Olympus esg-400 User Manual

Page 1
j
MAINTENANCE / ON-SITE - MANUAL
Electro Surgical Generator
ESG-400
Page 2
ESG-400
Indicates an imminently hazardous situation which, if not avoided, will result in
d
Indicates a potentially hazardous situation which, if not avoided, could result in
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices
The user must have received appropriate training in using, servicing,

Introduction

The intended use depends on the approval of the country. Refer to the instructions for use of the electrosurgical unit.
Maintenance instructions
This maintenance manual contains essential information on using and maintaining this electrosurgical generator safely and effectively. Instructions for the operation of this electrosurgical generator and related danger, warnings and cautions concerning electrosurgery are beyond the scope of this maintenance manual. Before using and maintaining, thoroughly review this manual and the instructions for use or maintenance manuals of all equipment which will be used during maintenance. Use the equipment as instructed. Keep this manual in a safe, accessible location. If you have any questions or comments about any information in this maintenance manual, cont act O ly mpus.
Signal words
The following signal words are used throughout this maintenance manual:
INTRODUCTION
DANGER
WARNING
CAUTION
NOTE
User qualifications
The user must have received appropriate training in using this electrosurgical generator. The following instructions are for use by qualified personnel only. Use of this maintenance manual by other individuals is prohibited. The training will be provided by authorized representatives of Olympus during installation and commissioning.
NOTE
Federal Law of the USA restricts this device to use by, or on the order of, a physician.
death or serious injury.
death or serious injury.
or potential equipment damage. Indicates additional helpful information.
adjustment, updating and upgrading this electrosurgical generator.
7.022.211 / ISSUE 5 2 / 110 Introduction
Page 3
ESG-400
User
The safety and effectiveness of electrosurgical interventions depend not only o the equipment used, but also to a major extent on factors which are under the control of the user. It is therefore extremely important to read, understand and follow the instructions supplied with the electrosurgical generator and the accessories in order to ensure safety and effectiveness.
Always use the electrosurgical generator as outlined in this will not only impede functions and prevent optimum performance, but may cause equipment damage and
ications. Before each use, always inspect the equipment as outlined in
this
The electrosurgical generator and the footswitch must undergo a safety check in yearly intervals in accordance with
Environmental
Be sure that this electrosurgical unit is not used adjacent to or stacked with other equipment (other than the components of this electrosurgical unit or system) to avoid electromagnetic interference.
Before use, thoroughly confirm the compatibility of all equipment.
To ensure electrical safety, the electrosurgical unit s hould not be used in conjunction w
The electrosurgical generator complies with the electromagnetic compatibility (EMC) standard. Never equipment. If an auxiliary computer system is in use together with the electrosurgical generator and endoscopic imaging techniques, the image on the monitor might f the instructions in “Electromagnetic Compatibility (EMC) information” in the Appendix of the instruction for use regarding electromagnetic ambient conditions.
Never loop the cords (active cord, bipolar cord, neutral electrode cor with cords belonging to other medical equipment. The high frequency s ignals or spark discharge noise generated by the unit may interfere with the operation of other medical equipment.
Do not use the electrosurgical unit in a location exposed to strong electromagnetic radiation (microwave or short-wave medical treatment equipment, Magnetic Resonance Imaging, radio or mobile phone equipment). Electrosurgical unit malfunction can occur.
The elect rosurgi cal
should only be used under the conditions as described in chapter 1 equipment damage.
Precautions
High frequency leakage c urrent or spark discharge may cause user burns.
Follow the dangers, warnings and cautions given below when handling and servicing this electrosurgical unit. This information is to be supplemented by the dangers, warnings and cautions given in each chapter.
-related error prevention
WARNING Improper use
n the design of
maintenance manual. Improper use
/ or compl
maintenance manual.
WARNING Annual safety checks / Inspection
the national statutory regulations (refer to chapter 7 “Inspection”).
conditions
CAUTION Interference of the unit with other equipment
ith:
Electrical equipment whose safety against leakage current is not guaranteed.
Electrosurgical equipment whose safety in com bined use is not guaranteed.
theless, when the electrosurgical generator is active it may disturb neighboring electronic
reeze or blackout. Follow
d) or bundle cords together
CAUTION Unsuitable temperature and humidity
generator
-3 (Limitations). Use under other conditions may impede normal performance and / or result in
7.022.211 / ISSUE 5 3 / 110 Introduction
Page 4
ESG-400
Accessories
Do not apply excessive bending, straining, or squeezing force to any cords. It may cause malfunction.
The electrosurgical generator shall only be used with compatible accessories. When connecting accessories (cords, electrodes, HF instruments) avoid output settings where the maximum output voltage of the electrosurgical generator may exceed the rated accessory voltage (refer to “Mode characteristics”, “Output characteristics” in the Appendix of the ins instruction manual of the accessory). For a list of compatible neutral electrodes, refer to “Specifications” in the Appendix of the instruction for use.
Before use, the electrosurgical unit and accessories must be examined for damage. All communication cables and its plugs must be free of scratches and cracks. Cables and accessories with damaged insulation or connections must not be used.
Electric shock
To prevent the risk of electric shock, the housing of the electrosurgical unit must be grounded. Always connect the power cord plug to a properly grounded wall outlet. Do not use a 3
To prevent user shock, malfunction and damage of the electrosurgical unit, keep liquids away from all electrical equipment.
When taking measurements or troubleshooting of the electrosurgical unit, take appropriate precautions, such as using isolated tools and equipment, using the “one han
When the housing is opened, there is a danger of electric shock. The unit must only be serviced by
Burns
The maximum output voltage characteristics of the electrosurgi diagrams in “Output characteristics” in the Appendix of the instruction for use. When setting the power level, first set it to a low level and increase it gradually. If the output is initially set to a high level, the electro
or patient burns. However, certain modes may present an unacceptable risk at low output power settings. For example, with the PulseCut fast mode or PulseCut slow mode, the ris k of an excessive thermal effect rises if the output power setting is too low. Therefore, it is recommended that you perform basic testing before using the electrosurgical generator. If the instruction manual of the HF instrument to be used stipulate a rated voltage, the output should exceed that voltage.
High frequency, high voltage signals that can cause severe burns are present in the monopolar when testing and troublesho
WARNING
CAUTION Non-compatible accessories and accessory damage
WARNING Grounding failure
-pin / 2-pin adapter, as it can impair safe operation of the unit.
WARNING User shock
Mechanical stress
truction for use, and the
CAUTION Injury during servicing
authorized technicians.
WARNING User
cal generator are shown in the
de’s insulation may be damaged and cause user and /
be set so that it does not
/ bipolar sockets descri bed in thi s maintenance manual. Take appropriate precautions
oting this area of the electrosurgical unit.
7.022.211 / ISSUE 5 4 / 110 Introduction
d rule,” etc.
Page 5
ESG-400
Fire
The risk of flammable gases or other materials being ignited exists with any contact of electrical energy. Precautionary measures must be taken to keep flammable materials and substances away from an active electrosurgical unit (do not use flammable oxygen). Otherwise, explosion or fire may result and cause serious injuries. This electrosurgical uni
Flammable agents used for cleaning and disinfection must be allowed to evaporate before the electrosurgical unit is used and servi
Non
Haz
The electrosurgical unit may be contaminated with infections materials; therefore, all surfaces of the unit’s housing should be cleaned before servicing according to chapter
Should any abnormal output be suspected during operation, immediately terminate the use of the equipment by releasing the footswitch. If the footswitch does not react, switch off the electrosurgical unit. Otherwise, malfunction of the equipment m in output.
Take additional precautions for service technicians, when using the unit’s service operation mode (see chapter 15, Service operation mode).
To prevent electrosurgic
circuit electrodes (accessories, neutral
electrodes).
In the event of a defect or malfunction in the unit, an undesirably high output power may be emitted.
Repairs must only be carried out by Olympus or a firm authorized by Olympus.
Preventive maintenance (inspection / periodic safety check) must only be carried out by a qualifie
/ Explosion
DANGER Ignitable anaesthetics / fire supporting gases
anesthetics, nitrous oxide or
t is not explosion-proof. Do not use the unit within an explosion zone.
WARNING Ignitable cleaning- and disinfection agents
ced.
-flammable agents should be used for cleaning and disinfection wherever possible.
WARNING Risk of fire
Replace fuses as marked. The fuses
authorized technicians.
ards and complications
WARNING Contamination
1-8 (Cleaning).
WARNING Output performance
ay cause an unintended increase
WARNING
CAUTION Unit defect
DANGER Unit defect
Never use the electrosurgical unit if an abnorm ali ty is suspe cted.
Service persons
al unit damage, never short-
Repair and Maintenance
CAUTION Repair
CAUTION Maintenance
d person / technician.
7.022.211 / ISSUE 5 5 / 110 Introduction
Page 6
ESG-400
Copyright
©2011 Olympus Winter & Ibe GmbH. All rights reserved. Unauthorized reproduction or distribution in part or in whole is prohibited.
Trademarks
OLYMPUS is a registered trademark of the Olympus Corporation. The company names, product names, and proprietary technic al terms in this document are the trademarks or
registered trademarks of their respective owners.
7.022.211 / ISSUE 5 6 / 110 Introduction
Page 7
ESG-400

CONTENT

INTRODUCTION ................................................................................................................ 2
CONTENT .......................................................................................................................... 7
CHAPTER 1: PRODUCT SPECIFICATIONS .................................................................... 9
1 OUTLINE ...................................................................................................................... 10
1-1 Intended Use .................................................................................................................................... 10
1-2 Compatibility ..................................................................................................................................... 10
1-3 Expected service life ........................................................................................................................ 10
2 FEATURES ................................................................................................................... 11
2-1 Application Modes ............................................................................................................................ 11
2-2 Accessories ...................................................................................................................................... 12
3 LIMITATIONS ................................................................................................................ 13
4 SPECIFICATIONS ........................................................................................................ 14
4-1 ELECTROSURGICAL GENERATOR ESG-400 (REF: WB91051W) .............................................. 14
4-2 Power cords (4.5 m angled plug) ..................................................................................................... 15
4-3 Footswitch (REF: WB50402W, double pedal) ................................................................................. 16
4-4 Footswitch (REF: WB50403W, single pedal, optional) .................................................................... 16
4-5 Neutral electrode cableP-cord” (REF: MAJ-814, optional) ............................................................ 16
4-6 Communication cable 0.25 m (REF: MAJ-18 71, optional) .............................................................. 17
4-7 Communication cable 10 m (REF: MAJ-1872, optional) ................................................................. 17
4-8 Adapter for UHI-2/3 (REF: MAJ-1873, optional) .............................................................................. 17
5 NAME AND FUNCTION OF EACH PART .................................................................... 18
5-1 Symbols and descriptions ................................................................................................................ 18
5-1-1 Safety related symbols ............................................................................................................ 18
5-1-2 Front panel .............................................................................................................................. 19
5-1-3 Touch screen ........................................................................................................................... 19
5-1-4 Rear panel ............................................................................................................................... 21
5-2 Front panel ....................................................................................................................................... 22
5-3 Rear panel ........................................................................................................................................ 24
5-4 Bottom panel .................................................................................................................................... 25
5-5 All screen ......................................................................................................................................... 25
5-6 Set screen ........................................................................................................................................ 26
5-7 Mode screen .................................................................................................................................... 27
5-8 Footswitch with two pedals .............................................................................................................. 28
5-9 Footswitch with one pedal (optional)................................................................................................ 28
5-10 Neutral electrode cable “P-cord” (optional) ...................................................................................... 29
6 CONNECTOR ............................................................................................................... 30
6-1 Docking Connector .......................................................................................................................... 30
6-2 Monopolar Standard 1 ...................................................................................................................... 30
6-3 Monoploar Standard 2 (Erbe) ........................................................................................................... 30
6-4 Bipolar Standard 3............................................................................................................................ 30
6-5 Monopolar Neutral Electrode ........................................................................................................... 31
6-6 Foot switch 1 (SIP/SOP) .................................................................................................................. 31
6-7 Foot switch 2 (SIP/SOP) .................................................................................................................. 32
7 SYSTEM DIAGRAM ..................................................................................................... 33
8 CLEANING, STORAGE AND DISPOSAL .................................................................... 34
8-1 Cleaning ........................................................................................................................................... 34
8-2 Storage ............................................................................................................................................. 35
8-3 Disposal of the unit .......................................................................................................................... 35
CHAPTER 2: BLOCK DESCRIPTION ............................................................................. 37
1 BLOCK DESCRIPTIONS .............................................................................................. 38
1-1 Motherboard ..................................................................................................................................... 39
1-2 HVPS Board ..................................................................................................................................... 41
1-3 Generator board ............................................................................................................................... 41
1-4 Relay Board ..................................................................................................................................... 41
1-5 Front Panel ....................................................................................................................................... 42
CHAPTER 3: REPAIR SYSTEM ...................................................................................... 43
7.022.211 / ISSUE 5 7 / 110 Content
Page 8
ESG-400
1 ESG-400 MAIN UNIT .................................................................................................... 44
2 BOARD COMPATIBILITY ............................................................................................. 44
3 OPTIONAL ACCESSORIES ......................................................................................... 44
3-1 WB50402W (Footswitch with two pedals) ....................................................................................... 44
3-2 WB50403W (Footswitch with one pedal) ......................................................................................... 44
3-3 MAJ-814 (Neutral electrode cable “P-cord”) .................................................................................... 44
4 PRECAUTIONS ON FUNCTION AND OPERATION SETTINGS ................................. 44
4-1 General Precautions ........................................................................................................................ 44
CHAPTER 4: TROUBLESHOOTING .............................................................................. 45
1 GENERAL ..................................................................................................................... 46
2 NEUTRAL ELECTRODE OPERATION ........................................................................ 47
3 ERROR SCREEN, CODES AND MEASURES ............................................................. 48
3-1 What to do when no error code is displayed .................................................................................... 50
3-2 What to do when an error code is displayed .................................................................................... 54
CHAPTER 5: INSPECTION ............................................................................................. 73
1 JIGS, TOOLS, AND MEASURING EQUIPMENT FOR INSPECTION .......................... 74
2 INSPECTION PROCEDURES ...................................................................................... 75
2-1 Visual inspection of the electrosurgical generator and accessories ................................................ 76
2-2 Verifying the contact quality monitor function .................................................................................. 80
2-3 Checking the DC resistance (according to IEC 60601-2-2) ............................................................. 82
2-4 Checking the earth resistance (according to IEC 60601-1 and IEC 62353) .................................... 82
2-5 Checking the earth leakage current (according to IEC 60601-1) .................................................... 82
2-6 Checking the patient leakage current (according to IEC 60601-1) .................................................. 83
2-7 Checking the current and power consumption and output waveform .............................................. 84
2-8 Checking the high frequency leakage current (according to IEC 60601-2-2) .................................. 86
2-8-1 Measurement of the monopolar high frequency leakage current under loaded condition ...... 86
2-8-2 Measurement of the monopolar high frequency leakage current under unloaded condition .. 88
2-8-3 Measurement of the bipolar high frequency leakage current under loaded condition ............ 90
2-8-4 Measurement of the bipolar high frequency leakage current under unloaded condition ........ 92
2-9 Checking the output power .............................................................................................................. 94
2-10 Checking for certain features and error messages .......................................................................... 94
2-11 Final test ........................................................................................................................................... 95
2-11-1 Self test.................................................................................................................................... 95
2-11-2 Display and sound check ........................................................................................................ 95
2-11-3 Functionality of push buttons .................................................................................................. 95
2-11-4 Communication test ................................................................................................................. 95
2-11-5 Restore of output power settings ............................................................................................ 96
2-12 Inspection label (For applicable markets) ........................................................................................ 96
3 INSPECTION CARD ..................................................................................................... 97
CHAPTER 6: DEVICE MENU ........................................................................................ 103
1 SAFETY TEST ............................................................................................................ 104
2 SOFTWARE VERSION ............................................................................................... 105
CHAPTER 7: REVISION HISTORY ............................................................................... 107
3 REVISION HISTORY .................................................................................................. 108
7.022.211 / ISSUE 5 8 / 110 Content
Page 9
ESG-400

CHAPTER 1: PRODUCT SPECIFICATIONS

1 OUTLINE ...................................................................................................................... 10
1-1 Intended Use .................................................................................................................................... 10
1-2 Compatibility ..................................................................................................................................... 10
1-3 Expected service life ........................................................................................................................ 10
2 FEATURES ................................................................................................................... 11
2-1 Application Modes ............................................................................................................................ 11
2-2 Accessories ...................................................................................................................................... 12
3 LIMITATIONS ................................................................................................................ 13
4 SPECIFICATIONS ........................................................................................................ 14
4-1 ELECTROSURGICAL GENERATOR ESG-400 (REF: WB91051W) .............................................. 14
4-2 Power cords (4.5 m angled plug) ..................................................................................................... 15
4-3 Footswitch (REF: WB50402W, double pedal) ................................................................................. 16
4-4 Footswitch (REF: WB50403W, single pedal, optional) .................................................................... 16
4-5 Neutral electrode cableP-cord” (REF: MAJ-814, optional) ............................................................ 16
4-6 Communication cable 0.25 m (REF: MAJ-18 71, optional) .............................................................. 17
4-7 Communication cable 10 m (REF: MAJ-1872, optional) ................................................................. 17
4-8 Adapter for UHI-2/3 (REF: MAJ-1873, optional) .............................................................................. 17
5 NAME AND FUNCTION OF EACH PART .................................................................... 18
5-1 Symbols and descriptions ................................................................................................................ 18
5-1-1 Safety related symbols ............................................................................................................ 18
5-1-2 Front panel .............................................................................................................................. 19
5-1-3 Touch screen ........................................................................................................................... 19
5-1-4 Rear panel ............................................................................................................................... 21
5-2 Front panel ....................................................................................................................................... 22
5-3 Rear panel ........................................................................................................................................ 24
5-4 Bottom panel .................................................................................................................................... 25
5-5 All screen ......................................................................................................................................... 25
5-6 Set screen ........................................................................................................................................ 26
5-7 Mode screen .................................................................................................................................... 27
5-8 Footswitch with two pedals .............................................................................................................. 28
5-9 Footswitch with one pedal (optional)................................................................................................ 28
5-10 Neutral electrode cable “P-cord” (optional) ...................................................................................... 29
6 CONNECTOR ............................................................................................................... 30
6-1 Docking Connector .......................................................................................................................... 30
6-2 Monopolar Standard 1 ...................................................................................................................... 30
6-3 Monoploar Standard 2 (Erbe) ........................................................................................................... 30
6-4 Bipolar Standard 3............................................................................................................................ 30
6-5 Monopolar Neutral Electrode ........................................................................................................... 31
6-6 Foot switch 1 (SIP/SOP) .................................................................................................................. 31
6-7 Foot switch 2 (SIP/SOP) .................................................................................................................. 32
7 SYSTEM DIAGRAM ..................................................................................................... 33
8 CLEANING, STORAGE AND DISPOSAL .................................................................... 34
8-1 Cleaning ........................................................................................................................................... 34
8-2 Storage ............................................................................................................................................. 35
8-3 Disposal of the unit .......................................................................................................................... 35
7.022.211 / ISSUE 5 9 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 10
ESG-400

1-1 Intended Use

The intended use depends on the approval of the country. Refer to the instructions for use of the electrosurgical generator.

1-2 Compatibility

This product can be used in combination with the products listed in compatibility table.

1-3 Expected service life

The expected service life is 10 years.

1 Outline

7.022.211 / ISSUE 5 10 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 11
ESG-400

2 Features

The ESG-400 is a reusable, non-sterile electrosurgical generator with different mono- and bipol ar cutting and coagulation modes. The maximum output power is 320 W.
On the front side it features a touch screen display that displays the connection status of accessories and peripherals connected to the electrosurgical generator. It is also used to show and modify the output settings (e.g. mode, output power, effect) as well as to control other functions (e.g. save settings).
In addition the ESG-400 has a bipolar socket, two monopolar sockets, a neutral electrode socket, and a universal socket to connect applicators with instrument recognition. The power switch turns the generator on and off.
Two contact quality indicators (one for split and one for non-split electrodes) are green illuminated if neutral electrodes are correctly connected. Three additional push buttons allow recalling a previously saved setting (Select Procedure), to assign the footswitches to specific output sockets (Footswitch), and to control several other functions (Menu), e.g. select language, touch tone control, output volume, or brightness.
On the rear panel the volum e control, a ventilation hole, the equipotential bonding point, the AC power socket, and two footswitch sockets can be found. Furthermore, for the connection of peripheral equipment 26-pin plugs respectively 14-pin plugs can be connected to the LINK-IN or to the LINK-OUT socket.
On the bottom panel, a docking socket is featured. It can be used to connect the ESG-400 directly to the USG-400 and upcoming devices . The ESG-400 is compatible with the new USG-400 ultrasonic generator to enable the use of combined (US + HF) instruments.

2-1 Application Modes

Monopolar Cut:
PureCut (Cutting of varying tissue structures; 3 Effects)
BlendCut (Cutting of varying tissue structures; 3 Effects)
PulseCut slow (Intermittent cutting; 3 Effects)
PulseCut fast (Intermittent cutting; 3 Effects)
Monopolar Coagulation:
SoftCoag (Coagulation of tissue with little sticking and carbonization; 3 Effects)
ForcedCoag (Fast and effective coagulation; 3 Effects)
SprayCoag (Contact-free surface coagulation with little penetration depth; 3 Effects)
PowerCoag (Fast and effective coagulation with increased dissection capabilities; 3 Effects)
7.022.211 / ISSUE 5 11 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 12
ESG-400
Bipolar Cut:
BipolarCut (All bipolar cutting procedures of tissue structures; 3 Effects)
SalineCut (Cutting in conductive fluid; 3 Effects; only available via UNIVERSAL socket)
PK PureCut (Cutting of varying tissue structures; 3 Effects; only available via UNIVERSAL socket)
PK SoftCut (Cutting of varying tissue stru cture s; 3 Effects; only available via UNIVERSAL socket)
PK LoopCut (Cutting of varying tissue structures, especially fibroid tissue; 3 Effects; on ly avai lab le via
UNIVERSAL socket)
PK MorceCut (Cutting of varying tissue structures, especially fibroid tissue; 3 Effects; only available via UNIVERSAL socket)
Bipolar Coagulation:
BiSoftCoag (Coagulation with little sticking and carbonization; 3 Effects)
AutoCoag (Coagulation with little sticking and carbonization; 3 Effects)
SalineCoag (Coagulation in conductive fluid; 3 Effects; only available via UNIVERSAL socket )
HardCoag (Controlled tissue coagulation; 3 Effects)
RFCoag (Controlled deep tissue coagulation; with and without RCAP)
FineCoag (Coagulation of tissue with little sticking and carbonization; 1 Effect)
PK Coag (Coagulation of tissue with little sticking and carbonization; 3 Effects)
PK SoftCoag (Coagulation of tissue with little sticking and carbonization; 3 Effects)
PK AutoCoag (Controlled tissue coagulation; 1 Effect)
The modes have preset power levels that may be customized by the user in a defined range.

2-2 Accessories

Footswitch Double Pedal (WB50402W): It has a blue pedal that is used to activate the selected coagulation mode and a yellow pedal that is used to activate the selected cutting mode.
Footswitch Single Pedal (optional; WB50403W): It has a blue pedal that is used to activate the selected coagulation mode
P-Cord (optional; MAJ-814): The P-cord is used to connect a patient plate to the ESG-400.
7.022.211 / ISSUE 5 12 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 13
ESG-400
Operation
Temperature
+ 10…+ 40°C
Relative humidity
30…85%, non-condensing
Atmospheric pressure
70…106 kPa
Transportation and
Temperature
- 25…+ 60°C
Relative humidity
10…85%, non-condensing
Atmospheric pressure
50…106 kPa

3 Limitations

(1) Use this product under the supervision of a doctor at a medical facility. (2) Do not use this product in combination with the products other than those designated by Olympus. (3) This product shoul be used, transported or stored in the following environment.
environment
storage environment
7.022.211 / ISSUE 5 13 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 14
ESG-400
Frequency
series from
Terminal for potential
Size, weight and
Volume
Weight of generator
Cardboard and expanded
Protection class according
Classification according to
Maximum high frequency
30 s

4 Specifications

4-1 ELECTROSURGICAL GENERATOR ESG-400 (REF: WB91051W)

Power supply Voltage range
Maximum input power
Power fuse
Power connection line
equalization
Width x Depth x Height
packaging
Weight of packaging
Type of packaging
Classification
to IEC 60601-1
100…120 V~ / 220…240 V~ 50 / 60 Hz 1500 VA 10 A (only FST-
Schurter) IEC 60320-1 / C13
Maximum length: 4.5 m Yes
370 × 465 × 156 mm 25752 cm³
12.5 kg
2.3 kg
polypropylene material CF, Class I
MDD 93/42/EEC
Output High frequency functions
High frequency
power
All modes
RFCoag (with or without RCAP)
IIb
Monopolar / Bipolar 430 kHz ±20%
320 W
25% duty cycle (e.g. 10 s activated / deactivated)
100% duty cycle
7.022.211 / ISSUE 5 14 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 15
ESG-400
Contact quality
Allowable resistance range for split type neutral
Allowable resistance range
split type neutral
USA, Canada and other
Kingdom and other
Sockets MONOPOLAR 1
MONOPOLAR 2
BIPOLAR
UNIVERSAL
Neutral electrode
monitor (CQM)
electrodes
for non­electrodes
3-pin ( 4 mm), Valleylab standard;
coaxial (∅ 8 mm), Bovie standard
3-pin ( 4 mm), V alley lab st andard;
coaxial (∅ 5 / 9 mm), Erbe standard
2-pin ( 4 mm, pin spacing 28.8 mm), V alley lab st andard; coaxial (
8 mm,
inner
4 mm),
outer
Erbe standard 7-pin, Olympus standard
Single or split, 10 mm pl ug
10...155 Ω ±15 Ω
< 10 Ω ±5 Ω

4-2 Power cords (4.5 m angled plug)

Power cords WA95621A
WA95622A
WA95623A
Many European countries Type E/F
countries Type B
United countries
Type G
7.022.211 / ISSUE 5 15 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 16
ESG-400
Protection class according
Size, weight and
Type of packaging
Protection class according
Size, weight and
Width x Depth x Height

4-3 Footswitch (REF: WB50402W, double pedal)

Classification
to IEC 60529
IPX8 (except the plug section)
Width x Depth x Height
packaging
Weight of footswitch
Length of cord
Weight of packaging
350 × 185 × 65 mm
1.9 kg 4 m
0.5 kg Cardboard material

4-4 Footswitch (REF: WB50403W, single pedal, optional)

Classification
to IEC 60529
packaging
Weight of footswitch
Length of cord
Weight of packaging
Type of packaging
IPX8 (except the plug section)
175 × 185 × 50 mm
1.6 kg 4 m
0.5 kg Cardboard material

4-5 Neutr al el ectrode cable “P-cord” (REF: MAJ-814, optional)

Size Weight
0.14 kg
Length of cord
3.1 m
7.022.211 / ISSUE 5 16 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 17
ESG-400

4-6 Communication cable 0.25 m (REF: MAJ-1871, optional)

Size Weight
0.05 kg
Length of cord
0.25 m

4-7 Communication cable 10 m (REF: MAJ-1872, optional)

Size Weight
Length of cord
0.5 kg
10 m

4-8 Adapter for UHI-2/3 (REF: MAJ-1873, optional)

Size Width x Depth x Height
Weight
Compatible cables
100 × 77 × 42 mm
0.35 kg
MAJ-1871, MAJ-1872
7.022.211 / ISSUE 5 17 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 18
ESG-400
Date of manufacture
CE marking
Non-ionizing electromagnetic radiation
Output insulated from earth
Caution, read
Defibrillation proof type CF applied
Waste electrical and electronic equipment
cTUVus marking
Manufacturer
Potential equalization terminal
Fuse rating
Type plate

5 Name and Function of each part

5-1 Symbols and descriptions

5-1-1 Safety related symbols

instructions Refer to
instructions
(connection for neutral electrode)
part (cardiac application)
7.022.211 / ISSUE 5 18 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 19
ESG-400

5-1-2 Front panel

Power on / off
Neutral electrode – non-split type
Neutral electrode – split type
Select procedure
Footswitch
Menu

5-1-3 Touch screen

BIPOLAR socket
MONOPOLAR 1 socket
UNIVERSAL socket
MONOPOLAR 2 socket
Double footswitch
Single footswitch
Autostart
Plus
Minus
Return
OK
Cancel
7.022.211 / ISSUE 5 19 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 20
ESG-400
Save procedure
Delete procedure
Languages
Touch tone on
Touch tone off
Software version
Safety test
Service
Volume
Brightness
Select procedure (in title line)
Menu (in title line)
Toggle
Previous
Next
Numeric
Alphabetic
7.022.211 / ISSUE 5 20 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Uppercase / lowercase
Backspace
Caution
Communication indicator
Page 21
ESG-400

5-1-4 Rear panel

RCAP
Reset
Resistance Controlled Automatic Power
Reference to BIPOLAR socket
Reference to MONOPOLAR 1 socket
Reference to UNIVERSAL socket
Reference to MONOPOLAR 2 socket
Volume
Footswitch
LINK-IN
LINK-OUT
LINK-IN socket
LINK-OUT socket
7.022.211 / ISSUE 5 21 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 22
ESG-400
. It is also used to show and m odify the output settings (e.g. mode, output power, effect) as
This button is used to open the “Footswitch screen” to assign one or two footswitch(es) or the autostart
This button is used to open the “Menu screen” to control several functions (save or delete a procedure,
8.
6.
1.
2.
3.
4.
7.
12.
11.
10.
9.
5.

5-2 Front panel

Power sw itch
1.
This switch turns the electrosurgical generator on and off.
BIPOLAR socket
2.
This socket connects the plug of a bipolar HF instrument (applied part).
MONOPOLAR 2 socket
3.
This socket connects the plug of a monopolar HF instrument ( appl ied p art) .
MONOPOLAR 1 socket
4.
This socket connects the plug of a monopolar HF instrument (applied part).
Touch-screen
5.
Displays the connection status of the accessories and peripherals connected to the electrosurgical generator well as to control other functions (e.g. save procedures, delete procedures).
FOOTSWITCH push button
6.
function to a specific output socket.
SELECT PROCEDURE push button
7.
This button is used to open the “Selec t Procedure screen” to recall saved settings. MENU push button
8.
control the touch tone, output volume and brightness as well as other functions).
7.022.211 / ISSUE 5 22 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 23
ESG-400
This indicator illuminates green if a split neutral electrode is connected and the contact resistance is
red if the split neutral electrode is not connected
or not applied properly (e.g. bad contact quality or partly dislocated) or no neutral electrode is
Contact quality monitor indicator for split neutral electrode
9.
within an acceptable range. The indicator illuminates
connected (in both cases the activation of monopolar output is disabled). Contact quality monitor indicator for non-split neut ral electrode
10.
This indicator illuminates green if a non-sp lit neutral electrode is connected.
Neutral electrode socket
11.
This socket connects the plug of a neutral electrode for monopolar application (applied part).
UNIVERSAL socket
12.
This socket connects the plug of an Olympus HF instrument with HF instrument recognition (applied part).
7.022.211 / ISSUE 5 23 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 24
ESG-400
Holes for air ventilation via a cooling fan; there are also ventilation holes on each side of the
oint is used for potential equalization. All equipment hous ings that
screen messages may depend on the
For a detailed explanation of the different types of
chapter 6 “Connection of neutral
electrode” and chapter 3.7 “Connection of HF
1.
2.
3.
4.
7.
6.
5.
1.

5-3 Rear panel

Footswitch socket s
1.
This socket connects the
Volume control
2.
This knob is used for adjusting the output volume. Ventilation hole
3.
plug of a single or double pedal footswitch.
electrosurgical generator.
Equipotential bonding point
4.
To increase electrical safety, this p come into contact with the patient are electrically connected in order to prevent low-frequency electrical currents from endangering the patient in the event of a defect in the conventi onal protective conductor system.
AC power socket
5.
This socket serves as a connection to the mains power supply via a power cord
LINK-OUT socket
6.
This socket connects the plug (14-pin) of a cable connected to peripheral equipment.
LINK-IN socket
7.
This socket connects the plug (26-pin) of a cable connected to peripheral equipment.
NOTE
The touch-
language setting of the electrosurgical generator.
sockets, refer to
instruments”.
7.022.211 / ISSUE 5 24 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 25
ESG-400
12.
This indicator shows the corresponding output socket where the same symbol is printed on the front This symbol indicates if the autostart function is assigned to the corresponding output socket. Blank if
1.
1.
2.
3.
4.
5.
8.
6.
11.
7.
9.
10.

5-4 Bottom panel

Docking socket
1.
This socket connects the plug (7-pin) of a docking connector to connect peripheral equipment. For more details, see chapter 1-6-1.

5-5 All screen

Reference to output sockets indicator
1.
panel.
Output socket name
2.
The name of the corresponding output socket is displ ay ed here. Autostart indicator
3.
7.022.211 / ISSUE 5 25 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 26
ESG-400
autostart or foots witch is not assigned. Refer to chapter 6.4, “Assign footswitch and autostart function”.
This symbol indicates if communication with peripheral equipment connected to the docking socket is
This symbol indicates if a connected double pedal footswitch is assigned to the corresponding output socket. Blank if autostart or footswitch is not assigned. Refer to chapter 6.4, “Assign footswitch and
The name of the output mode as selected in the “Mode sc reen” is displayed here. If “Off” is selected,
The number shows the effect as selected in the “Set screen”. For RFCoag mode the RCAP function
Each button covers the entire area including all output socket related information as described above (3. to 10.). Press the button, to switch to the corresponding “Set screen” to select the mode, power
This symbol indicates if a c onnected single pedal footswitch is assigned to the corresponding output
The name of the instrument or cable will be displayed instead of the output socket name
The name of the output mode as selected in the “Mode screen” is displayed here. Press this button to
1.
5.
2.
2
6.
3.
4.
Procedure name
4.
The name of the selected procedure is displayed here. Blank if no procedure is selected.
Communication indicator
5.
established. Footswitch indicator (double pedal)
6.
autostart function”.
Output mode
7.
“--“ will be displayed instead of power level and effect.
Output power level
8.
The number shows the output power level as selected in the “Set screen”. If an output power level is set to zero, “--” will be displayed instead of numbers.
Effect
9.
can be selected instead of an effect (refer to chapter 5.3, “O utput sett ing” ) .
Button area
10.
levels and effects for the corresponding output socket.
Footswitch indicator (single pedal)
11.
socket. Blank if autostart or footswitch is not assigned. Refer to chapter 6.4, “Assign footswitch and autostart function”.
UNIVERSAL / Instrument name
12.
“UNIVERSAL” if an instrument or cable with instrument recognition is connected to the UNIVERSAL socket.

5-6 Set screen

Mode button
1.
switch to the “Mode screen”. If “Off” is selected, “--“ will be displayed instead of power level and effect.
Plus button / Minus button
2.
These buttons increase / decrease the output power level.
7.022.211 / ISSUE 5 26 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 27
ESG-400
The number shows the selected effect. For RFCoag mode the RCAP function can be selected instead
These buttons allow the mode selection for a corresponding output socket as shown in the title line. If
Arrow button
Optional buttons to browse through the mode list. They are disabled if the number of available modes
1.
2.
3.
Toggle button
3.
This button switches to the next effect. Return button
4.
Press this button to save the settings and to return to the “All screen.” Output power level
5.
The number shows the selected output power level. If an output power level is set to zero, “--” will be displayed instead of numbers.
Effect
6.
of an effect (refer to chapter 5.3, “Output setting”)

5-7 Mode screen

Mode button
1.
a selection is already activated, this is indicated by a gray button. If no mode shall be selected, press the “Off button.”
Return button
2.
Press this button to return to the “Set screen.”
3.
fit to one screen.
7.022.211 / ISSUE 5 27 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 28
ESG-400
1.
2.
3.
1.
2.

5-8 Footswitch with two pedals

The footswitch with two pedals (Olympus REF: WB50402W) is included in deliv ery .
Cut pedal (yellow color)
1.
This pedal is used to activate the selected cutting mode.
Coagulation pedal (blue color)
2.
This pedal is used to activate the selected coagulation mode.
Footswitch plug
3.
Connects the footswitch with the electrosurgical generator on the rear panel.

5-9 Footswitch with one pedal (optional)

The footswitch with one pedal (Olympus REF: WB50403W) is an optional item which may be purchased separately.
Coagulation pedal (blue color)
1.
This pedal is used to activate the selected coagulation mode.
Footswitch plug
2.
Connects the footswitch with the electrosurgical generator on the rear panel.
7.022.211 / ISSUE 5 28 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 29
ESG-400
3.
Plug on the electrosurgical generator side
1.
2.
3.

5-10 Neutral electrode cable “P-cord” (optional)

The neutral electrode cable “P-cord” (Olympus REF: MAJ-814) is an optional item for the connection with a neutral electrode which may be purchased separately.
Lever-locking arm
1.
This arm secures the connector of the neutral electrode with the clamp. Clamp
2.
This clamp connects the neutral electrode to the “P-cord”.
This plug connects the “P-cord” to the electrosurgical generator.
7.022.211 / ISSUE 5 29 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 30
ESG-400
Pinning of connector 13 Docking Connector (bottom view) of ESG
1
2 4 3
1 Hand Cut 2 H 3 Active electrode 4 Active electrode
1 Hand Cut 2 Hand Coag 3 Active electrode 4 Cut+Coag+Active electrode (top)
1 2 3
4

6 Connector

6-1 Docking Connector

1) BNE – Bipolar Neutral Electrode
2) BAE – Bipolar Active Electrode
3) MAE – Monopolar Active Electrode
4) Common CD – Common ground for connection detection
5) MAPCAE – Monopolar Activ e Electrode
11) CD APC – Active pin for connection detection
12) CD NEO – Active pin for connection detection
- view of connector side
-400

6-2 Monopolar Standard 1

Type: 3 pin Valleylab, pin diameter = 4mm 1 pin BOVIE, pin diameter = 8 mm
Function: Monopolar output
Finger switch input (only for Valleylab: c ut and coag)

6-3 Monoploar Standard 2 (Erbe)

Type: 3 pin Valleylab, pin diameter = 4mm Coaxial ERBE, pin diameter = 5 mm (inner) and 9 mm (outer)
Function: Monopolar output
Finger switch input (cut and coag)

6-4 Bipolar Standard 3

Type: 2 pin socket, pin diameter = 4mm / pin distance 28.8 mm Coaxial socket, pin diameter = 4 mm (inner) and 8 mm (outer)
7.022.211 / ISSUE 5 30 / 110 Chapter 1: PRODUCT SPECIFICATIONs
and Coag
Page 31
ESG-400
1 Neutral electrode 2 Active and neutral electrode
1
2
Pin 3
Pin 2
Pin 4
Pin 5
Pin 6
Pin 1
B (COAG)
A (CUT)
(coding)
Pin 7 - NC
(coding)
Com
Function: Bipolar output

6-5 Monopolar Neutral Electrode

Type: 2 pins socket, Pin diameter = 2.5 mm, Pin distance = 10 mm Function: Monopolar output
CQM input
Principle sketch of connector 5 Neutral Elec trode

6-6 Foot switch 1 (SIP/SOP)

Type: Foot switch, 7-pol.
Pinning of foot switch connector
7.022.211 / ISSUE 5 31 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 32
ESG-400
Left Pedal
Right Pedal
L
Com
L/R
R
M

6-7 Foot switch 2 (SIP/SOP)

Type: Foot switch, 7-pol.
Activation detection of foot switch
7.022.211 / ISSUE 5 32 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 33
ESG-400

7 System Diagram

The recommended combinations of ancillary equipment and accessories that can be used with the electrosurgical generator are listed in the system chart below. In addition, new products released after the introduction of this product may also become compatible with this electrosurgical generator. For further details, contact Olympus.
WARNING
7.022.211 / ISSUE 5 33 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 34
ESG-400
After cleaning the electrosurgical unit, dry it thoroughly before storage or using
Patient debris and reprocessing chemicals are hazardous. During cleaning and disinfection, always wear appropriate personal protective equipment, such
perly so that your skin is not exposed. Always
When disconnecting plugs of instruments or power cords, always hold the
Never immerse the electrosurgical unit in water, clean or disinfect by
Do not clean the connectors or the alternating current power inlet. Cleaning
m or corrode the contacts, which could damage the
Do not wipe the external surface with hard or abrasive wiping material. The
The electrosurgical unit m ay be contaminated with infections materials, therefore, before servicing, perform the following cleaning procedures. For maintenance and storage of other items than those described below, refer to the respective instructions for use.

8-1 Cleaning

All surfaces of the unit’s housing can be cleaned and disinfected with the cleaning agents and surface disinfectants normally used for medical equipment (mild cleaning solution, e.g. 70 % isopropyl alcohol). No liquid must enter the connector or the unit duri ng cleaning.
1) Switch off the electrosurgical unit and disconnect the power cord from the grounded wall outlet.
2) If the equi pm en t and / or accessories are contaminated with blood or other potentially infectious materials, first wipe off all gross debris using neutral detergent, and then wipe its surface with a lint-free cloth moistened with a surface disinfectant.
3) To remove dust, dirt and non-patient debris, wipe the electrosurgical unit and footswitch using a soft, lint-free cloth moistened with 70 % ethyl or isopropyl alcohol.

8 Cleaning, sto rage and disposal

WARNING
CAUTION
it again. If it is used while still wet, there is a risk of electric shock.
as eye wear, face mask, moisture-resistant clothing and chemical-resistant waterproof gloves that fit pro remove contaminated protective clothing before leaving the reprocessing area.
plug. Pulling the cable m ay result in damaging of the wires.
immersion, gas sterilization or autoclaving. It may cause equipment damage.
them can defor electrosurgical unit.
surface will be scratched.
7.022.211 / ISSUE 5 34 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 35
ESG-400
Do not store the electrosurgical unit in a location exposed to direct sunlight,
adiation (e.g. near
wave medical treatment
equipment, magnetic resonance imaging equipment, radio or mobile phones).

8-2 Storage

Before storage of the electrosurgical unit, disconnect the power cord and store it properly according to the environmental conditi on s descr ibed in cha pter 1.4 (Technical dat a).
CAUTION
x-rays, radioactivity, liquids or strong electromagnetic r microwave medical treatment equipment, short-
Damage to the electrosurgical unit may result.

8-3 Disposal of the unit

When disposing of this electrosurgical unit, or any of its components (such as fuses), follow all applicable national and local laws and guidelines.
Waste electrical and electronic equipment
In accordance with European Directive 2002/96/EC on waste electrical and elec tronic equipment (WEE E), the product must not be disposed of as unsorted municipal waste, but s hould be collected separately.
Refer to Olympus for return and / or collection systems available in your country.
7.022.211 / ISSUE 5 35 / 110 Chapter 1: PRODUCT SPECIFICATIONs
Page 36
ESG-400
7.022.211 / ISSUE 5 36 / 110 Chapter 2: Block Description
Page 37
ESG-400

CHAPTER 2: BLOCK DESC RIPTION

1 BLOCK DESCRIPTIONS .............................................................................................. 38
1-1 Motherboard ..................................................................................................................................... 39
1-2 HVPS Board ..................................................................................................................................... 41
1-3 Generator board ............................................................................................................................... 41
1-4 Relay Board ..................................................................................................................................... 41
1-5 Front Panel ....................................................................................................................................... 42
7.022.211 / ISSUE 5 37 / 110 Chapter 2: Block Description
Page 38
ESG-400
Fig. 2.1.1. Block descriptions

1 Block Descriptions

7.022.211 / ISSUE 5 38 / 110 Chapter 2: Block Description
Page 39
ESG-400

1-1 Motherboard

Due to the containing embedded PC the Motherboard is the central unit of the ESG-400. The Motherboard controls the Relay Board, the HVPS Board and the Generator Board. It contains all input and output interfaces to the user as well as to other medical devices or computers. Additionally func tionali ties off the board are the low voltage supplies for the complete unit, the mains input including filters and the measuring part of the voltage line selection circuit for switching between 115 and 230 VAC.
Overview:
Embedded PC incl. periphery
Embedded PC with MPC5200 controller (incl. address and data bus, chip selects,
interrupt inputs, I 2C, SPI, in-/output ports, uarts, timer)
Watchdog circuit
Chip select decoder
Hardware reset
JTAG interface
Real time clock
POF interface for the spark monitor
Digital input and output circuits
D/A converters for controlling the HVPS
A/D converters for measuring different signals from Relay, HVPS and Generator, temperatures and
watching on important voltages
Connections/Interfaces:
To the PCBs Relay, HVPS and Generator
Push buttons for user inputs on the front panel
Volumeboard for changing the speaker volume
Power Indicator shows power-on of the unit on the front
CQM Indicator shows status of CQM on the front
Controlling and driving the main housing fan
Audio circuit incl. D/A converter and amplifier for sound
Graphic controller with driver and backlight for the front display
Touch controller for the touch display
Ethernet controller and connec tor for external connections
RS-232 with connector for external connections
USB host with transceiver and connector for external connections
FlexRay controller, transceiver and connector for external connections
Connectors for foots witch incl. detection and analysis circuit
Connectors for handswitches incl. activation detection circuits
Instrument recognition circuit for instruments connected to the universal socket
Detection circuit for devices connec ted to the dock ing co nne c tor
7.022.211 / ISSUE 5 39 / 110 Chapter 2: Block Description
Page 40
ESG-400
Low voltage supplies
Switching regulators for -12 VDC, +5 VDC and +3,3 VDC (5 V and 3,3 V c ascaded)
DC/DC converters for isolated +12 VDC and +5 VDC SIP/SOP voltages
Batteries for a permanent +3 V voltage for RTC and SRAM
Reference voltage of 8,192 V
Mains input
Input filters
Inrush current limiter
Mains voltage measurement and output signal for a selection circuit on the HVPS
7.022.211 / ISSUE 5 40 / 110 Chapter 2: Block Description
Page 41
ESG-400

1-2 HVPS Board

The high voltage power supply (HVPS) is a switching mode power supply with series resonance circuit. It provides a high DC voltage for the HF Generator. It contains:
voltage line selection circuit, f o r automatic change between 110/230 VAC
simple rectifier circuit
PWM driving circuit
driving circuits including a digital flip-flop stage for complete cycle driving
power FET half-bridge, a series resonance circuit, output transformer and rectifying stage
current and voltage monitors
discharge circuit

1-3 Generator board

The Generator Board generates the HF output energy from a DC input voltage and contains:
control circuit for generating st art and driv in g puls es of “one cy cle” sin us oscil lato r
driving stage for power FET, parallel resonance circuit and series resonance circuit
relays for switching between different transformer windings
HF output voltage monitor and redundant HF voltage monitor
HF output current monitor and redundant HF current monitor
HF output phase monitor
HF leakage current monitor
spark monitor (SPM) supply circuit
spark monitor for detecting positive and negative DC voltage offset

1-4 Relay Board

The Relay board is used to connect the active output socket to the generator board. It contains:
connectors to every single HF output socket
relays which are separating the non active output terminals from the active output terminals
separating relays are forced guided relays with read-back contact in secondary circuit to control the relay
status
contact quality monitor (CQM)
transient voltage suppression (TVS) diodes in applied part
7.022.211 / ISSUE 5 41 / 110 Chapter 2: Block Description
Page 42
ESG-400

1-5 Front Panel

The Front Panel is the main part of the user interface. It contains:
LCD touch screen
Push Buttons
Contact Quality Monitor
BIPOLAR socket
MONOPOLAR 1 socket Valleylab & Bovie
MONOPOLAR 2 socket Valleylab & Erbe
UNIVERSAL socket
Socket for neutral electrode
7.022.211 / ISSUE 5 42 / 110 Chapter 2: Block Description
Page 43
ESG-400

CHAPTER 3: REPAIR SYSTEM

1 ESG-400 MAIN UNIT .................................................................................................... 44
2 BOARD COMPATIBILITY ............................................................................................. 44
3 OPTIONAL A CCESSORIES ......................................................................................... 44
3-1 WB50402W (Footswitch with two pedals) ....................................................................................... 44
3-2 WB50403W (Footswitch with one pedal) ......................................................................................... 44
3-3 MAJ-814 (Neutral electrode cable “P-cord”) .................................................................................... 44
4 PRECAUTIONS ON FUNCTION AND OPERATION SETTINGS ................................. 44
4-1 General Precautions ........................................................................................................................ 44
7.022.211 / ISSUE 5 43 / 110 Chapter 3: Repair System
Page 44
ESG-400

1 ESG-400 Main Unit

(1) In general, the main unit mus t be shipped to a service center in the event of a malfunction. (2) Indiv id ual units can be replaced.

2 Board Compatibility

The compatibility of boards and components is dependent on the hardware version of the generator. The hardware version can be indentified by the serial number of the generator.
The serial number starting with 5 numbers, hardware version with WXX and followed by 3 numbers after the hyphen.
Example: XXXXXWYY-ZZZ, WYY will show the hardware version.

3 Optional Accessories

3-1 WB50402W (Footswitch with two pedals)

Supplied as a spare part subject to repair services in the event of a malfunction.

3-2 WB50403W (Footswitch with one pedal)

Supplied as a spare part subject to repair services in the event of a malfunction.

3-3 MAJ-814 (Neutral electrode cable “P-cord”)

Supplied as a spare part subject to repair services in the event of a malfunction.

4 Precautions on Function and Operation Settings

4-1 General Precautions

Before repair, it is generally advisable to record the function and operation settings as the basis for restoring these settings after service. If the original settings cannot be known due to mechanical problems present at the time the unit was accepted for repair, apply the f actory-set values or the safest settings (such as the lowest output levels). In this case, inform the user that the settings have been changed.
7.022.211 / ISSUE 5 44 / 110 Chapter 3: Repair System
Page 45
ESG-400

CHAPTER 4: TROUBLESHOOTING

1 GENERAL ..................................................................................................................... 46
2 NEUTRAL ELECTRODE OPERATION ........................................................................ 47
3 ERROR SCREEN, CODES AND MEASURES ............................................................. 48
3-1 What to do when no error code is displayed .................................................................................... 50
3-2 What to do when an error code is displayed .................................................................................... 54
7.022.211 / ISSUE 5 45 / 110 Chapter 4: Troubleshooting
Page 46
ESG-400

1 General

If the electrosurgical unit has visible damage, do not use the electrosurgical unit and contact the legal manufacturer. If the unit is not functioning properly, use the information in this chapter to identify and correct the malfunction. If the problem cannot be resolved by the described remedial action, stop using the electrosurgical unit and contact the legal manufacturer for repair.
DANGER
CAUTION
CAUTION
Never use the electrosurgical unit if an abnorm ali ty is suspected.
Repairs must only be carried out by Olympus or a firm authorized by Olympus.
Preventive maintenance (inspection / periodic safety check) must only be carried out by a qualified person / technician.
7.022.211 / ISSUE 5 46 / 110 Chapter 4: Troubleshooting
Page 47
ESG-400

2 Neutral electrode operation

Check the following table, to identify or correct failures regarding the neutral elec trode operation.
Contact quality monitor Mode Indication Bipolar application Standby and activation
Monopolar application
A non-spli t neutral electrode is connected. Activation is possible. Contact quality monitor detects connection of neutral elec trode.
If a split neutral electrode is connected, it has a short circuit. Immediately replace the neutral electrode!
A split neutral electrode is connected. Activation is possible. Contact quality monitor detects connectio n of neutral electrode and contac t to patients’ skin.
During standby: A split or a non-split neutral electrode is not connected or a split neutral electrode detaches.
Activation is disabled. During activation: A split or a non-split
neutral electrode has disconnected or a split neutral electrode detaches. The activation is stopped.
A neutral electrode is not required. Contact quality monitor indicator for split neutral electrode illuminates red.
Contact quality monitor indicator for non-split neutral electrode illuminates green.
Contact quality monitor indicator for split neutral electrode illuminates green.
Contact quality monitor indicator for split neutral electrode illuminates red.
During activation an alarm signal can be heard and the touch-screen will display an error window (E202).
Legend: Red illumination of the indicator
Green illumination of the indic ator
7.022.211 / ISSUE 5 47 / 110 Chapter 4: Troubleshooting
Page 48
ESG-400
tton is not available in software version lower
Error window
Fig. 4.3.1. Error Screen
Error title
Error code
Remedial actions
Caution symbol

3 Error screen, co des and measures

Follow the troubleshooting advices in this chapter, to identify or correct failures. The error window is configured as shown in figure below.
(Example: E002 Short circuit)
NOTE
If an error occurs (see Fig. 4.3.1): An error window will appear and an alarm signal is audible.
The OK bu than 4.00-A
A short message with the error code, error title and a description of the remedial action will be displayed.  The error code consists of an error number shown under the “caution” symbol.  Depending on the error priority, the condition of the audible signal and the “caution” symbol are different
(see Table 4.1).
Proceed with the described remedial action.  The error window disappears after a few seconds, if the error is cleared.  If the error window is still displayed, the error is not cleared. Proceed with the next remedial action if
available.
Error category Error condition priority Indicator (“caution”)
symbol condition
High priority Immediate user response
is required
Medium priority Prompt user response
is required
Low priority Awareness of the user
is required
Table 4.1: Error priorities and the corresponding indicator symbol condition
Flashes in red color
Flashes in yellow color
Constant on in yellow color
7.022.211 / ISSUE 5 48 / 110 Chapter 4: Troubleshooting
Page 49
ESG-400
The electrosurgical generator is equipped with an intelligent
multiple variables. Depending on the risk potential, alarms are classified in “high priority”, “medium priority” and “low priority” alarms. An alarm of higher priority overrides an existing alarm of
er priority. If more than one alarm situation of equal priority is
NOTE
alarm system which determines alarm conditions on the base of
low determined, the one that occurred first is displayed only. This electrosurgical generator complies with the IEC 60601-1-8: 2006.
7.022.211 / ISSUE 5 49 / 110 Chapter 4: Troubleshooting
Page 50
ESG-400

3-1 What to do when no error code is displayed

Perform the indicated remedial actions below. If the problem cannot be resolved by the described remedial action, contact the legal manuf actur er.
Situation Possible cause Remedial action The electrosurgical
generator does not respond after pressing the power switch.
The touch-screen remains dark after switching the electrosurgical generator on (sound is audible after switching on).
The touch-screen cannot be controlled.
Improper connection of the power cord to the AC power socket on the rear panel of the electrosurgical generator or to the grounded wall outlet.
The grounded wall outlet has wrong or not output voltage.
The power cord is damaged. Check the power cord for damages and, if
Malfunction of the electrosurgical generator.
Malfunction of the touch-screen. Contact the legal manufacturer.
An object is in contact with the touch-screen.
The touch-screen is not properly calibrated.
Check the power cord and the grounded wall outlet for correct connection.
Check the grounded wall outlet or use an alternative grounded wall outlet.
necessary, replace the power cord. Contact the legal manufacturer.
Remove the object.
Contact the legal manufactur e r.
The electrosurgical generator does not react when a push button on the front panel is pressed during standby.
The electrosurgical generator does not react when a (push) button on the front panel is pressed during activation.
Malfunction of the touch-screen. Stop using the electrosurgical generator a nd
press the power switch to turn off the electrosurgical generator. Contact the legal
manufacturer. A push button is already pressed. Release the push button. Malfunction of the electrosurgical
generator.
The (push) buttons are not available during activation.
Contact the legal manufactur e r.
Release the footswitch or hand switch to stop
the activation.
7.022.211 / ISSUE 5 50 / 110 Chapter 4: Troubleshooting
Page 51
ESG-400
Situation Possible cause Remedial action No sound is audible
during activation.
The volume can not be adjusted via the volume control within the “Menu screen” or at the rear panel.
The electrosurgical generator does not respond to footswitch or handswitch activation.
The volume is set to an inaudible level (e.g. due to high environmental noise).
Malfunction of the electrosurgical generator.
The volume of the error-related audible signal is not adjustable.
Malfunction of the electrosurgical generator.
Improper connection of the footswitch to one of the footswitch sockets on the rear panel of the electrosurgical generator or the HF instrument to the output socket.
The footswitch or the handswitch of the HF instrument and / or the connection cable are damaged.
Increase the volume either on the touch-screen
within the “Menu screen” or use the volume
control on the rear panel of the electrosurgical
generator.
Stop using the electr o surg ica l gener at or and
press the power switch to turn off the
electrosurgical generator. Contact the legal
manufacturer.
No action required.
Contact the legal manufactur e r.
Check the footswitch and the HF instrum ent f or
correct connection.
Check the footswitch or the handswitch of the
HF instrument and / or the connection cable for
damages and, if necessary, replace the
footswitch, the HF instrument or the connection
cable. The incorrect footswitch peda l or
handswitch button is pressed. The electrosurgical generator is not
switched on. Another footswitch pedal or handswitch
button of the HF instrument is pressed.
The output is activated by the peripheral equipment.
A window is displayed on the touch-screen.
The “All screen” or “Set screen is not displayed on the touch-screen.
The corresponding output mode has been deactivated in the “Mode screen” (“Off” is displayed) or the power level has been set to “--“.
Press the correct footswitch pedal or
handswitch button of the HF instrument.
Switch on the electrosurgical generator with the
power switch.
To activate the intended output, release the
current pressed footswitch pedal or handswitch
button of the HF instrument.
If the output of the peripheral equipment is
activated, the output of the electrosurgical
generator cannot be activated simultaneously.
Stop using the peri phera l equipment.
Press the “OK button” or “Cancel button” to
close the window or wait until the window
disappears automatically after a few seconds.
Return to the “All screen” or “Set screen”.
Select an output mode in the “Mode screen” or
increase the power level via the “Set screen”
(refer to chapter 5.3, “Output setting”).
Malfunction of the electrosurgical generator.
If the autostart function is selected, the electrosurgical
7.022.211 / ISSUE 5 51 / 110 Chapter 4: Troubleshooting
The autostart function is assigned to another output socket.
Contact the legal manufactur e r.
Check the correct assignment of the autostart
function (refer to chapter 6.4, “Assign foot switch
and autostart function”).
Page 52
ESG-400
generator does not
Situation Possible cause Remedial action
activate the output when the electrode has contact with the tissue.
If an HF instrument is connected to the UNIVERSAL socket, the electrosurgical generator does not recognize the connected HF instrument.
Footswitch or handswitch of the HF instrument is pressed and activation sound is audible but no output power is delivered.
Long time delay of the autostart function has been selected in the “Autostart screen”.
Malfunction of the electrosurgical generator.
Improper connection of the Olympus HF instrument plug with the UNIVERSAL socket on the front panel of the electrosurgical generator.
The HF instrument does not support Olympus HF instrument recognition.
The Olympus HF instrument and / or the connection cable is damaged.
Malfunction of the electrosurgical generator.
The footswitch is ass igned to another output socket.
Improper connection of the HF instrument plug with the output socket on the front panel of the electrosurgical generator.
Set a shorter time delay in the “Autostart
screen” (refer to chapter 6.5, “Menu - Autostart
setup”).
Contact the legal manufactur e r.
Check the Olympus HF instrument plug for
correct connection.
Confirm the use of an Olympus HF instrument
with HF instrument recognition capabilities.
Replace the Olympus HF instrument and / or
the connection cable.
Contact the legal manufactur e r.
Check the correct assignment of the foot sw itch
(refer to chapter 6.4, “Assign footswitch and
autostart function”).
Check the HF instrument plug for correct
connection.
Malfunction of the electrosurgical generator.
Contact the legal manufactur e r.
7.022.211 / ISSUE 5 52 / 110 Chapter 4: Troubleshooting
Page 53
ESG-400
Situation Possible cause Remedial action No output power is
delivered when RFCoag mode with or without RCAP is selected and end of activation signal is audible.
The output of the electrosurgical generator cannot be deactivated.
The electrosurgical generator cannot be switched off.
The electrode has no contact with the tissue.
Improper connection of the HF instrument plug with the output socket on the front panel of the electrosurgical generator.
Damaged HF instrument connecti on cable.
Malfunction of the electrosurgical generator.
The autostart function is selected to the current used output sock et and both electrodes touch the tis sue.
Malfunction of the footswitch or handswitch.
Malfunction of the electrosurgical generator.
Malfunction of the electrosurgical generator.
Check that the electrode has contact with the
tissue.
Check the HF instrument plug for correct
connection.
Replace the HF instrument connection cable.
Contact the legal manufacturer.
Remove the electrode from the tissue.
Immediately switch off the electrosurgical
generator by pressing the power switch.
Replace the footswitch or HF instrument with
handswitch.
Contact the legal manufactur e r.
Disconnect the power cord plug from the AC
power socket on the rear panel of the
electrosurgical generator or from the grounded
wall outlet. Contact the legal manufacturer.
Automatic mist & smoke evacuation system/function does not work.
The settings are erroneous. Correct the settings of the compatib le high flow
insufflation unit. The communication cable i s not
connected.
The connection of the communication cable is erroneous.
Compatible high flow insufflation unit malfunction.
Connect the communication cable. Refer to 3.5,
“Automatic mist & smoke evacuation
system/function (when usi ng the compatible
high flow insufflation unit)”.
Reconnect the communication cable. Refer to
Section 3.5, “Automatic mist & smoke
evacuation system/function (when using the
compatible high flow insufflation unit)”.
Contact Olympus.
7.022.211 / ISSUE 5 53 / 110 Chapter 4: Troubleshooting
Page 54
ESG-400

3-2 What to do when an error code is displayed

If an error code is displayed, perform the indicated remedial actions below. If the problem cannot be resolved by the described remedial action, contact the legal manufacturer.
The error messages frequently used are translated in table 4.2.
Description in Error Code Table Actual messages displayed on ESG-400
“Auto-restart”
“Contact OLY” If the problem persists, contact Olympus.
“Release FSW”
“Release HSW”
“Replace Instrument”
“Reconnect Instrument”
“Cable Connection”
“Cable Damage”
Table 4.2: frequently used error messages
Error no.
E001
NOTE
Error message Possible cause Remedial action
Open circuit
Check if the electrodes of the instrument have proper tissue contact.
The ESG-400 will be restarted automatic ally when the error with the message
of “Auto-restart” is occurred.
Electrodes of the HF instrument may have no proper tissue contact.
Malfunction of the HF instrument and / or the connection cable.
ESG-400 will automatically restart.
Release the footswitch pedal to continue. Release the handswitch of the instrument to contin ue. If the problem persists, replace the instrument. Reconnect the instrument to the UNIVERSAL socket. Check all communication c ables are connected correctly. Check all cables for damage. If necessary,
Ensure that the electrodes of the HF instrument have proper tissue contact.
Replace the HF instrument and / or the connection cable.
replace the cables.
E002
E003 E004
E006
Short circuit
Ensure the instrument electrodes do not come into contact with each other and proceed.
--­Error
„Auto-restart” „Contact OLY”
Non-conductive fluid
Ensure conductive fluid is used for bipolar resection.
Electrodes of the HF instrument may touch each other.
Malfunction of the HF instrument and / or the connection cable.
DockingConnectorError --­Internal software error. Send back to Olympus Service
Use of non-conductive fluid during a bipolar cutting procedure. The active and / or neutral electrode is within an air environment.
The bipolar Olympus HF instrument has not been properly connected to the UNIVERSAL socket or damaged
Ensure that the electrodes of the HF instrument do not touch each other.
Replace the HF instrument and / or the connection cable.
Ensure that conductive fluid is used during bipolar resection procedure. Always immerse the active and / or neutral electrode within the conductive fluid.
Check the connection of the bipolar Olympus HF instrument to the connection cable and the connection of the
7.022.211 / ISSUE 5 54 / 110 Chapter 4: Troubleshooting
Page 55
ESG-400
connection cable.
connection cable to the UNIVERSAL
E007
E013 E016
E017
Wrong or broken Footswitch,
E018
E019
E020
E023
E024
Error no.
[…] E011
E012
E014 E015
Error message Possible cause Remedial action
socket and / or replace the connection cable.
Error
„Auto-restart” „Contact OLY”
Adjustment incomplete „Contact OLY” Error
„Auto-restart” „Contact OLY”
Burn-in incomplete
„Contact OLY”
The electrode might be contaminated and encrusted.
Adjustment missing or incomplete. Complete the adjustment of the device.
Burn-in missing or incomplete. Complete the burn-in of the dev ice
Check the electrodes for conta minat io n and encrustation before use and, if necessary, clean the electrodes.
Refer to E004
Refer to E004
E021 E022
Error „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Footswitch combination
Connection of only one single pedal and/or only one double pedal footswitch is allowed.
„Contact OLY”
Error
„Contact OLY”
Invalid serial number
„Contact OLY”
damaged hardware.
Two single pedal or two double pedal footswitches have been connected.
Internal hardware error. Send back to Olympus Service
Wrong or missing serial number of the device.
1. Check the Footswitch connection.
2. Change the Footswitch.
3. Send back to Olympus Service
Refer to E004
Ensure that only one single pedal and/or only one double pedal footswitch are connected.
Send back to factory.
[…] E030
7.022.211 / ISSUE 5 55 / 110 Chapter 4: Troubleshooting
Error
„Auto-restart” „Contact OLY”
Refer to E004
Page 56
ESG-400
E031
E032
Wrong docking connector cod i ng E033
E035
E038
E039
E045
E48
Communication error with USG-400 E049
E052
Communication error with USG-400
Error no.
E034
E036 E037
Error message Possible cause Remedial action
Error
„Auto-restart” „Contact OLY”
Error „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error
Wrong docking connector cod i ng signal, broken hardware.
signal, broken hardware.
Internal hardware error. Send back to Olympus Ser vice
1. Check the docking connection to the USG-400.
2. Send back to Olympus Service
1. Check the docking connection to the USG-400.
2. Send back to Olympus Service
Refer to E020
Refer to E004
[…] E043
[…] E047
[…] E051
„Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error
„Contact OLY”
Refer to E004
Communication error with USG-400 or other devices.
or other devices.
Communication error with USG-400 or other devices.
1. Check the FlexRay connections.
2. Check if all other devices are working properly.
3. Send back to Olympus Service
1. Check the FlexRay connections.
2. Check if all other devices are working properly.
3. Send back to Olympus Service
1. Check the FlexRay connections.
2. Check if all other devices are working properly.
Error
„Auto-restart” „Contact OLY”
7.022.211 / ISSUE 5 56 / 110 Chapter 4: Troubleshooting
or other devices.
3. Send back to Olympus Service
1. Check the FlexRay connections.
2. Check if all other devices are working properly.
3. Send back to Olympus Service
Page 57
ESG-400
E053
Communication error with USG-400
E054
1. Check the FlexRay connections.
E055
Communication error with USG-400
E056
1. Check the FlexRay connections.
E058
E061 E062
E063
E067 E068
E074
E078
Error no.
E057
Error message Possible cause Remedial action
Error
„Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error „Contact OLY”
Error
„Auto-restart” „Contact OLY”
or other devices.
2. Check if all other devices are working properly.
3. Update the software.
or other devices.
2. Check if all other devices are working properly.
3. Update the software.
1. Check the FlexRay c onnections.
2. Check if all other devices are working properly.
3. Send back to Olympus Service
1. Check the FlexRay connections.
2. Check if all other devices are working properly.
3. Send back to Olympus Service
1. Check the FlexRay connections.
2. Check if all other devices are working properly.
3. Send back to Olympus Service
1. Check the FlexRay connections.
2. Check if all other devices are working properly.
3. Send back to Olympus Service
E059 E060
[…] E066
[…] E073
E075
Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error „Contact OLY” Error „Contact OLY” Error
„Auto-restart” „Contact OLY”
Error „Contact OLY” Error
„Auto-restart”
Internal software or hardwar e error. Send back to Olympus Service
Refer to E004
Refer to E004
Refer to E004
Refer to E045
Refer to E045
Refer to E045
[…]
7.022.211 / ISSUE 5 57 / 110 Chapter 4: Troubleshooting
„Contact OLY”
Error
Refer to E045
Page 58
ESG-400
E080
E081
E082
E085
Send back to Olympus Service
E100
E102
E103
E104
E105
E106
E107
Error no.
E083 E084
[…] E099
E101
Error message Possible cause Remedial action
„Contact OLY” Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart”
Internal hardware error.
Spark Monitor communication error. Send back to Olympus Service
Refer to E045
Refer to E004
Refer to E085
„Contact OLY”
Push button pressed
Release the push button to continue.
„Contact OLY”
Footswitch pedal pressed
„Release FSW” „Contact OLY”
Footswitch pedal pressed
„Release FSW” „Contact OLY”
Footswitch pedal pressed
„Release FSW” „Contact OLY”
A push button on the Front Panel is pressed while switching on.
A pressed Cut Pedal on Footswitch 1 has been detected during power-up.
Malfunction of the footswitch. Change the Footswitch 1.
A pressed Coag Pedal on Footswitch 1 has been detected during power-up.
Malfunction of the footswitch. Change the Footswitch 1.
A pressed Cut Pedal on Footswitch 2 has been detected during power-up.
Malfunction of the footswitch. Change the Footswitch 2.
Release the push button.
Release the Cut Pedal on Footswitch 1 and restart the device.
Release the Coag Pedal on Footswitch 1 and restart the device.
Release the Cut Pedal on Footswitch 2 and restart the device.
Footswitch pedal pressed
„Release FSW” „Contact OLY”
7.022.211 / ISSUE 5 58 / 110 Chapter 4: Troubleshooting
A pressed Coag Pedal on Footswitch 2 has been detected during power-up.
Malfunction of the footswitch. Change the Footswitch 2.
Release the Coag Pedal on Footswitch 2 and restart the device.
Page 59
ESG-400
E108
E109
E110
E111
E112
E113 E114
E115
E116 E120
Error no.
Error message Possible cause Remedial action
Handswitch pressed
„Release HSW” „Replace Instrument” „Contact OLY”
Handswitch pressed
„Release HSW” „Replace Instrument” „Contact OLY”
Handswitch pressed
„Release HSW” „Replace Instrument” „Contact OLY”
Handswitch pressed
„Release HSW” „Replace Instrument” „Contact OLY”
A pressed Cut Handswitch at the Monopolar 1 socket has been detected during power-up.
Malfunction of the HF instrument. Replace the HF instrument.
A pressed Coag Handswitch at the Monopolar 1 socket has been detected during power-up.
Malfunction of the HF instrument. Replace the HF instrument.
A pressed Cut Handswitch at the Monopolar 2 socket has been detected during power-up.
Malfunction of the HF instrument. Replace the HF instrument.
A pressed Coag Handswitch at the Monopolar 2 socket has been detected during power-up.
Malfunction of the HF instrument. Replace the HF instrument.
Release the Cut Handswitch at the Monopolar 1 socket and restart the device.
Release the Coag Handswitch at the Monopolar 1 socket and restart the device.
Release the Cut Handswitch at the Monopolar 2 socket and restart the device.
Release the Coag Handswitch at the Monopolar 2 socket and restart the device.
Handswitch pressed
„Release HSW” „Replace Instrument” „Contact OLY”
Handswitch pressed
„Release HSW” „Replace Instrument” „Contact OLY”
Touch-screen pressed
Do not touch the screen. „Contact OLY”
Application time limit exceeded
Release the footswitch or handswitch and reactivate to continue.
„Contact OLY”
A pressed Cut Handswitch at the Multifunction socket has been detected during power-up.
Malfunction of the HF instrument. Replace the HF instrument.
A pressed Coag Handswitch at the Multifunction socket has been detected during power-up.
Malfunction of the HF instrument. Replace the HF instrument.
The screen is touched while switching on.
The maximum time limit for the application has been exceeded.
Release the Cut Handswitch at the Multifunction socket and restart the device.
Release the Coag Handswitch at the Multifunction socket and restart the device.
Release the finger from the screen.
Release the footswitch or handswitch for about 15 sec. and activate again by repressing the footswitch or handswitch.
[…] E119
E121
7.022.211 / ISSUE 5 59 / 110 Chapter 4: Troubleshooting
Error
„Auto-restart” „Contact OLY”
Error
„Contact OLY”
Refer to E004
Refer to E004
Page 60
ESG-400
E122
E131
E132
Internal hardware error.
E133
Internal hardware error. E134
Internal hardware error.
Send back to Olympus Service
E135
E136
E137 E138
E139
Error no.
[…] E130
Error message Possible cause Remedial action
Error
„Auto-restart” „Contact OLY”
Unknown instrument
„Reconnect Instrument” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error
An Olympus HF instrument has not been properly connected to the UNIVERSAL socket.
Invalid default mode detected at the instrument connected to the Multifunction socket.
Refer to E004
Check the connection of the Olympus HF instrument to the UNIVERSAL socket.
Change the Instrument at the Multifunction socket
Send back to Olympus Service
Send back to Olympus Service
„Contact OLY”
Single pedal footswitch not assigned
Single pedal footswitch not assigned
„Contact OLY”
Double pedal footswitch not assigned
Double pedal footswitch not assigned
„Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error
„Contact OLY”
The single pedal footswitch has not been assigned to the corresponding output socket.
The double pedal footswitch has not assigned to the corresponding output socket.
The Auto Start feature is not assigned correctly.
Invalid Thunder Beat mode sett ing detected.
Assign the single pedal footswitch.
Assign the double pedal footswitch.
Reassign the Auto Start feature.
1. Check the Thunder Beat instrument connected to the USG-400.
2. Check the USG-400.
Error
„Contact OLY”
7.022.211 / ISSUE 5 60 / 110 Chapter 4: Troubleshooting
Invalid Thunder Beat seal mode setting detected.
1. Check the Thunder Beat instrument connected to the USG-400.
2. Check the USG-400.
Page 61
ESG-400
E140
No mode has been selected while
Select a mode via the “Mode screen”. E141
E142
E147 E148
E149 E150
E151
E152
Error no.
[…] E145
Error message Possible cause Remedial action
No mode selected
No mode selected Select a mode „Contact OLY”
Power set to zero (--)
Set a valid power level „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Unknown instrument
„Reconnect Instrument” „Replace Instrument” „Contact OLY”
activating.
The power level for the c hosen mode is set to zero.
An Olympus HF instrument has not been properly connected to the UNIVERSAL socket.
The instrument connected to the Multifunction socket can not be determined.
Increase the power level via the “Set screen”.
Refer to E004
Check the connection of the Olympus HF instrument to the UNIVERSAL socket.
Change the connected instrument at the Multifunction socket,
Unknown instrument
„Reconnect Instrument” „Replace Instrument” „Contact OLY”
Unknown instrument
„Reconnect Instrument” „Replace Instrument” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Unknown instrument
„Reconnect Instrument” „Replace Instrument” „Contact OLY”
An Olympus HF instrument has not been properly connected to the UNIVERSAL socket.
A short circuit at the Multifunction socket has been detected.
An Olympus HF instrument has not been properly connected to the UNIVERSAL socket.
A malfunc tion at the Multifunction socket has been detected.
Refer to E004
An Olympus HF instrument has not been properly connected to the UNIVERSAL socket.
Invalid power setting detected at the instrument connected to the Multifunction socket.
Check the connection of the Olympus HF instrument to the UNIVERSAL socket.
Change the connected instrum ent at the Multifunction socket
Check the connection of the Olympus HF instrument to the UNIVERSAL socket.
Change the connected instrument at the Multifunction socket
Check the connection of the Olympus HF instrument to the UNIVERSAL socket.
Change the Instrument at the Multifunction socket
Unknown instrument
„Reconnect Instrument” „Replace Instrument” „Contact OLY”
7.022.211 / ISSUE 5 61 / 110 Chapter 4: Troubleshooting
An Olympus HF instrument has not been properly connected to the UNIVERSAL socket.
Invalid effect setting detect ed at the instrument connected to the Multifunction socket.
Check the connection of the Olympus HF instrument to the UNIVERSAL socket.
Change the Instrument at the Multifunction socket
Page 62
ESG-400
E153
E154
E159
E165
E166 E167
E168
E171
E179
The generator (Generator Board) is
Error no.
Error message Possible cause Remedial action
Unknown instrument
„Reconnect Instrument” „Replace Instrument” „Contact OLY”
Unknown instrument
„Reconnect Instrument” „Replace Instrument” „Contact OLY”
Unknown instrument
„Reconnect Instrument” „Replace Instrument” „Contact OLY”
An Olympus HF instrument has not been properly connected to the UNIVERSAL socket.
Invalid default Cut Mode detected at the instrument connected to the Multifunction socket.
An Olympus HF instrument has not been properly connected to the UNIVERSAL socket.
Invalid default Coag Mode detected at the instrument connec ted to the Multifunction socket.
An Olympus HF instrument has not been properly connected to the UNIVERSAL socket.
Invalid Mode detected at the instrument connected to the Multifunction socket.
Check the connection of the Olympus HF instrument to the UNIVERSAL socket.
Change the Instrument at the Multifunction socket
Check the connection of the Olympus HF instrument to the UNIVERSAL socket.
Change the Instrument at the Multifunction socket
Check the connection of the Olympus HF instrument to the UNIVERSAL socket.
Change the Instrument at the Multifunction socket
E169 E170
[…] E178
Error
„Auto-restart” „Contact OLY”
Unknown instrument
„Reconnect Instrument” „Replace Instrument” „Contact OLY”
Unknown instrument
„Reconnect Instrument” „Replace Instrument” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Refer to E004
An Olympus HF instrument has not been properly connected to the UNIVERSAL socket.
Invalid Version Number detected at the instrument connected to the Multifunction socket.
An Olympus HF instrument has not been properly connected to the UNIVERSAL socket.
Internal software error. Change the Embedded PC, change the
Internal hardware error. Send back to Olympus Service
Check the connection of the Olympus HF instrument to the UNIVERSAL socket.
Change the Instrument at the Multifunction socket
Check the connection of the Olympus HF instrument to the UNIVERSAL socket.
Motherboard
Refer to E004
T emper atur e below limit
Switch off ESG-400 and wait until operating temperature is reached.
7.022.211 / ISSUE 5 62 / 110 Chapter 4: Troubleshooting
„Contact OLY”
too cold.
Switch off the electrosurgica l gener at or and wait until it has reached the specified operating temperature.
Page 63
ESG-400
E180
The generator (Generator Board) is
E181
The generator (HVPS Board) is too
E182
The generator (HVPS Board) is too
E183
E186
E187
E188
E189
E197
Error no.
Error message Possible cause Remedial action
T emper atur e above limit
Switch off the ESG-400 and wait until it has cooled down.
„Contact OLY”
Temperature below limit
Switch off ESG-400 and wait until operating temperature is reached.
„Contact OLY”
T emper atur e above limit
Switch off the ESG-400 and wait until it has cooled down.
„Contact OLY”
too hot.
cold.
hot.
Switch off the electrosurgica l gener at or and wait until it has cooled down or reached the specified operating temperature.
Switch off the electrosurgica l gener at or and wait until it has reached the specified operating temperature.
Switch off the electrosurgica l gener at or and wait until it has cooled down or reached the specified operating temperature.
E184 E185
[…] E196
E198
Error „Contact OLY” Error „Contact OLY” Increased HF leakage
current
Check if an instrument, the neutral electrode or patient is unintentionally grounded
„Contact OLY”
Excessive HF leakage current
Check if an instrument, the neutral electrode or patient is unintentionally grounded
„Contact OLY”
Error „Contact OLY” Error
„Auto-restart”
Refer to E132
Internal hardware error. Send back to Olympus Service
The high frequency leakage current has exceeded the limit of 150 mA for monopolar application or 100 mA for bipolar application.
The high frequency leakage current has exceeded the limit of 300 mA for monopolar application or 200 mA for bipolar application.
Check if an instrument, the neutral electrode or the patient is unintentionally grounded.
Check if an instrument, the neutral electrode or the patient is unintentionally grounded.
Refer to 132
Refer to 132
7.022.211 / ISSUE 5 63 / 110 Chapter 4: Troubleshooting
„Contact OLY”
Page 64
ESG-400
E199
E200
E202
E203 E212
E213 E214
E215
E216
E220
E222
E295
E296
Error no.
E201
[…] E211
Error message Possible cause Remedial action
Error
„Contact OLY”
Error „Contact OLY” Insufficient neutral
electrode contact
Check the connection and attachment of the neutral electrode.
If the problem persists, attach a new neutral electrode.
„Contact OLY”
Error
„Auto-restart” „Contact OLY”
The contact resistance of the neutral electrode is too high or the neutral electrode is not connected.
Malfunction of the neutral electrode and / or the neutral electrode cable.
Refer to 133
Refer to 132
Check the connection / attachment of the neutral electrode.
Replace the neutral electrode and / or the cable.
Refer to E020
E217
[…] E294
Error „Contact OLY” Error
„Auto-restart” „Contact OLY”
Low battery „Contact OLY” Error „Contact OLY” Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Refer to E020
Refer to E020
The batteries on the Motherboard are low and caused a loss of data.
Internal software or hardwar e error. Send back to Olympus Service
Send back to Olympus Service
Refer to E020
Refer to E058
Refer to E058
Error
„Auto-restart” „Contact OLY”
Error
7.022.211 / ISSUE 5 64 / 110 Chapter 4: Troubleshooting
Unable to read the data at the Multifunction socket.
Unable to write the data from the 1. Change the connected instrument at
1. Change the connected instrument at the Multifunction socket.
2. Send back to Olympus Service
Page 65
ESG-400
Multifunction socket.
the Multifunction socket.
E297 E299
E390
E391
E394
E396
E397
Error no.
[…] E387
Error message Possible cause Remedial action
„Auto-restart” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Communication error
„Cable Connection” „Cable Damage” „Contact OLY”
Unable to read the instrument name from the Multifunction s ocket.
Improper connection of the communication cables to the LINK-OUT / LINK-IN socket: “The communication to other devices has been aborted.”
2. Send back to Olympus Service
1. Change the connected instrument at the Multifunction socket.
2. Send back to Olympus Service
Refer to E058
1. Check all cable connections.
2. Check all other connected devices.
3. Send back to Olympus Service
E392 E393
Error „Contact OLY” Communication error
„Cable Connection” „Cable Damage” „Contact OLY”
T emper atur e below limit
Switch off ESG-400 and wait until operating temperature is reached.
„Contact OLY”
T emper atur e above limit
Malfunction or damage of the communication cables.
Refer to E058
Improper connection of the communication cables to the LINK-OUT / LINK-IN socket: “The ring connection to other devices has been detected.”
Malfunction or damage of the communication cables.
The generator (Generator Board) is too cold.
The generator (Generator Board) is too hot.
Check the cables for damages and, if necessary, replace the cables.
1. Check all cable connections.
2. Check all other connected devices.
3. Send back to Olympus Service. .
Check the cables for damages and, if necessary, replace the cables.
Place the generator at norm al room temperature and wait until it is warmed up before use.
Wait until the generator has cooled down.
7.022.211 / ISSUE 5 65 / 110 Chapter 4: Troub les hoot ing
Switch off the ESG-400 and wait until it has cooled down.
„Contact OLY”
Page 66
ESG-400
E398
The generator (HVPS Board) is too
Place the generator at normal room
E399
E400
Refer to E058
E409
E410
E411
E412
Refer to E058
Error no.
[…] E408
Error message Possible cause Remedial action
T emper atur e below limit
Switch off ESG-400 and wait until operating temperature is reached.
„Contact OLY”
Temperature above limit
Switch off the ESG-400 and wait until it has cooled down.
„Contact OLY”
Error
„Auto-restart” „Contact OLY”
Communication error
„Cable Connection” „Cable Damage” „Contact OLY”
cold.
The generator (HVPS Board) is too hot.
Improper connection of the communication cables to the LINK-OUT / LINK-IN socket: “The communication problem to other devices has been detected.”
temperature and wait until it is warmed up before use.
Wait until the generator has cooled down.
1. Check all cable connections.
2. Check all other connected devices.
3. Send back to Olympus Service
Error
„Auto-restart” „Contact OLY”
Communication error
„Cable Connection” „Cable Damage” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Malfunction or damage of the communication cables.
Improper connection of the communication cables to the LINK-OUT / LINK-IN socket: “The communication problem to other devices has been detected.”
Malfunction or damage of the communication cables.
Check the cables for damages and, if necessary, replace the cables.
Refer to E058
1. Check all cable connections.
2. Check all other connected devices.
3. Send back to Olympus Service
Check the cables for damages and, if necessary, replace the cables.
7.022.211 / ISSUE 5 66 / 110 Chapter 4: Troubleshooting
Page 67
ESG-400
E413
E414
The controlled output power at the
E415 E416
E419
E430
E431
E432
E434
E438
E439
Error no.
E417
[…] E429
Error message Possible cause Remedial action
Error „Contact OLY” Error
„Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error „Contact OLY” Error
„Auto-restart” „Contact OLY”
--­Error
The controlled output power at the Burn-In is below the limit.
Burn-In is above the limit.
--- (POWER_FAIL) ---
Check the connected resistors.
Send back to Olympus Service
Refer to E004
Refer to E004
Refer to E004
Refer to E004
E433
[…] E437
E440
„Auto-restart” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Procedure data error
One or more procedures have been deleted.
Press OK to continue. „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Internal hardware error. Send back to Olympus Service
Refer to E004
One or more saved procedures have been deleted. (Inconsistant Procedure data detected. One or more Procedure settings have been lost.)
Internal hardware error. Send back to Olympus Service
Press the “OK button” to close the error window and to continue.
7.022.211 / ISSUE 5 67 / 110 Chapter 4: Troubleshooting
Page 68
ESG-400
E441
All settings of the electrosurgical
The electrosurgical generator is ready to E443
E457 E458
E459
E461
E486
E488
Refer to E058
E490
Improper connection of the
Malfunction or damage of the
Check the cables for damages and, if
E491
Refer to E058
Error no.
[…] E456
E460
Error message Possible cause Remedial action
Device setting error
Device settings have been set to default.
„Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart”
generator have been set to default. (SRAM and EEPROM content has been initialized.)
Internal software or hardwar e error. Send back to Olympus Service
use after the error window disappeared.
Refer to E004
Refer to E058
Refer to E133
[…] E484
E489
„Contact OLY”
Error
„Auto-restart” „Contact OLY”
No instrument connected
Connect an instrument to the UNIVERSAL socket.
„Replace Instrument” „Contact OLY”
Error
„Contact OLY”
Communication error
„Cable Connection” „Cable Damage” „Contact OLY”
An Olympus HF instr ument and / or connection cable has not been properly connected to the UNIVERSAL socket.
Malfunction of the Olympus HF instrument and / or the connection cable.
communication cables to the LINK-OUT / LINK-IN socket: “The communication problem to other devices has been detected.”
Refer to E058
Ensure the proper connection of the Olympus HF instrument and / or the connection cable to the UNIVERSAL socket.
Replace the Olympus HF instrument and / or the connection cable.
1. Check all cable connections.
2. Check all other connected devices.
3. Send back to Olympus Service
[…]
7.022.211 / ISSUE 5 68 / 110 Chapter 4: Troubleshooting
Error
communication cables.
necessary, replace the cables.
Page 69
ESG-400
E504
E505
Refer to E058
E515
E516
E519
E525
E527
E539
E540
E541 E550
Error no.
[…] E514
Error message Possible cause Remedial action
„Auto-restart” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Communication error
„Cable Connection” „Cable Damage” „Contact OLY”
Communication error
„Cable Connection” „Cable Damage” „Contact OLY”
Improper connection of the communication cables to the LINK-OUT / LINK-IN socket: “The communication problem to other devices has been detected.”
Malfunction or damage of the communication cables.
Improper connection of the communication cables to the LINK-OUT / LINK-IN socket: “The communication problem to other devices has been detected.”
1. Check all cable connections.
2. Check all other connected devices.
3. Send back to Olympus Service
Check the cables for damages and, if necessary, replace the cables.
1. Check all cable connections.
2. Check all other connected devices.
3. Send back to Olympus Service
[…] E524
E526
[…] E538
Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart”
Malfunction or damage of the communication cables.
Check the cables for damages and, if necessary, replace the cables.
Refer to E058
Refer to E020
Refer to E058
Refer to E058
Refer to E058
„Contact OLY”
[…] E549
[…] E552
7.022.211 / ISSUE 5 69 / 110 Chapter 4: Troubleshooting
Error
„Auto-restart” „Contact OLY” Error
(<4.09) „Auto-restart”
Refer to E058
Refer to E058
Page 70
ESG-400
E553
E558
E560
E561
E564
E617
Tissue built up on tip causing
Clean the tip with saline solution. If
Malfunction of the Olympus HF
Replace the Olympus HF instrument and
E619
Error no.
[…] E557
Error message Possible cause Remedial action
„Contact OLY” Flash Memory Failure
( 4.09) The system is no longer
able to boot. Change the Power PC
module. *Message is different for
software versions lower than 4.09-A
Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Refer to E058
Refer to E058
E562
[…] E611
Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Excess coagulum on tip
Clean the instrument tip and proceed.
„Replace Instrument” „Contact OLY”
Data transfer
Wait for device recognition prior to activation.
Press OK. „Replace Instrument” „Contact OLY”
sparking.
instrument and / or the connection cable.
Device was activated by pressing the footswitch or handswitch while the instrument recognition data are read after connecting to the generator.
Malfunction of the Olympus HF instrument and / or the connection cable.
Refer to E058
Refer to E058
Refer to E058
necessary use soft tissue or brush to protect the coating of the instrument.
/ or the connection cabl e.
Wait for the completion of the data transfer (about 3 seconds) indicated by the display of the instrument name on the screen. Afterwards the device can be activated.
Replace the Olympus HF instrument and / or the connection cabl e.
7.022.211 / ISSUE 5 70 / 110 Chapter 4: Troubleshooting
Page 71
ESG-400
E620
E646 E647
E650
E652 E653
E654
E657
E658
E659
Error no.
[…] E645
[…] E649
E651
Error message Possible cause Remedial action
Error
„Auto-restart” „Contact OLY”
Mains voltage drop
(only visible in error log) „Auto-restart” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart”
Internal software error. Send back to Olympus Service
Mains voltage too low. Check the power cord and the connection
to the wall outlet. Check the condition of the power network of the facility.
Internal hardware error. Send back to Olympus Service
Internal hardware or softwar e error. Send back to Olympus Service
Malfunction of Olympus HF instrument.
Internal hardware or softwar e error. Send back to Olympus Service
Internal software error. Send back to Olympus Service
Replace the instrument.
[…] E656
„Contact OLY”
Error
„Contact OLY”
Error
„Auto-restart” „Contact OLY”
Short circuit
Re-grasp tissue and proceed.
Avoid having the jaws in contact with each other.
„Contact OLY”
Short circuit
Short circuit / Re-adjust loop and proceed.
Avoid contact betw een loop and other instruments or metal parts.
Internal software error. Send back to Olympus Service
Internal software error. Send back to Olympus Service
Electrodes of the HF instrument may touch each other.
Malfunction of the HF instrument and / or the connection cable.
Electrodes of the HF instrument may touch each other.
Malfunction of the HF instrument and / or the connection cable.
Ensure that the electrodes of the HF instrument do not touch each other.
Replace the HF instrument and / or the connection cable.
Ensure that the electrodes of the HF instrument do not touch each other.
Replace the HF instrument and / or the connection cable.
„Contact OLY”
Error
„Auto-restart” „Contact OLY”
7.022.211 / ISSUE 5 71 / 110 Chapter 4: Troubleshooting
Internal software error. Send back to Olympus Service
Page 72
ESG-400
E661
E665
Cut and coag switches were
Release the switches and proceed to
Malfunction of the HF instr um e nt
Replace the HF instrument and / or the
E669
E672
Error no.
[…] E671
E673
Error message Possible cause Remedial action
Short circuit
Re-grasp tissue and proceed. Avoid contact between the instrument tip and grasper.
„Replace Instrument” „Contact OLY”
Error
Application time limit exceeded
„Replace Instrument” „Contact OLY”
Error
„Auto-restart” „Contact OLY”
Error
„Auto-restart”
Electrodes of the HF instrument may touch each other.
Malfunction of the HF instrument and / or the connection cable.
pressed simultaneously.
and / or the connection cable.
Internal software error. Send back to Olympus Service
Internal software or hardware err or. Send back to Olympus Service
Ensure that the electrodes of the HF instrument do not touch each other.
Replace the HF instrument and / or the connection cable.
work by pressing only one switch (cut or coag).
connection cable.
„Contact OLY”
7.022.211 / ISSUE 5 72 / 110 Chapter 4: Troubleshooting
Page 73
ESG-400

CHAPTER 5: INSPECTION

1 JIGS, TOOLS, AND MEASURING EQUIPMENT FOR INSPECTION .......................... 74
2 INSPECTION PROCEDURES ...................................................................................... 75
2-1 Visual inspection of the electrosurgical generator and accessories ................................................ 76
2-2 Verifying the contact quality monitor function .................................................................................. 80
2-3 Checking the DC resistance (according to IEC 60601-2-2) ............................................................. 82
2-4 Checking the earth resistance (according to IEC 60601-1 and IEC 62353) .................................... 82
2-5 Checking the earth leakage current (according to IEC 60601-1) .................................................... 82
2-6 Checking the patient leakage current (according to IEC 60601-1) .................................................. 83
2-7 Checking the current and power consumption and output waveform .............................................. 84
2-8 Checking the high frequency leakage current (according to IEC 606 01-2-2) .................................. 86
2-8-1 Measurement of the monopolar high frequency leakage current under loaded condition ...... 86
2-8-2 Measurement of the monopolar high frequency leakage current under unloaded condition .. 88
2-8-3 Measurement of the bipolar high frequency leakage current under loaded condition ............ 90
2-8-4 Measurement of the bipolar high frequency leakage current under unloaded condition ........ 92
2-9 Checking the output power .............................................................................................................. 94
2-10 Checking for certain features and error messages .......................................................................... 94
2-11 Final test ........................................................................................................................................... 95
2-11-1 Self test.................................................................................................................................... 95
2-11-2 Display and sound check ........................................................................................................ 95
2-11-3 Functionality of push buttons .................................................................................................. 95
2-11-4 Communication test ................................................................................................................. 95
2-11-5 Restore of output power settings ............................................................................................ 96
2-12 Inspection label (For applicable markets) ........................................................................................ 96
3 INSPECTION CARD ..................................................................................................... 97
7.022.211 / ISSUE 5 73 / 110 Chapter 5: Inspection
Page 74
ESG-400
No.
Name
Model/REF
Specifications/Remarks
J001
Electrical safety tester
e.g.: Seculife ST, Unimet 1000 ST
J002
J003
J004
J005
J007
Connection cable: monopolar or bipolar
and the MONOPOLAR 2 socket may
J008
Connection cable: bipolar UNIVERSAL to various equipment (1 x 4 mm, banana
J009
handswitch to various equipment (1 x 4
J010
ndswitch to various equipment (1 x 4
J011
J012
J013
For connection to Potential Equalization
J014
J015
J016
Ordering depending on country. Refer to
J057
J058
J059
Connection cable: bipolar UNIVERSAL to
pment (3 x 4 mm, banana

1 Jigs, Tools, and Measuring Equipment for Inspection

(Bender), QA-90 (Metron)
Electrosurgical analyzer Digital multimeter
Load resistors (low pow e r)
Load resistors (high power, low inductive)
Cables, 4 mm to 4 mm (banana)
DC accuracy: < 1 %,
low inductive part, short time l oad 5 Ω, 140
WB979015
e.g.: QA-ES (Metron)
DC voltage range 500 V
Ω, 170 Ω all 0.5 W (or more) and 1 %
tolerance or alternatively: resistor decade box
Low inductive part, 75 Ω, 500 Ω (300 W, 5%) 7 Ω, 916 Ω, 10 kΩ (100 W, 1%) OR ESG-Testbox
output (1 x 4 mm) to various equipment (1 x 4 mm, banana plug), length 1 m
WARNING:
A connection of a 4 mm plug to any other receptacle except the right-hand-side receptacle of the MONOPOLAR 1 socket
Cable, UNIVERSAL to 4 mm (banana)
Cable, Monopolar 1 including handswitch
Cable, Monopolar 2 including hands witch
Cable, neutral electrode (P-cord) to 2 x 4 mm (banana)
Cable, Communication Crocodile clips (with 4 mm connection)
ESG-400 foot switch double pedal ESG-400 foot switch single pedal Power Cord
Oscilloscope High Vo ltage Probe
Adapter Universal Socket
destroy the socket during activation.
WB979008
plug), length 1 m
WB979013
WB979014
WB979002 Connection cable: neut ral electrode output
MAJ-1871 Communication cable, length 0.25 m
WB50402W WB50403W
e.g.: DSOX2024A (Agilent) Attenuation: 100:1
W5106278
Connection Cable: monopolar output incl.
mm, banana plug), length 3…4 m Connection Cable: monopolar output incl.
ha mm, banana plug), length 3…4
(2 x 2.5 mm, 10 mm, P-cord) to various equipment (2 x 4 mm, banana plug), length 1 m
bonding
Chapter 1-4-2
Max. Input Voltage 850 Vp
various equi plug)
7.022.211 / ISSUE 5 74 / 110 Chapter 5: Inspection
Page 75
ESG-400
or shunting of high frequency currents
from any metal surface including tabletops and other resistors. This is especially true if several resistors are connected in series or parallel to obtain
Keep test leads to the minimum length usable; lead inductance and stray
Carefully select suitable ground points to avoid ground loop error in

2 Inspection procedures

The electrosurgical generator and the footswitch must undergo an ins pection / safety check in yearly intervals in accordance with the national statutory regulations. Inspection is also mandatory after repair, adjustment, update and upgrade.
Generally the foots witches are inspected together with the electrosurgical unit. If a footswitch alone has to be checked, only following tests are mandatory. The Numbers used referring to the Inspection Card in chapter 5-3:
For a single footswitch test number 2, 39 and 55 are mandatory. For a double footswitch test number 2, 38, 39 and 55 are mandatory.
Follow these test instructions. All tests must be done with fully functional and calibrated test equipment and by technicians trained in the service / maintenance of electrical medical devices. Record the test results in the “Inspection Card” (refer to chapter 5-3, Inspection Card) for reference in future tests and provide the user of the electrosurgical unit with a signed report.
During service / maintenance take care of the different hardware and software versions which may be applicable. Information how to identify the hardware version can be found in chapter 3-2, Board Compatibility. Information how to identify the software version can be found in chapter 1-3, Software version.
If the electrosurgical unit fails to meet any of the checks, the unit has to be adjusted according to chapter 6 (Adjustment) or refer to chapter 4 (Troubleshooting). If the failure still occurs, contact the manufa cturer.
CAUTION
NOTE
To avoid inadvertent coupling and / around the resistor elements, keep the resistors at least 10 centimeters away
a specified value. Do not allow the resistor bodies to touch each other.
capacitance can adverse ly affect readings.
measurements. For tests and c he ck s r e qu ir in g a power cord, use the power cord provided with
the electrosurgical unit. Inspection is mandatory after repair, adjustment, update and upgrade.
7.022.211 / ISSUE 5 75 / 110 Chapter 5: Inspection
Page 76
ESG-400
12345W13-678 2011-08
12345W13-678 2014-08

2-1 Visual inspection of the electrosurgical generator and accessories

1) Check that the labels according to chapter 5-1 (Front panel, Rear panel) are present and legible. The product name should be clearly visible on the front panel. Verify that the type plate shows details about the type of the device, reference number, line voltage range, supply frequency, output power classification, output frequency, duty cycle, serial number, manufacturing date and manufacturer according to Fig. 5.2.1, Fig. 5.2.2, Fig. 5.2.3 or Fig. 5.2.4
Fig. 5.2.1. Sample of type plate of the electrosurgical unit
Fig. 5.2.2. Sample of type plate of the electrosurgical unit ( Standard PK version)
7.022.211 / ISSUE 5 76 / 110 Chapter 5: Inspection
Page 77
ESG-400
12345W13-678 2013-03
Fig. 5.2.3 Sample of type plate of the electrosurgical unit (Japan Version)
Fig. 5.2.4 Sample of type plate of the electrosurgical unit (China Version)
2) Check the electrosurgical unit and the accessories for external obstructions or damage. Verify that the
housing, the front panel and the rear panel have no serious destructions.
3) Record the test results in the safety check report (refer to chapter 5-3).
4) Copy the tables 7.1 and 7.2 on the next page and prepare them to be filled out while inspecting the
7.022.211 / ISSUE 5 77 / 110 Chapter 5: Inspection
Page 78
ESG-400
electrosurgical generator. Fill out the table 7.1 after the first startup of the generator. The field “Actual Procedure” can only be filled, if a procedure is chosen.
5) The table 7.2 is used every time before a mode is changed in order to perform a test. The settings of the
used mode are noted in one line of table 7.2. Restore the values of table 7.2 either after each performed test or afte r the ins pec ti on. This should be done, to make sure that the physician will be provided with the last output power setting.
7.022.211 / ISSUE 5 78 / 110 Chapter 5: Inspection
Page 79
ESG-400
Actual Procedure
<if applicable>
Actual Screen
Table 7.1: Template to note the actual procedure and the user assignment of the modes to the sock ets
Socket Mode Power Level Effect
BIPOLAR MONOPOLAR 1
Cut Mode Coag Mode Cut Mode Coag Mode
UNIVERSAL MONOPOLAR 2
Cut Mode Coag Mode Cut Mode Coag Mode
Table 7.2: Template to note the power settings
7.022.211 / ISSUE 5 79 / 110 Chapter 5: Inspection
Page 80
ESG-400

2-2 Verifying the contact quality monitor function

1) Connect the power cord and the footswitch to the electrosurgical generator (refer to chapter 1-5-3 and
1-5-8).
2) All steps for this test are described in the table 7.3 at the next page. Assign the footswitch according to the
column “Mode”. Connect the appropriate resistor R (J004) described in the column “Action” via the connection cable (J011) with the neutral electrode socket at the front panel (refer to chapter 1-5-2). For some tests leave the connector open or short circuit the conn ecti on cabl e (J011).
3) Follow the steps described in the column “Action” and verify the reaction of the corresponding contact
quality monitor indicator for non-split neutral electrode or split neutral electrode at the front panel (refer to chapter 1-5-2) according to table 7.3.
4) Repeat step 2 and 3 until all action items in table 7.3 have been checked.
5) Record the test results in the Inspection Card (refer to chapter 5-3).
7.022.211 / ISSUE 5 80 / 110 Chapter 5: Inspection
Page 81
ESG-400
split neutral
split neutral electrode
ion a warning tone
is displayed (“Insufficient neutral
NOTE
Mode Action Expected test result
Always disconnect the neutral electrode connector (J011) before connecting a new resistor.
Any bipolar mode
Any monopolar mode
During standby and activation
1) The neutral electrode socket is left open, nothing is connected.
2) Activate any bipolar output.
During standby:
1) Short circuit the
neutral electrode connector.
2) Connect a R = 5 Ω
with the neutral electrode socket
During standby:
1) Connect a R = 140 Ω
with the neutral electrode connector
The contact quality monitor is independent from the bipolar mode. The indicator for electrode is illuminated red. During activation the HF-output is working.
Pop-Up Window at the screen will appear.
Contact quality monitor indicator for non-
illuminates green.
Contact quality monitor indicator for split neutral electrode illuminates
green.
During standby and activation:
1) Connect a R = 170 Ω, with the neutral
electrode connector
2) Activate any monopolar output.
Table 7.3: Checking the contact quality monitor
Contact quality monitor indicator for split neutral electrode illuminates
red. During activat
can be heard, the error code E202
electrode contact”) and activation is disabled.
7.022.211 / ISSUE 5 81 / 110 Chapter 5: Inspection
Page 82
ESG-400

2-3 Checking the DC resistance (according to IEC 60601-2-2)

1) Activate the safety test function in the service menu (see chapter 6-1, Safety Test) and verify that the button “Relays on” is marked white.
2) Connect the digital multimeter (J003) with the bipolar connector on the front panel (see chapter 1-5-2).
3) Verify the resistance is ≥ 2 MΩ.
4) Deactivate the safety test function in the service menu.
5) Record the test results in the Inspection Card (see chapter 5-3).

2-4 Checking the earth resistance (according to IEC 60601-1 and IEC 62353)

1) Connect the electrosurgical generator with an electrical safety tester (J001) according to the tester’s instructions for use.
2) If the power cord and the electrosurgical unit are m easured together, verify the protective earth resistance ≤ 0.3 Ω against metal parts which can be touched.
3) Record the test results in the Inspection Card (see chapter 5-3).

2-5 Checking the earth leakage current (according to IEC 60601-1)

1) Connect the electrosurgical generator with an electrical safety tester (J001) according to the tester’s instructions for use.
2) Switch on the electrosurgical unit.
3) Verify under normal condition (NC) the earth leakage current is ≤ 0.5 mA.
4) Verify under single fault condition (SFC) the earth leakage current is ≤ 1.0 mA
5) Record the test results in the Inspection Card (see chapter 5-3).
7.022.211 / ISSUE 5 82 / 110 Chapter 5: Inspection
Page 83
ESG-400
For this test an interconnection of the three receptacles of MONOPOLAR 1
allowed. Before interconnect the three receptacles

2-6 Checking the patient leakage current (according to IEC 60601-1)

NOTE
1) Switch the electrosurgical generator on and activate the safety test function in the service menu (see chapter 8-2, Safety Test) and verify that the button “Relays on” is marked white.
2) Connect the UNIVERSAL cable (J008) to the UNIVERSAL socket and the connection cable P-cord (J011) to the neutral socket.
3) Connect two 4mm cables “banana” (J007) to every receptacle of the BIPOLAR socket.
4) Connect three 4mm cables “banana” to the 4mm receptacles of the MONOPOLAR 1 socket. Leave the 8 mm BOVIE connector open.
5) Connect three 4mm cables “banana” to the 4mm receptacles of the MONOPOLAR 2 socket. Leave the 5/9 mm ERBE connector open.
6) Verify again that the Button “Relays on” at the touch screen is marked white and short circuit all connectors.
and MONOPOLAR 2 is verify, that the safety test function is activated and the button “Relays on” is marked white.
7) Switch the electrosurgical generator off, cable it with the elec trical safety tester (J001) according to the tester’s instructions for use and switch it on again. Start the test with the electrical safety tester according to the tester’s instructions for use.
8) Verify under normal condition (NC) for AC the patient leakage current is ≤ 0.01 mA.
9) Verify under normal condition (NC) for DC the patient leakage current is ≤ 0.01 mA.
10) Verify under single fault condition (SFC, “open earth”) for AC the patient leakage current is ≤ 0.05 mA.
11) Verify under single fault c ondition (SFC, “open earth”) for DC the patient leakage current is ≤ 0.05 mA.
12) Disconnect all cables from the electrosurgical generator.
13) Deactivate the safety test function in the service menu.
14) Record the test results in the Inspection Card (see chapter 5-3).
7.022.211 / ISSUE 5 83 / 110 Chapter 5: Inspection
Page 84
ESG-400

2-7 Checking the current and power consumption and output waveform

CAUTION
1) Connect the load resistor RL = 500 Ω (J005) via a 4mm cable “banana” (J007) to the right receptacle of the MONOPOLAR 1 socket and via the connection cable “P-cord” (J011) to the neutral socket in the front panel.
2) Cable the electrosurgical generator with the electrical safety tester according to the testers instructions for use. Activate the function to measure current and power consumption.
3) Switch the electrosurgical generator on and select the PureCut (Effect 3) mode at the MONOPOLAR 1 socket. Set the level to 300. Assign the footswitch to the MONOPOLAR 1 socket. (Perform the setting according to the instruction for use.)
4) Activate the output power by pressing “CUT” at the footswitch.
5) Verify the current consu mption I
6) Connect high voltage probe to oscilloscope and to generator output socket (signal to right receptacle of the MONOPOLAR 1 socket and signal ground to neutral socket)
A connection of a 4 mm plug to any other receptacle except the right-hand-side receptacle of the MONOPOLAR 1 socket and the MONOPOLAR 2 socket may destroy the socket during activation.
is ≤ 10 A.
L
Recommended oscilloscope settings:
o Channel 1: high voltage probe 1000:1, DC, 200 V/div (minimum total voltage range
displayed on oscilloscope: -700 V … +700 V)
o Time:
minimum 2 ms/div (minimum total time range displayed on oscilloscope:
20 ms)
maximum 5 ms/div (maximum total time range displayed on oscilloscope:
50 ms)
7) Activate the output power by pressing “CUT” at the footswitch. Record the waveform in the time range after the high power cut support pulse (ca. 100 ms after HF output start) in steady state.
8) Verify output waveform: The amplitude level of the output voltage must be stable. No oscillation in the frequency range of 50 … 200 Hz. A maximum amplitude fluctuation from minimum peak voltage to maximum peak voltage of 110 V is allowed for steady state.
NOTE
Table 7.4 is showing good and bad examples how the output waveform should
look like. From the f igures in this table it is also clear to see, which parameter is to be measured.
9) Disconnect high voltage probe.
10) Connect the load resistor RL = 75 Ω (J005) via the connection cable (J008) with the UNIVERSAL socket in the front panel.
7.022.211 / ISSUE 5 84 / 110 Chapter 5: Inspection
Page 85
ESG-400
example for
example for with mains
voltage of
example for
h mains
voltage of
11) Select the SalineCut (Effect 3) mode at the UNIVERSAL socket. Set the level to 320. Assign the footswitch to the UNIVERSAL socket. (Perform the setting according to the instruction for use.)
12) Activate the output power by pressing “CUT” at the footswitch.
13) Verify the apparent power consumption S
14) Record the test results in the Inspection Card (see chapter 7-3)
15) Switch the electrosurgical generator off and disassemble the test set-up.
is ≤ 1500 VA.
L
Setup Picture of oscilloscope (2 ms / div) Picture of oscilloscope (5 ms / div)
“good” WB91051W
“bad” WB91051W
230 VAC
“bad” WB91051W
wit 115 VAC
Table 7.4: good and bad examples of the output waveform
7.022.211 / ISSUE 5 85 / 110 Chapter 5: Inspection
Page 86
ESG-400

2-8 Checking the high frequency leakage current (according to IEC 60601-2-2)

The monopolar high frequency leakage current has to be measured according to the procedure 19.3.101 a) 2) (neutral electrode isolated from earth at high frequency) as described in IEC 60601-2-2.
The bipolar high frequency leakage current has to be measured according to the procedure 19.3.101 a) 3) (bipola r application) as described in IE C 60601-2-2.
The high frequency current is measured from the appropriate output connector of each pole while the electrosurgical unit is operated at maximum output power setting in an appropriate mode and the output being unloaded and loaded at rated load. The high frequency current is measured with the electrosurgical analyzer through an internal resistance of 200 Ω.

2-8-1 Measurement of the monopolar high frequency leakage current under loaded condition

CAUTION
Fig. 5.2.4 and Fig. 5.2.5 are showing an example how to set up the test to measure the monopolar high frequency leakage current under loaded condition. The load resistor R connection cable (J007, banana) with the active pole MONOPOLAR 1 and via the connection cable J011 with the NEUTRAL connector. The high frequency leakage current is floating from appropriate pole to the equipotential bonding port. It is measured via the electrosurgical analyzer via a measuring resistance of 200 . Fig. 5.2.4 shows the measurement from the neutral pole; Fig. 5.2.5 shows the measurement from the active pole.
A connection of a 4 mm plug to any other receptacle except the right-hand-side receptacle of the MONOPOLAR 1 socket and the MONOPOLAR 2 socket may destroy the socket during activation.
= 500 Ω (J005) is connected via the
L
1) Assign the footswitch to the MONOPOLAR 1 socket and select the monopolar PureCut mode with effect 3.
Set the power level to 300. (Perform the setting according to the instruction for use.)
2) Set up the measurement according to Fig. 5.2.4 (Measuring of the high frequency leakage current under
loaded condition at the neut ral pole.)
3) Activate the output power by pressing the corresponding footswitch pedal.
4) Verify the leakage current is ≤ 150 mA.
5) Set up the measurement according to Fig. 5.2.5 (Measuring of the high frequenc y leakage current under
loaded condition at the act iv e pole.)
6) Activate the output power by pressing the corresp onding footswitch pedal.
7) Verify the leakage current is ≤ 150 mA.
8) Repeat step 2 - 7 for the monopolar SprayCoag mode with effect 3 and power level 120.
9) Record the test results in the Inspection Card (see chapter 5-3).
7.022.211 / ISSUE 5 86 / 110 Chapter 5: Inspection
Page 87
ESG-400
A connection of a 4 mm plug to any other receptacle except the
side receptacle of the MONOPOLAR 1 socket and the
Fig. 5.2.3. Example for the loaded measurement of the monopolar high frequency leakage current at the neutral pole
Fig. 5.2.4. Example for the loaded measurement of the monopolar high frequency leakage current at the active pole
CAUTION
7.022.211 / ISSUE 5 87 / 110 Chapter 5: Inspection
right-hand­MONOPOLAR 2 socket may destroy the socket during activation.
Page 88
ESG-400

2-8-2 Measurement of the monopolar high frequency leakage current under unloaded condition

CAUTION
Fig. 5.2.6 and Fig. 5.2.7 are showing an example how to set up the test to measure the monopolar high frequency leakage current under unloaded condition. In this test the output is unloaded, the load resistance used in the test before is removed. The high frequency leakage current is floating from appropriate pole to the equipotential bonding port. It is measured via the electrosurgical analyzer via a measuring resistance of 200 . Fig. 5.2.6 shows the measurement from the neutral pole; Fig. 5.2.7 shows the measurement from the active pole.
A connection of a 4 mm plug to any other receptacle except the right-hand-side receptacle of the MONOPOLAR 1 socket and the MONOPOLAR 2 socket may destroy the socket during activation.
1) Assign the footswitch to the MONOPOLAR 1 socket and select the monopolar PureCut mode with effect 3.
Set the power level to 300. (Perform the setting according to the ins truction for use.)
2) Set up the measurement according to Fig. 5.2.6 (Measuring of the high frequency leakage current under
unloaded condition at the neutral pole.)
3) Activate the output power by pressing the corresponding footswitch pedal.
4) Verify the leakage current is ≤ 150 mA.
5) Set up the measurement according to Fig. 5.2.7 (Measuring of the high frequency leakage current under
unloaded condition at the active pole.)
6) Activate the output power by pressing the corresponding footswitch pedal.
7) Verify the leakage current is ≤ 150 mA.
8) Repeat step 2 - 7 for the monopolar SprayCoag mode with effect 3 and power level 120.
9) Record the test results in the Inspection Card (see chapter 5-3).
7.022.211 / ISSUE 5 88 / 110 Chapter 5: Inspection
Page 89
ESG-400
A connection of a 4 mm plug to any other receptacle except the
side receptacle of the MONOPOLAR 1 socket and the
Fig. 5.2.5. Example for the unloaded m easurement of the monopolar high frequ ency leakage current at the neutral pole
Fig. 5.2.6. Example for the unloaded m easurement of the monopolar high frequ ency leakage current at the active pole
CAUTION
7.022.211 / ISSUE 5 89 / 110 Chapter 5: Inspection
right-hand­MONOPOLAR 2 socket may des troy the socket during activation.
Page 90
ESG-400

2-8-3 Measurement of the bipolar high frequency leakage current under loaded condition

CAUTION
A connection of a 4 mm plug to any other receptacle except the right-hand-side receptacle of the MONOPOLAR 1 socket and the MONOPOLAR 2 socket may destroy the socket during activation.
Fig. 5.2.8 and Fig. 5.2.9 are showing an example how to set up the test to m easure the bipolar high frequency leakage current under loaded condition. The load resistor R
= 500 Ω (J005) is connected via the connection
L
cables (J007, banana) with both BIPOLAR receptacles. The high frequency leakage current is floating from appropriate pole to the equipotential bonding port. It is measured via the electrosurgical analyzer via a measuring resistance of 200 . Fig. 5.2.8 shows the measurement from terminal 1; Fig. 5.2.9 shows the measurement from termin al 2.
1) Assign the footswitch to the BIPOLAR socket and select the BipolarCut mode with effect 3. Set the power
level to 100. (Perform the setting according to the instruction for use.)
2) Set up the measurement according to Fig. 5.2.8 (Measuring of the high frequency leakage current under
loaded condition at terminal 1.)
3) Activate the output power by pressing the corresponding footswitch pedal.
4) Verify the leakage current is ≤ 70 mA.
5) Set up the measurement according to Fig. 5.2.9 (Measuring of the high frequency leakage current under
loaded condition at term inal 2.)
6) Activate the output power by pressing the corresp onding footswitch pedal.
7) Verify the leakage current is ≤ 70 mA.
8) Record the test results in the Inspection Card (see chapter 5-3).
7.022.211 / ISSUE 5 90 / 110 Chapter 5: Inspection
Page 91
ESG-400
Fig. 5.2.7. Example for the loaded measurement of the bipolar high frequency leakage current at terminal 1
Fig. 5.2.8. Example for the loaded measurement of the bipolar high frequency leakage current at terminal 2
7.022.211 / ISSUE 5 91 / 110 Chapter 5: Inspection
Page 92
ESG-400

2-8-4 Measurement of the bipolar high frequency leakage current under unloaded condition

Fig. 5.2.10 and Fig. 5.2.11 are showing an example how to set up the test to measure the bipolar high frequency leakage current under unloaded condition. In this test the output is unloaded, the load resistance used in the test before is removed. The high frequency leakage current is floating from appropriate pole to the equipotential bonding port. It is measured via the electrosurgical analyzer via a measuring resistance of 200 . Fig. 5.2.10 shows the measurement from terminal 1; Fig. 5.2.11 shows the measurement from terminal 2.
9) Assign the footswitch to the BIPOLAR socket and select the BipolarCut mode with effect 3. Set the power
level to 100. (Perform the setting according to the instruction for use.)
10) Set up the measurement according to Fig. 5.2.10 (Measuring of the high frequency leakage current under
unloaded condition at terminal 1.)
11) Activate the output power by pressing the corresponding footswitch pedal.
12) Verify the leakage current is ≤ 70 mA.
13) Set up the measurement according to Fig. 5.2.11 (Measuring of the high frequency leakage current under
unloaded condition at terminal 2.)
14) Activate the output power by pressing the corresponding footswitch pedal.
15) Verify the leakage current is ≤ 70 mA.
16) Record the test results in the Inspection Card (see chapter 5-3 ).
7.022.211 / ISSUE 5 92 / 110 Chapter 5: Inspection
Page 93
ESG-400
Fig. 5.2.9. Example for the unloaded measurement of the bipolar high frequency leakage current at terminal 1
Fig. 5.2.10. Example for the unloaded measurement of the bipolar high frequency leakage current at terminal 2
7.022.211 / ISSUE 5 93 / 110 Chapter 5: Inspection
Page 94
ESG-400
For the measurement of this (high current) value, it is important, that the cable

2-9 Checking the output power

1) All constraints for these tests are listed in the inspection card in chapter 5-3 under number 31 to 47. The constraints are the output socket, the way of activation (e.g. footswitch, handswitch), the settings and the load resistance. For measuring and to simulate the appropriate resistance the electrosurgical analyzer (J002) is used.
2) Connect the appropriate output of the electrosurgical generator to the variable load of the electrosurgical analyzer (J002) (according to the inspection card in chapter 5-3). Set the electrosurgical generator to continuous measuring and set t he internal mea suri ng resistance to the appropriate value (according to the inspection card in chapter 5-3). Start the measuring with the electrosurgical generator. (Settings o f the electrosurgical analyzer are done according to the analyzer’s instruction for use.)
3) Select the mode and set the power level according to the inspection card i n chapter 5-3.
4) Activate the output power by pressing the corresponding footswitch pedal or handswitch button. During the activated output the output tone can be heard.
5) Verify the output power is in the range according to the inspection card in chapter 5-3.
6) Repeat step 2 - 5 until test number 30 to 45 in the inspection card in chapter 5-3 have been checked.
7) Record the test results in the Inspection Card (see chapter 5-3).
NOTE
of the instrument is not “forming an inductor”. So do not wrap it up like a coil. Either use it flat or unfold it completely.

2-10 Checking for certain features and error messages

1) All constraints for these tests are listed in the inspection card in chapter 5-3 under number 48 to 50. The constraints are the output socket, the way of activation (e.g. footswitch, handswitch), the settings and the load resistance. Low inductive high load resistors (J005) are used.
2) Connect the appropriate output of the electrosurgical generator to the appropriate load resistor (according to the inspection card in chapter 5-3).
3) Select the mode and set the power level according to the inspection card in chapter 5-3.
4) Activate the output power by pressing the corresponding footswitch pedal. During the activated output the output tone can be heard.
5) Verify the r eaction of the electrosurgical generator according to the inspection car d in chapter 5-3.
6) Repeat step 2 - 5 until test number 46 to 48 in the inspection card in chapter 5-3 have been checked.
7) Record the test results in the Inspection Card (see chapter 5-3).
7.022.211 / ISSUE 5 94 / 110 Chapter 5: Inspection
Page 95
ESG-400
Check the functionality of each push button
Record the test results in the Inspection
MENU
Fig. 5.2.11. Checking volume and brightness

2-11 Final test

2-11-1 Self test

1) Connect the electrosurgical generator via power cord (J016) to mains. Switch the generator on by pressing the “Power Switch”.
2) The generator should boot without any alarm. If an alarm occurs, refer to Chapter 4 (Troubleshooting).

2-11-2 Display and sound check

1) Check the adjustment of display brightness and volume via touch screen. Enter the “Select Menu” by pressing the push button “MENU push button”. Change the Volume via the roc ker switch. A sound should be carried according to the setting. Change the Brightness via the rocker switch. The brightness should change according to the setting.
2) Check the adjustment of the volume via the volume control knob at the rear of the housing.
There will be an acoustic feedback for the setting.
3) Record the test results in the Inspection Card (refer to chapter 5-3).

2-11-3 Functionality of push buttons

1) by pressing.
2) The buttons should perform according to the description in the instruction for use.
3) Card (refer to chapter 5-3).

2-11-4 Communication test

1) Connect the sockets “LINK-IN” and “LINK-OUT” at the rear of the housing by the Communication Cable
7.022.211 / ISSUE 5 95 / 110 Chapter 5: Inspection
Fig. 5.2.12. Position of the push buttons
Page 96
ESG-400
(J012) with each other. In this way a ring communication is established.
2) The Error Message “E394 Communication error” should appear.
3) Record the test results in the Inspection Card (refer to chapter 5-3).

2-11-5 Restore of output power settings

Restore the settings by using the tables created in chapter 7-2-1. When not done while inspecting, restore all power settings off the changed modes noted in table 7.2. Finally restore the las t active procedure if applic able and assign all modes to the sockets.

2-12 Inspection label (For applicable markets)

1) Cover or exchange an inspection label (as shown in Fig. 5.2.14) at the rear panel of the electrosurgical unit’s housing and mark the due date of the next periodic safety check (month / year). The electrosurgical unit must undergo a periodic safety check at annual intervals.
2) Record the due date in the safety check report (see chapter 5-3).
Fig. 5.2.13. Sample of an inspection label
7.022.211 / ISSUE 5 96 / 110 Chapter 5: Inspection
Page 97
ESG-400
Obstructions or damage of

3 Inspection Card

Product name:
REF no.:
Serial no.:
Software version:
Service / Maintenance manual version:
Nr. Test type / mode Pwr.
level
Effect Yes No
Visual inspection of the electrosurgical unit and accessories
1 Labels N/A N/A Procedure passed N/A
2
unit and accessories
N/A N/A Procedure passed N/A
Load
resistance
Requirement Measured
value
Test passed
Contact quality monitor function
3 Any bipolar mode N/A N/A Procedure passed N/A
4
5 Standby, split N/A 140 Ω Procedure passed N/A
6 Standby, split N/A 170 Ω Procedure passed N/A
7
8 DC resistance N/A N/A ≥ 2.0 MΩ
9
10 Normal condition (NC) N/A N/A 0.5 mA
Any monopolar mode
DC Resistance
Earth resistance (with power cord)
Earth leakage current
Standby, non-split
Activation, split
N/A 0 Ω, 5 Ω Procedure passed N/A
N/A 170 Ω Procedure passed N/A
N/A N/A 0.3 Ω
11 Single fault condition (SFC) N/A N/A 1.0 mA
7.022.211 / ISSUE 5 97 / 110 Chapter 5: Inspection
Page 98
ESG-400
Single fault condition (SFC),
Single fault condition (SFC),
Nr. Test type / mode Pwr.
level
Effect Yes No
Patient leakage current
12 Normal condition (NC), AC N/A N/A 0.01 mA
13 Normal condition (NC), DC N/A N/A 0.01 mA
14
15
16 PureCut
17 Amplitude fluctuation
AC
DC
Current and power consumption and output waveform
N/A N/A 0.05 mA
N/A N/A 0.05 mA
300
Eff. 3
300
Eff. 3
Load
resistance
500 Ω IL 10 A
500 Ω Δ ≤ 110 V
Requirement Measured
value
Test passed
18 SalineCut
High frequency leakage current
19
20
21
22
23
24
Monopolar PureCut (neutral electrode term inal)
Monopolar PureCut (active electrode terminal)
Monopolar SprayCoag (neutral electrode term inal)
Monopolar SprayCoag (active electrode terminal)
Monopolar PureCut (neutral electrode term inal)
Monopolar PureCut (active electrode terminal)
320
75 Ω SL 1500 VA
Eff. 3
300
500 Ω 150 mA
Eff. 3
300
500 Ω 150 mA
Eff. 3
120
500 Ω 150 mA
Eff. 3
120
500 Ω 150 mA
Eff. 3
300
unloaded 150 mA
Eff. 3
300
unloaded 150 mA
Eff. 3
25
26
7.022.211 / ISSUE 5 98 / 110 Chapter 5: Inspection
Monopolar SprayCoag (neutral electrode term inal)
Monopolar SprayCoag (active electrode terminal)
120
unloaded 150 mA
Eff. 3
120
unloaded 150 mA
Eff. 3
Page 99
ESG-400
Nr. Test type / mode Pwr.
level
Effect Yes No
27
28
29
30
BipolarCut (active electrode terminal 1)
BipolarCut (active electrode terminal 2)
BipolarCut (active electrode terminal 1)
BipolarCut (active electrode terminal 2)
Eff. 3
Eff. 3
Eff. 3
Eff. 3
Output power (all versions)
Monopolar PureCut
31
(hand switch@ MONOPOLAR 1)
Eff. 3
100
100
100
100
300
Load
resistance
Requirement Measured
value
Test passed
500 Ω 70 mA
500 Ω 70 mA
unloaded 70 mA
unloaded 70 mA
500 Ω 244 W ≤ P
366 W
out
32
33
34
35
36
37
Monopolar SprayCoag (hand switch @ MONOPOLAR 1)
Monopolar BlendCut (hand switch @ MONOPOLAR 2)
Monopolar BlendCut (hand switch @ MONOPOLAR 2)
Monopolar PowerCoag (hand switch @ MONOPOLAR 2)
Monopolar PowerCoag (hand switch @ MONOPOLAR 2)
Monopolar SoftCoag (foot switch @ MONOPOLAR 2 with J007 without J010)
120
Eff. 3
200
Eff. 3
200
Eff. 3
120
Eff. 3
60
Eff. 3
200
Eff. 3
500 Ω 93 W ≤ P
500 Ω 154 W ≤ P
2000 Ω 49 W ≤ P
500 Ω 93 W ≤ P
500 Ω 48 W ≤ P
50 Ω 154 W ≤ P
139 W
out
232 W
out
73 W
out
139 W
out
72 W
out
232 W
out
38
Monopolar ForcedCoag (foot switch @ MONOPOLAR 2 with J007 without J010)
120
Eff. 3
500 Ω 93 W ≤ P
139 W
out
7.022.211 / ISSUE 5 99 / 110 Chapter 5: Inspection
Page 100
ESG-400
Nr. Test type / mode Pwr.
level
Effect Yes No
BipolarCut
39
(double foot switch @ BIPOLAR)
Eff. 3
Bipolar AutoCoag (single foot switch (if
40
provided with the generator, otherwise double footswitch)
Eff. 3
@ BIPOLAR) Bipolar AutoCoag
41
(double foot switch @ BIPOLAR)
Eff. 3
Bipolar RFCoag
42
(without RCAP) (foot switch @ BIPOLAR)
RCAP
Bipolar RFCoag
43
(without RCAP) (foot switch @ BIPOLAR)
RCAP
100
120
120
50
w/o
50
w/o
Load
Requirement Measured
resistance
500 Ω 82 W ≤ P
25 Ω 48 W ≤ P
75 Ω 94 W ≤ P
75 Ω 39 W ≤ P
1000 Ω 0 W ≤ P
value
122 W
out
72 W
out
140 W
out
59 W
out
5 W
out
Test passed
44
45
Bipolar FineCoag (foot switch @ BIPOLAR)
Bipolar HardCoag (foot switch @ BIPOLAR)
46 Bipolar SalineCoag
(foot switch @ UNIVERSAL with J008)
39
Eff. 1
42
Eff. 1
200
Eff. 3
Or (hand switch @ UNIVERSAL
with J059) SW < 4.09 AE/NE
(Pin1) and NE (Pin6)
SW ≥ 4.09 AE/NE
(Pin2) and NE (Pin6)
Output power (only for software version ≥ 4.09)
47 Bipolar PK PureCut
(handswitch @ UNIVERSAL with J059 between AE (Pin1)
200
Eff. 3
and AE/NE (Pin2))
50 Ω 35 W ≤ P
500 Ω 0 W ≤ P
154 W ≤ P
Message “Use default
75 Ω
instrument settings? (UNIVERSAL socket)” can be acknowledged.
200 167 W ≤ P
43 W
out
12 W
out
230 W
out
249 W
out
7.022.211 / ISSUE 5 100 / 110 Chapter 5: Inspection
Loading...