and Humidity
Dimensions:About 250 (L) × 100 (W) × 45(H) mm
Weight:About 500g with battery
Printer:Interval print :
00 hour : 00 mi nute : 03 second 〜23 hour :
59 minute : 59 sec
Trigger print : Print three data
(date / time / Reading)
Accessories:Thermo carrying case, Test leads with alligator
clip, 2 rolls of thermo paper, 110V/220V AC to 9V
DC power adapter/power cord, RS-232 cable, P.C.
software disk, 6pcs 1.5V AA batteries, TL-A37
analog output cable, RS-232 convert header,
Operation manual.
4
2-2 Electrical Specification
Analog test range :DCV/ACV:2V,200mV
DCA: 20mA
Accuracy:(at 23℃±5℃,below 80%RH)
DCè(±1%rdg+3dgts)
ACè(±1%rdg+5dgts) at (50~450Hz)
Resolution:200mV/2V range is 0.1mV/1mV
20mA range is 0.01mA
A / D Sample Rate :2.5 times/sec
Power Source:6 pcs 1.5V,# 3 battery or 9V, DC adaptor
(inside negative, outside positive)
Battery Life:Approx. 100 hours (without printing)
Printer:Thermal printer (EPL1902 of Panasonic)
MEMORY:60000 records (up to 130000 records)
5
3. PARTS & CONTROLS
3-1 The Name of Each Parts and Positions:
200mV2V 200 mV 2V20mA
C¢XFcLuxdB
F
¢X
HiHiHiLoLoLo
DATE TIME START STOP
13456 789
20
,
3
SETTING
FEED
SETUP
LINEMIN
CLOCKALARM
PRINT
RECORD
,
1
TRIGGER
,
3 CONSECUTIVE
COM
mA
20mA
MAX
FUSED
600V
(Fig-2)
6
¢X
¢X
60V
24V
MAX
RH A A
UNITRANGE
MAX
V
TG
INTV
1.
V
A
2.
3.
4.
200mV
C / F
HiHi
DATA
0
2V
Lux/Fc
Hi
TIME
1/2/3
200mV
dB
LoLo
START
4/5/6
2
%RH
Lo
STOP
Toggle button for ( 200mV, ℃/℉, HiHi, DATA)
and digit “ o ”.
Toggle button for ( 2V, Lux/ Fc, Hi, TIME) and
digit “ 1/2/3 ”.
Toggle button for ( 200mV, dB, LoLo, START)
and digit “ 4/5/6 ”.
Toggle button for ( 2V, %RH, Lo, STOP) and
digit “ 7/8/9 ”.
5.
6.
7/8/9
20m
A/
G/T
INTV
3'SETTING
SETUP
Toggle button for ( 20mA, A / A,
G / T, INTV ) button and “ ” right shift
function.
Press and hold this button for 3 sec will into
setting mode, press again will end setting
mode.
7
7.
RANGE
FEED
LINE
Feed paper in normal mode, range function in
setting mode.
8.
9.
10.
11.
“ COM ” Measuring
12
13
14
“ V ” Measuring
“ mA ” Measuring
UNIT
MAX
MIN
CLOCK
PRINT
ALARM
=
RECORD
Socket
Socket
Socket
MAX/MIN function in normal mode “ UNIT ”
function in setting mode.
In normal mode, press 1 sec is trigger print
function, press and hold for 3 sec is start/stop
consecutive print function, but in setting mode
will be “ clock ” function.
In normal mode, press 1 sec is trigger record (
MAX 255 times), In setting mode is “ alarm ”
function, press and hold over 3 sec is start/stop
consecutive record function.
Power button.
To connect negative lead of voltage, current.
19 FEED ButtonFeed paper into printer
20 DC 9V SocketTo connect adaptor.
8
3-2 Power Status :
As battery power is not sufficient, the , will be shown on LCD
replace with six new 1.5V batteries is required.
3-3 Quick Manual for Easy Set-up
STEP1STEP2STEP3STEP4STEP5
Push this
button 3 second.
SETUP
Select one.
RANGE
UNIT
CLOCK
ALARM
Choose one from specified
function Bar.
2V
LUX / FC
TIME START
Hi
200mV
dB%RH
STOP
LoLo
Lo
200mV2V
F
c
DATE
HiHi
20mA
A /
INTV
G / T
A
Reconfirm
input setting.
RANGE
UNIT
CLOCK
ALARM
a. You could only set up one of the functions at each time.
b. To change position of point decimal, follow step1 & step2, then press button.
c. To input digit, follow step1 & step3 to set the number.
Press to
return
normal.
SETUP
3-4 ACV、DCV and DCA Test Setting :
1. ACV 2V Test setting sequence:Press and hold
à press
3'SETTING
RANGE
à press
, to end setting mode.
2. ACV 200mV Test setting sequence:Press and hold
sec à press
3'SETTING
press
RANGE
à press
, to end setting mode.
3. DCV 2V Test setting sequence:Press and hold
à press
3'SETTING
RANGE
à press
, to end setting mode.
2V
à press
200mV
2V
à press
9
3'SETTING
RANGE
à press
3'SETTING
RANGE
3'SETTING
for 3 sec
à press
for 3
RANGE
for 3 sec
à press
à
4. DCV 200mV Test setting sequence:Press and hold
sec à press
3'SETTING
press
RANGE
à press
, to end setting mode.
200mV
à press
3'SETTING
RANGE
for 3
à
5. DCA 20mA Test setting sequence:Press and hold
sec à press
3'SETTING
press
RANGE
à press
, to end setting mode.
20mA
3-5 UNIT Setting Step:
1. ℃/℉ Setting step:press and hold
UNIT
press
à press
3'SETTING
, to end setting mode.
C/ F
change℃/℉ à press
2. Lux/Fc Setting step:press and hold
UNIT
press
à press
3'SETTING
Lux/Fc
change Luc/Fc à press
, to end setting mode.
3. dB Setting step:press and hold
UNIT
à press
dB
à press
to end setting mode.
4. %RH Setting step:press and hold
UNIT
à press
%RH
à press
to end setting mode.
5. / A Setting step:press and hold
UNIT
à press change A/A à press
3'SETTING
, to end setting mode.
3'SETTING
3'SETTING
3'SETTING
UNIT
3'SETTING
UNIT
3'SETTING
à press
3'SETTING
for 3
RANGE
for 3 sec à press
UNIT
for 3 sec à press
UNIT
for 3 sec à press
à press
3'SETTING
for 3 sec à press
à press
3'SETTING
for 3 sec à press
UNIT
à press
à
à
à
,
,
3-6 ALARM Setting Step:
1. HIHI ALARM Setting step:press and hold
press
ALARM
à press
shift button setting “ALARM" à press
to end setting mode.
2. HI ALARM Setting step:press and hold
mode à press
ALARM
HiHi
à press
10
3'SETTING
for 3 sec à
à assort 0~9 number è right
ALARM
3'SETTING
Hi
à press
fo r 3 sec into setting
à assort 0~9 number
3'SETTING
,
è right shift button setting “ALARM" à press
3'SETTING
, to end setting mode.
ALARM
à press
3. LOLO ALARM Setting step:press and hold
setting mode à press
ALARM
à press
number è right shift button setting “ALARM" à press
3'SETTING
press
, to end setting mode.
4. LO ALARM Setting step:press and hold
setting mode à press
ALARM
à press
number è right shift button setting “ALARM" à press
3'SETTING
press
, to end setting mode.
3'SETTING
3'SETTING
LoLo
Lo
for 3 sec into
à assort 0~9
ALARM
for 3 sec into
à assort 0~9
ALARM
3-7 Real Time Clock , Time/ Date Setting Step:
1. Real time clock , date setting step:press and hold
sec into setting mode à press
CLOCK
à press
assort 0~9number è right shift button setting date à press
CLOCK
à press
3'SETTING
, to end setting mode.
2. Real time clock, time setting step:press and hold
sec into setting mode à press
CLOCK
à press
assort 0~9number è right shift button setting time à press
CLOCK
à press
3'SETTING
, to end setting mode.
3'SETTING
3'SETTING
for 3
DATE
for 3
TIME
3-8 Record/Print Start Time, Stop Time, Interval Time Setting Step:
à
à
à
à
1.
Start time setting step
setting mode
à press
:press and hold
CLOCK
à press
3'SETTING
START
number è right shift button setting start time à press
3'SETTING
press
2.Stop
time setting step
setting mode
0~9number è right shift button
3'SETTING
press
3. Interval time setting step:press and hold
setting mode à press
number è right shift button setting in terval time à press
to end setting mode.
:press and hold
à press
to end setting mode.
CLOCK
setting stop time à press
CLOCK
à press
11
3'SETTING
à press
3'SETTING
INTV
for 3 sec into
à assort
CLOCK
for 3 sec into
STOP
à
CLOCK
for 3 sec into
à assort 0~9
CLOCK
0~9
à
assort
à
à
3'SETTING
press
to end setting mode.
3-9 Print Mode Setting Step:
Printer work mode setting sequence:press and hold
sec into setting mode à press
G/T MODE à press
3'SETTING
ALARM
, to end setting mode.
à press
3-10 Clear Record Step:
3'SETTING
G/T
for 3
change
Press
=
RECORD
+
Power ON in the same time
The LCD should display “dEL”
3-11 Disable Start、Stop、Intv and Setting Step:
1.Disable START、STOP、INTV sequence:press and hold
3'SETTING
START
for 3 sec into
3'SETTING
start, stop, intv à press
for 3 sec into setting mode à press
3'SETTING
to end setting mode.
2.Disable ALARM sequence:press and hold
setting mode à press
STOP
disable alarm à press
to end setting mode.
3-12 Decimal Point Setting Step:
Shift point setting sequence:press and hold
enter setting mode à press
position à press
3'SETTING
to end setting mode.
INTV
3'SETTING
to change decimal point
3-13 Consecutive Record and Print Setting Step:
Setting sequence:press and hold
mode à press
TIME
setting / = à press
end setting mode.
※ If already in setting MODE, press
3'SETTING
TIME
for 3 sec to enter setting
3'SETTING
button only.
3-14 Reset System Setting Step:
disable
3'SETTING
for 3 sec to
, to
Press
3'SETTING
+ Power ON
12
3-15 Record / Print Start and Step Operation Method:
way)
b). Into setting mode, setting Intv.
c). Back to general mode.
d). Press
=
RECORD
PRINT
button 3 second, the LCD will be show
= or symbol, into consecutive Record / Print of hand
setting mode.
PRINT
or
(2). End Record/Print :
1. Automatic mode : The function will be stopped automatically while
the STOP time is up.
2. Hand setting mode :To stop consecutive Record/Print by user press
=
and hold
PRINT
RECORD
symbol will be disappear.
or
PRINT
button for 3 seconds. The LCD
3-16 Memory Full Warning :
The = symbol will be flash.
3-17 Alarm Print in Graphic Mode :
When printing in G mode, if you have setting any alarm points
(HiHi, Hi, LoLo, Lo) when the read value over the set points, the
printer the will print alarm value on the paper.
13
=
RECORD
or
3-18 Scale Change in Graphic Mode
HiHi
ALARMHiALARMLoALARM
üNo change
üNo change
üNo change
üüChange scale by HiHi Hi ALARM
üüChange scale by HiHi Lo ALARM
üüChange scale by HiHi LoLo ALARM
üüChange scale by Hi Lo ALARM
üüChange scale by Hi LoLo ALARM
üüChange scale by Lo LoLo ALARM
üüüChange scale by HiHi Lo ALARM
üüüChange scale by HiHi LoLo ALARM
üüüChange scale by HiHi LoLo ALARM
üüüChange scale byHi LoLo ALARM
LoLo
ALARM
üNo change
Scale change by alarm setting
üüüüChange scale by HiHi Lo ALARM
No setting ALARM, no change
ü : had setting
3-19 How to ConnectsTL-(A37) Analog Output Cable
14
200mV
2V
200mV
20mA
RH
Lux
F
Fc2VdB
C
¢X
¢X
HiHi
DATE
TIME
0
1
,
SETTING
3
SETUP
CLOCK
PRINT
,
1
TRIGGER
,
3
CONSECUTIVE
mA
A
A
¢X
¢X
Lo
LoLo
Hi
T
G
STOP
START
INTV
3
87
9
65
2
4
RANGE
UNIT
FEED
MAX
LINE
MIN
ALARM
RECORD
COM
V
20mA
60V
MAX
24V
FUSED
MAX
600V
ModelConnecting way (DC Input) Connecting way (AC Input)Remark
TES-1300
TES-1302
TES-1303
TES-1332
TES-1334
TES-1350
TES-1351
TES-1352
TES-1353
TES-1360
Red lead connect to (V),
Black lead connect to (COM)
Red lead connect to (V),
Black lead connect to (COM)
Red lead connect to (V),
Black lead connect to (COM)
Red lead connect to (V),
Black lead connect to (COM)
Red lead connect to (V),
Black lead connect to (COM)
Red lead connect to (V),
Black lead connect to (COM)
Red lead connect to (V),
Black lead connect to (COM)
White lead connect to (V),
Black lead connect to (COM)
White lead connect to (V),
Black lead connect to (COM)
Red lead connect to (V),
Black lead connect to (COM)
No (AC input)
No (AC input)
No (AC input)
No (AC input)
No (AC input)
White lead connect to (V),
Black lead connect to (COM)
White lead connect to (V),
Black lead connect to (COM)
White lead connect to (V),
Black lead connect to (COM)
White lead connect to (V),
Black lead connect to (COM)
White lead connect to (V),
Black lead connect to (COM)
AC,DC output
terminal same
AC,DC output
terminal same
AC,DC output
terminal difference
AC,DC output
terminal difference
3-20 Setup Process Transducer (4-20mA) Mode :
15
Operation way :
Press
INTV
+ Power ON into Process T ransducer (4 - 20 mA) mode.
First of all, settin g Ref1 “ MAX ” refer ence value (t he “MAX” symbo l will be flash) ,
then press button to inpu t “ MAX ” reference
0.001 ~ 9999, press
0 1/2/3 4/5/6 7/8/9
=
RECORD
button and wait for speaker sound. Then
the process transducer (4 - 20mA) mode setting is success.
Ref1-0
“ LCD ” display value:
16
´
(A/D Reading-4) = “ LCD ” display value
※Available for (4mA ~ 20mA) transducer only.
4. BATTERY & FUSE REPLACEMENT
4-1 Battery Check & Replacement:
1. As battery power is not sufficient, the , will be shown on LCD
replace with six new 1.5V batteries is required.
2.
a). Disconnecting test leads.
b).Turning off multimeter with screw-drive r.
c). Open battery cover.
d). Replace the batteries
e). Cover and secure the battery cover.
4-2 Fuse Replacement:
a). Disconnecting test leads.
b). Turn off multimeter.
c). Open fuse cover with screw-driver.
d). Remove old fuse and replace with new fuse.
e). Cover and secure the fuse cover.
※ Only 5ψ×20mm fast blow type fuse is allowable.
4-3 Note:
To prevent electrical hazard or shock, turn off multimeter and disconnect
test leads before removing any cover.
5. DATALOGGING
16
5-1 Software Installation:
1. PC Hardware requirements:q HDD 20 MB space, 1.44MB floppy drive r, se rial po rts, 4 86 PC or above,
with COM1, COM2 comports.
q EGA or higher display.
q 4M bytes or more memory size.
2.
Hardware installation :
Switch off all power related to your PC.
Connect the female socket of RS-232 cable to PC’s COM1 or COM2.
20mA
200mV
200mV
2V
2V
F¢X
Lux
A
¢X
dB
¢X C
Fc
RH
¢X
LoLo
Hi
GAT
HiHi
Lo
INTV
START
DATE
TIME
STOP
0,12 43
765 98
RANGE
SETTING3
UNIT
FEED
MAX
SETUP
LINE
MIN
ALARM
CLOCK
PRINT
RECORD
TRIGGER,1
,
CONSECUTIVE3
COM
mA
V
20mA
60V
MAX
24V
FUSED
MAX
600V
Power on.
Connect the male socket of RS-232 cable to Graphic Recorder.
3. Software setup:
Before setting up software of Graphic Recorder start up windows
95/98 operating system.
Close all application.
Insert the installation diskette into drive A: or drive B:.
Click “ Start ” from Start menu, m ove mouse po inter to “ Ru n ”, then clic k this butt on.
17
A “ Run ” window appears, then type A:\SETUP or B:\SETUP and press “ OK ”.
Setup program will run automatically.
18
Click “NEXT”
ê
ê
a). Click “ NEXT ” or
b). If willing to set up on a different folder, click “ Browse ”.
19
Click “ NEXT ”
20
5-2 Default Settings :
1. Start up
Run the Graphic Recorder software :
Click “Start” from the “Start” menu, move mouse pointer to “Programs”
and then move to “Omega” (Default Settings). Move mouse pointer to
“GRAPHER” and then click this button.
Move mouse pointer to available comport COM1, COM2 then click.
2. RS232 settings
Baud rate : 9600 Parity : None
Data bits : 8 Stop bit : 1
21
3. Time Setup : Move mouse pointer to “System Time Set” then click this button
to input time of PC to the
Graphic Recorder.
When “WAIT” (center) disappeared, setup of time is completed.
Also could be set up by “Setup” mode. Refer to P.31 for more details.
4. Memory size : If you have bought more memories :
Move mouse pointer to “Option” then click. Move mouse pointer to
“Memory” and then click this button.
22
In the window “MEMORY”, move mouse pointer to “UPDATE” and then
click this button. Input the No.s of IC, and then press OK.
5. ID CODE Setup :
Move mouse pointer to “Option” then click. Move mouse pointer to
“Memory” and then click this button.
23
In the window of “MEMORY” , double click or drag the mouse to
highlight the values of “ID Code”.
Input the expected values of “ID Code” and then click the button
“ID Code”.
While “PLEASE WAIT” disappeared, setup of “ID Code” is completed.
24
6. Setup of sampling rate of the meter :
Move mouse pointer to “Option” then click. Move mouse to “Setup” and
then click this button.
25
In the window “SETUP”, select the option button named “CLOCK” under
the title “MODE”.
Drag mouse to highlight the values of “INTERVAL”.
26
Input the expected values with “HOUR” , “MINUTE” , “SECOND” of
“INTERVAL” then click “SETUP”.
While the message “Start/Stop REC/PRT with timing ? ” appears, press
“Cancel” at this moment then escape Setup mode.
7. Setup of Input Voltage \ Current
27
Move mouse pointer to “Option” then click. Move mouse pointer to
“Setup” and click this button.
In the window “SETUP”, select the option button named “RANGE”
under the title “MODE”. Under the “RANGE” title, select the expected
input Voltage \ Current then escape “SETUP” mode.
8. Setup of Unit
Move mouse pointer to “Option” then click. Move mouse pointer to
“Setup” and click this button.
28
In the window “SETUP”, select the option button named “UNIT” under
the title “MODE”. Under the “UNIT” title, select the expected input unit
then escape “SETUP” mode.
9. Setup of Alarm :
Move mouse pointer to “Option” then click. Move mouse pointer to
“Setup” and click this button.
29
In the window “SETUP”, select the option button named “ALARM” under
the title “MODE”. Under the “ALARM” title, drag mouse to highlight the
values of HiHi, Hi, Lo, LoLo. Input the expected values then move
mouse pointer to “UPDATE” and click. Escape the “SETUP” window to
complete.
30
10. Setup of Record / Print / Time : (Open up “PANEL” window is required)
Move mouse pointer to “Option” then click. Move mouse pointer to
“Setup” and click this button.
In the window “SETUP”, select the option button named “CLOCK” under
31
the title “MODE”. Under the “CLOCK” title.
A. Setup time of the meter manually :
a). Move mouse pointer to “TIME SET” then click. While the warning
message appears, click “OK”.
b). Drag mouse to highlight the values of YEAR / MONTH / DATE /
HOUR / MINUTE / SECOND under the red title “DATE” or “TIME”.
32
c). Input the expected values and then click the right bottom button
“SETUP”.
d). While the “START/STOP REC/PRT with timing…….” appears, click
“Cancel” and then escape the “SETUP” mode.
B. Setup time of REC/PRT automatically
a). Drag mouse to highlight the values of HOUR \ MINUTE \ SECOND under
the title “START” then input the expected values.
b) . D rag mouse to highlight the values of HOU R \ M I N UTE \ SE CO N D u n de r
the title “STOP” then input the expected values.
c). Drag mouse to highlight the values of HOUR \ MINUTE \ SECOND under
the title “INTERVAL” then input the expected values.
d). Move mouse pointer to “SETUP” located at right bottom then click this button.
33
e). While the “START/STOP REC/PRT with timing…….” appears, click
“OK”.
f ). While the P / R signal appears (in the window of “PANEL”), move mouse
pointer to this signal and then click to set up the Start / Stop time of
record / print.
11. Setup the print mode of the meter :
Move mouse pointer to “Option” then click. Move mouse pointer to
“Setup” and click this button.
34
In the window “SETUP” select the option button named “PRINT MODE” under
the title “MODE”.
Under the “PRINT” title, select a print mode and then click that button.
Escape “SETUP” mode to complete.
35
5-3 Meter Stand alone Operation :
1. How to clear “ Data Logger ”
ðPress “ RECORD ” + “ POWER ON ”,
the LCD will be show “dEL”.
ðAfter “ POWER ON ”
2. How to do trigger record
ðPress “ RECORD ” button 1 second,
ðRelease “ RECORD ” button the
3. How to consecutive Record/Print (Ⅰ)
, release “ RECORD
”, button.
the LCD
= or
LCD
= or
will be display.
symbol will be cle ar .
ðPress and hold “ RECORD ” or “
PRINT ” button approximate 3
second, the
be display.
= or
symbol will
ðRelease “ RECORD ” or “ PRINT ”
button, the LCD
will be disappear.
36
= or
symbol
4. How to set consecutive Record/Print (Ⅱ)
r
ð Press “ SETUP ” button
approximate 3 second, the LCD
should show “ SET ” symbol.
ðRepeat press
Print/Record/Print, Record function,
the LCD will show
ðPress “ SETUP ” button, back to
general mode, the “ SET ” symbol
will be clear and shows
5. How to stop consecutive Record/Print (Ⅰ)
ðPress and hold “ RECORD ” or “
PRINT ” button for 3 seconds to
stop function.
ðStop consecutive Record/Print
123
= by choice LCD.
function the LCD
.
clear
Button, choose
=//=
=//
= or will be
6. How stop consecutive Record/Print (Ⅱ)
ðIf the “ START ” and “ STOP ”
37
function have been set, the printe
will be stopped when “STOP” time
is up.
5-4 Download Datalogger (Graphic Recorder èPC)
Move mouse pointer to “Option” then click. Move mouse pointer to “Memory”
and then click this button to read recorded data in memory.
ò
As above, the total memory size is 128K, 126.8K remained.
38
1. Move mouse pointer to “Numbers of Sets” then click this button to show how many
sets have been recorded in the memory of Graphic Recorder (EEPROM).
2. Move mouse pointer to “Time of Sets” then click this button.
There will be a list that shows the recording time of each set.
39
3. Move mouse pointer to the list and choose a set then click. The number of
records for this set will be shown.
4. For more details of the chosen set, move mouse pointer to “Show Data”
and then click.
SAVE : To save data recorded in HD and to apply for other applications such as
EXCEL, WORD. For example, P.42
a). Move mouse pointer to “SAVE” then click this button.
b). There comes up a dialog box. Input the name of the file if willing to
save as.
c). Click “OK”
40
5-5 Data Conversion
1. Save data by RS-232 communication port in PC.
Open Graphic Recorder software, check if Graphic Recorder connected and
PC is well. If not the “ NO COM ” signal will be displayed, please be sure
RS-232 cable is plugged and secured to COM1 or COM2. After a good
connection, the COM1 or COM2 symbol will be showed on the screen.
Move mouse pointer to “File”, then click. Move mouse pointer to “ SAVE
AS ” then click this button. You can find the “ open file ” dialog box,
please change to new file name : *.xls from original file *.dat then press
enter. For example : test.xls
ê
41
Move mouse pointer to “File”, then click. And move mouse pointer to “
START RECORDING ”, then click this button. The “RECORDING”
signal and No.s of PC record will display from 1 up to end.
42
To end this record function, move mouse pointe r to “ F il e ”, t hen click. Move
mouse pointer to “ STOP RECORDING ”, click the button. The “
RECORDING ” signal will disappear.
2.
Apply for Excel :
Find the EXCEL type. For example, test.xls. Select test.xls then click
Open Microsoft EXCEL, go to “open file ”, from the searching function.
“open” button.
43
From the “ Files of type ”, click the pull down pointer, then choose “All
Files ”, select sample.dat then click “open” button.
The “ Text Import Wizard ” then appears, click finish button, the selected
data will show in excel type.
44
5-6 On-Line Operation with PC
1. Switch off all power related to your PC.
2. Connect the socket (female) of RS-232 cable to PC’s COM1 or COM2.
3. Switch on all power.
4. Connect the male socket of RS-232 cable to Graphic Recorder.
5. Run the Grapher software
Click “ Start ” from Start menu, move mouse pointer to “ Programs ”, then
move pointer to “Omega” (default), move mouse pointer to “GRAPHER”
then click.
6.
Move mouse to available comport (COM1 , COM2) then click.
45
7. Main tableau
Time Setup:Move mouse pointer to “ System Time Set ” then click to
input the time of PC to the Graphic Recorder.
When “ WAIT ” (in the center of the windows) disappeared, setup of
time is completed.
46
RESET:Reset current maximum and minimum values. Move mouse
pointer to “RESET” then click this button to complete.
Save data in HDD:
a). Move mouse pointer to “File” then click. And move mouse pointer to
“ SAVE AS ” then click this button. There shows a dialog box to
open file. Input a file name then press “ OK ”.
ê
47
b). Move mouse pointer to “File” then click. And move mouse pointer to
“Start Recording”, then click this button. The “Recording” meanwhile
from 1 up to end also will show the Numbers of PC recording.
c). If willing to end this recording, please move mouse pointer to “File” then
click. And move mouse pointer to “Stop Recording”, then click this
button. The “Recording” signal will disappear.
48
d). Move mouse point to “File” then click. And move mouse pointer to
“OPEN FILE”, then click this button. There comes a dialog box to open
file. Input the saved file name to read.
To show reading values in “ GRAPH ” type, move mouse pointer to
”Display” then click. And move mouse pointer to “GRAPH” then click
this button.
ê
49
a). Print : to print graph.
b). Limit : Move mouse pointer to “ Limit ”, then click this button. There
comes a dialog box. Input the willing setting and then click “
OK ”. The graph page will show which records on each.
50
To show reading values in LIST type, move mouse pointer to “Display”
then click. And move mouse pointer to “List” and then click this button
to launch.
ê
51
PC SAMPLING:(Default : 2 seconds). Change the sampling time of
PC. Move mouse pointer to “Display” then click. And
move mouse pointer to “Sample Rate” then click this
button.
There shows a dialog box “ Input Sampling Time ”. Input the willing
seconds the click “ OK ” to complete this setting.
52
Operation of connecting with other meters :
A. Tes meters :
a). Move mouse pointer to “Option” then click. And move mouse pointer
to “Setup” then click this button.
b) In the window “SETUP”, move mouse pointer to “TES METER” which
is under the title “MODE” then click this option button.
c). Select an expected meter and then press “OK” to complete.
B. Other meters not in the list :
a). Move mouse pointer to “Option” then click. And move mouse pointer
to “Setup” then click this button.
b). Comply with the output voltage / current of which meters then refer
to P.27 (7).
c). Comply with the unit of which meters then refer to P.28 (8).
53
Process Transducer ( Od Mode )
a). Move mouse pointer to “ Option ” then click. Move mouse pointer to
“ Od Mode ” and then click this button. There shows an “ Input HiHi
Alarm ” dialog box. (as below)
b). In this dialog box, input the willing values and then press “ OK ” to
complete.
54
2
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.