Omega Products DP116-Mxx Installation Manual

http://www.omega.com
e-mail: info@omega.com
®
User ’s Guide
DP116-RTD
MINIATURE PANEL
THERMOMETER
MADE
IN
Servicing North America:
USA: One Omega Drive, P.O. Box 4047
ISO 9001 Cer
tified
Stamford, CT 06907-0047 Tel: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com
Canada: 976 Bergar
Laval (Quebec) H7L 5A1 Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.ca
For immediate technical or application assistance:
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
¨
Customer Service: 1-800-622-2378 / 1-800-622-BEST
¨
Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
¨
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Mexico: Tel: (001) 800-TC-OMEGA
¨
FAX: (001) 203-359-7807
En Espan÷ol: (001) 203-359-7803 e-mail: info@omega.com.mx
Servicing Europe:
Benelux: Postbus 8034, 1180 LAAmstelveen, The Netherlands
Tel: +31 (0)20 6418405 FAX: +31 (0)20 6434643 Toll Free in Benelux: 0800 0993344 e-mail: nl@omega.com
Czech Republic: RudŽ arm‡dy 1868, 733 01 Karvin‡ 8
Tel: +420 (0)69 6311899 FAX: +420 (0)69 6311114 e-mail: czech@omega.com
France: 9, rue Denis Papin, 78190 Trappes
Tel: +33 (0)130 621 400 FAX: +33 (0)130 699 120 Toll Free in France: 0800406342 e-mail: france@omega.com
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
Tel: +49 (0)7056 9398-0 FAX: +49 (0)7056 9398-29 Toll Free in Germany: 0800 TC-OMEGA
¨
e-mail: germany@omega.com
United Kingdom: One Omega Drive,
ISO 9002 Certified River Bend Technology Centre
Northbank, Irlam Manchester M44 5EX United Kingdom Tel: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622 Toll Free in United Kingdom: 0800 488 488 e-mail: sales@omega.co.uk
OMEGAnet¨On-Line Service Internet e-mail
http://www.omega.com info@omega.com
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, patient connected applications.
This device is marked with the international hazard symbol. It is important to read the Setup Guide before installing or commissioning this device as it contains important information relating to safety and EMC.
TABLE OF CONTENTS
SECTION PAGE
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ii
Models Available . . . . . . . . . . . . . . . . . . . . . . . . .iii
Section 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.2 Safety Considerations . . . . . . . . . . . . . . . . . . . . .2
Section 2 About the Meter . . . . . . . . . . . . . . . . . . . . . . . . .3
2.1 Front of the Meter . . . . . . . . . . . . . . . . . . . . . . . .3
2.2 Back of the Meter . . . . . . . . . . . . . . . . . . . . . . . .4
2.3 Connector Description . . . . . . . . . . . . . . . . . . . . .4
Section 3 Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . .5
3.1 Main Board Power Jumpers . . . . . . . . . . . . . . . .5
3.2 Changing the Meter Parameters . . . . . . . . . . . . .7
3.3 Installation and Panel Mounting . . . . . . . . . . . . .10
3.4 Sensor Input Connections . . . . . . . . . . . . . . . . .12
3.5 Analog Output Connections . . . . . . . . . . . . . . . .14
3.6 Display Hold Connections . . . . . . . . . . . . . . . . .15
3.7 Main Power Connections . . . . . . . . . . . . . . . . .16
3.8 Disassembly/Assembly . . . . . . . . . . . . . . . . . . .18
Section 4 Operation and Calibration . . . . . . . . . . . . . . . . .20
4.1 Equipment Required . . . . . . . . . . . . . . . . . . . . .20
4.2 Calibration Procedure . . . . . . . . . . . . . . . . . . . .21
Section 5 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . .23
Section 6 Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
i
ii
PREFACE
Manual Objectives: This manual shows you how to set up and use the RTD Panel Thermometer.
This meter is an economical Miniature Temperature Panel Thermometer featuring a large display with a linearized analog output that is supplied as a standard feature.
Each of the models* can be converted by the user to display in degrees Fahrenheit or Celsius between one degree and a tenth degree resolution.
RTD is an abbreviation for resistive temperature detector. The meter is designed to measure 100 ohm platinum resistance with
0.00385 temperature coefficient, also known as the European curve. The RTD will change resistance with a change in temperature. When using a 2-wire measurement configuration, lead wire impedance can contribute significant error to the temperature measurement. When using long lead wires a 3- or 4-wire measurement configuration is recommended.
The meter is available in platinum RTD with 2-, 3-, or 4-wire connections. All of the models listed in Table 1-1 come standard with red LEDs, and can be ordered with a green LED display as an option. The part numbers would then end with Ò-GR".
Example: DP116-MF1-GR
Also, these meters can be ordered with different power configurations. Refer to Table 1-3 for available choices.
*Refer to Table 1-1 for the complete listing of models available.
iii
MODELS AVAILABLE
TABLE 1-1
RTD MODELS AVAILABLE
The accuracies and ranges are listed in Section 5.
TABLE 1-2
Power Options Available
MODEL POWER
DP116-Mxx 115 Vac ±15%, 50/60 Hz
DP116-Mxx type-230 230 Vac ±15%, 50/60 Hz
DP116-Mxx type-9/26 9-26 Vdc @ 110mA max.
DP116-Mxx type-100 100 Vac ±15%, 50/60 Hz
DP116-Mxx type-24 24 Vac ±15%, 50/60 Hz
The following 3-1/2 digit mini RTD panel meters are available and discussed in this OperatorÕs Manual.
MODEL TYPE ¡C or ¡F RESOLUTION
DP116-MF1 RTD F 1.0¡F DP116-MF2 RTD F 0.1¡F DP116-MC1 RTD C 1.0¡C DP116-MC2 RTD C 0.1¡C
Note
iv
TABLE 1-3
Other Models Available
The following 3-1/2 digit mini thermocouple panel thermometers are discussed in a separate manual.
MODEL TYPE ¡C or ¡F RESOLUTION
DP116-JF1 J F 1.0¡F DP116-JF2 J F 0.1¡F DP116-JC1 J C 1.0¡C DP116-JC2 J C 0.1¡C
DP116-KF1 K F 1.0¡F DP116-KF2 K F 0.1¡F DP116-KC1 K C 1.0¡C DP116-KC2 K C 0.1¡C
DP116-TF1 T F 1.0¡F DP116-TF2 T F 0.1¡F DP116-TC1 T C 1.0¡C DP116-TC2 T C 0.1¡C
DP116-EF1 E F 1.0¡F DP116-EF2 E F 0.1¡F DP116-EC1 E C 1.0¡C DP116-EC2 E C 0.1¡C
v
NOTES, WARNINGS and CAUTIONS
Information that is especially important to note is identified by these labels:
NOTE: provides you with information that is important to successfully setup and use the Programmable Digital Meter.
CAUTION or WARNING: tells you about the risk of electric shock.
CAUTION, WARNING or IMPORTANT: tells you of circumstances or practices that can effect the meter's functionality and must refer to accompanying documents.
Note
vi
NOTES
1
SECTION 1 INTRODUCTION
1.1 UNPACKING
Remove the Packing List and verify that all equipment has been received. If there are any questions about the shipment, use the phone number for the Customer Service Department nearest you.
Upon receipt of shipment, inspect the container and equipment for any signs of damage. Take particular note of any evidence of rough handling in transit. Immediately report any damage to the shipping agent.
The carrier will not honor any claims unless all shipping material is saved for their examination. After examining and removing contents, save packing material and carton in the event reshipment is necessary.
Verify that you received the following items in the shipping box:
QTY DESCRIPTION
1 Panel Thermometer with 3 small connectors plugged
into the rear of the meter.
1 OperatorÕs Manual
Note
2
1.2 SAFETY CONSIDERATIONS
This device is marked with the international caution symbol. It is important to read this manual before installing or commissioning this device as it contains important information relating to Safety and EMC (Electromagnetic Compatibility).
Unpacking & Inspection
Unpack the instrument and inspect for obvious shipping damage. Do not attempt to operate the unit if damage is found.
This instrument is a panel mount device protected in accordance with Class I of EN 61010 (115/230 AC power connections). Installation of this instrument should be done by Qualified personnel. In order to ensure safe operation, the following instructions should be followed.
This instrument has no power-on switch. An external switch or circuit-breaker shall be included in the building installation as a disconnecting device. It shall be marked to indicate this function, and it shall be in close proximity to the equipment within easy reach of the operator. The switch or circuit-breaker shall not interrupt the Protective Conductor (Earth wire), and it shall meet the relevant requirements of IEC 947Ð1 and IEC 947-3 (International Electrotechnical Commission). The switch shall not be incorporated in the mains supply cord.
Furthermore, to provide protection against excessive energy being drawn from the mains supply in case of a fault in the equipment, an overcurrent protection device shall be installed.
¥ The Protective Conductor must be connected for safety reasons.
Check that the power cable has the proper Earth wire, and it is properly connected. It is not safe to operate this unit without the Protective Conductor Terminal connected.
¥ Do not exceed voltage rating on the label located on the top of the
instrument housing. ¥ Always disconnect power before changing signal and power connections. ¥ Do not use this instrument on a work bench without its case for safety
reasons. ¥ Do not operate this instrument in flammable or explosive atmospheres. ¥ Do not expose this instrument to rain or moisture. ¥ Unit mounting should allow for adequate ventilation to ensure instrument
does not exceed operating temperature rating. ¥ Use electrical wires with adequate size to handle mechanical strain and
power requirements. Install without exposing bare wire outside the
connector to minimize electrical shock hazards.
EMC Considerations
¥ Whenever EMC is an issue, always use shielded cables. ¥ Never run signal and power wires in the same conduit. ¥ Use signal wire connections with twisted-pair cables. ¥ Install Ferrite Bead(s) on signal wires close to the instrument if EMC
problems persist.
Note
Note
3
SECTION 2 ABOUT THE METER
2.1 Front of the Meter
Figure 2-1 shows the panel thermometer.
Figure 2-1. Panel Thermometer
Features:
Display:3 1/2 Digit, 7-Segment Red or Green LED
Full-size:14.2 mm (0.56") LED Display
Analog Output Standard
3/64 DIN Standard Panel Cutout
Removable Screw-Clamp Cable Connector
Display Hold Capability