ISS-2
Integrated Sampling System
Installation and Operation Instructions
Description
The ISS-2 Integrated Sampling System is a fully integrated 1-cm cuvette holder and tungsten halogen
light source. The ISS-2 couples to Ocean Optics spectrometers via an optical fiber to create a smallfootprint system for VIS-NIR (approximately 360 to 1100 nm) measurements.
The ISS-2 light source features a 900-hour bulb. Additionally, the fan in the ISS-2 is exposed (not
enclosed in the base) and requires particular care when handling.
134-00000-000-01-201010 1

ISS-2 Installation and Operation Instructions
Parts Included
The ISS-2 Integrated Sampling System ships with the following components:
ISS-2 cuvette holder and light source assembly
12 VDC power supply
1-cm square plastic cuvette
Screwdriver (for adjusting cuvette fit)
Allen wrench (for adjusting collimating lens on light source)
WARNING
The Light Source in the ISS-2 gets extremely hot during operation. After use, allow
sufficient time for the lamp to cool before handling the ISS-2.
Installing the Sampling System
Assembling the ISS-2
Ocean Optics packages the cuvette holder and the ISS-2 light source as separate components that you
must manually connect.
To assemble the ISS-2, screw the cuvette holder onto the SMA-905 connector of the ISS-2 light
source until snug.
Attaching Fibers to the ISS-2
► Procedure
Follow the steps below to attach fibers to the ISS-2:
1. Connect an SMA 905-terminated optical fiber to the collimating lens of the ISS-2’s cuvette
holder.
2. Connect the other end of the fiber (the read fiber) to the SMA 905 connector of the
spectrometer.
Adjusting the Fit of the Cuvette
The ISS-UV-VIS is designed to hold 1-cm square cuvettes. When properly adjusted, the cuvette
should fit snugly into the holder.
2 134-00000-000-01-201010