Omega HH802U User Manual

OPERATING INSTRUCTIONS
MODEL HH802U
DUAL INPUT K/J
DIGITAL THERMOMETER
HH802U
MADE IN TAIWAN
INTRODUCTION
This instrument is a 5 digit, compact-sized portable digital thermometer de­signed to use external K type and J type thermocouple sensor. Temperature indication follows Reference Temperature/Voltage Tables (N.I.S.T. Mono­graph 175 Revised to ITS-90) for K type and J type thermocouples. Two K type thermocouple are supplied with the thermometer.
SAFETY INFORMATION
It is recommended that you read the safety and operation instructions before
using the thermometer.
WARNING
To avoid electrical shock, do not use this instrument when working voltages
at the measurement surface over 24V AC or DC.
WARNING
To avoid damage or burns, do not make temperature measurement in mi-
crowave ovens.
CAUTION
Repeated sharp flexing can break the thermocouple leads. To prolong lead
life, avoid sharp bends in the leads, especially near the connector.
Input Connector:
Accepts standard miniature thermocouple connectors (flat blades spaced
7.9mm, center to center).
ENVIRONMENTAL
Ambient Operating Ranges:
0°C to 50°C (32°F to 122°F) <70% R.H.
Storage Temperature:
-20°C to 60°C (-4°F to 140°F) <80% R.H.
GENERAL
Display: 5 digit liquid crystal display (LCD). Overload: “----.-“ or “OL” is display. Battery: 1.5V x 4 PCS (SIZE AAA) UM-4 R03. Battery Life: 200 hours typical with carbon zinc battery. Auto power off: 30 minutes, press power key to resume operation. Dimensions: 160mm(H) x 83mm(W) x 38mm(D) Weight: Approx. 265g including batteries. Supplied Thermocouples (1 per input):
1 meter (40”) type K insulated beaded wire thermocouple. Maximum insula­tion temperature is 482°C (900°F). Thermocouple accuracy is ±1.1°C or 0.4% of reading (whichever is greater) from 0°C to 1250°C.
SPECIFICATIONS
ELECTRICAL
Temperature Scale: Celsius or Fahrenheit user selectable. Measurement Range:
J-TYPE -200°C to 1050°C, (-328°F to 1922°F) K-TYPE -200°C to 1370°C, (-328°F to 2498°F)
Resolution: 0.1°C or 0.2°F Accuracy: Accuracy is specified for operating temperatures over the range of
18°C to 28°C (64°F to 82°F), for 1 year, not including thermocouple error. ±(0.05% rdg + 0.3°C) -50°C to 1370°C ±(0.05% rdg + 0.7°C) -50°C to -200°C
±(0.05% rdg + 0.6°F) -58°F to 2498°F ±(0.05% rdg + 1.4°F) -58°F to -328°F
Temperature Coefficient:
0.1 times the applicable accuracy specification per °C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to122°F).
Input Protection:
24V dc or 24V ac rms maximum input voltage on any combination of input pins.
Maximum Differential Common Mode Voltage (Maximum Voltage be­tween T1 and T2 during measurement): 1volt. Reading Rate: 1 sample/second.
OPERATING INSTRUCTIONS
1. Power Switch
key turns the thermometer on or off. In the SET mode the unit can-
The not be powered off. Exit SET mode to power off.
APO function mode
Press
power key for more than 6 seconds to disable the auto-power
function. The display will show “APO OFF”.
2. °C/°F Selecting the Temperature Scale
Readings are displayed in either degrees Celsius(°C) or degrees Fahren­heit(°F). When the thermometer is turned on, it is set to the temperature scale that was in use when the thermometer was last turned off. To change the temperature scale, press the °C/°F key.
3. Button (only Main display)
Press this key to enter the Data Hold mode, the “HOLD” annunciator is dis­played. When HOLD mode is selected, the thermometer holds the present
readings and stops all further measurements. Press the key again to cancel HOLD mode and resume measurement. In the MIN/MAX recording
mode, press key to stop the recording. Press key again to resume recording. (Previously recorded readings are not erased).
Backlight function mode
Press the button for more than two seconds to turn on the backlight Press the button again for more than two seconds to turn off the backlight. The backlight will switch-off automatically after 30 seconds.
4. T1 T2/T1-T2 Main Display Input Selection
The input selection determines which input is selected for the main display; T1 thermocouple, T2 thermocouple or the difference between the two ther­mocouples (T1-T2), when the thermometer is turned on it is set to T1.
5. K/J T1 Input Type Selection (Main display)
The K/J key selects the T1 thermocouple type, when the main display is showing T1. When the thermometer is turned on, it is set to the type se­lected when the thermometer was last turned off.
6. MIN MAX with Elapsed Time Mode (only Main display)
Press MIN MAX key to enter the MIN MAX Recording mode, (displays the Maximum reading with time, Minimum reading with time and Average reading stored in record mode). In the this mode the automatic power-off
feature is disabled and key and main display T1 T2 T1-T2 key, K/J key are disabled. The beeper emits a tone when a new minimum or maximum value is recorded.
key, °C/°F key, REL key, SET key, Hi/Lo Limits
Push MIN MAX key to cycle through the MAX, MIN and AVG readings. If an overload is recorded, the averaging function is stopped. In this mode, press the HOLD key to stop the recording of readings, all values are held, press again to restart recording. To prevent accidental loss of MIN, MAX and AVG data, this mode can only be cancelled by pressing and holding the MIN MAX key for 2 seconds. All recorded readings are erased.
7. REL Relative mode (only Main display)
Press the REL key to enter the Relative mode, zero the display, and store the displayed Reading as a reference value. REL is shown on the display. Press REL key again to exit the relative mode. The relative reference value can also be entered by the user. (See “SET mode” later in this manual.) When the desired Relative value has been entered, press REL key to enter the Relative mode and then press SET key use the entered Relative value as a reference value. Press REL key again to exit the relative mode. In the Relative mode, the value (can not >±3000.0 counts) shown on the LCD is always the difference between the stored reference and the present reading.
8. Sec. Minu. Selecting the Elapsed Time scale
Press this key to display the elapsed time on the third display in either hours and minutes or minutes and seconds. When the thermometer is turned on, it is set to seconds. To change the time scale, press sec. Minus. key. Maxi­mum elapsed time reading is 100 hours. If 100 hours is exceeded, the elapsed time resets to zero.
Press the SET key to enter Relative value SET mode (Press ENTER key
to skip setting relative value). = = = =.= is displayed on the main display. The Relative value is entered via overlay numbers, press overlay ENTER key to store the relative value, and advance to elapsed time set mode.
Elapsed Time set mode, (Press ENTER key to skip Elapsed Time set
mode) =.= = = : = = is displayed in second and third display. Time (hours, minutes, seconds) value is entered via overlay numbers, press overlay
ENTER key to store Time value. Elapsed Time starts from set time value.
Hi Limit value set mode, is displayed (Press ENTER key to skip
Hi Limit value set mode), = = = =.= is displayed in main display, Hi Limit value is entered via overlay numbers, then press ENTER key to store the Hi Limit value. = = = =.= is displayed in main display, Lo Limit value is entered via overlay numbers, then press overlay ENTER key to store the Lo Limit value and exit SET mode.
When the thermometer is turned on it uses the Relative value and Hi/Lo
Limits values that were entered when thermometer was last in use.
10.T1/T2 T1-T2 second display Input Selection
The input selection indicates which input is selected for second display; T1 thermocouple, T2 thermocouple or the difference between the two thermo­couples (T1-T2). When the thermometer is turned on, it is set to T2 input. To select a different type of thermocouple use the K/J button.
11.K/J T2 Thermocouple type select (second display)
The K/J button for T2 input selects K-type or J-type thermocouple as input, when the second display has T2 shown. When the thermometer is turned on, it is set to the type selected when the thermometer was last turned off.
12.Hi/Lo Limits mode (only Main display)
Press Hi/Lo Limits key to enter the Hi/Lo Limits comparative mode. is displayed. When the input temperature value exceeds the Hi or Lo Limit value, the beeper emits a continuous or pulse tone. Press Hi/Lo Limits key again to exit the Hi/Lo Limits mode.
It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification. The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, human applications.
OPERATOR MAINTENANCE
To avoid possible electrical shock, disconnect the thermocouple connectors
from the thermometer before removing the cover.
Battery Replacement
1. Power is supplied by 4pcs 1.5V (SIZE AAA) R03P.
2. The “ ” appears on the LCD display when replacement is needed. To replace battery remove screw from back of meter and lift off the battery cover.
3. Remove the battery from battery contacts and replace.
4. When not in use for long periods the batteries should be removed.
5. Do not ations with high temperatures, or high humidity.
Cleaning
Periodically wipe the case with a damp cloth and detergent, do not use abra sives or solvents.
OMEGAnet®On-Line Service Internet e-mail
omega.com info@omega.com
U.S.A.: ISO 9001 Certified
One Omega Drive, Box 4047 Stamford, CT 06907-0047 Tel: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com
U.S.A. and Canada:
Sales Service: 1-800-826-6342/1-800-TC-OMEGA Customer Service: 1-800-622-2378/1-800-622-BEST Engineering Service: 1-800-872-9436/1-800-USA-WHEN
Czech Republic:
Frystatska 184, 733 01 Karvina´, Czech Republic Tel: +420 (0)59 6311899 FAX: +420 (0)59 6311114 Toll Free: 0800-1-66342 e-mail: info@omegashop.cz
Germany/Austria:
Daimlerstrasse 26, D-75392 Deckenpfronn, Germany Tel: +49 (0)7056 9398-0 FAX: +49 (0)7056 9398-29 Toll Free in Germany: 0800 639 7678 e-mail: info@omega.de
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY the following information available BEFORE contacting OMEGA:
1. Purchase Order number under which
2. Model and serial number of the product
3. Repair instructions and/or specific
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering. OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 2007 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine­readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
For immediate technical or application assistance:
RETURN REQUESTS/INQUIRIES
RETURNS, please have
the product was PURCHASED,
under warranty, and
problems relative to the product.
WARNING
Servicing North America:
Servicing Europe:
Canada:
976 Bergar Laval (Quebec) H7L 5A1, Canada Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.ca
®
®
®
United Kingdom
ISO 9002 Certified
One Omega Drive River Bend Technology Centre Northbank, Irlam Manchester M44 5BD
United Kingdom Tel: +44 (0)161 777 6611
FAX: +44 (0)161 777 6622 Toll Free in United Kingdom: 0800-488-488 e-mail: sales@omega.co.uk
FOR NON-WARRANTY REPAIRS, OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST of the repair,
2. Model and serial number of the
product, and
3. Repair instructions and/or specific problems
relative to the product.
-
Mexico:
En Espan~ol: (001) 203-359-7803 FAX: (001) 203-359-7807 e-mail: espanol@omega.com
info@omega.com.mx
consult
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components in which wear is not warranted, include but are not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY / DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop online at omega.com
WARRANTY/DISCLAIMER
TEMPERATURE
䡺⻬
Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies
䡺⻬
Wire: Thermocouple, RTD & Thermistor
䡺⻬
Calibrators & Ice Point References
䡺⻬
Recorders, Controllers & Process Monitors
䡺⻬
Infrared Pyrometers
PRESSURE, STRAIN AND FORCE
䡺⻬
Transducers & Strain Gages
䡺⻬
Load Cells & Pressure Gages
䡺⻬
Displacement Transducers
䡺⻬
Instrumentation & Accessories
FLOW/LEVEL
䡺⻬
Rotameters, Gas Mass Flowmeters & Flow Computers
䡺⻬
Air Velocity Indicators
䡺⻬
Turbine/Paddlewheel Systems
䡺⻬
Totalizers & Batch Controllers
pH/CONDUCTIVITY
䡺⻬
pH Electrodes, Testers & Accessories
䡺⻬
Benchtop/Laboratory Meters
䡺⻬
Controllers, Calibrators, Simulators & Pumps
䡺⻬
Industrial pH & Conductivity Equipment
DATA ACQUISITION
䡺⻬
Data Acquisition & Engineering Software
䡺⻬
Communications-Based Acquisition Systems
䡺⻬
Plug-in Cards for Apple, IBM & Compatibles
䡺⻬
Datalogging Systems
䡺⻬
Recorders, Printers & Plotters
HEATERS
䡺⻬
Heating Cable
䡺⻬
Cartridge & Strip Heaters
䡺⻬
Immersion & Band Heaters
䡺⻬
Flexible Heaters
䡺⻬
Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL
䡺⻬
Metering & Control Instrumentation
䡺⻬
Refractometers
䡺⻬
Pumps & Tubing
䡺⻬
Air, Soil & Water Monitors
䡺⻬
Industrial Water & Wastewater Treatment
䡺⻬
pH, Conductivity & Dissolved Oxygen Instruments
V1. 032707
M4498/0607
Loading...