Oakton T-100IR Instruction Manual

- 1 -
T-100IR Portable Turbidity Meter
- 2 -
Table of Contents
1 Introduction ................................................................................................................................... - 3 -
2 Technical Specifications .............................................................................................................. - 4 -
3 Instrumentation Illustration ......................................................................................................... - 5 -
3.1 Summary ................................................................................................................................. - 5 -
3.2 Configuration ......................................................................................................................... - 5 -
3.3 Display Mode .......................................................................................................................... - 6 -
3.4 Keypad .................................................................................................................................... - 7 -
3.5 Battery Installment ................................................................................................................ - 7 -
3.6 Data Storage, Recall, and Deletion ...................................................................................... - 7 -
4 Calibration...................................................................................................................................... - 8 -
4.1 Preparation for Calibration ................................................................................................... - 8 -
4.2 Calibration Procedure (Take 0 NTU and 20 NTU as an example) ..................................... - 9 -
4.3 Notes for Calibration ........................................................................................................... - 10 -
5 Turbidity Measurement ............................................................................................................... - 11 -
5.1 Sample Vial Handling .......................................................................................................... - 11 -
5.2 Preparation for Measurement ............................................................................................. - 11 -
5.3 Normal Measurement .......................................................................................................... - 11 -
5.4 Continuous Measurement .................................................................................................. - 11 -
5.5 Notes for Measurement ....................................................................................................... - 12 -
6 Parameter Setup .......................................................................................................................... - 13 -
6.1 Operation .............................................................................................................................. - 13 -
6.2 Parameter Setup Reference Table ..................................................................................... - 13 -
6.3 Parameters Description ...................................................................................................... - 13 -
7 Data Output via USB Port ........................................................................................................... - 14 -
8 Lamp Replacement ..................................................................................................................... - 16 -
9 Warranty ....................................................................................................................................... - 17 -
- 3 -
1 INTRODUCTION
Thank you for purchasing the Oakton T-100IR Portable Turbidimeter (hereafter referred to as the instrument). The instrument is designed using the principle of scattering and conforms to the ISO 7027, (DIN EN 27027) methods for measuring the turbidity of aqueous solutions and is suitable for use in laboratories or on site.
The instrument provides accurate results with the simple calibration and testing, giving you unparalleled ease of use and accuracy. The instrument has intelligent functions such as automatic calibration, operation
navigation, parameter Setup, real-time clock display, calibration date check, automatic power off, low voltage indication, and a replaceable tungsten light source.
Large TFT color screen with blue background for measurement mode, green
background for calibration mode, and operation guides for easy calibration and measurement, optional English & Spanish menu for convenient operation.
500 sets of data storage with USB data output to PC. Normal and continuous two kinds of measurement mode. Continuous measurement
mode with the option to select the number of continuous measurements, automatically generating average, maximum and minimum measured values in a data storage interface, suitable for sample solutions with rapid settling and continuous measurement changes.
It meets IP67 water-resistant grade requirements and is suitable for field use.
Product comes in a rugged carrying case, including calibration solutions, sample test
cuvettes, and more.
- 4 -
2 TECHNICAL SPECIFICATIONS
Specification
Description
Measurement
Method
ISO 7027, (DIN EN 27027) compliant
Light Source
Infrared LED
Measuring Range
0 ~ 1000 NTU (FNU), automatic range switch
0.01~19.99 NTU(FNU)
20.0~99.9 NTU(FNU) 100~1000 NTU(FNU)
Accuracy
±2% of reading+ stray light
Repeatability
± 1% of reading or 0.02 NTU(FNU) (the greater of the two)
Resolution
0.01/0.1/1 NTU (FNU)
Stray Light
0.02 NTU (FNU)
Calibration
Standard
Distilled water or deionized water: 0 NTU (FNU)
Polymer or Formazin Solution20, 100, 400 and 800NTU (FNU)
Detector
Silicon photovoltaic
Measuring Mode
Normal measurement and continuous measurement
Data Storage
500 groups
Data output
USB to PC
Calibration Record
Calibration date and time
Display
TFT color screen
Sample Vial
Φ25×60 mm,high borosilicate glass with cap
Vial Volume
18 ml
Working
Temperature: 0 to 50℃(32~122℉);Relative humidity:
0~90% at 30, 0~80%at 40, 0~70% at 50, no condensation
Instrument sealing
grade
IP67
Certificates
ISO9001:2015 and CE
Warranty
2 years
Dimension and
Weight
Meter: (90×203×80) mm/385g
Carrying case: (310×295×110) mm/1.5kg
- 5 -
3 INSTRUMENTATION ILLUSTRATION
3.1 Summary
3.2 Configuration
Flip cover of the sampling well (Close the cover when measuring)
Dust proof plug (Remove for measurements)
Housing
Sample well
Display
Calibration vials or sample vials
Keypad
Positioning mark (Align the mark with the mark on the calibration vial or the sample vial)
USB Port
Battery compartment
Lamp cover
0 NTU calibration vial distilled water self-prepared by users Calibration Solution:20, 100, 400, 800NTU
Carrying case
T-100 Turbidimeter
Microfiber cloth
AA Alkaline batteries×4
Silicone oil (10 ml)
Sample vials×6
Screwdriver (for opening battery cap)
User Manual
USB Cable & CD (under the meter)
Figure-1
Graph-2
- 6 -
3.3 Display Mode
Measurement Mode Measured parameter, units Operation Guide Low battery indicator Date and time
Measuring
Mode
Calibration Mode - Green Measured parameter Progress bar Operation guide
Calibration Mode Standard values Operation guide Finished calibration indicator
Parameter Setup Menu Operation guide
Data Recall Mode Data set numberings Saved measurements
Operation guide
Calibration
Mode
Calibration
Menu
Parameter
Setup
Data
Recall