Ohaus D32XW30VR, D32XW300VXOLT, D32XW60VL, D32XW150VXAC, D32XW150VX Data Sheet

...
3000 Series Indicators Instruction Manual
T31P Indicator
T31XW Indicator
i
ii
Compliance to the following standards is indicated by the corresponding marking on the product.
This product conforms to the EMC Directive 89/336/EEC, the Low Voltage Directive 73/23/EEC and the Non-automatic Weighing Instruments Directive 90/384/EEC. The complete Declaration of Conformity is available from Ohaus Corporation.
AS/NZS4251.1, AS/NZS4252.1
Important Notice for verified weighing instruments
Weighing Instruments verified at the place of manufacture bear one of the preceding marks on the pack­ing label and the green ‘M’ (metrology) sticker on the descriptive plate. They may be put into service immediately.
Weighing Instruments to be verified in two stages have no green ‘M’ (metrology) on the descriptive plate and bear one of the preceding identification mark on the packing label. The second stage of the initial verification must be carried out by the approved service organization of the authorized representative within the EC or by the national weights & measures (W+M) authorities.
The first stage of the initial verification has been carried out at the manufacturer’s work. It comprises all tests according to the adopted European standard EN 45501:1992, paragraph 8.2.2.
If national regulations limit the validity period of the verification, the user of the weighing instrument must strictly observe the re-verification period and inform the respective W+M authorities.
iii
Disposal
In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment.
If you have any questions, please contact the responsible authority or the distributor from which you purchased this device.
Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related.
Thank you for your contribution to environmental protection.
FCC Note
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Industry Canada Note
This Class A digital apparatus complies with the Canadian ICES-003. Cet appareil numérique de la classe A est conforme à la Norme NMB-003 du Canada.
ISO 9001 Registration
In 1994, Ohaus Corporation, USA, was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International (BVQI), confirming that the Ohaus quality management system is compliant with the ISO 9001 standard’s requirements. On May 15, 2003, Ohaus Corporation, USA, was re-registered to the ISO 9001:2000 standard.
iv
TABLE OF CONTENTS
1. INTRODUCTION ..........................................................................................................................................EN-4
1.1 Safety Precautions .....................................................................................................................................EN-4
1.2 Overview of Parts and Controls ...................................................................................................................EN-5
1.3 Control Functions .......................................................................................................................................EN-9
2. INSTALLATION .........................................................................................................................................EN-10
2.1 Unpacking ..............................................................................................................................................EN-10
2.2 External Connections ................................................................................................................................EN-10
2.2.1 RS232 Interface Cable to T31P........................................................................................................EN-10
2.2.2 AC Power to T31P ........................................................................................................................EN-10
2.2.3 AC Power to T31XW .......................................................................................................................EN-10
2.2.4 Battery power ................................................................................................................................EN-11
2.2.5 Mounting Bracket to T31XW ............................................................................................................EN-11
2.3 Internal Connections .................................................................................................................................EN-11
2.3.1 Opening the Housing ....................................................................................................................... EN-11
2.3.2 Scale Base to T31P or T31XW .......................................................................................................EN-12
2.3.3 RS232 Interface Cable to T31XW ......................................................................................................EN-12
2.4 T31P Rear Cover Orientation .....................................................................................................................EN-13
2.5 Direct Wall Mounting (T31P only) .............................................................................................................EN-13
2.6 Mounting Bracket (T31XW only) ................................................................................................................EN-13
EN-13000 Series Indicators
3. SETTINGS ................................................................................................................................................EN-14
3.1 Menu Structure ........................................................................................................................................EN-14
3.2 Menu Navigation .....................................................................................................................................EN-15
3.3 Calibration Menu .....................................................................................................................................EN-15
3.3.1 Span Calibration ...........................................................................................................................EN-16
3.3.2 Linearity Calibration ......................................................................................................................EN-16
3.3.3 Geographical Adjustment Factor .....................................................................................................EN-17
3.3.4 End Calibration .............................................................................................................................EN-17
3.4 Setup Menu .............................................................................................................................................EN-19
3.4.1 Reset ...........................................................................................................................................EN-19
3.4.2 Legal for trade .............................................................................................................................. EN-19
3.4.3 Calibration Unit .............................................................................................................................EN-19
3.4.4 Capacity ......................................................................................................................................EN-19
3.4.5 Graduation ...................................................................................................................................EN-21
3.4.6 Power On Unit ..............................................................................................................................EN-21
3.4.7 Zero Range ..................................................................................................................................EN-21
3.4.8 End Setup ....................................................................................................................................EN-21
EN-2 3000 Series Indicators
TABLE OF CONTENTS (Cont.)
3.5 Readout Menu ......................................................................................................................................... EN-21
3.5.1 Reset ...........................................................................................................................................EN-21
3.5.2 Filter ............................................................................................................................................EN-22
3.5.3 Auto-Zero Tracking ........................................................................................................................EN-22
3.5.4 Backlight ..................................................................................................................................... EN-22
3.5.5 Auto Off Timer ..............................................................................................................................EN-22
3.5.6 End Readout ................................................................................................................................. EN-22
3.6 Mode Menu .............................................................................................................................................EN-23
3.6.1 Reset ...........................................................................................................................................EN-23
3.6.2 Parts Counting Mode .....................................................................................................................EN-23
3.6.3 End Mode ....................................................................................................................................EN-23
3.7 Unit Menu ...............................................................................................................................................EN-24
3.7.1 Reset ...........................................................................................................................................EN-24
3.7.2 Kilogram Unit ...............................................................................................................................EN-24
3.7.3 Pound Unit ...................................................................................................................................EN-24
3.7.4 Gram Unit ....................................................................................................................................EN-24
3.7.5 Ounce Unit ...................................................................................................................................EN-24
3.7.6 Pound Ounce Unit .........................................................................................................................EN-24
3.7.7 End Unit ......................................................................................................................................EN-24
3.8 Print Menu ..........................................................................................................................................EN-25
3.8.1 Reset ...........................................................................................................................................EN-25
3.8.2 Baud ........................................................................................................................................... EN-25
3.8.3 Parity .......................................................................................................................................... EN-25
3.8.4 Stop Bit .......................................................................................................................................EN-25
3.8.5 Handshake ..................................................................................................................................EN-26
3.8.6 Print Stable Data Only ...................................................................................................................EN-26
3.8.7 Auto Print .....................................................................................................................................EN-26
3.8.8 Content ........................................................................................................................................EN-26
3.8.9 End Print .....................................................................................................................................EN-26
3.9 Menu Lock Menu .....................................................................................................................................EN-27
3.9.1 Reset ...........................................................................................................................................EN-27
3.9.2 Lock Calibration ...........................................................................................................................EN-27
3.9.3 Lock Setup ...................................................................................................................................EN-27
3.9.4 Lock Readout ............................................................................................................................... EN-27
3.9.5 Lock Mode ...................................................................................................................................EN-27
3.9.6 Lock Unit .....................................................................................................................................EN-27
3.9.7 Lock Print ...................................................................................................................................EN-28
3.9.8 End Lock .....................................................................................................................................EN-28
TABLE OF CONTENTS (Cont.)
3.10 Security Switch .......................................................................................................................................EN-28
4. OPERATION ............................................................................................................................................. EN-28
4.1 Turning Indicator On/Off ............................................................................................................................EN-28
4.2 Zero Operation .........................................................................................................................................EN-28
4.3 Manual Tare ............................................................................................................................................EN-28
4.4 Changing Units of Measure .......................................................................................................................EN-29
4.5 Printing Data ........................................................................................................................................... EN-29
4.6 Application Modes ...................................................................................................................................EN-29
4.6.1 Weighing .....................................................................................................................................EN-29
4.6.2 Parts Counting .............................................................................................................................EN-29
5. SERIAL COMMUNICATION ..........................................................................................................................EN-31
5.1 Interface Commands ................................................................................................................................EN-31
5.2 Output Format ......................................................................................................................................... EN-32
6. LEGAL FOR TRADE ................................................................................................................................... EN-33
6.1 Settings .................................................................................................................................................. EN-33
6.2 Verification .............................................................................................................................................. EN-33
6.3 Sealing ................................................................................................................................................... EN-33
EN-33000 Series Indicators
7. MAINTENANCE .........................................................................................................................................EN-34
7.1 Model T31P Cleaning ...............................................................................................................................EN-34
7.2 Model T31XW Cleaning ............................................................................................................................ EN-34
7.3 Troubleshooting .......................................................................................................................................EN-34
7.4 Service Information .................................................................................................................................. EN-35
8. TECHNICAL DATA .....................................................................................................................................EN-36
8.1 Specifications .......................................................................................................................................... EN-36
8.2 Accessories and Options ...........................................................................................................................EN-37
8.3 Drawings and Dimensions ........................................................................................................................EN-38
EN-4 3000 Series Indicators
1. INTRODUCTION
This manual contains installation, operation and maintenance instructions for the T31P and T31XW Indicators. Please read this manual completely before installation and operation.
1.1 Safety Precautions
For safe and dependable operation of this equipment, please comply with the following safety precautions:
• Verify that the input voltage range printed on the data label matches the local AC power to be used.
• Make sure that the power cord does not pose a potential obstacle or tripping hazard.
• Use only approved accessories and peripherals.
• Operate the equipment only under ambient conditions specified in these instructions.
• Disconnect the equipment from the power supply before cleaning.
• Do not operate the equipment in hazardous or unstable environments.
• Do not immerse the equipment in water or other liquids.
• Service should only be performed by authorized personnel.
• The T31XW is supplied with a grounded power cable. Use only with a compatible grounded power outlet.
1.2 Overview of Parts and Controls
EN-53000 Series Indicators
TABLE 1-1. T31P PARTS.
2
1
3
4
5
6 7
8
Item Description
1 Data Label
2 Front Housing
3 Control Panel
4 Security Screw
5 Key Hole (4) for wall
mounting
6 Screw (4)
7 Data Label
8 Rear Housing
9 Power Receptacle
10 Strain Relief for Load Cell
Cable
11 RS232 Connector
910
Figure 1-1. T31P Indicator.
11
EN-6 3000 Series Indicators
1.2 Overview of Parts and Controls (Cont.)
1
2
3
4
5
6
TABLE 1-2. T31XW PARTS.
Item Description
1 Data Label
2 Front Housing
3 Adjusting Knob (2)
4 Control Panel
5 Mounting Bracket
6 Screw (4)
7 Rear Housing
8 Data Label
9 Strain Relief for RS232
10 Strain Relief for Load Cell
Cable
11 Power cord
7
8
910
Figure 1-2. T31XW Indicator.
11
1.2 Overview of Parts and Controls (Cont.)
EN-73000 Series Indicators
J4
1
2
3
LOAD CELL WIRING
J5
W1 W2
45
Figure 1-3. Main PC Board.
6
OFF
J7 J6
7 8
TABLE 1-3. MAIN PC BOARD.
Item Description
1 Keypad Connector J4 T31XW Model only
2 Battery Connector
ON
9
+EXC +SIG -SIG -EXC +SENS GND -SENS
RS232 WIRING
RXD GND
TXD
3 Line Power Input
4 Sense Jumper W1
5 Load Cell Terminal Block J5
6 Sense Jumper W2
7 RS232 Terminal Block J7 T31XW Model only
8 RS232 Connector J6 T31P Model only
9 LFT On / Off Switch
EN-8 3000 Series Indicators
1.2 Overview of Parts and Controls (Cont.)
1
14
13
12
Figure 1-4. Controls and Indicators.
TABLE 1-4. CONTROL PANEL.
No. Designation
1 Capacity Label Window
2 Pound symbol
3 Ounce symbol
4 Kilogram, gram symbols
5 TARE
6 TARE function symbol
7 FUNCTION
8 NET function symbol
9 PCS function symbol
10 PRINT
11 Battery function symbol
12 ON/ZERO
13 Center of Zero Indicator
14 Stable weight Indicator
8910 711
Menu
Units
Off
button
Mode
button
6
button
button
2
3
4
5
1.3 Control Functions
Button
EN-93000 Series Indicators
TABLE 1-5. CONTROL FUNCTIONS.
Primary Function (Short Press)
Secondary Function (Long Press)
Menu Function (Short Press)
ON/ZERO
If Indicator is On, sets zero.
Off
Turns the Indicator on or off.
Yes
Accepts the current setting on the display.
PRINT
Sends the current value to the COM port if AUTOPRINT is set to Off.
Units
Changes the weighing Unit.
No
Advances to the next menu or menu item.
Rejects the current setting on the display and advances to the next available setting.
Increments the value.
FUNCTION
Initiates an application mode.
Mode
Allows changing the application mode.
Press and hold allows scrolling through modes.
Back
Moves Back to previous menu item.
Decrements the value.
TARE
Performs a tare operation.
Menu
Enter the User menu.
Exit
Exits the User menu.
Aborts the calibration in progress.
EN-10 3000 Series Indicators
2. INSTALLATION
2.1 Unpacking
Unpack the following items:
• T31P or T31XW Indicator
• AC Adapter (T31P only)
• Mounting Bracket (supplied with T31XW only)
• Knobs (2) (supplied with T31XW only)
• Capacity Label Sheet
• Instruction Manual CD
• Warranty Card
• LFT sealing Kit
2.2 External Connections
2.2.1 RS232 interface Cable to T31P
Connect the optional RS232 cable to the RS232 connector Figure 1-1, item 13).
Pin Connection
1 N/C
2 TXD
3 RXD
4 N/C
5 GND
6 N/C
7 N/C
8 N/C
9 N/C
4
5
9
Figure 2-1. RS232 Pins.
2
3
1
78
6
2.2.2 AC Power to T31P
Connect the AC Adapter to the power receptacle (Figure 1-1, item 8), then plug the AC Adapter into an electrical outlet.
2.2.3 AC Power to T31XW
Connect the AC plug to a properly grounded electrical outlet.
EN-113000 Series Indicators
2.2.4 Battery Power
The indicator can be operated on the internal rechargeable battery when AC power is not available. The indicator will automatically switch to battery operation if there is a power failure or the power cord is removed.
Note: Before using the indicator for the first time, the internal rechargeable battery should be fully charged for up to 12 hours. The indicator can be operated during the charging process. The battery is protected against over charging and the indicator can remain connected to the AC power line.
Connect AC power to the indicator and allow it to charge. While the battery is charging, the triangle above the battery function symbol will light. When the battery is fully charged, this triangle will disappear.
The indicator can operate for up to 100 hours on a fully charged battery.
During battery operation, a flashing triangle above the battery function symbol indicates the battery is low and requires recharging. Approximately 60 minutes of operation will remain when the battery symbol starts to blink. The indicator will display Lo.BAT and automatically turn off when the battery is fully discharged.
CAUTION
BATTERY IS TO BE REPLACED ONLY BY AN AUTHORIZED OHAUS SERVICE DEALER.
RISK OF EXPLOSION CAN OCCUR IF REPLACED WITH THE WRONG TYPE OR CONNECTED IMPROPERLY.
Dispose of the lead acid battery according to local laws and regulations.
2.2.5 Mounting Bracket to T31XW
Align the mounting bracket over the threaded holes in the side of the indicator and install the knobs. Adjust the indicator to the desired angle and tighten the knobs.
2.3 Internal Connections
Some connections require the housing to be opened.
2.3.1 Opening the Housing
CAUTION: ELECTRICAL SHOCK HAZARD. REMOVE ALL POWER CONNECTIONS TO THE INDICATOR BEFORE SERVICING OR MAKING INTERNAL CONNECTIONS. THE HOUSING SHOULD ONLY BE OPENED BY AUTHORIZED AND QUALIFIED PERSONNEL, SUCH AS AN ELECTRICAL TECHNICIAN.
EN-12 3000 Series Indicators
T31P
Remove the four Phillips head screws from the rear housing. Open the housing being careful not to disturb the internal connections. Once all connections are made, reattach the front housing.
T31XW
Remove the four hex head screws from the rear housing. Open the housing by carefully pulling the top of the front housing forward. Once all connections are made, reattach the front housing. The screws should be tightened fully to maintain a watertight seal.
2.3.2 Scale Base to T31P or T31XW
Pass the load cell cable through the strain relief (Figure 1-1, item 9 or Figure 1-2, item 10) and attach it to terminal block J5 (Figure 1-3, item 5). Re-tighten the strain relief to ensure a watertight seal.
Jumper Connections
For a 4-wire load cell with no sense wires: Jumpers W2 and W3 must be shorted. For a 6-wire load cell that includes sense wires, see Figure 2-2. Jumpers W2 and W3 must be opened. For load cells with an extra ground shield wire: Connect the shield to the center position (GND) of J5.
Pin Connection
J5-1 +EXCITATION
J5-2 +SENSE
J5-3 +SIGNAL
J5-4 GND
J5-5 -SIGNAL
J5-6 -SENSE
J5-7 -EXCITATION
Figure 2-2. Jumper Connections.
After wiring is completed and jumpers are in place, replace the indicator housing screws. Make sure the strain relief is properly tightened.
2.3.3 RS232 Interface Cable to T31XW
Pass the optional RS232 cable through the strain relief (Figure 1-2, item 9) and attach it to terminal block J7 (Figure 1-3, item 7). Re-tighten the strain relief to ensure a water tight seal.
Pin Connection
J7-1 RXD
J7-2 TXD
J7-3 GND
Loading...