Omega SMCJ-K, SMCJ-J, SMCJ-T User Manual

Super MCJ
SMCJ-J, SMCJ-K, SMCJ-T
INSTRUCTION
SHEET
M2813-0103
Thermocouple-to-Analog Connector
This is a Thermocouple to Analog Connector. It converts a thermocouple input signal (Type J, K, or T) to a linear, compensated, amplified analog output. The unit provides 1mV/°F or 1mV/°C analog output. The analog output can go to a strip chart recorder or a digital voltmeter for temperature monitoring.
The unit is powered by a single AA size lithium battery. In order to install the battery, remove the back cover and place the battery on the PC board. The battery polarity is indicated on the PC board. If the battery is installed reverse, the unit will not power up. The back cover is mounted to the front cover via two mounting holes.
The input connection can be SMP or OST type connections. The output connection is standard banana plug or jack. The plastic case and the rubber gasket Bumper Band
®
provides a rugged and sealed connector design. Figure 1 provides a functional flow chart of how the unit operates. This unit is operated using a single tactile feedback button .
There are 3 LEDs built into the connector. The °C green LED flashes once every 2 seconds when the output provides 1mV/°C. The °F green LED flashes once every 2 seconds when the output provides 1mV/°F. If the input is open the °C and °F green LEDs
both flash alternately once a second. When the battery voltage drops below a
certain level, a red LED starts to flash once a second. This is a low battery indication. The battery must be replaced with a fresh battery immediately.
Field Adjustments
The unit is field adjustable by removing the back cover and accessing the three potentiometers. R17 provides offset adjustment when the unit is in mV/°F mode. R42 provides offset adjustment when the unit is in mV/°C mode. R32 provides span (gain) adjustment either in mV/°F or mV/°C mode.
Press
SPECIFICATIONS
Measuring Temperature Range:
Type J: -100 to +750°C (-148 to +1382°F) Type K: -100 to +1250°C (-148 to +2282°F) Type T: -100 to +350°C (-148 to +662°F)
Accuracy @ 75°F ambient temperature across the range:
Type J: ±1.4°C (± 2.5°F) Type K: ±3°C (± 5°F) Type T: ±2°C (±3°F)
Input Connection: SMP or OST connector Input Impedance: 3M-ohms Output Connection: Standard Banana
Plug or Jack
Thermocouple Break Protection:
Output goes Upscale (std.) °F & °C Green LEDs flash once a second
Output: 1mV/°F or 1mV/°C, °F or °C Green LEDs flash once every 2 seconds
Output Load: 1K-Ohms min. Power: One AA size Lithium Battery, 3.4 to
3.6 VDC 1900 mAH
Battery Life @ 75°F Ambient Temperature:
1440 hours (60 Days), continuous, no load
Low Battery Indication:
Red LED flashes once a second,
Response Time:
(0 to 63% of Final value): 150 msec
Operating Temperature:
0 to 50°C (32 to 122°F) Thermal Zero Drift: 0.05 mV/°C
(0.05 mV/°F) ambient temperature change Operating Relative Humidity: 95% or
less, without condensation
Dimensions:
4"L x 2"W x 1"H (10L x 5W x 2.5H cm) Weight: 2.4 oz (67g)
Press
Press
1mV/°C Mode
1 sec Delay
1mV/°F Mode
Input
Open
Input Open
Output Upscale/Downscale
Output Upscale/Downscale
Figure 1, Functional Flow Chart
°C
°F
°C °F
°C
°F
°C °F
YES
YES
NO
NO
1mV/°C Output
1mV/°F Output
Start Sleep Mode
Press
+
Analog Output, 1 mV per degree
Thermo­couple Input
(SMP or OST Connectors)
®
°C °F
OFF
Press
Super MCJ
K
Output
Low Batt
Thermocouple to Analog connector
Model No. Description
OM-NOMAD-BATT Replacement Battery SMCJ-CAB Cable Wire (Red & Black) 36" Long
Accessories
GENERAL DESCRIPTION
Follow all the safety precautions and
operating instructions outlined in the manual.
The installation category is one (1). There are no user replaceable fuse in
this product.
Maximum input voltage to earth: 30
Vrms (42.4 peak) or 60 Vdc.
Avoid contact with hazardous live parts. The thermocouple probe should not be connected to hazardous live voltages (less than max. input voltage to earth).
The output terminals of this product are
for use with equipment (digital multimeter, chart recorder, etc.) which have no accessible live parts. Such equipment should comply with all the applicable safety requirements.
Do not open batteries, dispose of in fire, heat above 100°C (212°F), expose contents to water, recharge, put in backwards, mix with used or other battery types. It may cause explosion, leakage or personal injury.
Use cloth damped with warm water and soap to clean the outside case and label. Do not use flammable liquid solutions for cleaning. Do not immerse the product in water.
OTHER IMPORTANT CONSIDERATIONS BEFORE OPERATION
CAUTION
Shop online at: omega.com e-mail: info@omega.com For latest product manuals: omegamanual.info
It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, human applications.
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components in which wear is not warranted, include but are not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/ DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
Servicing North America:
U.S.A.: One Omega Drive, Box 4047
ISO 9001 Certified Stamford, CT 06907-0047
Tel: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com
Canada: 976 Bergar
Laval (Quebec) H7L 5A1, Canada Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.ca
For immediate technical or application assistance:
U.S.A. and Canada: Sales Service: 1-800-826-6342/1-800-TC-OMEGA
®
Customer Service: 1-800-622-2378/1-800-622-BEST
®
Engineering Service: 1-800-872-9436/1-800-USA-WHEN
®
Mexico: En Espan˜ol: (001) 203-359-7803 e-mail:espanol@omega.com
FAX: (001) 203-359-7807 info@omega.com.mx
OMEGAnet®Online Service Internet e-mail
omega.com info@omega.com
Servicing Europe:
Czech Republic: Frystatska 184, 733 01 Karviná, Czech Republic
Tel: +420 (0)59 6311899 FAX: +420 (0)59 6311114 Toll Free: 0800-1-66342 e-mail: info@omegashop.cz
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
Tel: +49 (0)7056 9398-0 FAX: +49 (0)7056 9398-29 Toll Free in Germany: 0800 639 7678 e-mail: info@omega.de
United Kingdom: One Omega Drive, River Bend Technology Centre
ISO 9002 Certified Northbank, Irlam, Manchester
M44 5BD United Kingdom Tel: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622 Toll Free in United Kingdom: 0800-488-488 e-mail: sales@omega.co.uk
RETURN REQUESTS / INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR W
ARRANTY RETURNS, please have the following information
available BEFORE contacting OMEGA:
1. Purchase Order number under which the product was PURCHASED,
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the product.
FOR NON-WARRANTY REPAIRS,
consult OMEGA for current repair charges.
Have the following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST of the repair,
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 2006 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic
medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
Loading...