Ocean Optics HR2000+ User Manual

Page 1
HR2000+ High-speed Fiber Optic Spectrometer
Installation and Operation Manual
Document Number 294-00000-000-02-201012
Offices: Ocean Optics, Inc. World Headquarters
830 Douglas Ave., Dunedin, FL, USA 34698
Phone 727.733.2447 Fax 727.733.3962
E-mail: Info@OceanOptics.com (General sales inquiries)
Orders@OceanOptics.com (Questions about orders) TechSupport@OceanOptics.com (Technical support)
000-00000-000-02-A
Page 2
Additional Offices:
Regional Headquarters
Maybachstrasse 11 73760 Ostfildern
Phone 49-711 34 16 96-0 Fax 49-711 34 16 96-85 E-Mail Sales@Mikropack.de
Ocean Optics Asia
666 Gubei Road, Kirin Tower, Suite 601B, Changning District, Shanghai, PRC. 200336
Phone 86.21.5206.8686
Fax 86.21.5206.8686 E-Mail Sun.Ling@OceanOptics.com
Ocean Optics EMEA
Sales and Support Center Geograaf 24, 6921 EW DUIVEN, The Netherlands
Phone 31-(0)26-3190500
Fax 31-(0)26-3190505 E-Mail
Info@OceanOptics.eu
Copyright © 2010 Ocean Optics, Inc.
All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted, by any means, electronic, mechanical, photocopying, recording, or otherwise, without written permission from Ocean Optics, Inc. This manual is sold as part of an order and subject to the condition that it shall not, by way of trade or otherwise, be lent, re-sold, hired out or otherwise circulated without the prior consent of Ocean Optics, Inc. in any form of binding or cover other than that in which it is published.
Trademarks
All products and services herein are the trademarks, service marks, registered trademarks or registered service marks of their respective owners.
Limit of Liability
Every effort has been made to make this manual as complete and as accurate as possible, but no warranty or fitness is implied. The information provided is on an “as is” basis. Ocean Optics, Inc. shall have neither liability nor responsibility to any person or entity with respect to any loss or damages arising from the information contained in this manual.
Page 3
Table of Contents
About This Manual......................................................................................................... iii
Document Purpose and Intended Audience..............................................................................iii
What’s New in this Document ................................................................................................... iii
Document Summary..................................................................................................................iii
Product-Related Documentation.................................................................................... iii
Upgrades....................................................................................................................... iv
Chapter 1: Introduction .....................................................................1
Product Overview .......................................................................................................... 1
System Requirements ................................................................................................... 2
EEPROM Utilization .................................................................................................................. 2
About SpectraSuite.................................................................................................................... 2
Sampling System Overview....................................................................................................... 2
How Sampling Works............................................................................................................2
Modular Sampling Accessories............................................................................................. 3
Interface Options ........................................................................................................... 3
Breakout Box ............................................................................................................................. 3
Shipment Components .................................................................................................. 4
Chapter 2: Installing the HR2000+....................................................5
Overview ....................................................................................................................... 5
HR2000+ Installation ..................................................................................................... 5
Configuring the HR2000+ with OOIBase32 .............................................................................. 6
Operator and Serial Number Dialog Box .............................................................................. 6
Default Spectrometer Configuration File ............................................................................... 6
Configure Hardware Screen .................................................................................................6
Spectrometer Configuration Screen......................................................................................7
Connect Spectroscopic Accessories.............................................................................. 8
External Triggering Options........................................................................................... 8
294-00000-000-02-201012 i
Page 4
Table of Contents
Chapter 3: Troubleshooting..............................................................9
Overview ....................................................................................................................... 9
HR2000+ Connected to PC Prior to Software Installation .............................................. 9
Windows Operating Systems .................................................................................................... 9
Remove the Unknown Device from Windows Device Manager ...........................................10
Remove Improperly Installed Files........................................................................................10
Mac Operating Systems ............................................................................................................11
Linux Operating Systems .......................................................................................................... 11
Older Version of OOIBase32 Installed........................................................................... 11
Appendix A: Calibrating the Wavelength of the HR2000+ ..............13
Overview ....................................................................................................................... 13
About Wavelength Calibration ....................................................................................... 13
Calibrating the Spectrometer ......................................................................................... 14
Preparing for Calibration............................................................................................................ 14
Calibrating the Wavelength of the Spectrometer ......................................................................14
Saving the New Calibration Coefficients: USB Mode ..................................................... 16
Saving the New Calibration Coefficients: Serial Mode ................................................... 17
Appendix B: Specifications...............................................................19
Overview ....................................................................................................................... 19
How the HR2000+ Works .............................................................................................. 19
HR2000+ Components Table .................................................................................................... 20
HR2000+ Specifications ................................................................................................ 21
CCD Detector Specifications ..................................................................................................... 21
HR2000+ Spectrometer............................................................................................................. 21
System Compatibility ..................................................................................................... 22
Compatibility for Desktop or Notebook PCs .............................................................................. 22
30-Pin Accessory Connector Pinout .............................................................................. 23
30-Pin Accessory Connector Pinout Diagram ........................................................................... 23
30-Pin Accessory Connector – Pin Definitions and Descriptions.............................................. 23
30-Pin J2 Accessory Connector - Part Numbers.......................................................................25
HR2000+ 15-Pin Accessory Cable Pinout ..................................................................... 26
Index ................................................................................................... 27
ii 294-00000-000-02-201012
Page 5
About This Manual
Document Purpose and Intended Audience
This document provides the user of HR2000+ Spectrometer with instructions for setting up, calibrating and performing experiments with their spectrometer.
What’s New in this Document
This version of the HR2000+ High-speed Fiber Optic Spectrometer, Installation and Operation Manual updates information for the new triggering options for firmware versions 3.0 and above.
Document Summary
Chapter Description
Chapter 1: Introduction
Chapter 2: Installing the HR2000+
Chapter 3: Troubleshooting
Appendix A: Calibrating the
Wavelength of the HR2000+
Appendix B: Specifications
Contains descriptive information about the HR2000+ Spectrometer and how sampling works. It also provides a list of system requirements, interface options, and shipment components.
Provides installation and configuration instructions.
Contains recommended steps to isolate and correct common problems.
Provides instructions for calibrating the HR2000+ Spectrometers.
Contains technical specifications and connector pinouts for the HR2000+ Spectrometers.
Product-Related Documentation
You can access documentation for Ocean Optics products by visiting our website at
http://www.oceanoptics.com. Select Technical → Operating Instructions, then choose the appropriate
document from the available drop-down lists. Or, use the of the web page.
Search by Model Number field at the bottom
Detailed instructions for SpectraSuite Spectrometer Operating Software are located at:
http://www.oceanoptics.com/technical/SpectraSuite.pdf.
Detailed instructions for the OOIBase32 Spectrometer Operating Software are located at:
http://www.oceanoptics.com/technical/ooibase32.pdf
294-00000-000-02-201012 iii
Page 6
About This Manual
Detailed instructions for the Breakout Box are located at:
http://www.oceanoptics.com/technical/HR4_breakout.pdf.
Detailed instructions for External Triggering for firmware versions below 3.0 are located at:
http://www.oceanoptics.com/technical/external-triggering.pdf.
Detailed instructions for External Triggering for firmware versions 3.0 and above are located at:
http://www.oceanoptics.com/technical/external-triggering2.pdf.
Engineering-level documentation is located on our website at
You can also access operating instructions for Ocean Optics products from the Software and Technical
Resources
CD that ships with the product.
Technical → Engineering Docs.
Upgrades
Occasionally, you may find that you need Ocean Optics to make a change or an upgrade to your system. To facilitate these changes, you must first contact Customer Support and obtain a Return Merchandise Authorization (RMA) number. Please contact Ocean Optics for specific instructions when returning a product.
iv 294-00000-000-02-201012
Page 7
Chapter 1
Introduction
Product Overview
The HR2000+ High-Speed Miniature Fiber Optic Spectrometer provides optical resolution as good as
0.035 nm (FWHM). The HR2000+ is responsive from 200-1100 nm, but the specific range and resolution depends on your grating and entrance slit selections. With its capability of transferring 1ms spectra continuously, the HR2000+ is the fastest spectrometer available from Ocean Optics.
The HR2000+ is perfect for applications where fast reactions need to be monitored and high resolution is necessary, such as chemistry and biochemistry applications.
Data programmed into a memory chip on each HR2000+ includes wavelength calibration coefficients, linearity coefficients, and the serial number unique to each spectrometer. Our spectrometer operating software simply reads these values from the spectrometer — a feature that enables hot swapping of spectrometers among PCs.
The HR2000+ Spectrometer connects to a notebook or desktop PC via USB port or serial port. When connected to the USB port of a PC, the HR2000+ draws power from the host PC, eliminating the need for an external power supply.
Ocean Optics HR2000+ High-speed Fiber Optic Spectrometer
294-00000-000-02-201012 1
Page 8
1: Introduction
System Requirements
You can use the HR2000+’s USB connectivity with any PC that meets the following requirements:
Windows 98/Me/2000/XP operating system (or Windows CE 2.11 or later for handheld PCs)
Ocean Optics SpectraSuite or OOIBase32 software application installed and configured for use
with the HR2000+. See
instructions.
Alternately, the HR2000+ has serial port adaptability for connecting to PCs, PLCs, and other devices that support the RS-232 communication protocol. However, this connection method requires an external power supply to power the HR2000+, the Breakout Box (HR4-BREAKOUT), and a serial cable.
EEPROM Utilization
An EEPROM memory chip in each HR2000+ contains wavelength calibration coefficients, linearity coefficients, and a serial number unique to each individual spectrometer. The software application reads these values directly from the spectrometer, enabling the ability to “hot-swap” spectrometers between PCs without entering the spectrometer coefficients manually on each PC.
Chapter 2: Installing the HR2000+ for specific configuration
About SpectraSuite
SpectraSuite is the latest generation of operating software for all Ocean Optics spectrometers. It is a completely modular, Java-based spectroscopy software platform that operates on Windows, Macintosh and Linux operating systems. The software can control any Ocean Optics USB spectrometer and device, as well as any other manufacturer’s USB instrumentation using the appropriate drivers.
SpectraSuite is a user-customizable, advanced acquisition and display program that provides a real-time interface to a variety of signal-processing functions. With SpectraSuite, you have the ability to perform spectroscopic measurements (such as absorbance, reflectance, and emission), control all system parameters, collect and display data in real time, and perform reference monitoring and time acquisition experiments. Consult the SpectraSuite manual for hardware requirements when using SpectraSuite (see
Product-Related Documentation).
Sampling System Overview
How Sampling Works
Ocean Optics components function in a sampling system as follows:
1. The user stores reference and dark measurements to correct for instrument response variables.
2. The light transmits through an optical fiber to the sample.
3. The light interacts with the sample.
4. Another optical fiber collects and transmits the result of the interaction to the spectrometer.
2 294-00000-000-02-201012
Page 9
1: Introduction
5. The spectrometer measures the amount of light and transforms the data collected by the spectrometer into digital information.
6. The spectrometer passes the sample information to SpectraSuite.
7. SpectraSuite compares the sample to the reference measurement and displays processed spectral information.
Modular Sampling Accessories
Ocean Optics offers a complete line of spectroscopic accessories for use with the HR2000+. Most of our spectroscopic accessories have SMA connectors for application flexibility. Accordingly, changing the sampling system components is as easy as unscrewing a connector and replacing an accessory.
Interface Options
The HR2000+ has both USB and serial port connectors (with the use of an adapter), enabling you to connect the spectrometer to a desktop or notebook PC via a USB port.
Computer Interface
Desktop or Notebook PC via USB Port
Desktop or Notebook PC via Serial Port
Operating System Requirements
Windows 98/Me/ 2000/XP
Any 32-bit Windows operating system
Part Needed Description of Part
USB-CBL-1 (included)
HR4­BREAKOUT (not included)
Cable that connects from USB port on HR2000+ to USB port on desktop or notebook PC
Adapter block that enables connection from serial port on HR2000+ to serial port on desktop or notebook PC; comes with 5 VDC power supply (required when connecting to serial port). User must supply own software.
Breakout Box
Ocean Optics also offers the Breakout Box (HR4-BREAKOUT), a passive module that separates the signals from their 30-pin port to an array of standard connectors and headers, enabling easy access to a variety of features found in Ocean Optics’ HR2000+ Spectrometer. In addition to the accessory connector, the breakout box features a circuit board based on a neutral breadboard pattern that allows custom circuitry to be prototyped on the board itself.
294-00000-000-02-201012 3
Page 10
1: Introduction
Shipment Components
The following information and documentation ships with the HR2000+ Spectrometer:
Packing List
The packing list is inside a plastic bag attached to the outside of the shipment box (the invoice arrives separately). It lists all items in the order, including customized components in the spectrometer (such as the grating, detector collection lens, and slit). The packing list also includes the shipping and billing addresses, as well as any items on back order.
Spectrometer Installation Instructions
A sheet of paper that contains the information that you need to get your spectrometer system up and running. Further information can be found on the Ocean Optics
Library
Wavelength Calibration Data Sheet
CD (see below).
Each spectrometer is shipped with a Wavelength Calibration Data Sheet that contains information unique to your spectrometer. SpectraSuite reads this calibration data from your spectrometer when it interfaces to a PC via the USB port. Any other interface requires that you manually enter the calibration data in OOIBase32 (select
Spectrometer | Configure | Wavelength Calibration
tab). See the OOIBase32 documentation for more information (refer to Product-Related
Documentation for instructions on accessing OOIBase32 documentation).
Software and Resources
Note
Please save the Wavelength Calibration Data Sheet for future reference.
Software and Technical Resources CD
Each order ships with the Ocean Optics Software and Technical Resources CD. This disc contains software, operating instructions, and product information for all Ocean Optics software, spectrometers, and spectroscopic accessories. You must have Adobe Acrobat Reader version 6.0 or higher to view these files. Ocean Optics includes the Adobe Acrobat Reader on the
and Technical Resources CD
.
Ocean Optics software requires a password during the installation process. You can locate passwords for the other software applications on the back of the
Software and Resources Library
CD package.
Software
4 294-00000-000-02-201012
Page 11
Chapter 2
Installing the HR2000+
Overview
You must install the software application prior to connecting the HR2000+ Spectrometer to the PC. The software installation installs the drivers required for HR2000+ installation. If you do not install the software first, the system will not properly recognize the HR2000+.
If you have already connected the HR2000+ to the PC prior to installing the software, consult
Troubleshooting
for information on correcting a corrupt HR2000+ installation.
HR2000+ Installation
To connect the HR2000+ to a PC via the USB port, the PC must be running the Windows 98/ME/2000/XP operating system.
Note
The USB port on a PC can power up to five HR2000+ spectrometer channels. Systems with more than five channels require a powered USB hub.
Procedure
Follow the steps below to connect the HR2000+ to a PC via the USB port:
1. Install SpectraSuite on the destination computer.
2. Locate the USB cable (USB-CBL-1) provided with the HR2000+.
3. Insert the square end of the cable into the side of the HR2000+.
Chapter 3:
4. Insert the rectangular end of the cable into the USB port of the PC.
If you installed SpectraSuite prior to connecting the HR2000+, SpectraSuite installs the HR2000+ drivers. If the drivers do not successfully install (or if you connected the HR2000+ to the PC before installing SpectraSuite), consult
If you have followed the previous steps and started SpectraSuite, the spectrometer is already acquiring data. Even with no light in the spectrometer, there should be a dynamic trace displayed in the bottom of the graph. If you allow light into the spectrometer, the graph trace should rise with increasing light intensity. This means the software and hardware are correctly installed.
Note the spectrometer(s) that you have installed are listed in the Data Sources pane.
294-00000-000-02-201012 5
Chapter 3: Troubleshooting.
Page 12
2: Installing the HR2000+
Once you install the SpectraSuite software and the hardware, and establish your sampling system, you are ready to take measurements.
Configuring the HR2000+ with OOIBase32
Once you install the HR2000+ with OOIBase32 software, you must configure OOIBase32’s Configure Spectrometer
OOIBase32 Spectrometer Operating Software Operating Instructions for detailed instructions on configuring the spectrometer in OOIBase32 (see
The following sections contain instructions on initially configuring the HR2000+ the first time you start OOIBase32. Additional features are available for this spectrometer. See the
Operating Software Operating Instructions
Operator and Serial Number Dialog Box
The Operator and Serial Number screen prompts you to enter a user name and software serial number into OOIBase32. Some data files created by OOIBase32 during sampling procedures use this information in the file headers.
options so that OOIBase32 recognizes the HR2000+ Spectrometer. Consult the
Product-Related Documentation).
OOIBase32 Spectrometer
for detailed information on these HR2000+ features.
Default Spectrometer Configuration File
The Default Spectrometer Configuration File screen prompts you to select a spectrometer configuration (.SPEC) file for use with the HR2000+. The unique serial number of the HR2000+ precedes the file extension (for example, HR2A0162.SPEC).
Navigate to the OOIBase32 installation directory and select the spectrometer configuration file.
Configure Hardware Screen
The Configure Hardware screen prompts you to enter spectrometer-specific information into OOIBase32 the first time you run the program. Typically, you need only enter this information the first time you run OOIBase32. However, you can alter the hardware configuration at any time using the Spectrometer Configuration screen. Select Spectrometer | Configure from the OOIBase32 menu bar to access the
Spectrometer Configuration screen.
6 294-00000-000-02-201012
Page 13
Procedure
2: Installing the HR2000+
To configure hardware in USB Mode:
1. Specify S2000 in the
2. Specify HR2000+ in the
3. Specify the serial number of the HR2000+ under the
Spectrometer Type drop-down menu.
A/D Converter Type drop-down menu.
USB Serial Number drop-down menu.
Note
The system pre-fills this drop-down menu with the serial numbers of all discovered HR2000+ Spectrometers.
4. Click the OK button to accept the selected options.
5. The spectrometer should now be able to acquire data and respond to light. Exit and restart OOIBase32 to save configuration data to disk.
Spectrometer Configuration Screen
The Spectrometer Configuration screen prompts you to configure specific channel-level spectrometer information, if necessary.
Procedure
1. Select Spectrometer | Configure from the menu and set system parameters.
2. Select the
Wavelength Calibration tab. OOIBase32 pre-fills the coefficients for the HR2000+
from information on a memory chip in the spectrometer.
294-00000-000-02-201012 7
Page 14
2: Installing the HR2000+
3. Verify that the calibration coefficients match the coefficients from the Wavelength Calibration Data Sheet that accompanied the spectrometer. If necessary, modify these values using the USB Programmer utility.
4. Additionally, ensure that you select both the
5. In the
A/D Interface tab, enter the same values as in the Configure Hardware screen.
OOIBase32 stores this information for future use once you close the program.
Master and Channel Enabled boxes.
Connect Spectroscopic Accessories
To find operating instructions for HR2000+-compatible products (such as light sources, sampling chambers, and probes), consult the
http://www.oceanoptics.com/technical/operatinginstructions.asp.
Software and Technical Resources CD or the Ocean Optics website at
External Triggering Options
You can trigger the HR2000+ using a variety of External Triggering options through the 30-pin Accessory Connector on the spectrometer. For HR2000+ firmware versions below 3.0, see the External Triggering Options document located at
triggering.pdf
document located at ???. The triggering document contains instructions for configuring External Triggering options for the HR2000+.
. For HR2000+ firmware version 3.0 and above, see the External Triggering Options
http://www.oceanoptics.com/technical/external-
8 294-00000-000-02-201012
Page 15
Chapter 3
Troubleshooting
Overview
The following sections contain information on troubleshooting issues you may encounter when using the HR2000+ Spectrometer.
HR2000+ Connected to PC Prior to Software Installation
Windows Operating Systems
If you connected your Ocean Optics USB device to the computer prior to installing your Ocean Optics software application, you may encounter installation issues that you must correct before your Ocean Optics device will operate properly.
Follow the applicable steps below to remove the incorrectly installed device, device driver, and installation files.
Note
If these procedures do not correct your device driver problem, you must obtain the Correcting Device Driver Issues document from the Ocean Optics website:
http://www.oceanoptics.com/technical/engineering/correctingdevicedriverissues.pdf.
294-00000-000-02-201012 9
Page 16
3: Troubleshooting
Remove the Unknown Device from Windows Device Manager
Procedure
1. Open Windows Device Manager. Consult the Windows operating instructions for your computer for directions, if needed.
2. Locate the sign to the immediate left.
Other Devices option and expand the Other Devices selection by clicking on the "+"
Note
Improperly installed USB devices can also appear under the Universal Serial Bus Controller option. Be sure to check this location if you cannot locate the unknown device.
3. Locate the unknown device (marked with a large question mark). Right-click on the Unknown
Device
4. Click the Device. Click the
5. Disconnect the HR2000+ from your computer.
6. Locate the section in this chapter that is appropriate to your operating system and perform the steps in the following
listing and select the Uninstall or Remove option.
OK button to continue. A warning box appears confirming the removal of the Unknown
OK button to confirm the device removal.
Remove Improperly Installed Files section.
Remove Improperly Installed Files
Procedure
1. Open Windows Explorer.
2. Navigate to the
Windows | INF directory.
Note
If the INF directory is not visible, you must disable the Hide System Files and Folders and Hide File Extensions for Known File Types options in Windows Folder Options. Access Windows Folder Options from Windows Explorer, under the
Options
3. Delete the OOI_USB.INF in the INF directory. If your computer is running either the Windows 2000 or XP operating system, you must also delete the
4. Navigate to the
5. Delete the
6. Reinstall your Ocean Optics application and reboot the system when prompted.
10 294-00000-000-02-201012
menu selection.
OOI_USB.PNF file in the INF directory.
Windows | System32 | Drivers directory.
EZUSB.SYS file.
Tools | Folder
Page 17
3: Troubleshooting
7. Plug in the USB device.
The system is now able to locate and install the correct drivers for the USB device.
Mac Operating Systems
Since there are no device files for the USB2000-FLG Spectrometer in a Mac operating system, you should not encounter any problems if you installed the spectrometer before the SpectraSuite software.
Linux Operating Systems
For Linux operating systems, all you need to do is install the SpectraSuite software, then unplug and replug in the spectrometer. Technically, the driver files for Linux simply give nonprivileged users permission to use newly connected hardware. There isn’t any long-term harm to plugging in the device before installing the software.
Older Version of OOIBase32 Installed
If the PC to be used to interface to your HR2000+ already has an older version of OOIBase32 software installed, you must install the latest version of OOIBase32. You can download the latest version of OOIBase32 from the Ocean Optics website at
http://www.oceanoptics.com/technical/softwaredownloads.asp.
You do not need to uninstall previous versions of OOIBase32 when upgrading to the latest version.
294-00000-000-02-201012 11
Page 18
3: Troubleshooting
12 294-00000-000-02-201012
Page 19
Appendix A
Calibrating the Wavelength of
the HR2000+
Overview
This appendix describes how to calibrate the wavelength of your spectrometer. Though each spectrometer is calibrated before it leaves Ocean Optics, the wavelength for all spectrometers will drift slightly as a function of time and environmental conditions. Ocean Optics recommends periodically recalibrating the HR2000+.
About Wavelength Calibration
You are going to be solving the following equation, which shows that the relationship between pixel number and wavelength is a third-order polynomial:
= I + C1 p + C2 p2 + C3 p
p
Where:
= the wavelength of pixel p
= the wavelength of pixel 0
I
C1 = the first coefficient (nm/pixel)
C2 = the second coefficient (nm/pixel2)
C
= the third coefficient (nm/pixel3)
3
R = the reference intensity at wavelength
You will be calculating the value for I and the three Cs.
3
294-00000-000-02-201012 13
Page 20
A: Calibrating the Wavelength of the HR2000+
Calibrating the Spectrometer
Preparing for Calibration
To recalibrate the wavelength of your spectrometer, you need the following components:
A light source capable of producing spectral lines
Note
Ocean Optics’ HG-1 Mercury-Argon lamp is ideal for recalibration. If you do not have an HG-1, you need a light source that produces several (at least 4-6) spectral lines in the wavelength region of your spectrometer.
An HR2000+ spectrometer
An optical fiber (for spectrometers without a built-in slit, a 50-m fiber works best)
A spreadsheet program (Excel or Quattro Pro, for example) or a calculator that performs third-
order linear regressions
Note
If you are using Microsoft Excel, choose Tools | Add-Ins and check AnalysisToolPak and AnalysisTookPak-VBA.
Calibrating the Wavelength of the Spectrometer
Procedure
Perform the steps below to calibrate the wavelength of the spectrometer:
1. Place the software into Scope mode and take a spectrum of your light source. Adjust the integration time (or the A/D conversion frequency) until there are several peaks on the screen that are not off-scale.
2. Move the cursor to one of the peaks and position the cursor so that it is at the point of maximum intensity.
3. Record the pixel number that is displayed in the status bar or legend (located beneath the graph). Repeat this step for all of the peaks in your spectrum.
4. Use the spreadsheet program or calculator to create a table like the one shown in the following figure. In the first column, place the exact or true wavelength of the spectral lines that you used.
In the second column of this worksheet, place the observed pixel number. In the third column, calculate the pixel number squared, and in the fourth column, calculate the pixel number cubed.
14 294-00000-000-02-201012
Page 21
A: Calibrating the Wavelength of the HR2000+
Independent
Variable
True Wavelength (nm) Pixel # Pixel #
253.65
296.73
302.15
313.16
334.15
365.02
404.66
407.78
435.84
546.07
576.96
579.07
696.54
706.72
727.29
738.40
751.47
175 296 312 342 402 490 604 613
694 1022 1116 1122 1491 1523 1590 1627 1669
Dependent
Variables
30625 87616
97344 116964 161604 240100 364816 375769 481636
1044484 1245456 1258884 2223081 2319529 2528100 2647129 2785561
2
1067462648 1389928896 1412467848 3314613771 3532642667 4019679000 4306878883 4649101309
Pixel #
5359375 25934336 30371328 40001688 64964808
117649000 220348864 230346397 334255384
3
Values Computed
from the Regression
Output
Predicted
Wavelength
253.56
296.72
302.40
313.02
334.19
365.05
404.67
407.78
435.65
546.13
577.05
579.01
696.70
706.62
727.24
738.53
751.27
Difference
0.09
0.01
-0.25
0.13
-0.05
-0.04
-0.01
0.00
0.19
-0.06
-0.09
0.06
-0.15
0.10
0.06
-0.13
0.19
5. Use the spreadsheet or calculator to calculate the wavelength calibration coefficients. In the spreadsheet program, find the functions to perform linear regressions.
If using Quattro Pro, look under Tools | Advanced Math
If using Excel, look under Analysis ToolPak
6. Select the true wavelength as the dependent variable (Y). Select the pixel number, pixel number squared, and the pixel number cubed as the independent variables (X). After executing the regression, you will obtain an output similar to the one shown below. Numbers of importance are noted.
Regression Statistics
Multiple R 0.999999831 R Square 0.999999663 R Squared Adjusted R Square 0.999999607 Standard Error 0.125540214 Observations 22
Intercept
Coefficients Standard Error
Intercept 190.473993 0.369047536 First coefficient X Variable 1 0.36263983 0.001684745 X Variable 2-1.174416E-05 8.35279E-07 X Variable 3-2.523787E-09 2.656608E-10 Second coefficient
Third coefficient
294-00000-000-02-201012 15
Page 22
A: Calibrating the Wavelength of the HR2000+
7. Record the Intercept, as well as the First, Second, and Third Coefficients. Additionally, look at the value for R squared. It should be very close to 1. If not, you have most likely assigned one of your wavelengths incorrectly.
Keep these values at hand.
Saving the New Calibration Coefficients: USB Mode
Ocean Optics programs wavelength calibration coefficients unique to each HR2000+ onto an EEPROM memory chip in the HR2000+.
You can overwrite old calibration coefficients on the EEPROM if you are using the HR2000+ via the USB port. If you are using the HR2000+ via the serial port, consult the
Coefficients: Serial Mode section later in this appendix.
Procedure
To save wavelength calibration coefficients using the USB mode, perform the following steps:
1. Ensure that the HR2000+ is connected to the PC and that you have closed all other applications.
Saving the New Calibration
2. Point your browser to scroll down to
3. Save the setup file to your computer.
4. Run the
5. Click the
6. Accept the default installation location, or click the click the
7. Click the
8. Click the
Complete
9. Click the
10. Navigate to the
11. Click on the desired HR2000+ device displayed in the left pane of the
12. Double-click on each of the calibration coefficients displayed in the right pane of the USB Programmer screen and enter the new values acquired in Steps 5 and 6 of the
Wavelength of the Spectrometer section in this appendix.
13. Repeat Step 12 for all of the new values.
14. Click on the software.
Microcode. Select USB EEPROM Programmer.
Setup.exe file to install the software. The Welcome screen appears.
Next button. The Destination Location screen appears.
Next button. The Program Manager Group screen appears.
Next button. The Start Installation screen appears.
Next button to begin the installation. Once the installation finishes, the Installation
screen appears.
Finish button and reboot the computer when prompted.
Save All Values button to save the information, and then Exit the USB Programmer
http://www.oceanoptics.com/technical/softwaredownloads.asp and
Browse button to specify a directory. Then,
USB EEPROM Programmer from the Start menu and run the software.
USB Programmer screen.
Calibrating the
The new wavelength calibration coefficients are now loaded onto the EEPROM memory chip on the HR2000+.
16 294-00000-000-02-201012
Page 23
A: Calibrating the Wavelength of the HR2000+
Saving the New Calibration Coefficients: Serial Mode
If you are connecting the HR2000+ Spectrometer to the serial port of the PC, you need to save the new wavelength calibration coefficients to the
You cannot save the calibration coefficients to the EEPROM memory chip on the HR2000+ when using the serial mode.
Procedure
To save Wavelength Calibration Coefficients using the Serial mode, perform the following steps:
1. Open the OOIBase32 application.
.SPEC file that OOIBase32 accesses when opened.
Note
2. Select
3. Select the
4. Enter in the new values acquired from Steps 5 and 6 of the
5. Click the
Spectrometer | Configure from the OOIBase32 menu bar. The Configure Spectrometer
screen appears.
Wavelength Calibration tab to update the wavelength coefficients within OOIBase32.
Spectrometer section in this appendix.
OK button to save the information in OOIBase32.
Calibrating the Wavelength of the
294-00000-000-02-201012 17
Page 24
A: Calibrating the Wavelength of the HR2000+
18 294-00000-000-02-201012
Page 25
Appendix B
Specifications
Overview
This appendix contains information on spectrometer operation, specifications, and system compatibility. It also includes accessory connector pinout diagrams and pin-specific information.
How the HR2000+ Works
Below is a diagram of how light moves through the optical bench of an HR2000+ Spectrometer. The optical bench has no moving parts that can wear or break; all the components are fixed in place at the time of manufacture. Items with an asterisk (*) are user-specified.
HR2000+ Spectrometer with Components
See HR2000+ Components Table on the following page for an explanation of the function of each numbered component in the HR2000+ Spectrometer in this diagram.
294-00000-000-02-201012 19
Page 26
B: HR2000+ Specifications
HR2000+ Components Table
Ocean Optics permanently secures all components in the HR2000+ at the time of manufacture. Only Ocean Optics technicians can replace interchangeable components, where noted.
Item Name Description
1
2 Slit
3 Filter
4
5 Grating
SMA Connector
Collimating Mirror
Secures the input fiber to the spectrometer. Light from the input fiber enters the optical bench through this connector.
A dark piece of material containing a rectangular aperture, which is mounted directly behind the SMA Connector. The size of the aperture regulates the amount of light that enters the optical bench and controls spectral resolution.
You can also use the HR2000+ without a Slit. In this configuration, the diameter of the fiber connected to the HR2000+ determines the size of the entrance aperture.
Only Ocean Optics technicians can change the Slit.
Restricts optical radiation to pre-determined wavelength regions. Light passes through the Filter before entering the optical bench. Both bandpass and longpass filters are available to restrict radiation to certain wavelength regions.
Only Ocean Optics technicians can change the Filter.
Focuses light entering the optical bench towards the Grating of the spectrometer.
Light enters the spectrometer, passes through the SMA Connector, Slit, and Filter, and then reflects off the Collimating Mirror onto the Grating.
Diffracts light from the Collimating Mirror and directs the diffracted light onto the Focusing Mirror. Gratings are available in different groove densities, allowing you to specify wavelength coverage and resolution in the spectrometer.
Only Ocean Optics technicians can change the Grating.
6
7
8
20 294-00000-000-02-201012
Focusing Mirror
L2 Detector Collection Lens
CCD Detector (UV or VIS)
Receives light reflected from the Grating and focuses the light onto the CCD Detector or L2 Detector Collection Lens (depending on the spectrometer configuration).
An optional component that attaches to the CCD Detector. It focuses light from a tall slit onto the shorter CCD Detector elements.
The L2 Detector Collection Lens should be used with large diameter slits or in applications with low light levels. It also improves efficiency by reducing the effects of stray light.
Only Ocean Optics technicians can add or remove the L2 Detection Collection Lens.
Collects the light received from the Focusing Mirror or L2 Detector Collection Lens and converts the optical signal to a digital signal.
Each pixel on the CCD Detector responds to the wavelength of light that strikes it, creating a digital response. The spectrometer then transmits the digital signal to the OOIBase32 application.
Page 27
B: HR2000+ Specifications
HR2000+ Specifications
The following sections provide specification information for the CCD detector in the HR2000+, as well as the HR2000+ Spectrometer itself. HR2000+CG-UV-NIR specifications are listed in
Error! Reference source not found..
CCD Detector Specifications
Specification Value
Detector Sony ILX-511B linear silicon CCD array
No. of elements 2048 pixels
Appendix C:
Sensitivity
Pixel size 14 µm x 200 µm
Pixel well depth 65 Ke-
Signal-to-noise ratio 250:1 (at full signal)
A/D resolution 14 bit
Dark noise 3.2 RMS counts
Corrected linearity >99.8%
Maximum pixel rate Rate at which pixels are digitized is 1 MHz
75 photons per count at 400 nm 41 photons per count at 600 nm
HR2000+ Spectrometer
Specification Value
Dimensions 148.6 mm x 104.8 mm x 45.1 mm
Weight 570 g
Power consumption 220 mA @ 5 VDC
Detector 2048-element linear silicon CCD array
Detector range 200-1100 nm
Gratings 14 gratings available
294-00000-000-02-201012 21
Page 28
B: HR2000+ Specifications
Specification Value
Entrance aperture 5, 10, 25, 50, 100 or 200 µm wide slits
Order-sorting filters Installed longpass and bandpass filters
Focal length f/4, 101 mm
Optical resolution Depends on grating and size of entrance aperture
Stray light <0.05% at 600 nm; <0.10% at 435 nm
Dynamic range 2 x 108 (system); 1300:1 for a single acquisition
Fiber optic connector SMA 905 to single-strand optical fiber (0.22 NA)
Data transfer rate:
USB 2.0 Port
Serial Port
Integration time 1 ms to 65 seconds
Interfaces
Operating systems:
USB Port
Serial Port
Onboard GPIO 10 user-programmable digital I/Os
Analog channels One 13-bit analog input and one 9-bit analog output
Full scans into memory every 1 millisecond
Full scans into memory every 600 milliseconds
USB 2.0, 480 Mbps (USB 1.1 compatible); RS-232 (2-wire); SPI (3-wire); I
Windows 2000/XP, Mac OS X, and Linux
Any 32-bit Windows operating system
2
C Inter-Integrated Circuit 2-wire serial bus
System Compatibility
The following sections provide information on hardware and software requirements for the HR2000+.
Compatibility for Desktop or Notebook PCs
To use the HR2000+, you must have a PC that meets the following minimum requirements:
Operating systems: Windows 2000/XP, Mac OS X, or Linux with USB port; any 32-bit or 64-bit
Windows OS with serial port
Computer interfaces: USB 2.0 @ 480 Mbps; RS-232 (2-wire) @ 115.2 K baud
Peripheral interfaces: I2C inter-integrated circuit
22 294-00000-000-02-201012
Page 29
B: HR2000+ Specifications
30-Pin Accessory Connector Pinout
The HR2000+ features a 30-pin Accessory Connector, located on the side of the unit as shown:
Location of HR2000+ 30-Pin Accessory Connector
30-Pin Accessory Connector Pinout Diagram
When facing the 30-pin Accessory Connector on the front of the vertical wall of the HR2000+, pin numbering is as follows:
USB
Port
2 4 6 8 10 12 14 16 18 20 22 24 26 28 30
1 3 5 7 9 11 13 15 17 19 21 23 25 27 29
30-Pin Accessory Connector Pinout Diagram
30-Pin Accessory Connector – Pin Definitions and Descriptions
The following table contains information regarding the function of each pin in the HR2000+’s 30-Pin Accessory Connector:
Pin
Function Input/Output Description
#
1 RS232 Rx Input RS232 receive signal – Communicates with a PC over DB9 Pin 3
2 RS232 Tx Output RS232 transmit signal – Communicates with a PC over DB9 Pin 2
3 GPIO (2) Input/Output
General purpose software-programmable, digital input/output (channel number) Reserved
4 V5_SW Output Regulated 5 Volt power pin – Supplies 50 mA (maximum)
294-00000-000-02-201012 23
Page 30
B: HR2000+ Specifications
Pin
Function Input/Output Description
#
5 Ground Input/Output Ground
6 I2C SCL Input/Output I2C clock signal for communication to other I2C peripherals
7 GPIO (0) Input/Output
General purpose software-programmable, digital input/output (channel number) Reserved
8 I2C SDA Input/Output I2C data signal for communication to other I2C peripherals
9 GPIO (1) Input/Output
10
Ext. Trigger In
Input
11 GPIO (3) Input/Output
V
, V
12
CC
or 5V
IN
USB
,
Input or Output
General purpose software-programmable, digital input/output (channel number) Reserved
TTL input trigger signal – See External Triggering Options document for information.
General purpose software-programmable, digital input/output (channel number) Reserved
Input power pin for HR2000+ – When operating via USB, this pin can power other peripherals – Ensure that peripherals comply with USB specifications
13 Reserved Output Reserved
Input power pin for HR2000+ – When operating via USB, this pin can power other peripherals – Ensure that peripherals comply with USB specifications
14
V
, V
CC
or 5V
IN
USB
,
Input or Output
15 Reserved Input Reserved
16 GPIO (4) Input /Output
17
Single Strobe
Output
18 GPIO (5) Input/Output
General purpose software-programmable, digital input/output (channel number) Reserved
TTL output pulse used as a strobe signal – Has a programmable delay relative to the beginning of the spectrometer integration period
General purpose software-programmable, digital input/output (channel number) Reserved
19 Reserved Output Reserved
Continuous
20
Strobe
SPI Chip
21
Select
24 294-00000-000-02-201012
Output
Output
TTL output signal used to pulse a strobe – Divided down from the master clock signal
SPI Chip/Device Select signal for communication to other SPI peripherals
Page 31
Pin
Function Input/Output Description
#
B: HR2000+ Specifications
22 GPIO (6) Input/Output
Analog In
23
(0-5V)
Analog Out
24
(0-5V)
Lamp
25
Enable
26 GPIO (7) Input/Output
27 Ground Input/Output Ground
28 GPIO (8) Input/Output
29 Ground Input/Output Ground
30 GPIO (9) Input/Output
Input 13-bit analog-to-digital input with a 0-5V range. For future use.
Output 9-bit programmable output voltage with a 0-5V range
Output
General purpose software-programmable, digital input/output (channel number) Reserved
TTL signal driven Active HIGH when the Lamp Enable command is sent to the spectrometer
General purpose software-programmable, digital input/output (channel number) Reserved
General purpose software-programmable, digital input/output (channel number) Reserved
General purpose software-programmable, digital input/output (channel number) Reserved
30-Pin J2 Accessory Connector - Part Numbers
The part numbers for the 30-pin accessory connector on the HR2000+ Spectrometer are as follows:
The connector is Pak50™ model from 3M Corp. Headed Connector – Part Number
P50–030P1–RR1–TG.
The mating connector is Part Number P50–030S–TGF.
Mating the two components requires two 1.27 mm (50 mil) flat ribbon cables (3M 3365 Series is
recommended).
If you are customizing your HR2000+ Spectrometer system or configuring External Triggering, you may need these part numbers to complete your setup.
294-00000-000-02-201012 25
Page 32
B: HR2000+ Specifications
HR2000+ 15-Pin Accessory Cable Pinout
Pin # Description Pin # Description
1 Single_strobe 9 GPIO-9
2 ContStrobe 10 GND_SIGNAL
3 V5_SW 11 SDA
4 ExtTrigIn 12 SCL
5 ExtTrigIn 13 LampEnable
6 GPIO-8 14 A_IN
7 A_OUT 15 GPIO-7
8 ExtTrigIn
26 294-00000-000-02-201012
Page 33
Index
Numbers
15-pin accessory cable
pinouts, 26
30-pin accessory connector
diagram, 23 part numbers, 25 pin definitions, 23
A
accessories, 8 Accessories, 3 accessory connector
pinout, 23
Adobe Acrobat Reader, 4
B
breakout box, 3
C
Calibrating, iii, 13 calibration, 13
preparing for, 14 procedure, 14
calibration coefficients
saving in Serial mode, 17
saving in USB mode, 16 CCD, 21 CCD Detector, 20 collimating mirror, 20 compatibility, 22
Desktop or Notebook PCs, 22 Components Table, 20 Configure Hardware, 6 configuring, 6
D
Default Spectrometer Configuration File, 6 detector, 21 Detector Collection Lens, 20 document
audience, iii purpose, iii summary, iii
E
EEPROM, 2 External Triggering, 8
F
filter, 20 focusing mirror, 20
G
grating, 20
I
Installation, 5
USB mode, 5 installed filter, 20 Interface, 3
L
L2 Detector Collection Lens, 20 Lens, 20
294-00000-000-02-201012 27
Page 34
Index
M
memory chip, 2 mirror, 20
O
OOIBase32
configuring, 6
Options
Interface, 3
P
packing list, 4 passwords, 4 power supply (external), 2 product-related documentation, iii
S
Sampling
Accessories, 3
System, 2 serial number, 6 setup, 5 shipment components, 4 slit, 20 SMA Connector, 20 Software and Resources Library CD, 4 specifications, 19, 21
detector, 21
spectrometer, 21 SpectraSuite, 2 Spectrometer Configuration Screen, 7 spectrometer operation, 19 spectroscopic accessories, 8 System Requirements, 2
T
Triggering, 8 troubleshooting
Linux systems, 11 Mac systems, 11 OOIBase32, 11 Windows systems, 9
Troubleshooting, 9
U
upgrades, iv USB mode, 5 USB-ADP-PC, 3 USB-CBL-1, 3
W
Wavelength Calibration
about, 13 Wavelength Calibration Data File, 4 Wavelength Calibration Data Sheet, 4 what's new, iii
28 294-00000-000-02-201012
Loading...