Olympus PSD-30 User manual

INSTRUCTIONS

ELECTROSURGICAL UNIT

PSD-30

Contents

Contents
Labels and Symbols..................................................................... 1
Intended use ............................................................................................ 3
Applicability of endoscopic treatment ....................................................... 3
Instruction manual..................................................................................... 4
User qualifications .................................................................................... 4
Instrument compatibility ........................................................................... 4
Repair and modification ............................................................................ 5
Signal words ............................................................................................. 5
Dangers, warnings and cautions .............................................................. 5
Chapter 1 Checking the Package Contents............................ 12
Chapter 2 Nomenclature and Functions................................. 14
2.1 Symbols and descriptions ............................................................... 14
2.2 Front panel...................................................................................... 17
2.3 Rear panel ...................................................................................... 21
2.4 Foot switch (MAJ-811) .................................................................... 22
2.5 S-cord (MAJ-813)............................................................................ 23
2.6 P cord (MAJ-814) ............................................................................ 23
Chapter 3 Installation and Connection ................................... 24
3.1 Installation of equipment ................................................................. 24
3.2 Connection to an AC mains power supply ...................................... 25
3.3 Connection of the foot switch.......................................................... 26
3.4 Connection of the patient plate ....................................................... 27
3.5 Connection of the S-cord (when using an endoscope with a S-cord
connector mount) ............................................................................ 32
3.6 Connection of the A-cord ................................................................ 33
Chapter 4 Inspection ................................................................ 35
4.1 Inspection of the power supply ....................................................... 35
4.2 Inspection of the foot switch............................................................ 38
4.3 Inspection of the warning functions................................................. 39
4.4 Inspection of the cut mode and output ............................................ 40
4.5 Inspection of the coagulation mode and output .............................. 41
ELECTROSURGICAL UNIT PSD-30
i
Contents
Chapter 5 Operation.................................................................. 43
5.1 Turn the power ON ......................................................................... 44
5.2 Selection of cut mode ..................................................................... 45
5.3 Selection of coagulation mode ........................................................ 46
5.4 Setting output .................................................................................. 47
5.5 Program mode ................................................................................ 48
5.6 Electrosurgery ................................................................................. 51
5.7 Stand-by mode................................................................................ 53
5.8 Adjusting the volume ....................................................................... 55
5.9 Procedure after use ........................................................................ 56
Chapter 6 Care, Storage and Disposal.................................... 58
6.1 Care ................................................................................................ 58
6.2 Storage ........................................................................................... 59
6.3 Disposal .......................................................................................... 59
Chapter 7 Troubleshooting ...................................................... 60
7.1 Troubleshooting guide .................................................................... 60
7.2 Alarm functions ............................................................................... 69
7.3 Returning the electrosurgical unit for repair .................................... 73
Appendix........................................................................................ 74
Construction ............................................................................................. 74
System chart ............................................................................................. 75
Operating environment ............................................................................. 77
Specifications ............................................................................................ 78
EMC information........................................................................................ 86
ii
ELECTROSURGICAL UNIT PSD-30

Labels and Symbols

Safety-related labels and symbols are attached to the instrument at the location
shown below.
If labels or symbols are missing or illegible, contact Olympus.
Rear side
Labels and Symbols
Fuse rating
CE marking
Potential equalization terminal
Foot switch connector
ELECTROSURGICAL UNIT PSD-30
Electrical rating
Product name, rated voltage and current are displayed.
Serial number
Crossed-out wheeled bin’ symbol
1
Labels and Symbols
Electrical rating
ELECTROSURGICAL UNIT
MODEL PSD-30 INPUT 220V – 240V
2A 50/60Hz
MONOPOLAR : 50W/300
(SOFT/AUTO STOP 100Ω)
350kHz INT.10S/30S
Back cover of this instruction manual
Manufacturer
Authorised representative in the European Community
2
ELECTROSURGICAL UNIT PSD-30
Important Information — Please Read Before Use
Important Information — Please Read Before Use
High frequency leakage current or spark discharge may cause patient,
operator and assistant burns. Follow the dangers, warnings and
cautions described in this instruction manual when using this
electrosurgical unit.
Always prepare for an emergency operation in case of unintentional
patient burn, bleeding and perforation.

Intended use

This electrosurgical unit has been designed to be used with an Olympus endoscope (fiberscope or videoscope) compatible with electrosurgery, light
source, electrosurgical accessories and other ancillary equipment for general
and endoscopic electrosurgery (cutting and coagulation).
Do not use this electrosurgical unit for any purpose other than its intended use.

Applicability of endoscopic treatment

If there is an official standard on the applicability of endoscopic treatment as
defined by a national or local medical administration, or other institution, such as
an academic society, follow that standard when performing the procedure.
Before performing endoscopic treatment, thoroughly study the properties,
purposes, effects and possible risks (the nature, extent, probability and
imminence) of the planned treatment and any alternative therapeutic method that can be performed. Carry out endoscopic treatment only when its benefits
outweigh its risks. Fully explain to the patient the possible benefits and risks of
endoscopic treatment as well as those of any therapeutic method(s) that can be
performed instead of endoscopy, and perform endoscopic treatment only after
patient consent is granted. During endoscopic treatment, continue to evaluate
the potential benefits and risks, and stop the treatment if the risks become greater than the possible benefit to the patient.
ELECTROSURGICAL UNIT PSD-30
3
Important Information — Please Read Before Use

Instruction manual

This instruction manual contains essential information on using this
electrosurgical unit safely and effectively.
Before use, thoroughly review this manual and the manuals of all equipment
which will be used during the procedure and use the equipment as instructed.
Keep this and all related instruction manuals in a safe, accessible location.
If you have any questions or comments about any information in this manual, please contact Olympus.
Terms used in this manual
Wall mains outlet
An electrical outlet that has a terminal used exclusively for grounding.

User qualifications

If there is an official standard that outlines the qualifications required of medical personnel performing endoscopic treatments as defined by a national or local
medical administration or other institution, such as an academic society, follow
that standard. If there is no such standard, the operator of this instrument must
be a physician who has been approved by the hospital’s medical safety manager
or surgical department manager.

Instrument compatibility

Refer to the “System chart” in the Appendix to confirm that this electrosurgical
unit is compatible with the ancillary equipment being used.
Using incompatible equipment can result in patient injury and/or equipment
damage.
4
ELECTROSURGICAL UNIT PSD-30

Repair and modification

This electrosurgical unit does not contain any user-serviceable parts. Do not
disassemble, modify or attempt to repair it; patient or user injury and/or
equipment damage can result. This instrument is to be repaired by Olympus
technicians only.
Some problems that appear to be malfunctions may be correctable by referring
to Chapter 7, “Troubleshooting”. If the problem cannot be resolved using the information in Chapter 7, contact
Olympus.

Signal words

The following signal words are used throughout this manual:
Important Information — Please Read Before Use
Indicates an imminently hazardous situation which, if not
avoided, will result in death or serious injury.
Indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not
avoided, may result in minor or moderate injury. It may also
be used to alert against unsafe practices or potential
equipment damage.
Indicates additional helpful information.

Dangers, warnings and cautions

Follow the dangers, warnings and cautions described below when handling this
electrosurgical unit. This information is to be supplemented by the dangers,
warnings and cautions given in each chapter.
ELECTROSURGICAL UNIT PSD-30
5
Important Information — Please Read Before Use
The HF equipment, when applied to a patient with a
To prevent shock hazards, never apply the PSD-30 to the
When using the electrosurgical unit on or in the vicinity of the
Never install or operate the electrosurgical unit in locations
pacemaker implanted, may cause malfunctioning or failure of
the pacemaker and may seriously affect the patient. Always
confirm that it is safe to proceed with a cardiologist and the manufacturer of the pacemaker before proceeding.
heart in combination with TYPE B or TYPE BF applied parts.
heart, be sure to use it with the minimum necessary output.
Spark discharge during operation may affect the heart.
where:
the concentration of oxygen is high.
oxidizing agents (such as nitrous oxide (N
O)) are present
2
in the atmosphere.
flammable anesthetics are present in the atmosphere.
flammable gases are present in the atmosphere.
Otherwise, explosion or fire may result and cause damage to
human body. Because this electrosurgical unit is not explosion-proof.
The PSD-30 has been designed for flexible endoscopic
applications. Therefore, do not use it for surgical operations.
Always use the S-cord when endoscopic treatment is performed with an endoscope equipped with the S-cord
connector mount. When using an endoscope without the
S-cord connector mount, refer to the instruction manual and
confirm that the electrosurgery is feasible with this
endoscope. If the electrosurgical unit is used in conjunction
with an endoscope that is not feasible with the electrosurgery, a burn to the patient, the operator or the
assistant may occur.
If the electrosurgical unit fails and the output is stopped
during treatment, it may be impossible to continue treatment due to adhesion of the endo-therapy accessory to a mucous
tissue or other accident. Prepare a spare electrosurgical unit
as backup to ensure that the procedure can be completed
without complication in case of a malfunction.
6
ELECTROSURGICAL UNIT PSD-30
Important Information — Please Read Before Use
Always have a defibrillator ready in case of a cardiac
emergency. During operation of the defibrillator, remove the
endoscope from the patient.
To respond to possible patient bleeding, prepare at least the
following three hemostatic procedures: coagulation, clipping
and local injection.
To prepare for possible accidents, emergency equipment for
life-saving, intubation and appropriate pharmaceuticals
should be located in or near the procedure room.
Never attach the patient plate in the vicinity of a metal
implant. The tissue in the vicinity of the metal implant may be
burned.
Keep the electrosurgical unit and ancillary cords (A-cord,
P cord, S-cord) as far away as possible from other electromedical equipment and cords. If placed too close, high
frequency signals or spark discharge noise from the
electrosurgical unit may interfere with the operation of other
equipment.
To prevent patient burns, the electrosurgical unit and ancillary
cords (A-cord, P cord, S-cord) should not come in contact
with the patient or metal parts of the operating table.
Furthermore, patients should also keep away from metallic
parts of the operating table or other devices.
Never loop the cords (A-cord, P cord, S-cord) or bundle
cords together with cords belonging to other medical
equipment. The high frequency signals or spark discharge
noise generated by the electrosurgical unit may interfere with
the operation of other medical equipment.
The electrosurgical unit may interfere with electromedical
equipment used in conjunction with it. Before use, thoroughly
confirm the compatibility of all the equipment.
Do not bring the tip of the electrosurgical accessory close to
a metal clip or other accessories. The tissue around the
metal clip or the accessory may be burned.
To prevent electric shock, the electrosurgical unit is designed
to function with insulated connectors that have Type CF
electromedical equipment classifications (A-cord, P cord,
P plate, S-cord). Therefore, to prevent shock caused by
leakage current from other electrical equipment applied to
the patient, the connectors from the other equipment should
not be grounded.
ELECTROSURGICAL UNIT PSD-30
7
Important Information — Please Read Before Use
Should any abnormal output be suspected during operation,
To prevent operator shock, electrosurgical unit malfunction
To prevent patient burns, the patient’s skin surfaces should
To prevent patient burns, the patient’s clothes must be dry.
To prevent operator and assistant burns, be sure to wear
immediately terminate the use of the equipment and turn
OFF the electrosurgical unit.
and electrosurgical unit damage, keep liquids away from all
electrical equipment. If liquid gets on or into the
electrosurgical unit, terminate operation immediately and
contact Olympus.
not touch each other (e.g. bare arm and side of chest) or any
metal items in the procedure room. Remove any metallic
items from the patient (wristwatches, jewelry, etc.) before starting the procedure.
chemical-resistant gloves during operation.
If the intestines contain a flammable gas, replace this gas
with air or a non-flammable gas before performing the
operation, to minimize the risk of fire or explosion.
The endoscopic treatment performed should not include a
therapy with which part of the treated tissue (poly head, etc.)
or part of the endoscope distal end or endo-therapy
accessory is in contact with or close to normal tissue. Otherwise, current flows to the normal tissue through the part
of the treated tissue, the metallic parts at the endoscope’s
distal end or endo-therapy accessory and may cause burns.
Be sure that the distal end of endoscope or accessory does not contact bridging fluids surrounding the target tissue.
Electric current may flow to the surrounding normal tissue via
the fluids and cause burns.
Never grasp the target tissue with non-insulated grasping forceps. Non-insulated grasping forceps will disperse the
electric current and normal operation may be impeded.
To ensure electrical safety, the electrosurgical unit should not
be used in conjunction with:
Electrical equipment whose safety against leakage
current is not guaranteed.
Electrosurgical equipment whose safety in combined use
is not guaranteed.
8
ELECTROSURGICAL UNIT PSD-30
Important Information — Please Read Before Use
If the electrosurgical unit is used in conjunction with a
non-Olympus electrosurgical unit, keep the electrosurgical
accessory away from the target area while the electrosurgical
unit is in operation. Do not activate output of both units simultaneously. Patient or operator injury may occur due to
the concentration of electric current.
When using an electrocardiograph or other physiological
monitoring equipment simultaneously with the electrosurgical unit on a patient, any monitoring electrodes should be placed
as faraway as possible from the electrodes and patient plate
used with the electrosurgical unit. If placed too close, high
frequency signals or spark discharge noise from the
electrosurgical unit may interfere with the operation of an
electrocardiograph or other physiological monitoring equipment. Needle monitoring electrodes should not be
used, as they may cause patient burns. Physiological
monitoring equipment incorporating high frequency current
limiting devices is recommended.
Use physiological monitoring equipment throughout the
entire procedure, for continuous observation of the patient’s
condition.
The open-circuit voltage output characteristics of the
electrosurgical unit are shown below. When setting the output level, first set it to a low output level and increase it
gradually. If the output is initially set to an excessive level, the
electrode’s insulation could be damaged, which could cause
operator and/or patient burns. Furthermore, it is
recommended that you perform basic experiments before
using the electrosurgical unit. If the instruction manual for the endoscopic accessory to be used stipulates a rated voltage,
the output should be set so that it does not exceed that
voltage.
ELECTROSURGICAL UNIT PSD-30
9
Important Information — Please Read Before Use
1000
800
600
400
200
0
060
(1)
(3)
(6)
(5)
(4)
20 40
OUTPUT (W)
(2)
(1) PURE (2) BLEND1 (3) BLEND2 (4) SOFT (5) AUTO STOP (6) FORCED
Flammable agents used for cleaning and disinfection must be allowed to evaporate before the electrosurgical unit is used.
Also ensure that flammable solutions are neither on the
patient’s skin nor in the patient’s body cavity when the
electrosurgical unit is used.
To prevent patient burns when the electrosurgical unit is used
in conjunction with a non-Olympus electrosurgical unit,
ensure that each unit is supplied with power from wall mains
outlets. Also ensure that each unit is supplied with power
from separate breakers.
During operation, temporarily unused electrodes should be
stored in an electrically insulated container. Unused
electrodes should never be placed on the patient otherwise it
may cause operator and/or patient burns.
The electrosurgical unit should not be used for performing
circumcisions.
To prevent patient burns, never use cords with scratches or cracks.
Studies have shown that smoke generated during
electrosurgical procedures can be irritating and potentially
harmful to surgical personnel. These studies recommend the use of surgical masks and adequate ventilation of smoke by
the use of surgical smoke evacuators or other means.
10
ELECTROSURGICAL UNIT PSD-30
Important Information — Please Read Before Use
William S.Sawchuk, et al., “Infections Papilomavirus in the
Vapor of Warts Treated With Laser or Electrocoagulation:
Detection and Protection”, Journal of American Academy
of Dermatology, Vol.21, No.1 (July, 1989): 41-49. Yoshifumi Tomita, et al., “Mutagencity of Smoke
Condensates Induced by CO
-Laser Irradiation and
2
Electrocauterization”, Mutation Research, Vol.89 (1989):
145-149.
Evaluation Report No.85-126. (National Technical
Information Service, 1985): 1-28.
When the patient is moved after the dispersive electrode has
been attached, confirm that the dispersive electrode is still in
proper contact with the patient otherwise it may cause
operator and/or patient burns.
Do not use the electrosurgical unit in a location exposed to
strong electromagnetic radiation (microwave medical
treatment equipment, short wave medical treatment
equipment, MRI, radio or mobile phone). Electrosurgical unit
malfunction can occur.
To prevent electrosurgical unit damage, never short-circuit
electrodes (accessories, patient plate).
Repairs should be performed only by Olympus.
The electrosurgical unit should only be used in accordance
with the conditions described in “Operating environment” in
the Appendix. The use under other conditions may impede
normal performance, and/or result in equipment damage.
During treatment, be sure that the distal (metallic) end of the
endoscope, and metallic parts of accessory do not touch the
normal tissue surrounding the target tissue. This action may
cause severe burns.
ELECTROSURGICAL UNIT PSD-30
11

Chapter 1 Checking the Package Contents

Chapter 1 Checking the Package
Contents
Match all items in the package with the components shown below. Inspect each
item for damage. If the electrosurgical unit is damaged, a component is missing
or you have any questions, do not use the electrosurgical unit; immediately
contact Olympus.
Electrosurgical unit (PSD-30)
Foot switch (MAJ-811)
Power cord
P cord (MAJ-814)
S-cord (MAJ-813)
12
Instruction manual
ELECTROSURGICAL UNIT PSD-30
Spare fuse (MAJ-1052, 2 pcs)
Pplate (MAJ-897, 10 pcs)
Chapter 1 Checking the Package Contents
The following Olympus items are optional and may be purchased separately:
A-cord (MA-255, MH-969)
P plate (MB-574)
P cord (MAJ-3)
P adapter (MAJ-812)
For other combinations, refer to the “System chart” in the Appendix.
Place the instruction manual next to the electrosurgical unit
or in another easily accessible place.
Before using an optional item, thoroughly review and
understand the instruction manual provided with that item.
ELECTROSURGICAL UNIT PSD-30
13

Chapter 2 Nomenclature and Functions

Chapter 2 Nomenclature and Functions

2.1 Symbols and descriptions

Warning section
Refer to instructions
Patient plate
S-cord
High frequency output
Output timer
Coag. section
Soft
Auto stop
Forced
14
ELECTROSURGICAL UNIT PSD-30

Cut section

Pure
Blend1
Blend2
Power switch
Power ON/OFF
Chapter 2 Nomenclature and Functions
Front panel
Type CF applied parts (defibrillation protected equipment)
STAND-BY
Program
ELECTROSURGICAL UNIT PSD-30
15
Chapter 2 Nomenclature and Functions

Connector section

Active
S-cord
Patient plate
Refer to instructions
Connection for neutral electrode (isolated from earth at high
frequency)
Rear panel
Potential equalization terminal
Foot switch
Fuse
Alternating current
Serial number
16
ELECTROSURGICAL UNIT PSD-30

2.2 Front panel

Chapter 2 Nomenclature and Functions
2. Program switch
1. Stand-by switch
3. Cut mode selector
4. Output control buttons
9. Power switch
ELECTROSURGICAL UNIT PSD-30
7. Active connector
5. Coagulation mode selector
8. Patient plate connector
6. S-cord connector
17
Chapter 2 Nomenclature and Functions
10. Patient/S-cord warning indicator
11. Output indicators
12. High frequency output warning indicator
13. Output time warning indicator
19. Power indicator
18. CQM indicator
17. Cut mode indicators
18
14. Coagulation output setting indicator
15. Coagulation mode indicators
16. Cut output setting indicator
ELECTROSURGICAL UNIT PSD-30
Chapter 2 Nomenclature and Functions
1. Stand-by switch
This switch disables high frequency output and change of setting. In the
stand-by mode, the output tone volume can be adjusted.
2. Program switch
This switch is pressed to store output setting and output mode and to recall
stored output setting and output mode.
3. Cut mode selector
This selector is pressed to select the cut mode.
4. Output control buttons
When the “ ” button is pressed, the value of the output setting indicator increases. When the “ ” button is pressed, the value of the output setting
indicator decreases.
5. Coagulation mode selector
This selector is pressed to select the coagulation mode.
6. S-cord connector
This connector connects the electrosurgical unit to the endoscope via the
S-cord.
7. Active connector
This connector connects the electrosurgical unit to the active electrode via
the A-cord.
8. Patient plate connector
This connector connects the electrosurgical unit to the patient plate via the
P cord.
9. Power switch
This switch is pressed to turn ON/OFF the electrosurgical unit.
10. Patient/S-cord warning indicator
This indicator lights to indicate a defective connection of cords or broken
wires in the P cord, patient plate or S-cord.
11. Output indicators
During output, the indicator corresponding to the selected output (cut,
coagulation) is lit.
12. High frequency output warning indicator
This indicator lights to indicate that the output level is not in accordance with the set output level.
13. Output time warning indicator
This indicator lights if output continues for longer than the specified time.
14. Coagulation output setting indicator
This indicator displays the value set with the output control buttons.
15. Coagulation mode indicators
These indicators light when the corresponding coagulation mode selector is
pressed.
ELECTROSURGICAL UNIT PSD-30
19
Chapter 2 Nomenclature and Functions
16. Cut output setting indicator
This indicator displays the value set with the output control buttons.
17. Cut mode indicators
These indicators light when the corresponding cut mode selector is pressed.
18. CQM indicator
When a split-type patient plate is connected, the CQM indicator lights to
indicate that the connection between the patient and the patient plate is
correct.
19. Power indicator
This indicator lights when the electrosurgical unit is turned ON.
20
ELECTROSURGICAL UNIT PSD-30

2.3 Rear panel

Chapter 2 Nomenclature and Functions
4. Potential equalization terminal
3. Foot switch connector
1. Fuse box
This box prevents short-circuits within the unit if equipment malfunctions.
2. AC power inlet
This inlet accepts the power cord.
3. Foot switch connector
This connector connects the foot switch (MAJ-811) to the electrosurgical
unit.
4. Potential equalization terminal
To be equipotential, connect this terminal to a potential equalization busbar
of the electrical installation.
1. Fuse box
2. AC power inlet
ELECTROSURGICAL UNIT PSD-30
21
Chapter 2 Nomenclature and Functions

2.4 Foot switch (MAJ-811)

2. Coagulation pedal
1. Cut pedal
3. Foot switch plug
1. Cut pedal
This pedal starts and stops the cut output.
2. Coagulation pedal
This pedal starts and stops the coagulation output.
3. Foot switch plug
This plug connects the foot switch to the electrosurgical unit.
22
ELECTROSURGICAL UNIT PSD-30

2.5 S-cord (MAJ-813)

1. Plug on the unit side
1. Plug on the unit side
This plug connects to the electrosurgical unit to the endoscope via S-cord.
Chapter 2 Nomenclature and Functions
2. Plug on the endoscope side
2. Plug on the endoscope side
This plug connects the endoscope to the electrosurgical unit via the S-cord.
2.6 P cord (MAJ-814)
3. Lever-locking arm
2. Clamp
1. Plug on the unit side
This plug connects the patient plate to the electrosurgical unit via the P cord.
1. Plug on the unit side
2. Clamp
This clamp connects the patient plate to the electrosurgical unit via the
P cord.
3. Lever-locking arm
This arm locks the clamp securely in the connector of the patient plate.
ELECTROSURGICAL UNIT PSD-30
23

Chapter 3 Installation and Connection

Chapter 3 Installation and Connection
Prepare the electrosurgical unit and compatible equipment (shown in the
“System chart” in the Appendix) before each use, and refer to the instruction
manuals of each system component. Install and connect the equipment as
follows:

3.1 Installation of equipment

Install the unit on a stable, level surface. Otherwise, the unit
could fall, causing equipment damage and/or injury to the
operator or patient.
If the electrosurgical unit is placed on a cart, the cart must be of adequate strength and size to hold the unit securely.
Never place the electrosurgical unit on its side or upside
down. Otherwise, the electrosurgical unit does not work
correctly.
Place the instruction manual near the electrosurgical unit or
in another easily accessible place.
1. Ensure that the operation of the electrosurgical unit is in accordance with
the precautions described in “Dangers, warnings and cautions” on page 5.
2. Place the electrosurgical unit on a level, stable bench or cart.
24
ELECTROSURGICAL UNIT PSD-30
Chapter 3 Installation and Connection

3.2 Connection to an AC mains power supply

Connect the power plug of the power cord directly to a
grounded wall mains outlet. If the electrosurgical unit is not
grounded properly, it can cause an electric shock and/or fire.
Do not bend, pull or twist the power cord. Electric shock,
equipment damage and/or fire can result.
To prevent the risk of electric shock, the housing of the
electrosurgical unit must be grounded. Always connect the
power cord plug to a properly grounded hospital grade AC
receptacle (wall mains outlet).
Always plug the power cord into a 3-pin outlet. Do not use a
3-pin/2-pin adapter, as it can impair the safe operation of the
unit.
Do not allow the power cord to become wet. Electric shock,
equipment damage or fire can result.
Always use the power cord provided with the electrosurgical
unit. Never attempt to modify the power cord.
If the same circuit breaker is used to supply power to other
electrosurgical equipment, carefully consider the power
requirements of the additional equipment and use circuit
breakers that have ample capacity. Otherwise, the
electrosurgical unit does not work correctly.
1. Confirm that the electrosurgical unit is turned OFF.
2. Connect the power cord to the AC power inlet of the electrosurgical unit and
a wall mains outlet.
ELECTROSURGICAL UNIT PSD-30
25
Loading...
+ 67 hidden pages