Ohaus TROOPER COMPACT COUNTING SCALES User Manual [en, es, fr]

Trooper Count Scale Instruction Manual
Balanza Trooper Count Manual de Instrucciones
Balance Trooper Count Guide d'Utilisation
Ohaus Corporation, 19A Chapin Road, P.O. Box 2033 Pine Brook, New Jersey, 07058, USA
0103
M
Declaration of Conformity We, Ohaus Corporation, declare under our sole responsibility that the balance models listed below marked with CE - are in conformity with the directives and standards mentioned. Konformitätserkärung Wir, die Ohaus Corporation, erklären in alleiniger Verantwortung, dass die untenstehenden Waagentypen, gekennzeichnet mit CE - mit den genannten Richtlinien und Normen übereinstimmen. Déclaration de conformité Nous, Ohaus Corporation, déclarons sous notre seule responsabilité, que les types de balance ci-dessous cité ­munis de la mention «CE» - sont conformes aux directives et aux normes mentionnées ci-après. Declaración de Conformidad Nosotros, Ohaus Corporation, declaramos bajo responsabilidad exclusiva que los modelos de balanzas indicados a continuación - con el distintivo ,CE - están conformes con las directivas y normas citadas. Dichiarazione di conformità Noi, Ohaus Corporation, U.S.A, dichiariamo sotto nostra unica responsabilità, che i tipi di bilance specificati di seguito - contrassegnati con la marcatura CE - sono conformi alle direttive e norme citate.
Instrument Type/Waagentyp/Type de instrument/Tipo de instrumento/Tipo di strumento: Trooper Count Series
Marked with: Directive Standard Gekennzeichnet mit: Richtlinie Norm Munis de la mention: Directive Norme Con el distintivo: Directiva Norma Contrassegnati con la Direttiva Norma marcatura:
EU 73/23/EEC EN61010-1:1993 + A2: 1995
Low Voltage Safety Regulations Niederspannung Sicherheitsbestimmungen Basse tension Consignes de sécurité Baja tensión Disposiciones sobreseguridad Bassa tensione
Prescrizioni di sicurezza
EU 89/336/EEC EN55011: 1991 (class B) Emissions; EN61000-3-2 Electromagnetic compatibility EN50082-2:1995 Immunity; EN61000-3-3 Elektromagnetische EN55011: 1991 (class B) Funkstörungen; EN61000-3-2 Verträglichkeit EN50082-2:1995 Immunität; EN61000-3-3 Compatibilité électromagnétique EN55011: 1991 (class B) Emissions parasites; EN61000-3-2 Compatibilidad electromagnética EN50082-2:1995 Immunité; EN61000-3-3 Compatibilità elettromagnetica EN55011: 1991 (class B) Radiointerferencias; EN61000-3-2
EN50082-2:1995 Inmunidad; EN61000-3-3 EN55011: 1991 (class B) Radiointerferenze; EN61000-3-2 EN50082-2:1995 Immunità; EN61000-3-3
EU 90/384/EEC EN45501 1)
NAWI Non Automatic Weighing Instruments FNSW Für nicht selbsttätige Waagen BFNA Balances à fonctionnement non automatique PBNA Para balanzas no automátäcas BFNA Per bilance a funzionamento non automatics
1) Applies only to certified non-automatic weighing instruments Betrifft nur zertifizierte nicht selbsttätige Waagen Sapplique uniquement aux instruments de pesage à fonctionnement non automatique approuvés Applicabile solamente a strumenti di pesatura a funzionamento non automatico Aplicable solamente a instrumentos de pesaje aprobados de funcionamiento no automático
Date: March 28, 2003
Ted Xia
President
Ohaus Corporation
Pine Brook, NJ USA
Johan Dierbach
General Manager
Ohaus Europe
Greifensee, Switzerland
ISO 9001 Registration for Ohaus Corporation. Ohaus Corporation, USA, was examined and evaluated in 1994 by the Bureau Veritas Quality International, BVQI, and was awarded ISO 9001 registration. This certifies that Ohaus Corporation, USA, has a quality system that conforms with the international standards for quality management and quality assurance (ISO 9000 series). Repeat audits are carried out by BVQI at intervals to check that the quality system is operated in the proper manner.
ISO 9001-Zertifikat für Ohaus Corporation. Die Firma Ohaus Corporation, USA, wurde 1994 durch das Bureau Veritas Quality Interna- tional BVQI geprüft, und erhielt das ISO 9001 Zertifikat. Dieses bescheinigt, dass Ohaus Corporation, USA über ein Qualitätssystem verfügt, welches den internationalen Normen für Qualitätsmanagement und Qualitátssicherung (ISO 9000er-Reihe) entspricht. Anlässlich von Wiederhol-Audits durch das BVQI wird periodisch überprüft, ob das Qualitätssystem zweckmässig gehandhabt wird.
Certificat ISO 9001 pour Ohaus Corporation. La société Ohaus Corporation, USA, a été contrôlée en 1994 par Bureau Veritas Quality International BVQI et a obtenu le certificat, degré ISO 9001. Celui-ci atteste que Ohaus Corporation, USA, dispose dun système qualité correspondant aux normes internationales pour la gestion de la qualité et pour Iassurance qualité (degré ISO 9000). Des audits réguliers effectués par la BVQI vérifient si le système qualité est appliqué de facon appropriée.
Certificado ISO 9001 para Ohaus Corporation. La firma Ohaus Corporation, USA, ha sido inspeccionada por la Bureau Veritas Quality International (BVQI) y ha obtenido el certificado ISO 9001. Esto acredita que Ohaus Corporation, USA, dispone de un sistema de calidad que cumple las normas internacionales para gestión y garantfa de calidad (ISO serie 9000). Con ocasión de las inspecciones de repetibilidad por parte de la BVQI, se comprueba periódicamente si el sistema de calidad se manipula de forma correcta.
Certificato ISO 9001 per la Ohaus Corporation. ll sistema di garanzia della qualità della Società Ohaus Corporation, USA è certificato ISO 9001 sin dal 1994 dal Bureau Veritas Quality International BVQI, e così fomisce la dimostrazione che il suo sistema di Garanzia Qualità soddisfa i massimi requisiti. Verifiche periodiche del BVQI garantiscono che il sistema qualità opera correttamente.
NOTE: THIS EQUIPMENT HAS BEEN TESTED AND FOUND TO COMPLY WITH THE LIMITS FOR A CLASS A DIGITAL DEVICE, PURSUANT TO PART 15 OF THE FCC RULES.
THESE LIMITS ARE DESIGNED TO PROVIDE REASONABLE PROTECTION AGAINST HARMFUL INTERFERENCE WHEN THE EQUIPMENT IS OPERATED IN A COMMERCIAL ENVIRONMENT. THIS EQUIPMENT GENERATES, USES, AND CAN RADIATE RADIO FREQUENCY ENERGY AND, IF NOT INSTALLED AND USED IN ACCORDANCE WITH THE INSTRUCTION MANUAL, MAY CAUSE HARMFUL INTERFERENCE TO RADIO COMMUNICATIONS. OPERATION OF THIS EQUIPMENT IN A RESIDENTIAL AREA IS LIKELY TO CAUSE HARMFUL INTERFERENCE IN WHICH CASE THE USER WILL BE REQUIRED TO CORRECT THE INTERFERENCE AT HIS OWN EXPENSE.
THIS DIGITAL APPARATUS DOES NOT EXCEED THE CLASS A LIMITS FOR RADIO NOISE EMISSIONS FROM DIGITAL APPARATUS AS SET OUT IN THE INTERFERENCE-CAUSING EQUIPMENT STANDARD ENTITLED DIGITAL APPARATUS, ICES-003 OF THE DEPARTMENT OF COMMU­NICATIONS.
CET APPAREIL NUMERIQUE RESPECTE LES LIMITES DE BRUITS RADIOELECTRIQUES APPLICABLES AUX APPAREILS NUMERIQUES DE CLASSE A PRESCRITES DANS LA NORME SUR LE MATERIEL BROUILLEUR : APPAREILS NUMERIQUES, NMB-003 EDICTEE PAR LE MINISTRE DES COMMUNICATIONS.
Unauthorized changes or modifications to this equipment are not permitted.
TABLE OF CONTENTS
1. GETTING TO KNOW YOUR TROOPER COUNT SCALE....................................................................................................... 3
1.1 Introduction .............................................................................................................................................................. 3
1.1.1 Features ......................................................................................................................................................... 3
2. INSTALLATION ........................................................................................................................................................... 3
2.1 Unpacking and Checking ............................................................................................................................................ 3
2.2 Selecting the Location ................................................................................................................................................ 4
2.3 Connecting Power ..................................................................................................................................................... 4
2.3.1 Battery Installation ............................................................................................................................................ 4
2.3.2 AC Power......................................................................................................................................................... 4
2.4 Leveling the Scale ...................................................................................................................................................... 4
2.5 Stabilization ............................................................................................................................................................. 4
3. OPERATION............................................................................................................................................................... 5
3.1 Overview of Controls and Display Functions ................................................................................................................. 5
3.2 Basic Weighing ......................................................................................................................................................... 7
3.2.1 Turning On the Scale......................................................................................................................................... 7
3.2.2 Turning Off the Scale ......................................................................................................................................... 7
3.2.3 Zero Operation ................................................................................................................................................. 7
3.2.4 Weighing ......................................................................................................................................................... 7
3.2.5 Manual Tare Operation ...................................................................................................................................... 8
3.2.6 Pre-Set Tare Operation ...................................................................................................................................... 8
3.2.7 Gross/Tare Recall Operation............................................................................................................................... 8
3.2.8 Unit Switch Operation ........................................................................................................................................ 8
3.3 Basic Counting .......................................................................................................................................................... 9
3.3.1 Positive Sampling............................................................................................................................................. 9
3.3.2 Negative Sampling ........................................................................................................................................... 9
3.3.3 Positive Counting............................................................................................................................................ 10
3.3.4 Negative Counting .......................................................................................................................................... 10
3.3.5 Using Keypad to Edit Counting Data ................................................................................................................. 11
3.4 Library Mode ........................................................................................................................................................... 11
3.4.1 Library Operation ............................................................................................................................................ 11
3.4.2 Entering Data Directly into Library..................................................................................................................... 11
3.4.3 Saving Library Data ........................................................................................................................................ 13
3.4.4 Saving Library Data in Another Location ............................................................................................................ 13
3.4.5 Using the Library ............................................................................................................................................ 14
3.5 Auto Optimization .................................................................................................................................................... 14
3.6 Auto Add Mode ........................................................................................................................................................ 14
1
TABLE OF CONTENTS (Cont.)
3.7 RS232 Communications .......................................................................................................................................... 15
3.7.1 RS232 User Command Table........................................................................................................................... 15
3.7.2 Output Formats .............................................................................................................................................. 15
3.7.3 RS-232 Pin Out.............................................................................................................................................. 15
4. SETUP............ ......................................................................................................................................................... 16
4.1 Setup Protection ...................................................................................................................................................... 16
4.2 Menu Structure ........................................................................................................................................................ 16
4.3 Control Functions .................................................................................................................................................... 17
4.3.1 Menu Operation.............................................................................................................................................. 17
4.4 Checking the Menu Settings ...................................................................................................................................... 17
4.5 Setup Menu ............................................................................................................................................................ 18
4.6 Read Menu ............................................................................................................................................................. 19
4.7 Print Menu ............................................................................................................................................................. 19
4.8 Count Menu ............................................................................................................................................................ 21
4.9 Lockout Menu ......................................................................................................................................................... 22
5. CALIBRATION AND SEALING ...................................................................................................................................... 23
5.1 Span Calibration ...................................................................................................................................................... 24
5.2 Linearity Calibration ................................................................................................................................................. 25
5.3 Legal for Trade Operation and Sealing ....................................................................................................................... 27
6. CARE AND MAINTENANCE ........................................................................................................................................ 28
6.1 Troubleshooting ...................................................................................................................................................... 28
6.2 Error Codes List ....................................................................................................................................................... 29
6.3 Service Information .................................................................................................................................................. 29
6.4 Accessories ............................................................................................................................................................ 29
7. TECHNICAL DATA ..................................................................................................................................................... 28
7.1 Counting Limits ....................................................................................................................................................... 28
7.2 Specifications.......................................................................................................................................................... 30
2
Trooper Count Scale
1. GETTING TO KNOW YOUR TROOPER COUNT SCALE
1.1 Introduction
Thank you for deciding to purchase a Trooper Count Scale from Ohaus. The Ohaus Trooper Count Scale is a rugged, reliable, electronic counting/weighing scale designed for easy operation.
Behind your instrument stands Ohaus, a leading manufacturer of precision Indicators, Scales and Balances. An After­market Department with trained instrument technicians is dedicated to providing you with the fastest service possible in the event your instrument requires servicing. Ohaus also has a Customer Service Department to answer any inquiries regarding applications and accessories.
To ensure you make full use of the possibilities offered by your Trooper Count Scale, please read the manual completely before installation and operation.
1.1.1 Features
Major features include:
 Three backlit LCDs  Durable cast aluminum internal frame, stainless steel weighing pan, ABS housing  21 button operation with numeric keypad  Flexible unit switching-lb/kg/oz/g  RS232 serial communication interface  Battery operation with internal 6 C-cell batteries  AC power adapter  Auto shut off for power saving  Library feature saves counting data for 60 items. (Part number, APW, tare values, over/under values,
and count totals.)
 Positive or negative sampling and counting
2. INSTALLATION
2.1 Unpacking and Checking
Open the package and remove the instrument and the accessories. Check the instrument for transport damage. Imme­diately inform your Ohaus dealer if you have complaints or if parts are missing. Your Trooper Count Scale package should contain:
 Trooper Count Scale  Instruction Manual  Weighing Platform  Warranty card  AC power adapter  Lead seal kit and capacity labels
Store all parts of the packaging. This packaging ensures the best possible protection for the transport of your instrument.
NOTE: Remove the two shipping screws if present on top of the unit.
3
Trooper Count Scale
2.2 Selecting the Location
The Scale should be used in an environment which is free from dust, water, corrosives, vibration or temperature extremes as these factors can affect scale accuracy. The Scale should be located on a stable level surface and kept away from vibrating sources such as large machinery.
2.3 Connecting Power
2.3.1 Battery Installation (If desired)
Remove battery cover and insert 6 Alkaline C-type batteries into the battery holder making sure that the batteries are properly orientated. The negative battery terminals should face the spring contacts.
NOTE: It is recommended that when the Trooper Count Scale is operated from batteries, the Auto-Off Timer feature be
turned on to extend battery life. When setting up the Scale, refer to the Setup menu, section 4.5.
2.3.2 AC Power
Verify that the AC power source matches the AC adapter rating. Connect the Adapter power cord at the rear of the Scale and plug the Adapter into a convenient power outlet.
2.4 Leveling the Scale
To compensate for small irregularities or inclinations at the location, the Scale can be leveled. The Scale is equipped with a level indicator located at the rear. Adjust the leveling feet until the air bubble in the indicator is centered as shown.
NOTE: The instrument should be leveled each time its location is changed.
2.5 Stabilization
Before initially using the Scale, allow time for it to adjust to its new environment. The recommended warm up period is five (5) minutes if the Scale is already stabilized to room temperature.
4
3. OPERATION
3.1 Overview of Controls and Display Functions
Trooper Count Scale
1
ozkg g lb Net B/G %Ac T PT APW MR M < > PCS
ON/ZERO
Off
14
23 4
The counting feature is not legal for trade
DATAWEIGHT COUNT
PRINT
Units
SAMPLE
Menu
1213
LIBRARY
Save
11
TARE
10
No. Designation Function
1 Weight Display Displays current weight on scale.
789 456 123
0.
Data
Clear
M+
M-
Enter
5 6
7 8
9
Indicators identify active weight unit and Net or Gross weight:
kg (Kilogram)  Unit of measure. g (Gram) - Unit of measure. lb (Avoirdupois Pound) - Unit of measure. oz (Avoirdupois Ounce) - Unit of measure. Net (Net weight)  Gross weight minus Tare.
2 Data Display Displays user selected data.
Indicators show active data mode:
B/G (Brutto/Gross)  Total weight of items. %AC (Percent Count accuracy)  Estimated accuracy of the count based on sample
size and scale resolution.
T (Tare value)  Weight of the container . PT (Preset Tare value)  Tare value recalled from memory or entered via keypad. APW (Average Piece Weight)  Average weight of one item. MR (Memory Recall)  Accumulated quantity stored in memory.
3 Count Display Displays current count of items on the scale.
Indicators show count status:
M (Memory)  Accumulated quantity in memory. < (Under)  Current quantity is less than lower limit. > (Over)  Current quantity is more than upper limit. < > (Target range)  Current quantity is between upper and lower limits. PCS (Pieces)  A count value is being shown.
5
Trooper Count Scale
3.1 Overview of Controls and Display Functions (Cont.)
No. Designation Function
4 Numeric Keypad Enter specific values for tare, APW and other data entries.
5 Data button Selects between data modes in the Data display.
In edit mode, serves as a Go Back or backspace key.
6 Clear button During editing, clears the blinking entry.
7 M+ button Adds the displayed quantity into the accumulating memory.
During editing, selects the next higher option or number.
8 M- button Subtracts the keyboard entry from the accumulating memory. In sampling mode,
enters negative sampling.
During editing, selects the next lower option or number.
9 Enter button Enters edit function of current mode. If current mode does not support edit func-
tions, an error beep will occur.
In edit mode, accepts entry.
10 TARE button Tares the weight on the pan or accepts the keypad tare entries.
11 LIBRARY Save button Short press displays Library ID (blinking) and the part number. Repeat press
returns to weigh/count mode.
Long press (3 seconds) saves library changes.
12 SAMPLE Menu button Short press enters the sampling mode.
Long press (3 seconds) enters scale setup menu. In menu mode, a short press will return to weigh mode.
13 PRINT Units button Short press sends data to RS232 port.
Long press (3 seconds) steps through active weigh units. Release on desired unit.
14 ON/ZERO Off button Short press turns the scale on or zeros the display.
Long press (3 seconds) turns the scale off.
6
Trooper Count Scale
3.2 Basic Weighing
The Trooper Count Scale has been factory calibrated and is ready for operation. Ohaus recommends that the Scale be re-calibrated before use to ensure optimal accuracy. You can operate the scale at this point using the factory default settings. You can check the default menu settings on page 14. All bolded items on the menu are the factory default settings. The scale will operate in the default mode. If you want to change the settings, continue with Section 4, Setup.
When the Scale is positioned for operation, follow the operational procedure outlined next.
3.2.1. Turning On the Scale
Press and hold ON/ZERO Off button until the LCD display appears. The display cycles as shown below displaying "Ohaus", with model number, serial num­ber, software revision number and weighing mode display.
3.2.2 Turning Off the Scale
To turn the Scale off, press and hold the ON/ZERO Off button until OFF is displayed.
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
Net
Net
Net
Net
Net
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
MR
APW
PT
MR
APW
PT
APW
PT
APW
PT
APW
PT
M
MR
MR
MR
<>
M
<>
<>
M
<>
M
M
<>
PCS
PCS
PCS
PCS
PCS
3.2.3 Zero Operation
Press the ON/ZERO Off button to zero the weight dis- play. The gross weight is also shown in the DATA display.
NOTE: A pre-set tare value will not be zeroed.
3.2.4 Weighing
Place item to be weighed on the Scale. The display indicates a sample of 1.5kg, gross weight.
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
Net
Net
B/G %Ac
B/G %Ac
DATA COUNT
T
DATA COUNT
T
PT
MR
APW
PT
APW
MR
M
<>
M
<>
PCS
PCS
7
Trooper Count Scale
3.2.5 Manual Tare Operation
When weighing a sample that must be held in a container, taring stores the container weight memory.
Place the container on the Scale. Display shown is
0.5kg.
Press TARE button. The container is tared, the Net weight is shown in the WEIGHT display, and the Tare weight is shown in the DATA display.
To clear the tare with an empty pan, press the TARE button.
3.2.6 Pre-Set Tare Operation
Pre-Set Tare is a tare value that is entered using the numeric keypad. To use Pre-Set Tare, repeately press the Data button until the PT indicator is shown in the DATA display. The DATA display indicates OFF.
Press the Enter button to toggle the Pre-Set Tare on or off. The current Pre-Set Tare value (if any) is displayed. Enter a numeric value using the keypad, (the COUNT display is blinking), then press the Enter button. The Net weight is shown in the WEIGHT display and the tare weight is shown in the DATA display. Display shown is 0.5kg.
Alternate Method:
Pre-Set Tare can also be entered using the numeric keypad, then press the TARE button.
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
Net
Net
Net
Net
Net
Net
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
MR
APW
PT
MR
APW
PT
APW
PT
APW
PT
APW
PT
APW
PT
M
MR
MR
MR
MR
<>
M
<>
M
<>
M
<>
M
<>
M
<>
BLINKING
PCS
PCS
PCS
PCS
PCS
PCS
NOTE: The Pre-Set Tare value is overwritten when a new value is entered.
3.2.7 Gross/Tare Recall Operation
Repeatedly press the Data button to cycle through Gross and Tare readings.
3.2.8 Unit Switch Operation
Press and hold PRINT Units button until display changes to selected measuring unit. Display indica­tors at the bottom of the WEIGHT display indicate the active weighing unit. Depending on which units are enabled in the Read menu, you have a choice of g, lb, kg or oz. The display sample indicates a 1kg load changed to lb.
8
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
Net
Net
Net
B/G %Ac
B/G %Ac
B/G %Ac
T
DATA COUNT
T
DATA COUNT
T
MR
APW
PT
APW
PT
APW
PT
MR
MR
M
M
M
<>
<>
<>
PCS
PCS
PCS
Trooper Count Scale
3.3 Basic Counting
Read this procedure entirely before counting pieces. Piece sampling and counting can be further enhanced using menu selections in the Count menu. There are several methods that are available for sampling.
 Positive Sampling - given number of pieces are placed on the scale, the weight is divided by the number of pieces to calculate the Average Piece Weight (APW). Additional pieces are then placed on the scale and counted and dis­ played.
 Negative Sampling - calculates the APW when a given number of pieces are removed from the total quantity.
 The APW can also be entered through the numeric keypad. As parts are placed on the scale, the count is immediately displayed.
 The Trooper Count will support positive counting or negative counting.
3.3.1 Positive Sampling
If a weighing container is to be used:  If the tare value is known, enter a Pre-Set Tare.  If the tare value is not known, place the container on the pan and press the TARE button. Press the SAMPLE Menu button.
WEIGHT
kg
g
oz
lb
Net
B/G %Ac
DATA COUNT
T
MR
APW
PT
<>
M
PCS
WEIGHT
kg
g
oz
lb
Net
B/G %Ac
DATA COUNT
T
MR
APW
PT
<>
M
PCS
COUNT
<>
M
PCS
COUNT
<>
M
PCS
The display indicates to place 10 samples on the scale. Place the desired number of sample pieces in the container. If you have placed 10 pieces in the container, press the Enter button, or by pressing the numeric keys you may directly enter an alternate sample size. Increased sample size results in better results. If you make a mistake, press the Clear button and start again. Press the Enter button. The example indicates a sample size of 10 pieces.
WEIGHT
DATA COUNT
NOTES:
1. If APW is less than minimum, [Err3] [LOW]
kg
g
oz
lb
Net
B/G %Ac
T
MR
APW
PT
<>
M
[APW] is displayed. See section 6.2.
PCS
2. If Ref. Wt. is less than minimum, [Err4] [LOW] [REF] is displayed. See section 6.2.
3.3.2 Negative Sampling
Negative sampling permits the determination of the total quantity on the pan by removing a known number of sample pieces. The total quantity may be placed directly on the Scale or in a container. The Trooper Count Scale automatically calculates the remaining pieces after the sample amount is removed. When a container is used, you may enter a preset tare weight or simply tare the empty container before starting.
9
Trooper Count Scale
3.3.2 Negative Sampling (Cont.)
If a weighing container is to be used:  If the tare value is known, enter a Pre-Set Tare.  If the tare value is not known, place the container on the pan and press the TARE button.
Press the SAMPLE Menu button.
WEIGHT
DATA COUNT
COUNT
COUNT
kg
g
oz
lb
Net
B/G %Ac
T
MR
APW
PT
<>
M
PCS
<>
M
PCS
<>
M
PCS
Press the M- button.
WEIGHT
kg
g
oz
lb
Net
B/G %Ac
DATA COUNT
T
MR
APW
PT
<>
M
PCS
COUNT
<>
M
PCS
COUNT
<>
M
PCS
Place the total number of pieces on the pan, then press the Enter button.
WEIGHT
kg
g
oz
lb
Net
B/G %Ac
DATA COUNT
T
MR
APW
PT
<>
M
PCS
COUNT
<>
M
PCS
COUNT
<>
M
PCS
Remove 10 pieces as instructed, or use the keypad to change the number of pieces to be removed, then press Enter button.
WEIGHT
DATA COUNT
Place the pieces that were removed back on the scale to display the total number of pieces.
kg
g
oz
lb
Net
B/G %Ac
T
MR
APW
PT
<>
M
PCS
NOTE: If AUTO-ADD is ON, the Scale may direct you to remove additional pieces if the reference weight is too small.
3.3.3 Positive Counting
After sampling, place additional pieces on the scale and view the piece count in the COUNT display.
3.3.4 Negative Counting
After sampling, place the entire quantity of pieces on the scale. Press the ON/ZERO Off button. As pieces are removed, the quantity removed is shown in the COUNT display as a negative number.
10
Trooper Count Scale
BLINKING
3.3.5 Using Keypad to Edit Counting Data
The APW, Pre-Set Tare (PT) and the accumulated count (MR) can be viewed and edited in the DATA window. While in weigh mode repeatedly press the Data button until the APW indicator is lit. A new APW value may be entered via the keypad and accepted with an Enter press. PT and MR can also be similarly edited.
3.4 Library Mode
The Trooper Count scale allows the storage of part/counting data in 60 locations, designated ID 00 to ID 59. Each location includes: part number, Average Piece Weight (APW), a pre-set tare value, over and under count values and an accumulated piece count.
Data in the library is saved in a manner similar to a computer file. Temporary editing of the data will not affect the saved data. When desired, the temporary edits can be saved, allowing the scale to be turned off without losing the changes. Data in ID 00 may not be saved permanently but the data can be easily copied to another location.
The Library, ID 01 to ID 59, can be locked (see 4.9 Lockout Menu) to prevent unauthorized changes to the data. The accumulated count will not be locked allowing it to be updated. When turned on, the scale defaults to ID 00.
3.4.1 Library Operation
To enter the Library, momentarily press the LIBRARY Save button. A short press in Library mode will exit to weigh mode.  M+ and M- buttons advance through the blinking choices.  The Enter button accepts the blinking choice.  The Data button returns to the previous choice.  The Clear button will clear blinking data or back space during data entry.  Three rapid beeps indicates an error, an improper button press or data value.
3.4.2 Entering Data Directly into Library
The following example is used to illustrate how to enter data in a particular library location.
Short press on the LIBRARY Save button.
Short presses on the M+ or M- buttons will display library locations which are already used.
Using the numeric keys, enter the desired library num­ber. The sample shown is an empty location, ID 22. Press the Enter button to accept ID 22.
Press the Enter button, to edit the part number or press M+ to skip.
WEIGHT
kg
g
oz
lb
BLINKING
WEIGHT
kg
g
oz
lb
Net
Net
B/G %Ac
B/G %Ac
DATA COUNT
T
APW
PT
DATA COUNT
T
APW
PT
MR
MR
<>
M
<>
M
PCS
PCS
11
Trooper Count Scale
BLINKING
BLINKING
BLINKING
BLINKING
BLINKING
3.4.2 Entering Data Directly into Library (Cont.)
Using the numeric keys 0-9, enter the desired part number. For this example, we will use 1230000000. You will notice that the part number can be up to 12 digits and utilizes the DATA and COUNT display win­dows. Press the Enter button to confirm the entry.
Press the M+ button, the display advances to APW. Press the Enter button to access APW value.
Using the numeric keypad, enter the APW. Sample shown indicates an average piece weight of 2g. Press the Enter button to accept.
Press the M+ button to advance to setting the preset tare. Press the Enter button to access the tare value.
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
Net
Net
Net
Net
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
MR
APW
PT
MR
APW
PT
MR
APW
PT
MR
APW
PT
M
<>
M
<>
M
<>
M
<>
BLINKING
PCS
PCS
PCS
BLINKING
PCS
Using the keypad, enter the desired tare weight. The sample shown is 50g. Press the Enter button to ac-
cept.
Press the M+ button to advance to Under count. Press Enter button to access the under count value.
Using the numeric keypad, enter the desired under count limit. For this sample, we will use 95. Press the Enter button to accept.
Press M+ to advance to over count. In this sample, we are using 105 pieces.
Press Enter button to access over count value. Using the numerical keypad, enter the desired over count limit (must be greater than under count value). Press the Enter button to accept.
12
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
Net
Net
Net
Net
Net
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
MR
APW
PT
APW
PT
APW
PT
APW
PT
APW
PT
MR
MR
MR
MR
M
M
<>
M
M
<>
M
BLINKING
<>
BLINKING
<>
BLINKING
<>
PCS
PCS
PCS
PCS
PCS
BLINKING
3.4.2 Entering Data Directly into Library (Cont.)
Press M+ to advance to count. Count is the accumu­lated piece count in memory. Press Enter button to edit count value.
Trooper Count Scale
WEIGHT
kg
g
oz
lb
Net
B/G %Ac
DATA COUNT
T
MR
APW
PT
<>
M
PCS
Using the numeric keypad, enter the desired count. The sample illustrates 250 pieces. Press the Enter button to accept.
To exit library editing, momentarily press the Library
3.4.3 Saving Library Data
Temporary changes to a library location need to be permanently saved or the changes will be lost when the scale is turned off. The scale will recommend saving temporary changes when the library location is changed or when the scale is turned off.
To save the settings, use a long press on the Library Save button. The library location (ID 22) and UPDATE YES is displayed.
Press Enter button to accept save location. Press Enter button again to save changes.
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
BLINKING
WEIGHT
Net
Net
B/G %Ac
B/G %Ac
DATA COUNT
T
DATA COUNT
T
DATA COUNT
MR
APW
PT
MR
APW
PT
<>
M
BLINKING
<>
M
PCS
PCS
The UPDATE DONE display confirms a successful save.
3.4.4 Saving Library Data in Another Location
You have the option of copying all data from one li­brary location into another library location. In the following example, we will save data from library lo­cation 22 to 52. Library 22 is shown.
Make a long press on LIBRARY Save button until display indicates UPDATE.
Enter the desired new location directory using the nu­meric keypad or by pressing the M+ button to ad­vance to the next open location. The sample is loca­tion 52. Press the Enter button.
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
BLINKING
WEIGHT
kg
g
oz
lb
BLINKING
Net
Net
Net
Net
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
MR
APW
PT
MR
APW
PT
MR
APW
PT
MR
APW
PT
<>
M
M
M
BLINKING
<>
<>
<>
M
PCS
PCS
PCS
PCS
13
Trooper Count Scale
3.4.4 Saving Library Data in Another Location (Cont.)
Press the Enter button to save data.
UPDATE DONE, then the ID number with IN USE will
WEIGHT
kg
g
oz
lb
WEIGHT
Net
B/G %Ac
DATA COUNT
T
DATA COUNT
MR
APW
PT
<>
M
BLINKING
PCS
be displayed momentarily followed by the counting display.
kg
g
lb
WEIGHT
oz
Net
B/G %Ac
T
DATA COUNT
MR
APW
PT
<>
M
PCS
At this point, all data from Library 22 has been copied into Library location 52.
kg
g
oz
lb
Net
B/G %Ac
T
MR
APW
PT
<>
M
PCS
3.4.5 Using the Library
To enter the Library, momentarily press the LIBRARY Save button. Use the keypad or the M+ button to enter a location ID that you wish to use. Press the LIBRARY Save button again to select the new ID. ID XX IN USE will display indicating that the data stored in that memory is now active. If an APW and a Preset Tare was set, placing a sample filled container on the scale will give an immediate count.
If the Library is not locked, it is possible to edit the data in the current ID. The changes need not be saved if they were for temporary use. If the Library is locked, you must change to ID 00 for simple counting use.
3.5 Auto Optimization
Automatic reference optimization results in a more accurate count by increasing the reference weight without the need to count additional parts. A higher reference weight is important when there is a risk of inconsistent piece weights or if the reference weight is close to the minimum (See section 7.1).
Auto Optimization uses an initial APW to count additional pieces that are placed on the scale. After a few seconds, the message A-OPt appears in the Data display as the new higher reference weight is used to re-calculate the APW. The process can be repeated as long as the additional weight is less than the previous reference weight. Once this limit is exceeded Auto-Opt is turned off. This function can be activated or deactivated in the Count menu. The factory default setting is on.
3.6 Auto Add Mode
Auto Add Mode is setup for a more accurate count procedure by increasing the minimum reference weight and the minimum APW limits (see section 7.1). The scale requests additional pieces if the reference weight is less than the recommended Auto-Add minimum weight in Table 7-2. The instructions to add these additional pieces may be ig­nored. This function can be activated or deactivated in the Count menu. The factory default setting is off.
Place the stipulated number of pieces on the scale (e.g. 2 pieces). The scale then determines the new APW based on the total pieces
WEIGHT
kg
g
oz
lb
Net
B/G %Ac
DATA COUNT
T
MR
APW
PT
<>
M
PCS
14
Trooper Count Scale
3.7 RS232 Communication
All communication is accomplished using standard ASCII format. Characters shown in the following table are acknowl­edged by the Scale. Invalid command response "ES" error indicates the Scale has not recognized the command. Commands sent to the Scale must be terminated with a carriage return (CR) or carriage return-line feed (CRLF). Data output by the Scale is always terminated with a carriage return-line feed (CRLF). See section 4.7 for RS232 and Print Setup.
3.7.1 RS232 User Command Table
Command Description Character (LFT is OFF)
P Same as pressing PRINT button. Sends print data to peripheral device. V Output software version. For example: "TC30 Sr. 1.20" T Same as pressing TARE. If the Scale accepts the command, it will respond "OK".
If the Scale is in motion, it will output "CANT TARE" until the Scale is stable. When stable, tare is accepted. (NOTE: If LFT is ON, special tare rules apply.)
Z Same as pressing ZERO button. If the Scale accepts the command, it will respond "OK".
NOTE: Any other command will produce the answer "ES".
3.7.2 Output Formats
Data output can be initiated by: 1) By pressing PRINT Units button or, 2) Sending a print command (P) from a computer. The two available output styles are shown below.
OHAUS TROOPER COUNT TC3RS SN: 222222
PART# = 123456789000 COUNT = 38 TOTAL = 188 APW = 5.000 NET = 190 g TARE = 10 g GROSS = 200
g
g
OHAUS TROOPER COUNT TC3RS SERIAL No. : 222222
(WEIGHT UNIT = g) PART No. COUNT TOTAL APW NET TARE GROSS 123456789000 138 326 5.000 690 10 700
"Column" Print Style "Line" Print Style
3.7.3 RS-232 Pin Out.
The adjacent table illustrates the pin-out connections on the Female 9-pin D-sub connector.
1 N/C 2 Data In (RXD) 3 Data Out (TXD) 4 N/C 5 Ground 6, 7, 8, 9 N/C
15
Trooper Count Scale
4. SETUP
For first time setup, step through all menus and set the parameters as desired.
4.1 Setup Protection
The Trooper Count Scale is equipped with menus which permit certain functions to be locked out (not changed) during operation. If you intend to physically lock out changes to the setup selections, you will need to access the hardware Lock Switch located on the main circuit board following the setup procedure.
4.2 Menu Structure
The following table illustrates the menu structure in the Trooper Count Scale.
CAL
Span
Lin
SETUP
Reset yes,no
LFT on,
off
Cal Unit lb, 1000g
Cal Point 1/3, 2/3, Full
Beeper All, signal, off
Backlight on, off, auto
Auto Off Timer on (5 min), off
Library
On,OFF
READ
Unit g on, off
Unit kg (1000g) on, off
Unit lb on, off
Unit oz on, off
Auto Zero Tracking
0.5*, 1, 3, off
PRINT
Baud Rate
2400, 4800,
1200, 9600, 19200
Data Length 7,8
Parity Bit none, Even, Odd
Print Part No. on, off
Print COUNT on, off
Print APW on, off,
Print Net on, off
Print Tare on, off
Print TOTAL pcs on, off
Print Gross on, off
Print On Stable
off, 0.5d, 1d*, 3d
COUNT
Auto Add off, on
Print Clear on, off
Auto Opt. on, off
M requires 10d chg.
on, off
LOCSW
CAL Lock on, off
Setup Lock on, off
Read Lock on, off
Print Lock on, off
Count Lock on, off
Library Lock on, off
NOTE: Bolded items indicate factory defaults. * LFT on setting
16
Print Header on, off
Print STYLE
Col, Line
Trooper Count Scale
4.3 Control Functions
The Scale has six menus; CAL, SETUP, READ, PRINT, COUNT and LOCSW which are entered by pressing and holding the SAMPLE Menu button until CAL MENU is displayed, then releasing it. During setup, the M-, M+, Enter and Data buttons are used to traverse and enter the menus.
4.3.1 Menu Operation
 To enter menus, press and hold SAMPLE Menu button. A short press in menu mode will exit to the weigh mode.  M+ and M- buttons advances through blinking choices.  Enter button accepts blinking choice.  Data button returns to the next higher menu level. At top level, it exits the menu operation.  Three rapid beeps indicates an improper button press.
4.4 Checking the Menu Settings
Each of the menus in the Trooper Count Scale can be quickly checked to see if any of the parameters need to be changed to match your requirements. The default settings are shown as bolded entries on the menu structure, section
4.2. Review the menu structure to determine what settings have to be changed.
To enter the menus, press and hold the SAMPLE Menu button until CAL MENU is displayed.
Repeated pressing of the M- button will traverse through the (6) menus in the order shown. You may go back by using the M+ button. To leave the menus, press the Data button.
To enter a selected menu (blinking), press the Enter button, then scroll through all settings by repeately pressing the M+ or M- buttons. In this manner, you can quickly check all settings for any particular menu.
All menu settings are saved once they are changed.
WEIGHT
kg
g
oz
lb
BLINKING
WEIGHT
kg
g
oz
lb
BLINKING
WEIGHT
kg
g
oz
lb
BLINKING
WEIGHT
Net
Net
Net
B/G %Ac
B/G %Ac
B/G %Ac
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
MR
APW
PT
MR
APW
PT
MR
APW
PT
<>
M
<>
M
<>
M
PCS
PCS
PCS
LOC ON indicates that the menu has been locked in the Loc Menu. You may view the menu parameters, but they may not be changed. See section 4.9 Lock­out menu for details.
The LOC MENU SW ON appears if the lockswitch un­derneath the scale is set on. This prevents the set­tings from being changed (as in a Legal For Trade application). See section 5.3 for details.
kg
g
oz
lb
BLINKING
WEIGHT
kg
g
oz
lb
BLINKING
WEIGHT
kg
g
oz
lb
BLINKING
Net
Net
Net
B/G %Ac
B/G %Ac
B/G %Ac
T
DATA COUNT
T
DATA COUNT
T
MR
APW
PT
MR
APW
PT
MR
APW
PT
<>
M
<>
M
<>
M
PCS
PCS
PCS
17
Trooper Count Scale
4.5 Setup Menu Procedure
With the Scale ON, press and hold theSAMPLE Menu button until CAL MENU is displayed.
NOTE: The calibration menu is covered in Section 5.
Press M- button to advance to SETUP MENU, then press Enter button. Navigate through the menu as explained in section 4.3.1.
Press Enter button to get into the RESET sub-menu. YES will change all menu settings to their default set­tings. Library data will not be affected.
ON will set the Scale to LFT configuration. In the ON setting, the zero setting will be limited to +/- 2%. In the OFF setting, the zero setting is 100%. Also, in the ON setting, the LFT readability is set. In the OFF posi­tion, the readability is set to the Enhanced value.
Selected calibration units are either lb or 1000g (kg).
WEIGHT
kg
g
oz
lb
BLINKING
WEIGHT
kg
g
oz
lb
BLINKING
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
Net
Net
Net
Net
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
MR
APW
PT
MR
APW
PT
MR
APW
PT
MR
APW
PT
<>
M
<>
M
<>
M
<>
M
PCS
PCS
PCS
PCS
A selection of 1/3, 2/3 or full span calibration values are available. The numeric values of the calibration points are displayed for the specific Scale capacity and calibration unit.
Three setting are available, ALL, SIGNAL and OFF. ALL = the beeper sounds when buttons are pressed or when error or target signals are needed. SIGNAL = beeps on errors (three beeps) or at count target (con­tinuous). OFF = is always off.
Backlight has three available settings; ON, OFF, AUTO. ON = backlight is always on. AUTO = backlight turns on with any key press or a weight change (5d) and remains on for 5 seconds. OFF = backlight is always off. OFF or AUTO is recommended for battery opera­tion.
ON = Scale turns off in 5 minutes if no key press or weight changes (5d) are detected. OFF = auto off feature is disabled. ON is recommended for battery operation.
18
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
Net
Net
Net
Net
Net
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
MR
APW
PT
MR
APW
PT
APW
PT
APW
PT
APW
PT
MR
MR
M
M
M
MR
<>
M
<>
<>
<>
<>
M
PCS
PCS
PCS
PCS
PCS
4.6 Read Menu
Trooper Count Scale
Procedure
If you have just finished changing settings in the SETUP menu, Press the Data button to go back to the SETUP MENU, SETUP should be blinking.
Press M- button to advance to the READ menu. In this menu, you are able to turn the measuring units on or off and set the Auto Zero tracking. Press the Enter button. Navigate through the menu as explained in section 4.3.1.
Grams (Unit g) - Select ON or OFF.
Kilograms (1000 g) - Select ON or OFF .
Avoirdupois Pounds (lb) - Select ON or OFF.
WEIGHT
kg
g
oz
lb
BLINKING
WEIGHT
kg
g
oz
lb
BLINKING
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
Net
Net
Net
Net
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
MR
APW
PT
MR
APW
PT
MR
APW
PT
MR
APW
PT
<>
M
<>
M
<>
M
<>
M
PCS
PCS
PCS
PCS
Ounces (oz) - Select ON or OFF.
Auto Zero Tracking - Scale will re-zero at 0.5d, 1d or 3d per second (0.5d is the LFT setting). Select 0.5d, 1d, 3d or OFF.
4.7 Print Menu Procedure
If you have just finished changing settings in the READ menu, press the Data button to go back to the READ MENU. READ should be blinking.
Press M- button to advance to the PRINT MENU. In this menu, you are able to change communication parameters and printing functions. Press the Enter button. Navigate through the menu as explained in section 4.3.1.
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
BLINKING
WEIGHT
kg
g
oz
lb
BLINKING
Net
Net
Net
Net
Net
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
MR
APW
PT
MR
APW
PT
MR
APW
PT
MR
APW
PT
MR
APW
PT
<>
M
<>
M
<>
M
<>
M
<>
M
PCS
PCS
PCS
PCS
PCS
19
Trooper Count Scale
4.7 Print Menu (Cont.)
Procedure (Cont.)
Baud rate - Select either 1200, 2400, 4800, 9600 or 19200 baud.
Data Length - Select either 7 or 8.
Parity - Select either NONE, EVEN or ODD. NOTE: No stop bit settings are required.
Print Part Number - Select ON or OFF
Print Count Quantity - Select ON or OFF.
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
Net
Net
Net
Net
Net
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
MR
APW
PT
MR
APW
PT
APW
PT
APW
PT
APW
PT
MR
MR
MR
M
M
M
M
<>
M
<>
<>
<>
<>
PCS
PCS
PCS
PCS
PCS
Print APW - Select ON or OFF.
Print Net weight - Select ON or OFF.
Print Tare Value - Select ON or OFF.
Print Accumulated Count - Select ON or OFF.
Print Gross (Brutto) Weight - Select ON or OFF.
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
Net
Net
Net
Net
Net
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
MR
APW
PT
MR
APW
PT
MR
APW
PT
MR
APW
PT
MR
APW
PT
<>
M
<>
M
<>
M
M
<>
M
<>
PCS
PCS
PCS
PCS
PCS
20
4.7 Print Menu (Cont.)
Trooper Count Scale
Procedure (Cont.)
Select stability required for printing at 0.5d, 1d, 3d or OFF. OFF allows the printing when the weight is not stable. In LFT mode, stability is 1d.
Print Header - Allows printing of preset header that includes OHAUS, Model No. and unit serial number. Select ON or OFF.
Printout Style - See section 3.7.2 for examples. Se­lect COL (24 column) or LINE (line).
4.8 Count Menu
Procedure
If you have just finished changing settings in the Print menu, press the Data button to go back to the PRINT MENU. Press M- button to advance to the Count menu, then press Enter button. Navigate through the menu as explained in section 4.3.1.
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
Net
Net
Net
Net
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
MR
APW
PT
MR
APW
PT
APW
PT
APW
PT
M
MR
MR
<>
M
<>
<>
M
M
<>
PCS
PCS
PCS
PCS
Auto Add - Auto Add instructs the operator to add x pieces to reference count (see section 3.6). Select ON or OFF.
Print Clear - When Print clear is set to ON, the Scale prints the accumulated count memory and then clears the accumulated memory when the PRINT Units but- ton is pressed. Select ON or OFF.
Auto Reference Optimization - If additional reference pieces less than or equal to the existing reference count are added to the pan, the Scale will recompute the APW based on the new reference weight (see section
3.5). Select ON or OFF.
Memory Change - A 10d weight change is required before the accumulated memory can be changed by pressing M+ or M-. Select ON or OFF.
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
Net
Net
Net
Net
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
MR
APW
PT
MR
APW
PT
MR
APW
PT
MR
APW
PT
<>
M
M
<>
M
<>
M
<>
PCS
PCS
PCS
PCS
21
Trooper Count Scale
4.9 Lockout Menu
The Lockout menu (LOC) is a software-controlled feature that allows the Calibration, Setup, Read, Print and Count menus and the Library data to be locked to restrict tampering. When a menu or the Library is locked, the data can be viewed but not changed. To further restrict tampering, the Lock Switch on the circuit board can be set to prevent changes to the Lockout menu. See section 5.3 for Lock Switch information.
Procedure
If you have just finished entering data in the Count menu, press the Data button to go back to the COUNT MENU. COUNT should be blinking. Press M- button to advance to the LOC menu. Press the Enter button. Navigate through the menu as explained in section
4.3.1.
LOC CAL - When set to ON, locks the CAL calibration sub-menu. Select ON or OFF.
LOC SETUP - When set to ON, locks the SETUP sub­menu. Select ON or OFF.
LOC READ - When set to ON, locks the READ sub­menu. Select ON or OFF.
LOC PRINT - When set to ON, locks the PRINT sub­menu. Select ON or OFF.
LOC COUNT - When set to ON, locks the COUNT sub­menu. Select ON or OFF.
LOC LIB - When set to ON, locks the Library records but still allows the accumulated memory to update (if activated). Select ON or OFF.
WEIGHT
kg
g
oz
lb
BLINKING
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
Net
Net
Net
Net
Net
Net
Net
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
B/G %Ac
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
PT
DATA COUNT
T
DATA COUNT
T
DATA COUNT
T
MR
APW
PT
APW
PT
APW
PT
APW
APW
PT
APW
PT
APW
PT
MR
MR
MR
MR
MR
MR
M
<>
M
<>
M
<>
M
<>
M
<>
M
<>
M
<>
PCS
PCS
PCS
PCS
PCS
PCS
PCS
NOTE: Location 00 cannot be locked.
22
Trooper Count Scale
5. CALIBRATION AND SEALING
Although your Trooper Count scale has been calibrated at the factory, many conditions may require it to be re-calibrated when first put into use and periodically thereafter. Temperature change, geographic gravity variation, altitude change and abuse are a few reasons why a scale may need calibration.
Span calibration uses two points to adjust the scale. One point is called the zero value where there is no weight on the scale. The other point is the Span value that is often but not necessarily at full capacity. Trooper Count lets you choose between three Span values. These values are shown in the table below. The highest weight value generally gives the best performance over the entire weight range. Other span values may be chosen due to calibration weight availability or for improved accuracy at a lower weight range.
Linearity calibration uses three points to adjust the scale. In addition to the zero and full scale values an additional point is chosen between the two. The values are factory set and are shown in the table below. Linearity calibration should only be performed if determined to be necessary after a Span calibration.
Before calibrating a scale first determine that it is necessary. Place available calibration masses on the scale, as each mass is added the display should indicate the correct weight within +/- 1 scale division. If the scale is within this tolerance there is no need to calibrate the scale.
If the scale must be calibrated, use the table below to select the calibration value(s) that will be used. Do not start the
calibration procedure until you are certain that you have the appropriate calibration masses.
Verify that the Cal menu is not locked (see section 4.9) and that the correct calibration unit (pound or kilogram) has been chosen in the Setup menu (See section 4.5).
CALIBRATION MASSES
Cal in kg: Span Cal choices Linearity Cal (fixed)
TC3RS 1, 2, 3 kg 2 & 3 kg TC6RS 2, 4, 6 kg 4 & 6 kg TC15RS 5, 10, 15 kg 10 & 15 kg TC30RS 10, 20, 30 kg 20 & 30 kg
Cal in lb: Span Cal choices Linearity Cal (fixed)
TC3RS 2, 4, 6 lb 4 & 6 lb TC6RS 5, 10, 15 lb 10 & 15 lb TC15RS 10, 20, 30 lb 20 & 30 lb TC30RS 20, 40, 60 lb 40 & 60 lb
Recommended tolerance Class for the calibration masses: ASTM Class 4 or OIML F2
23
Trooper Count Scale
5.1 Span Calibration Procedure
The following example is for a 6kg scale. With the Scale ON, press and hold the SAMPLE Menu button until CAL MENU is displayed.
Press the Enter button twice, NO should be blinking.
Press the M- button to change to YES, then press the Enter button.
WEIGHT
kg
g
oz
lb
Net
B/G %Ac
DATA COUNT
T
MR
APW
PT
<>
M
Clear the pan, then press the Enter button.
(Scale is determining zero value)
kg
kg
kg
PCS
WEIGHT
g
BLINKING
WEIGHT
g
WEIGHT
g
WEIGHT
DATA COUNT
oz
lb
oz
lb
oz
lb
B/G %Ac
Net
B/G %Ac
Net
B/G %Ac
Net
COUNT
M
<>
T
DATA COUNT
T
DATA COUNT
T
PCS
DATA COUNT
MR
APW
PT
MR
APW
PT
MR
APW
PT
<>
M
M
M
M
<>
BLINKING
<>
BLINKING
COUNT
<>
PCS
PCS
PCS
PCS
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
Net
B/G %Ac
DATA COUNT
T
MR
APW
PT
<>
M
PCS
B/G %Ac
Net
M
T
COUNT
<>
PT
PCS
APW
MR
<>
M
COUNT
<>
M
PCS
PCS
Place the indicated calibration mass on the pan, then press the Enter button. The sample illustration indicates a 6kg Scale using a 2kg mass.
WEIGHT
DATA COUNT
(Scale is determining span value) If the calibration was successful, SPAN TEST DONE is
displayed momentarily followed by CLEAR PAN THEN PRESS ENTER.
kg
g
lb
WEIGHT
oz
Net
B/G %Ac
T
DATA COUNT
MR
APW
PT
<>
M
PCS
Clear the pan, then press the Enter button.
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
Net
B/G %Ac
DATA COUNT
T
MR
APW
PT
<>
M
PCS
B/G %Ac
Net
M
T
COUNT
<>
PT
PCS
APW
MR
<>
M
COUNT
<>
M
PCS
PCS
The scale will restart and return to weighing. If calibration was unsuccessful, refer to the troubleshooting section.
24
BLINKING
BLINKING
5.2 Linearity Calibration
Procedure
The following example is with a 6kg Scale. With the Scale ON, press and hold the SAMPLE Menu button until CAL MENU is displayed.
Press the Enter button.
Trooper Count Scale
WEIGHT
kg
g
oz
lb
BLINKING
WEIGHT
Net
B/G %Ac
DATA COUNT
T
DATA COUNT
MR
APW
PT
<>
M
PCS
Press the M- button.
Press the Enter button.
Press the M+ button, then press the Enter button. Follow display instructions.
WEIGHT
kg
g
oz
lb
Net
B/G %Ac
DATA COUNT
T
MR
APW
PT
<>
M
(Scale is determining zero value)
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
PCS
WEIGHT
B/G %Ac
Net
B/G %Ac
Net
B/G %Ac
Net
B/G %Ac
Net
M
COUNT
<>
T
DATA COUNT
T
PT
DATA COUNT
T
PT
DATA COUNT
T
PT
PCS
DATA COUNT
MR
APW
PT
MR
APW
APW
APW
MR
MR
M
M
M
M
<>
M
<>
<>
BLINKING
<>
BLINKING
COUNT
<>
PCS
PCS
PCS
PCS
PCS
kg
g
oz
lb
WEIGHT
kg
g
oz
lb
Net
B/G %Ac
DATA COUNT
T
MR
APW
PT
<>
M
PCS
B/G %Ac
Net
COUNT
M
<>
T
PCS
MR
APW
PT
M
<>
COUNT
<>
M
PCS
PCS
Place the indicated calibration mass on the pan. Press the Enter button. The sample illustration indicates a 6kg Scale using a 4kg mass for the first linearity entry.
WEIGHT
DATA COUNT
(Scale is determining first linearity value)
kg
g
oz
lb
Net
B/G %Ac
T
MR
APW
PT
<>
M
PCS
25
Trooper Count Scale
5.2 Linearity Caibration (Cont.)
Procedure (Cont.)
WEIGHT
kg
g
oz
lb
Net
B/G %Ac
DATA COUNT
T
MR
APW
PT
<>
M
PCS
COUNT
<>
M
PCS
COUNT
<>
M
PCS
Place the indicated calibration mass on the pan. Press the Enter button. The sample illustration indicates a 6kg Scale using a 6kg mass.
WEIGHT
DATA COUNT
(Scale is determining second linearity value)
kg
g
lb
WEIGHT
oz
Net
B/G %Ac
T
DATA COUNT
MR
APW
PT
<>
M
PCS
If the calibration was successful, CAL TEST DONE is displayed momentarily followed by CLEAR PAN THEN PRESS ENTER.
kg
g
oz
lb
Net
B/G %Ac
T
MR
APW
PT
<>
M
PCS
Clear the pan, then press the Enter button.
WEIGHT
kg
g
oz
lb
Net
B/G %Ac
DATA COUNT
T
MR
APW
PT
<>
M
PCS
COUNT
<>
M
PCS
COUNT
<>
M
PCS
The scale will restart and return to weighing. If calibration was unsuccessful, refer to the troubleshooting section.
26
Loading...
+ 74 hidden pages