EN-1
INTRODUCTION
These instructions contain information on the installation, setup, and operation of the RS485/422 Serial Interface Option Kit used with Ohaus CKW Series Indicators and Scales. The kit includes all necessary components to modify the CKW-55 Indicator to communicate to external equipment using RS485/422 formats.
Each kit contains a PC board, 4 mounting screws, 2 interconnecting cables and this instruction manual.
APPLICATIONS
The RS485/422 Serial Interface provides a wider flexibility over basic RS232 serial communication systems. Typical uses include creation of multiple scale systems and networks, operation of long distance point-to-point serial communication. Other benefits include:
• Communication and control of multiple scales using a single cable system.
• Communications requiring long cable runs, (up to 4000 feet, 1200 meters).
• Operation in environments that are prone to high electrical noise.
• Multi-drop and Multi-point system connection capability.
TECHNICAL INFORMATION Operating Standard
The CKW RS485/422 Serial Interface follows the EIA-485 Industry Standard.
Interface Cables
For RS485/422 systems, high quality, twisted pair cables should be used for all but the shortest cable distances. Cable impedance should be close to 120 ohms. Category 5 cables with an impedance of 100 ohms are generally acceptable.
The CKW RS485/422 option supports 4-wire duplex or 2-wire half duplex operation. Examples of these configurations are shown in Appendix A.
Termination Resistance
The CKW RS485/422 Option can provide a termination resistance if required. Larger RS485 systems may require a termination resistance to improve the reliability of the cable system. Adding a termination resistance helps absorb excess signal energy that might otherwise reflect back into the system and cause undesirable noise. When used, a termination resistance is usually placed on the last RS485 device located at the end of a cable system.
EN-2
CKW Serial Interface Option Kit
Ground Connections
A signal ground connection is recommended but not always required. The use of a signal ground connection will help improve system reliability and noise immunity. Systems without a signal ground can tolerate common mode voltage differences of –7V to +12V. To reduce high common mode voltages, larger multi­point systems may require the use of a signal ground wire with series resistors inserted between each connection. The resistors reduce the currents that could flow in systems with high common mode voltage differences. Resistor values of 100 ohms are typical.
Use of RS485 Addresses
The Industry Standard EIA-485 allows for driving up to 32 devices on a single cable. The CKW RS485/422 interface option will allow up to 100 unique addresses to be configured. The actual number of devices able to operate reliably on a single cable will be dependent on the Master device’s drive capability, the cable quality, the data speed, and the cable length.
The use of scale addresses above the actual number of network connections would allow for simplified swapping of scales on an existing network. Scales could be swapped with no risk of repeated addresses on a network. Multiple network systems could also be created with unique scale addresses to give absolute scale identification.
The address may be disabled to allow compatibility with RS422 point-to-point or receive only multi-drop systems.
CKW RS485/422 and RS232 Interfaces
When the CKW RS485/422 option is setup, the original RS232 output cannot be used. This is because part of the original RS232 system is also used to support the RS485/422 interface. Cable connections to both serial interfaces can remain in place, but only one serial interface may be used at a time. Before use, the required interface must be setup in the COM menu.
INSTALLATION
CAUTION: ELECTRICAL SHOCK HAZARD. REMOVE ALL POWER CONNECTIONS TO THE INDICATOR BEFORE SERVICING OR MAKING INTERNAL CONNECTIONS. THE HOUSING SHOULD ONLY BE OPENED BY AUTHORIZED AND QUALIFIED PERSONNEL, SUCH AS AN ELECTRICAL TECHNICIAN.
CKW Serial Interface Option Kit
INSTALLATION (Cont.)
Serial Interface Board Installation
To install the optional Serial Interface board, proceed as follows:
• Unfasten the side screws of the indicator housing and open the front cover.
• Mount the Serial Interface board to the rear indicator housing as shown in Figure 1 using 4 of the small screws supplied.
. .
EN-3
Figure 1. Serial Interface Board shown with Relay Output Board Connections.
Internal Cable Installation
• Connect the long cable supplied with the kit to connector J10 on the main PC board. Connect the other end of the cable to either connector at the top of the RS485/422 Board. (See Figure 1). Using the cable tie supplied, attach the cable to the plastic mount at the bottom of the housing. (Other cables may already be attached to this same mount.) If the optional Relay board has been installed, use the supplied short cable in place of the long cable. Connect the short cable between the RS485/422 Board and the similar connector on the Relay board.
EN-4
CKW Serial Interface Option Kit
Wiring to the Serial Interface Board
The wire set used for connection to the Serial Interface option PCB should be part of a multi-conductor cable with a round outer insulating jacket. The overall cable diameter will need to be between 0.157 in. to 0.315 in., (4 mm to 8 mm).
• The terminal connections require wire sizes in the range of 26 to 16 AWG,
(0.14 mm2 to 1.5 mm2). Wire ends are stripped to 0.25 in. (6 mm).
• The cable set from the RS485/422 may be routed directly to the liquid-tight
fitting at the left side of the Indicator housing. If this fitting is not available, use the fitting to the far right instead.
• Connect the cable wires as indicated by the signal names on the PCB.
• Install and securely tighten a cable tie on to the cable just above the liquid-
tight fitting.
• After making the proper connections, tighten the liquid-tight connector
securely around the cable outside of the case.
• Reinstall the front cover and make sure the liquid-tight connector is tight.
• Refer to the CKW-55 Indicator Instruction manual for operational details.
CAUTION: Do not run high voltage wiring within the same cable set that connects
to the serial communication circuits. To maintain a liquid-tight enclosure, do not run more than one cable through each liquid-tight fitting.
SETUP Jumpers
Two shorting jumpers are provided on the RS485/422 PCB. Jumper W1 is used to select the 2 or 4-wire operating mode. Jumper W2 provides for an optional termination resistance. Refer to Figure A.1 in Appendix A for W1 and W2 jumper locations. Figure 2 below shows the proper positioning of the jumper components.
Figure 2. Open and Shorted Jumpers.
CKW Serial Interface Option Kit
EN-5
Jumpers (Cont.)
The jumpers are used as follows:
W1: Open = 4-wire operating mode. Shorted = 2-wire operating mode. W2: Open = No Termination Load. Shorted = 120 Ohm Termination Load.
Note: When W2 is shorted, the termination resistance is placed between the Rx+ and RX- lines.
CAUTION: The use of multiple termination resistors may reduce the total number of devices that can reliably operate on a single cable system.
Other Consideration
Avoid running RS485/422 system cables near other cables that may carry high currents. High currents on nearby cables could induce noise voltages on the RS485/422 cabling. If close contact is unavoidable, cross cables at right angles to minimize interference. The use of high quality shielded cable will also minimize undesired noise pickup.
On a typical long cable system, tap points or “T” connections should be kept to a minimum length to reduce signal reflections.
Some RS485 systems may require additional bias resistors on the signal lines. The CKW RS485/422 Option does not provide for these connections. If this type of connection is required it will need to be provided by the main controlling device or host system.
Menu Settings
This section is similiar to the information appearing in the CKW Instruction manual. Figure 3 illustrates the structure of the CKW Communication Menu.
The Communication menu is used to configure communications parameters of the serial port.
COM
Reset: No, Yes Baud Rate: 300, 600, 1200, 2400,
4800, 9600,19200 Parity: None, Even, Odd Data Length: 7, 8 Stop Bit: 1, Handshake: None, XON/XOFF Type: RS232, RS485 Alt Command: Print, Tare, Zero
End
22
2
22
Figure 3. Communication Menu.
EN-6
CKW Serial Interface Option Kit
Menu Settings (Cont.)
Only commands shown in Table 1, Serial Interface Command Table, are acknowledged by the indicator. The default serial output format from the RS485 interface is shown as follows:
Field: @ Address Polarity Space Weight Space Units Stability CR LF
Length: 1 2 1 1 7 1 5 1 1 1
Definitions: “@”: The “@” character and 2 digit address are automatically
added when a RS485 address is enabled. Address: A 2 digit number from 00 to 99. Polarity: “-” sign if negative, blank if positive. Weight: Up to 6 digits and 1 decimal, leading zero blanking. Units: Up to 5 characters. Stability: “?” character is printed if not stable, blank if stable.
Reset
Used to reset the Communication menu to factory defaults. Selections are: No, Yes.
Baud Rate
Used to define Baud rate. Selections are: 300, 600, 1200, 2400, 4800, 9600 or 19200
..
.
..
Parity
Used to define parity. Selections are:
None, Odd or Even.
Data Length
Used to define data length. Selections are:
7 or 8.
Stop Bit
Used define the number of stop bits. Selections are: 1 or 2.
CKW Serial Interface Option Kit
Handshake
Used to define the flow control method during serial communication. Selections are: None or XON/XOFF (ON-OFF).
Type
Used to define the type of serial communication interface to be used. Selections are:
RS232 = Use only the default RS232 interface. RS485 = Use only the optional RS485/422 interface.
When RS485 is selected, the next sub-menu allows turning the address mode ON or OFF. The selections are:
AdrON = The user may enter a two digit address from 00
to 99. The default address is 01.
AdrOFF = The interface operation will be compatible with
RS422 standards with no address requirements.
Note: During RS485 operation, the selected address is used to identify the scale that is sending or receiving serial data. Refer to section “OPERATION” for more information.
EN-7
Alternate Command
Used to define alternate command characters for Print, Tare and Zero.
END
Represents the end of the menu.
EN-8
CKW Serial Interface Option Kit
TABLE 1. SERIAL INTERFACE COMMAND TABLE.
Command Description Character
PU Print current weighing unit: kg, g, lb, oz, lb:oz, %. P Same as pressing PRINT button. T Same as pressing TARE button. Z Same as pressing ZERO button. xS Enable print stable data only. Where x=0 for Off, x=1 for On. xP Auto print condition. Where x = 1 to 3600 second intervals,
x=0 for Off, x=C for Continuous, x=S for On Stable.
xCM Checkweigh Input Mode, where x=0 for Range, 1=Variance,
2=Percent (%).
xCU Set Under value, as a Range, Variance or Percent value
(depending on Checkweigh Input Mode setting).*
xCO Set Over value, as a Range, Variance or Percent value
(depending on Checkweigh Input Mode setting).*
xCT Set Target value (used only in Variance and Percent Input
Modes, not used in range Mode)*.
xU Set weighing unit: Where x=1 to 6, 1=g, 2=kg, 3=lb, 4=oz,
5=lb:oz, 6=%, if x=blank, then increment to next available mode. (The unit can not be selected if it is not enabled).
PV Print model name, software revisions, and LFT status - if on, (eg.
“CKW-55 Sr1.28 RS485 Sr1.25”
xT Set a pre-set tare value of x, in present weighing units*
* Up to 6 numeric characters, an optional “-” sign, and single decimal point
(e.g., 1.000CU, 2:3.2CU) Use colon “:” as a separator for lb:oz data entry.
Note: All commands must be followed by a carriage return, <cr>, or carriage return-line feed, <cr><lf>, sequence. For addressed RS485 communication, all commands must be preceded by the “@” character and a two digit address.
CKW Serial Interface Option Kit
EN-9
OPERATION Sending Commands to the RS485 Interface
All commands to be sent to the specific RS485 must be preceded by the “@” character and the two digit address set up in the COM menu. For example, to send a print command to the RS485 that is setup as address 01, the user must send the following sequence: “@01P”.
Receiving Data from the RS485 Interface
Each response from the RS485 is preceded by the @ character and the two digit address set up in the COM menu. For example, the same setup as above might respond to the print command with the following reply: “@01 10.005 kg “.
Broadcast Commands
Using the address “00” with a command will act as a broadcast command, and received by all systems on the RS485 network. Under normal circumstances there is no response from a broadcast command. This is to insure that there is no uncontrolled data being placed on the line.
If the address of a system is intentionally set to 00, then there may be a response from this one system following a broadcast command. This option should be used with care since the response will also contain the 00 broadcast address.
REPLACEMENT PARTS
Description Part Number Serial Interface Printed Circuit Board 71169391 Long Option Cable Assembly 71170778 Short Option Cable Assembly 71170779
Service Information
For Service assistance in the United States and Canada, please call Ohaus Corporation toll free at (800) 526-0659 from 8 AM to 5 PM Eastern Time. An Ohaus Product Service Specialist will be available to help you.
EN-10
APPENDIX A Typical RS485 Connections
Figure A.1. Typical 4-Wire Duplex Connection.
CKW Serial Interface Option Kit
Figure A.2. Typical 2-Wire Half Duplex Connection (W1 Shorted).
Kit opcional de interfase serial CKW
ES-1
INTRODUCCIÓN
Estas instrucciones contienen información sobre la instalación, configuración y operación del kit opcional de interfase serial RS485/422 utilizado con los Indicadores y Básculas de la serie CKW de Ohaus. El kit incluye todos los componentes necesarios para modificar el Indicador CKW-55 para que se comunique con el equipo externo utilizando formatos RS485/422.
Cada kit contiene una tarjeta PC, 4 tornillos de montaje, 2 cables de interconexión y este manual de instrucciones.
APLICACIONES
La interfase serial RS485/422 proporciona una gran flexibilidad sobre los sistemas básicos de comunicación serial RS232. Los usos típicos incluyen la creación de sistemas de múltiples básculas y redes de trabajo, y operación de comunicación serial punto a punto de larga distancia. Otros beneficios incluyen:
• Comunicación y control de múltiples básculas utilizando un sistema de cable individual.
• Comunicaciones que requieren acometidas de cables largos (hasta 4000 pies, 1200 metros).
• Operación en ambientes propensos a ruidos eléctricos altos.
• Capacidad de conexión de sistema multipunto.
INFORMACIÓN TÉCNICA Estándar de operación
La interfase serial CKW RS485/422 obedece al Estándar de la Industria EIA-485.
Cables de interfase
Para sistemas RS485/422, deben utilizarse cables de pares trenzados de alta calidad para todas las distancias de cable menos las más cortas. La impedancia del cable debe ser cercana a los 120 ohmios. Los cables de categoría 5 con una impedancia de 100 ohmios son generalmente aceptables.
La opción CKW RS485/422 soporta la operación dúplex de 4 cables o la operación semi-dúplex de 2 cables. Ejemplos de estas configuraciones se muestran en el Apéndice A.
Resistencia de la terminación
La opción CKW RS485/422 puede proporcionar una resistencia de terminación si se requiere. Los sistemas RS485 más grandes pueden requerir una resistencia de terminación para mejorar la confiabilidad del sistema de cableado. Adicionar una resistencia de terminación ayuda a absorber el exceso de energía de la señal que podría de otra forma rebotar dentro del sistema y causar un ruido indeseable. Cuando se le utiliza, una resistencia de terminación es colocada usualmente en el último dispositivo RS485 ubicado al final de un sistema de cableado.
ES-2
Kit opcional de interfase serial CKW
Conexiones a tierra
Se recomienda una conexión a tierra de la señal pero no siempre se requiere. El uso de una conexión a tierra de la señal ayudará a mejorar la confiabilidad del sistema y la inmunidad al ruido. Los sistemas sin una tierra de la señal pueden tolerar diferencias de voltaje de modo común de –7 V a +12 V. Para reducir los voltajes de modo común altos, los sistemas multipunto más grandes pueden requerir el uso de un cable a tierra de la señal con resistencias en serie insertadas entre cada conexión. Las resistencias reducen las corrientes que podrían fluir en sistemas con diferencias altas del voltaje de modo común. Son típicas las resistencias con valores de 100 ohmios.
Uso de las direcciones RS485
El Estándar de la Industria EIA-485 permite un manejo hasta de 32 dispositivos en un cable individual. La opción de interfase CKW RS485/422 permitirá que se configuren hasta 100 direcciones individuales. El número real de dispositivos capaces de operar confiablemente en un cable individual dependerá de la capacidad de manejo del dispositivo Maestro, la calidad del cable, la velocidad de datos y la longitud del cable.
El uso de direcciones de báscula por encima del número real de conexiones de red permitiría un intercambio simplificado de básculas en una red de trabajo existente. Las básculas podrían intercambiarse sin riesgo de direcciones repetidas en una red de trabajo. Los sistemas de múltiples redes de trabajo también podrían crearse con direcciones únicas de báscula para permitir una identificación absoluta de la báscula.
La dirección puede desactivarse para permitir la compatibilidad con RS422 punto a punto o recibir sólo sistemas multipunto.
Interfases CKW RS485/422 y RS232
Cuando está configurada la opción CKW RS485/422, no puede utilizarse la salida original RS232. Esto se debe a que parte del sistema RS232 original también es utilizado para soportar la interfase RS485/422. Las conexiones de cable en ambas interfases seriales pueden permanecer en su puesto, pero sólo puede utilizarse una interfase serial en un momento dado. Antes de usarse, la interfase requerida debe configurarse en el menú COM.
INSTALACIÓN
PRECAUCIÓN: EXISTE EL RIESGO DE DESCARGA ELÉCTRICA. RETIRE TODAS LAS CONEXIONES DE ALIMENTACIÓN AL INDICADOR ANTES DE DAR SERVICIO O HACER CONEXIONES INTERNAS. LA CAJA SÓLO DEBE SER ABIERTA POR PERSONAL AUTORIZADO Y CALIFICADO, COMO UN TÉCNICO ELÉCTRICO.
Kit opcional de interfase serial CKW
INSTALACIÓN (continuación)
Instalación de la tarjeta de interfase serial
Para instalar la tarjeta de interfase serial opcional, proceda como se indica a continuación:
• Afloje los tornillos laterales de la caja del indicador y abra la cubierta frontal.
• Utilizando los 4 tornillos pequeños suministrados, instale la tarjeta de interfase serial a la parte posterior de la caja del indicador como se muestra en la Figura 1.
CAJA DEL INDICADOR
CABLE CORTO OPCIONAL
ALIMENTACIÓN
TARJETA
RS-422/485
. .
TARJETA
DE SALIDA
DE RELEVO
OPCIONAL
ELÉCTRICA
ES-3
AMARRE DE CABLE
CABLE LARGO AL CONECTOR
J10 EN LA TARJETA PC
PRINCIPAL
A LOS CIRCUITOS
DE COMUNICACIÓN
EXTERNA
POSICIÓN DE CABLE
OPCIONAL
Figura 1. Tarjeta de interfase serial mostrada con conexiones
a la tarjeta de salida de relevo.
Instalación interna del cable
• Conecte el cable largo suministrado con el kit al conector J10 en la tarjeta PC principal. Conecte el otro extremo del cable a cualquier conector en la parte superior de la tarjeta RS485/422. (Vea Figura 1). Utilizando el amarre de cable suministrado, fije el cable a la montura plástica en la parte inferior de la caja. (Otros cables pueden estar ya fijos a ésta misma montura.) Si la tarjeta de relevo opcional ha sido instalada, utilice el cable corto suministrado en lugar del cable largo. Conecte el cable corto entre la Tarjeta RS485/422 y el conector similar en la tarjeta de relevo.
ES-4
Kit opcional de interfase serial CKW
Cableado a la tarjeta de interfase serial
El juego de cables utilizado para la conexión a la PCB de interfase serial opcional debe ser parte de un cable multi-conductor con una envoltura aislante. El diámetro general del cable necesitará ser de 0.157 in. a 0.315 in., (4 mm a 8 mm).
• Las conexiones de la terminal requieren tamaños de cable en el rango
26 a 16 AWG, (0.14 mm cortados (pelados) a 0.25 in. (6 mm).
• El conjunto de cables de la RS485/422 puede ser encaminado directamente
al acople hermético en la parte lateral de la caja del Indicador. Si no está disponible este acople, utilice el cierre hermético del extremo derecho.
• Conecte los alambres del cable como se indica por los nombres de las
señales en la PCB.
• Instale y fije firmemente un amarre de cable al cable justo encima del acople
hermético.
• Después de hacer las conexiones apropiadas, apriete firmemente el conector
hermético sobre el cable fuera de la caja.
• Reinstale la cubierta frontal y asegúrese de que el conector hermético está
apretado.
• Refiérase al manual de instrucciones del indicador CKW-55 para detalles
operativos.
PRECAUCIÓN: No coloque cableados de alto voltaje dentro del mismo conjunto
de cables que se conecta a los circuitos de comunicación serial. Para mantener un cierre hermético, no coloque más de un cable a través del acople hermético
2
a 1.5 mm2). Los extremos del cable necesitan ser
CONFIGURACIÓN Puentes
Se proporcionan dos puentes para corto con la PCB RS485/422. El puente W1 es utilizado para seleccionar el modo de operación con 2 ó 4 cables. El puente W2 proporciona una resistencia de terminación opcional. Refiérase a la Figura A.1 en el Apéndice A para las ubicaciones de los puentes W1 y W2. La Figura 2 abajo muestra el posicionamiento adecuado de los componentes del puente.
Ó
PUENTES ABIERTOS PUENTE EN CORTO
Figura 2. Puente abierto y en corto.
Kit opcional de interfase serial CKW
ES-5
Puentes (Cont.)
Los puentes son utilizados como sigue:
W1: Abierto = modo de operación En corto = modo de operación
con 4 cables. con 2 cables.
W2: Abierto = sin carga En corto = carga de terminación
de terminación. de 120 ohmios.
Nota: cuando el W2 está en corto, la resistencia de terminación está colocada entre las líneas Rx+ y RX-.
PRECAUCIÓN: el uso de múltiples resistencias de terminación puede reducir el número total de dispositivos que pueden operar confiablemente en un sistema de cable individual.
Otra consideración
Evite la acometida de cables del sistema RS485/422 cerca de otros cables que puedan llevar corrientes altas. Las corrientes altas en la cercanía de los cables pueden inducir voltajes de ruido sobre el cableado RS485/422. Si es inevitable el contacto estrecho, cruce los cables en ángulos rectos para minimizar la interferencia. El uso de cables con blindaje de alta calidad también minimiza la absorción de ruido indeseado.
En un típico sistema de cable largo, los puntos de conexión intermedia o conexiones “T” deben mantener una longitud mínima para reducir los rebotes de la señal.
Algunos sistemas RS485 pueden requerir resistencias de polarización adicionales en las líneas de la señal. La opción CKW RS485/422 no permite estas conexiones. Si se requieren estas conexiones, será necesario que sean proporcionadas por el dispositivo de control principal o el sistema central.
COM
Opciones de configuración del menú
Esta sección es similar a la información que aparece en el manual de instrucciones CKW. La Figura 3 ilustra la estructura del menú de comunicación de CKW.
El menú de comunicación es utilizado para configurar los parámetros de comunicación del puerto serial.
Reestablecer: No, Sí Velocidad de 300, 600, 1200, 2400, transmisión : 4800, 9600, 19200 Paridad: No, Par, Impar Longitud de datos: 7, 8 Bit de parada: 1, Contacto para intercambio de datos: No, XON/XOFF Tipo: RS232, RS485 Comando alterno: Imprimir, Tara, Cero Final
Figura 3. Menú de comunicación.
22
2
22
Loading...
+ 36 hidden pages