INTSMA Slit
Installation Instructions
Description
Ocean Optics Jaz, Torus and QE65 Pro Spectrometers offer a special feature that make them even
more customizable for your unique application. Part of the versatility of these spectrometers lies in
their replaceable laser-cut slit and aperture assembli es (I N TSMA). Your spec tr om ete r arrived w ith the
slit size specified in your order. However, if your needs have changed, you can order additional
replacement slits (in various widths from 5 µm to 200 µm) and install the new slit using these
instructions. Be sure to use a fiber with a diameter that matches the slit size to optimize throughput.
The INTSMA slit kits offer you the following advantages:
• Adjustable Throughput
The spectrometer throughput can be set for applications such as absorbance and fluorescence.
For example, QE65 Pro users who need great sensitivity for low light applications can change
slits to avoid saturation in absorbance applications.
• Adjustable Op t ica l Re s ol u tio n
Optical resolution is a function of grating groove density and slit size. By changing slit size,
you can observe different Raman shift resolution with a single QE65Pro spectro m eter.
• Multiple Uses for Your Spectr om ete r
For example, a Jaz spectrometer can be changed from an absorbance spec tropho tometer in the
lab to a fluorescence unit in the field. Or a Torus configured to resolve closely aligned
emission peaks for one application and make low light level measurements for another.
• Field-Changeable
Our laser-cut slit and aperture assemblies can be changed in the field, eliminating the need to
send us your spectrometer for rework. Changing slits can be as easy as removing a few
screws. You can also send us your spectrometer to change the slit size and take advantage of
the opportunity to purchase one of our calibration and maintenance packages.
Note
If you are changing to a smaller slit size, you will need to have your spectrometer
recalibrated for accurate measurements. Best practice would be to validate the
spectrometer calibration using an HG-1 Mercury Argon Calibration Source (or similar
light source) whenever the slit size is altered.
013-INTSMA-000-01-201209 1
INTSMA Slit Installation Instructions
Additional Resources
For more information on Jaz, see the Jaz Installation and Operation Manual.
For more information on Torus, see the Torus Installation and Operation Manual.
For more information on the QE65 Pro, see the QE65 Pro Installation and Operation Manual.
To speak to a Technical Support represen tativ e or an Ocean Opt ics Appl ic at ions Scien tist, ple ase ca ll
(727) 733-2447. Our office hours are 8 a.m. – 8 p.m. (Mon-Thu), 8 a.m.– 6 p.m. (Fri) EST.
INTSMA Kits Available
The INTSMA kits available include the following:
• INTSMA-000: Interchangeable round SMA with no slit installed
• INTSMA-005: Interchangeable round SMA 5 µm
• INTSMA-010: Interchangeable round SMA 10 µm
• INTSMA-010S: Interchangeable round SMA 5 µm small aperture
• INTSMA-025: Interchangeable round SMA 25 µm
• INTSMA-050: Interchangeable round SMA 50 µm
• INTSMA-100: Interchangeable round SMA100 µm
• INTSMA-200: Interchangeable round SMA 200 µm
Package Contents
The INTSMA kit ships with the following components:
• INTSMA slit
• Four (4) 0-80 screws (2 for slit and 2 spares)
Tools Needed
• 0.050 inch hex driver (Allen wrench)
2 013-INTSMA-000-01-201209