Oakton T-100WL Instruction Manual

- 1 -
T100WL Portable Turbidity Meter
35635-10
- 2 -
Table of Contents
1 Introduction ................................................................................................................................... - 3 -
2 Technical Specifications .............................................................................................................. - 4 -
3 Instrumentation Illustration ......................................................................................................... - 5 -
3.1 Summary ................................................................................................................................. - 5 -
3.2 Configuration ......................................................................................................................... - 6 -
3.3 Display Mode .......................................................................................................................... - 6 -
3.4 Keypad .................................................................................................................................... - 7 -
3.5 Power supply .......................................................................................................................... - 8 -
3.6 Data Save, Recall, and Deletion ........................................................................................... - 8 -
3.7 Setting for the first time use. .............................................................................................. - 10 -
4 Calibration.................................................................................................................................... - 10 -
4.1 Preparation for Calibration ................................................................................................. - 10 -
4.2 Calibration Procedure (Take 0 NTU and 20 NTU as an example) ................................... - 11 -
4.3 Notes for Calibration ........................................................................................................... - 12 -
5 Turbidity Measurement ............................................................................................................... - 13 -
5.1 Sample Vial Handling .......................................................................................................... - 13 -
5.2 Measurement Preparation................................................................................................... - 14 -
5.3 Measurement Mode ............................................................................................................. - 14 -
5.4 Notes for Measurement ....................................................................................................... - 15 -
6 Parameter Setup .......................................................................................................................... - 15 -
6.1 Operation .............................................................................................................................. - 15 -
6.2 Parameter Setup Reference Table ..................................................................................... - 16 -
6.3 Parameters Description ...................................................................................................... - 16 -
7 Data Output via USB Port ........................................................................................................... - 17 -
8 Lamp Replacement ..................................................................................................................... - 19 -
9 Warranty ....................................................................................................................................... - 20 -
*Fully charge the battery before first-time use.
- 3 -
1 INTRODUCTION
Thank you for purchasing Oakton T100WL Portable Turbidity Meter (hereafter referred to as the instrument). The instrument uses tungsten filament lamp as the light source and 90° scattering method, which is compliant with U.S EPA 180.1 method for the determination of turbidity in drinking, ground, surface, and saline waters, domestic and industrial wastes. It is suitable for on-site and laboratory use.
The instrument provides accurate results with simple measurement, calibration, and data management, giving you unparalleled confidence in test results and ease of use. Intelligent functions such as automatic calibration, operation navigation, parameter setup, real-time
clock display, calibration date check, automatic power off, low voltage indication, and a replaceable tungsten filament lamp.
Large TFT color screen with blue background for measurement mode, green background for
calibration mode.
Operation guidance and reminders in the process of calibration, measurement and parameter
setting.
TruRead measurement mode automatically takes multiple consecutive readings, and calculate their
average, minimum and maximum value, as well as displaying each set of data in a brief test report. TruRead is a much better way to determine the true turbidity of your samples, rather than relying on just one reading. It's especially suitable for sample solutions with rapid settling and continuous measurement changes.
Zero turbidity error reminder: the instrument has zero-point calibration and zero-point error reminder
to ensure measurement accuracy for low turbidity solutions.
200 sets of data storage with USB data output to PC (Windows-based). ⚫ 3.7V rechargeable lithium battery can supply power for over 20 hours of continuous measurement.
The battery life is 5 times longer compared with tungsten lamp turbidimeter in the market using AA alkaline batteries.
Multi-language operating system, including English, Spanish & Simplified Chinese. ⚫ The instrument’s structure meets IP67 water-resistant grade and is suitable for use in harsh
environments.
Everything comes in a rugged carrying case, including accessories such as calibration solutions,
test vials, USB charger, and more.
- 4 -
2 TECHNICAL SPECIFICATIONS
Specification
Description
Measurement
Method
90° scattering measurement, U.S EPA180.1 compliant
Light Source
Tungsten filament lamp
Measuring Range
0 to 1000 NTU (FNU), automatic range switch
0.01 to 19.99 NTU(FNU)
20.0 to 99.9 NTU(FNU) 100 to 1000 NTU(FNU)
Accuracy
≤ ±2% of reading+ stray light
Repeatability
≤ ± 1% of reading or 0.02 NTU(FNU) (the greater of the two)
Resolution
0.01/0.1/1 NTU (FNU)
Stray Light
≤0.02 NTU (FNU)
Calibration
Standard
AMCO Polymer or Formazin Solution0, 20, 100, 400 and 800NTU (FNU)
Detector
Silicon photovoltaic
Measuring Mode
Normal measurement and TruRead measurement
Data Storage
200 groups
Data Output
USB to PC
Calibration Record
Calibration date and time
Display
TFT color screen
Sample Vial
Φ25×60 mm,high borosilicate glass with lid
Sample Vial
Volume
18 ml
Power supply
3.7V Rechargeable lithium battery
Working Condition
Temperature: 0 to 50°C(32°F to 122°F); Relative humidity
0 to 90% at 30°C, 0 to 80%at 40°C, 0 to 70% at 50°C, no condensation
Instrument sealing
grade
IP67
Certificates
ISO9001:2015 and CE
Limited Warranty
2 years
Dimension and
Weight
Meter: (90×203×80) mm/385g
Test Kit: (310×295×110) mm/1.5kg
- 5 -
3 INSTRUMENTATION ILLUSTRATION
3.1 Summary
Flip cover of the sample vial holder (Close the cover when measuring)
Dust proof plug (Take off the plug when measuring)
Housing
Sample vial holder
Display
Calibration vials or sample vials
Keypad
Positioning mark (Align the mark with the mark on the calibration vial or sample vial)
USB Port
Built-in lithium battery (Rechargeable)
Lamp cover
Diagram-1
- 6 -
3.2 Configuration
3.3 Display Mode
Calibration Solutions
0.0, 20, 100, 400, 800 NTU
Carrying case
T100WL Turbidimeter
Microfiber cloth
Power adaptor (5V 1A)
Software flash drive
Silicone oil (10 ml)
Sample vials×6
USB Cable (under the meter)
User Manual
Diagram-2
Measurement Mode Measurements and unit Operation Guide Battery icon Date and time
Measuring
Mode
Calibration solution Measurements and unit Progress bar Operation guide
Calibration
Mode