Omega Products CNiTH-i16D Installation Manual

Page 1
User’s Guide
USA
MADE IN
www.omega.com
e-mail: info@omega.com
iSeries info: www.omega.com/specs/iseries
Temperature & Humidity
Controller Manual
Shop on line at
®
®
Page 2
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, patient-connected applications.
!
This device is marked with the international caution symbol. It is important to read the Setup Guide before installing or commissioning this device as the guide contains important information relating to safety and EMC.
®
®
OMEGAnet® On-Line Service
www.omega.com
Internet e-mail
info@omega.com
Servicing North America:
USA: One Omega Drive, P.O. Box 4047
ISO 9001 Certified Stamford CT 06907-0047
TEL: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com
Canada: 976 Bergar Laval (Quebec) H7L 5A1 TEL: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.ca
For immediate technical or application assistance:
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA Customer Service: 1-800-622-2378 / 1-800-622-BEST Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
Mexico and TEL: (001)800-TC-OMEGA® FAX: (001) 203-359-7807 Latin America: En Español: (001) 203-359-7803
e-mail: espanol@omega.com
®
®
®
Servicing Europe:
Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands TEL: +31 20 3472121 FAX: +31 20 6434643 Toll Free in Benelux: 0800 0993344 e-mail: sales@omegaeng.nl
Czech Republic: Frystatska 184, 733 01 Karviná TEL: +420 59 6311899 FAX: +420 59 6311114 e-mail: info@omegashop.cz
France: 11, rue Jacques Cartier, 78280 Guyancourt TEL: +33 1 61 37 29 00 FAX: +33 1 30 57 54 27 Toll Free in France: 0800 466 342
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany TEL: +49 7056 9398-0 FAX: +49 7056 9398-29 Toll Free in Germany: 0800 639 7678 e-mail: info@omega.de
United Kingdom: One Omega Drive
ISO 9002 Certified River Bend Technology Centre
Northbank, Irlam Manchester M44 5BD United Kingdom TEL: +44 161 777 6611 FAX: +44 161 777 6622 Toll Free in England: 0800 488 488 e-mail: sales@omega.co.uk
e-mail: sales@omega.fr
Page 3
TABLE OF CONTENTS
Part 1: Introduction............................................................................................2
1.1 Description .................................................................................2
1.2 Safety Considerations ...............................................................3
1.3 Before You Begin .......................................................................4
Part 2: Setup.......................................................................................................5
2.1 Front Panel .................................................................................6
2.2 Rear Panel Connections............................................................6
2.3 Electrical Installation .................................................................7
2.3.1 Power Connections........................................................7
2.3.2 Humidity and Temperature Probe.................................8
2.3.3 Wiring Outputs - Wiring Hookup...................................9
2.3.4 Dual Display Color Setup.............................................11
Part 3: Operation: Configuration Mode .........................................................12
3.1 Introduction ..............................................................................12
Turning your Instrument On for the First Time Buttons Functions in Configuration Mode
3.2 Menu Configuration ................................................................13
3.2.1 ID Number .....................................................................14
3.2.2 Setpoints .......................................................................15
3.2.3 Configuration Menu ....................................................15
3.2.4 Reading Configuration Menu .....................................16
3.2.5 Alarm 1 Menu ...............................................................17
3.2.6 Analog Output (Retransmission) Menu......................21
3.2.7 Alarm 2 Menu................................................................24
3.2.8 Loop Break Time Menu/Field Calibration...................25
3.2.9 Output 1 Menu ..............................................................27
3.2.10 Output 2 Menu ..............................................................34
3.2.11 Ramp and Soak Menu ..................................................36
3.2.12 ID Code Menu ...............................................................38
3.2.13 Communication (Options) Menu.................................40
3.2.14 Display Color Selection Menu.....................................46
Part 4: Specifications ......................................................................................49
Part 5: Factory Preset Values.........................................................................52
CE APPROVAL INFORMATION .......................................................................54
i
Page 4
LIST OF FIGURES:
Figure 2.1 Front Panel Display..............................................................................5
Figure 2.2 Rear Panel Power and Output Connections ......................................6
Figure 2.3 Rear Panel Input Connections ............................................................6
Figure 2.4 Main Power Connections.....................................................................7
Figure 2.5 Probe Wiring Hookup...........................................................................8
Figure 2.6 Output Connections:
a) Mechanical Relay and SSR Outputs – Wiring Hookup...............9
b) Pulse and Analog Outputs – Wiring Hookup...............................9
Figure 2.7 Typical Application...............................................................................9
Figure 2.8 Communication Output:
a) RS-232 Output – Wiring Hookup ................................................10
b) RS-485 Output – Wiring Hookup ................................................10
Figure 2.9 Excitation Output................................................................................10
Figure 2.10 Snubber Circuits Wiring Hookup ......................................................10
Figure 2.11 i/8D Locations of S1 and Selectable Jumper Positions..................11
Figure 2.12 i/16D Location of S1 and Selectable Jumper Positions..................11
Figure 3.1 Flow Chart for ID and Setpoints........................................................13
Figure 3.2 Flow Chart for Configuration Menu ..................................................15
Figure 3.3 Flow Chart for Reading Configuration Menu ...................................16
Figure 3.4 Flow Chart for Alarm 1.......................................................................17
Figure 3.5 Flow Chart for Analog Output (Retransmission).............................21
Figure 3.6 Flow Chart for Alarm 2.......................................................................24
Figure 3.7 Flow Chart for Loop Break Time.......................................................25
Figure 3.8 Flow Chart for Output 1 .....................................................................27
Figure 3.9 Flow Chart for Output 2 .....................................................................34
Figure 3.10 Flow Chart for Ramp and Soak .........................................................36
Figure 3.11 Flow Chart for ID Code.......................................................................38
Figure 3.12 Flow Chart for Communication Option ............................................40
Figure 3.13 Flow Chart for Display Color Selection............................................46
LIST OF TABLES:
Table 2.1 Front Panel Annunciators....................................................................5
Table 2.2 Rear Panel Connector..........................................................................6
Table 2.3 Fuse Requirement ................................................................................7
Table 3.1 Button Function in Configuration Mode...........................................12
Table 3.2 Command Letters and Suffix for iTH ...............................................41
Table 5.1 Factory Preset Values........................................................................52
ii
Page 5
NOTES, WARNINGS and CAUTIONS
Information that is especially important to note is identified by following labels:
• NOTE
• WARNING or CAUTION
• IMPORTANT
• TIP
NOTE: Provides you with information that is important to successfully
setup and use the Programmable Digital Meter.
CAUTION or WARNING: Tells you about the risk of electrical shock.
CAUTION, WARNING or IMPORTANT: Tells you of circumstances or
practices that can effect the instrument’s functionality and must refer to accompanying documents.
TIP: Provides you helpful hints.
1
Page 6
PART 1
INTRODUCTION
1.1 Description
The iTH Series instruments monitor and control both temperature and relative humidity.
• Used with an iTH probe the controller comes with dual displays. The top displays relative humidity and the bottom displays temperature. Relative Humidity can be toggled with Temperature readings (by pressing the bbutton) or Dewpoint readings (by pressing the cbutton). Dewpoint is the temperature at which water vapor from the air begins to form droplets and condenses on surfaces that are colder than the dewpoint of air.
• The temperature and humidity control can be achieved by using on/off or PID heat/cool control strategy. Control can be optimized with an auto tune feature. The instrument offers a ramp to setpoint with timed soak period before switching off the output.
The instruments are simple to configure and use, while providing tremendous versatility and a wealth of powerful features. The iTH Series instruments are available either as monitors or controllers. The monitors are extremely accurate programmable digital panel meters displaying humidity, temperature, or dew point. The controllers also provide dual loop control for both humidity and temperature and are easily programmed for any control or alarming requirement from simple on-off to full autotune PID control.
The iTH family of meters and controllers are available in four true DIN Sizes with NEMA 4, IP65 splash resistant bezels: the ultra compact 1/32 DIN (the world's smallest dual loop Humidity + Temperature controller); the popular midsize 1/16 DIN square bezel with dual display; the 1/8 DIN vertical, and the 1/8 DIN horizontal with the big bright 21mm digits.
The iTH series LED displays can be programmed to change color between Green, Amber, and Red at any set point or alarm point.
The iTH controller models offer a choice of two control or alarm outputs in almost any combination: solid state relays (SSR); Form "C" SPDT (Single Pole Double Throw) relays; pulsed 10 Vdc output for use with an external SSR; or Analog Output selectable for control or retransmission of the process value. Universal power supply accepts 90 to 240 Vac. Low voltage power option accepts 24 Vac or 12 to 36 Vdc.
The Networking and Communications options include direct Ethernet LAN connectivity with an Embedded Web Server, and serial communications. The -C24 serial communications option includes both RS-232 and RS-485. Protocols include both MODBUS and a straightforward ASCII protocol. The -C4EI option includes both Ethernet and RS-485 ASCII/MODBUS on one device.
The iTH Series meters and controllers are designed for easy integration with popular industrial automation, data acquisition and control programs as well as Microsoft Visual Basic and Excel. provides free configuration and data acquisition software and demos which makes it fast and easy to get up and running with many applications.
2
Page 7
1.2 Safety Considerations
This device is marked with the international caution symbol. It is important to read this manual before installing or commissioning this device as it contains important information relating to Safety and EMC (Electromagnetic Compatibility).
This instrument is a panel mount device protected in accordance with EN 61010-1:2001, electrical safety requirements for electrical equipment for measurement, control and laboratory. Installation of this instrument should be done by qualified personnel. In order to ensure safe operation, the following instructions should be followed.
This instrument has no power-on switch. An external switch or circuit­breaker shall be included in the building installation as a disconnecting device. It shall be marked to indicate this function, and it shall be in close proximity to the equipment within easy reach of the operator. The switch or circuit-breaker shall meet the relevant requirements of IEC 947–1 and IEC 947-3 (International Electrotechnical Commission). The switch shall not be incorporated in the main supply cord.
Furthermore, to provide protection against excessive energy being drawn from the main supply in case of a fault in the equipment, an overcurrent protection device shall be installed.
Do not exceed voltage rating on the label located on the top of the instrument housing.
Always disconnect power before changing signal and power connections.
Do not use this instrument on a work bench without its case for safety reasons.
Do not operate this instrument in flammable or explosive atmospheres.
Do not expose this instrument to rain or moisture.
Unit mounting should allow for adequate ventilation to ensure instrument does not exceed operating temperature rating.
Use electrical wires with adequate size to handle mechanical strain and power requirements. Install without exposing bare wire outside the connector to minimize electrical shock hazards.
EMC Considerations
Whenever EMC is an issue, always use shielded cables.
Never run signal and power wires in the same conduit.
Use signal wire connections with twisted-pair cables.
Install Ferrite Bead(s) on signal wires close to the instrument if EMC problems persist.
Failure to follow all instructions and warnings may result in injury!
3
Page 8
1.3 Before You Begin
Inspecting Your Shipment:
Remove the packing slip and verify that you have received everything listed. Inspect the container and equipment for signs of damage as soon as you receive the shipment. Note any evidence of rough handling in transit. Immediately report any damage to the shipping agent. The carrier will not honor damage claims unless all shipping material is saved for inspection. After examining and removing the contents, save the packing material and carton in the event reshipment is necessary.
Customer Service:
If you need assistance, please call the nearest Customer Service Department, listed in this manual.
Manuals, Software:
The latest Operation and Communication Manual as well as free configuration software and ActiveX controls are available from
the website listed in this manual or on the CD-ROM enclosed with your shipment
.
For first-time users: Refer to the QuickStart Manual for basic operation and set-up instructions.
If you have the Serial Communications/Ethernet Option you can easily configure the controller on your computer or on-line.
To Disable Outputs:
To ensure that menu changes are properly stored, Standby Mode should be used during setup of the instrument. During Standby Mode, the instrument remains in a ready condition, but all outputs are disabled. Standby Mode is useful when maintenence of the system is necessary.
When the instrument is in "RUN" Mode, push d twice to disable all outputs and alarms. It is now in "STANDBY" Mode. Push d once more to resume "RUN" Mode.
PUSH d TWICE to disable the system during an EMERGENCY.
To Reset the Meter:
When the controller is in the "MENU" Mode, push c once to direct controller one step backward of the top menu item.
Push c twice to reset controller, prior to resuming "Run" Mode except after "Alarms", that will go to the "Run" Mode without resetting the controller.
4
Page 9
PART 2
SETUP
2.1 Front Panel
Figure 2.1 Front Panel Display
The upper display may be RH, Temperature or Dewpoint readings depending on your Reading Configuration selections. Factory defaults are shown in Figure 2.1.
The Dual Display allows the user to observe the Relative Humidity or Dewpoint (upper display) and Temperature Value (lower display), at the same time.
Table 2.1 Front Panel Annunciators
1
Output 1/Setpoint 1/ Alarm 1 indicator
2
Output 2/Setpoint 2/ Alarm 2 indicator
°C
°C unit indicator for Temperature or Dewpoint
°F
°F unit indicator for Temperature or Dewpoint
%RH
Display shows the Percent Relative Humidity
D
Display shows the Dewpoint
a
Changes display to Configuration Mode and advances through menu items*
b
Used in Program Mode: On Dual Display: swaps the upper and lower displays from RH readings to Temperature readings. Note: this eliminates the small LED ennuciators* On Single Display units: replaces RH readings to Temperature readings*
c
Used in Program Mode: On Dual Display: changes upper display from RH readings to Dewpoint readings* On Single Display units: replaces RH readings to Dewpoint readings*
d
Accesses submenus in Configuration Mode
and stores selected values*
* See Part 3 Operation: Configuration Mode.
5
Refer to the Quick Start Guide for assembly and disassembly instructions.
Relative Humidity
Upper Display
Temperature
Lower Display
iTH-i32
1 2
%RH
D
1 2
%RH
D
1
1.00
[25.4]
1 2
%RH
D
1
1 2 C F
2
1
%RH
D
iTH-i16D
iTH-i8DH
iTH-i8DV
Page 10
2.2 Rear Panel Connections
The rear panel connections are shown in Figures 2.2 and 2.3.
Figure 2.2 Rear Panel Power and Output Connections
Figure 2.3 Rear Panel Input Connections
Table 2.2 Rear Panel Connector
POWER AC/DC Power Connector: All models INPUT iTH-PROBE OUTPUT 1 Based on one of the following models:
Relay SPDT Solid State Relay Pulse Analog Output (Voltage and Current)
OUTPUT 2 Based on one of the following models:
Relay SPDT Solid State Relay Pulse
OPTION Based on one of the following models:
RS-232C or RS-485 programmable Excitation
6
6 5 4 6 5 4 6 5 4 3 2 1 3 2 1
8 7 6 5 4 3 2 1
N/C
SHIELD
RED(VCC)
WHITE(CLK)
BLACK(RTN)
GREEN(DATA)
Page 11
2.3 Electrical Installation
2.3.1 Power Connections
Caution: Do not connect power to your device until you have completed all
input and output connections. Failure to do so may result in injury!
Connect the main power connections as shown in Figure 2.4.
Figure 2.4 Main Power Connections
Table 2.3 Fuse Requirement (see specifications)
FUSE Connector Output Type For 115Vac For 230Vac DC FUSE 1 Power N/A 100 mA(T) 100 mA(T) 100 mA(T) FUSE 2 Power N/A N/A N/A 400 mA(T)
For the low voltage power option, in order to maintain the same degree of protection as the standard high voltage input power units (90 - 240 Vac), always use a Safety Agency Approved DC or AC source with the same Overvoltage Category and pollution degree as the standard AC unit (90 ­240 Vac).
The Safety European Standard EN61010-1 for measurement, control, and laboratory equipment requires that fuses must be specified based on IEC127. This standard specifies for a Time-lag fuse, the letter code “T”. The above recommended fuses are of the type IEC127-2-sheet III. Be aware that there are significant differences between the requirements listed in the UL 248-14/CSA 248.14 and the IEC 127 fuse standards. As a result, no single fuse can carry all approval listings. A 1.0 Amp IEC fuse is approximately equivalent to a 1.4 Amp UL/CSA fuse. It is advised to consult the manufacturer’s data sheets for a cross-reference.
7
Use copper conductors only for power connections
Page 12
2.3.2 Humidity and Temperature Probe
The figure below shows the wiring hookup for the temperature and humidity probe.
Choose one which gives the best signal integrity-
1) Connect Probe’s Shield to RTN if Probe Housing is not
connected to
Earth Ground.
OR
2) Connect Probe’s Shield to Earth Ground if Probe Housing is not connected to Earth Ground.
Figure 2.5 Probe Wiring Hookup
8
METAL HOUSING
IS CONNECTED
TO SHIELD WIRE
N/C
RED(VCC)
WHITE(CLK)
BLACK(RTN)
GREEN(DATA)
SHIELD(See Note)
Page 13
9
2.3.3 Wiring Outputs
This meter has two factory installed outputs. The SPDT Mechanical Relay, SPST Solid State Relay, Pulse and Analog Output Connection are shown below.
Figure 2.6 a) Mechanical Relay and SSR b) Pulse and Analog
Outputs Wiring Hookup Outputs Wiring Hookup
Figure 2.7 Typical Applications
Use copper conductors only for power connections
ac CONTROLLED SSR USED WITH TEMPERATURE CONTROLLER WITH dc VOLTAGE SSR DRIVER OUTPUT
Vac
TEMPERATURE
CONTROLLER
CONTROL
SIDE
0 or 5 Vdc,
TYPICALLY
dc INPUT
SSR
431
LOAD
SIDE
2
HEATER
FAST BLOW
FUSE
Vac
ac CONTROLLED SSR USED WITH TEMPERATURE CONTROLLER WITH MECHANICAL RELAY OUTPUT
TEMPERATURE
CONTROLLER
Vac
ac CONTROLLED SSR USED WITH TEMPERATURE CONTROLLER WITH TRIAC OUTPUT
TEMPERATURE
CONTROLLER
Vac
CONTROL
SIDE
Vac
DRIVING
SSR
CONTROL SIDE
LOAD
RESISTOR
Vac
DRIVING
SSR
ac INPUT
SSR
431
ac INPUT
SSR
431
LOAD
SIDE
2
LOAD
SIDE
2
HEATER
FAST BLOW
FUSE
HEATER
FAST BLOW
FUSE
Vac
Vac
Page 14
10
This device may have a programmable communication output. The RS-232 and RS-485 Output Connection are shown below.
Figure 2.8
a) RS-232 Output Wiring Hookup b) RS-485 Output Wiring Hookup
This device may also have an excitation output.
Excitation is not available if communication option is installed.
If the Dual Display model has a Low Voltage power supply option, then excitation is not available.
This device has snubber circuits designed to protect the contacts of the mechanical relays when it switches to inductive loads (i.e. solenoids, relays). These snubbers are internally connected between the Common (C) and Normally Open (NO) relay contacts of Output 1 and Output 2.
If you have an inductive load connected between Common (C) and Normally Closed (NC) contacts of the mechanical relays and you want to protect them from the rush current during the switching period, you have to connect an external snubber circuit between Common (C) and Normally Closed (NC) contacts as indicated in
Figure 2.10.
Figure 2.10
Snubber Circuits Wiring
Hookup
Figure 2.9
Excitation Output
Page 15
11
2.3.4 Dual Display Color Setup
The dual display option allows the user to change the color of the upper and lower displays.
To change the color of the upper display, see Section 3.2.14 (Display Color section).
To change the color of the lower display follow the instructions below: The unit should be removed from the panel and opened.
Refer to the Quick Start Guide for assembly and disassembly instructions.
The S1 jumper is located on the back side of the display board. The location of S1 and pin selection jumpers are shown below.
Use a jumper for GREEN or RED, never leave S1 open.
Figure 2.11 i/8D Location of S1 and Selectable Jumper Positions
Figure 2.12 i/16D Location of S1 and Selectable Jumper Positions
i/8DH Dual Horizontal
i/8DV Dual Vertical
Page 16
12
PART 3
OPERATION: Configuration Mode
3.1 Introduction
The instrument has two different modes of operation. The first, Run Mode, is used to display Temperature and Relative Humidity. The other mode, Menu Configuration Mode, is used to navigate through the menu options and configure the controller. Part 3 of this manual will explain the Menu Configuration Mode. For your instrument to operate properly, the user must first "program" or configure the menu options.
Turning your Controller On for the First Time
The device becomes active as soon as it is connected to a power source. It has no On or Off switch. The device at first momentarily shows the software version number, followed by reset
RST
, and then proceeds to the Run Mode.
For first-time users: Refer to the QuickStart Manual for basic operation and set-up instructions.
If you have the Serial Communications/Ethernet Option you can easily configure the controller on your computer or on-line.
Table 3.1 Button Function in Configuration Mode
• To enter the Menu, the user must first press abutton.
• Use this button to advance/navigate to the next menu item. The user can navigate through all the top level menus by pressing a.
• While a parameter is being modified, press ato escape without saving the parameter.
• Press the up bbutton to scroll through “flashing” selections. When a numerical value is displayed press this key to increase value of a parameter that is currently being modified.
• Pressing the bbutton for approximately 3 seconds will speed up the rate at which the set point value increments.
• In the Run Mode, pressing the b button changes display from RH readings to Temperature readings.
• Press the down c button to go back to a previous Top Level Menu item.
• Press this button twice to reset the controller to the Run Mode.
• When a numerical value is flashing (except set point value) press cto scroll digits from left to right allowing the user to select the desired digit to modify.
• When a setpoint value is displayed press c to decrease value of a setpoint that is currently being modified. Pressing the c button for approximately 3 seconds will speed up the rate at which the setpoint value is decremented.
• In the Run Mode, pressing the c button changes from RH readings to Dewpoint readings.
• Press the enter d button to access the submenus from a Top Level Menu item.
• Press d to store a submenu selection or after entering a value — the display will flash a
STRD
message to confirm your selection.
• In the Run Mode, press d twice to enable Standby Mode with flashing
STBY
.
Reset: Except for Alarms, modifying any settings of the menu configuration will reset the instrument prior to resuming Run Mode.
a
MENU
b
(UP)
c
(DOWN)
d
ENTER
Page 17
13
3.2 Menu Configuration
It is required that you put the controller in the Standby Mode for any configuration change other than Setpoints & Alarms.
Figure 3.1 Flow Chart for ID and Setpoints
RdG
Page 18
14
3.2.1 ID Number
TO ENABLE/DISABLE OR CHANGE ID CODE, SEE SECTION 3.2.12.
If ID Code is Disabled or set as Default (0000) the menu will skip ID step to Setpoint Menu.
If ID Code is set to Full Security Level and user attempts to enter the Main Menu, they will be prompted for an ID Code.
If ID Code is set to Setpoint/ID Security Level and user attempts to enter the Configuration Menu, they will be prompted for an ID Code.
ENTERING YOUR NON-DEFAULT FULL SECURITY ID NUMBER.
Press a 1) Display shows ID. Press d 2) Display advances to
____
.
Press b & c 3) Press bto increase digit 0-9. Press cto activate next digit
(flashing). Continue to use band c to enter your 4-digit ID code.
Press d 4) If the correct ID code is entered, the menu will advance to
the Setpoint 1 Menu, otherwise an error message
ERRo
will
be displayed and the instrument will return to the Run Mode.
To change ID Code, see ID Menu in the Configuration section.
ENTERING YOUR NON-DEFAULT SETPOINT/ID SECURITY ID NUMBER.
Press a 5) Display shows
SP1
Setpoint 1 Menu.
Press a 6) Display shows
SP2
Setpoint 2 Menu.
Press a 7) Display shows IDID Code Menu. Press d 8) Display advances to
____
.
Press b & c 9) Use b and c to change your ID Code. Press d 10) If correct ID Code is entered, the display will advance to
the
INPT
Input Menu, otherwise the error message
ERRo
will be displayed and the controller will return to the Run Mode.
To prevent unauthorized tampering with the setup parameters, the instrument provides protection by requiring the user to enter the ID Code before allowing access to subsequent menus. If the ID Code entered does not match the ID Code stored, the controller responds with an error message and access to subsequent menus will be denied.
Use numbers that are easy for you to remember. If the ID Code is forgotten or lost, call customer service with your serial number to access and reset the default to
0000
.
Page 19
15
3.2.2 Set Points
SETPOINT 1:
Press a 1) Press a, if necessary until
SP1
prompt appears.
Press d 2) Display shows previous value of “Setpoint 1”. Press b & c 3) Press b and c to increase or decrease Setpoint 1
respectively.
Holding b & c buttons down for approximately 3 seconds will speed up the rate at which the Setpoint value increments or decrements.
Press b & c 4) Continue to use b and c to enter your 4-digit Setpoint 1 value. Press d 5) Display shows
STRD
stored message momentarily and then
advances to
SP2
only, if a change was made, otherwise press
a to advance to
SP2
Setpoint 2 Menu.
SETPOINT 2:
Press d 6) Display shows previous value of “Setpoint 2”. Press b & c 7) Press b and c to increase or decrease Setpoint 2
respectively.
Holding b & c buttons down for approximately 3 seconds will speed up the rate at which the setpoint value increments or decrements.
Press d 8) Display shows
STRD
stored message momentarily and then
advances to
CNFG
only, if a change was made, otherwise press
a to advance to
CNFG
Configuration Menu.
3.2.3 Configuration Menu
Figure 3.2 Flow Chart for Configuration Menu
Page 20
16
3.2.4 Reading Configuration
It is required that you put the controller in the Standby Mode for any
configuration change other than Set Points & Alarms.
Figure 3.3 Flow Chart for Reading Configuration Menu
ENTER READING CONFIGURATION MENU:
Press
a
1) Press a, if necessary, until
CNFG
prompt appears.
Press
d
2) Display advances to
RDG
Reading Configuration Menu.
Press
d
3) Display advances to
SNSR
Sensor.
SENSOR SUBMENU:
Press
d
4) Sensor selection for Autotune, Loop, or Ramp and Soak
°F.°C
is for temperature and
DRRh
is for Humidity.
TEMPERATURE UNIT SUBMENU:
Press
d
5) Display flashes previous Temperature Unit selection.
Press
b
6) Scroll though the available selections to the Temperature Unit of your choice: °For °C.
Press
d
7) Display shows
STRD
stored message momentarily and then
advances to
FLTR
Filter Constant.
FILTER CONSTANT SUBMENU:
Press
d
8) Display flashes previous selection for Filter Constant.
Press
b
9) Scroll though the available selections:
0001,0002,0004,0008,0016,0032,0064,0128
Press
d
10) Display shows
STRD
stored message momentarily only, if change was made, otherwise press ato advance to the next menu.
The Filter Constant Submenu allows the user to specify the number of readings stored in the Digital Averaging Filter.
For PID control select filter value 0001-0004. A filter value of 2 is approximately equal to 1 second RC low pass time constant.
Page 21
17
3.2.5 Alarm 1
This unit is equipped with two physical outputs that can only be configured as follows: Alarm 1 & Alarm 2, Alarm 1 & Output 2, Output 1 & Alarm 2, Output 1 & Output 2, Analog Out 1 & Alarm 2, Analog Out 1 & Output 2. Analog Out available only if Analog Output Option board is factory installed. Alarm1 will only work for Humidity, not Temperature.
If Analog Output Option is installed, the controller will skip Alarm 1 Menu item to Analog Output.
Alarm must be DISABLED if Ramp is ENABLED.
Figure 3.4 Flow Chart for Alarm 1
ENTER ALARM 1 MENU:
Press a 1) Press a, if necessary, until
CNFG
prompt appears.
Press d 2) Display advances to
RDG
Reading Configuration Menu.
Press a 3) Press a, until Display advances to
ALR1
Alarm 1 Menu.
Press d 4) Display advances to Alarm 1
ENBL
Enable or
DSBL
Disable
Submenu and flashes the previous selection.
Alarm 1 is designed to monitor the humidity value around Setpoint 1 and Alarm 2 is designed to monitor the temperature value around Setpoint 2.
Page 22
18
ALARM 1 ENABLE/DISABLE SUBMENU:
Press b 5) Scroll though the available selection until
ENBL
displays to
use Alarm 1.
Press d 6) Display shows
STRD
stored message momentarily and then
advances to
ABSo
only if it was changed, otherwise press a to
advance to
ABSo
Alarm 1 Absolute/Deviation Submenu.
If
DSBL
Alarm 1 Disabled was selected, all submenus of Alarm 1
Menu will be skipped and meter advances to
ALR2
Alarm 2
Menu. If
ENBL
Alarm 1 Enabled was selected, Output 1 would
be automatically Disabled.
ALARM 1 ABSOLUTE/DEVIATION SUBMENU:
Press
d
7)
Display flashes previous selection. Press bto
ABSo
Absolute or
_DEV
Deviation.
Press
d
8) Display shows
STRD
stored message momentarily (only if it
was changed) and then advances to
LTçH
.
Absolute Mode allows Alarm 1 to function independently from Setpoint 1. If the process being monitored does not change often, then "Absolute" Mode is recommended. The absolute Setpoint values are placed in the Alarm 1 Low and Hi values.
Deviation Mode allows Alarm 1 trigger at the Setpoint 1 value, as long as the Alarm 1 Low and Hi values equal 0. Deviation mode is typically the ideal mode if the process temperature changes often.
ALARM 1 LATCH/UNLATCH SUBMENU:
Press
d
9) Display flashes previous selection. Press bto
LTçH
Latched or
UNLT
Unlatched.
Press
d
10) Display shows
STRD
stored message momentarily (only if it
was changed) and then advances to
CT.CL
.
Latched Mode: Relay remains "latched" until reset. To reset already latched alarm, select Alarm Latch and press Max twice (i.e. Unlatch and then back to Latch) or from a Run Mode, push dtwice to put the controller in Standby Mode and then push done more time to return to the Run Mode.
Unlatched Mode: Relay remains latched only as long as the alarm condition is true.
Page 23
19
CONTACT CLOSURE SUBMENU:
Press
d
11) Display flashes previous selection. Press b to
N.ç.
Normally
Closed or
N.o.
Normally Open.
Press
d
12) Display shows
STRD
stored message momentarily (only if it
was changed) and then advances to
AçTV
.
Normally Open: If this feature is selected, then the relay is "energized" only when an alarm condition occurs.
Normally Closed: "Fail Safe" Mode. Relay is energized under "normal" conditions and becomes de-energized during alarm or power failure.
ACTIVE SUBMENU:
Press
d
13) Display flashes previous selection. Press bto scroll through the available selections:
ABoV
Above,
BELo
Below,
HI.Lo
HI/Low and
BAND
Band. (Band is active if
_DEV
Deviation was
selected).
Press
d
14) Display shows
STRD
stored message momentarily (only if it was changed) and then advances to
A.P.oN
.
Above: In Absolute Mode, Alarm 1 is triggered when the process variable is greater than the Alarm Hi Value (Low value ignored). In Deviation Mode
, Alarm 1 is
triggered at the Setpoint plus any offset placed in the Alarm Hi value.
Below: In Absolute Mode
, Alarm 1 is triggered when the process variable is less than the Alarm Low Value (Hi value ignored). In Deviation Mode, Alarm 1 is triggered at the Setpoint plus any offset placed in the Alarm Low value.
Hi/Low: In Absolute Mode, Alarm 1 is triggered when the process variable is less than the Alarm Low Value or above the Hi Value. In Deviation Mode
, Alarm 1 is triggered when the process variable is less than the Setpoint minus any offset in the Alarm Low or greater than the Setpoint plus any offset in the Alarm Hi value.
Band: Alarm 1 is triggered when the process variable is above or below the "band" set around Setpoint 1. Band equals Hi Value (Low Value ignored). A "band" is set around the Setpoint by the instrument only in the "Deviation" Mode.
Page 24
20
ALARM ENABLE/DISABLE AT POWER ON:
Press
d
15) Display flashes previous selection. Press bto
ENBL
enable
or
DSBL
disable.
Press d 16)
Display shows
STRD
stored message. momentarily
(only if it
was changed)
and then advances to
ALR.L
.
If Alarm at Power On is enabled, the alarm will be active when an alarm condition occurs. If Alarm at Power On is disabled, the alarm will not be active (even if an alarm condition exists) unless the process value moves into a non-alarm condition and back into an alarm condition.
ALARM 1 LOW VALUE SUBMENU:
Press
d
17) Display flashes 1stdigit of previous value. Use band cto
enter new value. Press b& c18) Use band cto enter Alarm 1 Low Value. Press
d
19) Display shows
STRD
storage message momentarily (only if
it was changed) and then advances to
ALR.H
.
ALARM 1 HI VALUE SUBMENU:
Press
d
20) Display flashes 1stdigit of previous value. Use band cto
enter new value. Press b& c21) Use band cto enter Alarm1 Hi Value. Press
d
22) Display shows
STRD
stored message momentarily (only if it
was changed) and then advances to the next menu.
Page 25
21
3.2.6 Analog Output (Retransmission)
Analog Output works only for Humidity Readings and can be configured as Retransmission or Control outputs. This section will explain Retransmission Output.
This unit is equipped with two physical outputs that can only be configured as follows: Alarm 1 & Alarm 2, Alarm 1 & Output 2, Output 1 & Alarm 2, Output 1 & Output 2, Analog Out 1 & Alarm 2, Analog Out 1 & Output 2. Analog Output is available only if Analog Output Option board is factory installed.
If Analog Output Option is not installed, the instrument will skip to Alarm 2 Menu.
Figure 3.5 Flow Chart for Analog Output (Retransmission)
ENTER ANALOG OUTPUT MENU:
Press
a
1) Press a, if necessary, until
CNFG
prompt appears.
Press
d
2) Display advances to
RDG
Reading Configuration Menu.
Press
a
3) Press a, if necessary, until Display advances to
ANLG
Analog Output Menu. Press
d
4) Display advances to Analog Output
ENBL
Enable or
DSBL
Disable Submenu and flashes the previous selection.
Page 26
22
ANALOG OUTPUT ENABLE/DISABLE SUBMENU:
Press
b
5) Scroll though the available selection until
ENBL
displays to use Analog Output Retransmission (output proportional to the input signal).
Press
d
6) Display shows
STRD
stored message momentarily and then
advances to
CURR
or
VoLT
Submenu only if it was changed,
otherwise press a to advance to
CURR
or
VoLT
Current/Voltage Submenu.
If
DSBL
Analog Output Disabled was selected, all submenus of Analog
Output Menu will be skipped and the meter will advance to
ALR2
Alarm 2
Menu. If
ENBL
Analog Output Enabled was selected, Output 1 would be
automatically Disabled, and reassigned as Analog Output.
CURRENT/VOLTAGE SUBMENU:
Press
d
7) Display flashes
CURR
Current or
VoLT
Voltage.
Press
b
8) Scroll through the available selection: Current or Voltage (Example
VoLT
).
Press
d
9) Display shows
STRD
stored message momentarily and then
advances to
RD1
Submenu only if it was changed, otherwise
press ato advance to
RD1
Reading 1 Submenu.
READING 1:
Press
d
10) Display flashes 1
st
digit of previous “Reading 1” value. Press b& c11) Enter “Reading 1” value. (Example 0000) Press
d
12) Display advances to
OUT.1
Out 1 Submenu.
OUT 1:
Press d 13) Display flashes 1
st
digit of previous “Out 1” value. Press b & c 14) Enter “Out 1” value. (Example 00.00) Press d 15) Display advances to RD!2 Reading 2 Submenu.
READING 2:
Press
d
16) Display flashes 1
st
digit of previous “Reading 2” value. Press b& c17) Enter “Reading 2” value. (Example 100.0) Press
d
18) Display advances to
OUT.2
Out 2 Submenu.
OUT 2:
Press
d
19) Display flashes 1
st
t
digit of previous “Out 2” value. Press b& c20) Enter “Out 2” value. (Example 10.00) Press
d
21) Display advances to the
ALR2
Alarm 2 Menu.
The above example (also the factory default) is for 0-10 V of the entire range of the Humidity Input and Analog Output. For 0-20 mA output you need to set “Analog Type” to Current and OUT 2 to 20.00.
Page 27
23
Accuracy of Analog Output board is +/-1% of FS (Full Scale) when following conditions are satisfied:
1. The input is not scaled below 1% of Input FS (10 mV @ 1 V or 0.2 mA @ 20 mA input ranges).
2. Analog Output is not scaled below 3% of Output FS (300 mV @ 10 V or
0.6 mA @ 20 mA output ranges).
Otherwise certain corrections need to be applied.
For example:
For entire range of Humidity Input, the Analog Output on 10 V FS scaled for 300 mV output range:
Rd1 = 0000, Out1 = 00.00 RD2 = 100.0, Out2 = 00.30
The measured output will be as follows:
Rd1 = 0000, Out1 = -0.07 V Rd2 = 100.0, Out2 = 0.23 V
This means that for 300 mV output range we have -70 mV offset at zero and at full scale. In order to compensate this 70 mV offset the correct scaling will be as follows:
Rd1 = 0000, Out1 = 00.07 Rd2 = 100.0, Out2 = 00.37
The above corrections need to be applied only for Input scaled below 1% of FS and Output scaled below 3% of FS or if you need the Analog Output
accuracy to be better than 1% of FS.
Page 28
24
3.2.7 Alarm 2
This unit is equipped with two physical outputs that can only be configured as follows: Alarm 1 & Alarm 2, Alarm 1 & Output 2, Output 1 & Alarm 2, Output 1 & Output 2, Analog Out 1 & Alarm 2, Analog Out 1 & Output 2. Analog Out available only if Analog Output Option board is factory installed. Alarm 2 only works for Temperature, not Humidity.
Alarm must be DISABLED if Ramp is ENABLED.
Figure 3.6 Flow Chart for Alarm 2
ENTER ALARM 2 MENU:
Press
a
1) Press a, if necessary, until
CNFG
prompt appears.
Press
d
2) Display advances to
RDG
Reading Configuration Menu.
Press
a
3)
Press a, if necessary, until Display advances to
ALR2
Alarm 2 Menu.
Press
d
4)
Display advances to Alarm 2
ENBL
Enable or
DSBL
Disable
Submenu.
ALARM 2 ENABLE/DISABLE SUBMENU:
5) Display flashes previous selection. Press buntil
ENBL
displays
to use Alarm 2.
Press
d
6) Display shows
STRD
stored message momentarily and then
advances to
ABSo
only if it was changed, otherwise press ato
advance to
ABSo
Absolute/Deviation Submenu.
If
DSBL
Alarm 2 Disabled was selected, all submenus of Alarm 2 will
be skipped and meter advances to
LOOP
Loop Break Time Menu. If
ENBL
Alarm 2 Enabled was selected, Output 2 will automatically
Disabled, and reassigned as Alarm 2.
The remaining Alarm 2 menu items are identical to Alarm 1 Menu. Modifying Alarm Settings will not reset the instrument.
Page 29
25
3.2.8 Loop Break Time
Figure 3.7 Flow Chart for Loop Break Time
ENTER LOOP BREAK TIME MENU:
Press
a
1) Press a, if necessary, until
CNFG
prompt appears.
Press
d
2) Display advances to
RDG
Reading Configuration Menu.
Press
a
3) Press a, if necessary, until Display advances to
LOOP
Loop
Break Time Menu.
Press
d
4) Display advances to Loop Break Time
ENBL
Enable or
DSBL
Disable Submenu and flashes the previous selection.
LOOP BREAK ENABLE/DISABLE SUBMENU:
Press b 5) Scroll through the available selections:
ENBL
or
DSBL
.
Press
d
6) Display shows
STRD
stored message momentarily and then
advances to
B.TIM
Loop Break Time Value Submenu.
Loop Break is an additional safety feature intended to monitor the rate of change of the process value, while approaching the SP1, if on the Sensor Menu
DRRh
is selected, or SP2 if
°F.°C
is selected. It is strictly intended as an additional warning system, therefore its use is entirely optional. An active Loop Break will cause the Process Value digits to blink in a rotating pattern. If the process value reaches the set point the blinking will stop and
B.TIM
is
completed successfully, otherwise
BR.AL
Break Alarm warning will flash, and
Output 1 will be turned off.
LOOP BREAK TIME VALUE SUBMENU:
Press
d
7) Display flashes 1stdigit of previous Loop Value.
Press b& c8)
Press band cbuttons to enter a new Loop Value (0 to
99.59).
Press
d
9) Display shows
STRD
stored message momentarily and then
advances to
R.ADJ
Reading Adjust Menu.
Loop Break Time Value allows the user to determine the time interval in MM:SS (from zero to 99 minutes and 59 seconds) that the Process Value changes at least 10 counts. At the specified time interval, if the process value change is less than the stated rate, flashing
B.TIM
will be displayed, the output will be de­energized, and Alarm energized. Loop break time will be disabled when the Process Value (PV) enters the control band.
Loop Break Alarm can work on the RH or Temperature function of the selection made on the Sensor Menu.
It is required that you put
the controller in the Standby Mode for any configuration change other than Set Points & Alarms.
Page 30
26
READING ADJUST SUBMENU:
Press
d
10) Display flashes 1stdigit of previous reading adjust value.
Press b& c11) Press band cbuttons to enter a new Reading Adjust value
(-1999 to 9999).
Press
d
12) Display shows
STRD
stored message momentarily and then
advances to
SP.DV
Setpoint Deviation Menu.
Reading Offset Adjust
For Relative Humidity, the controller allows the user to fine tune a minor error of the transducer, however some applications may require a large offset adjust. (Displayed Process Value = Measured Process Value ± R.ADJ). R.ADJ is adjustable between -19.99 to 99.99
SETPOINT DEVIATION ENABLE/DISABLE SUBMENU:
Press
d
13) Display advances to Setpoint Deviation
ENBL
Enable or
DSBL
Disable Submenu and flashes the previous selection.
Press
b
14) Scroll through the available selections:
ENBL
or
DSBL
.
Press
d
15) Display shows
STRD
stored message momentarily and
then advances to
OUT1
Menu.
Setpoint Deviation Submenu, if “enabled”, allows changes to Setpoint 1 to be made automatically to Setpoint 2. This mode is very helpful if the Process Value changes often. In Setpoint Deviation Mode, set SP2 a certain number of degrees or counts away from SP1 - this relation remains fixed when SP1 is changed. For instance: Setting SP1=200 and SP2=20 and enabling
SP.DV
means that the absolute value of SP2=220. Moving SP1 to 300, the absolute value of SP2 becomes 320.
Page 31
27
3.2.9 Output 1
Alarm 1 and Output 1 or Analog Output (Retransmission) share the same contacts on the rear panel connector. If Alarm 1 or Analog Output (Retransmission) is
Enabled, Output 1 is automatically Disabled.
Figure 3.8 Flow Chart for Output 1
It is required that you
put the controller in the Standby Mode for any configuration change other than Set Points & Alarms. Alarm 1 and Output 1 will work for Humidity only.
Page 32
28
ENTER OUTPUT 1 MENU:
Press
a
1) Press a, if necessary, until
CNFG
prompt appears.
Press
d
2) Display advances to
RDG
Reading Configuration Menu.
Press
a
3) Press a, if necessary, until Display advances to
OUT1
Output
1 Menu.
Press
d
4) Display advances to
SELF
Self Submenu.
SELF SUBMENU:
The Self Option allows the output of the instrument to be controlled manually from the front panel.
Press
d
5) Display flashes the current setting of Self,
ENBL
Enabled or
DSBL
Disabled.
Press
b
6) Press the bbutton to select between Enable and Disable.
Press
d
7) If Self
ENBL
Enabled was selected, display shows
STRD
stored message momentarily and then advances to the next menu (Output 1 setting is completed).
The output is now under the direct control of the operator and can be adjusted in the Run Mode (
M00.0
to
M99.9
), by pressing the band cbuttons, where M calls for the Manual (Self) Control. For example, setting of
M50.0
of an Analog Output of 0
to 10 Vdc would produce roughly 5 Vdc at the output.
8) If Self
DSBL
Disabled was selected, display shows
STRD
stored message momentarily and then advances to
oPLO
Minimum/Percent Low Submenu of Output 1 Menu.
There is a shorter way to Enable or Disable Self Mode. From a Run Mode, press dand then press a. Self Mode is Enabled now. Press bor
c
to display MXX.X. To disable Self, press dand then press a. Display
goes to the Run Mode. Self Mode is Disabled now.
MINIMUM/PERCENT LOW SUBMENU:
Specify in percent, the minimum value (0000) for control output. If the output is analog proportional (Current or Voltage), then the minimum voltage or current, in percent, is specified. If the output is time proportional (Relay, SSR or Pulse), then the minimum duty-cycle, in percent, is specified.
Press
d
9) Display flashes 1st digit of previous “Percent Low” setting.
Press b& c10) Use band cbuttons to enter a new value for “Percent Low”. Press
d
11) Display shows
STRD
stored message momentarily and then
advances to
oPHI
Maximum/Percent High Submenu.
Page 33
29
MAXIMUM/PERCENT HIGH SUBMENU:
Specify in percent, the maximum value (99) for control output. If the output is analog proportional (Current or Voltage), then the maximum voltage or current, in percent, is specified. If the output is time proportional (Relay, SSR, or Pulse), then the maximum duty-cycle, in percent, is specified.
Press
d
12) Display flashes 1
st
digit of previous “Percent High” setting. Press b& c13) Use band cbuttons to enter a new value for “Percent High”. Press
d
14) Display shows
STRD
stored message momentarily and then
advances to
CTRL
Control Type Submenu.
Example: On an Analog Output of 0~10 Vdc, a setting of %LO = 10 and %HI = 90, cause the minimum on the control output to be 1 V and the maximum on the control output to be 9 V. The same setting on a time proportional output, will cause 10% duty cycle for the minimum control output and 90% duty cycle for maximum control output. To disable %LO/HI, set LO to 00 and HI to 99. If
%LO/HI
is at other values than the default (%LO = 00, %HI = 99),
SOAK
is disabled.
*CONTROL TYPE OUTPUT:
(Relay, SSR, Pulse or Analog) Press
d
15) Display flashes
ON.OF
On/Off or
PID
Proportional, Integral,
Derivative.
Press
b
16) Scroll through the available selections: “ON/OFF” or “PID”.
Press
d
17) Display flashes
STRD
stored message momentarily and
then advances to
AçTN
only, if it was changed, otherwise press
a to advance to
AçTN
Action Type Submenu.
The ON/OFF control is a coarse way of controlling the process. The “Dead Band” improves the cycling associated with the On/Off control. The PID control is best for processes where the Setpoint is continuously changing and/or a tight control of the process variable is required. PID control requires tuning and adjustment of the "Proportional", "Integral or Reset" and "Derivative or Rate" terms by a trial-and-error method. The instrument provides an "Auto Tuning" feature making the tuning process automatic, possibly optimum.
* If Analog Output (Current/Voltage) is your control Output 1, this menu i.e.
CTRL
type will not appear, instead
4-20
Current will be displayed. Select
ENBL
for a
4-20 mA current (2-10 V Voltage) outputs or
DSBL
for a 0-20 mA current (0-10 V
Voltage) outputs. If 4-20 mA is enabled, %HI/LO setting will have no effect.
Both Current and Voltage control outputs are active simultaneously.
Page 34
30
ACTION TYPE SUBMENU:
The error that results from the measurement of the Process Variable may be positive or negative since it may be greater or smaller than the Setpoint. If a positive error should cause the instrument output to increase (i.e. cooling), it would be called Direct Acting. If a negative error should cause the output to increase (i.e. heating), it would be called Reverse Acting.
Press
d
18) Display flashes
DRçT
Direct or
RVRS
Reverse.
Press
b
19) Scroll through the available selections: “Direct” or “Reverse”.
Press
d
20) Display shows
STRD
stored message momentarily and then
advances to
AUTo
only, if it was changed, otherwise press ato
advance to
AUTo
Auto PID Submenu (if PID Control Type was
selected).
If “ON/OFF” was selected in the Control Type, the display skips to the Dead Band Submenu.
AUTO PID SUBMENU:
Press
d
21) Display flashes
ENBL
or
DSBL
.
Press
b
22) Scroll through the available selections: “Enable” or “Disable”.
Press
d
23) Display shows
STRD
stored message momentarily and then
advances to
ANTL
only, if it was changed, otherwise press ato
advance to
ANTL
Anti Integral Submenu.
If “Enabled”, the controller can determine, by enabling Start PID, the optimum values for the three adjustments — Proportional, Reset and Rate corresponding to P, I, and D. These values may be changed once the auto tuning is complete.
If “Disabled” is selected, the user will manually enter these three adjustment values. If you want the instrument to do the auto PID and the P , PI or PID, first select auto disable and enter 0000 for unwanted parameter. i.e. for PI enter 0000 for the rate.
ANTI INTEGRAL SUBMENU:
Press
d
24) Display flashes
ENBL
or
DSBL
.
Press
b
25) Scroll through the available selections: “Enable” or “Disable”.
Press
d
26) Display shows
STRD
stored message momentarily and then
advances to
STRT
only, if it was changed, otherwise press ato
advance to
STRT
to Start Auto Tune PID Submenu (If auto PID
was Enabled).
If Auto PID was disabled display advances to
PRoP
Proportional Band Submenu.
Page 35
31
If Anti Integral (Anti Windup) Submenu “Enabled”, this feature allows the error term outside the proportional band to be calculated and accumulated for integration. This may be an important feature in applications where fast response time is desirable.
START AUTO TUNE PID:
Press
d
27) Display flashes
ENBL
or
DSBL
.
Press
b
28) Scroll through the available selections: “Enable” or “Disable”.
Press
d
29) Display shows
STRD
stored message momentarily and then
advances to
CYCL
only, if it was changed, otherwise press ato
advance to
CYCL
Cycle Time Submenu.
If “Enabled”, an output needs to be selected from the Reading Configuration Menu
°F.°C
, or
DRRh
. This way the microcontroller knows which one of the outputs to Autotune for Temperature or Humidity. Also, the controller is ready to calculate P, PI or PID parameters. The instrument performs this by activating the output and observing the delay and rate at which the Process Value changes. The setpoints must be at least 18°F (10°C) or 10%RH above the (PV) Process Value in order to perform Auto Tune, otherwise an error message will be displayed.
To start Auto Tune PlD select PID, enable Auto PID and enable Start PID. Sometimes Auto PID parameter needs fine tuning i.e. for each 5°F over shoot increase the Proportional Band (PB) by 15% and for each ±1°F fluctuation at the Setpoint (SP) increase reset by 20%.
Once started, display shows A.TUN with letters blinking in the rotating pattern. When auto tune stops, display will show process value. Do not perform any operations or settings before first stopping Auto Tune. Any alarms or other output is disabled during Auto Tune.
If “AUTO PID” was “DISABLED”, the display will show the following three submenus. This allows the user to manually enter values for Proportional, Reset and Rate terms corresponding to P, I, and D. It also can be used for auto PID for disabling unwanted parameter i.e. PI enter 0000 for rate.
PROPORTIONAL BAND SUBMENU:
Press
d
30) Display flashes 1stdigit of the previous P
PRoP
Proportional
band value.
Press b& c31) Press band cbuttons to enter a new “Proportional Band”
value.
Press
d
32) Display shows
STRD
stored message momentarily and then
advances to
REST
only, if it was changed, otherwise press ato
advance to
REST
Reset Setup Submenu.
Proportional band is in degrees of temperature or counts of process. Proportional band is defined, as the change in the instrument input to cause a 100% change in the controller output.
Page 36
32
RESET SETUP SUBMENU:
Press
d
33) Display flashes 1stdigit of the previous I
REST
Reset value. Press b& c34) Press band cbuttons to enter a new “Reset” value. Press
d
35) Display shows
STRD
stored message momentarily and then
advances to
RATE
only, if it was changed, otherwise press ato
advance to
RATE
Rate Setup Submenu.
Reset unit is in seconds 0-3999.
RATE SETUP SUBMENU:
Press
d
36) Display flashes 1stdigit of previous D
RATE
Rate value.
Press b& c37) Press band cbuttons to enter a new
RATE
value.
Press
d
38) Display shows
STRD
stored message momentarily and then
advances to the
CYCL
only, if it was changed, otherwise press
a
to advance to
CYCL
Cycle Time submenu for RTD and
Thermocouple types.
Rate unit is in seconds 000.0-399.9.
If the Output 1 is Analog Option the display skips to Damping Factor.
CYCLE TIME SUBMENU:
Press
d
39) Display flashes 1stdigit of the previous
CYCL
Cycle Time
value.
Press b& c40) Press band cbuttons to enter a new “Cycle Time” value.
(1 to 199 seconds)
Press
d
41) Display shows
STRD
stored message momentarily and then
advances to
DPNG
only, if it was changed, otherwise press ato
advance to
DPNG
Damping Factor Submenu.
A Cycle Time selected between 1 and 199 seconds determines the total On/Off time of each proportional cycle. For example, a 15 second cycle time means that every 15 seconds the output will turn on for part or all of the cycle. For Relay control outputs, do not select a cycle time of less than 7 seconds or the relays’ lifetime will be shortened. For a cycle time of less than 7 seconds select SSR or DC pulse. Use an external SSR with the DC pulse option for higher currents (higher than 1 Amp).
Page 37
33
DAMPING FACTOR SUBMENU:
Press
d
42) Display flashes the previous “Damping Factor” selection.
Press
b
43) Scroll through the available selections:
0000,0001,0002
,
0003,0004,0005,0006,0007
.
Press
d
44) Display flashes
STRD
stored message and then advances
to
OUT2
only, if it was changed, otherwise press ato advance
to
OUT2
Output 2 Menu.
Damping Factor is a measure of speed, overshoot, and undershoot in which the process variable responds to the output changes of the instrument, which were used during the Auto Tune. This value is typically set to the ratio of Rate to Reset. This Default value is (0003). For fast response time, this value should be decreased while for slow response time it should be increased.
The "DEADBAND" Submenu will only appear if "ON/OFF" was selected from the "Control Type" Menu.
DEADBAND SUBMENU:
Press
d
45) Display flashes 1stdigit of the previous
DEAD
Deadband
value. Press b& c46) Press band cbuttons to enter a new “Deadband” value. Press
d
47) Display shows
STRD
stored message and then advances to
OUT2
only, if it was changed, otherwise press ato advance to
OUT2
Output 2 Menu.
Dead Band units are the same as Proportional Band units.
The Dead Band or neutral zone is the percentage (humidity) or number of degrees around the Setpoint which the Process Variable must pass above or below the Setpoint, before the output changes state.
Page 38
34
3.2.10 Output 2
Output 2 and Alarm 2 share the same contacts on the rear panel connector. If Alarm 2 is Enabled, Output 2 is automatically Disabled.
Figure 3.9 Flow Chart for Output 2
ENTER OUTPUT 2 MENU:
Press
a
1) Press a, if necessary, until
CNFG
prompt appears.
Press
d
2) Display advances to
RDG
Reading Configuration Menu..
Press
a
3) Press a, if necessary, until Display advances to
OUT2
Output 2 Menu.
Press
d
4) Display advances to
CTRL
Control Type Submenu.
CONTROL TYPE SUBMENU:
Press
d
5) Display flashes
ON.OF
ON/OFF, or
PID
PID.
Press b 6) Scroll through the available selections: “ON/OFF” or “PID”. Press
d
7) Display shows
STRD
stored message momentarily and then
advances to
AçTN
only, if it was changed, otherwise press ato
advance to
AçTN
Action Type Submenu.
The ON/OFF control is a coarse way of controlling the Process. The “Dead Band” improves the cycling associated with the ON/Off control. The PID control is best for processes where the Setpoint is continuously changing and/or tight control of the Process Variable is required.
It is required that you put the controller in the Standby Mode for any configuration change other than Set Points & Alarms. Alarm 2 and Output 2 works for Temperature only.
RESET
RATE
RES/VALUE
RATE/VALUE
Page 39
35
ACTION TYPE SUBMENU:
The error that results from the measurement of the Process Variable may be positive or negative since it may be greater or smaller than the Setpoint. If a positive error should cause the instrument output to increase (i.e. cooling), it would be called Direct Acting. If a negative error should cause the output to decrease (i.e. heating), it would be called Reverse Acting.
Press
d
8) Display flashes
DRçT
Direct or
RVRS
Reverse.
Press
b
9) Scroll through the available selections: “Direct” or “Reverse”.
Press
d
10) Display shows
STRD
stored message momentarily and then
advances to
AUTo
only, if it was changed, otherwise press ato
advance to
AUTo
Auto PID Submenu (If PID Control type was
selected).
If ON/OFF was selected in the Control Type, the display skips to
the Dead Band Submenu.
AUTO PID SUBMENU:
Press
d
11) Display flashes
ENBL
Enable or
DSBL
Disable.
Press
b
12) Scroll through the available selections: “Enable” or “Disable”.
If "Enabled", the PID parameter of Output 1 will be copied to
Output 2.
Press
d
13) Display shows
STRD
stored message momentarily and then advances to the next submenu only, if it was changed, otherwise press ato advance to the next submenu.
If AUTO PID was ENABLED", the display skips to the
CYCL
CYCLE TIME submenu. If "AUTO PID" was "DISABLED", the display will show
PRoP
PROPORTIONAL BAND Submenu
allowing the user to manually enter the Proportional Band value.
Refer to “Proportional Band”, “Reset”, “Rate”, and “Cycle Time” Submenus of “Output 1” Menu.
DEADBAND SUBMENU: The DEADBAND Submenu will only appear if the ON/OFF was selected from the "Control Type" Submenu.
Press
d
14) Display flashes 1stdigit of the previous “Dead Band” value.
Press b& c15) Press band cbuttons to enter a new “Dead Band” value. Press
d
16) Display shows
STRD
stored message momentarily and then
advances to
RAMP
only, if it was changed, otherwise press ato
advance to
RAMP
Ramp Value Menu.
Dead Band units are the same as Proportional Band units.
The Dead Band is the number of degrees or counts around the Setpoint which the Process Variable must pass through before the output changes state.
Page 40
36
3.2.11 Ramp & Soak
Alarm must be DISABLED if Ramp is ENABLED.
Figure 3.10 Flow Chart for Ramp and Soak
ENTER RAMP AND SOAK MENU:
Press
a
1) Press a, if necessary, until
CNFG
prompt appears.
Press
d
2) Display advances to
RDG
Reading Configuration Menu.
Press
a
3) Press a, if necessary, until Display advances to
RAMP
Ramp
and
SOAK
Soak Menu.
RAMP ENABLE/DISABLE SUBMENU:
Press
d
4) Display advances to “Ramp Enable/Disable” Submenu and flashes
ENBL
or
DSBL
.
Press
b
5) Scroll through the available selections: “Enable” or “Disable”.
Press
d
6) Display shows
STRD
stored message momentarily and then
advances to
SOAK
Soak Enable/Disable Menu.
If
RAMP
Disable was selected, display skips to the next menu
item (ID Code).
It is required that you put the controller in the Standby Mode for any
configuration change other than Set Points & Alarms.
If
DRRh
is selected in the Reading Configuration Menu the unit will
ramp only on humidity, but if
°F.°C
is selected, both humidity and
temperature will ramp to Setpoint.
Page 41
37
SOAK ENABLE/DISABLE SUBMENU:
Press
d
7) Display flashes
ENBL
or
DSBL
.
Press
b
8) Scroll through the available selections: “Enable” or “Disable”.
Press
d
9) Display shows
STRD
stored message momentarily and then
advances to “Ramp Value” Submenu.
Ramp & Soak provides users with the flexibility to slowly bring the Process Variable (PV) to the desired setpoint. Ramp & Soak values are specified in HH.MM format. The Ramp value indicates the time specified to bring the process variable to Setpoint 1 (SP1). Once the set point is reached, the PID takes over and the Process Variable will be controlled at the desired set point indefinitely. If Soak is enabled, PID will control the Process Variable at the specified Setpoint for the duration of Soak time and then will turn off Output 1. To start a new Ramp/Soak cycle, reset the instrument by pressing aand then
c
button.
An active Ramp/Soak will change SP1 one degree above the PV and will cause the most significant digit to blink. The SP1 will be incremented by one degree until it reaches the original SP1. The minimum Ramp time must be at least twice the time that it will take the PV to reach the Setpoint Value (SV) with OUT 1 fully ON.
RAMP VALUE SUBMENU:
Press
d
10) Display flashes 1stdigit of previous stored “Ramp Value”.
Press b& c11) Press band cbuttons to enter a new “Ramp Value”. Press
d
12) Display shows
STRD
stored message momentarily and then
advances to “Soak Value” Submenu.
SOAK VALUE SUBMENU:
Press
d
13) Display flashes 1stdigit of previous stored “Soak Value”.
Press b& c14) Press band cbuttons to enter a new “Soak Value”. Press
d
15) Display shows
STRD
stored message and advances to the
ID
ID Code Menu.
The Ramp and Soak time is 00:00 to 99:59 i.e. HH.MM. (from zero to 99 hours and 59 minutes) During Ramp & Soak do not perform any operations or settings before first stopping it. Any alarms or other output are disabled during this time. To stop Ramp & Soak first put instrument into Standby Mode, then go to Ramp & Soak Menu and disable it.
Page 42
38
3.2.12 ID CODE
Figure 3.11 Flow Chart for ID Code
ENTER ID CODE MENU:
Press
a
1) Press a, if necessary, until
CNFG
prompt appears.
Press d 2) Display advances to
RDG
Reading Configuration Menu.
Press
a
3) Press a, if necessary, until Display advances to IDID Code Menu.
ENTERING OR CHANGING YOUR (NON-DEFAULT) ID CODE:
Press d 4) Display advances to
____
with 1stunder score flashing.
Press b& c5) Press band cto enter your 4-digit “ID Code” number. Press d 6) Display advances to
CH.ID
Change ID Code Submenu.
If entered “ID Code” is incorrect display shows
ERRo
Error
message momentarily and then skips to the Run Mode.
Press
d
7) Display flashes the first digit of previous entered “ID Code” number.
Press b& c8) Press band cbuttons to enter your new “ID Code” number. Press
d
9) Display shows
STRD
stored message momentarily and then
advances to the
FULL
Full Security Submenu.
Page 43
39
ENTERING OR CHANGING YOUR (DEFAULT) ID CODE:
Enter IDmenu (Repeat steps from 1 to 3).
Press
d
10) Display advances to
CH.ID
Change ID Code Submenu.
Press
d
11) Display shows
0000
message with flashing 1stdigit.
If you want to change your default “ID Code” you can do it now, otherwise press aand menu will skip to
FULL
Full Security
Submenu.
Press b& c12) Press band cbuttons to enter your new “ID Code” number. Press
d
13) Display shows
STRD
stored message momentarily and then
advances to the
FULL
Full Security Submenu.
FULL SECURITY LEVEL SUBMENU:
Press
d
14) Display flashes
ENBL
Enable or
DSBL
Disable.
Press
b
15) Scroll through the available selections: “Enable” or “Disable”.
Press
d
16) Display shows
STRD
stored message momentarily and then
advances to
SP.ID
Setpoint/ID Submenu.
If "Full" Security Level is "Enabled" and the user attempts to enter the Main Menu, they will be prompted for an ID Code. The ID Code should be correct to enter the instrument Menu item.
SETPOINT/ID SECURITY LEVEL SUBMENU:
This Security Level can be functional only if
FULL
Security
Level is Disabled.
Press
d
17) Display flashes
ENBL
Enable or
DSBL
Disable.
Press
b
18) Scroll through the available selections: “Enable” or “Disable”.
Press
d
19) Display shows
STRD
stored message momentarily and then
advances to
COMM
Communication Submenu.
If "Setpoint/ID" Security Level is "Enabled" and the user attempts to advance into the
CNFG
Configuration Menu, he will be prompted for ID Code number. The ID Code should be correct to proceed into the Configuration Menu, otherwise display will show an Error and skip to the Run Mode.
If “Full” and “Setpoint/ID” Security Levels are "Disabled", the ID code will be “Disabled” and user will not be asked for ID Code to enter the Menu items (“ID” Submenu will not show up in “ID/Setpoint” Menu).
Page 44
40
3.2.13 COMMUNICATION OPTION
Purchasing the controller with Serial Communications permits an instrument to be configured or monitored from an IBM PC compatible computer using software available from the website or on the CD-ROM enclosed with your shipment.
For complete instructions on the use of the Communications Option, refer to the Serial Communications Reference Manual.
Figure 3.12 Flow Chart for Communication Option
With the exception of DATA FORMAT Sub Menu, all other command formats are the same in the Serial
Communications Manual.
Page 45
41
The following table are the exceptions to the Serial Communication Manual’s Table 5.3
Table 3.2 Command Letters and Suffix for iTH
ENTER COMMUNICATION OPTION MENU:
Press
a
1) Press a, if necessary, until
CNFG
prompt appears.
Press
d
2) Display advances to
RDG
Reading Configuration Menu.
Press
a
3) Press a, if necessary, until Display advances to
COMM
Communication Options Menu.
Press
d
4) Display advances to
C.PAR
Communication Parameters
Submenu.
If Communication Option is not installed, the display shows
NONE
and skips to the Color Display Menu.
COMMUNICATION PARAMETERS SUBMENU:
Allows the user to adjust Serial Communications Settings of the instrument. When connecting an instrument to a computer or other device, the Communications Parameters must match. Generally the default settings (as shown in Section 5) should be utilized.
Press
d
5) Display advances to
BAUD
Baud Submenu.
BAUD SUBMENU:
Press
d
6) Display flashes previous selection for
BAUD
value.
Press
b
7) Scroll through the available selections:
300_, 600_, 1200
,
2400, 4800
,
9600, 19.2K
.
Press
d
8) Display shows
STRD
stored message momentarily and then
advances to
PRTY
only, if it was changed, otherwise press ato
advance to
PRTY
Parity Submenu.
PARITY SUBMENU:
Press
d
9) Display flashes previous selection for “Parity”.
Press
b
10) Scroll through the available selections: NO, ODD, EVEN.
Press
d
11) Display shows
STRD
stored message momentarily and then
advances to
DATA
only, if it was changed, otherwise press ato
advance to
DATA
Data Bit Submenu.
DATA BIT SUBMENU:
Press
d
12) Display flashes previous selection for “Data Bit”.
Press
b
13) Scroll through the available selections: 7-BIT, 8-BIT.
Press
d
14) Display shows
STRD
stored message and then advances to
STOP
only, if it was changed, otherwise press ato advance to
STOP
Stop Bit Submenu.
Command Command Function Command # Of Default
Index Bytes Characters Value
X 01 Send RH Reading 0 0 ­X 02 Send Temperature Reading 0 0 ­X 03 Send Dewpoint Reading 0 0 -
Page 46
42
STOP BIT SUBMENU:
Press
d
15) Display flashes previous selection for “Stop Bit”.
Press
b
16) Scroll through the available selections:
1-BIT, 2-BIT
.
Press
d
17) Display shows
STRD
stored message momentarily and then
advances to
BUS.F
only, if it was changed, otherwise press ato
advance to
BUS.F
Bus Format Submenu.
BUS FORMAT SUBMENU:
Determines Communications Standards and Command/Data Formats for transferring information into and out of the controller via the Serial Communications Bus. Bus Format submenus essentially determine how and when data can be accessed via the Serial Communications of the device.
Press
d
18) Display advances to
M.BUS
Modbus Submenu.
MODBUS PROTOCOL SUBMENU:
Press
d
19) Display flashes previous selection for
M.BUS
.
Press
b
20) Scroll through the available selections: NO, YES.
Press
d
21) Display shows
STRD
stored message momentarily and then
advances to
_LF_
only, if it was changed, otherwise press ato
advance to
_LF_
Line Feed submenu.
To select iSeries Protocol, set Modbus submenu to “No”. To select Modbus Protocol, set Modbus submenu to “Yes”.
If Modbus Protocol was selected, the following Communications Parameters must be set as: No Parity, 8-bit Data Bit, 1-Stop Bit. Do not attempt to change these parameters.
LINE FEED SUBMENU:
Determines if data sent from the instrument will have a Line Feed appended to the end - useful for viewing or logging results on separate lines when displayed on communications software at a computer.
Press
d
22) Display flashes previous selection for “Line Feed”.
Press
b
23) Scroll through the available selections: NO, YES.
Press
d
24) Display shows
STRD
stored message momentarily and then
advances to
ECHO
only, if it was changed, otherwise press ato
advance to
ECHO
Echo Submenu.
ECHO SUBMENU:
When valid commands are sent to the instrument, this determines whether the command will be echoed to the Serial Bus. Use of echo is recommended in most situations, especially to help verify that data was received and recognized by the controller.
Page 47
43
Press
d
25) Display flashes previous selection for “Echo”.
Press
b
26) Scroll through the available selections: NO, YES.
Press
d
27) Display flashes
STRD
stored message momentarily and then
advances to
STND
only if it was changed, otherwise press ato
advance to
STND
Communication Standard Submenu.
COMMUNICATION INTERFACE STANDARD SUBMENU:
Determines whether device should be connected to an RS-232C serial port (as is commonly used on IBM PC-compatible computers) or via an RS-485 bus connected through appropriate RS-232/485 converter. When used in RS-485 Mode, the device must be accessed with an appropriate Address Value as selected in the Address Submenu described later.
Press
d
28) Display flashes previous selection for “Standard”.
Press
b
29) Scroll through the available selections: 232C, 485.
Press
d
30) Display shows
STRD
stored message momentarily and then
advances to
MoDE
only, if it was changed, otherwise press ato
advance to
MoDE
Data Flow Mode Submenu.
DATA FLOW MODE SUBMENU:
Determines whether the instrument will wait for commands and data requests from the Serial Bus or whether the instrument will send data automatically and continuously to the Serial Bus. Devices configured for the RS-485 Communications Standard operate properly only under Command Mode.
Press
d
31) Display flashes previous selection for “Mode”.
Press
b
32) Scroll through the available selections:
CMD
__
“Command”,
CoNT
“Continuous”.
Press
d
33) Display shows
STRD
stored message momentarily and then
advances to
SEPR
only, if it was changed, otherwise press ato
advance to
SEPR
Data Separation Submenu.
DATA SEPARATION CHARACTER SUBMENU:
Determines whether data sent from the device in Continuous Data Flow Mode will be separated by spaces or by Carriage Returns.
Press
d
34) Display flashes previous selection for “Separation” Submenu.
Press
b
35) Scroll through the available selections:
SPCE
“Space” or
_çR_
“Carriage Return”.
Press
d
36) Display shows
STRD
stored message momentarily and then
advances to
DAT.F
only, if it was changed, otherwise press ato
advance to
DAT.F
Data Format Submenu.
Page 48
44
DATA FORMAT SUBMENU:
Preformatted data can be sent automatically or upon request from the controller. Use the Data Format Submenus to determine what data will be sent in this preformatted data string. Refer to the iSeries Communications Manual for more information about the data format. At least one of the following suboptions must be enabled and hence output data to the Serial Bus.
This menu is applicable for Continuous Mode of RS-232 communication.
Press
d
37) Display advances to
STAT
Alarm Status Submenu.
ALARM STATUS SUBMENU:
Includes Alarm Status bytes in the data string.
Press
d
38) Display flashes previous selection for “Status” (alarm status).
Press
b
39) Scroll through the available selections: NO, YES.
Press
d
40) Display shows
STRD
stored message momentarily and then
advances to
HUMD
only, if it was changed, otherwise press ato
advance to
HUMD
Humidity Submenu.
HUMIDITY READING SUBMENU:
Includes Humidity Reading in the data string.
Press
d
41) Display flashes
HUMD
.
Press
b
42) Scroll through the available selections: NO, YES.
Press
d
43) Display shows
STRD
stored message momentarily and then
advances to
TEMP
only, if it was changed, otherwise press ato
advance to
TEMP
Temperature Submenu.
TEMPERATURE READING SUBMENU:
Includes Temperature Reading in the data string.
Press
d
44) Display flashes
TEMP
.
Press
b
45) Scroll through the available selections: NO, YES.
Press
d
46) Display shows
STRD
stored message momentarily and then
advances to
DEU
only, it was changed, otherwise press ato
advance to
DEU
Dewpoint Submenu.
DEWPOINT READING SUBMENU:
Includes Dewpoint Reading in the data string.
Press
d
47) Display flashes
DEU
.
Press
b
48) Scroll through the available selections: NO, YES.
Press
d
49) Display shows
STRD
stored message momentarily and then
advances to
UNIT
only, if it was changed, otherwise press ato
advance to
UNIT
Temperature Unit Submenu.
Page 49
45
TEMPERATURE UNIT SUBMENU:
Includes a byte in the data string to indicate whether reading is in Celsius or Fahrenheit.
Press
d
50) Display flashes previous selection for
UNIT
.
Press
b
51) Scroll through the available selections: NO, YES.
Press
d
52) Display shows
STRD
stored message momentarily and then
advances to
ADDR
only, if it was changed, otherwise press ato
advance to
ADDR
Address Setup Submenu.
ADDRESS SETUP SUBMENU:
This menu is applicable to the RS-485 Option only.
Press
d
53) Display advances to “Address Value” (0000 to 0199) Submenu.
ADDRESS VALUE SUBMENU:
Press
d
54) Display flashes 1
st
digit of previously stored Address Value. Press b& c55) Press band cto enter new “Address Value”. Press
d
56) Display shows
STRD
stored message momentarily and then
advances to
TR.TM
only, if it was changed, otherwise press ato
advance to
TR.TM
Transmit Time Interval Submenu.
TRANSMIT TIME INTERVAL SUBMENU:
This menu is applicable if “Continuous” Mode was selected in the “Data Flow Mode” Submenu and the device is configured as an RS-232C Standard device. Also, one or more options under the Data Format Submenu must be enabled.
Press
d
57) Display advances to “Transmit Time Value” Submenu.
TRANSMIT TIME INTERVAL VALUE SUBMENU:
Determines the interval at which data will be emitted to the RS-232 Serial Bus when the instrument is in Continuous Data Flow Mode.
Press
d
58) Display flashes 1
st
digit of previous “Transmit Time Value” in
seconds.
Press b& c59) Press band cto enter new “Transmit Time Value”, e.g.
0030 will send the data every 30 seconds in Continuous Mode.
Press
d
60) Display shows
STRD
stored message momentarily and then
advances to
COLR
only, if it was changed, otherwise press ato
advance to
COLR
Color Display Selection Menu.
For more details, refer to the Serial Communication Manual available at the website listed in the cover page of this manual.
Page 50
46
3.2.14 DISPLAY COLOR SELECTION
This submenu allows the user to select the color of the upper display.
Figure 3.13 Flow Chart for Display Color Selection
ENTER DISPLAY COLOR SELECTION MENU:
Press
a
1) Press a, if necessary, until
CNFG
prompt appears.
Press
d
2) Display advances to
RDG
Reading Configuration Menu.
Press
a
3) Press a, if necessary, until Display advances to
COLR
Display Color Selection Menu.
Press
d
4) Display advances to
N.CLR
Normal Color Submenu.
NORMAL COLOR DISPLAY SUBMENU:
Press
d
5) Display flashes the previous selection for “Normal Color”.
Press
b
6) Scroll through the available selections:
GRN, RED
or
AMBR
.
Press
d
7) Display shows
STRD
stored message momentarily and then
advances to
1.CLR
only, if it was changed, otherwise press ato
advance to
1.CLR
Alarm 1 Display Color Submenu.
The menu below allows the user to change the color of the upper display when Alarm 1 is triggered.
ALARM 1 DISPLAY COLOR SUBMENU:
Press
d
8) Display flashes previous selection for “Alarm 1 Color Display”.
Press
b
9) Scroll through the available selections:
GRN, RED
or
AMBR
.
Press
d
10) Display shows
STRD
stored message momentarily and then
advances to
2.CLR
only, if it was changed, otherwise press ato
advance to
2.CLR
Alarm 2 Display Color Submenu.
A color change is based on the Relative Humidity value only. If this value causes an alarm condition, the upper display will change to the selected Alarm 1 Color. A color change will occur whether Alarm 1 is enabled or disabled.
Page 51
47
ALARM 2 DISPLAY COLOR SUBMENU:
Press
d
11) Display flashes previous selection for “Alarm 2 Color Display”.
Press
b
12) Scroll through the available selections:
GRN, RED
or
AMBR
.
Press
d
13) Display shows
STRD
stored message momentarily and then
momentarily shows the software version number, followed by
RST
Reset, and then proceeds to the Run Mode.
IN ORDER TO DISPLAY ONE COLOR, SET THE SAME DISPLAY COLOR ON ALL THREE SUBMENUS ABOVE.
Example 1:
Output 1 & Output 2 = SSR Alarm Setup: Absolute, Above, Alarm 2 HI Value “ALR.H” = 200, Alarm 1 HI Value “ALR.H” = 400 "Color Display" Setup: Normal Color “N.CLR” = Green, Alarm 1 Color “1.CLR” = Amber, Alarm 2 Color “2.CLR” = Red
Display Colors change sequences:
GREEN RED AMBER
--➤------------------------------•-----------------------------•------------------------------
0 AL2.H = 200 AL1.H = 400
Example 2:
Output 1 & Output 2
= Pulse Alarm Setup: Absolute, Below, Alarm 2 Low Value “ALR.L” = 300, Alarm 1 Low Value “ALR.L” = 100 Color Display Setup: "N.CLR" = Green, "1.CLR" = Amber, "2.CLR" = Red
Display Colors change sequences:
AMBER RED GREEN
--------------•----------------------------------•------------------------------------------- --
0 AL1.L = 100 AL2.L = 300
Page 52
48
Example 3:
Output 1 = Analog Output (Alarm 1 disabled), Setpoint 1 = 300, Output 2 = Relay, Setpoint 2 = 200 Alarm 1 & 2 Setup: Deviation, Band, “ALR.H” = 10 Color Display Setup: “N.CLR” = Green, “1.CLR” = Amber, “2.CLR” = Red
Display Colors change sequences:
RED RED RED GREEN RED
---------------•------•------•--------------------------------•------•-------•----------------
0 190 200 210 290 300 310
Alarm 1 is designed to monitor the Process Value around the Setpoint 1. Alarm 2 is designed to monitor the Process Value around the Setpoint 2. If Analog Output Option board is installed (Alarm 1 is disabled), only Alarm 2 is active and only two colors are available.
Example 4:
Output 1
= Relay, Setpoint 1 = 200 Output 2 = Relay, Setpoint 2 = 200 Alarm 1 Setup
: Deviation, Band, “ALR.H” = 20
Alarm 2 Setup
: Deviation, Hi/Low, “ALR.H” = 10, “ALR.L” = 5
Color Display Setup: “N.CLR” = Green, “1.CLR” = Amber, “2.CLR” = Red
Display colors change sequences:
AMBER RED GREEN GREEN RED AMBER
--➤---------------•----------------•-------------•--------------•-------------•---------------------
0 180 195 200 210 220
Reset: The instrument automatically resets after the last menu of the Configuration Mode has been entered. After the instrument resets, it advances to the Run Mode.
Page 53
49
PART 4
SPECIFICATIONS
SENSOR SPECIFICATIONS
Relative Humidity (RH) Accuracy/Range: ±2% for 10 to 90% RH
±3% for 0 to 10%RH and 90 to 100%RH
Non-linearity: ±3% Hysteresis: ±1% RH Response Time:
4 sec (63% slowly moving air)
Repeatability: ±0.1% Resolution: 0.03%, 12bit NOTE: Reconditioning of the probe may
be necessary if the probe is stored for a period of time in a harsh environment (e.g. high humidity or exposure to chemicals). To recondition the probe: heat probe for 1 day at 100°C to return it to calibration conditions.
Temperature (T) Accuracy/Range*:
±1°C (±2°F) for
-40 to 0°C and 80 to 123.8°C (-40 to 32°F and 176 to 254°F)
±0.5°C (±1°F) for 0 to 80°C (32 to 176°F) *NOTE: extended temperature range is
for Probe only, the Controller’s operating temperature is 0-50°C
Response Time: 5 sec (63% slowly moving air)
Repeatability: ±0.1°C Resolution: 0.01°C, 14 bit
METER SPECIFICATIONS
NMRR: 60 dB CMRR: 120 dB A/D Conversion:
12 bit RH and 14 bit Temperature
Reading Rate:
2 samples per seconds max.
Digital Filter: Programmable Decimal Selection:
None, 0.1 for temp and humidity
Warm up to Rated Accuracy: 30 min.
Display
4-digit, 9-segment LED
10.2mm (0.40"): i32, i16D,
i8DV
• 10.2mm (0.40”) & 21mm (0.83”): i8DH red, green and amber programmable
colors for process variable, set point and temperature units
CONTROL
Action Reverse (heat) or direct (cool)
Modes
Time and Amplitude Proportional Control Modes; selectable Manual or Auto PID, Proportional, Proportional with Integral, Proportional with Derivative with Anti-reset Windup and ON/OFF
Rate
0 to 399.9 seconds
Reset
0 to 3999 seconds
Cycle Time
1 to 199 seconds; set to 0 for ON/OFF operation
Gain
0.5 to 100% of span; Setpoints 1 or 2
Damping
0000 to 0008
Soak
00.00 to 99.59 (HH:MM), or OFF
Ramp to Setpoint
00.00 to 99.59 (HH:MM), or OFF
Auto Tune
Operator initiated from front panel for one input at a time only
Page 54
50
CONTROL OUTPUT 1 & 2 Relay
250 Vac or 30 Vdc @ 3 A (Resistive Load); configurable for on/off, PID and Ramp and Soak
Output 1: SPDT type, can be configured as Alarm 1 output
Output 2: SPDT type, can be configured as Alarm 2 output
SSR
20-265 Vac @ 0.05-0.5 A (Resistive Load); continuous
DC Pulse
Non-Isolated; 10 Vdc @ 20 mA Analog Output (Output 1 only) Non-Isolated, Proportional 0 to 10
Vdc or 0 to 20 mA; 500 Ω max
NETWORK AND COMMUNICATIONS (Optional -C24, -C4EI, -EI)
Ethernet: Standards Compliance
IEEE 802.3 10Base-T Supported Protocols: TCP/IP,
ARP, HTTPGET
RS-232/RS-422/RS-485/MODBUS:
Selectable from menu; both ASCII and modbus protocol selectable from menu.
Programmable 300 to 19.2 K baud; complete programmable setup capability; program to transmit current display, alarm status, min/max, actual measured input value and status.
RS-485
Addressable from 0 to 199
Connection
Screw terminals
ALARM 1 & 2 (programmable): Type
Same as Output 1 & 2
Operation
High/low, above/below, band, latch/unlatch, normally open/normally closed and process/deviation; front panel configurations
ANALOG OUTPUT (programmable)
Non-Isolated, Retransmission 0 to 10 Vdc or 0 to 20 mA, 500 Ω max (Output 1 only). Accuracy is +
1% of FS when
following conditions are satisfied.
1) Input is not scaled below 1% of Input FS.
2) Analog Output is not scaled below 3% of Output FS.
EXCITATION (optional in place of Communication)
24 Vdc @ 25 mA Not available for Low Power Option
INSULATION Power to Input/Output
2300 Vac per 1 min. test 1500 Vac per 1 min. test
(Low Voltage/Power Option)
Power to Relays/SSR Outputs
2300 Vac per 1 min. test
Relays/SSR to Relay/SSR Outputs
2300 Vac per 1 min. test
RS-232/485 to Inputs/Outputs
500 Vac per 1 min. test
APPROVALS
UL, C-UL, and see CE Approval Section
Page 55
51
GENERAL Line Voltage/Power
90-240 Vac +/-10%, 50-400 Hz* 110-375 Vdc, equivalent voltage
4 W, power for i32 Models 5 W, power for i8DV, i8DH, i16D Models
* No CE compliance above 60 Hz
Low Voltage/Power Option
12-36 Vdc or 24 Vac
**
+/-10%
, 3 W
External power source must meet Safety Agency Approvals.
** Units can be powered safely with 24 Vac
power but, no Certification for CE/UL are claimed.
External Fuse Required
Time-Delay, UL 248-14 listed:
100 mA/250 V 400 mA/250 V
(Low Voltage/Power Option)
Time-Lag, IEC 127-3 recognized:
100 mA/250 V 400 mA/250 V
(Low Voltage/Power Option)
Environmental Conditions
i32: 0 to 55°C (32 to 131°F), 90% RH non-condensing
i8DV, i8DH, i16D: 0 to 50°C (32 to 122°F), 90% RH non-condensing
Cable: operating temperature 0-105°C (32 to 221°F)
Protection
NEMA-4x/Type 4/IP65 front bezel: i32, i16D
NEMA-1/Type 1 front bezel: i8DH, i8DV
Dimensions i/8 Series:
48 H x 96 W x 127 mm D (1.89 x 3.78 x 5")
i/16 Series:
48 H x 48 W x 127 mm D (1.89 x 1.89 x 5")
i/32 Series:
25.4 H x 48 W x 127 mm D (1.0 x 1.89 x 5")
Industrial Probe iTHP-2:
16mm Dia. x 51mm Long (0.63” x 2”) with 0.9m long (3’ ) cable
Industrial Probe iTHP-5:
16mm Dia. x 137mm Long (0.63” x 5”) with 6.1m long (20’ ) cable
Material: Probe Housing, SS316
Panel Cutout i/8 Series:
45 H x 92 mm W (1.772" x 3.622 "), 1/8 DIN
i/16 Series:
45 mm (1.772") square, 1/16 DIN
i/32 Series:
22.5 H x 45 mm W (0.886" x 1.772"), 1/32 DIN
Weight
i/8 Series: 295 g (0.65 lb) i/16 Series: 159 g (0.35 lb) i/32 Series: 127 g (0.28 lb)
Page 56
52
PART 5
FACTORY PRESET VALUES
Table 5.1 Factory preset value
MENU ITEMS FACTORY PRESET VALUES NOTES
Set Point 1 (SP1) 000.0 Set Point 2 (SP2) 000.0
Reading Configuration (RDG):
Sensor (SENS) %RH Decimal Point FFF.F
not menu selectable
Temperature unit (tEMP) °F Filter value (FLtR) 0004
Alarm 1 & 2
:
Alarm 1 (ALR1), Alarm 2 (ALR2) Disable (dSbL) Absolute/Deviation (AbSO/dEV) Absolute (AbSO) Latch/Unlatch (LtCH/UNLt) Unlatch (UNLt) Contact Closure (Ct.CL) Normally Open (N.O.) Active (ACtV) Above (AbOV) Alarm At Power On (A.P.ON) Disable (dSbL) Alarm 1 only Alarm Low (ALR.L) 000.0 Alarm High (ALR.H) 80.0
LOOP:
Loop Break Time (LOOP) Disable (dSbL) Loop Value (B.tIM) 00:59 Reading Adjust Value (R.AdJ) 000.0 Setpoint Deviation (SP.dV) Disabled (dSbL)
ANALOG OUTPUT (Retransmission)
:
Analog Output (ANLG) Enabled (ENbL) Current/Voltage (CURR/VOLt) Voltage (VOLt) Scale and Offset Reading: 0 - 100.0 cts,
Output: 0 - 10 V
OUTPUT 1 & 2
:
Self (SELF) Disabled (dSbL) Output 1 only % Low Value (%LO) 0000 Output 1 only % High Value (%HI) 0099 Output 1 only Control Type (CtRL) On/Off Action Type (ACtN) Reverse (RVRS) Dead Band (dEAd) 020.0 PID Auto (AUtO) Disable (dSbL) Anti Integral (ANtI) Disable (dSbL) Output 1 only Proportion Value (PROP) 020.0 Reset Value (RESt) 0180 Rate Value (RAtE) 0000 Cycle Value (CYCL) 0007 Damping Factor (DPNG) 0003
Page 57
53
MENU ITEMS FACTORY PRESET VALUES NOTES Ramp & Soak (RAMP)
:
Ramp (RAMP) Disable (dSbL) Soak (SOAK) Disable (dSbL) Ramp Value (RAMP) 00:00 Soak Value (SOAK) 00:00
ID
:
ID Value 0000 Full ID (FULL) Disable (dSbL) Set Point ID (Id.SP) Disable (dSbL)
Communication Parameters:
Baud Rate (BAUd) 9600 Parity (PRtY) Odd Data bit (DAtA) 7 bit Stop Bit (StOP) 1 bit Modbus Protocol (M.bUS) No Line Feed (_LF_) No Echo (ECHO) Yes Standard Interface (StNd) RS-232 (232C) Command Mode (MOdE) Command (CMd_) Separation (SEPR) Space (SPCE) Alarm Status (StAt) No Humidity (HUMd) Yes Temperature (TEMP) No Dewpoint (dEU) No Units (UNIt) No Multipoint Address (AddR) 0001 Transmit Time (tR.tM) 0016
Display Color (COLR):
Normal Color (N.CLR) Green (GRN) Alarm 1 Color (1.CLR) Red (RED) Alarm 2 Color (2.CLR) Amber (AMbR)
Page 58
54
PART 6
CE APPROVALS INFORMATION
This product conforms to the EMC directive 89/336/EEC amended by 93/68/EEC, and with the European Low Voltage Directive 72/23/EEC.
Electrical Safety EN61010-1:2001
Safety requirements for electrical equipment for measurement, control and laboratory.
Double Insulation Pollution Degree 2 Dielectric withstand Test per 1 min
Power to Input/Output: 2300Vac (3250Vdc)
Power to Input/Output: 1500Vac (2120Vdc)
(Low Voltage dc Power Option*)
Power to Relays/SSR Output: 2300Vac (3250Vdc)
Ethernet to Inputs: 1500Vac (2120Vdc)
Isolated RS232 to Inputs: 500Vac (720Vdc)
Isolated Analog to Inputs: 500Vac (720Vdc)
Analog/Pulse to Inputs: No Isolation
Measurement Category I
Category I are measurements performed on circuits not directly connected to the Mains Supply (power). Maximum Line-to-Neutral working voltage is 50Vac/dc. This unit should not be used in Measurement Categories II, III, IV.
Transients Overvoltage Surge (1.2 / 50uS pulse)
Input Power: 2500V
Input Power: 1500V
(Low Voltage dc Power Option*)
Ethernet: 1500V
Input/Output Signals: 500V
Note: *Units configured for external low power dc voltage, 12-36Vdc
EMC EN61326:1997 + and A1:1998 + A2:2001
Immunity and Emissions requirements for electrical equipment for measurement, control and laboratory.
EMC Emissions Table 4, Class B of EN61326
EMC Immunity** Table 1 of EN61326
Note: **I/O signal and control lines require shielded cables and these cables
must be located on conductive cable trays or in conduits. Furthermore, the length of these cables should not exceed 30 meters
Refer to the EMC and Safety installation considerations (Guidelines) of this manual for additional information.
Page 59
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of one (1) year from the date of purchase. In addition to OMEGA’s standard warranty period, OMEGA Engineering will extend the warranty period for four (4) additional years if the warranty card enclosed with each instrument is returned to OMEGA.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA:
1. Purchase Order number under which the product was PURCHASED,
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the product.
FOR NON-WARRANTY REPAIRS,
consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST of the repair,
2. Model and serial number of product, and
3. Repair instructions and/or specific problems relative to the
product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
© Copyright 2005 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
TRADEMARK NOTICE:
®
, omega.com
®
, , and ®are Trademarks of OMEGA ENGINEERING, INC.
PATENT NOTICE: This product is covered by one or more of the following patents: U.S. Pat. No. Des. 336,895; 5,274,577; 6,243,021 / CANADA 2052599; 2052600 / ITALY 1249456; 1250938 / FRANCE BREVET No. 91 12756 / SPAIN 2039150; 2048066 / UK PATENT No. GB2 249 837; GB2 248 954 / GERMANY DE 41 34398 C2. The “Meter Bezel Design” is a Trademark of NEWPORT Electronics, Inc. Used under License. Other US and International Patents pending or applied for.
®
Page 60
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop on line at www.omega.com
TEMPERATURE

Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies

Wire: Thermocouple, RTD & Thermistor

Calibrators & Ice Point References

Recorders, Controllers & Process Monitors

Infrared Pyrometers
PRESSURE, STRAIN AND FORCE

Transducers & Strain Gauges

Load Cells & Pressure Gauges

Displacement Transducers

Instrumentation & Accessories
FLOW/LEVEL

Rotameters, Gas Mass Flowmeters & Flow Computers

Air Velocity Indicators

Turbine/Paddlewheel Systems

Totalizers & Batch Controllers
pH/CONDUCTIVITY

pH Electrodes, Testers & Accessories

Benchtop/Laboratory Meters

Controllers, Calibrators, Simulators & Pumps

Industrial pH & Conductivity Equipment
DATA ACQUISITION

Data Acquisition & Engineering Software

Communications-Based Acquisition Systems

Plug-in Cards for Apple, IBM & Compatibles

Datalogging Systems

Recorders, Printers & Plotters
HEATERS

Heating Cable

Cartridge & Strip Heaters

Immersion & Band Heaters

Flexible Heaters

Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL

Metering & Control Instrumentation

Refractometers

Pumps & Tubing

Air, Soil & Water Monitors

Industrial Water & Wastewater Treatment

pH, Conductivity & Dissolved Oxygen Instruments
M4000/1105
Loading...