Quick Start Guide CL Series Scales
PREPARATION FOR USE
Release Locks
Before using the scale, release the
Transportation Lock. On scale bottom,
slide the locking pin to the position.
Check Power
Ensure voltage identification label matches
local line voltage. If not, do not connect the
unit. Contact Ohaus or your dealer.
To use battery power, remove battery cover
and insert three LR6 (AA) batteries.
Allow at least 30-60 seconds warm-up time,
for scale to stabilize before operation.
SELECTING THE LOCATION:
Place on a firm, steady surface with sufficient space.
Remove
Battery
Cover to
insert 3
Batteries:
LR6 (AA).
Power Input Jack
Transportation
Lock
Avoid excessive air current, vibrations, heat sources, or rapid temperature changes.
SAFETY PRECAUTIONS
For dependable operation of this product, observe the following precautions:
• Verify that the input voltage printed on the power Adapter and the plug type
matches the local power supply (if using AC power).
• Make sure that the power cord does not pose an obstacle or tripping hazard.
• Disconnect the product from the power supply when cleaning.
• Do not operate this product in hazardous or unstable environments.
• Do not drop loads on the pan.
• Use only approved accessories and peripherals, as available.
• Operate the product only under ambient conditions specified in the Instruction
Manual.
• Service should be performed by authorized personnel only.
Ohaus Corporation www.ohaus.com i CL Series Scales Quick Start Guide

CL Series Scales Quick Start Guide
DISPLAY
Negative sign
Stability
indicator
OPERATION
Place the scale on a horizontal flat surface.
Power Up test: All display segments appear briefly on power up.
Once 0 appears, the scale is ready for use.
The on left indicates the a stable reading.
Press UNIT/CAL to switch weighing units: ounces, troy ounces, penny weight, grams
Overload: If the applied load exceeds the capacity of the scale,
E appears. Remove the load immediately.
The scale will return to normal operation.
TARE: Values can be added progressively to a sample:
Press ON/ZERO OFF, the scale display returns to zero.
Add more weight, press ON/ZERO OFF to zero the scale again.
Add weight up to the capacity of the scale.
Symbols for:
Ounces,
Troy ounces,
Pennies,
Grams,
Place a container on the platform (eg, 20 grams).
Press ON/ZERO OFF. The display shows 0.00.
Add material to the container (eg, 25 grams).
Display shows weight of material – not the container.
When container and material are removed, display will show
weight of container as negative value. To return to normal
operation, press ON/ZERO OFF again.
Low Battery: The display shows Lo when the batteries are weak
and need to be replaced.
*80253049*
P/N 80253049
CL Series Scales Quick Start Guide ii Ohaus Corporation www.ohaus.com