel push buttons. The following example assumes load cells with these specifications:
Maximum Load: 100.0 lbs Output: 3.1 mV/V Sensor Excitation: 10 Vdc Output: 31mV = (3.1 mV/V) x (10 V)
1.
Determine the correct values for IN1 and IN!2 based on the load cell specifications. In most cases, RD1 and RD!2 are equal to the minimum and maximum of the transducer output span. The example assumes RD1 and RD!2 are equal to the range of the load (RD1 = 0 and RD!2 = 100.0). Calculate IN1 and IN!2 using the
loadcell output span and the following equation:
IN = (Sensor Output) x (Natural Gain) x (Multiplier).
Table 4-3. Natural Gain
2. Determine the multiplier by the Input Resolution setting (R.2 in the RD.CF menu) and the input range selected. Typically R.2=4 is suitable for most applications.
Table 4-4. Input Resolution Multiplier
Input Range R.2=4 R.2=3 R.2=2 R.2=1 R.2=0
0 to 100 mV 1.000 2.000 3.333 5.000 10.00 0 to 10 V 1.000 2.000 3.333 5.000 10.00 0 to 20 mA 1.000 2.000 3.333 5.000 10.00 ± 50 mV 1.000 1.000 1.667 2.500 5.000 ± 5 V 1.000 1.000 1.667 2.500 5.000
Input Range Span Units Natural Gain
0 to 100 mV Millivolts 100 cts/mV
±50 mV Millivolts 40 cts/mV
0 to 10 V Volts 1000 cts/V
± 5 V Volts 400 cts/V
0 to 20 mA Milliamps 500 cts/mA
Page 37
27
Configuring The Meter
4
3. Determine IN1 and IN!2 input range and resolution. The example selects the 0 to 100 mV range and 10 uV resolution (R.2=4 ).
Example: IN1 = (0 mV) x (100 cts/mV) x (1.000) = 0
IN!2 = (31 mV) x (100 cts/mV) x (1.000) = 3100 RD1 = 0000 RD!2 = 100.0
4. Press MENU button until the meter shows RD.S.O .
5. Press the ∂TARE button. The meter shows IN1 .
6. Press the ∂TARE button again, the meter shows the last Input 1 value, with the fourth digit flashing.
7. Press the ßNT/GRS button to change the value of your digits.
8. Press the ∂TARE button to scroll horizontally to the next digit.
9. Press the MENU button to store this value. The meter shows RD1 .
10. Press the TARE button. The meter shows the last value for Read 1.
Repeat steps 7, 8 and 9 until RD1 , IN!2 and RD!2 have been displayed, verified, changed (if necessary) and stored.
4.4 USING READING CONFIGURATION RD.CF
Refer to Table 6-1 for a summary list of menu configuration.
You may use Reading Configuration RD.CF to configure your meter for the following:
To select ratiometric or non-ratiometric operation
To set the input resolution of your meter
To display the filtered/unfiltered signal input value
To select gross/net vs. peak reading
Note
Page 38
4
Configuring The Meter
28
4.4.1 Selecting Ratiometric/Non-Ratiometric Operation
1. Press the MENU button until RD.CF displays.
2. Press the TARE button. The meter shows one of the following:
R.1=R (Ratiometric reading) (Default - for strain meters)
R.1=N (Non-ratiometric reading - typically for voltage & current transducers)
3. Press the ßNT/GRS button to view last stored selection. Press the ßNT/GRS
button to toggle between selections.
4. Press the TARE button to select input resolution or press the MENU button to
store your selection and shows COLR menu.
4.4.2 Setting Input Resolution
To set the input resolution of your meter, follow these steps:
1. Press the MENU button until RD.CF displays, then press the TARE button twice.
or Press the TARE button from R.1 .
One of the following displays (default is R.2=4 ):
R.2=4 = 10 µV for Unipolar inputs. 25 µV for Bipolar inputs R.2=0 = 1 µV for Unipolar inputs. 5 µV for Bipolar inputs. R.2=1 = 2 µV for Unipolar inputs. 10 µV for Bipolar inputs R.2=2 = 3 µV for Unipolar inputs. 15 µV for Bipolar inputs. R.2=3 = 5 µV for Unipolar inputs. 25 µV for Bipolar inputs
Example: 3 µV resolution means that if you input 0-30 mV, at 30 mV the display shows
9999 .
2. Press the ßNT/GRS button to scroll through available selections.
3. Press the TARE button to display the filtered/unfiltered signal input or press the
MENU button to store your selection and shows COLR menu.
Page 39
29
Configuring The Meter
4
4.4.3 Displaying the Filtered/Unfiltered Input Signal
To display the filtered/unfiltered signal input, follow these steps:
1. Press the MENU button until RD.CF displays, then press the TARE button three times. or
Press the TARE button from R.2 .
One of the following displays:
R.3=F (Filtered value) (Default)
R.3=U (Unfiltered value)
2. Press the ßNT/GRS button to toggle between available choices.
3. Press the MENU button to store your selections. STRD momentarily displays, followed by COLR Menu.
4.4.4 Selecting Gross/Net or Peak Display
To select gross/net or peak display:
1. Press the MENU button until RD.CF displays, then press the TARE button three times. or
Press the TARE button from R.3.
One of the following displays:
R.4=G Gross/Net Display (Default)
R.4=P Peak Display
2. Press the ßNT/GRS button to toggle between available choices.
3. Press the MENU button to store your selections. STRD momentarily displays, followed by COLR Menu.
Page 40
4
Configuring The Meter
30
4.5 SELECTING A DISPLAY COLOR COLR
Refer to Table 6-1 for a summary list of menu configuration.
Selecting Display Color is not active unless your meter is a Version “B”.
To select a display color, follow these steps:
1. Press the MENU button until the meter shows COLR.
2. Press the TARE button. The meter shows one of the following:
GRN
RED
AMBR
3. Press the ßNT/GRS button to scroll between available choices.
4. Press the MENU button to store your choice. The meter momentarily shows
STRD , followed by the next menu S1.CF (Setpoint 1 Configuration). Or you can press the RESET button to abort and go back to the RD.CF menu.
4.6 USING SETPOINT 1 CONFIGURATION S1.CF
Refer to Table 6-1 for a summary list of menu configuration.
Setpoint 1 Configuration S1.CF is not active unless your meter has dual relay output capabilities. The LED's will display whether the S1.CF is active or not. You may use Setpoint 1 Configuration S1.CF for the following:
To set the setpoint's active band above or below your chosen value
To select whether the setpoint operation is latched or unlatched
Assigning setpoint values to the net or gross reading
Note
Note
Page 41
31
Configuring The Meter
4
4.6.1 Setting Setpoint 1's Active Band
1. Press the MENU button until the meter shows S1.CF .
2. Press the TARE button. The meter shows one of the following:
S.1=A (Active above the setpoint) (Default)
S.1=B (Active below the setpoint)
3. Press the ßNT/GRS button to toggle between available choices.
4. Press the TARE button to select if Setpoint 1 is latched/unlatched or press the
MENU button to store your selection. The unit shows S2.CF.
4.6.2 Selecting if Setpoint 1 is Latched or Unlatched
1. Press the MENU button until S1.CF displays, then press the TARE button twice. or Press the TARE button from S.1 .
The meter shows one of the following:
S.2=U Setpoint 1 to be unlatched (Default)
S.2=L Setpoint 1 to be latched
2. Press the ßNT/GRS button to toggle between available choices.
3. Press the TARE button to assign Setpoint 1 values to net or gross reading or press the MENU button to enter S2.CF (Setpoint 2 Configuration).
Page 42
4
Configuring The Meter
32
4.6.3 Assigning Setpoint 1 Values to Net or Gross Readings
1. Press the MENU button until S1.CF displays, then press the TARE button twice.
or Press the TARE button from S.2.
The meter shows one of the following:
S.3=N Setpoint 1 assigned to net reading (Default)
S.3=G Setpoint 1 assigned to gross reading
2. Press the ßNT/GRS button to toggle between available choices.
3. Press the MENU button to store your selection(s). The meter momentarily shows
STRD, followed by S2.CF (Setpoint 2 Configuration).
4.7 USING SETPOINT 2 CONFIGURATION S2.CF
Refer to Table 6-1 for a summary list of menu configuration.
Setpoint 2 Configuration S2.CF is not active unless your meter has dual relay output capabilities. The LED's will display whether the S2.CF is active or not. You may use Setpoint 2 Configuration S2.CF for the following:
• To set the setpoint's active band above or below your chosen value
• To select whether the setpoint operation is latched or unlatched
• To assign setpoint values to the net or gross reading
4.7.1 Setting Setpoint 2's Active Band
1. Press the MENU button until the meter shows S2.CF.
or
2. Press the TARE button. The meter shows one of the following:
S.1=A (Active above the setpoint) (Default)
S.1=B (Active below the setpoint)
3. Press the ßNT/GRS button to toggle between available choices.
4. Press the TARE button to select if Setpoint 2 is latched/unlatched or press the
MENU button to store your selection. The meter shows S1.DB.
Note
Page 43
33
Configuring The Meter
4
4.7.2 Selecting if Setpoint 2 is Latched or Unlatched
1. Press the MENU button until S2.CF displays, then press the TARE button twice. or Press the TARE button from S.1.
The meter shows one of the following:
S.2=U Setpoint 2 to be unlatched (Default)
S.2=L Setpoint 2 to be latched
2. Press the ßNT/GRS button to toggle between available choices.
3. Press the MENU button to store your selection(s). The meter momentarily shows STRD , followed by S1.DB (Setpoint 1 Deadband).
4.7.3 Assigning Setpoint 2 Values to Net or Gross Readings
1. Press the MENU button until S2.CF displays, then press the TARE button twice. or Press the TARE button from S.2.
The meter shows one of the following:
S.3=N Setpoint 2 assigned to net reading (Default)
S.3=G Setpoint 2 assigned to gross reading
2. Press the ßNT/GRS button to toggle between available choices.
3. Press the MENU button to store your selection(s). The meter momentarily shows STRD, followed by S1.DB (Setpoint 1 Deadband).
Page 44
4
Configuring The Meter
34
4.8 SETTING THE SETPOINT 1 DEADBAND S1.DB
Refer to Table 6-1 for a summary list of menu configuration.
Setpoint 1 Deadband S1.DB is not active unless your meter has dual relay output capabilities. The LED's will display whether the S1.DB is active or not. The Setpoint 1
Default deadband is 0003. To change the deadband (hysteresis) of Setpoint 1, follow these steps:
1. Press the MENU button until the meter shows S1.DB .
2. Press the TARE button. The meter shows the last previously stored 4-digit
number (0000 through 9999) with flashing 4th digit.
3. Press the ßNT/GRS button to change the value of the flashing digit. If you continue to press the ßNT/GRS button, the flashing digit's value continues to change.
4. Press the TARE button to scroll to the next digit.
5. Press the MENU button to store your selection. The meter momentarily shows STRD , followed by S2.DB (Setpoint 2 Deadband).
Note
Page 45
35
Configuring The Meter
4
4.9 SETTING THE SETPOINT 2 DEADBAND S2.DB
Refer to Table 6-1 for a summary list of menu configuration.
Setpoint 2 Deadband S2.DB is not active unless your meter has dual relay output capabilities. The LED's will display whether the S2.DB is active or not. The Setpoint 2
default deadband is 0003. To change the deadband (hysteresis) of Setpoint 2, follow these steps:
1. Press the MENU button until the meter shows S2.DB .
2. Press the TARE button. The meter shows the last previously stored 4-digit
number (0000 through 9999) with flashing 4th digit.
3. Press the ßNT/GRS button to change the value of the flashing digit. If you continue to press the ßNT/GRS button, the flashing digit's value continues to change.
4. Press the TARE button to scroll to the next digit.
5. Press the MENU button to store your selection. The meter momentarily shows STRD , followed by OT.CF (Output Configuration) if you have analog output
capabilities.
Note
Page 46
4
Configuring The Meter
36
Figure 4-1. Alarm Example
To reset latched alarms you must:
1. Input a signal OUT of the alarm zone
2. Then press SETPTS and then, RESET button
SIGNAL LEVEL
ON ONOFF
ACTIVE BELOW ACTIVE ABOVE
DEADBAND
SIGNAL LEVEL
ON
NOTE: DEADBAND WORKS AS HYSTERISIS
SIGNAL LEVEL
ON ON ON ON ON
OFF
ACTIVE BELOW
WITH DEADBAND 3
ON
ON
OFF OFF
OFF
ON
ONOFF OFF
ACTIVE ABOVE
WITH DEADBAND 3
SETPOINT
3
SETPOINT
3
SETPOINT
Note
ACTIVE BELOW LATCHED ACTIVE ABOVE LATCHED
Page 47
37
Configuring The Meter
4
4.10 USING OUTPUT CONFIGURATION OT.CF
Refer to Table 6-1 for a summary list of menu configuration.
Output Configuration OT.CF is not active unless your meter has analog output capabilities. The menu will display whether analog output is present or not. Analog output must be ordered at the time of purchase.
Use Output Configuration OT.CF to select the following:
To enable or disable the analog output
To select if the analog output is current or voltage
To assign te output to the net or gross reading
4.10.1 Enabling or Disabling the Analog Output
To enable or disable the analog output, follow these steps:
1. Press the MENU button until the meter shows OT.CF .
2. Press the TARE button. The meter shows one of the following:
O.1=E (Analog output enabled) (Default)
O.1=D (Analog output disabled)
3. Press the ßNT/GRS button to toggle between available choices.
4. Press the TARE button to select analog output as current or voltage or press the MENU button to store your selection and enter OT.S.O (Output Scale and Offset).
Note
Page 48
4
Configuring The Meter
3838
4.10.2 Selecting Analog Output as Current or Voltage
1. Press the MENU button until it shows OT.CF , then press the TARE button twice.
or Press the TARE button from 0.1.
The meter shows one of the following:
O.2=C (Analog output = current) (Default)
O.2=V (Analog output = voltage)
2. Press the ßNT/GRS button to toggle between available choices.
3.
Press the TARE button to select analog output or proportional control or press the
MENU
button to store your selection and enter OT.S.O (Output Scale and Offset).
4.10.3 Selecting Analog Output or Proportional Control
1. Press the MENU button until it shows OT.CF , then press the TARE button twice.
or Press the TARE button from 0.2.
The meter shows one of the following:
O.3=N (Net Reading) (Default)
O.3=G (Gross Reading)
2. Press the ßNT/GRS button to toggle between available choices.
3.
Press the
MENU
button to store your selection.
The meter momentarily shows
STRD , followed by
OT.S.O
(Output Scale and Offset).
Page 49
39
Configuring The Meter
4
39
4.11 USING OUTPUT SCALE AND OFFSET OT.S.O
Refer to Table 6-1 for a summary list of menu configuration.
Output Scale and Offset OT.S.O is not active unless your meter has analog output capabilities. The menu will display whether analog output is present or not. Output Scale
and Offset OT.S .O scales your analog output to be equal to the meter's display and/or any engineering units you require. You may scale the output for direct (4-20 mA, 0-10 V, etc) or reverse acting (20-4 mA, 10-0 V, etc).
1. Press the MENU button until OT.S.O displays.
2. Press the TARE button. RD1 (Read 1) displays.
This is your first point of display reading.
3. Press the TARE button again. The meter shows the last previously stored 4-digit number (-1999 through 9999) with flashing 4th digit.
4. Press the ßNT/GRS button to change the digits.
5. Press the TARE button to scroll to the next digit.
6. Press the MENU button to store your selection. OUT.1 (Output 1) displays.
This starting analog signal corresponds to your Read 1 display.
7. Press the TARE button. Selected output displays.
If you select O.2=V for voltage, the maximum signal you may select is
10.00 for an 0-10 Vdc signal output. If you select O.2=C for current, the maximum signal you may select is 20.00.
8. Press the ßNT/GRS button to enter the Output 1 signal selection. If you continue to press the ßNT/GRS button, the flashing digit's value continues to change.
9. Press the TARE button to scroll to the next digit.
10. Press the MENU button to store your selection. RD!2 (Read 2) displays.
This is your second point of display reading.
Note
Note
Note
Note
Note
Page 50
4
Configuring The Meter
404040
4.11 USING OUTPUT SCALE AND OFFSET OT.S.O (Continued)
11. Press the TARE button. The meter shows last previously stored 4-digit number
(-1999 through 9999) displays with flashing 4th digit.
12.
Press the ßNT/GRS button to change the value of the flashing digit. If you continue to press the ßNT/GRS button, the flashing digit's value continues to change.
13. Press the TARE button to scroll to the next digit.
14.
Press the
MENU
button to store your selection. The meter shows OUT.2 (Output 2).
This analog signal should correspond to your Read 2 display.
15. Press the TARE button. The meter shows selected output.
If you select O.2=V for voltage, the maximum signal you may select is 10.00 for an 0-10 Vdc signal output. If you select O.2=C for current, the maximum
signal you may select is 20.00 for a 0-20 or 4-20 mA DC signal output.
16. Press the ßNT/GRS button to change the value of the flashing digit. If you continue to press the ßNT/GRS button, the flashing digit's value continues to change.
17. Press the TARE button to scroll to the next digit.
18. Press the MENU button to store your selection. The meter momentarily shows STRD , followed by LK.CF (Lockout Configuration).
WARNING: If the meter displays all flashing values on any item, the value has overflowed. Press the ßNT/GRS button to start new values.
Note
Note
Page 51
41
Configuring The Meter
4
4.11.1 Examples for Output Scale and Offset
Example: You want to send 4-20 mA output for 0 to 100.0. The meter has 0.1 degree resolution. Complete the following steps:
1. Press the MENU button until the meter shows
OT.S.O
.
2. Press the TARE button. The meter shows
RD.1
(Read 1).
3. Press the TARE button to show the existing value.
4. Change the value of Read 1 to 000.0 by pressing the ßNT/GRS and TARE
buttons.
5. Press the MENU button to store your selection. The meter shows
OUT.1
(Output 1).
6. Press the TARE button to show the existing value.
7.
Change the value of Output 1 to 04.00 by pressing the ßNT/GRS and TARE buttons.
8.
Press the
MENU
button to store your selection. The meter shows
RD.2
(Read 2).
9. Press the TARE button to show the existing value.
10. Change the value of Read 2 to 100.0 by pressing the ßNT/GRS and TARE
buttons.
11. Press the MENU button to store your selection. The meter shows
OUT.2
(Output 2).
12. Press the TARE button to show the existing value.
13. Change the value of Output 2 to 20.0 by pressing the ßNT/GRS and TARE
buttons.
14. Press the MENU button to store your selection. The meter shows
LK.CF
(Lock Out Configuration).
Page 52
4
Configuring The Meter
42
4.12 USING LOCK OUT CONFIGURATION LK.CF
Refer to Table 6-1 for a summary list of menu configuration.
Use Lock Out Configuration LK.CF for the following:
To enable or disable setpoint changes
To enable or disable the RESET button in the Run Mode
To enable or diable displaying meter firmware version
4.12.1 Enabling or Disabling the RESET button in the Run Mode
1. Press the MENU button until the meter shows LK.CF (after OT.S.O ).
2.
Press the TARE button. The meter shows one of the following:
RS.=E To enable the RESET button in the Run Mode (Default)
RS.=D To disable the RESET button in the Run Mode
4. Press the ßNT/GRS button to toggle between available choices.
5. Press the MENU button to store the changes. The meter shows STRD if the new value is different otherwise the meter shows BRIT and returns to the Run Mode.
4.12.2 Enabling or Disabling SETPOINT Changes
1. Press the MENU button until the meter shows LK.CF (after OT.S.O).
2. Press the TARE button twice. The meter shows one of the following:
SP.=E To enable setpoint changes (Default)
SP.=D To disable setpoint changes
3. Press the ßNT/GRS button to toggle between available choices.
4. Press the MENU button to store the changes. The meter shows STRD if the new value is different otherwise the meter shows BRIT and returns to the Run Mode.
Note
Page 53
43
Configuring The Meter
4
4.12.3 SETPOINT Display Function: Firmware version or Setpoint value
1. Press the MENU button until the meter shows LK.CF (after OT.S.O ).
2. Press the ∂TARE button three times. The meter shows one of the following:
L.3=0 SETPTS button will display setpoint values.
L.3=1 SETPTS button will display the meter's firmware version.
3. Press the ßNT/GRS button to toggle between the choices above.
4. Press the MENU button to store the changes. The meter shows STRD if the new value is different otherwise the meter shows BRIT and returns to the Run Mode.
If your meter does not have the relay option, setpoint menu items above will not be available and SETPTS button will always display the meter's firmware
version. These units will have +OL (overload) or +OPN memory indicated by Alarm 1 & 2 LED displays. LEDs can be reset by pressing MENU then RESET button or by Power OFF then ON.
4.13 USING DISPLAY BRIGHTNESS CONFIGURATION
4.13.1 Changing Brightness Level
Changing Display Brightness is not active unless your meter is a Version “B”.
1. Press the MENU button until the meter shows BRIT(after LKCF).
2. Press the TARE button from BRIT. The meter shows one of the following:
M.BrT Medium Brightness
L.BrT Low Brightness
H.BrT High Brightness (Default)
3. Press the ßNT/GRS button to toggle between available choices.
4. Press the MENU button to store your selection. The meter momentarily shows STRD followed by STRD, RST, STRN , then measured value.
Note
Page 54
5
Display Messages
44
Table 5-1. Display Messages
MESSAGE DESCRIPTION
STRN Strain Meter
RST Hard (power on) Reset
INPT Input Type
DEC.P Decimal Point Position
RD.S.O Reading Scale and Offset
RD.CF Reading Configuration
COLR Display Color S1.CF Setpoint 1 Configuration S2.CF Setpoint 2 Configuration
S1.DB Setpoint 1 Deadband S2.DB Setpoint 2 Deadband
OT.CF Output Configuration
OT.S.O Output Scale and Offset
LK.CF Lock Out Configuration
BRIT Display Brightness
+OL + Overload Signal
-OL - Overload Signal
RS.OF Resolution Overflow
+999 Value Overflow in Setpoint/Menu Routines
I999 Value Overflow in Setpoint/Menu Routines NT.OF Net Value Overflow GT.OF Gross Value Overflow
ER1 2 Coordinate Format Programming Error
PEAK Peak Value PK.RS Peak Reset
T.RS Tare Reset
SP.RS Setpoint Reset
NET Net Value
GROS Gross Value
SP1 Setpoint 1 Value SP2 Setpoint 2 Value
R.OV.S Resolution Over Scale
V.-8.8 Firmware Version (where 8is 0 ~ 9)
RUN Operating Mode
-
Page 55
45
Menu Configuration Displays
6
SECTION 6. MENU CONFIGURATION DISPLAYS
Not all menu items display on standard meters.
Table 6-1. Menu Configuration Displays
(Defaults in Bold and Italics)
MENU TARE ßNT/GRS
Show input choices: 100M (Default
)
INPT
±50M 10V ±5V 0-20
Show current decimal FFFF (Default
)
DEC.P
point position F.FFF
FF.FF FFF.F
(Reading Scale & Offset)
Shows IN1
RD.S.O
Shows prior value entered and flashing digit. Changes the value Scrolls to the next digit. of the flashing digit
• If
TARE is pressed, actual input
is shown and can not be changed with
ßNT/GRS
.
• If ßNT/GRS
is pressed, unit can
scroll through digits with TARE.
Enter new value and Shows prior value entered and flashing digit. Changes the value show RD1 Scrolls to the next digit. of the flashing digit
Enter new value and Shows prior value entered and flashing digit. Changes the value show IN2 Scrolls to the next digit. of the flashing digit
• If
TARE is pressed, actual input
is shown and can not be changed with
ßNT/GRS
.
• If ßNT/GRS
is pressed, unit can
scroll through digits with TARE.
Enter new value and Shows prior value entered and flashing digit. Changes the value show RD2 Scrolls to the next digit. of the flashing digit
1
3
2
4
6
7
5
Page 56
6
Menu Configuration Displays
46
SECTION 6. MENU CONFIGURATION DISPLAYS (Continued)
Table 6-1. Menu Configuration Displays (Continued)
(Defaults in Bold and Italics)
MENU TARE
ßNT/GRS
Reading Configuration
R.1= R.1=R (Ratiometric Reading)
R.1=N (Non-ratiometric reading)
RD.CF
R.2= R.2=0
(1 µV resolution for unipolar & 5 µV resolution for bipolar)
R.2=1 (2 µV resolution for unipolar
& 10 µV resolution for bipolar)
R.2=2 (3 µV resolution for unipolar
& 15 µV resolution for bipolar)
R.2=3 (5 µV resolution for unipolar
& 25 µV resolution for bipolar)
R.2=4 (10 µV resolution for
unipolar & 25 µV resolution for bipolar)
Note: 3 µV resolution means if
your input is 0-30 mV, at 30 mV the display shows 9999.
R.3= R.3=F (Filtered value)
R.3=U (Unfiltered value)
R.4= R.4=G (Gross/Net Display)
R.4=P (Peak Display)
Show input choices:
GRN
(
Green
)
Display Color Selection
RED
(
Red
)
COLR
AMBR
(
Amber
)
Setpoint 1 Configuration S.1= S.1=A (Active above)
S.1=B (Active below)
S1.CF S.2= S.2=U (Unlatched)
S.2=L (Latched)
S.3= S.3=N (Net Reading)
S.3=G (Gross Reading)
Setpoint 2 Configuration S.1= S.1=A (Active above)
S.1=B (Active below)
S2.CF S.2= S.2=U (Unlatched)
S.2=L (Latched)
S.3= S.3=N (Net Reading)
S.3=G (Gross Reading)
Page 57
47
Menu Configuration Displays
6
SECTION 6. MENU CONFIGURATION DISPLAYS (Continued)
Table 6-1. Menu Configuration Displays (Continued)
(Defaults in Bold and Italics)
MENU TARE
ßNT/GRS
Press to scroll to the next Press to change the value of the flashing
Setpoint 1 Deadband
digit to the right digit
S1.DB
Press to scroll to the next Press to change the value of the flashing
Setpoint 2 Deadband
digit to the right digit
S2.DB
0.1 = 0.1=E (Analog output is enabled)
Output Configuration
0.1=D (Analog output is disabled)
OT.CF
0.2 = 0.2=C (Analog output is current)
0.2=V (Analog output is voltage)
0.3 = 0.3=N (Net reading)
0.3=G (Gross Reading)
Output Scale & Offset
Shows RD1 Changes the value of the flashing digit
OT.S.O
Shows prior value entered and flashing digit. Scrolls to the next digit.
Enter new value and Shows prior value entered Changes the value of the flashing digit show OUT1 and flashing digit.
Scrolls to the next digit.
Enter new value and Shows prior value entered Changes the value of the flashing digit show RD2 and flashing digit.
Scrolls to the next digit.
Enter new value and Shows prior value entered Changes the value of the flashing digit show OUT2 and flashing digit.
Scrolls to the next digit.
1
3
2
4
6
7
5
(
Shown if O.3=A in Output
Configuration Menu OT.CF
)
Page 58
6
Menu Configuration Displays
48
SECTION 6. MENU CONFIGURATION DISPLAYS (Continued)
Table 6-1. Menu Configuration Displays (Continued)
(Defaults in Bold and Italics)
Table 6-2. Run Mode Displays
Display
TARE ßNT/GRS RESET Description
Displays NET Peak Reading
NET / GROS
or GROSS reading. Once reading shows, Toggle between respective value Net and Gross shows.
values.
Press to Will reset your Tare Reset
T.RST
activate. tare when
viewing this function.
Reset Latched
SP.RS
Alarms
Pressing the RESET button resets your latched alarms.
MENU TARE
ßNT/GRS
RS= RS=E (Enable RESET button in the Run Mode)
Lock Out Configuration
RS=D (Disable RESET button in the Run Mode)
LK.CF
SP= SP=E (Enable setpoint changes)
SP=D (Disable setpoint changes)
L3= L3=0
(
SETPTS button display setpoint values
)
L3=1
(
SETPTS button display firmware version
V.-8.8
where 8 is 0 ~ 9
)
M.BrT(Medium Brightness
)
Brightness Configuration
BRIT
L.BrT (Low Brightness)
H.BrT
(High Brightness)
Shows input choices
Page 59
49
Setpoint Configuration Displays
7
SECTION 7. SETPOINT CONFIGURATION DISPLAYS
Table 7-1. Setpoint Configuration Displays
MENU
TARE
ßNT/GRS
Description
SETPOINT 1
SP1 Press to scroll to Press to change Select from -1999
the next digit to the the value of the through 9999 right flashing digit
SETPOINT 2
SP2 Press to scroll to Press to change Select from -1999
the next digit to the the value of the through 9999 right flashing digit
Page 60
8
Specifications
50
SECTION 8. SPECIFICATIONS
SIGNAL INPUT
Input Ranges: 0-100 mV, ± 50 mV, 0-10 V, ± 5 V, 0-20 mA, 4-20 mA
Isolation: Dielectric strength to 2500V transient per 3mm spacing
based on EN 61010 for 260Vrms or DC working voltage
Noise Rejection: Normal Mode Rejection (NMR) = 60 dB
Common Mode Rejection (CMR) = 120 dB
Resistance: 100 Meg ohms for 100 mV or ±50 mV input range
1 Meg ohm for 10 V or +5 V input range 5 ohms for 20 mA current input range
“Big” Display:
4-digit, three color programmable 9-segment, LED 21 mm (0.83")
Symbol: (
-1.9.9.9~9.9.9.9
)
Standard Display: 4-digit, 14-segment LED, 13.8 mm (0.54”) Symbol: (
-1.9.9.9. ~9.9.9.9.
)
ANALOG TO DIGITAL
Technique: Dual slope
Internal resolution: 15 bits
Read Rate: 3/sec Polarity Automatic
ACCURACY AT 25°C
Max Error Strain: ±0.03% of reading, ±1 count
Span Tempco: 50 ppm/°C
Step Response: 1 sec
Warm Up to Rated Accuracy:
30 min
Excitation Voltage: AC
power units DC power units
24 V @ 25 mA, Refer to Table 8-1 12 V @ 50 mA, Color chart for DC Output 10 V @ 120 mA, Excitation 5 V @ 60 mA
Load Regulation: 1.1%
Line Regulation: 0.02% per Vac
Page 61
51
Specifications
8
SECTION 8. SPECIFICATIONS (Continued)
ALARM OUTPUTS (if applicable)
2 Form "C" on/off relays. Configurable for latched and unlatched by software.
Max current: 5 AMPS, resistive load Max voltage: 250 Vac or 30 Vdc
ANALOG OUTPUT (if applicable)
Signal Type: Current or voltage
Signal Level: Current: 10 V max compliance at 20 mA output
Voltage: 20 mA max for 0-10 V output
Function: May be assigned to a display range or proportional
control output with Setpoint #1 when used as a control output.
Linearity: 0.2%
Step Response Time: 2 - 3 seconds to 99% of the final value
ISOLATED ANALOG OUTPUT (TB5, if applicable)
Same as non-isolated analog output except isolated.
Signal Type: Current or voltage
Signal Level: Current: 10 V max compliance at 20 mA output
Voltage: 20 mA max for 0-10 V output
Function: May be assigned to a display range or proportional
control output with Setpoint #1 when used as a control output.
Linearity: 0.2%
Step Response Time: 2 - 3 seconds to 99% of the final value
Isolation 130 Vrms working voltage, 1000 V/60sec Dielectric test
• Only one analog output is available on each unit and it must be factory installed.
Note
Page 62
8
Specifications
52
SECTION 8. SPECIFICATIONS (Continued)
INPUT POWER INFORMATION
AC units 115/230 V~(AC) + 10%, 50/60 Hz
9.5 W max, power consumption (Non-Isolated Analog Out)
11.0 W max, power consumption (Isolated Analog Out)
DC units 10-32 Vdc or 26-56 Vdc, 8 W
Do not use a combination of dc power and internal
excitation
or Isolated Analog Out, unless using dc power of 20-32
Vdc.
Refer to Table 8-1 below.
External Fuse Required:
IEC 127-2/III
Power Fuse
115 V 125 mA @ 250 (T)
230 V 63 mA @ 250 (T)
UL 248-14 (Listed Fuse)
Power Fuse
115 V 175 mA @ 250 V Slow-Blow
230 V 80 mA @ 250 V Slow-Blow
ENVIRONMENT
Operating temperature: 0° to 50°C (32° to 122°F)
Storage temperature: -40° to 85°C (-40° to 185° F)
Relative humidity: 90% at 40°C (non-condensing)
MECHANICAL
Panel cutout: 1/8 DIN 3.62 x 1.78" (45 x 92mm)
Weight: 1.27 lb (575 g)
Case material: Polycarbonate, 94 V-O UL rated
Protection: NEMA-4/Type 4 Front Bezel
Page 63
53
Specifications
8
Warning:
• Do not use Internal Excitation. Use External Excitation.
Do not use Isolated Analog Output. Use Non-Isolated Analog Ouput.
SECTION 8. SPECIFICATIONS (Continued)
Table 8-1. COLOR CHART FOR DC POWER
COLOR HIGH BRIGHTNESS MEDIUM & LOW BRIGHTNESS
RED Sensor Excitation: Any combination of Sensor Excitation
24 V @ 25 mA, and Analog Output 12 V, 10 V, 5 V @ 35 mA Max 24 V @ 25 mA,
Analog Output: 12 V @ 35 mA Max
Non-Isolated option only 10 V @ 35 mA Max
5 V @ 35 mA Max
GREEN
Analog Output:
Non-Isolated options or Isolated Analog option
AMBER
HIGH/LOW Brightness and AMBER are only available on Version “B” meters. Standard display meters are MEDIUM Brightness.
Note
Page 64
8
Specifications
54
SECTION 8. SPECIFICATIONS (Continued)
Figure 8-1 Meter Dimensions/ Panel Cutout
96,0 (3.78)48,0 (1.89)
FRONT BEZEL
20,3 (.80)
RETAINER
CASE
151,4 (5.96)
REAR COVER
SIDE VIEW
PANEL THICKNESS
6,4 (.25) MAX 0,8 (.03) MIN
1,5
R(.06)
4 PLCS
92,00 + 0,81/–0,00
(3.622 + .032/–.000)
NOTE: Dimensions in Millimeters (Inches)
45,00 + 0,61/-0,00
(1.772 + .024/–.000)
TOP VIEW
Page 65
55
Factory Preset Values
9
SECTION 9. FACTORY PRESET VALUES
Table 9-1. Factory Preset Values
MENU ITEM FACTORY PRESET VALUES
INPT Input Type: 0-100 (0-100 mV) input DEC.P Decimal Point Position: FFFF RD.S.O Reading Scale and Offset:
0-100 mV = 0-1000
RD.CF Reading Configuration:
R.1=R (Ratiometric) R.2=4 (10 µV resolution for unipolar &
25 µV resolution for bipolar)
R.3=F (Filtered value) R.4=G (Gross/Net reading)
COLR Normal Color Display:
RED or GRN (Note: depending how unit was ordered)
S1.CF Setpoint 1 Configuration:
S.1=A (Setpoint is Active above) S.2=U (Setpoint is Unlatched) S.3=N (Net reading)
S2.CF Setpoint 2 Configuration:
S.1=A (Setpoint is Active above) S.2=U (Setpoint is Unlatched)
S.3=N (Net reading) S1.DB Setpoint 1 Deadband: 0003 S2.DB Setpoint 2 Deadband: 0003 OT.CF Output Configuration:
O.1=E (Analog output is Enabled)
O.2=C (Analog output is Current)
O.3=N (Analog output follows the Net value) OT.S.O Output Scale and Offset:
0-1000 = 4-20 mA DC
LK.CF Lock Out Configuration
RS=E (Enable the RESET button in the Run Mode)
SP=E (Enable setpoint changes)
L3=0 (Display setpoint values) BRIT H.BrT (Brightness Level) SP1 Setpoint 1 Value: 0000 SP2 Setpoint 2 Value: 0000
Sensor Excitation 10 Vdc
Page 66
56
CE APPROVALS INFORMATION
This product conforms to the EMC directive 89/336/EEC amended by
93/68/EEC, and with the European Low Voltage Directive 72/23/EEC.
Electrical Safety EN61010-1:2001
Safety requirements for electrical equipment for measurement, control and laboratory.
Double Insulation Pollution Degree 2 Dielectric withstand Test per 1 min
Power to Input/Output: 2300 Vac (3250 Vdc)
Power to Input/Output: 500 Vac (720 Vdc)
(Low Voltage dc Power Option*)
Power to Relays Output: 2300 Vac (3250 Vdc)
Relay 1 to Relay 2: 2300 Vac (3250 Vdc)
Isolated Analog to Inputs: 1000 Vac (1420 Vdc)
Analog to Inputs: No Isolation
Measurement Category I
Category I are measurements performed on circuits not directly connected to the Mains Supply (power). Maximum Line-to-Neutral working voltage is 50 Vac/dc. This unit should not be used in Measurement Categories II, III, IV.
Transients Overvoltage Surge (1.2 / 50uS pulse)
Input Power: 2500 V
Input Power: 500 V
(Low Voltage dc Power Option*)
Isolated Analog: 500 V
Input/Output Signals: 500 V
Note: *Units configured for external low power dc voltage, 10-32 Vdc (Basic Insulation)
EMC EN61326:1997 + and A1:1998 + A2:2001
Immunity and Emissions requirements for electrical equipment for measurement, control and laboratory.
EMC Emissions Table 4, Class B of EN61326
EMC Immunity** Table 1 of EN61326
Note: **I/O signal and control lines require shielded cables and these cables
must be located on conductive cable trays or in conduits. Furthermore, the length of these cables should not exceed 30 meters
Refer to the EMC and Safety installation considerations (Guidelines) of this manual
for additional information.
Page 67
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of one (1) year from the date of purchase. In addition to OMEGA’s standard warranty period, OMEGA Engineering will extend the warranty period for four (4) additional years if the warranty card enclosed with each instrument is returned to OMEGA.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA:
1. Purchase Order number under which the product was PURCHASED,
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the product.
FOR NON-W
ARRANTY REPAIRS,
consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST
of the repair,
2. Model and serial number of product, and
3. Repair instructions and/or specific problems
relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
© Copyright 2002 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
TRADEMARK NOTICE:
®
,
omega.com
®
and are trademarks of OMEGA Engineering, Inc.
PATENT NOTICE:
This product is covered by one or more of the following patents: U.S. Pat. No. Des. 336,895; 5,274,577; 6,243,021 / CANADA 2052599; 2052600 / ITALY 1249456; 1250938 / FRANCE BREVET No. 91 12756 / SPAIN 2039150; 2048066 / UK PATENT No. GB2 249 837; GB2 248 954 / GERMANY DE 41 34398 C2.
The “Meter Case Bezel Design” is a trademark of
NEWPORT Electronics, Inc., registered in the U.S.
USED UNDER LICENSE. Other US and International Patents pending or
applied for.
®
Page 68
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop on line at omega.com
TEMPERATURE

Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies

Wire: Thermocouple, RTD & Thermistor

Calibrators & Ice Point References

Recorders, Controllers & Process Monitors

Infrared Pyrometers
PRESSURE, STRAIN AND FORCE

Transducers & Strain Gauges

Load Cells & Pressure Gauges

Displacement Transducers

Instrumentation & Accessories
FLOW/LEVEL

Rotameters, Gas Mass Flowmeters & Flow Computers

Air Velocity Indicators

Turbine/Paddlewheel Systems

Totalizers & Batch Controllers
pH/CONDUCTIVITY

pH Electrodes, Testers & Accessories

Benchtop/Laboratory Meters

Controllers, Calibrators, Simulators & Pumps

Industrial pH & Conductivity Equipment
DATA ACQUISITION

Data Acquisition & Engineering Software

Communications-Based Acquisition Systems

Plug-in Cards for Apple, IBM & Compatibles

Datalogging Systems

Recorders, Printers & Plotters
HEATERS

Heating Cable

Cartridge & Strip Heaters

Immersion & Band Heaters

Flexible Heaters

Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL

Metering & Control Instrumentation

Refractometers

Pumps & Tubing

Air, Soil & Water Monitors

Industrial Water & Wastewater Treatment

pH, Conductivity & Dissolved Oxygen Instruments
M3598/0906
Loading...