CHAPTER 1 GETTING STARTED
1. INTRODUCTION
This service manual contains instructions for the diagnosis and repair work to be performed by
Ohaus Dealers or Ohaus authorized service centers. Knowledge of the operation of the Scale
is assumed. Instruction manuals may be required with this service manual. For complete
information on operation, refer to the Instruction Manual.
This manual covers maintenance on the following:
EB Series Scales, Models EB3, EB6, EB15 and EB30
The contents of this manual are contained in five chapters and two Appendixes with service
menu and calibration instructions.
Chapter 1 Introduction
test masses, service strategy, specifications, and setup.
Chapter 2 Diagnosis
preliminary checks, troubleshooting tables, testing procedures, and an interconnection diagram.
Chapter 3 Repair Procedures
Chapter 4 Final Testing
Chapter 5 Parts Lists
components with associated parts lists.
Appendix A Service Modes
Series Scales in a service mode.
Appendix B Service Calibration
- Contains information about service facilities, tools, test equipment,
- Contains information on problem verification, scale examination,
- Contains detailed repair procedures for all major components.
- Contains performance tests, and calibration.
- Contains overall views identifying all major serviceable replacement
- Contains Configuration, Capacity and Readability for the EB
- Contains Service Calibration instructions.
1.1 SERVICE FACILITIES
To service a scale, the service area should meet the following requirements:
Meet scale specifications for temperature environmental requirements.
Be free of vibrations such as fork lift trucks close by, large motors, air currents or drafts
from air conditioning/heating ducts, open windows, people walking by, fans, etc.
Be clean and free of excessive dust.
Work surface must be stable and level.
Scale must not be exposed to direct sunlight or radiating heat sources.
Use an approved Electro-Static Device.
EB Series Scales Service Manual1-1Ohaus Corporation www.ohaus.com
CHAPTER 1 GETTING STARTED
MODEL
Calibration Masses
Totaling
EB3
EB6
EB15
EB30
3kg
6kg
15kg
30kg
1.2 TOOLS AND EQUIPMENT
1.2.1 Standard Tools and Test Equipment
The service shop should contain the following equipment:
1. Digital Voltmeter (DVM).
2. Standard Electronics tool kit.
3. Desk magnifier on a stand.
4. Grounding mat and clip.
5. Razor blades.
1.2.2 Special Tools
To service the Ohaus EB Series Scales, the following equipment is recommended:
: Adapters required for servicing are location dependent.
1.3 CALIBRATION MASSES REQUIRED
The masses required to test the EB Series Scales must meet the requirements of ASTM Class
4 or OIML Class F2. The mass values are listed in Table 1-1.
TABLE 1-1. CALIBRATION MASSES
Note: The EB Scales can be calibrated over a wide range, however, it is recommended that the
calibration mass used is at or near the full capacity of the scale for maximum accuracy.
1.4 SERVICE STRATEGY
The repair method for the EB Series Scales is the direct substitution of major assemblies.
The available repair parts are listed in Table 5-1.
The EB Series Scales contain 6 major replaceable assemblies: Top Housing, Membrane
Switch, Load Cell Assembly with Frame, Lead Acid Rechargeable Battery, Display PC Board
with LCD and the Main PC Board.
The Load Cell and LCD may also be ordered and replaced separately.
Note:
Ohaus Corporation www.ohaus.com 1-2 EB Series Scales Service Manual
Full Capacity by Subtraction (except EB15: up to 10kg only)
Stabilization Time
< 2 seconds
Operating Temperature
0° to 40° C
Humidity Range
≤ 90% relative humidity, non-condensing
Power
AC Adapter 12V DC / 800mA
Internal rechargeable lead acid battery
Battery Life
80 hours continuous use with 12 hour recharge time
Calibration
Automatic external with kg/g mass, factory calibration recovery
Shipping Protection
Shipping screw to avoid damage to sensitive components
Safe Overload Capacity
120% of capacity
Platform Size
294 x 226 mm / 11.6 x 8.9 in.
Scale dimensions
325 x 330.5 x 114 mm / 12.8 x 13 x 4.5 in. (W x D x H)
Shipping dimensions
440 x 360 x 160 mm / 17.3 x 14.2 x 6.3 in (W x D x H)
Net Weight
4.2 kg / 9.3 lb
Shipping Weight
5.3 kg / 11.7 lb
Other Features
Auto-Zero Tracking, Filtering Level
CHAPTER 1 GETTING STARTED
TABLE 1-2. SPECIFICATIONS
EB Series Scales Service Manual1-3Ohaus Corporation www.ohaus.com
CHAPTER 1 GETTING STARTED
inC 01EL AUAC 2dFil 1
AU
ON
OFF
0.5d
1d
2d
4d
0
1
2
3
(Levels)
(d=scale division)
(mode-dependent
settings)
inC 01
EL AU
AC 2d
FiL 1
Scale increment
- Sets the displayed scale increment (also known as readability or
graduation). Selectable setup values will be model dependent and are equivalent
to within 30,000 to 3,000 scale divisions.
Backlight - Sets the activation mode of the backlight.
The following settings are available: AU (Auto-on with items > 9d placed on the
Platform or any button pressed; turns off after 5 seconds of inactivity), On, oFF.
Zero Display Range - Sets the range in which the Zero reading is maintained.
The following settings are available: 0.5d, 1d, 2d, 4d (d= scale division).
Filtering - Sets the level at which the Stable indication turns on; the higher the
setting, the faster the stabilization time.
The following settings are available: 0, 1, 2, 3 (levels).
1.6 SCALE SETTINGS
The EB scale contains four user-selectable scale settings (Setup mode). These are as follows:
Scale Increment, Backlight, Zero Tracking Range, and Filtering. A description of each setting is
explained below.
1. To enter the setup mode, press and then release COUNT and M+ at the same time to
enter into the user-selectable scale settings.
2. In the Setup mode, press ▲ to step through available settings.
3. Press Enter to accept the displayed setting and proceed to the next parameter.
4. Press Exit to proceed to the next parameter without saving any changes.
Table 1-3. MENU STRUCTURE
Ohaus Corporation www.ohaus.com 1-4 EB Series Scales Service Manual
CHAPTER 2 DIAGNOSIS
TABLE 2-1. SCALE WILL NOT TURN ON WITH AC ADAPTER
SYMPTOM
PROBABLE CAUSE
REMEDY
Unit will not turn on
with AC adapter
supplied
Adapter defective
Membrane Switch defective
Check the AC adapter voltage output
(Section 2.5). If voltage is low or
nonexistent, replace the AC adapter.
If OK, proceed.
Test the Membrane Switch (Section
2.6). Replace if necessary
(Section 3.2).
If the scale fails to turn on with a new
Membrane Switch, test the Main PC
Board (Section 2.8). If defective,
replace the Main PCB (Section 3.3.1).
After repair, proceed with Performance
Tests in Chapter 4.
2. DIAGNOSIS
This section contains information needed to properly evaluate the reported problem and
diagnose its cause.
2.1 SCALE SETUP AND EXAMINATION
Set up the scale according to the Instruction manual. Allow the scale to stabilize to room
temperature. Examine the scale for signs of corrosion or physical damage.
2.2 PRELIMINARY CHECKS
Power up the scale using the customer’s ac adapter. In the case where the scale will not power
up, check the ac adapter. Observe and record the error codes and software revision.
Record all menu settings as received.
2.3 TROUBLESHOOTING TABLES
Troubleshooting tables 2-1 through 2-7 identify actual types of problems that could be
encountered with the scale.
The troubleshooting tables refer to the EB Scales Interconnection Diagram, Figure 2-2, to assist
in locating the problem.
EB Series Scales Series Service Manual2-1Ohaus Corporation www.ohaus.com
CHAPTER 2 DIAGNOSIS
TABLE 2-2. SCALE WILL NOT TURN ON USING BATTERY POWER
SYMPTOM
PROBABLE CAUSE
REMEDY
Scale will not turn on
using battery power.
Battery discharged or
defective.
Wiring harness defective or
battery clips connection
broken.
Membrane Switch defective.
Main PC Board is defective.
Test the battery (Section 2.9). Replace
if necessary (Section 3.6).
Open the scale (Section 3.1). Check
voltage at battery contacts first. Check
DC voltage at the battery connector on
the Main PC Board. Voltage should
read approximately 6 Volts DC
minimum. If voltage is not present at
the connector, disconnect the leads
from the battery and make a continuity
test between the connectors on the
Harness to the connector on the Main
PC Board. Resistance should be 0
ohms for the red lead and 0 ohms for
the black lead. If an open condition
exists, replace wiring and connector as
necessary. If OK, proceed.
Test the Membrane Switch (Section
2.6). Replace if necessary
(Section 3.2).
If the scale fails to turn on with a new
Membrane Switch, test the Main PC
Board (Section 2.8). If defective,
replace the Main PCB (Section 3.3.1).
After repair, proceed with Performance
Tests in Chapter 4.
Ohaus Corporation www.ohaus.com 2-2 EB Series Scales Series Service Manual
CHAPTER 2 DIAGNOSIS
TABLE 2-3. SCALE DOES NOT RESPOND TO FRONT PANEL CONTROLS.
SYMPTOM
PROBABLE CAUSE
REMEDY
Scale does not
respond to front panel
controls.
Membrane Switch defective
Test the Membrane Switch (Section
2.6). Replace if necessary (Section
3.2). If membrane switches are OK,
test the Main PC Board (Section 2.8).
If defective, replace the Main PCB
(Section 3.3.1).
After repair, proceed with Performance
Tests in Chapter 4.
TABLE 2-4. NO DISPLAY OR PARTIAL DISPLAY.
SYMPTOM
PROBABLE CAUSE
REMEDY
Display is not on or
partial characters are
displayed.
Display PC Board is
defective or LCD may be
defective.
NOTE: If the LCD shows
signs of damage such as
cracked, dim or partial
characters, the LCD should
be replaced.
The Display PC Board is replaced as a
whole assembly. Check procedures in
Tables 2-1, 2-2 and 2-3 first and verify
that other problems do not exist.
Replace Display PC Board or LCD.
See Repair Procedures 3.3.2 and
3.3.3.
After repair, proceed with Performance
Tests in Chapter 4.
EB Series Scales Series Service Manual2-3Ohaus Corporation www.ohaus.com
CHAPTER 2 DIAGNOSIS
TABLE 2-5. BALANCE READING INCORRECT OR UNSTABLE.
SYMPTOM
PROBABLE CAUSE
REMEDY
Balance reads
incorrectly.
Balance reading
unstable.
Needs calibration.
Load Cell is damaged.
Main PC Board defective.
Perform Service Calibration in
Appendix B. If calibration fails, refer to
Table 2-6.
Test the Load Cell (Section 2.7). If
defective replace it (Section 3.4).
After repair, proceed with Performance
Tests in Chapter 4.
Test the Main PC Board (Section 2.8).
If defective, replace it (Section 3.3).
After repair, proceed with Performance
Tests in Chapter 4.
TABLE 2-6. CANNOT CALIBRATE THE SCALE.
SYMPTOM
PROBABLE CAUSE
REMEDY
Scale can be turned
on but will not
calibrate.
Incorrect weights.
Load cell assembly
defective.
Main PC Board defective
Verify that proper weights are used.
Test the Load Cell (Section 2.7). If
defective replace it (Section 3.4).
After repair, proceed with Performance
Tests in Chapter 4.
Test the Main PC Board (Section 2.8).
If defective, replace it (Section 3.3).
After repair, proceed with Performance
Tests in Chapter 4.