All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted, by any means, electronic,
mechanical, photocopying, recording, or otherwise, without written permission from Ocean Optics, Inc.
This manual is sold as part of an order and subject to the condition that it shall not, by way of trade or otherwise, be lent, re-sold, hired out or
otherwise circulated with out the pr ior consent of Ocean Optics, Inc. in any form of binding or cover other th an t ha t in wh ich it is published.
Trademarks
Microsoft, Windows, Wind ow s 95, Windows 98, Windows Me, Wi ndows NT, Windows 2000, W ind ow s XP an d Excel are either registered
trademarks or trademarks of Microsoft Corporation.
Limit of Liability
Every effort h as been made to make this manua l as c omplete and as a ccurate as possible, but no warranty or fitn ess is implied. T he information
provided is on an “ as is” basis. Oce an Optics, Inc. s hall have neither liability nor responsibility to any person or entity with respect to any loss or
damages arising from the inform a t i on contained in this manual.
Important Safety Notices
1. Do not remove or modify any installed safety device on this equipment. Doing so will void your
warrant y and create an unsafe ope rating environment.
2. Dangerous voltages are present in t his device. There are NO user serviceabl e parts inside the
power supply.
3. Only allow qualified personnel to service thi s unit.
4. Inspect this unit and its power supply before u s ing it f or the fi rs t tim e. Do n ot use the unit if it is
damaged in any way. Contact your dealer for repair or replacement information.
000-10000-190-02-201308A
Important Safety Notices
B 000-10000-190-02-201308
Table of Contents
About This Manual .............................................................................................................iii
Document Purpose and Intended Audience .............................................................................. iii
Document Summary .................................................................................................................. iii
Product-Related Documentation ............................................................................................... iii
Calibration for Absolute Irradiance: Models HL-3 plus-CAL, HL-3 plus-CAL-EXT, HL-3-CAL
and HL-3-CAL-EXT ............................................................................................................9
Calibration for Radiant Flux ....................................................................................................... 9
Uncertainty Level Calculation for HL-3 plus Family ...........................................................10
Appendix B: HL-3 Use Log .................................................................. 13
HL-3 Series Use Log ..........................................................................................................13
Index ...................................................................................................... 15
000-10000-190-02-201308i
Table of Contents
ii 000-10000-190-02-201308
About This Manual
Document Purpos e and Intended Audience
This document provides you with an installation section to get your system up and running, basic
information ab out th e calibration if the light source, and a log to save the time of the light source’s use.
Document Summary
Chapter Description
Chapter 1: Setup
Chapter 2: HL-3 Specifications
Appendix A: Calibration Basics
Appendix B: HL-3 Use Log
Contains a list of package contents and unpacking
instructions. Also contains procedures for connecting to a
spectrometer.
Contains operating environment specifications, as well as
other physical details of the product.
Provides an overview of the physics for radiometrically
calibrated systems.
Provides a sample log for recording lamp use.
Product-Related Documentation
You can access documentation for Ocean Optics products by visiting our website at
http://www.oceanoptics.com. Select Technical → Operating Instructions, then choose the appropriate
document from the available drop-down lists. Or, use the Search by Model Number field at the bottom
of the web page.
• Detailed ins tructions for OceanView Spectrometer Operating Software is located at:
You can also access operating instructions for Ocean Optics products on the Software and Technical
Resources CD includ ed with the system.
Engineering-level documentation is located on our website at Technical → Engineering Docs.
.
000-10000-190-02-201308iii
About This Manual
Upgrades
Occasionally, you may find that you need Ocean Optics t o make a change or an upgrade to your system.
To facilitate these changes, you must first contact Customer Support and obtain a Return Merchandise
Authorization (RMA) number. Please contact an Ocean Optics Application Scientist for specific
instructions when returning a product.
iv 000-10000-190-02-201308
Chapter 1
Setup
Overview
The HL-3 Series of Halogen Calibrated Light Sources for the VIS-Shortwave NIR (350nm-1100nm) is a
tungsten-halogen light source that provides known absolute intensit y values at several wavelengths,
expressed in µW/cm²/nm. Since the spectral intensity of the HL-3 Series can be traced to an intensity
standard traceable to the N ational Inst itute of Standards and Te chnology (N IST), it is specifically
designed for calibrating the absolute spectral response of your system.
Note
If you have a spectr o met er se tup tha t is high ly sensitive, you may oversatura te it.
HL-3 plus-CAL Calibrated Light Source
000-10000-190-02-2013081
1: Setup
Models Available
HL-3 Series light source models are available in regular (350 – 1100 nm) and extended (EXT models)
(350 – 2400 nm) wavelength ranges. The INT models are used with an i ntegrating sphere (rat her than a
CC-3 cosine-corrected ir radiance probe). The following models are available in the HL-3 Series:
Model Calibrated Wavelength Range Minimum Uncertainty
Level
HL-3-CAL 350 – 1100 nm 10% No
HL-3-CAL-EXT 350 – 2400 nm 10% No
HL-3-INT-CAL 350 – 1100 nm 15% No
HL-3-INT-CAL-EXT 350 – 2400 nm 15% No
In addition the following plus ver sions are available which reduce the percentage of uncertainty to 3%
and offer a shutter :
Model Calibrated Wavelength Range Minimum Uncertainty
1. Unpack your lamp assembly and power supply carefully. Although the lamp is rigidly mounted,
dropping this instrument can cause permanent damage.
2. Inspect the outside of th e instrument and make sur e that there is no dama ge. Do not us e the
instrument if damage is present.
3. Use this instrument in a clean laboratory environment (see Operating Envi ronment
2 000-10000-190-02-201308
).
1: Setup
Contents
Your HL-3 Series Light Source package should contain the following:
HL-3 light source unit
1 AC adapter for stabilizing power
Power cord
Lamp Calibration Report for usin g the HL-3 light source with a CC-3 cosine-corrected irradiance
probe or integ ra tin g sph e re (INT mod el s)
1 CD that holds files for both lamp calibration reports
1 Allen wrench for adjusting the inner barrel of the SMA connector and CC-3 cosine corrector
Connecting a Spectrometer to the HL-3
Before using an HL-3 S eries light source for the first time, ch eck for tran sport damage. Be sure to adhere
to all warnings on the unit and in this operat ional manual. Make sure you have your HL-3 light source,
your spectro met er, a CC -3 cosine-corrected irradi ance probe o r i ntegrating sphere, and Ocean Optics
spectromet er op erating software.
The procedure will differ, depending on whether you are using a cosine corrector or an integrating sphere
(INT models).
Connecting a Spectrometer to the HL-3-CAL Models When
Using a Cosine Corrector
► Procedure
Use the following pro cedure to connect you r HL-3 light source to a spectro meter when using a co sine
corrector:
1. Loosen the set screw on the HL-3's SMA connector.
2. Screw the CC-3 cosine corrector onto the end of the fiber. The connection should be tight. Insert
the CC-3 all the w a y into the HL -3's SMA connector, taking the place of the SMA's inner barrel.
3. Tighten the setsc rew on the SMA conn e cto r of the H L-3 with an Allen wrench . Connect the other
end of the fiber into the SMA connector of the spectrometer.
4. Plug the switching AC adapte r into the bac k of the HL-3. The adapter stabilizes power coming
into the la mp to ensure constant spectral intensity.
5. Plug one end of the p ower cord into the AC adapter and the other end into a standard outlet.
6. Turn the lamp on. Allow the lamp to warm up for at least 15 minutes before using it. Pla ce th e
HL-3 horizontally.
000-10000-190-02-2013083
1: Setup
7. Insert the cal ibration disk that came with your light s ource into your computer. This disk contains
one ASCII file with the same in formation as the Lamp Calibration Reports that came with your
HL-3. The file name con tains the l amp's serial number, followed b y .LMP. Copy these files to
the desired directory on your computer.
8. Start your spect r om eter op e ra tin g software. S ee your spectrometer manual for instructions on
calibrating the spectrometer with the HL-3.
Connecting a Spectrometer to the HL-3-INT-CAL Models
When Using an I ntegrating Sphere
► Procedure
Use the following pro cedure to connect you r HL-3 light source to a spectrometer when using an
integratin g sph e re ( INT models):
1. The HL-3-CAL-INT and HL-3-CAL-INT-EXT h a ve a diffu s or for li ght ou tpu t. P l ace the HL-3-
CAL-INT output diffusor a t the aperture of your integrating sphere. The minimum aperture
diameter of your integrating sphere should be 6 mm. The HL-3-CAL-INT is validat ed f or any
larger Aperture diameters up to 30 mm. The HL-3-CAL-INT has to be placed in a vertical
orientation. Screw the CC-3 cosine corrector onto the end of the fiber. The connection should be
tight. Insert the CC-3 all the way into the HL-3's SMA connect or, taking the place of the SMA's
inner barr el.
2. Tighten the setsc rew on the SMA conn e cto r of the HL-3 with an Allen wrench. Connect the other
end of the fiber into the SM A connector o f the spectrometer.
3. Plug the switching AC adapter into the back of the H L-3. The adap ter stabilizes power coming
into the la mp to ensure constant spectral intensity.
4. Plug one end of the power cord into the AC adapter and the other end into a standard outlet.
5. Turn the lamp on. Allow the lamp to warm up for at least 15 minutes before using it.
6. Insert the cal ibration disk that came with your light source into your computer. This disk contains
one ASCII file with the same information as the Lamp Calibrat ion Reports that came with your
HL-3. The file n ame contains the lamp's serial number, followed by .LMP. C opy these files to
the desired directory on your computer.
7. Start your spectro met er op era tin g so ftw a re. Se e you r spec tr om ete r man ual f or in st ru cti ons on
calibrating the spectro meter with the HL-3.
4 000-10000-190-02-201308
1: Setup
Tips
When you perform measurement s with the calibrated spectrometer system, record a new
dark spectr um before e ach measurement. Th is minimizes the influence of a shifted dark
spectrum.
Use a log to monitor the op eration time of the light source. Recal ibration of the light
source after 50 hours of operati on is recommended. A sample log is provided in
Appendix B: HL-3 Use Log
.
000-10000-190-02-2013085
1: Setup
6 000-10000-190-02-201308
Chapter 2
HL-3 Specifications
This secti on provides i nformation on the operating environment and specifi cations of the HL-3 Ser ies
light sources.
Operating Environment
The following table provides informa tion on optim izing the operating environment of your HL-3.
Operating Environment The HL-3 Unit . . .
Moisture Is designed for operation in dry rooms only.
Ventilation Should be placed so that its location or pos i tio n does not interf er e with proper
ventilation.
Heat Should be placed away from any device that emits excessive heat.
Object and Liquid Entry Should be positioned so that objects do not fall on top of the unit. Additionally,
ensure that no liquids are spilled into the enclosure through openings.
Specifications
Specification Value
Spectral Range (calibrated):
Regular Models
EXT Models
Dimensions (LWH):
HL-3-CAL and HL-3-CAL-EXT
HL-3-INT-CAL and HL-3-INT-CAL-EXT
HL-3 plus-CAL and HL-3plus-CAL-EXT
HL-3 plus-INT-CAL and HL-3 plus-INT-CAL-EXT
350 – 1100 nm
350 – 2400 nm
13.5 cm x 5.8 cm x 5.9 cm; 5.3" x 2.3" x 2.3"
12.8 cm x 5.8 cm x 5.9 cm; 5.0" x 2.3" x 2.3"
14.5 cm x 6.3 cm x 6.1 cm; 5.7" x 2.5" x 2.4"
13.8 cm x 6.3 cm x 6.1 cm; 5.4" x 2.5" x 2.4"
Power Input 12 VDC/800 mA – 1100 mA (regulated)
Power Output 7 W
Bulb Life 10,000 hours (recalibrate after 50 hours of use)
Output to Bulb 5 V / 1.0 A
000-10000-190-02-2013087
2: HL-3 Specifications
Specification Value
Time to Stabilized Output 15 minutes
Connector:
Regular Models
INT Models
Shutter:
Regular Models
plus Models
CC-3 cosine corrector
Integrating Sphere
No
Yes
8000-10000-190-02-201308
Appendix A
Calibration Basics
Overview
Calibrated systems, independent of the particular calibrated un it, are always traceable to national
standards. A c ali br at ed i tem is c omp ar ab l e to all calibrated items which are calibrated f or the same
calibrated u nit.
Each measu rement result has an un certainty, and the level of uncertainty is given either for one single
standard deviation (k = 1) or for two standard de viations (k = 2), which is often named as expanded
uncertainty.
A calibrated device should give trusted results; therefore, the ca lib ra ti on h as to be done b y the metrologist
following a previously defined and proven calibration procedure. Clear documentation of the calibration
setup and of the calibrated devices used is an indispen sable require men t.
Calibration for Absolute Irradiance: Models HL-3
plus-CAL, HL-3 plus-CAL-EXT, HL-3-CAL and HL3-CAL-EXT
Absolute irradiance has the physical u nit µW/nm/c m^2. The interpretation is as follows: This is the
wavelength resolved electromagnetic radiation which is emitted through an area [cm^2].
Radiant flux has the physical unit µW/nm. It reflects the total amount of emitted electromagnetic
radiation from a light source and is typically r ealized by placing a light source inside an integrating
sphere. Th e HL-3 plus-INT models are used to calibrate the integrating sphere with an attached
spectrometer.
000-10000-190-02-2013089
A: Calibration Basics
Uncertainty Level Calculati on f or HL-3 plus
Family
Each realist ic me a sur e men t has an un cer ta in t y and ther ef o re, a calibration has an uncert aint y. The
uncertainty should be validated seriously since th is reflects how reliab le the calibration is. T he following
internation al st an d ards define the uncertainty and are describing how the uncertainty should be derived:
• IEC Guide 115 Application of uncertainty of measurement to conformity asses sment a ctiv it i es in
the electrot echn i c al sec t or.
• JCGM100:2008: GUM 1995 with minor corrections
Ocean Optics calibrations are done in relation to these international stand ards.
Ocean Optics provides a calibration of the HL-3 plus-CAL on an ab so lut e irr ad i a nce s ca le in
µW/nm/cm^2. The sources of uncertainties taken into account are listed below in the example. The actual
values are l isted in th e individual calibration certificate for your light source. The calibration certificate
contains m or e detailed information.
Sources of Uncertainties
Variable Source of uncertainty
1. R Uncertainty in reproducibility [R], caused by switching on/off
_1.1 S Stability[S] within 50 hours of operation
Influence of temperature [T] and humidity [F] in the air. The temperatures during
calibration have a temperature range of 19 °C to 25 °C. The humidity might
2. T and F
3. d
4. ROT Uncertainty from rotating [ROT] the Cosine Corrector.
5. D_Lambda
6. D_AbsIrrad
7. StrayL
8. T_TEC_
9. Dark
10. Rep Reproducibility of the used spectrometer system for calibration.
influence the calibration in the NIR for wavelength larger than 1100 nm.
Uncertainty in mounting distance [d] of the used Cosine Corrector. 1.8 % per 0.1
mm deviation from aperture plane.
Uncertainty caused by wavelength calibration of the spectrometer which was
used to calibrate the Light source. [D_Lamba[QE]=0,3 nm,
D_Lambda[NirQ512]= 1 nm]
Uncertainty of the calibration light source used. This is in most cases the Ocean
Optics working standard. See calibration certificate for details.
Influence from spectrometers internal stray light which was used to calibrate the
light source.
Influence of the sensor temperature [TEC] of the spectrometer which was used
to calibrate the light source.
Influence from uncertainties of the measured dark level of the spectrometer
used to calibrate the light source.
Linearity of the spectrometer system which was used to calibrate the light
11. SP_Lin
12. Bend Uncertainty caused by bending [Bend] of the fibers.
10000-10000-190-02-201308
source.
A: Calibration Basics
These factors are taken into account for deriving the wavelength for resolved levels of uncertainty (listed
below).
Standard Uncertainties and Uncertainty Budgets in % at Wavelength (E xa mple)