Omega Products DP1610-GN-R1 User Manual

Thermocouple/Process
OMEGA
CARE
SM
Extended Warranty Program
US
C
1
8 DIN Digital Indicator
Shown Actual Size
DP1610 Series
$
205
Basic Unit
U Large Four-Digit Red or
Green LED Display
U Universal Sensor Input:
Thermocouple, 3-Wire RTD, DC Voltage or Current
U One Latching or
Non-Latching Relay
U Max Hold, Min Hold
and Time Elapse
U Universal Power Supply U IP65/NEMA-4 Front
Panel
U Engineering Unit Labels
Supplied for Faceplate
Optional Features U 2nd and 3rd Alarm
Relays
U Remote Reset for
Alarms
U Analog Output U Transmitter Power
Supply
U RS-485 Communications U Low Voltage Power U Fully Configurable with
Laptop or PC
The DP1610 digital indicator is an easy-to-operate microprocessor­based instrument, incorporating the latest in surface mount and CMOS technology.
The instrument is a high visibility, full range digital indicator with the flexibility to accommodate thermocouple, RTD and dc inputs. These inputs can be scaled and ranged from the front panel without calibration. Excellent accuracy combined with small size and useful output options make the DP1610 a highly attractive, low cost solution to industrial measurement and indication.
Using the maximum and minimum hold feature, the DP1610 allows maximum and minimum measured variables to be stored and subsequently viewed, even after power down.
The units are supplied with a face plate that has a clear window to display the engineering units being measured. Engineering unit labels are supplied for: %, °F, °C, mV, V, mA, A, kW, W, lb, rpm, atmos, ppm, ft., %RH, inHg, psi, GAL, GPS, GPM, GPH, GPD, and MGD.
Specifications Input
Thermocouple:
R, S, J, T, K, L, B and N
RTD: 100 (.00385 ohm/ohm/°C) Volts: 0 to 5 Vdc, 1 to 5 Vdc,
0 to 10 Vdc and 2 to 10 Vdc Millivolts: 0 to 50 mV dc and 10 to 50 mV dc
M-45
Milliamps:
0 to 20 mA dc and 4 to 20 mA dc
Sensor Fault Detection:
Displays “LL” or “HH” for thermocouple or RTD inputs and sensor break SnSr. Alarms operate as if the process variable has gone over-range (TC & RTD) and under-range (V, mV, mA)
Remote Reset:
Voltage free or TTL compatible
Alarm Outputs Output 1 Relay Only:
SPDT, 2 A resistive @120/240 Vac
Output 2 Relay:
SPDT 2 A, resistive @ 120/240 Vac
Analog Outputs Current Output (retransmission only):
0 to 20 mA dc into 500 ohm max 4 to 20 mA dc into 500 ohm max.
Volts DC Output: (retransmission only):
0 to 10 Vdc 500 ohm minimum 0 to 5 Vdc 500 ohm minimum Output 3 Relay: SPDT, 2 A resistive @ 120/240 Vac
Transmitter Power Supply:
20 to 28 Vdc (24 Vdc nominal) 910 ohm (22 mA @ 20 Vdc)
Display Digital Display:
Four 7-segment LED’s, 9.9 mm (.39") high
Alarm Adjustment Process Alarm: ± Input Span Performance Measurement Accuracy (T/C & RTD):
±0.25% of span, ±1 LSD @ 20°C (68°F)
Note: Reduced performance with Type "B" thermocouple from 100 to 600°C (212 to 1112°F)
J
K
T
R
S
B
RTD
Measurement Accuracy (dc linear):
N
±0.05% of span, ±1 digit
Ambient Temperature Error:
0.01% of span/°C change in ambient
Linearization Accuracy (TC and RTD):
Better than: ±0.2°C any point, any 0.1°C range (±0.05°C (typical). Better than: ± 0.5°C any point, any 1°C range
Cold Junction Compensation:
Better than ± 0.7°C
Noise Rejection Common Mode: >120 dB at
50/60 Hz, giving negligible effect at up to 264 V, 50/60 Hz Series Mode: >500% of span (at 50/60 Hz) causes negligible effect Line Voltage: 90-264 Vac, 50/60 Hz (standard) 20-50 Vac 50/60 Hz or 22-65 Vdc (optional)
Operating Temperature:
0 to 55°C (32 to 131°F)
Storage Temperature:
-20 to 80°C (-4 to 176°F)
Humidity: 20 to 95% non-condensing Source Resistance:
1000 max. (thermocouple)
Lead Resistance:
50 per lead max balanced (Pt100)
EMI Susceptibility:
Meets EN50082 Part 2 and EN50081 Part 2
Companion Products: PTC1632 series 1/32 DIN
Timer/Controllers also available. See page M-129
Shown actual size
CN1632 and DP1632 controllers and meters available. See pages P-19 and M-43
Shown actual size
Dimensions Front panel:
96 x
100 mm
deep
(1.89 x 3.78 x
48 x
3.94") Weight: .48 kg (1.06 lb) max
Front Panel Sealing: IP65
1
8 DIN
DP1610
Shown Actual Size
1
32 DIN
DP1632
Input Types and Ranges
Input Type Range
Iron- 0 to 450°C 32 to 1471°F Constantan 0 to 761°C 32 to 571°F
CHROMEGA®- -200 to 760°C -328 to 1399°F ALOMEGA
Copper -200 to 262°C -328 to 503°F Constantan 0.0 to 260.6°C 32.0 to 501.0°F
Iron- 0 to 450°C 32 to 841°F Constantan 0 to 762°C 32 to 1403°F
L
(J DIN) 0.0 to 205.7°C 32.0 to 402.2°F OMEGALLOY Pt/13% Rh-Pt 0 to 1650°C 32 to 3002°F Pt/10% Rh-Pt 0 to 1649°C 32 to 3000°F Pt/30%Rh-Pt6%Rh Pt, 385 100 V -101.0 to 300.5°C -149.8 to 572.9°F
Process Voltage Process Current 0 to 20 mA, 4 to 20 mA
®
®
0.0 to 205.4°C 32.0 to 401.7°F
-200 to 1399°C -328 to 2503°F
0 to 1399°C 32 to 2550°F
100 to 1824°C 211 to 3315°F
0 to 50 mV dc, 10 to 50 mV dc, 0 to 5 Vdc,
1 to 5 Vdc, 0 to 10 Vdc, 1 to 10 Vdc, 2 to 10 Vdc
IN STOCK FOR FAST DELIVERY!
To Order
Model Number Price Description DP1610-R1 $2051⁄8 DIN digital indicator with red display DP1610-GN-R1 2051⁄8 DIN digital indicator with green display
Note: Basic unit comes with 1 relay output built-in. Up to 2 more outputs and 1 add' option
can be added. OMEGACARESMextended warranty program is available for models shown on this page.
Ask your sales representative for full details when placing an order. Ordering Example: DP1610-R1-R2-C4-24LV, 1⁄8 DIN meter with dual relay outputs, RS-485 communications and 24V power supply, $205 + 26 + 62 + 15 = $308 OCW-1 OMEGACARESMextends standard 3-year warranty to a total of 4 years ($30), $308+ 30 = $338
Output Options:
Output 2 Suffix Price Description
-R2 $26 Alarm Relay
-F2 47 Analog
Output 3 Suffix Price Description
-R3 $26 Alarm Relay
-TPS3 60 20-28 Vdc Trans.
Note: only 1 option can be selected from
each output ordering box above.
Optional Power Supply
Ordering Add’l Suffix Price Description
-24LV $15
(Specify Model Number)
Optional Plug-in Boards
Model No. Price Description
BD1600-R $26 2 A relay BD1600-F 47 Analog
Retransmission
Power Supply
BD1600-TPS 60 Trans. power BD1600-C4 62 RS-485
Option “R” above is for output slots 2 & 3 only. Option “F” above is for output slot 2 only. Option “TPS” above is for output slot 3 only.
Additional Options:
Ordering Add’l Suffix Price Description
-C4 $62
20 to 50 Vac or 22 to 65 Vdc
M-46
For Additional Controllers, See Section P
-RR 26
supply comm.
RS-485 communications
Remote reset
M
Loading...