All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted, by any means, electronic,
mechanical, photocopying, recording, or otherwise, without written permission from Ocean Optics, Inc.
This manual is sold as part of an order and subject to the condition that it shall not, by way of trade or otherwise, be lent, re-sold, hired out or
otherwise circulated without the prior consent of Ocean Optics, Inc. in any form of binding or cover other than that in which it is published.
Trademarks
Microsoft, Windows, Windows 95, Windows 98, Windows Me, Windows NT, Windows 2000, Windows XP and Excel are either registered
trademarks or trademarks of Microsoft Corporation.
Limit of Liability
Every effort has been made to make this manual as complete and as accurate as possible, but no warranty or fitness is implied. The
information provided is on an “as is” basis. Ocean Optics, Inc. shall have neither liability nor responsibility to any person or entity with respect
to any loss or damages arising from the information contained in this manual.
System Requirements ...................................................................................................................................................................................1
About OOIBase32 .........................................................................................................................................................................................2
Sampling System Overview ..........................................................................................................................................................................2
Packing List ...................................................................................................................................................................................................3
Wavelength Calibration Data Sheet ..............................................................................................................................................................3
Software and Technical Resources CD ........................................................................................................................................................4
Other Documentation ....................................................................................................................................................................................4
Modification and Repair Policy ......................................................................................................................................................................4
How the HR2000 Works ........................................................................................................................................ 5
System Compatibility.....................................................................................................................................................................................8
20-Pin J2 Accessory Connector - Part Number and Compatibility................................................................................................................9
Chapter 3: Installing the HR2000........................................................................................................10
Configuring the HR2000 in OOIBase32 ......................................................................................................................................................11
HR2000 Connected to PC Prior to OOIBase32 Installation ................................................................................ 14
Remove the Unknown Device from Windows Device Manager ..................................................................................................................14
Windows 98:................................................................................................................................................................................................15
Windows 2000/XP: ......................................................................................................................................................................................16
Troubleshooting the Serial Port Configuration..................................................................................................... 17
Older Version of OOIBase32 Installed ................................................................................................................ 17
Preparing for Experiments................................................................................................................................... 18
Time Acquisition Experiments ............................................................................................................................. 28
Appendix A: Calibrating the Wavelength of the HR2000 .................................................................31
About Wavelength Calibration ............................................................................................................................. 31
Calibrating the Spectrometer............................................................................................................................... 31
Preparing for Calibration .............................................................................................................................................................................31
Calibrating the Wavelength of the Spectrometer.........................................................................................................................................32
Saving the New Calibration Coefficients: USB Mode..................................................................................................................................34
Saving the New Calibration Coefficients: Serial Mode ................................................................................................................................35
HR2000CG-UV-NIR Features ............................................................................................................................. 36
Index ......................................................................................................................................................37
HR2000 Introduction
1 HR2000 Introduction
The following chapter contains introductory information about the HR2000 Spectrometer, shipment information,
and spectrometer connectivity.
Product Overview
The HR2000 High-resolution Miniature Fiber Optic Spectrometer provides optical resolution as good as 0.065 nm
(FWHM). The HR2000 is responsive from 200-1100 nm, but the specific range and resolution depends on your
grating and entrance slit selections.
The HR2000 is perfect for applications where high resolution is necessary, such as absorbance of gases or
atomic emission lines (for solution chemistry or for color measurements, the S2000, PC2000 and USB2000 are
more appropriate spectrometers).
Data programmed into a memory chip on each HR2000 includes wavelength calibration coefficients, linearity
coefficients and the serial number unique to each spectrometer. Our spectrometer operating software simply
reads these values from the spectrometer -- a feature that enables hot swapping of spectrometers among PCs.
The HR2000 Spectrometer connects to a notebook or desktop PC via USB port or serial port. When connected to
the USB port of a PC, the HR2000 draws power from the host PC, eliminating the need for an external power
supply.
You can use the HR2000’s USB connectivity with any PC that meets the following requirements:
• Windows 98/Me/2000/XP operating system (or Windows CE 2.11 or later for handheld PCs)
• Ocean Optics OOIBase32 software application installed and configured for use with the HR2000
(OOIPS2000 software if using a handheld PC). Consult the “Configuring the HR2000 in OOIBase32”
section of Chapter 3: Installing the HR2000 for specific configuration instructions.
Alternately, the HR2000 has serial port adaptability for connecting to PCs, PLCs, and other devices that support
the RS-232 communication protocol. However, this connection method requires an external power supply to
power the HR2000, as well as a serial port adapter.
HR2000 High-Resolution Fiber Optic Spectrometer 1
HR2000 Introduction
EEPROM Utilization
An EEPROM memory chip in each HR2000 contains wavelength calibration coefficients, linearity coefficients, and
a serial number unique to each individual spectrometer. The OOIBase32 software application reads these values
directly from the spectrometer, enabling the ability to “hot-swap” spectrometers between PCs without entering the
spectrometer coefficients manually on each PC.
About OOIBase32
OOIBase32 is the latest generation of operating software for all Ocean Optics spectrometers and is available free
to all customers. OOIBase32 is a user-customizable, advanced acquisition and display program that provides a
real-time interface to a variety of signal-processing functions. With OOIBase32, you have the ability to perform
spectroscopic measurements (such as absorbance, reflectance, and emission), control all system parameters,
collect and display data in real time, and perform reference monitoring and time acquisition experiments.
Note: When using a handheld PC, you will use the OOIPS2000 software instead of OOIBase32. The
functionality of OOIPS2000 is similar to OOIBase32, but it is specifically for the handheld PC.
Sampling System Overview
Ocean Optics fiber optic spectrometer systems consist of low-cost, modular data acquisition components. A
typical HR2000-based sampling system contains four core elements:
• HR2000 Spectrometer
• OOIBase32 operating software
• Light source
• Sampling optics (varying, depending on application requirements)
How Sampling Works
The following list explains the function of Ocean Optics sampling components in the sampling process:
1. The user stores reference and dark measurements to correct for instrument response variables.
2. The light from the light source transmits through an optical fiber to the sample.
3. The light interacts with the sample.
4. Another optical fiber collects and transmits the result of the interaction to the spectrometer.
5. The spectrometer measures the amount of light and transforms the data collected by the spectrometer
into digital information.
6. The spectrometer passes the sample information to OOIBase32.
7. OOIBase32 compares the sample to the reference measurement and displays processed spectral
information.
Modular Sampling Accessories
Ocean Optics offers a complete line of spectroscopic accessories for use with the HR2000. Most of our
spectroscopic accessories have SMA connectors for application flexibility. Accordingly, changing the sampling
system components is as easy as unscrewing a connector and replacing an accessory.
HR2000 High-Resolution Fiber Optic Spectrometer 2
HR2000 Introduction
Interface Options
The HR2000 has both USB and serial port connectors (with the use of an adapter), enabling you to connect the
spectrometer to a desktop or notebook PC via a USB port or to a desktop, notebook, or to a handheld PC via a
serial port.
Computer
Interface
Desktop or
Notebook PC
via USB Port
Desktop or
Notebook PC
via Serial Port
Handheld PC
via Serial Port
Operating System
Requirements
Windows 98/Me/
2000/XP
Any 32-bit
Windows operating
system
Windows CE 2.11
or higher
Part Needed Description of Part
USB-CBL-1
(included)
USB-ADP-PC
(not included)
USB-ADP-H
(not included)
Cable that connects from USB port on HR2000 to USB port on
desktop or notebook PC
Adapter block that enables connection from serial port on
HR2000 to serial port on desktop or notebook PC; comes with 5
VDC power supply (required when connecting to serial port)
Adapter block that enables connection (with standard 9-pin
serial cable) from serial port on HR2000 to serial port on
handheld PC; comes with 5 VDC power supply (required when
connecting to serial port)
Shipment Components
The following information and documentation ships with the HR2000 Spectrometer:
Packing List
The packing list is inside a plastic bag attached to the outside of the shipment box (the invoice arrives separately).
The packing slip lists all items in the order, including customized components in the spectrometer (such as the
grating, detector collection lens, and slit). The packing list also includes the shipping and billing addresses, as well
as any items on back order.
Wavelength Calibration Data Sheet
Each spectrometer is shipped with a Wavelength Calibration Data Sheet that contains information unique to your
spectrometer. OOIBase32 Operating Software reads this calibration data from your spectrometer when it
interfaces to a PC via the USB port. Any other interface requires that you manually enter the calibration data in
OOIBase32 (select Spectrometer | Configure | Wavelength Calibration tab). See the OOIBase32
documentation for more information (refer to Other Documentation
documentation).
Note: Please save the Wavelength Calibration Data Sheet for future reference.
HR2000 High-Resolution Fiber Optic Spectrometer 3
for instructions on accessing OOIBase32
HR2000 Introduction
Software and Technical Resources CD
Each order ships with the Ocean Optics Software and Technical Resources CD. This disc contains software,
operating instructions, and product information for all Ocean Optics software, spectrometers, and spectroscopic
accessories. You will need Adobe Acrobat Reader version 6.0 or higher to view these files. Ocean Optics includes
Adobe Acrobat Reader on the Software and Technical Resources CD.
With the exception of OOIBase32 Spectrometer Operating Software, all Ocean Optics software requires a
password during the installation process. You can locate passwords for the other software applications on the
back of the Software and Technical Resources CD package.
Other Documentation
You can find detailed instructions for the OOIBase32 Spectrometer Operating Software at
http://www.oceanoptics.com/technical/operatinginstructions.asp.
Note: You can find instructions for the OOIPS2000 Operating Software for the handheld PC at the following
To find operating instructions on some of Ocean Optics’ most popular spectroscopic accessories (including light
sources, sampling chambers, sampling optics, and probes), consult the operating instructions for Ocean Optics
products from the Software and Technical Resources CD that ships with the product, or visit our website at
http://www.oceanoptics.com/technical/operatinginstructions.asp.
Modification and Repair Policy
In the event that you need to return an item for modification or service, you must obtain a Return Merchandise
Authorization (RMA) number prior to shipping the item back to Ocean Optics. Contact an Ocean Optics
Applications Sales Engineer for specific instructions on returning an item.
HR2000 High-Resolution Fiber Optic Spectrometer 4
HR2000 Specifications
2 HR2000 Specifications
This chapter contains information on spectrometer operation, specifications, and system compatibility. It also
includes accessory connector pinout diagrams and pin-specific information.
How the HR2000 Works
Below is a diagram of how light moves through the optical bench of an HR2000 Spectrometer. The optical bench
has no moving parts that can wear or break; all the components are fixed in place at the time of manufacture.
Items with an asterisk (*) are user-specified.
Figure 2-1: HR2000 Spectrometer with Components.
The HR2000 Component Table on the following page explains the function of each numbered component in the
HR2000 Spectrometer diagram (Figure 2-1).
HR2000 High-Resolution Fiber Optic Spectrometer 5
HR2000 Specifications
HR2000 Specifications Table
Ocean Optics permanently secures all components in the HR2000 at the time of manufacture. Only Ocean Optics
Technicians can replace interchangeable components, where noted.
Item Name Description
1 SMA Connector
2 Slit*
3 Filter*
4 Collimating Mirror
5 Grating*
6 Focusing Mirror
L2 Detector
7
Collection Lens*
CCD Detector
8
(UV or VIS)
The SMA Connector secures the input fiber to the spectrometer. Light from the
input fiber enters the optical bench through this connector.
The Slit is a dark piece of material containing a rectangular aperture, which is
mounted directly behind the SMA Connector. The size of the aperture regulates the
amount of light that enters the optical bench and controls spectral resolution.
You can also use the HR2000 without a Slit. In this configuration, the diameter of
the fiber connected to the HR2000 determines the size of the entrance aperture.
Only Ocean Optics technicians can change the Slit.
The Filter is a device that restricts optical radiation to pre-determined wavelength
regions. Light passes through the Filter before entering the optical bench. Both
bandpass and longpass filters are available to restrict radiation to certain
wavelength regions.
Only Ocean Optics technicians can change the Filter.
The Collimating Mirror focuses light entering the optical bench towards the Grating
of the spectrometer.
Light enters the spectrometer, passes through the SMA Connector, Slit, and Filter,
and then reflects off the Collimating Mirror onto the Grating.
The Grating diffracts light from the Collimating Mirror and directs the diffracted light
onto the Focusing Mirror. Gratings are available in different groove densities,
allowing you to specify wavelength coverage and resolution in the spectrometer.
Only Ocean Optics technicians can change the Grating.
The Focusing Mirror receives light reflected from the Grating and focuses the light
onto the CCD Detector or L2 Detector Collection Lens (depending on the
spectrometer configuration).
The L2 Detector Collection Lens (optional) attaches to the CCD Detector. It
focuses light from a tall slit onto the shorter CCD Detector elements.
The L2 Detector Collection Lens should be used with large diameter slits or in
applications with low light levels. It also improves efficiency by reducing the effects
of stray light.
Only Ocean Optics technicians can add or remove the L2 Detection Collection Lens.
The CCD Detector collects the light received from the Focusing Mirror or L2
Detector Collection Lens and converts the optical signal to a digital signal.
Each pixel on the CCD Detector responds to the wavelength of light that strikes it,
creating a digital response. The spectrometer then transmits the digital signal to the
OOIBase32 application.
HR2000 High-Resolution Fiber Optic Spectrometer 6
HR2000 Specifications
HR2000 Specifications
The following sections provide specification information for the CCD detector in the HR2000, as well as the
HR2000 Spectrometer itself.
CCD Detector Specifications
Detector: Sony ILX511 linear silicon CCD array
No. of elements: 2048 pixels
Pixel size: 14 µm x 200 µm
Pixel well depth: 62,500 electrons
Signal-to-noise ratio: 250:1 (at full signal)
A/D resolution: 12 bit
Dark noise: 2.5 RMS counts
Corrected linearity: >99.8%
HR2000 Spectrometer Specifications
Dimensions: 148.6 mm x 104.8 mm x 45.1 mm
Weight: 570 g
Power consumption: 95 mA @ 5 VDC
Detector range: 200-1100 nm
Gratings: 13 gratings available
Entrance aperture: 5, 10, 25, 50, 100 or 200 µm wide slits or fiber (no slit)
Order-sorting filters: Installed longpass and bandpass filters
Focal length: f/4, 101 mm
Optical resolution: Depends on grating and size of entrance aperture
Stray light: <0.05% at 600 nm; <0.10% at 435 nm
Dynamic range: 2 x 108 (system); 2000:1 for a single scan
Fiber optic connector: SMA 905 to single-strand optical fiber (0.22 NA)
Data transfer rate: Full scans into memory every 13 milliseconds
Integration time: 3 milliseconds to 65 seconds
Windows 98/Me/2000/XP when using the USB interface on desktop or
Operating systems:
notebook PCs
Any 32-bit Windows operating system when using the serial port on
desktop or notebook PCs
HR2000 High-Resolution Fiber Optic Spectrometer 7
HR2000 Specifications
System Compatibility
The following sections provide information on hardware and software requirements for the HR2000:
Compatibility for Desktop or Notebook PCs
To use the HR2000, you must have a PC that meets the following minimum requirements:
• IBM-compatible PC with Pentium (or higher) processor
• 32 MB RAM
• OOIBase32 Spectrometer Operating Software
• Windows 98/ME/2000/XP operating system (when connecting the HR2000 to a PC via USB port)
or
Any 32-bit version of Windows (when connecting the HR2000 to a PC via serial port)
Compatibility for Handheld PCs
To use the HR2000 with your handheld PC, the computer must meet the following minimum requirements:
• Handheld PC running Windows CE 2.11 or later
• 32 MB RAM
• OOIPS2000 Spectrometer Operating Software
• Serial port connectivity
20-Pin Accessory Connector Pinout
The HR2000 features a 20-pin Accessory Connector, located on the side of the unit as follows:
20-Pin Connector
Figure 2-2: Location of HR2000 20-Pin Accessory Connector
HR2000 High-Resolution Fiber Optic Spectrometer 8
HR2000 Specifications
20-Pin Accessory Connector Pinout Diagram
When facing the 20-pin Accessory Connector on the HR2000, pin numbering is as follows:
The following table contains information regarding the function of each pin in the HR2000’s 20-Pin Accessory
Connector:
Pin # Description
1 V
2 RS232 Tx
3 RS232 Rx
4 Lamp Enable
5 Continuous Strobe
6 Ground
7 External Trigger In
8 Single Strobe
9 I2C SCL
10 I2C SDA
11 No connect
12 No connect
13 No connect
14 No connect
15 No connect
16 No connect
17 No connect
18 No connect
19 No connect
20 No connect
USB
or 5Vin
20-Pin J2 Accessory Connector - Part Number and Compatibility
The part numbers for the 20-pin accessory connector on the HR2000 Spectrometer are as follows:
• The connector is Samtec model IPT1-110-01-S-D-RA.
• The mating right-angle connector is Samtec model IPS1-110-01-S-D-RA.
If you are customizing your HR2000 Spectrometer system or configuring External Triggering, you may need these
part numbers to complete your setup.
HR2000 High-Resolution Fiber Optic Spectrometer 9
Installing the HR2000
3 Installing the HR2000
This chapter contains instructions in parallel for connecting the HR2000 via both USB and serial modes.
Note: You must install the OOIBase32 software application prior to connecting the HR2000 Spectrometer to the
PC. The OOIBase32 software installation installs the drivers required for HR2000 installation. If you do
not install OOIBase32 first, the system will not properly recognize the HR2000.
If you have already connected the HR2000 to the PC prior to installing OOIBase32, consult Chapter 4: Troubleshooting for information on correcting a corrupt HR2000 installation.
HR2000 Installation
USB Mode
To connect the HR2000 to a PC via the USB port, the
PC must be running the Windows 98/ME/2000/XP
operating system.
Note: The USB port on a PC can power up to five
HR2000 spectrometer channels. Systems with
more than five channels require a powered
USB hub.
Follow the steps below to connect the HR2000 to a
PC via the USB port:
1. Install OOIBase32 on the destination PC.
2. Locate the USB cable (USB-CBL-1) provided
with the HR2000.
3. Insert the square end of the cable into the
side of the HR2000.
4. Insert the rectangular end of the cable into the
USB port of the PC.
If you installed OOIBase32 prior to connecting the
HR2000, the Add New Hardware Wizard appears
and installs the HR2000 drivers. If the drivers do not
successfully install (or if you connected the HR2000 to
the PC before installing OOIBase32), consult Chapter 4: Troubleshooting.
Serial Port Mode
To use the serial port capacity of the HR2000
Spectrometer, the PC must be running a 32-bit
version of the Windows operating system (or Windows
CE 2.11 or higher for handheld PCs).
Follow the steps below to connect the HR2000 to the
PC via serial port:
1. Connect the serial cable adapter block to the
10 pins on the right side of the HR2000’s 20Pin Accessory Connector (pins 1-10).
2. Connect one end of the 9-pin serial cable to
the adapter block on the HR2000, and then
connect the other end to a serial port on the
PC.
3. Note the number of the serial port (COM Port)
to which you connected the HR2000 (some
PCs may not have numbered ports; handheld
PCs typically have only one serial port).
4. Plug the 5 VDC external power supply into an
outlet and connect it to the HR2000.
Note: Connecting the spectrometer to the PC’s
serial port requires that you manually enter
the calibration coefficients from the
Wavelength Calibration Data Sheet into
OOIBase32 software (select Spectrometer | Configure | Wavelength Calibration tab).
See the OOIBase32 documentation for more
information.
Once you install the HR2000, you must configure OOIBase32’s Configure Spectrometer options so that
OOIBase32 recognizes the HR2000 Spectrometer.
Note: Consult the OOIBase32 Spectrometer Operating Software Manual for detailed instructions on configuring
the spectrometer in OOIBase32. You can find these instructions on the Software and Technical
Resources CD or on the Ocean Optics web site at the following location:
http://www.oceanoptics.com/technical/ooibase32.zip.
The following sections contain instructions on initially configuring the HR2000 the first time you start OOIBase32:
Operator and Serial Number Dialog Box
The Operator and Serial Number screen prompts you to enter a user name and software serial number into
OOIBase32. Some data files created by OOIBase32 during sampling procedures use this information in the file
headers.
Default Spectrometer Configuration File
The Default Spectrometer Configuration File screen prompts you to select a spectrometer configuration
(.SPEC) file for use with the HR2000. The unique serial number of the HR2000 precedes the file extension (for
example, HR2A0162.SPEC).
Navigate to the OOIBase32 installation directory and select the spectrometer configuration file.
Configure Hardware Screen
The Configure Hardware screen prompts you to enter spectrometer-specific information into OOIBase32 the first
time you run the program. Typically, you will only enter this information the first time you run OOIBase32.
However, you can alter the hardware configuration at any time using the Spectrometer Configuration screen.
Select Spectrometer | Configure from the OOIBase32 menu bar to access the Spectrometer Configuration
screen.
Follow the steps on the next page to configure the HR2000:
Note: You do not need to configure the spectrometer hardware in the OOIPS2000 handheld PC operating
software. Most handheld PCs have only one serial connector and do not have USB ports. Thus,
OOIPS2000 will communicate with the spectrometer via this port at a fixed baud rate. There is no way to
customize the HR2000 configuration with OOIPS2000.
menu with the serial numbers of all
discovered HR2000 Spectrometers.
4. Click the OK button to accept the selected
options.
Serial Port Mode
1. Select the S2000/PC2000/USB2000/HR2000
option from the Spectrometer Type dropdown menu.
2. Select the Serial (RS-232) A/D option from the
A/D Converter Type drop-down menu. This
selection enables serial-specific options in the
lower portion of the Configure Hardware
screen.
3. Select the COM port that the HR2000 is
connected to in the SAD500 Serial Port dropdown menu. Consult Chapter 4: Troubleshooting for information on identifying
serial ports.
4. Select the speed at which the HR2000 will
operate from the SAD500 Baud Rate dropdown menu (115,200 baud is recommended).
5. Specify the pixel resolution (from 1 to 500) in
the SAD500 Pixel Resolution box. This value
specifies that every nth pixel of the
spectrometer will transmit from the HR2000 to
the PC.
Note: You can sacrifice pixel resolution to
gain speed. The transfer of one
complete spectra requires ~0.3
seconds at 115,200 baud.
6. Enable the Compress SAD500 Data
function to minimize the amount of data
transferred over the RS-232 connection.
The transmission of spectral data over the
serial port is a relatively slow process.
Enabling this function ensures that the
HR2000 compresses every scan that it
transmits. This greatly increases the data
transfer speed of the HR2000.
7. Click the OK button to complete setup.
The spectrometer should now be able to acquire data and respond to light. Exit and restart OOIBase32 to save
configuration data to disk.
The Spectrometer Configuration screen prompts you to configure specific channel-level spectrometer information,
if necessary.
Select Spectrometer | Configure | Wavelength Calibration tab from the menu and set system parameters. If
you have connected your spectrometer to the PC’s USB port, OOIBase32 pre-fills the coefficients for the HR2000
from information on a memory chip in the spectrometer. Otherwise, you must manually type the coefficients as
they are printed on the Wavelength Calibration Data Sheet that accompanied your spectrometer. Verify that the
calibration coefficients match the coefficients from the Wavelength Calibration Data Sheet. If necessary, modify
these values using the USB Programmer utility. Additionally, ensure that you select both the Master and Channel Enabled boxes.
In the A/D Interface tab, enter the same values as in the Configure Hardware dialog box. OOIBase32 stores this
information for future use once you close the program.
Note: For information on using the HR2000 with the OOIPS2000 Operating Software for the handheld PC, visit
To find operating instructions for HR2000-compatible products (such as light sources, sampling chambers, and
probes), consult the Software and Technical Resources CD or the Ocean Optics web site at:
The following sections contain information on troubleshooting issues you may encounter when using the HR2000
Spectrometer.
Note: For issues encountered when using a handheld PC, consult the OOIPS2000 manual.
HR2000 Connected to PC Prior to OOIBase32 Installation
If you connect your Ocean Optics USB device to the computer prior to installing your Ocean Optics software
application, you may encounter installation issues that you must correct before your Ocean Optics device will
operate properly.
Follow the applicable steps in this document to remove the incorrectly installed device, device driver, and
installation files.
Note: If these procedures do not correct your device driver problem, you will need to obtain the “Correcting
Device Driver Issues” document from the Ocean Optics website. Visit the following web address:
4. Click the OK button to continue. A warning box appears confirming the removal of the Unknown Device.
Click the OK button to confirm the device removal.
5. Disconnect the HR2000 from your computer.
6. Locate the section in this chapter that is appropriate to your operating system and perform the steps in
the “Remove Improperly Installed Files” section.
Windows 98:
Remove Improperly Installed Files:
1. Open Windows Explorer.
2. Navigate to the Windows | INF directory. If the INF directory is not visible, you will need to disable the
“Hide System Files and Folders” option on in Windows Folder Options.
Note: If the INF directory is not visible, you will need to disable the “Hide System Files and Folders” and
“Hide File Extensions for Known File Types” options in Windows Folder Options.
You can access Windows Folder Options from Windows Explorer, under the View | Options
menu selection.
3. Delete the OOI_USB.INF file in the INF directory.
4. Navigate to the Windows | System32 | Drivers directory.
5. Delete the EZUSB.SYS file.
6. Reinstall your Ocean Optics application and reboot the system when prompted.
7. Plug in the USB device.
The system will now be able to locate and install the correct drivers for the USB device.
2. Navigate to the Windows | INF directory. If the INF directory is not visible, you will need to disable the
“Hide System Files and Folders” option on in Windows Folder Options.
Note: If the INF directory is not visible, you will need to disable the “Hide System Files and Folders” and
“Hide File Extensions for Known File Types” options in Windows Folder Options.
You can access Windows Folder Options from Windows Explorer, under the Tools | Folder Options menu selection.
3. Delete the OOI_USB.INF and OOI_USB.PNF files in the INF directory.
4. Navigate to the Windows | System32 | Drivers directory.
5. Delete the EZUSB.SYS file.
6. Reinstall your Ocean Optics application and reboot the system when prompted.
7. Plug in the USB device.
The system will now be able to locate and install the correct drivers for the USB device.
Occasionally, you may encounter problems with the serial port connection and/or software. Perform the following
steps to troubleshoot the serial port connection:
1. Cycle the power on the HR2000 and restart the OOIBase32 software. This ensures that the software and
the HR2000 are synchronized.
2. Determine the serial port (COM port) number:
Operating System Instructions
1. Right-click on My Computer
Windows 95/98/ME
Windows 2000/XP
Windows NT
2. Select Properties
3. Click on the Device Manager tab
1. Select Start | Settings | Control Panel | System
3. Double-click on the Ports (COM & LPT) option to display COM port numbers. Ensure that no warning icon
appears next to the HR2000’s COM port.
4. Verify that the COM port to which the HR2000 is interfaced is active. If the ports on the PC are not
labeled and you do not know the COM port number, use trial-and-error to find the correct COM port.
Open OOIBase32 and view the displayed graph. If the correct COM port is selected, you will see a
dynamic trace responding to light near the bottom of the graph. If the correct COM port is not selected,
you will see a straight line at zero counts.
5. Disable virus protection to ensure timely and complete data transfer (optional – some computers require
this step).
Older Version of OOIBase32 Installed
If the PC you want to use to interface to your HR2000 already has an older version of OOIBase32, you need to
install the latest version of OOIBase32. You can download the latest version of OOIBase32 from the Software and Technical Resources CD or from the Ocean Optics web site at the following URL:
The following sections contain information on conducting sample experiments using the HR2000 and OOIBase32.
For information on experiments with OOIPS2000, consult the OOIPS2000 Operating Instructions.
Preparing for Experiments
Follow the steps below to configure the HR2000 and OOIBase32 for experiments:
1. Double-check that you have correctly installed the HR2000, installed OOIBase32, and configured the light
source and other sampling optics.
2. Open the OOIBase32 application, select Spectrometer | Configure from the menu bar, and doublecheck that A/D Interface settings are correct.
3. Check your spectrometer setup configurations in OOIBase32:
Locate the Wavelength Calibration Data sheet that came with the HR2000. Select Spectrometer | Configure from the menu and choose the Wavelength Calibration page. For each spectrometer
channel in the system, enable the channel and make sure the First Coefficient, Second Coefficient, Third
Coefficient and Intercept correspond to those of the system.
4. Adjust the acquisition parameters using the Acquisition Parameters dialog bar or select Spectrum | Configure Data Acquisition from the menu.
If you have followed the previous steps and started OOIBase32, the spectrometer is already acquiring
data. Even with no light in the spectrometer, there should be a dynamic trace displayed in the bottom of
the graph. If you allow light into the spectrometer, the graph trace should rise with increasing light
intensity. This means the software and hardware are correctly installed.
Once you install the hardware, configure the software, and establish your sampling system, you are ready to take
measurements.
There are four basic optical measurements from which to choose:
• Absorbance
• Transmission
• Reflection
• Relative irradiance
The type of measurement you will take determines the configuration of the sampling optics for your system.
Furthermore, your choice of reference and data analysis determines how the OOIBase32 presents the results.
Note: For each measurement, you must first take a reference and dark spectrum. After you take a reference
and a dark spectrum, you can take as many measurement scans as needed. However, if you change any
sampling variable (integration time, averaging, smoothing, angle, temperature, fiber size, etc.), you must
store a new dark and reference spectrum.
If the signal you collect is saturating the spectrometer (intensity greater than 4000 counts), you can decrease the
light level on scale in scope mode by:
• Decreasing the integration time
• Attenuating the light going into the spectrometer
• Using a smaller diameter fiber
• Using a neutral density filter with the correct optical density
If the signal you collect has too little light, you can increase the light level on scale in scope mode by:
Absorbance spectra are a measure of how much light a sample absorbs. For most samples, absorbance is
linearly related to the concentration of the substance. OOIBase32 calculates absorbance (A
equation…
S
- D
λ
Aλ = - log10
(
Rλ - D
λ
)
λ
) using the following
λ
…where S
intensity at wavelength λ.
Figure 6-1: Typical absorbance setup. The light source (far right) sends light via an input fiber into a cuvette in a cuvette
is the sample intensity at wavelength λ, Dλ is the dark intensity at wavelength λ, Rλ is the reference
λ
holder (bottom center). The light interacts with the sample. The output fiber carries light from the sample to the
spectrometer (top center), which is connected to the PC (far left).
Absorbance is also proportional to the concentration of the substance interacting with the light (this is known as
Beer’s Law). Common absorption applications include the quantification of chemical concentrations in aqueous or
gaseous samples.
Follow the steps below to take an absorbance measurement using OOIBase32:
1. Place OOIBase32 in scope mode by clicking the scope mode icon on the toolbar or selecting Spectrum | Scope Mode from the menu bar.
2. Ensure that the entire signal is on scale. The intensity of the reference signal should peak at about 3500
counts. If necessary, adjust the integration time until the intensity is approximately 3500 counts.
3. Place a sample of the solvent into a cuvette and take a reference spectrum. You must take a reference
spectrum before measuring absorbance.
Note: Do not put the sample itself in the path when taking a reference spectrum, only the solvent.
Click the Store Reference spectrum icon on the toolbar or select Spectrum | Store Reference from the
menu bar to store the reference. This command merely stores a reference spectrum in memory. You
must select File | Save | Reference from the menu bar to permanently save the spectrum to disk.
4. Block the light path to the spectrometer. Then, take a dark spectrum by clicking the Store Dark Spectrum icon on the toolbar or by selecting Spectrum | Store Dark from the menu bar. This command
merely stores a dark spectrum in memory. You must select File | Save | Dark from the menu to
permanently save the spectrum to disk.
Note: If possible, do not turn off the light source when taking a dark spectrum. If you must turn off your
light source to store a dark spectrum, allow enough time for the lamp to warm up again before
continuing your experiment. After the lamp warms up again, store a new reference (Step 3).
You must take a dark spectrum before measuring absorbance.
5. Put the sample in place and ensure that the light path is clear. Then, take an absorbance measurement
by clicking on the Absorbance Mode icon on the toolbar or selecting Spectrum | Absorbance Mode
from the menu. To permanently save the spectrum to disk, click the Save icon on the toolbar or select
File | Save | Processed from the menu bar.
Note: If you change any sampling variable (integration time, averaging, smoothing, angle, temperature,
fiber size, etc.), you must store a new dark and reference spectrum.
Transmission is the percentage of energy passing through a sample relative to the amount that passes through
the reference. Transmission Mode also displays the portion of light reflected from a sample, since transmission
and reflection measurements use the same mathematical calculations. The transmission is expressed as a
percentage (%T
following equation…
) relative to a standard substance (such as air). OOIBase32 calculates %Tλ (or %Rλ) with the
λ
S
- D
λ
λ
%Tλ =
R
λ
- D
x 100%
λ
…where S
is the sample intensity at wavelength λ, Dλ is the dark intensity at wavelength λ, Rλ is the reference
λ
intensity at wavelength λ.
Figure 6-2: Typical transmission setup. The light source (far right) sends light via the input leg of a transmission probe
into a container (bottom center). The light interacts with the sample. The output leg of the transmission probe carries the
information to the spectrometer (top center), which transmits the information to the PC (far left).
Common transmission applications include measuring light through solutions, optical filters, optical coatings, and
other optical elements (such as lenses and fibers).
Perform the following steps to take a transmission measurement using OOIBase32:
1. Place OOIBase32 in scope mode by clicking the Scope Mode icon on the toolbar or by selecting
Spectrum | Scope Mode from the menu bar.
2. Ensure that the entire signal is on scale. The intensity of the reference signal should peak at about 3500
counts. If necessary, adjust the integration time until the intensity is approximately 3500 counts.
6. Place a sample of the solvent into a cuvette and take a reference spectrum. You must take a reference
spectrum before measuring transmission.
Note: Do not put the sample itself in the path when taking a reference spectrum, only the solvent.
Click the Store Reference spectrum icon on the toolbar or select Spectrum | Store Reference from the
menu bar to store the reference. This command merely stores a reference spectrum in memory. You
must select File | Save | Reference from the menu bar to permanently save the spectrum to disk.
3. Block the light path to the spectrometer. Then, take a dark spectrum by clicking the Store Dark Spectrum icon on the toolbar or by selecting Spectrum | Store Dark from the menu bar. This command
merely stores a dark spectrum in memory. You must select File | Save | Dark from the menu to
permanently save the spectrum to disk.
Note: If possible, do not turn off the light source when taking a dark spectrum. If you must turn off your
light source to store a dark spectrum, allow enough time for the lamp to warm up again before
continuing your experiment.
You must take a dark spectrum before measuring transmission.
4. Put the sample in place and verify that the light path is clear. Then, take a transmission measurement by
clicking the Transmission Mode icon on the toolbar or selecting Spectrum | Transmission Mode from
the menu bar. To save the spectrum to disk, click the Save icon on the toolbar or select File | Save |
Processed from the menu bar.
Note: If you change any sampling variable (integration time, averaging, smoothing, angle, temperature,
fiber size, etc.), you must store a new dark and reference spectrum.
Reflection is the return of radiation by a surface, without a change in wavelength. Reflection can be:
••Specular (the angle of incidence is equal to the angle of reflection)
Diffuse (the angle of incidence is not equal to the angle of reflection)
Every surface returns both specular and diffuse reflections. Some surfaces may return mostly specular reflection,
while others may return mostly diffuse reflection. Specular reflection increases proportionately with the amount of
gloss on a surface.
Reflection is expressed as a percentage (%R
…where S
is the sample intensity at wavelength λ, Dλ is the dark intensity at wavelength λ, Rλ is the reference
λ
intensity at wavelength λ.
) relative to the reflection from a standard reference substance…
λ
S
- D
λ
λ
%Rλ =
R
λ
- D
x 100%
λ
Figure 6-3: Typical reflection setup. A light source (far right) sends light via the input leg of a reflection probe onto a
sample (bottom center). A reflection probe holder holds the probe in either a 90 or 45-degree angle from the surface. The
output leg of the reflection probe carries light from the sample to the spectrometer (top center), which is connected to the
PC (far left).
Common reflection applications include measuring the properties of mirrors and coatings. Other applications
include measuring the visual properties of the color in paints, plastics, and food products.
Perform the following steps to take reflection measurements using OOIBase32:
1. Place OOIBase32 is in scope mode by clicking the Scope Mode icon on the toolbar, or by selecting
Spectrum | Scope Mode from the menu bar.
2. Ensure that the entire signal is on scale. The intensity of the reference signal should peak at about 3500
counts.
3. Take a reference spectrum with the WS-1 Diffuse Reflectance Standard or the STAN-SSH Highreflectivity Reference Standard. You must take a reference spectrum before measuring reflection.
Click the Store Reference spectrum icon on the toolbar or select Spectrum | Store Reference from the
menu bar to store the reference. This command merely stores a reference spectrum in memory. You
must select File | Save | Reference from the menu bar to permanently save the spectrum to disk.
4. Block the light path to the spectrometer. Then, take a dark spectrum by clicking the Store Dark Spectrum icon on the toolbar or by selecting Spectrum | Store Dark from the menu bar. This command
merely stores a dark spectrum in memory. You must select File | Save | Dark from the menu to
permanently save the spectrum to disk.
Note: If possible, do not turn off the light source when taking a dark spectrum. If you must turn off your
light source to store a dark spectrum, allow enough time for the lamp to warm up again before
continuing your experiment.
You must take a dark spectrum before measuring transmission.
5. Put the sample is in place and ensure that the light path is clear. Then, take a reflection measurement by
clicking on the Transmission Mode icon on the toolbar or selecting Spectrum | Transmission Mode
from the menu bar (since the mathematical calculations used to calculate transmission and reflection are
identical). To save the spectrum to disk, click the Save icon on the toolbar or select File | Save |
Processed from the menu bar.
Note: If you change any sampling variable (integration time, averaging, smoothing, angle, temperature,
fiber size, etc.), you must store a new dark and reference spectrum.
Irradiance is the amount of energy at each wavelength emitted from a radiant sample. In relative terms, it is a
comparison of the fraction of energy the sample emits and the energy the sampling system collects from a lamp
with a blackbody energy distribution (normalized to 1 at the energy maximum). OOIBase32 calculates relative
irradiance with the following equation…
S
- D
λ
Iλ = B
λ
(
Rλ - D
λ
)
λ
…where B
the sample intensity at wavelength λ, D
wavelength λ.
Figure 6-4: Typical Relative Irradiance Setup. A light source with a known color temperature (such as the LS-1 or
LS-1-LL - lower right) is used to take a reference spectrum. The light to measure (lower left) is accumulated through a
CC-3 Cosine Corrector (or FOIS integrating sphere) into an input fiber, which carries the light information to the
spectrometer (upper right). The spectrometer then transmits the information to the PC (upper left), which compares the
measured spectra against the reference spectrum, thus removing wavelength-dependent instrument response from the
is the relative energy of the reference (calculated from the color temperature) at wavelength λ, Sλ is
λ
is the dark intensity at wavelength λ, Rλ is the reference intensity at
λ
measurement.
Common applications include characterizing the light output of LEDs, incandescent lamps, and other radiant
energy sources such as sunlight. Relative irradiance measurements also include fluorescence measurements,
which measure the energy given off by materials that have been excited by light at shorter wavelengths.
Perform the following steps to take a relative irradiance measurement using OOIBase32:
1. Place OOIBase32 is in scope mode by clicking the Scope Mode icon on the toolbar, or by selecting
Spectrum | Scope Mode from the menu bar.
2. Ensure that the entire signal is on scale. The intensity of the reference signal should peak at about 3500
counts.
Note: The light source must be a blackbody of known color temperature.
3. In the Reference Color Temperature dialog box, enter the color temperature of the light source (in
Kelvin) and click the OK button.
4. Take a reference spectrum using a light source with a black body of a known color temperature, such as
the LS-1.
Click the Store Reference spectrum icon on the toolbar or select Spectrum | Store Reference from the
menu bar to store the reference. This command merely stores a reference spectrum in memory. You
must select File | Save | Reference from the menu bar to permanently save the spectrum to disk.
5. Block the light path to the spectrometer. Then, take a dark spectrum by clicking the Store Dark Spectrum icon on the toolbar or by selecting Spectrum | Store Dark from the menu bar. This command
merely stores a dark spectrum in memory. You must select File | Save | Dark from the menu to
permanently save the spectrum to disk.
Note: If possible, do not turn off the light source when taking a dark spectrum. If you must turn off your
light source to store a dark spectrum, allow enough time for the lamp to warm up again before
continuing your experiment.
You must take a dark spectrum before measuring relative irradiance.
6. Position the fiber at the light source you wish to measure. Then, choose the Irradiance mode icon on the
toolbar or select Spectrum | Relative Irradiance Mode from the menu bar.
7. Click the Save icon on the toolbar or select File | Save | Processed from the menu bar to save the
spectrum to disk.
Note: If you change any sampling variable (integration time, averaging, smoothing, angle, temperature,
fiber size, etc.), you must store a new dark and reference spectrum.
OOIBase32 allows you to perform time acquisition experiments. Time acquisition experiments track processes,
perform kinetic analyses, and monitor spectral events all as a function of time. You can collect, as a function of
time, spectral data from up to six single wavelengths (designated as Channels A through F) and up to two
mathematical combinations of these wavelengths (designated as Combinations 1 and 2). Additionally, you can
acquire data in any mode (transmission, absorbance, etc.).
For more details about this and other OOIBase32 functions, refer to the
Software Online Help System
Follow the steps below to perform a time series experiment in OOIBase32:
1. Enter scope mode and store a reference spectra and dark spectra.
2. Choose the measurement mode (absorbance, transmission, etc.) and select Time Acquisition | Configure | Configure Time Channels from the menu bar to access the Time Acquisition Channel
Configuration screen.
Proceed to the Configuring the Time Acquisition Channel Configuration Screen section below.
Configuring the Time Acquisition Channel Configuration Screen
1. Perform the following steps on the Time Acquisition Channel Configuration screen:
a. Select Enabled to set the time acquisition calculation for the wavelength. The time acquisition
process will not calculate data if you do not select this option for at least
b. Select Plotted to see a real-time graph of the acquired data in a spectral window.
c. Select a Spectrometer Channel for the time acquisition process
d. Specify the analysis wavelength in the Wavelength (nm) box.
e. Specify the number of pixels around the analysis wavelength to average in the Bandwidth
(pixels) box.
f. Select a multiplicative factor to apply to the data before plotting or storing. Then, select an
additive constant or offset to apply to the data. OOIBase32 applies the additive constant or offset
after applying the factor but before plotting or storing data.
The equation for the Factor and Offset functions is:
Results = (Factor * Data) + Offset
.
OOIBase32 Spectrometer Operating
2. Configure a time acquisition process for the second single wavelength (if desired). Select the Channel B
page and repeat Steps 1-3 for Channel B.
To configure a time acquisition process for the third, fourth, fifth, and sixth single wavelengths, select the
Channel C, Channel D, Channel E, and Channel F pages, respectively, and set the necessary
parameters.
3. Configure a time acquisition process for a combination of two time channels (if desired) by selecting
Combination 1.
Perform the steps below to configure a combination:
a. Select Enabled to set the time acquisition calculation for the wavelength.
b. Enable Plotted to see a real-time graph of the acquired data in a spectral window.
c. Specify Time Channel A through F for the First Channel.
d. Select the mathematical operation to produce the data for Combination 1.
e. Specify Time Channel A through F for the Second Channel.
f. Select a multiplicative factor to apply to the data before plotting or storing. Then, select an
additive constant or offset to apply to the data. OOIBase32 applies the additive constant or offset
after applying the factor but before plotting or storing data.
The equation for the Factor and Offset functions is:
Results = (Factor * Data) + Offset
4. Configure a time acquisition process for the Combination 2 page, if desired. This page is virtually identical
to the Combination 1 page, with the exception that you can choose Combination 1 for the first or second
channel in Combination 2.
5. Click the Apply button to apply the changes, and then click the OK button to close the Time Acquisition
Channel Configuration screen.
Proceed to the Configuring the Configure Acquisition Screen section below.
Configuring the Time Acquisition Configuration Screen
1. Select Time Acquisition | Configure | Configure Acquisition from the menu bar to open the Time
Acquisition Configuration screen.
2. Enable Stream Data to Disk to save time acquisition data.
3. Enter a value in the Write Data to Disk Every X Acquisitions box to set the frequency for data saves.
OOIBase32 saves data more frequently if the number is smaller, or less frequently if the number is larger.
Entering a large number enhances the performance of the time acquisition process.
Note: At specified time intervals, OOIBase32 stores data into time acquisition channels or combination
channels. OOIBase32 can plot the data in a spectral window, or stream the data to disk, or both.
OOIBase32 can display up to 2048 acquisitions in a spectral window. If OOIBase32 collects more
than 2048 acquisitions, it only displays the last 2048. To store more than 2048 acquisitions, you
must stream the data to disk.
Writing data to the disk is a slow process (relative to the speed of some spectral acquisitions) and
causes a decrease in system performance. However, writing data to disk more frequently gives a
larger margin of safety.
4. Enable Show Values in Status Bar to see the time acquisition values in the status bar. These values
replace the cursor values.
5. Name the Stream Filename for the time acquisition process. Clicking on the ellipsis to the right of this
box opens a file save dialog box, allowing you to navigate to a designated folder.
6. Enable Save Every Acquisition to store data for every spectral acquisition during a time acquisition
process (optional).
Note: OOIBase32 has options to either store data for each acquisition, or to collect data only after a
specified delay. Several factors affect the minimum time acquisition frequency, including
integration time, number of spectrometer channels, samples averaged, and computer speed. If
you instruct OOIBase32 to store data every 100 milliseconds, the delay between data
acquisitions will be 100 milliseconds or more, depending on your experimental configuration.
OOIBase32 spends a large amount of time calculating, rendering, and displaying the spectra in a
spectral window. You can suspend the graph display, which greatly improves the performance of
OOIBase32.
7. Enter an Initial Delay to set the delay preceding the time acquisition process. Keep in mind that the delay
countdown does not begin until you start the time acquisition process. Be sure to select Hours, Minutes,
Seconds, or Milliseconds immediately to the right of the initial delay entry.
8. Enter a value to set the Frequency of the data collected in a time acquisition process. OOIBase32
stamps data from a time acquisition with a time accurate to one millisecond. Be sure to select Hours,
Minutes, Seconds, or Milliseconds immediately to the right of the frequency entry. You can enable the
Save Every Acquisition box to store the acquisitions that occur at this frequency. See Step 6 for more
information.
9. Enter a value to set the Duration for the entire time acquisition process. Be sure to select Hours, Minutes,
Seconds, or Milliseconds to the right of the duration entry. Click the OK button to close the Time
Acquisition Configuration dialog box. Then, enable Continue Until Manually Stopped, which instructs
OOIBase32 to store data until you manually stop the acquisition process (optional).
This Appendix describes how to calibrate the wavelength of your spectrometer. Although each spectrometer is
calibrated before it leaves Ocean Optics, the wavelength for all spectrometers will drift slightly as a function of
time and environmental conditions. Ocean Optics recommends periodically recalibrating the HR2000.
About Wavelength Calibration
You are going to be solving the following equation, which shows that the relationship between pixel number and
wavelength is a third-order polynomial…
λ
= I + C1 p + C2 p2 + C3 p
p
3
λ
...where
second coefficient (nm/pixel
the three Cs.
is the wavelength of pixel p, I is the wavelength of pixel 0, C1 is the first coefficient (nm/pixel), C2 is the
2
), and C3 is the third coefficient (nm/pixel3). You will be calculating the value for I and
Calibrating the Spectrometer
Preparing for Calibration
To re-calibrate the wavelength of your spectrometer, you will need the following:
•A light source capable of producing spectral lines
Note: Ocean Optics’ HG-1 Mercury-Argon lamp is ideal for recalibration. If you do not have an HG-1,
you will need a light source that produces several (at least 4-6) spectral lines in the wavelength
region of your spectrometer.
• An HR2000 spectrometer
• An optical fiber (for spectrometers without a built-in slit, a 50-
• A spreadsheet program (Excel or Quattro Pro, for example) or a calculator that performs third-order linear
regressions
Note: If you are using Microsoft Excel, choose Tools | Add-Ins and check AnalysisToolPak and
Perform the steps below to calibrate the wavelength of the spectrometer:
1. Place OOIBase32 into Scope Mode and take a spectrum of your light source. Adjust the integration time
(or the A/D conversion frequency) until there are several peaks on the screen that are not off-scale.
2. Move the cursor to one of the peaks and position the cursor so that it is at the point of maximum intensity.
3. Record the pixel number that is displayed in the status bar or legend (located beneath the graph). Repeat
this step for all of the peaks in your spectrum.
4. Use the spreadsheet program or calculator to create a table like the one shown below. In the first column,
place the exact or true wavelength of the spectral lines that you used.
In the second column of this worksheet, place the observed pixel number. In the third column, calculate
the pixel number squared, and in the fourth column, calculate the pixel number cubed.
5. Use the spreadsheet or calculator to calculate the wavelength calibration coefficients. In the spreadsheet
program, find the functions to perform linear regressions.
• If using Quattro Pro, look under Tools | Advanced Math
• If using Excel, look under Analysis ToolPak
6. Select the true wavelength as the dependent variable (Y). Select the pixel number, pixel number squared,
and the pixel number cubed as the independent variables (X). After executing the regression, you will
obtain an output similar to the one shown below.
Regression Statistics
Multiple R 0.999999831
R Square 0.999999663
Adjusted R Square 0.999999607
Standard Error 0.125540214
Observations 22
Intercept
Coefficients Standard Error
Intercept 190.473993 0.369047536 First coefficient
X Variable 1 0.36263983 0.001684745
X Variable 2 -1.174416E-05 8.35279E-07
X Variable 3 -2.523787E-09 2.656608E-10 S
Third coefficient
econd coefficient
The figure above notes the numbers of importance.
7. Record the Intercept, as well as the First, Second, and Third Coefficients. Additionally, look at the value
for R squared. It should be very close to 1. If not, you have most likely assigned one of your wavelengths
incorrectly.
Ocean Optics programs wavelength calibration coefficients unique to each HR2000 onto an EEPROM memory
chip in the HR2000.
You can overwrite old calibration coefficients on the EEPROM if you are using the HR2000 via the USB port. If
you are using the HR2000 via the serial port, consult the Saving the New Calibration Coefficients: Serial Mode
section later in this Appendix.
To Save Wavelength Calibration Coefficients Using the USB Mode
1. Ensure that the HR2000 is connected to the PC and that you have closed all other applications.
2. Point your browser to
http://www.oceanoptics.com/technical/softwaredownloads.asp and scroll down to
Microcode. Click on the USB EEPROM Programmer selection.
3. Save the setup file to your computer.
4. Run the Setup.exe file to install the software. The Welcome screen appears.
5. Click the Next button. The Destination Location screen appears.
6. Accept the default installation location, or click the Browse button to specify a directory. Then, click the
Next button. The Program Manager Group screen appears.
7. Click the Next button. The Start Installation screen appears.
8. Click the Next button to begin the installation. Once the installation finishes, the Installation Complete
screen appears.
9. Click the Finish button and reboot the computer when prompted.
10. Navigate to the USB EEPROM Programmer from the Start menu and run the software.
11. Click on the appropriate HR2000 device displayed in the left pane of the USB Programmer screen.
12. Double-click on each of the calibration coefficients displayed in the right pane of the USB Programmer
screen and enter the new values acquired in Steps 5 and 6 of the Calibrating the Wavelength of the Spectrometer section in this Appendix.
13. Repeat Step 12 for all of the new values.
14. Click on the Save All Values button to save the information, and then Exit the USB Programmer
software.
The new wavelength calibration coefficients are now loaded onto the EEPROM memory chip on the HR2000.
Saving the New Calibration Coefficients: Serial Mode
If you are connecting the HR2000 Spectrometer to the serial port of the PC, you need to save the new wavelength
calibration coefficients to the .SPEC file that OOIBase32 accesses when opened.
Note: You cannot save the calibration coefficients to the EEPROM memory chip on the HR2000 when using the
serial mode.
To Save Wavelength Calibration Coefficients Using the Serial Mode
1. Open the OOIBase32 application.
2. Select Spectrometer | Configure from the OOIBase32 menu bar. The Configure Spectrometer screen
appears.
3. Select the Wavelength Calibration tab to update the wavelength coefficients within OOIBase32.
4. Enter in the new values acquired from Steps 5 and 6 of the Calibrating the Wavelength of the Spectrometer section in this Appendix.
5. Click the OK button to save the information in OOIBase32.
The HR2000CG-UV-NIR Composite Grating Spectrometer has a new proprietary grating and order-sorting filter to
provide a 200-1100 nm wavelength range with 1.0 nm optical resolution in one spectrometer.
The HR2000-CG-UV-NIR is functionally similar to the standard HR2000 Spectrometer. Follow the instructions in
Chapter 3 to configure the HR2000CG-UV-NIR.
HR2000CG-UV-NIR Features
The HR2000CG-UV-NIR contains the following features that differ from the standard HR2000:
New HC-1 Landis Composite Grating
The HR2000CG-UV-NIR uses the new HC-1 Landis grating designed to provide a 200-1100 nm wavelength
range. The HC-1 is fixed in place at the time of manufacture.
Variable Order-Sorting Filter
The HR2000CG-UV-NIR contains a new OFLV-200-1000 variable order sorting filter to eliminate second and third
order effects.
HR2000CG-UV-NIR Specifications
Dimensions: 148.6 mm x 104.8 mm x 45.1 mm
Weight: 570 g
Power consumption: 95 mA @ 5 VDC
Detector: 2048-element linear silicon CCD array
Wavelength range: 200-1100 nm
Gratings: HC-1, 300 lines per nm grating
Entrance aperture:
Order-sorting filters: OFLV-200-100 installed
Focal length: f/4, 101 mm
Optical resolution: 1.0 nm FWHM
Stray light: <0.05% at 600 nm; <0.10% at 435 nm
Fiber optic connector: SMA 905 to single-strand optical fiber (0.22 NA)
Data transfer rate:
Integration time: 3 milliseconds to 65 seconds
Operating systems:
5 µm slit
Full scans into memory every 13 milliseconds with the USB port, even 300
milliseconds with the serial port
Windows 98/Me/2000/XP when using the USB interface on desktop or
notebook PCs
Any 32-bit Windows operating system when using the serial port on
desktop or notebook PCs