The reproduction, distribution and utilization of this operating manual as well as the communication of
its contents to others without express authorization is prohibited. Offenders will be held liable for the
payment of damages. All rights reserved in the
5.6 Test Task .............................................................................................................. 36
5.6.1 Name ........................................................................................................... 37
5.6.2 Data acquisition ........................................................................................... 37
event of the grant of a patent, utility model or design.
Series TCL-3M165E2 About this operating manual
e-mail: info@omega.com
DANGER! Risk of death due to electric current!
WARNING! / CAUTION! Risk of injury!
cause considerable damage to property.
CAUTION! High temperature!
considerable damage to property.
CAUTION! Material damage!
damage.
NOTICE!
NO DOMESTIC WASTE!
Check the specified points or notices.
0 About this operating manual
• The operating manual is aimed at specialists and semi-skilled personnel.
• Before each step, read through the relevant advice carefully and keep to the specified
order.
• Thoroughly read and understand the information in the section “Safety Instructions”.
• For calibrators with a cooling function, the term "Cooling" is also used for temperatures
below room temperature, in the meaning of "Heating".
If you have any problems or questions, please contact your supplier or contact us directly
at:
Omega.com
Tel: (203) 359-1660
Hazard signs and other symbols used:
This sign indicates dangers which could lead to serious health defects or to death.
This sign indicates dangers that cause personal injuries that can lead to health defects or
This sign indicates dangers resulting from high temperature that can lead to health defects or
This sign indicates actions which could lead to possible damage to material or environmental
ADHERE TO OPERATING MANUAL!
The device must not be disposed of
together with domestic waste.
This symbol indicates important notices,
tips or information.
Pay attention to and comply with
information that is marked with this symbol.
Follow the specified instructions and steps.
Adhere to the given order.
Technical changes reserved - 5 -
Reference to another section, document or
source.
•Item.
Device description Series TCL-3M165E2
Type plate example
1 Device description
The TCL-3M165E2 series of calibrators test and calibrate different temperature measuring
instruments and temperature sensors, and measure temperatures.
The calibrator can be very easily operated using the touch-sensitive screen or touch screen.
Pre-defined or self-defined testing tasks, by which the testing process is governed, are
employed for this purpose.
The portable instruments are of compact and robust construction and thus allow use directly
on-site or in a laboratory. The calibrators meet the requirements of EN 61326-1, class A
(industrial sector).
This TCL-3M165E2 series is used for service purposes and for different industrial and
laboratory tasks.
Thus, for example, thermometers, temperature switches/thermostats, resistance
thermometers and thermo-elements can be directly connected and checked.
Versions:
The TCL-3M165E2 is available as a micro bath, dry block, infrared, surface or multi-function
calibrator.
It can be supplied both with a heating function as well
as a cooling and heating function.
The series and the various types are different in their
construction (tank or metal block, the temperature
range, the accuracy, the adapter sleeves and
calibration inserts, the fittings and the available
accessories.
Further information on the series, the models of the
individual types that can be supplied, and their fittings
can be found in our catalogues.
Type plate:
You find the type plate on the rear of the device.
It includes the type designation, the serial number and the key electric
specifications ( example).
- 6 -
Series TCL-3M165E2 Device description
TCL-3M165E2
Transport case.
1.1 Delivery, unpacking and accessories
All units have been carefully checked for their operational reliability before shipment.
Immediately after receipt, please check the outer packaging for damages or any signs of
improper handling.
Report any possible damages to the forwarder and your responsible sales
representative. In such a case, state a description of the defect, the type and the serial
number of the device.
Report any in-transit damage immediately. Damage reported at a later date shall not be
recognized.
Unpacking:
Carefully unpack the unit to prevent any damage.
Check the completeness of the delivery based on the delivery note.
Save the packaging!
Temperature calibrators are delivered in special protective packaging.
Save the packaging for returning the instrument safely to the manufacturer for
recalibration or repair.
Scope of delivery and accessories:
Test certificate.
Power cable.
Sleeve exchange tool.
PC- and network cable.
Operating manual.
Protective packaging / transport
protection.
Sensor cage.
Drain syringe.
Transport cover.
Magnetic stirrer with magnet lifter.
Work cover with 5 silicone plugs.
IMPORTANT!
Use the type plate to check if the delivered unit corresponds to your order.
In particular, for devices with electrical components, check to see if the correct power
supply voltage is specified.
Scope of delivery Accessories (optional)
according to the order data.
Adapter sleeve(s).
Infrared insert.
Surface insert and exchange tool.
Tub insert.
External reference sensor TF 255-3-300.
PC software.
Network switch, barcode reader, WLAN
router.
Works certificate.
Technical changes reserved - 7 -
Device description Series TCL-3M165E2
1.2 Intended use
The TCL-3M165E2 series of calibrators may only be used for testing and calibration of
suitable temperature measuring instruments, temperature sensors and for measuring
temperatures.
The calibrators may not be used for warming up or heating other parts or gases.
The calibrators have been designed for indoor use only.
The micro baths may only be used with suitable media. Permitted liquids are silicone oils,
mineral oils and water ( § 2.1 "Safety instructions for the application of calibration liquids").
Hazardous media (flammable or explosive liquids or gases) may not be used.
The operational safety of the device supplied is only guaranteed by intended use. The
specified limits ( § 10 "Technical data") may under no circumstances be exceeded.
CAUTION! Risk of burns!
The calibrator can become very hot when in operation. Touching hot parts can result in
serious injuries.
Never touch the metal block, the tank, the adapter sleeve or the test specimen at
temperatures above 35 °C or below 10 °C (95 °F or below 50 °F).
Allow the calibrator to cool before you remove the test specimen, clean the tank,
change the adapter sleeve or the calibration insert or switch off the calibrator.
CAUTION! Material damage!
The opening in the metal block of the calibrator is only intended to be used with adapter
sleeves or calibration inserts.
Using heat transfer media (oil, thermal paste or other media) can lead to incorrect
measurements and damage to the calibrator.
Never fill the metal block opening with a heat transfer medium.
Only micro baths are suitable for use with heat transfer medium.
It is your responsibility to select the instrument which is suitable for your specific application,
to connect it correctly, to carry out tests and to maintain all the components.
1.3 Exclusion of liability
Omega accepts no liability for any damage or malfunctions resulting from incorrect
installation, in-appropriate use of the device or failure to follow the instructions in this
operating manual.
- 8 -
Series TCL-3M165E2 Safety Instructions
2 Safety Instructions
Before you install the TCL-3M165E2, read through this operating manual carefully. If the
instructions contained within are not followed, in particular the safety guidelines, this
could result in danger for people, the environment, and the device and the system it is
connected to.
The TCL-3M165E2 corresponds to the state-of-the-art technology. This concerns the
accuracy, the operating mode and the safe operation of the device.
In order to guarantee that the device operates safely, the operator must act competently and
be conscious of safety issues.
OMEGA ENGINEERING provides support for the use of its products either personally or via
relevant literature. The customer verifies that our product is fit for purpose based on our
technical information. The customer performs customer-specific and application-specific tests
to ensure that the product is suitable for the intended use. With this verification all hazards and
risks are transferred to our customers; our warranty is not valid.
Qualified personnel:
The personnel in charge of the installation, operation and maintenance of the TCL-
3M165E2 must hold a relevant qualification. This can be based on training or relevant
instructions.
Personnel must be aware of this operating manual and have access to it at all times.
The electrical connection should only be carried out by a fully qualified electrician.
General safety instructions:
In all work, the existing National regulations for accident prevention and safety in the
workplace must be complied with. Any internal regulations of the operator must also be
complied with, even if these are not mentioned in this manual.
Ensure that the complete operating instructions are always available in excellent condition
at the calibrator installation site.
Degree of protection according to EN 60529:
Ensure that the ambient conditions at the site of use does not exceed the requirements
for the stated protection rating ( § 10.1 "Characteristics TCL-3M165E2").
Structural safety in accordance with EN 61010-1:
The calibrator must be installed in such a way that the requirements for structural safety
are met.
Only use the TCL-3M165E2 if it is in perfect condition. Damaged or faulty devices must
be checked without delay and, if necessary, replaced.
If problems cannot be cleared, immediately shut down the calibrator and ensure that it
cannot be started up accidentally.
Never leave the calibrator unattended when it is in operation or in the cooling phase. Do not remove or destroy type plates or other markings on the device, or the warranty is
rendered null and void.
Technical changes reserved - 9 -
Safety Instructions Series TCL-3M165E2
Special safety instructions:
Thermal fuse!
The calibrator is equipped with a temperature fuse that works independently. If there is an
over-temperature in the inside of the housing, the power supply to the heating system is
cut off. The calibrator cannot then be started any more.
After it has cooled, send the calibrator for inspection to OMEGA ENGINEERING.
Risk of injury from hazardous gases!
When liquids are heated, the evaporation can result in dangerous gases being formed.
The calibrator may not be used in an explosion-endangered atmosphere (ignitable or
explosive atmosphere).
Remove all the easily flammable media from the vicinity of the calibrator.
Ensure that the calibrator cannot come in contact with easily flammable or explosive
media.
Operate the calibrator only in the temperature range permissible for the test sample. Ensure that the test sample is securely fixed in the calibrator.
Use only suitable adapter sleeves or calibration inserts.
When doing so, also ensure that the structural safety of the calibrator is retained.
Expert mode!
When selecting the function, test sample and testing tasks, you can switch into the expert
mode for administration and configuration. The settings that are made in this mode
require detailed knowledge of the calibration and method of working of calibrators.
If the settings are incorrect, the calibrator can get damaged!
IMPORTANT NOTICE!
The transport cover is equipped with a safety valve, which is activated once the pressure
reaches ~1.5 bar (21.75 psi). This can result in hot steam being released.
Always unscrew the transport cover before putting the micro bath into service, in
order to avoid excessive pressure.
Wait until the micro calibration bath has cooled down before screwing on the transport
cover.
Further warnings that are specifically relevant to individual operating procedures or activities
can be found at the beginning of the relevant sections of this operating manual.
- 10 -
Series TCL-3M165E2 Safety Instructions
Calibration liquid
Calibration Range
Flashpoint
Distilled water
2…95 °C (36…203 °F)
none
Silicone oil from XIAMETER®:
PMX-200 SILICONE FLUID 10 CS
-35 °C (-31 °F)
155 °C (311 °F)
165 °C (329 °F)
2.1 Safety instructions for the application of calibration liquids
Before using calibration liquids, read the entire safety data sheet thoroughly. Pay
particular attention to the information on the physical and chemical properties.
Only use calibration liquids that are suitable for the required temperature range and which
are not flammable.
Always wear safety goggles for the eyes when handling calibration liquids.
We recommend the following calibration liquids for the various temperature ranges:
Water:
Only use distilled water, otherwise excessive limescale and soiling will build up in the
calibrator tank.
Silicone oil:
Use only the silicone oil recommended here. Always read the safety data sheet supplied with the silicone oil before using it. Always ensure adequate ventilation when working with silicone oil, since hazardous
substances can be released.
Prevent silicone oil from coming into contact with your eyes. Since silicone oil is hygroscopic, always use the transport cover to close the calibration
bath after use.
Mineral oil:
OMEGA ENGINEERING supplies the calibrators only with silicone oil. Using mineral oil is possible, but must be done on your own responsibility. The danger
and the risk must be borne by the customer and our warranty will be rendered null and
void.
Please follow the safety data sheet of the mineral oil used. The safety instructions for silicone oil apply analogously for mineral oils as well. The same
also applies to the corresponding sections for silicone oil in this operating manual.
Technical changes reserved - 11 -
Construction, Function and Test Task Series TCL-3M165E2
3 Construction, Function and Test Task
3.1 Construction
The calibrator consists of a robust, black-and-red painted steel housing and has a carrying
handle at the top .
The front part of the housing holds the electronics of the TCL-3M165E2 for controlling the
reference temperature and for the touch screen .
On the front side are the main switch with fuse and power plug connection , the
connection for the external reference , as well as the ports for the PC, the network and
service .
The grille for the exhaust air is located, in the TCL-3M calibrators, in both the side walls of
the housing.
The rear part of the housing holds a heat-insulated calibration block with heating or cooling
elements and an integrated sensor for the reference temperature. In the case of the dry block
and multifunction calibrator, this is a metal block and in the micro bath calibrator, it is a
tank .
In the base, there are grilles for cooling the calibrator. The built-in fans control the inlet air of
the housing and of the calibration block .
An overview of the most important components of the TCL-3M165E2:
Carrying handle.
Steel housing.
Touch screen for operation and
measurement value display.
Main switch with fuse and power plug
socket.
Tank (Micro bath).
Calibrator ventilation:
Exhaust through lateral and upper
housing grilles.
Metal block (Dry block).
Connection for external reference.
Ports for PC, network and service.
Calibrator, bottom view:
Calibrator ventilation
Inlet air for housing cooling.
Inlet air for tank / metal block cooling.
- 12 -
Series TCL-3M165E2 Construction, Function and Test Task
Function
Type
Measuring insert:
Adapter sleeve
Infrared insert
Surface insert
Calibration liquid
TCL-3M165E2
3.2 Function
IMPORTANT NOTICE!
For your safety, manual operation of the calibrator, i.e. directly starting it up and its
remaining at a certain temperature, is not possible.
The testing procedure is always started with a testing task ( § 3.3).
This ensures that the calibrator is always started with a defined "Behavior at test end"
( p. 25).
The calibrators can be used with dry block, infrared, surface or micro bath function.
The function is determined by the calibrator type and the used measuring insert.
Dry block Infrared Surface Micro bath
Method of working:
Adapter sleeves or calibration inserts that are suitable for the temperature sensors or
temperature measuring instruments to be tested are inserted in the calibrator
( § 4.3 "Preparing of the calibrator").
They give the test specimen a secure fit and ensure optimum heat transfer.
Once all the preparations have been carried out, the calibrator can be switched on ( § 4.4).
The calibrator heats or cools the metal block or tank to the set temperature. As soon as this
temperature has been reached and is stable, the calibrator signals accordingly.
The calibration of the test specimen can now be carried out. Then, the next test point is
approached or the testing procedure is ended.
Functionality:
The measuring inserts ( p. 17) of the calibrator have different characteristic values, which
were determined in our laboratory through calibration. These characteristic values are
defined at the factory and stored in the calibrator as a protected function.
To be able to use a measuring insert, the relevant "Functionality of the calibrator" ( § 5.4)
must be selected.
IMPORTANT! Protected list entries!
The functions, test specimens and test tasks marked with the symbol are factorydefined. They can be neither deleted nor configured.
They serve as templates for your own functions, test specimens and test tasks, which
are marked with the symbol.
Technical changes reserved - 13 -
Construction, Function and Test Task Series TCL-3M165E2
3.3 Test tasks
Test tasks are containers for defined test conditions. They are helpful for recurring testing
processes, for standardizing test sequences and for generating measurement logs.
All the required settings and configurations for the calibration of a test specimen are compiled
in a test task. The parameters of the testing task are saved and linked to the selected
function and test specimen ( § 5.6).
The operation concept of the calibrator is based on pre-installed and self-defined test tasks.
They are a central component of the function and operation of the TCL-3M165E2.
The emphasis of the operating manual is on the use of test tasks during operation of the
calibrator.
The calibrator is shipped with functions, test samples and testing tasks defined in the factory,
in which the basic settings of the calibrator are stored. If desired, customer-specific test tasks
can also be pre-set.
These protected test tasks can be neither deleted nor edited. They serve as templates for
your own self-defined test tasks. They can be copied and can then be configured.
You can define your own test tasks for different test specimens or test sequences. These test
tasks are directly saved in the calibrator and can then be easily activated. This makes quick
access to recurring test tasks possible.
Upon switching on the calibrator, the first test task of the selection list is loaded automatically,
with the relevant parameters.
IMPORTANT! The function and test specimen are independent!
During the configuration of test tasks ( § 5.6) when selecting the function or the test
sample, you can switch to the respective expert mode for administration and
configuration.
Note here that changes to existing functions and test specimens always apply to all
the test tasks linked to them.
- 14 -
Series TCL-3M165E2 Commissioning and operation
4 Commissioning and operation
CAUTION! Risk of injury or material damage!
The calibrator can become very hot during operation. If the calibrator is operated without
supervision, third-party persons in the vicinity could get injured. Moreover, flammable
material could get into the calibrator and cause significant damage to property.
Never leave the calibrator unattended when it is in operation or in the cooling phase.
For safe operation of the calibrators of the TCL-3M165E2 series, a proper commissioning
procedure is necessary.
Commissioning includes the installation, the electrical connections, the preparation for the
test task as well as correct switching on and off of the calibrator.
Further, a visual inspection for damage is required before use.
The required steps are described in the following sections.
IMPORTANT!
Please also follow the instructions for intended use ( § 1.2), the safety instructions
( § 2) and the information on the ambient conditions ( p. 56).
4.1 Operating conditions
Select a safe installation site for commissioning the calibrator.
Installation site and operating position:
Only suitable for indoor use, do not use outdoors.
Operate only in the vertical position on an even surface. The surface must be stable,
clean, and dry.
If the operation positions do not conform to the above, the structural safety and the
specified properties of the calibrator are not guaranteed.
At higher testing temperatures, please use a sufficiently big, fire-resistant supporting
surface.
Sufficient clearance around the calibrator
On the front side > 1 m, behind and to the sides > 0.5 m. Head clearance and sufficient
clear space above the calibrator.
Sufficient ventilation must be ensured.
Do not operate in the vicinity of flammable materials.
Do not install in a cupboard or other similar location.
The ventilation openings must not be blocked or covered.
The calibrator must be installed so that it can be switched off at any time.
IMPORTANT! "KILL" switch!
The plug of the power connecting cable serves as a "KILL" switch.
Ensure that the plug is always easily accessible and easy to reach.
In an emergency, pull the plug, so that the calibrator is isolated from the power.
Technical changes reserved - 15 -
Commissioning and operation Series TCL-3M165E2
4.2 Electrical connection
The electrical connections are made with the supplied power connecting cable.
DANGER! Risk of death due to electric current!
The TCL-3M165E2 is operated at power voltages up to 240 VAC. Contact with the power
voltage can result in serious or fatal injuries.
Switch off the TCL-3M165E2 and remove the power connecting cable before you start
any work on live parts.
Check the following points before you connect the calibrator:
Operate the calibrator only with the approved supply voltage ( § 10.1). Ensure that the
power voltage is the same as that specified on the type plate.
Please follow the local regulations of the energy supplier.
Connect the calibrator only to a properly installed and earthed 3-pole outlet.
Do not use any extension cables or adapter plugs.
IMPORTANT! Power connecting cable!
The power connecting cable may only be replaced by an equivalent cable.
Use only original cables from OMEGA ENGINEERING or approved cables of the
same type with the correct design as replacements ( "Electrical characteristics").
Connect TCL-3M165E2:
Connect the power connecting cable with the connector plug of the TCL-3M165E2.
Insert the plug of the power connecting cable in a suitable outlet.
- 16 -
Series TCL-3M165E2 Commissioning and operation
4.3 Preparing of the calibrator
The preparations for the test task must be carried out with the calibrator switched off and
cooled to ambient temperature.
CAUTION! Risk of burns!
The calibrator can become very hot when in operation. Touching hot parts can result in
serious burn injuries.
Never touch the metal block, the tank, the adapter sleeve or the test specimen at
temperatures above 35 °C or below 10 °C (95 °F or below 50 °F).
Allow the calibrator to cool before you remove the test specimen, clean the tank,
change the adapter sleeve or the calibration insert or switch off the calibrator.
IMPORTANT! Empty the tank and clean it!
The tank must be emptied and cleaned to prevent the adapter sleeve or the calibration
insert from getting stuck.
Empty and clean the tank before you insert another adapter sleeve or calibration
insert.
Measuring insert:
The function of the calibrator is determined by the measuring insert. The required measuring
insert is inserted in the opening of the metal block or the tank.
It is thus easily possible to switch between the dry block, infrared, surface and micro bath
functions.
Each measuring insert has its own characteristic values, which are defined at the factory.
These characteristic values are influenced by different factors, such as: Diameter, type and
material of the measuring insert, number and diameter of the bore holes or the physical
properties of the calibration liquid.
These characteristic values are stored as protected functions for the accompanying
measuring inserts of the calibrator ( § 5.4).
Technical changes reserved - 17 -
Commissioning and operation Series TCL-3M165E2
Type
TCL-3M165E2
Dry block
Adapter sleeves and sleeve exchange tool
4.3.1 Dry block calibrator
Adapter sleeves with single or multiple holes are used for the calibration of straight
temperature sensors.
To achieve the specified accuracy of the calibrators ( § 10.1) the temperature
sensor (test specimen) and the adapter sleeve must be matched to one another:
•The borehole of the adapter sleeve may be a maximum of 0.5 mm (0.12 in)
larger than the diameter of the test specimen.
•The measurement element of the test specimen must be located in the
homogeneous temperature zone of the adapter sleeve.
Inserting:
The appropriate adapter sleeve is inserted
into the metal block with the aid of the
sleeve exchange tool.
External reference sensor (optional):
Align the adapter sleeve so that the hole
for the external reference sensor is located
at 12 o’clock.
Removing and cleaning:
Let the calibrator cool before you remove
the adapter sleeve.
Pull the adapter sleeve out of the metal
block with the help of the sleeve exchange tool.
Clean the adapter sleeve and the metal block.
This prevents the adapter sleeve from getting stuck in the metal block.
- 18 -
Loading...
+ 42 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.